STARFAST: a Wireless Wearable EEG Biometric System based

on the ENOBIO Sensor


H.S.PAVAN KUMAR C.ABDUL KHALEEK
Madina engineering college Madina engineering college

ABSTRACT—we present a wearable, Rate (FAR) of 0.82%. The obtained


wireless biometry system based on the new performance measures improve the results
ENOBIO 4- channel electrophysiology of similar systems presented in earlier work.
recording device developed at Star lab. I. INTRODUCTION
Features extracted from The present paper introduces a method to
electroencephalogram (EEG) and authenticate people from their physiological
electrocardiogram (ECG) recordings have activity, concretely the combination of ECG
proven to be unique enough between and EEG data. We call this system
subjects for biometric applications. We STARFAST (STAR Fast Authentication
show here that biometry based on these bio-Scanner Test). Several biometric
recordings offers a novel way to robustly modalities are already being exploited
authenticate or identify subjects. In this commercially for person authentication:
paper, we present a rapid and unobtrusive voice recognition, face recognition and
authentication method that only uses 2 fingerprint recognition are among the more
frontal electrodes and a wrist worn electrode common modalities nowadays. But other
referenced to another one placed at the ear types of biometrics are being studied as
lobe. Moreover, the system makes use of a well: ADN analysis, keystroke, gait, palm
multistage fusion architecture, which print, ear shape, hand geometry, vein
demonstrates to improved system patterns, iris, retina and written signature
performance. The performance analysis of [33]. Although these different techniques for
the system presented in this paper stems authentication exist nowadays, they present
from an experiment with 416 test trials, some problems. Typical biometric traits,
where an Equal Error Rate (EER) of 0% is such as fingerprint, voice and retina, are not
obtained after the EEG and ECG modalities universal, and can be subject to physical
fusion and using a complex boundary damage (dry skin, scars, loss of voice...). In
decision. If a lineal boundary decision is fact, it is estimated that 2-3% of the
used we obtain a True Acceptance Rate population is missing the feature that is
(TAR) of 97.9% and a False Acceptance required for authentication, or that the
provided biometric sample is of poor problem of entity authentication in body
quality. Furthermore, these systems are area sensor network (BASN) for m-Health.
subject of attacks such as presenting a Physiological signals detected by biomedical
registered deceased person, dismembered sensors have dual functions: (1) for a
body part or introduction of fake biometric specific medical application, and (2) for
samples. It also compares the performance sensors in the same BASN to recognize each
of the system with the Active Two system other by biometrics. A feasibility study of
from Biosemi. The Active Two system is proposed entity authentication scheme was
state of the art commercial equipment that is carried out on 12 healthy individuals, each
proven to have very good performance and with 2 channels of photoplethysmogram
has been used in many studies. A list of (PPG) captured simultaneously at different
some works can be found in [1]. So far 3 parts of the body. The beat-to-beat heartbeat
demonstration applications have been interval is used as a biometric characteristic
developed using the system. EOG based to generate identity of the individual. The
Human Computer Interface (HCI), EEG and results of statistical analysis suggest that it is
ECG based Biometry for Authentication a possible biometric feature for the entity
(presented at pHealth as a poster) and an authentication of BASN.
EEG based Sleepiness Prediction system for Advantages Since every
drivers. In all cases the applications are living and functional person has a recordable
designed around the same 4 channel system. EEG/ECG signal, the EEG/ECG feature is
ENOBIO has been developed with the help universal. Moreover brain or heart damage
of the SENSATION FP6 IP 507231. is something that rarely occurs, so it seems
With the evolution of m-Health, an to be quite invariant across time. Finally it
increasing number of biomedical sensors seems very difficult to fake an EEG/ECG
will be worn on or implanted in an a signature or to attack an EEG/ECG
individual in the future for the monitoring, biometric system. An ideal biometric
diagnosis, and treatment of diseases. For the system should present the following
optimization of resources, it is therefore characteristics: 100% reliability, user
necessary to investigate how to interconnect friendliness, fast operation and low cost. The
these sensors in a wireless body area perfect biometric trait should have the
network, wherein security of private data following characteristics: very low intra-
transmission is always a major concern. This subject variability, very high inter-subject
paper proposes a novel solution to tackle the variability, very high stability over time and
universality. In the next section we show the four channels plus electrical reference can
general architecture and the global be recorded. The hardware is power with a
performance of the system we have rechargeable Li-Ion battery for long
developed. duration. The communication of ENOBIO
with the Laptop is done via ZigBee to USB
SYSTEM DESCRIPTION AND receiver. The electrodes can be active
SPECIFICATIONS: Ag/AgCl or optionally dry. We can see in
ENOBIO is a wearable, wireless, 4- Figure 1 ENOBIO recording EEG data and
channel, all-digital electrophysiology transmitting them wirelessly to the laptop to
recording system that has been optimized for be processed for an on-line application.
dry electrodes. The system has been METHODS AND RESULTS:
designed in a modular way to allow In this paper, we present a rapid and
researchers to build systems tailored to their unobtrusive authentication/identification
needs. Robust, snap-in cables of any length method that only uses 2 frontal electrodes
allow any combination of EEG, EOG and (for EEG recording) with another electrode
ECG in a single wearable system without placed on the left wrist (for ECG
loss of signal quality. ENOBIO employs recording)and all 3 referenced to another
three fundamental enabling aspects to bring one placed at the right earlobe. Moreover the
electrophysiology into daily life: system makes use of a multi-stage fusion
- Active digital electrodes digitize architecture, which has been demonstrated
the signal on site in order to reduce to improve system performance. The
environmental noise. EEG/ECG recording device is ENOBIO [23,
24, 32, 35], a product developed at
- Modular design allows users to customize STARLAB BARCELONA SL. It is a
a single system for various applications wireless 4 channel (plus the common mode)
device with active electrodes. It is therefore
- Wireless communication allows wearable quite unobtrusive, fast and easy to place.
recording of data away from the lab or Even though ENOBIO can work in dry
controlled environments. The precision of mode, in this study conductive gel has been
the signal that can be acquired with used (see fig. 1).
ENOBIO is 1-3 uVpp to 1 mV from 0.1 to
100 Hz, and the output is a digital signal at
256 Samples/s per channel, 16 bit. Up to
decision. The enrolment and the
authentication test follow a common path
until the feature extraction. Then the red
arrows indicates the enrolment path and the
green ones the authentication path. Briefly,
the preprocessing module segments the
recorded takes into 4 seconds epochs. From
each one of those epochs we extract 5
Figure1: ENOBIO sensor embedded in a
different features: auto-regression
cap. We can see the 4 channel inputs (red
coefficients, fast Fourier transform, mutual
pins). In the picture, only two channels are
information, coherence and cross correlation
plugged. The data are transmitted wirelessly
[10, 11, 12, 19, 20, 21]. The classifier used
to the laptop. The general schema used is
in both the signature extraction module and
described below: Fig. 2 General scheme of
in the recognition module is the Fisher
the method. The data acquisition module is
Discriminant Analysis [13], with 4 different
the software that controls the ENOBIO
discriminant functions (linear, diagonal
sensor in order to capture the raw data.
linear, quadratic and diagonal quadratic).
Remember that 4 channels are recorded: 2
During the enrolment process, we need four
EEG channels placed in the forehead, 1
3-minutes takes, and these are used to select
ECG channel placed in the left wrist and 1
the best combinations of features and
electrode placed in the right earlobe for
classifiers, and then to train the classifiers
referencing the data. At this point the data is
for each subject. This new feature is what
separate in EEG data and ECG data and sent
we call ‘personal classifier’ approach, and it
to two parallel but different biometric
improves the performance of the system
modules for EEG and ECG. (the difference
considerably. The authentication test takes
is not shown in the scheme for simplicity).
are 1-minute long. The preprocessing and
For the signature extraction module, four 3-
the feature extraction are equivalent to the
minutes takes are needed. Once the
enrolment case. The recognition module
signatures are extracted, they are both stored
loads the trained classifiers for each
in the database for further retrieval when an
modality (ECG and EEG) and an
authentication process takes place. Then the
independent score is then provided for each.
recognition module provides an
authentication score for both modalities and INITIAL SYSTEM
finally the fusion module provides the final BENCHMARKING:
For the signal analysis we compared
the signal recorded with ENOBIO and
Biosemi simultaneously. The electrodes of
every system where placed in Fp1 of 1 cm
between them. We analyzed the signal
referenced to left ear lobe. We recorded
spontaneous EEG, 1 minute with eyes open
And 1 minute with the eyes closed for 2
subjects, using wet Ag/AgCl electrodes for Figure 3: Spectral density of the signal
both systems. Figures 2 to 5 show the power recorded with ENOBIO and Biosemi in fp1
spectral density of the signals recorded with referenced to left ear lobe. Sample was 1
ENOBIO (blue line) and Biosemi (red line). minute of spontaneous EEG with eyes close.
We can see a very good match in the DEMONSTRATION
spectrums of both systems. Just a small peak APLICATIONS:
of 50 Hz appears for ENOBIO for subject 1. Tree different applications have been
built on ENOBIO system. In this section we
make a fast overview of them. A EOG based
Human Computer Interface. Star lab
developed a system to track eye movements.
When we look at a scenario, we focus our
objects of interest changing our view with
fast eye movements. These movements are
detected by the system and their amplitude
And direction extracted from the EOG
Figure 2: Power spectral density of the signal. The signal is acquired by ENOBIO
signal recorded with ENOBIO and Biosemi and sent wirelessly to the ENOBIO server
in fp1 referenced to left ear lobe. Sample installed in a laptop. This raw signal can be
was 1 minute of spontaneous EEG with eyes sent via TCP/IP to any application, in that
open. case to the EOG HCI system.
The application processes the
data in real time and outputs a vector that
represents the movement on the screen. The
System is set on a cap as shown in Figures 1 himself), 350 impostor situations (when an
and 8. enrolled subject claims to be another subject
from the database) and 16 intruder situations
(when a subject who is not enrolled in the
system claims to be a subject belonging to
the database). Once the EEG and the ECG
biometrics results are fused, using a
complex boundary decision (dotted line 2 in
fig. 3), we can obtain an ideal performance,
Figure 4: EOG-HMI system. 3 electrodes
that is True Acceptance Rate (TAR) = 100%
are placed on the forehead; the 4th one and
and False Acceptance Rate (FAR) = 0%. If a
the electrical reference are placed in the ear
linear boundary decision is used (line 1 in
lobes with a clip. Figure 7 shows a graphical
fig. 3), we obtain a TAR = 97.9% and a
representation of the output of a movement.
FAR = 0.82%. The results are summarized
in table I. Bidimensional decision space. An
ordinate represent the ECG probabilities and
abscises the EEG probabilities. Red crosses
represent impostor/intruder cases and green
crosses represent legal cases. Two decision
functions are represented this system has
been tested as well to validate the initial

Figure 5: The vector of the picture state of users, and has been proved sensitive

represents the eye gaze movement of a user enough to detect it. If a subject has suffered

of the EOG Human Computer Interface. The from sleep deprivation [28], alcohol intake

absolute error in the prediction for the [26, 27] or drug ingestion when passing an

movements is 5, 9 % with a standard authentication test, the authentication

deviation of 6.6 for the X axe, and 9, 9% performance decreases. This fact provides

with a standard deviation of 13, 1 for the Y evidence that such a system is able to detect

axe. Then the fusion module [22] provides a not only the identity of a subject but his state

final decision about the subject as well.

authentication. In order to test the TABLE I

performance of our system we use 48 legal


situations (when a subject claims to be
validate in a continuous way that the person
supposed to be tele-present in an audiovisual
interactive space is actually the person that
is supposed to be. This could facilitate the
personalization of the reaction of the virtual
environment [31], or secure interactions that
CONCLUSION AND FUTURE
guarantee the authenticity of the person
WORK:
behind. Although the system described here
These results show that the
was oriented towards person identification,
authentication of people from physiologic
its performance has been significantly
data can be achieved using techniques of
modified by introducing physiological data
machine learning. Concretely it shows that
obtained in altered states such as the ones
the fusion of two (or more) independent
resulting from sleep deprivation. This
biometric modules increases the
indicates that such a system could be used to
performance of the system by applying a
extract dynamically changing, such as
fusion stage after obtaining the biometric
physiological activity related to mood or to
scores. This result shows that processing the
intense cognitive activity. For example, it
different physiological modalities separately
could be used to extract the features
on different processing modules, and
described (sleep deprivation, alcohol intake),
introducing a data fusion step, the resulting
but also information about basic emotions
performance can be increased. Applying a
(see, for example, [34]). The dynamic
very similar approach, we could easily adapt
extraction of such features could be used to
the system to do emotion recognition [29,
evaluate the response of people in virtual
30, and 34] from physiological data, or
environments, as well as adapt the behavior
develop a Brain Computer Interface, just
of such environments to the information
starting from different ground truth data.
extracted dynamically. The advantage that
From our point of view, this easy to extend
such a system would present compared to
feature of our system is the more interesting
currently existing approaches is that –by
part of our study along with the ‘personal
having built a decisional system that is
classifier’ approach which improves
completely modular- we could estimate each
considerably the performance of the system.
class independently of the other classes
The system described could have different
obtained, which would allow us to have a
applications for Virtual Reality. It can
modular approach to define the interaction
between the human and the machine. Finally [4] Mohammad G. et al. (2006) Person
we would like to highlight the usefulness of identification by using AR model
using a single system for user state For EEG signals. Proc. 9th International
monitoring and user personalization. Conference on Bioengineering
Assuming that the system is worn anyway Technology (ICBT2006), Czech Republic,5
for physiological feedback, this same system pp.
can personalize your experience because it [5] Para jape R. et al.(2001)The
knows who you are without any interaction electroencephalogram as a biometric.
required by the user. Proc. Canadian Conf. On Electrical and
ENOBIO is an electrophysiology recording Computer Engineering, pp.
wireless system. The quality of the signal is 1363-1366.
proven to be as good as the signal that can [6] Poulos M. et al. (1998) Person
be acquires with state of the art wired identification via the EEG using
equipment. Different applications can be Computational geometry algorithms.
built on the ENOBIO system. We presented Proceedings of the Ninth
a biometric application, a Human Machine European Signal Processing, EUSIPCO’98,
Interface and a Sleepiness prediction system. Rhodes, Greece.
REFERENCES September1998, pp. 2125-2128.

[1] Eischen S., Luctritz J. and Polish J. [7] Poulos M. et al. (1999) parametric

(1995) Spectral analysis of EEG person identification from EEG

From Families. Biological Psycholology, Using computational geometry. Proc. 6th

Vol. 41, pp. 61-68. International Conference on

[2] Hazarika N., Tsoi A. and Sergejew A. Electronics, Circuits and Systems

(1997) nonlinear considerations (ICECS’99),v. 2, pp. 1005-1008.

in EEG signal Classification. IEEE


Transactions on signal processing,
Vol. 45, pp.829-836.
[3] Marcel S., Mill J. (2005) Person
authentication using brainwaves
(EEG) and maximum a posteriori model
adaptation. IDIAP Research
Report 05-81, 11 pp.