This Joint Australian/New Zealand Standard was prepared by Joint Technical Committee
HT/23, Medical Diagnostic Imaging Equipment. It was approved on behalf of the Council
of Standards Australia on 21 December 1994 and on behalf of the Council of Standards
New Zealand on 7 February 1995. It was published on 5 May 1995.
STANDARDS AUSTRALIA
1 The Crescent,
Homebush NSW 2140 Australia
PREFACE
This Standard was prepared by the Joint Standards Australia/Standards New Zealand Committee
HT/23 on Medical Diagnostic Imaging Equipment, as a Joint Standard under the terms of the Active
Cooperation Agreement between Standards Australia and Standards New Zealand.
This Standard is identical with and has been reproduced from IEC 336:1993, X-ray tube assemblies
for medical diagnosis — Characteristics of focal spots.
Under arrangements made between Standards Australia/Standards New Zealand and the IEC, as well
as certain other Standards organizations, users of this Standard are advised that the number of this
Standard is not reproduced on each page; its identity is shown only on the cover and title page.
The sources for terms in capital letters are listed in Annex C.
For the purpose of this Standard, the IEC text should be modified as follows:
This is a free 7 page sample. Access the full version at http://infostore.saiglobal.com.
(a) Terminology The words ‘this Australian/New Zealand Standard’ should replace the words ‘this
International Standard’ wherever they appear.
(b) Decimal marker Substitute a full point for a comma where it appears as a decimal marker.
(c) References The references to international Standards should be replaced by references to the
following Australian or Joint Australian/New Zealand Standards:
Reference to International Standard Australian/New Zealand Standard
or other Publication
IEC AS/NZS
417 Graphical symbols for use on —
equipment. Index, survey and
compilation of the single sheets
601-2-28 Part 2-28: Particular requirements 3200.2.28 Part 2.28: Particular requirements
for the safety of X-ray source for safety—X-ray source assemblies
assemblies and X-ray tube and X-ray tube assemblies for
assemblies for medical diagnosis medical diagnosis
The term ‘informative’ has been used in this Standard to define the application of the Annexes to
which they apply. An ‘informative’ appendix is only for information and guidance.
CONTENTS
Clause SECTION 1: GENERAL Page
Figures
1 Essential dimensions of the slit diaphragm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2 Alignment of the SLIT CAMERA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
3 Reference dimensions and planes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
4 Directions of evaluation over distorted FOCAL SPOTS . . . . . . . . . . . . . . . . . . . . . . . . 6
5 Essential dimensions of a pinhole diaphragm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
6 Alignment of the PINHOLE CAMERA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
7 Essential dimensions of the test pattern . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
8 Alignment of the STAR PATTERN CAMERA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
9 Alignment of the slit of the microdensitometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
10 Limitation of input data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
11 Image of the test pattern . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
B.1 Projection of the ACTUAL FOCAL SPOT on the REFERENCE PLANE . . . . . . . . . . . . . . . . 29
Tables
1 Magnification for FOCAL SPOT SLIT RADIOGRAMS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2 LOADING FACTORS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3 Dimensions of the pinhole . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
4 Magnification for FOCAL SPOT PINHOLE RADIOGRAMS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
5 Permissible values of FOCAL SPOT dimensions for NOMINAL FOCAL SPOT VALUES . . . . . . 15
6 Standard magnification for MODULATION TRANSFER FUNCTIONS . . . . . . . . . . . . . . . . . . . . . . 21
7 Standard magnification for STAR PATTERN RESOLUTION LIMIT . . . . . . . . . . . . . . . . . . . . . . . . 24
8 LOADING FACTORS for the determination of the BLOOMING VALUE . . . . . . . . . . . . . . . . . . . . . 25
Annexes
A Supply of SLIT CAMERAS, PINHOLE CAMERAS and STAR PATTERN CAMERAS . . . . . . . . . . . . . 27
B Alignment to the REFERENCE AXIS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
C Index of defined terms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
1
SECTION 1: GENERAL
This International Standard applies to FOCAL SPOTS in diagnostic X- RAY TUBE ASSEMBLIES for
medical use, operating at X- RAY TUBE VOLTAGES up to and including 200 kV.
This International Standard describes measuring methods and requirements for test
This is a free 7 page sample. Access the full version at http://infostore.saiglobal.com.
arrangements to be applied for evaluating compliance with this standard together with the
means of indicating compliance both of the specified characteristics of FOCAL SPOTS and the
RADIOGRAMS prepared in checking these characteristics.
The following normative documents contain provisions which, through reference in this text,
constitute provisions of this International Standard. At the time of publication, the editions
indicated were valid. All normative documents are subject to revision, and parties to
agreements based on this International Standard are encouraged to investigate the
possibility of applying the most recent editions of the normative documents indicated below.
Members of IEC and ISO maintain registers of currently valid International Standards.
IEC 417:1973, Graphical symbols for use on equipment. Index, survey and compilation of
the single sheets
IEC 417G: 1985, Graphical symbols for use on equipment. Index, survey and compilation of
the single sheets — Seventh supplement
IEC 601-2-28:1993, Medical electrical equipment — Part 2: Particular requirements for the
safety of X-ray source assemblies and X-ray tube assemblies for medical diagnosis
IEC 613: 1989, Electrical, thermal and loading characteristics of rotating anode X-ray tubes
for medical diagnosis
1.3 TermInology
In this standard: