Professional Documents
Culture Documents
Daily
Monthly
DOSIMETRY
x-ray/electron flatness
x-ray/electron symmetry
SRS arc rotation*
(range: 0.5 10 MU/deg)
Annually
1%
+ or 1%
MU set vs delivered:
1MU or 2% (whichever is
greater)
Gantry arc set vs delivered:
1 or 2% depending on which is
greater
Within 1% deviation (absolute)
2% for field size < 4x4 cm2
1% 4x4 cm2
Within 2% deviation from
baseline
Within 1% deviation from
baseline
Within 1mm deviation
+ or 2%
2%
+ or 2% 5 MU
+ or 2% from baseline
+ or 1% from baseline
+ or 1% from baseline
2%
2% for both IMRT and
SRS/SBRT*
1%
2% or 2 mm
MECHANICAL
Within 1mm deviation
Functional
Within 2mm deviation
1mm deviation*
Within 2mm deviation
1
Within 2mm deviation
Functional
2 mm or 1% on a side
1mm or 1% on a side for
asymmetric fields
1 mm
1
2mm
2mm (symmetric)
1mm (asymmetric)
1mm
2mm/1
1mm/0.5*
2mm
1mm
Functional
+ or 1mm
2mm (non-IMRT)
1.5mm (IMRT)
1mm*
2mm (all techniques)
2mm
1mm*
SAFETY
Functional
Functional
Functional
Functional*
Functional
Functional
Functional
Functional
RESPIRATORY GATING
2%
Functional
Functional
Functional
2%
Functional
100 ms of expected
100 ms of expected
respiratory phase/amplitude
*Applies to SRS/SBRT procedures
Daily
Monthly
PLANAR KV AND MV (EPID) IMAGING
Collision interlocks
Functional
Functional
Functional*
Functional*
Positioning/repositioning
2mm
1mm*
Imaging and treatment
2mm
coordinate coincidence (single 1mm*
gantry angle)
Imaging and treatment
2mm
coordinate coincidence (four
1mm*
cardinal angles)
Scaling
2mm
2mm*
Spatial resolution
Baseline
Baseline*
Contrast
Baseline
Baseline*
Uniformity and noise
Baseline
Baseline*
Full range of travel SSF
Imaging dose
Beam quality/energy
CONE-BEAM CT (KV AND MV)
Annual
Collision interlocks
Imaging and treatment
coordinate coincidence
Positioning/repositioning
Functional
Functional*
2mm
1mm*
1mm
1mm*
2mm
1mm*
Baseline
Baseline*
Baseline
Baseline*
Baseline
Baseline*
Baseline
Baseline*
Baseline
Baseline*
Geometric distortion
Spatial resolution
Contrast
HU constancy
Uniformity and noise
Imaging dose
*Applies to SRS/SBRT procedures
Daily
Functional
Functional
Monthly
SAFETY
SYSTEM OPERATION
Laser/image/treatment
isocentric coincidence OR
Phantom localization and
repositioning with couch shift
Within 2mm
Within 2mm
Annually
IMAGE QUALITY
Baseline
Baseline
2mm (or 5 lp/cm)
Baseline
Baseline
DOSE
Imaging dose
Baseline
IMAGING SYSTEM PERFORMANCE
Baseline
SYSTEM OPERATION
Long and short term
planning of resources (disk
procedures
Tolerance Limits
NCRP recommendations or
applicable regulatory limits
Within 20% deviation of
manufacturer specifications
Within 2 mm deviation
Within 2mm deviation over the
length of laser projection
Within 2 mm deviation
longitudinal motion
Table indexing and position
Gantry tilt accuracy
Gantry tilt position accuracy
Scan localization
Radiation profile width
Sensitivity profile width
Generator tests
range of table
Annually
Annually
Annually
Annually
Annually (This test is optional
if the CTDI accuracy has been
verified)
Semiannually
After replacement of major
generator component
Tolerance limits
For water, 05 HU
Manufacturer specifications
Within 1mm deviations
Field uniformity
Electron density to CT
number conversion
Spatial resolution
Contrast resolution
Annually
Annually
Within 5 HU
Acceptance Test
Create an anatomical description based on a standard set of CT scans
Create a patient model based on the standard CT data discussed above
Verify that all beam technique functions work, using a standard beam description provided by the
vendor
Perform dose calculations for a standard photon beam dataset. Tests should include various open
fields, different SSDs, blocked fields, MLC-shaped fields, inhomogeneity test cases, multi-beam
plans, asymmetric jaw fields, wedged fields, and others
Perform a set of dose calculations for a standard electron beam data set
Perform a set of dose calculations for single sources of each type, as well as several multi-source
implant calculations, including standard implant techniques such as GYN insertion tandem &
ovoids
Display dose calculation results. Use a standard dose distribution provided by the vendor to verify
the DVH code works as described
Print out all hardcopy documentation for a given series of plans
Imaging Input Test
Document and verify parameters used to determine geometric description of each image (e.g.,
number of pixels, pixel size, slice thickness)
Left-right and head-foot orientations
Text information
Imaging data
Image unwarping (removing
distortions)
Structure attributes
Relative ED definition
Display characteristics
Capping (how end of structure
is based on contours)
Structure definition
MU calc
Leaf width
Number of leaves
Distance over midline that can
be traveled by a leaf
Leaf transmission
Minimum gap between
opposing leaves
Leaf labels
Leaf editing capabilities
dMLC
Machine library
Machine/beam accessories
Parameter limitations
Wedges
Compensators
Block type
Block Transmission
MLC leaf fits
Electron applicators
Orientation and angle
specifications
2-D Display
Confirm the proper method to calculate MUs wen the bolus is used
MLC Parameters
Leaf travel (min, max) field size min and max
Overlap between leaves (the tongue and groove design of most MLC system affect the parameter)
Max extension between leaves
Leaf readout resolution
Jaw algorithm
Leaf end design
Design of the side of leaves
Leaf synchronization for dMLC
Beam Configuration Tests
Verify that the library of available machines and beams is correct
Verify that the availability of machine and beam-specific accessories (electron cone or wedge)
Verify that limitations are correct for jaws, MLC, FS with wedges, compensators and electron
applicators. Verify MU limits, MU/deg limits, angle limits.
Verify that wedge characterizations such as coding, directions, and FS limitations, and availability
are correct
Verify correct use and display
Field Shape Design Test
Verify that the system distinguishes between island blocks, in which the aperture delineates the
block shape, and aperture or conformal blocks.
Verify correct specification of transmission or block thickness for full blocks and partial
transmission blocks
Document and test all methods used to fit the MLC leaves to desired field shape
Verify availability and size of electron applicators
Wedge Tests
Confirm that wedge orientation and angle specifications are consistent throughout the planning
system
Check display of wedge in different 2-D planes (parallel, orthogonal, oblique) for different
directions, collimator rotations, and wedge orientations
3-D Display
Orientation and FS limitations
Autowedges (wedges inside the
head of the machine)
Dynamic wedge
Dose points
Interactive point doses
Consistency
Dose grids
2D dose displays
Isodose Surfaces
Beam display
cases
Source display
1. Klein EE, Hanley J, Bayouth J, et al. TG-142: Quality assurance of medical accelerators. Med Phys. 2009;36(9):4197-4212.
http://dx.doi.org/10.1118/1.3190392
2.
Bissonnette J-P, Balter PA, Dong L, et al. TG-179: Quality assurance for image-guided radiation
therapy utilizing CT-based technologies. Med Phys. 2012;39(4):1946-1963.
http://dx.doi.org/10.1118/1.3690466
3.
Mutic S, Palta JR, Butker EK, et al. TG-66: Quality assurance for computed-tomography simulators
and the computed-tomography-simulation process. Med Phys. 2003;30(10):2762-2792.
http://dx.doi.org/10.1118/1.1609271
4.
Fraass B, Doppke K, Hunt M, et al. American Association of Physicists in Medicine Radiation Therapy
Committee Task Group 53: Quality assurance for clinical radiotherapy treatment planning. Med Phys.
1998;25(10):1773-1829. http://dx.doi.org/10.1118/1.598373