ACUSON X300

US

Software
System
Loading Software

Siemens AG

Siemens Medical Solutions


ejt
Ultrasound

Print No.: US11-101.816.01.01.02


Replaces: n.a.
Part No.:
sd

2006

The reproduction, transmission or use


of this document or its contents is not
permitted without express written
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.

English
Doc. Gen. Date: 09.06
n.a.

Revision / Disclaimer

1Revision / Disclaimer

Copyright
Copyright 2006 by Siemens Corporation. All rights reserved. No part of this publication
may be reproduced, transmitted, transcribed, stored in retrieval systems, or translated into
any language or computer language, in any form or by any means, electronic, mechanical,
magnetic, optical, chemical, manual, or otherwise, without the prior written permission of
Siemens Corporation. Siemens Corporation reserves the right to change its products and
services at any time. In addition, this manual is subject to change without notice. Siemens
Corporation welcomes customer input on corrections and suggestions for improvements
to this manual.Although Siemens Corporation has attempted to ensure accuracy throughout this manual, Siemens Corporation assumes no liability for any errors or omissions, nor
for any damages resulting from the application or use of this information.
Trademarks
ACUSON, Sequoia, Cypress, AEGIS, Aspen, 128XP, XP, AcuNav, CV70, SONOLINE,
SONOVISTA, Adara, Antares, G20, G40, G50, G60S, X300, X500, Omnia, Premier and
KinetDx are trademarks of Siemens Corporation registered in the U.S. Patent and Trademark Office. CWS3000, DS3000, WS3000, and WebPro are trademarks of Siemens Corporation.
Windows XP, 2000 and NT are registered trademarks of Microsoft Corporation. Internet
Explorer is a trademark of Microsoft Corporation. Netscape and Netscape Navigator are
registered trademarks of Netscape Communications Corporation. Java and the JavaScript
name are registered trademarks of Sun Microsystems, Inc. Adobe Acrobat is a registered
trademark of Adobe Systems, Inc.
Other products and brand names are trademarks of their respective owners.
Disclaimer
The service of equipment described herein is to be performed by qualified personnel who
are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are directed to contact one of the
local offices of Siemens or one of its affiliates before attempting installation or service procedures.
License Agreement
All computer programs copyright 1990-2006 by Siemens Corporation or its suppliers. Programs are licensed under the following agreement:
Siemens or its suppliers retain(s) ownership of and title to any computer program supplied
with the equipment and to the trade secrets embodied in such computer programs. Subject to the Buyers acceptance and fulfillment of the obligations in this paragraph, Siemens
grants the Buyer a personal, non-transferable, perpetual, non-exclusive license to use any
computer program supplied with the Equipment that is necessary to operate the Equipment solely on the medium in which such program is delivered for the purpose of operating the equipment in accordance with the instructions set forth in the operators manuals
supplied with the Equipment and for no other purpose whatsoever. Buyer may not
reverse-assemble, reverse-compile or otherwise reverse-engineer such computer programs nor may Buyer make a copy of such program or apply any techniques to derive the

ACUSON X300

US11-101.816.01.01.02
09.06

Page 2 of 32
sd

Siemens
Medical Solutions USA, Inc.

Revision / Disclaimer

trade secrets embodied therein. In the event of a failure by Buyer to comply with the terms
of this license, the license granted by this paragraph shall terminate. Further, because
unauthorized use of such computer programs will leave Siemens without an adequate
remedy at law, Buyer agrees that injunctive or other equitable relief will be appropriate to
restrain such use, threatened or actual. Buyer further agrees that (i) any of Siemens"s
suppliers of software is a direct and intended beneficiary of this end-user sublicense and
may enforce it directly against Buyer with respect to software supplied by such supplier,
and (ii) No supplier of Siemens shall be liable to buyer for any general, special, direct, indirect, consequential, incidental or other damages arising out of the sublicense of the computer programs supplied with the equipment.

Siemens
Medical Solutions USA, Inc.

US11-101.816.01.01.02
09.06

Page 3 of 32
sd

ACUSON X300

Table of Contents

1-

0Table of Contents

1 _______ Loading System Software _________________________________________ 5


Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Disk Partition Map . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Setup Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Tools and Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Reference Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Time Required. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6
6
6
6

2 _______ Verify System Software ___________________________________________ 7


Verify the System Software Version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

3 _______ Loading System Software _________________________________________ 9


Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Setup Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Tools and Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Reference Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Time Required. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

10
10
10
10

Verify the System Software Version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11


Backing Up Data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Installing the System Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Configuring Peripheral Video and Language Format . . . . . . . . . . . . . . . . . . . . . . . . 21
Verifying Software Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Check/Update the Firmware and BIOS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Verify the Software Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Confirming the Product Serial Number . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Restore the Presets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

23
23
23
24
25

Sleep Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
System Configuration: Hardware Version 2.0 - Software Version 2.0. . . . . . . . . . . . 27
System Configuration: Hardware Version 1.0 - Software Version 2.0. . . . . . . . . . . . 28

4 _______ Changes to Previous Version _____________________________________ 31


Changes to Previous Version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31

ACUSON X300

US11-101.816.01.01.02
09.06

Page 4 of 32
sd

Siemens
Medical Solutions USA, Inc.

Loading System Software

2-

Overview
1Loading System Software

The instructions below describe how to load system software on the system.
Observe the following Caution alerts before proceeding.

CAUTION

If the system software will be upgraded (or downgraded to an earlier version) for testing or to resolve an operational problem, the
feature keys for options that are not supported by the installed version will be lost.
To avoid this problem, uninstall all feature keys before upgrading
or downgrading the system software. See (US07-102.892.01 / Uninstalling the Product Feature Keys) for details.
Be sure that all feature keys are properly re-installed
after upgrading (or downgrading) software.

Disk Partition Map

The table shows the disk partition map.


Tab. 1

System Disk Partition Map

Partition Name Partition Siz

File System

Contents

C:

NTFS

Windows XP
Software (in-

Remains

cluding DIMAQ,
syngo, SCR
3D)
D:

Siemens
Medical Solutions USA, Inc.

60 GB

US11-101.816.01.01.02
09.06

NTFS

Page 5 of 32
sd

Patient data
and image files
(including Quick
Set and Imaging Preset).

ACUSON X300

Loading System Software

Setup Requirements

Tools and Materials

Standard service tools


Installation CD / DVD

Reference Documents

None

Time Required

Approximately 30 minutes is required to load software. More time will be required if the
customers data is backed up.

ACUSON X300

US11-101.816.01.01.02
09.06

Page 6 of 32
sd

Siemens
Medical Solutions USA, Inc.

Verify System Software

3-

Verify the System Software Version


2Verify System Software

Use the following instructions to verify the system software version of ACUSON X300
ultrasound system.
1. Power on the system and wait for the system to boot up.
2. Press Presets. The Preset Main Menu displays.

Fig. 1:

Preset Main Menu

Siemens
Medical Solutions USA, Inc.

US11-101.816.01.01.02
09.06

Page 7 of 32
sd

ACUSON X300

Verify System Software


3. Select Service. The Main Service Menu displays.

Fig. 2:

Service Main Screen

4. Verify the system SW version information.


5. Record the system SW Version information.
6. Select Cancel to exit.

NOTE

ACUSON X300

Before starting a software installation, make a backup copy of the


user-defined settings and the patient database. After loading the
system software, customize the setting of video and language format by following the procedure in US11-101.816.01 Installing the
System Software.

US11-101.816.01.01.02
09.06

Page 8 of 32
sd

Siemens
Medical Solutions USA, Inc.

Loading System Software

4-

Overview
3Loading System Software

The instructions below describe how to load system software on the ACUSON X300.
Observe the following Caution alerts before proceeding.

CAUTION

If the system software will be upgraded (or downgraded to an earlier version) for testing or to resolve an operational problem, the
feature keys for options that are not supported by the installed version will be lost.
To avoid this problem, uninstall all feature keys before upgrading
or downgrading the system software. See (US11-102.816.01 / License Key Installation) for details.
Be sure that all feature keys are properly re-installed
after upgrading (or downgrading) software.

Feature keys do not have to be uninstalled. Experienced customer service engineers


(CSEs) may choose to leave the keys installed.
The software generally consists of system Presets and system software.

System Presets (on CD-1)


System software (on CD-1 and CD-2)

Siemens
Medical Solutions USA, Inc.

US11-101.816.01.01.02
09.06

Page 9 of 32
sd

ACUSON X300

10

Loading System Software

Setup Requirements

Tools and Materials

Standard service tools

Reference Documents

None

Time Required

Approximately 30 minutes is required to load software. More time will be required if the
customers data is backed up.

ACUSON X300

US11-101.816.01.01.02
09.06

Page 10 of 32
sd

Siemens
Medical Solutions USA, Inc.

Loading System Software

11

Verify the System Software Version

Use (Verify System Software / p. 7)to verify the software version before proceeding.

Siemens
Medical Solutions USA, Inc.

US11-101.816.01.01.02
09.06

Page 11 of 32
sd

ACUSON X300

12

Loading System Software

Backing Up Data

Back up the data, see (US11-104.816.01 / Backing Up Data)


Back up the presets, see (US11-104.816.01 / Backing Up Data)
Back up the customers data, see ((US11-104.816.01 / Backing Up Data)

ACUSON X300

US11-101.816.01.01.02
09.06

Page 12 of 32
sd

Siemens
Medical Solutions USA, Inc.

Loading System Software

13

Installing the System Software

1. Power on the system and wait for it to boot up completely.


2. Press Preset. The Preset Main Menu displays.

Fig. 3:

Preset Main Menu

Siemens
Medical Solutions USA, Inc.

US11-101.816.01.01.02
09.06

Page 13 of 32
sd

ACUSON X300

14

Loading System Software


3. Select Service from the menu on the left side of the screen.

Fig. 4:

ACUSON X300

Service Main Menu

US11-101.816.01.01.02
09.06

Page 14 of 32
sd

Siemens
Medical Solutions USA, Inc.

Loading System Software

15

4. Select Service from the Main Menu.

Fig. 5:

Select Service from the Service Menu

5. The Authentication Screen displays.

Siemens
Medical Solutions USA, Inc.

US11-101.816.01.01.02
09.06

Page 15 of 32
sd

ACUSON X300

16

Loading System Software


6. Enter the password, then select OK. The Service Main Menu displays.

Fig. 6:

ACUSON X300

Authentication Screen

US11-101.816.01.01.02
09.06

Page 16 of 32
sd

Siemens
Medical Solutions USA, Inc.

Loading System Software

17

7. Select Utilities to display the Utilities menu.

Fig. 7:

Select Utilities from the Service Menu

Siemens
Medical Solutions USA, Inc.

US11-101.816.01.01.02
09.06

Page 17 of 32
sd

ACUSON X300

18

Loading System Software


8. Select Install Software to install software.

Fig. 8:

Select Install Software from the Utilities Menu

The following message displays while the software is installing.

Fig. 9:

Installation Message

9. Insert the installation CD / DVD into the CD / DVD-R/RW drive.

ACUSON X300

US11-101.816.01.01.02
09.06

Page 18 of 32
sd

Siemens
Medical Solutions USA, Inc.

Loading System Software

19

10. Click OK. The system automatically reboots.

Fig. 10: Reboot Screen

11. After the automatic reboot, the system prompts for the type of upgrade to perform.

Fig. 11: Software Install DOS Screen

- Upgrade/Update Installation retains the current presets.


- Clean/Factory Installation initializes the presets and all patient studies are erased.

Siemens
Medical Solutions USA, Inc.

US11-101.816.01.01.02
09.06

Page 19 of 32
sd

ACUSON X300

20

Loading System Software


NOTICE

Selecting Clean/Factory Installation erases all patient studies from


the hard disk drive.
Be sure to back up all patient studies before selecting
Clean/Factory Installation.

12. Choose the desired upgrade. If a choice is not made in 30 seconds, the system will default to Upgrade/Update Installation.
13. After approximately 7-8 minutes, the system will display a pop-up window asking for the
second CD. The CD tray will open. Remove the first CD and insert the second.

Fig. 12: Symantec Ghost CD Install

14. After approximately five more minutes, the upgrade is complete, the system will reboot.
After the system completes the reboot, verify the system software version as described
in (Verify the System Software Version / p. 7).
15. Press Preset, then select Service --> Configuration --> License & Components to
confirm newly installed software version.

ACUSON X300

US11-101.816.01.01.02
09.06

Page 20 of 32
sd

Siemens
Medical Solutions USA, Inc.

Loading System Software

21

Configuring Peripheral Video and Language Format

1. Press Preset and select Basic System 1 from the left side of the Preset Main screen.

Fig. 13: Language Format Selection

2. Select the Language Format from the Format pane according to customer need.
Be sure that all feature keys are properly re-installed after upgrading (or downgrading)
software.
3. Select Save.

Siemens
Medical Solutions USA, Inc.

US11-101.816.01.01.02
09.06

Page 21 of 32
sd

ACUSON X300

22

Loading System Software


4. Select Peripheral from the left side of the Preset Main screen.

Fig. 14: Peripherals Main Menu

5. Select the Video format from the Peripheral pane according to customer need.
6. Select Save.
7. Restart the system.

ACUSON X300

US11-101.816.01.01.02
09.06

Page 22 of 32
sd

Siemens
Medical Solutions USA, Inc.

Loading System Software

23

Verifying Software Operation

Use the following instructions to verify that the software is operational and to restore data.

Check/Update the Firmware and BIOS

During this last phase of the software installation, the system checks the revision level of
the firmware and BIOS, updating them if necessary.

Verify the Software Installation

Before continuing, verify the system is now running the new software.
1. Power on the system and wait for the system to boot up.
2. Press the Preset key to display the Preset Main Menu.

Fig. 15: Preset Main Menu

3. Select Service to display the Service Main Menu.

Siemens
Medical Solutions USA, Inc.

US11-101.816.01.01.02
09.06

Page 23 of 32
sd

ACUSON X300

24

Loading System Software


4. The system information section of the service preset menu displays the current software version. Verify the system software version.

Fig. 16: Verify System Software Version

Confirming the Product Serial Number

1. Confirm that no warning message is displayed on the monitor.


2. Press the Preset key on the keyboard.
3. Select Service in the left pane of the Preset menu.
4. Verify that the correct software version is displayed.

ACUSON X300

US11-101.816.01.01.02
09.06

Page 24 of 32
sd

Siemens
Medical Solutions USA, Inc.

Loading System Software

25

5. At the product information screen, verify that the correct product Serial No. is displayed: the number displayed on the product Information screen should be identical to
the product Serial Number recorded on the label attached to the rear of the product.

Fig. 17: Location of System Serial Number

6. If necessary, restore the Presets according to (US11-104.816.01 / Backing Up Data)


and perform the instructions in reverse fashion.
7. If necessary, restore the service data. Refer to (US11-104.816.01 / Backing Up Service
Data) and perform the instructions in reverse fashion.
8. If necessary, restore the Feature keys according to the instructions found in
(US11-102.816.01 / License Key Installation).

Restore the Presets


NOTE

Siemens
Medical Solutions USA, Inc.

Note: This procedure is only required when if a backup CD was


created prior to software installation.

US11-101.816.01.01.02
09.06

Page 25 of 32
sd

ACUSON X300

26

Loading System Software


1. Display the Service menu in the Preset.
-

Press the Preset key.


Select Service in the left pane of the Preset menu.
Select Go.
Select Backup & Restore icon in the Service menu window.
Select Restore. The system will request the back-up disc you have made.

2. Insert the Backup CD in the CD-R/W drive and wait approximately ten seconds for the
CD to spin up.
3. Select Restore on the pop-up window to start. The system will display Restore is
completed message.
4. Select Ok and remove the backup CD.
- The system will display Shutdown before the system starts restoring
message.
5. Select Ok to confirm.
6. Close Backup &Restore and select Exit twice to return to the preset menu.
7. Shutdown the system manually and wait for it to shut down
8. Restart the system.

ACUSON X300

US11-101.816.01.01.02
09.06

Page 26 of 32
sd

Siemens
Medical Solutions USA, Inc.

Loading System Software

27

Sleep Mode

The Sleep Mode option is a part of software version 2.0. It can be installed on systems
configured with hardware versions 1.0 or 2.0. The Sleep Mode setup is different for each
system hardware version. Use the set of instruction that apply to the system software and
hardware configuration.

System Configuration: Hardware Version 2.0 - Software Version 2.0


NOTE

After installing software version 2.0, the default status of Sleep


Mode is inactive. To activate Sleep Mode, select the Show field of
the Preset Sleep Mode Option.

1. Press the Preset key. The system displays the Preset Main Menu.
2. Select Service from the Presets menu. The system displays the Service Authentication screen.
3. Enter the password and select Go. The system displays the Service Main Menu.
4. Select Configuration. The system displays the Configuration screen.
5. Select System from the Configuration menu.
6. Select Preset Sleep Mode Option. See (Fig. 18 / p. 27) for location of the option.

Fig. 18: System Information Screen > Preset Sleep Mode Option

Siemens
Medical Solutions USA, Inc.

US11-101.816.01.01.02
09.06

Page 27 of 32
sd

ACUSON X300

28

Loading System Software


The system will display the Enable Sleep Mode option on Preset Main Menu, Basic
System 1 screen.
7. Select the Save button. The system saves the changes.
8. Select the Close button. The system displays to the Preset Main Menu.
9. Verify the Enable Sleep Mode option is selected. See (Fig. 19 / p. 28) for location of
the option.

Fig. 19: Preset Menu > Basic System 1 > Enable Sleep Mode

System Configuration: Hardware Version 1.0 - Software Version 2.0


NOTE

Systems with hardware configuration version 1.0 do not have the


sleep mode function. Therefore, there is no need to display the
Enable Sleep Mode function on Preset Main Menu screen.
If the system hardware includes the DC/DC power supply
(MN: 10010385) and control panel firmware, the Sleep Mode function can be deactivated by deselecting Enable Sleep Mode.

1. Press the Preset key. The system displays the Preset Main Menu.

ACUSON X300

US11-101.816.01.01.02
09.06

Page 28 of 32
sd

Siemens
Medical Solutions USA, Inc.

Loading System Software

29

2. Select Service from the Preset Menu. The system displays the Service Authentication
screen.
3. Enter the password and select Go. The system displays the Service Main Menu.
4. Select Configuration. The system displays the Configuration screen.
5. Select System from the Configuration menu.
6. Deselect the Preset Sleep Mode Option.

Fig. 20: System Information Screen > Preset Sleep Mode Option

The system hide the Enable Sleep Mode option unavailable on Preset Main Menu,
Basic System 1 screen.
7. Select the Save button. The system saves the changes.
8. Select the Close button. The system displays to the Preset Main Menu.

Siemens
Medical Solutions USA, Inc.

US11-101.816.01.01.02
09.06

Page 29 of 32
sd

ACUSON X300

30

Loading System Software


9. Verify the Enable Sleep Mode option is not displayed. See (Fig. 21 / p. 30) for location
of the hidden option.

Fig. 21: Preset Menu > Basic System 1 > Disable Sleep Mode

ACUSON X300

US11-101.816.01.01.02
09.06

Page 30 of 32
sd

Siemens
Medical Solutions USA, Inc.

Changes to Previous Version

31

5-

Changes to Previous Version


4Changes to Previous Version

Change(s)

Description

New document

Initial document.

Siemens
Medical Solutions USA, Inc.

US11-101.816.01.01.02
09.06

Page 31 of 32
sd

ACUSON X300

32

ACUSON X300

Changes to Previous Version

US11-101.816.01.01.02
09.06

Page 32 of 32
sd

Siemens
Medical Solutions USA, Inc.