You are on page 1of 39

AMS2750 E

Changes from AMS2750D


Feb 2013

Overview on Changes
AMS2750 Specification Updated to Revision E
- New Revision of Spec 2750E revised July 2012;
- Replaces Spec - 2750D (revised Sept 2005)
- Updated to include wireless sensor technology
- Added requirements to document As found instrument
calibration and offset conditions (procedural impacts)
- Calibration changes like check min. of 6 simulated sensor @
min. 4 points, calibrated for each type input/output used
- Provides more details on Temperature Uniformity Survey
Testing (TUS) procedures, furnace repairs and modifications,
data collection requirements, survey frequency, number and
location of sensors, Pass/Fail requirements
- Added Refrigeration & Quench Furnace class to tables

General Update
2

HONEYWELL - CONFIDENTIAL

File Number

Standards
AMS2750E Pyrometry
- Covers Temperature Sensor, Instrumentation, Thermal
processing equipment, System Accuracy Tests, Temperature
Uniformity Survey requirements
- Ensures that parts or raw material are heat treated in
accordance to applicable specifications

AS7102 Nadcap Requirements for Heat Treating


- Documents the requirements for the accreditation program
- Industry managed program for continuous improvement
Standard industry audit program for special processes
Focuses on overall supplier performance
Auditing overseen by PRI (Performance Review Institute)

HONEYWELL - CONFIDENTIAL

File Number

AMS 2750E Section 2 Definitions Changes


- Par. 2.2.15 DIGITAL INSTRUMENT:
Clarification was added to definition of digital instruments with
digital temperature display and with temperature trace on
paper chart:
An instrument that presents process measurements in a digital
(numeric) display format. Typical examples are mechanical
recorders with digital displays, digital indicators, controllers with
digital displays, etc. In the case of a recorder with digital
temperature display and a recorder trace (line) the record
(recorder trace) is analog while the display is digital. A
recorder trace on a paper chart is an analog record.

- Par. 2.2.25 INTERVAL


Interval definition was added to AMS2750E to avoid confusion
by some suppliers.
4

HONEYWELL - CONFIDENTIAL

File Number

AMS 2750E Section 2 Definitions Changes


- Par. 2.2.27 LOAD SENSORS :
Clarification regarding correct usage of load sensors/
thermocouples. (in rev. D it was not included in the load
sensor definition but stated in par. 3.1.8.1):
Sensors that are attached to or in contact with production
material, a representation of production material, or are buried
in the load of production material and which supply temperature
data of the production material to process instrumentation.

HONEYWELL - CONFIDENTIAL

File Number

AMS 2750E Temp Sensor Changes


Temperature Sensor - Section 3.1 Changes
- This section deals with the requirements for the temperature
sensors used in these heat treatment processes
AMS2750E
Section
Reference
3.1.1.5.1
3.1.2.2.1
3.1.2.2.6

3.1.2.5.2

AMS2750E Change
Allows use of Wireless transmitters as alternative to
extension wire
Certs for sensors now require Date of Calibration
Correction factor for each calibration temperature
traceable to NIST or other recognized National Standard.
Alternatively, deviation (error) for each calibration
temperature may be listed. The certificate shall clearly
state if it shows deviation or correction factor.
Extrapolation of calibration correction factors above the
highest calibration temperature and below the lowest
calibration temperature is prohibited by any calibration
source except NIST or other National Standards
calibration agency.

HONEYWELL - CONFIDENTIAL

Comment
New
New

Added this

Added except NIST or other


National Standards calibration
agency.

File Number

AMS 2750E Temp Sensor Changes


Temperature Sensor - Section 3.1 Changes
AMS2750E
Section
Reference
3.1.2.6

3.1.3.5

3.1.5.2

3.1.5.3

AMS2750E Change
When calibrating thermocouples rolls, AMS2750 Rev. E
requires sampling from both ends regardless of roll
length. (1000 feet/305 meters limit from AMS2750D for
one end sampling only is no longer acceptable). Other
requirements for rolls calibration remains the same.
Nonexpendable base metal TUS TCs used below 1200F
(650C) do not need to be permanently installed on TUS
rack between its usage like it was stated in Rev. D
Added time of 90 days
Expendable base metal load sensors may be used for
up to 90 calendar days after first use and for no more
than 30 uses
Changed the requirement on the life of nonexpendable
base metal load thermocouples
2200 F (1205 C) to 2299 F (1260 C)
3 months or 10 uses
1801 F (980 C) to 2199 F (1205 C)
3 months or 90 uses
HONEYWELL - CONFIDENTIAL

Comment
Added Rolls shall be sampled at
both ends of the roll

TUS T/Cs can be reinserted for each


TUS through ports in the furnace;
need to be used at the same location
and depth of insertion.
Previously was just 30 uses

Previously for 2200F was just 10


uses, no time period specified
1801F was 30 days or 90 uses;
now 90 days or 90 uses
File Number

AMS 2750E Temp Sensor Changes


Temperature Sensor - Section 3.1 Changes
AMS2750E
Section
Reference
Figure 1

Comment

Added Table of Reuse and Recalibration of Test & Load


Sensors
Sensor - Secondary Standard Max. Permitted Error
changes

Summarizes Reuse & Recalibration


Primarily for clarification
Increased max. permitted error

Figure 1

Sensor System Accuracy Test


Max. Permitted Error changes
Recalibration permited if Types E & K are used below
500F (260C) only (Note 11)

Increased max. permitted error

3.4.5.2.1

Type B TCs are allowed as resident SAT TCs for


temperatures above 500F (260C)

Type B T/Cs were not allowed by


Rev D.

Figure 1

AMS2750E Change

HONEYWELL - CONFIDENTIAL

File Number

AMS 2750E Instruments


Controlling, Monitoring or Recording Instruments

Instrumentation - Section 3.2 Changes


- This section deals with the requirements of the
instrumentation used in these heat treatment processes
AMS2750E
Section
Reference
3.2.1

3.2.1.1.1

3.2.1.1.2

AMS2750E Change

Comment

AMS2750D allowed to use instruments meeting


requirements of AMS2750C (but not meeting rev D
requirements) if these instruments were purchased
before September 2006 (one year after rev. D issue). If
you have such instruments, now you have 3 years you
upgrade your equipment to meet Rev. E requirements
Temperature resolution requirements for chart
recorders shall be in accordance with Table 4. (See next
slide for table)

Exemption will run out for older


instrumentation, now have to
upgrade within 3 years of issuing of
AMS2750E (issued July 2012)

Process Recorder Print and Chart Speeds shall be in


accordance with Table 5. (See next slide for table)

Added metric speeds and updated


notes

HONEYWELL - CONFIDENTIAL

New requirements for refrigeration &


quench system recorders added

File Number

AMS 2750E Instruments


Controlling, Monitoring or Recording Instruments

New - requriements for refrigeration


and quench system recorders were
not defined in rev D.

10

HONEYWELL - CONFIDENTIAL

File Number

AMS 2750E Instruments


Controlling, Monitoring or Recording Instruments
Secondary Standard & Field Inst. Calibration Changes
AMS2750E
Section
Reference
3.2.1.5.4

3.2.1.5.4.1

3.2.1.5.4.2

11

AMS2750E Change

Comment

Calibration of secondary standard instruments and field


test instruments shall be performed in accordance with
the manufacturers instructions. These instruments shall
be tested at a minimum of 6 simulated sensor inputs.
These shall include the minimum and maximum of the
operating range used for test or calibration and a
minimum of 4 points in between either representing
areas of normal operation or spaced at approximately
equal intervals across the range in which the instrument
is used for test or calibration.
The instrument shall be calibrated for each type of input
and/or output used (e.g., each thermocouple type in use
and for mV, mA, etc. if the instrument is used for these
methods).
Calibration shall be performed on each channel in use
that can be altered or adjusted individually, or on each
group of channels that can be altered or adjusted as a
group. It is recommended that channels not in use be
blocked or tagged to stop unintentional use.

New requirements regarding


secondary standard instruments and
field instruments calibration

HONEYWELL - CONFIDENTIAL

New requirements regarding


secondary standard instruments and
field instruments calibration
New requirements regarding
secondary standard instruments and
field instruments calibration

File Number

AMS 2750E Instruments


Controlling, Monitoring or Recording Instruments
Calibration Controlling/Monitoring/Recording Inst. Changes
AMS2750E
Section
Reference
3.2.5.5

3.2.5.5.1

3.2.5.5.2

12

AMS2750E Change
Calibration shall be performed on instrument(s) defined
under the Instrument Type column in Table 3.
Regardless of the calibration procedure used, it shall
comply with the requirements of Table 3. Prior to any
instrument alteration or calibration, document the
instrument(s) as-found condition
Check as-found condition at a minimum of three
simulated sensor inputs which represent the minimum,
maximum and at least one point in the middle 1/3rd of
the furnace Qualified Operating Temperature Range. If
necessary, calibrate the instrument in accordance with
the manufacturers instructions.
Following any adjustments test the instrument at a
minimum of three simulated sensor inputs which
represent the minimum, maximum and at least one
point in the middle 1/3 rd of the furnace Qualified
Operating Temperature Range to document the as-left
condition.
HONEYWELL - CONFIDENTIAL

Comment
Not documenting as found
condition is often an audit failure
point during Nadcap audits

New requirements regarding


calibration of controlling, monitoring
and recording instruments

New requirements regarding


calibration of controlling, monitoring
and recording instruments

File Number

AMS2750E Instruments and Instrument Calibration


Table 3 Instruments and Instrument Calibration updates compared to rev. D:

New

13

Table 3 Notes
(1) Instruments of equivalent or greater accuracy are acceptable.
(2) Comparison checks among the cells shall be performed monthly.
(3) Cells having temperature control of 0.02 F (0.01 C).
(4) Furnace class 1 & 2 instruments shall have a minimum sensitivity of 1
F (1 C). Furnace class 3 through 6
instruments shall have a minimum sensitivity of 3 F (2 C).
(5) Controlling, monitoring and recording instrument calibration interval
shall be:
Digital
Electromechanical
Furnace Class 1 - Monthly
Furnace Class 1 - Monthly
Furnace Class 2 - Quarterly
Furnace Class 2 - Monthly
Furnace Class 3 - Quarterly
Furnace Class 3 - Quarterly
Furnace Class 4 - Quarterly
Furnace Class 4 - Quarterly
Furnace Class 5 - Semiannually Furnace Class 5 - Quarterly
Furnace Class 6 - Semiannually Furnace Class 6 Quarterly
Refrigeration and quenchant monitor (digital or analog) semiannually.
Exception: If the quenchant/refrigerator sensor is connected to a channel
in a furnace process recorder, the interval for calibration of that channel
in the instrument shall be the same interval as for the other channels of
the instrument.
New
(6) Furnace Control, Recording, Monitoring or Data Acquisition
Instruments.
(7) Field Test instruments meeting accuracy requirements of secondary
standards may be used to calibrate SAT and TUS test instruments in the
field.
(8) Over-temperature instruments used solely for furnace over
temperature protection do not need to be calibrated.
(9) NIST or other equivalent National Standard.
New
(10) Includes wireless transmission systems
(11) Instruments are operated in a controlled environment and are not
used on the shop floor.

HONEYWELL - CONFIDENTIAL

File Number

AMS 2750E Instruments


Controlling, Monitoring or Recording Instruments
Calibration Controlling/Monitoring/Recording Inst. Changes
AMS2750E
Section
Reference
3.2.5.5.3

3.2.5.5.5

3.2.5.5.6

14

AMS2750E Change
Chart recorder (circular and strip) speed(s) shall be
verified annually and shall be accurate within 3 minutes
per hour. Result of chart speed measurement shall be
documented.
For furnaces used only at a single temperature, the
calibration of controlling, monitoring or recording
instruments may either be performed per
manufacturers instructions at the single temperature
of use, or with a minimum of 3 points: the temperature
of use, and two (or more) points bracketing the
temperature of use.
Wireless equipment which performs the analog to
digital conversion at the furnace and transmits a digital
signal to the recorder are allowed, but require use of
the complete wireless system (wireless sender,
wireless receiver and associated controlling,
monitoring and recording instruments) during the
calibration process (See 3.1.1.5.1).

HONEYWELL - CONFIDENTIAL

Comment
Added Chart Speed documentation
requirement

New requirement

New requirement

File Number

AMS 2750E Instruments


Controlling, Monitoring or Recording Instruments
Electronic Records Changes
AMS2750E
Section
Reference
3.2.7.1

3.2.7.1.1.3.1

15

AMS2750E Change

Comment

An electronic record is any combination of text,


graphics, data, audio, pictorial, or other information
representation in digital form that is created, modified,
maintained, archived, retrieved, or distributed by a
computer system. When using a system (furnace
control, recording, monitoring or data acquisition) that
creates electronic records, systems purchased after
Sept 2006 or any electronic record created after 2015
(3 years after release of this revision) shall meet the
following requirements detailed in section 3.2.7.1.3
The system shall be capable of providing evidence the
record was reviewed such as by recording an
electronic review, or a method of printing the record for
a physical marking indicating review.

Added additional requirement,


3.2.7.1.3.1, to provide a means of
providing evidence that the electronic
record was reviewed.
This can be accomplished in the
Trendview recorder by the mark on
chart feature.

HONEYWELL - CONFIDENTIAL

New requirement
This can be accomplished in the
Trendview recorder by the mark on
chart feature.

File Number

AMS 2750E Instruments - Thermal Processing


Furnace Instrumentation Type Classification Changes
AMS2750E
Section
Reference
3.3.2 Table 3

3.3.2 Table 3

16

AMS2750E Change
The temperature indicated by the control sensor in each
control zone shall be recorded by a recording
instrument.
Alternatively, the recording instrument may be
connected to a second sensor contained in the same
sheath or holder as the control sensor, and separated
from the control sensor by no more than 0.38 inch (10
mm).
At least 1 recording load sensor in each zone
During production in multi-zone furnaces, empty zones
(i.e., no material is placed in or intrudes into the zone)
do not require a load sensor. However, a notation
must be made to the furnace load record that the zone
was entirely empty.

HONEYWELL - CONFIDENTIAL

Comment
For Instrument Types A, B, C, D
Added alternative means of
connecting a control sensor, this was
not allowed in AMS 2750 D

For Instrument Types A, B, C, D


Added clarification for multi-zone
furnaces

File Number

AMS 2750E Instruments - Thermal Processing


Furnace Instrumentation Type Changes
AMS2750E
Section
Reference
3.3.4

3.3.5

3.3.6

17

AMS2750E Change
Additional Sensors - There is no limit on additional
recording or load sensors in any control zone, but their
use must be defined in controlled operating instructions.
High and low temperature sensors, when required
(instrument types A and C See Figure 3), may be
inserted in heat sinks (See 2.2.24), positioned at the
high and low temperature locations of the control zone
based on the most recent TUS when heat sinks of
similar configuration are used during the TUS.
For multiple zone vacuum furnaces with 225 cubic feet
(6.4 m3) or less in working volume that are instrumented
with Types A, B, or C instrumentation, it is acceptable to
treat the furnace working volume as a single work zone
for the purpose of locating high and low temperature
recording thermocouples and determining the number of
load thermocouples required regardless of the number of
control thermocouples or instruments within the vacuum
furnace. The minimum sensors required are shown in
Figure 4. Each furnace shall have over-temperature
protection in the work zone.
HONEYWELL - CONFIDENTIAL

Comment
Added clarification for use of
additional sensors
Improvement to aid the heat treater
Also refer to Section 2.2.24
definition of a heat sink

Added clarification multi-zone


furnaces

File Number

AMS 2750E Instruments - Thermal Processing


Furnace Instrumentation Type Changes
AMS2750E
Section
Reference
3.3.6.1

18

AMS2750E Change

Comment

Vacuum furnaces of volume greater than 225 cubic feet Added clarification multi-zone
(6.4 m3) may be treated in the same fashion by dividing furnaces
the furnace into working zones that shall not exceed 225
cubic feet (6.4 m3) each of these zones shall contain
all the required sensors as outlined in Figure 4 for a
single zone.

HONEYWELL - CONFIDENTIAL

File Number

AMS 2750E Instruments - Thermal Processing


Instrumentation for Refrigeration & Quench Systems
AMS2750E
Section
Reference
3.3.3.1

3.3.3.2

19

AMS2750E Change

Comment

Refrigeration equipment shall have a temperature


controller. This temperature controller requirement is not
applicable to liquid nitrogen, dry ice and dry ice/liquidcooled containers. All refrigeration equipment shall be
equipped with a temperature recorder where time-attemperature (minimum or maximum) is required. The
above requirements are not applicable during
transportation of materials at sub-ambient temperatures.
Quench systems used for heat treatments that include a
quenchant temperature requirement (minimum,
maximum or both) shall be equipped with a recording
instrument. Existing installations installed prior to
September 2005 do not require recording instruments
until 3 years after the release of this revision.

AMS2750D did not require to have


temperature recording instrument
for quench systems that were set up
before AMS2750D

HONEYWELL - CONFIDENTIAL

AMS2750D did not require a


temperature recording instrument
for quench systems that were set up
before (Sept. 2005). If you still have
a quench system without a
recording instrument, you have 3
years to upgrade per AMS2750E to
include a recorder.

File Number

AMS 2750E System Accuracy Test


System Accuracy Test Changes Section 3.4

- Defines what equipment requires SAT, the frequency, waiver


conditions and the Test Procedure to use

- The SAT is an on-site comparison of the instrument/leadwire/


sensor readings or values, with the readings or values of a
calibrated test instrument/leadwire/sensor to determine if the
measured temperature deviations are within applicable
requirements. The SAT is performed to assure the accuracy of
the furnace control and recording system in each control zone.

20

HONEYWELL - CONFIDENTIAL

File Number

AMS 2750E System Accuracy Test


System Accuracy Test Changes
AMS2750E
Section
Reference

21

AMS2750E Change

Comment

3.4.2.1

Two sensors in each control zone are Type B, N, R, or S.

3.4.5.2.1

Resident SAT thermocouples shall be restricted to Type B, B Type T/C added


N, R or S at temperatures exceeding 500 F
(260 C) and shall be nonexpendable if exposed to
temperatures above 1000 F (538 C).

HONEYWELL - CONFIDENTIAL

B Type T/C added

File Number

AMS 2750E System Accuracy Test


AMS2750 rev. E updates Tables 6 and 7 for SAT requirements:

New

In rev D there was


mistake and
Monthly/Quarterly
was stated

New

AMS 2750E System Accuracy Test Records


AMS2750 rev. E par. 3.4.8: SAT Records
3.4.8.1 The system accuracy test report for each sensor shall include:
a.

Identification of the sensor being tested

b.

Identification of the test sensor

c.

Identification of the test instrument

d.

Date and time of day of the test

e.

Set point of the furnace during test

f.

Observed furnace instrument reading

g.

Observed test instrument reading

h.

Test sensor and test instrument correction factors

i.

Corrected test instrument reading

j.

Calculated system accuracy difference

k.

Indication of test acceptance or failure

l.

Identification of technician performing the test

New

m. SAT company (if not performed in-house)


n.

Signature of the calibration company (if not performed in-house)

o.

Quality Organization approval.


HONEYWELL - CONFIDENTIAL

AMS 2750E Temp Uniformity Survey


TUS Data Collection Changes
AMS2750E
Section
Reference
3.5.13.2

24

AMS2750E Change
Once data collection begins, temperature data shall be
recorded from all TUS sensors at a frequency of at least
one set of all readings every 2 minutes for the duration of
the survey. Data from furnace sensors required by the
applicable instrumentation type (See 3.3) shall be
recorded as follows. (Sensors whose only function is overtemperature protection do not need to be recorded.)
If the normal frequency of temperature data recording in
production is 2 minutes or less, or is continuous as in the
case of analog recorders, the data shall be documented
in the normal production format. If the normal
frequency of temperature data recording in production is
greater than 2 minutes, the recording frequency interval
during the TUS shall not exceed 6 minutes.

HONEYWELL - CONFIDENTIAL

Comment
No change with this part of the
requirement

Added statement for those


processes that record at less than 2
minutes

File Number

AMS 2750E Temp Uniformity Survey Records


AMS2750 rev. E par. 3.5.21: Temperature Uniformity Survey Report
3.5.21.1 The following items shall be included in the temperature uniformity survey report:
a.
b.
c.

Furnace identification name or number


Survey temperatures
TUS sensor and location identification including a detailed diagram, description or photograph(s)
of any load or rack used
d. Time and temperature data from all recorded sensors required for furnace
instrumentation type for all zones tested (3.5.13.3.2)
e. Correction factors for TUS sensors at each survey temperature.
New
f. As found and as left TUS offsets (if used in production)
g. Corrected or uncorrected readings of all TUS sensors at each survey temperature. Readings shall
be identified as corrected or uncorrected.
h. Testing company identification (if not performed in-house)
i. Signature for the testing company (if not performed in-house)
j. Identification of technician performing survey
k. Survey start date and time
l. Survey end date and time
m. Survey test instrument identification number
n. Indication of test pass or test fail
o. When required, documentation of furnace survey sensor failures (See 3.5.16)
p. Summary of corrected plus and minus TUS readings at each test temperature after stabilization.
q. Quality Organization approval.
HONEYWELL - CONFIDENTIAL

AMS 2750E Temp Uniformity Survey Records


AMS2750 rev. E par. 3.5.21: Temperature Uniformity
Survey Report (continued)
3.5.21.2 Although not a required part of the uniformity survey report, the following shall be
accessible on site:
a.
b.
c.
d.

Control instrument tuning parameters


TUS sensor calibration report
Control and recording sensor calibration report
Diagrams of control and recording sensors, load and TUS sensor locations in
three dimensional space.

HONEYWELL - CONFIDENTIAL

AMS 2750E Rounding


AMS2750 rev. E par. 3.8: Rounding
Added: Rounding shall be in accordance with ASTM E 29 or other national standard
a.
b.
c.
d.

Control instrument tuning parameters


TUS sensor calibration report
Control and recording sensor calibration report
Diagrams of control and recording sensors, load and TUS sensor locations in
three dimensional space.

HONEYWELL - CONFIDENTIAL

AMS 2750E Instruments


3.2.4 Controlling, Monitoring or Recording Instruments
Requirements for Controlling, Monitoring or Recording Instruments
AMS 2750E Ref.

Requirement

Comment

Readability
3.2.3.1

At least one recording and/or


controlling instrument for each zone
shall have a minimum readability of
1F or 1C

Honeywell Recorders (X-Series Paperless,


DR4500, DPR180, DPR250), HC900 OI and
Controllers (UDC2500, UDC3200 & UDC3500)
have a readability of at least 1F or 1C, most
read down to 0.1.

Installation
3.2.3.2

Installation shall conform to


manufacturers recommendation

Manuals provide detailed information on


proper installation for each product.

Signal Inputs
3.2.3.3

Instruments shall receive an


unmodified signal from sensors except
for A to D and D to A conversions, or a
digitally processed, error-checked
equivalent representation of a direct
measured value.

Honeywell Recorders, Controllers and HC900


all accept direct sensor inputs of the T/C
listed in Table 2 of AMS 2750D, many of the
recorders can accept and document input
values over communications.

Offsets
3.2.3.4

If offsets are used, a documented


procedure shall exist describing
when and how to perform offsets.

Honeywell Recorders, Controllers and HC900


allow offsets to be easily entered and shows
the offsets as part of the configurations.

Honeywell Instruments meet the requirements


for readability, installation, offsets & signal input
28

HONEYWELL - CONFIDENTIAL

File Number

AMS 2750E Instruments


3.2.5 Instrument Calibration
Requirements for Instrument Calibration
AMS 2750E Ref.

Requirement

Comment

Calibration
compliance
3.2.5.1

Calibration shall be performed on the


instruments in the Instrument Types
listed in Table 3 & comply with the
Table 3 requirements. Prior to any
instrument alteration or calibration,
document the instrument(s) as-found
condition

Honeywell Recorders (X-Series Paperless,


DR4500, DPR180 & DPR250), HC900 and
Controllers (UDC2500, UDC3200 & UDC3500) are
considered control, monitoring and recording
instruments.

Accuracy & Freq.


3.2.5.2

Calibration accuracy & frequency


shall be in accordance with Table 3

Recorders, HC900 & Controllers can meet the 2F


(1.1C) for Digital Instruments or 2F (1.1C)
/0.3% of Max Survey Temp for electro-mech. inst.

Calibration
Performance
3.2.5.4

Calibration shall be to Mfgs


instructions or a minimum of 6
simulated sensor inputs at the Min. &
Max. of the Operating Range used for
test or calib. and a minimum of 4 points
in between ...

Honeywell Recorders, HC900 & Controllers


are typically calibrated at the zero and span
values for each range. The calibration
procedure is documented as part of the
product manual.

Calibration
Performance
3.2.5.5.4

Calibration may be performed with a


load in process if furnace temp remains
within the processing tolerance and the
record is appropriately annotated to
indicated that a calibration occurred.

Honeywell Recorders (X-Series Paperless


and DPRs) provide easy annotation of the
chart to aid the documentation process.

The recorders can also be used as Field Test


instruments for doing Temp Uniformity Surveys.

We can meet the requirements

29

HONEYWELL - CONFIDENTIAL

File Number

AMS 2750E Instruments


3.2.5 Instrument Calibration
Requirements for Instrument Calibration
AMS 2750E Ref.

Requirement

Comment

Calibration
Performance
3.2.5.5.2

Calibration shall be performed on each


channel or group of channels that can
be altered or adjusted.

Honeywell Recorders, HC900 & Controllers


allow each individual channel to be
calibrated.

Chart Speed
3.2.5.5.3

Chart recorder (circular & strip)


speed(s) shall be verified annually
and shall be accurate within 3
minutes per hour

DR4500 Clock accuracy 32PPM (~1.4


min/mo), DPR180 & DPR250 Clock accuracy
10PPM (~45sec/mo)
X-Series Paperless 30PPM (~1.3min/mo)
These clock accuracies are @ reference
conditions.

Sensitivity
3.2.5.3

Sensitivity shall be checked during


calibration per Table 3 footnote 4

Sensitivity for Honeywell Recorders, HC900 &


Controllers is negligible and is covered by the
accuracy specification.

Class 1 & 2 minimum sensitivity of


1F (1C), Classes 3 6 a minimum
sensitivity of 3F (2C),

Sensitivity is the temperature change in the input that is require to activate a change, either
upscale or downscale on the instrumentation. It is normally one-half the numerical value of the
dead band, such as a sensitivity of 1F (0.6C) is equivalent to a dead band of 2F (1.1C).

Honeywell Recorders & Controllers


can meet the requirements
30

HONEYWELL - CONFIDENTIAL

File Number

AMS 2750E Instruments


3.2.6 Instrumentation Records
Requirements for Documentation of Instrument Calibration
AMS 2750E Ref.

Requirement

Comment

Instrument
Sticker
3.2.6.1

A sticker affixed to the instrument


shall indicate the most recent
successful calibration. As a
minimum the sticker shall include:

The user must establish adequate traceability


that the instrumentation has been calibrated.
The X-Series Paperless Recorder Status
Maintenance & Diagnostic screens can provide
an aid for doing this.

Date Cal performed, Due date of next


Calibration, Technician who
performed calibration, and any limits
or restrictions of the calibration
Calibration
Results
3.2.6.2

The results of calibration shall be


documented. Minimum requirements:

The user must establish a report format to


satisfy these requirements.

Inst or Furnace No., Make & Model of


instrument, Standard used, method of
calibration, Required Accuracy, As Found &
as Left data for each cal point, Offset as
found & as left, Sensitivity, Statement of
acceptance or rejection, Any limitations or
restriction, Date of Cal, Next Due Date,
Technician name, Calibration Co Name (if
not done in house, Signature of Calibration
Co , Quality Organization approval

The user must document instrument calibrations


31

HONEYWELL - CONFIDENTIAL

File Number

AMS 2750E Instruments


3.2.7 Electronic Records
Requirements for Electronic Data Recording
AMS 2750E Ref.

Requirement

Electronic
Record
Definition
3.2.7.1

An electronic record is any combination of text, graphics, data, audio, pictorial or


other information representation in digital form that is created, modified, maintained,
archived, retrieved, or distributed by a computer system.

Record Security
3.2.7.1.1

The system must create electronic


records that cannot be altered
without detection

The X-Series Paperless recorders & HC900 OI


create encrypted write-once, read only records.
Requires TrendManager Software to read data.

Software &
Playback
3.2.7.1.2

The system software and playback


utilities shall provide a means of
examining and/or compiling the record
data, but shall not provide any means
for altering the source data.

The TrendManager Pro software provides


means to import data, review & examine
data, export to other spreadsheets and
generate graphs and reports. The Software
cannot alter the source data.

Review Copies
of records
3.2.7.1.3

The system shall provide the ability to


generate accurate and complete copies
of records in both human readable and
electronic form suitable for inspection,
review and copying

The TrendManager Pro software can


generate graphs and reports of the data. The
graphs use the data as received from the
recorder for completeness and accuracy.
These are in human readable form or
readable using TrendManager Software

Comment

X-Series Recorders create secure electronic data


32

HONEYWELL - CONFIDENTIAL

File Number

AMS 2750E Instruments


3.2.7 Electronic Records
Requirements for Electronic Data Recording
AMS 2750E Ref.
Review
evidence
3.2.7.1.4

Requirement
The system shall be capable of
providing evidence the record was
reviewed such as by recoding an
electronic review, or a method of
printing the record for a physical
marking indicating review

Comment

Retention &
retrieval
3.2.7.1.4

The system shall support protection,


retention, and retrieval of accurate
records throughout the record retention
period. The hardware and/or software
shall operate throughout the retention
period as specified in section 3.7

The TrendManager Software does not alter the


source information and supports the
protection, retention and retrieval of records.
A data base Management Tool provides the
means to move, copy and mange the files for
storage on secure servers. TrendManger
Software is compatible with current PC
Operating Systems and can read data from all
generations of Trendview Paperless recorders

System Access
(Passwords)
3.2.7.1.5

The system shall provide methods (e.g.


passwords) to limit system access to
only individuals whose authorization is
documented

The X-Series Paperless Recorders provides


full password system to comply with 21CFR
Part 11. 50 Users with unique ID & Password,
time outs, expirations of passwords,
password definitions and audit trail info. The
HC900 OI provides for secured access via a
User specified numerical code.

The X-Series Paperless Recorders provides


mark on chart, ability to create a report or data
can be imported to the AMS2750 Report Tool
to create a report.

TMP provides secure analysis & data archiving


33

HONEYWELL - CONFIDENTIAL

File Number

Section 3.3 Thermal Processing Equipment


Furnace Class Uniformity & Recorder Resolution
- Classes based on Temperature Uniformity requirements
- Resolution requirements affect the recorder and chart used
- Paperless recorders record actual values eliminating issues
with the chart and the chart resolution

34

Furnace
Class

Temp
Uniformity
F

Temp
Uniformity
C

Max per Inch


(CM) of Chart
F/Inch

Max per Inch


(CM) of Chart
C/CM

Max Chart
Recording
Increment
F/Line

50

11

10

150

33

15

150

33

20

10

250

55

10

25

14

250

55

10

30

28

350

77

25

15

HONEYWELL - CONFIDENTIAL

Max Chart
Recording
Increment
C/Line

File Number

AMS 2750E System Accuracy Tests


3.4.1 System Accuracy Tests
AMS 2750E Ref.

Requirement

Comment

Requirement
3.4.1

SAT shall be performed on the


temperature control and recording
systems in each control zone of each
piece of thermal processing equipment
used for production heat treatments

System accuracy checks are required for the


controllers and recorders used for
production heat treating.

SATs shall also be performed on


additional systems that qualify
instrumentation as Types A, B, or C
Requirements
3.4.1.1

SAT shall be performed using a test


instrument meeting requirements of
Table 3 and a test sensor meeting the
requirements of Table 1

A test instrument is one used to perform system


accuracy tests, temperature uniformity surveys
or calibrations of controllers, recorders, data
acquisition instruments or monitoring
instruments. The Test Instrument calibrated
accuracy allowed is 1F (0.6C) or 0.1% of
reading in F, which ever is greater
Based on this requirement Mintitrend QX,
Multitrend SX, DPR180, DPR250 and DR4500
could be used as a test instrument, field
calibration is typically required to meet this.

Instruments used for System Accuracy tests


have a higher accuracy requirement
35

HONEYWELL - CONFIDENTIAL

File Number

Recorders That Meet AMS 2750E


Honeywell has Paper and Paperless Recorders that meet
AMS 2750E requirements for Controlling, Monitoring and
Recording Instruments & Field Test
- Key requirement for Controlling, Monitoring & Recording instruments is a
calibrated accuracy of ( 2F/1.1C) or 0.2% of reading whichever is greater for
digital instruments, 2 F (1.1 C) or 0.3% of the maximum survey temperature of
the equipment, whichever is greater for electromechanical

- Key requirement for Field Test instruments is a calibrated accuracy of (1F/0.6C)

DPR 180

DPR 250

TrendManager
Software
DR4500 Circular
Chart
36

X-Series Paperless
HONEYWELL - CONFIDENTIAL

File Number

UDC Controllers That Meet AMS 2750E


Honeywell has Controllers that meet AMS 2750E
requirements for Controlling, Monitoring and
Recording Instruments
- Key requirement for Controlling, Monitoring & Recording
instruments is a calibrated accuracy ( 2F/1.1C) or 0.2% of
reading whichever is greater for digital instruments
UDC 3200
UDC 3500

UDC 2500

37

HONEYWELL - CONFIDENTIAL

File Number

HC900 Meets AMS 2750D


The Honeywell HC900 and associated 559 & 1042
Operator Interfaces (OI) meet AMS 2750D
requirements for Controlling, Monitoring and
Recording Instruments
- Key requirement for Controlling, Monitoring & Recording
instruments is a calibrated accuracy ( 2F/1.1C)

38

HONEYWELL - CONFIDENTIAL

File Number

www.honeywell.com

You might also like