System should have capability to use Ultra Inert Column of conventional size and Ultra Inert inlet
liner.
Recommended maximum sustained column flow to MS should be 50mL/min
One unit of Split/Split less Sample Injection System with EPC:
Injection system should have gas saver mode to reduce gas consumption without
compromising performance.
System should have Electronic flow control to eliminate ghost peaks.
Split/splitless inlet should be change liners quickly and easily without special tools.
Split ratios range should be up to 7,500:1 to avoid column overload.
Pressure Range: 0 to 680 kPa.
Maximum operating temperature: 400C
Suited for diameter of columns should be 0.200 mm in 0 to 100psig pressure &
< 0.200 mm in 0 to 150 psig pressure.
Total flow setting range should be:
0 to 200 mL/min N2
0 to 1,250 mL/min H2 or He
Flame ionization detector(FID) with advanced electronic control modules, for packed &
capillary column
-Flame ionization detector (FID) which will responds to most organic compounds.
-Detector should have minimum detectable level (for tridecane): < 1.4 pg C/s
-Linear dynamic range: >107 ( 10%).
-Data rates up to 500 Hz accommodate peaks as narrow as 10 msec at half height.
-advanced electronic control modules for three gases:
- Air: 0 to 800 mL/min
- H2 : 0 to 100 mL/min
- Makeup gas (N2 or He): 0 to 100 mL/min
- will be available in two versions: capillary column optimized or adaptable for either packed or
capillary columns.
- Should be flameout detection and automatic reignition
-Capable for 450 C maximum operating temperature
Software:
The Data Processing software should file be automatic conversions on the MS Data
Analysis.
Long file name drag and drop function can be fully utilized.
System start-stop should be fully automated and controlled via the software. Reporting
format functions must be flexible format is based in users design. Like MS-words,
complete flexible function introduces ease and rapidness for making report and tidying up
the analysis result.
This software must be fully GLP/GMP and access software.
Strong security function, validation function for hardware &software and audit trial
function for data Instability are built in as a standard.
Reproducibility and recovery of concentration can be calculated and judgment of pass or
fail can be given. It is also possible to stop the run schedule based judgment.
Qualitative tools provide deconvolution and library searching across a wide selection of
general and application-specific libraries.
Quantitation capabilities to shorten path to results
Qualitative tools to extract the important information from your samples
The library search mode should be either by similarly search or index search Automatic
calculation of Retention Index for target compounds.
Should have Retention Time-Locking software, where retention time repeatability
will be <0.008% or < 0.0008 min
Should have sleep/wake mode(GC/MS) option for power saving
Outlier flagging and integration metrics help focus on problem compounds
Additionally Automatic adjustment of Retention Time in the analysis.
The instrument should be designed and manufactured under a quality system registered to ISO
9001. The instrument complies with international regulatory, safety, and electromagnetic
compatibility requirements.
Ambient operating temperature: 15 C to 35 C
Ambient operating humidity: 5% to 95%
Storage extremes: 40 C to 70 C
Line voltage requirements: 10% of nominal
Guaranteed support for the system shall have minimum ten year: value guarantee such that
vendor shall have ten years guaranteed parts and comprehensive service available or give
residual value credit for replacement
Standard: Instrument should be ISO & CE Certified.
Computer: Local or supply from manufacturer:
Intel Pentium IV Processor, Corei5
Original Intel Motherboard with Network Card
Operating System Windows7.
MS Office Original Licensed version
Printer: Laser Printer, Canon or HP
Gas & gas regulator: Required He gas with regulator should be supply by supplier.
UPS: