PIC101 VERY LOW COST PROCESS INDICATOR

SPECIFICATIONS
1. DISPLAY: 4-digit (7 segment LED) 0.5” high.
Display Messages:
“Or” - Appears when measurement exceeds
range or open sensor is detected.
“rE” - TC reverse.
2. POWER:
48x96
85 to 270 VAC/DC, 50 or 60 Hz, 5 VA.
Optional: 24VDC (Consult Factory).
t
R oH

ia n

Sc l
omp
3. SETTINGS: Via three keys on front panel.
4. MEMORY: Nonvolatile EEPROM retains all
programmable parameters and values.
FEATURES
5. MAIN SENSOR INPUT:
! Inputs: Thermocouple inputs
TC / RTD. J : -200 to 750oC.
Current : 4 to 20mADC K : -200 to 1350oC.
Voltage: 0 to 10VDC T : -200 to 400oC.
! High Indication Accuracy: ± 0.25% R : -50 to 1750oC.
! Sensor Error Compensation (offset) and Break S : -50 to 1750oC.
Detection. RTD input (2 wire or 3 wire)
! 24 VDC Sensor Supply. PT100 : -100 to 850oC.
! 85 to 270 VAC/DC Supply Voltage (also available Voltage : 0 to 10 VDC.
24 VDC models - Consult factory) Current : 0 to 20 mADC.
! Compliance: CE, RoHS o
Temperature units C/ oF
:
Resolution :
Sensor: 1 / 0.1.
PRODUCT CODE Signal input: 0.001,
0.01, 0.1 & 1 decimal
MODEL SUPPLY VOLTAGE point selectable.
6. SENSOR SUPPLY: 24 VDC to power the sensors.
PIC101-N 85 to 270 VAC/DC 7. INDICATION ACCURACY:
0.25% of Span ±1oC (After 20min. Warmup).
8. ENVIRONMENTAL CONDITIONS:
NOTE : For 24VDC models add suffix ‘- 24’
Operating Range : 0 to 50oC.
Storage Range : -5 to 50oC.
DIMENSIONS (in mm)
Humidity : 85% max.
Panel Cutout
9. ISOLATION BREAKDOWN RATINGS:
50 45 46 AC line with respect to all inputs and outputs:
2000 Volts.
97.5
88 92 10. CONNECTION: Wire clamping screw terminals.
10
11. WEIGHT: 210 grams.
TERMINAL CONNECTIONS

mV/TC/RTD
+ - V mA CONFIG
A B B’ + - + -

1 2 3 4 5 6 7 8 9 10
22 24
21 23
11 12 13 14 15 16 17 18 N L
+24V GND
sensor supply - +

Industrial Supply Syndicate, 54, Ezra Street, Kolkata - 700 001


Phone: 32916080, 22356676 Fax: 033 3022 2923 Email: info@industrialindia.com