You are on page 1of 116

Oil and Gas Measuring Instruments

1
Course Aim

The aim of this training course is to build up the procedural and
declarative knowledge required to be recognized by projects engineer that
do not have past background of oil and gas measuring instruments. This
will help them to supervise projects dealing with instrumentation in plants
with a strong background.
In this course, the training cycle is divided in five steps that necessitate
the cooperation between the instructor and the trainees. These steps are
shown in figure below, they are summarized as follows:
1. Define the knowledge and skills required to be developed.
2. Define the elements of each knowledge or skill.
3. Formulate a verbal phrase for the learning objective of each
element.
4. Choose an adequate instructional activity to present each element.
5. Set up an indicator to measure the outcomes of the course and
modify the training skills to adapt the vocational needs.












Determine
Elements
Learning
Objectives
Define
Knowledge
& Skills
Instruction
Activity
Measure
& Correction
Determine
Elements
Learning
Objectives
Define
Knowledge
& Skills
Instruction
Activity
Measure
& Correction

Training Cycle.
Oil and Gas Measuring Instruments
2
Knowledge and Elements

Introduction to measurements.
Introduce general terms.
Introduce quantities and units.

Distinguish between different gauges and switches.
Introduce how quantity is measured.
Illustrate main components of instrument.
Classify different types of measuring instruments.

Develop knowledge about different transmitters and sensing
elements.
Establish knowledge base about transmitter technology.
Introduce Sensing Element.
Introduce theory of operation.

Introduce some analyzers.
Gas Chromatography.
Moisture Analyzer.
Oxygen Analyzer.








Oil and Gas Measuring Instruments
3
Table of Contents

Section I
Chapter 1 Introduction to Measuements 5
Chapter 2 Transmitters 16
Section II
Chapter 3 Mechanical Transducers 25
Chapter 4 Electric Transducers 36
Chapter 5 Flowmeters 73
Section III
Chapter 6 Analyzers 102
Chapter 7 Basic Considerations 109
















Oil and Gas Measuring Instruments
4























Oil and Gas Measuring Instruments
5
Chapter 1
Introduction to Measurement

1.1 Learning objectives
1. Introduce measurements and instruments.
2. Classify instruments and functions.
3. Understand instruments characteristics.

1.2 Measurements

The measurement of a given quantity is an act or the result of
comparison between the quantity and a predefined standard. Since two
quantities are compared, the result is expressed in numerical values. In
fact, the measurement is the process by which one can convert physical
parameters to meaningful numbers. In order that the results are
meaningful, there are two basic requirements:
1. The standard used for comparison purposes must be accurately
defined and should be commonly accepted.
2. The apparatus used and the method adopted must be proved.

1.2.1 Significance of Measurements

The advancement of science and technology is dependent upon a
parallel progress in measurement techniques. There are two major
functions in all branches of engineering:
1. Design of equipment and processes.
2. Proper operation and maintenance of equipment and processes.
Both of these functions require measurements.
Oil and Gas Measuring Instruments
6
1.2.2 Methods of Measurements

- Direct Method: The unknown quantity is directly compared against
a standard.
- Indirect Method: Measurement by direct methods are not always
possible, feasible and practicable. These methods in most of the
cases are inaccurate because of human factors. They are also less
sensitive.

1.2.3 Instruments

In simple cases, an instrument consists of a single unit which gives
an output reading or signal according to the unknown variable applied to
it. In more complex situations, a measuring instrument consists of several
separate elements. These elements may consist of transducer elements
which convert the measurand to an analogous form. The analogous signal
is then processed by some intermediate means and then fed to the end
devices to present the results for the purposes of display and or control.
These elements are:
- A detector.
- An intermediate transfer device.
- An indicator.

The history of development of instruments encompasses three phases:
- Mechanical.
- Electrical.
- Electronic.

Oil and Gas Measuring Instruments
7
1.2.4 Classification of Instruments

- Absolute instruments: These instruments give the magnitude of the
quantity under measurement in terms of physical constants of the
instrument. Example: Galvanometer.
- Secondary Instrument: These instruments are constructed that the
quantity being measured can only be measured by observing the
output indicated by the instrument.

1.2.4.1 Deflection Type

The deflection of the instrument provides a basis for determining
the quantity under measurement as shown in figure (1.1).

Figure 1.1 Deflection Type
1.2.4.2 Null Type

A zero or null indication leads to determination of the magnitude of
measured quantity as shown in figure (1.2).
Oil and Gas Measuring Instruments
8

Figure 1.2 Null Type
1.2.4.3 Contact Type

Often when a measured pressure reaches a certain max or min
value, it is desirable to have an alarm sound a warning, a light to
give a signal, or an auxiliary control system to energize or de-energize. A
micro switch is the device commonly used for this purpose.


Figure 1.3 Contact Type
Oil and Gas Measuring Instruments
9
1.2.5 Analog and Digital Modes of Operation

- Analog Signal: signals that vary in a continuous fashion and take
an infinite number of values in any given range.
- Digital signal: signals that vary in discrete steps and thus take only
finite different values in a given range.

1.2.6 Functions of Instruments

- Indicating function.
- Recording function.
- Controlling Function.

1.3 Characteristics of Instruments

1.3.1 Performance

It is to define a set of criteria that gives a meaningful description of
quality of measurement. Performance characteristics are obtained in one
form or another by a process called calibration. The calibration of all
instruments is important since it affords the opportunity to check the
instrument against a known standard.

1.3.2 Errors in Measurement

Measurements always involve errors. No measurement is free from
errors. An understanding and thorough evaluation of the errors is
essential.
Oil and Gas Measuring Instruments
10

Figure 1.4 Visual error
1.3.3 True Value

True Value: The true value of quantity to be measured may be
defined as the average of an infinite number of measured values when the
average deviation due to the various contributing factors tends to zero.

1.3.4 Ranges

- Scale range: it is defined as the difference between the largest and
the smallest reading of the instrument, i.e. scale range from 200 to
500 degree C.
- Scale Span: It is may be confusing with scale range but it is given
to be 300 degree C.
- Effective Range: It is defined as the range over which it meets
some specified accuracy requirements.
- Rangeability (turndown): If the effective range is from A to B, then
the rangeability is defined by B/A.

1.3.5 Discrimination, Accuracy, Error, Precision and Sensitivity

- Discrimination (Resolution): It is used to describe how finely an
instrument can measure. For example, the discrimination of a
Oil and Gas Measuring Instruments
11
digital electronic timer reading in milliseconds is a hundred times
as great as that of a stopwatch graduated in tenths of seconds. It is
often wrongly referred as sensitivity.
- Accuracy: It is the closeness with which the instrument reading
approaches the true value of the quantity. Thus accuracy means
conformity to truth.
- Error: It is defined as the difference between the measured value
and the true value. One kind of error is observational error.
- Precision: It is a measure of the degree of agreement within a group
of measurements. High precision means a tight cluster and repeated
results while low precision indicates a broad scattering of results.
- Certainty: It is often used as a synonym for accuracy. However,
Uncertainty is the property of a measurement rather than the
instrument used to make the measurement.
- Sensitivity: It is a measure of how an instrument is sensitive to the
measured quantity variation. It is the ability to produce detectable
output.

Figure 1.5 Accuracy and Repeatability

1.3.6 Reproducibility, Repeatability and Hysteresis

- Reproducibility: It is the closeness of agreement among repeated
measurements of the output for the same value of input mode under
Oil and Gas Measuring Instruments
12
the same operating condition over a period of time, approaching
from both directions.
- Repeatability: It is the closeness of agreement among a number of
consecutive measurements of the output for the same value of input
under the same operating conditions, approaching from the same
direction.

Figure 1.6 Repeatability

- Hysteresis and Dead Band: It is the maximum difference for the
same input between the upscale and downscale output values
during a full range transverse in each direction.
- Dead Time: It is defined as the time required by an instrument to
begin to respond to a change in the measurand.
- Dead Zone: It is defined as the largest change in which there is no
output from the instrument.
Oil and Gas Measuring Instruments
13

Figure 1.7 Hysteresis and Dead band

1.3.7 Drift

Perfect Reproducibility means no drift. No drift means that with a
given input the measured values do not vary with time.
- Zero Drift: if the whole calibration gradually shifts.
- Span Drift: If there is a proportional change in the indication all
along the upward scale.
- Zonal Drift: In case the drift occurs only over a portion of the span.

Figure 1.8 Drift
Oil and Gas Measuring Instruments
14
1.3.8 Noise

A spurious current or voltage extraneous to the current or voltage
of interest in an electrical or electronic circuit is called noise.

1.3.9 Linearity

It is the closeness to which a curve approximates a straight line. It
is a measure of the extent to which the instrument calibration curve over
its effective range departs from the best fitting straight line.

Figure 1.9 Linearity

1.3.10 Loading Effects

The ideal situation in a measuring system is that when an element
used for any purpose, the original signal should remain undistorted. In
practical conditions, it has been found that any element in the system
extracts energy and thereby distorting the original signal.

Oil and Gas Measuring Instruments
15
1.3.11 Other Effects

- Temperature Effect
- Pressure Effect
- Vibration Effect

1.4 Role Play

Each Trainee should speak thoroughly about one of the learning objective
elements.


















Oil and Gas Measuring Instruments
16
Chapter 2
Transmitters

2.1 Learning Objectives

1. Introduce history of transmitter technology.
2. Understand analog transmitters.
3. Understand smart transmitters with HART protocol.

2.2 Transmitter Technology

Transmitters are instruments that transfer measured output signal to
distance places where it is needed. The technology development through
years is:

1. Pneumatic and Hydraulic.
2. Electrical (Analog 4-20 mA).
3. Electronic (Analog 4-20 mA + Digital HART protocol).
4. Electronic (All digital Foundation Fieldbus).

Figure 2.1 Pneumatic Transmitter
Oil and Gas Measuring Instruments
17
2.3 Analog Transmitters

Analog transmitter uses a variable conversion element to translate
and accommodate the physical non-electrical measurand to electrical
analog signal (4-20 mA).

Figure 2.2 Analog Transmitter

2.3.1 Measurement Converters of Electrical Quantities

- Measuring amplifiers: demands on measuring amplifiers, negative
feedback, ideal operational amplifier, basic circuits of measuring
amplifiers using operational amplifiers (OAs)
- Measurement of low voltages and currents using OAs, estimating
uncertainty of measurement (including influence of input voltage
offset and input bias).
- Rectifiers (converters of the rectified mean value).

2.3.2 Ideal Operational Amplifiers

Figure 2.3 Ideal OP-Amp
Oil and Gas Measuring Instruments
18
2.3.3 Inverting amplifier

Figure 2.4 Inverting Amplifier
2.3.4 Current to Voltage Converter

Figure 2.5 Current to Voltage converter
2.3.5 Voltage Controlled Current Source

Figure 2.6 Voltage controlled Current source
2.3.6 Rectifiers

Figure 2.7 Rectifier
Oil and Gas Measuring Instruments
19
2.3.7 Adders

Figure 2.8 Adders

2.3.8 Differential Amplifiers

Figure 2.9 Differential Amplidier

2.3.9 Integrators

Figure 2.10 Integrators

Oil and Gas Measuring Instruments
20
2.4 HART Protocol

2.4.1 HART Overview

For many years, the field communication standard for process
automation equipment has been a milliamp analog current signal. HART
field communications protocol extends the 4-20 mA standards to enhance
communication with smart field instruments. It was designed for use with
intelligent measurement and control instruments which traditionally
communicate using mA analog signals. HART preserves the 4-20 mA
signals and enables two way digital communications to occur without
affecting the integrity of 4-20 mA signal.

Figure 2.11 Hart Digital Signal
HART, highway addressable remote transducer, makes use of Bell
202 FSK standard to superimpose digital signal at a low level on top of
analog signal; i.e. 1200 Hz for logic 1 and 2200 Hz for logic 0. HART
communicates 1200 bps without interrupting the mA signal and allows a
host application to get two or more digital updates per second from a field
device.
Oil and Gas Measuring Instruments
21

Figure 2.12 HART Connection
HART is a master/slave protocol which means that a field device
(slave) only speaks when spoken to by a master. HART provides for up to
two masters, primary and secondary, as shown in figure (2.12).

Figure 2.13 Master/Slave
The most commonly employed communication mode is the
master/slave, figure (2.13). The optional burst communication mode
where a slave device can continuously broadcast a HART reply message,
figure (2.14).

Figure 2.14 Burst
2.4.2 HART Benefits

2.4.2.1 35-40 data items Standard in every HART device
- Device Status & Diagnostic Alerts;
- Process Variables & Units;
- Loop Current & % Range;
- Basic Configuration Parameters;
- Manufacturer & Device Tag;
Oil and Gas Measuring Instruments
22
2.4.2.2 Increases control system integrity
- Get early warning of device problems;
- Use capability of multi-variable devices;
- Automatically track and detect changes (mismatch) in Range
or Engineering Units;
- Validate PV and Loop Current values at control system
against those from device;

2.4.2.3 HART is Safe, Secure, and Available
- Tested and Accepted global standard;
- Supported by all major instrumentation manufacturers;

2.4.2.4 Saves Time and Money
- Install and commission devices in fraction of the time;
- Enhanced communications and diagnostics reduce
maintenance & downtime;
- Low or no additional cost by many suppliers;

2.4.2.5 Improves Plant Operation and Product Quality
- Additional process variables and performance indicators
- Continuous device status for early detection of warnings and
errors
- Digital capability ensures easy integration with plant
networks
2.4.2.6 Protects Your Asset Investments
- Compatible with existing instrumentation systems,
equipment and people
- Allows benefits to be achieved incrementally
- No need to replace entire system
Oil and Gas Measuring Instruments
23
2.5 Role Play

Each Trainee should speak thoroughly about one of the learning objective
elements.
- Analog Transmitters
- Smart Transmitters and HART Protocol.























Oil and Gas Measuring Instruments
24






















Oil and Gas Measuring Instruments
25
Chapter 3
Mechanical Transducers

3.1 Learning objectives

1. Understand the theory of operation of different sensing elements.

3.2 Springs

Most mechanical input instruments employ mechanical springs of
one form or another. Various common types of springs are shown in
figure (3.1). These range from cantilever, helical and spiral springs.








Figure 3.1 Springs

3.3 Pressure Sensing Elements

Most pressure devices use elastic elements for sensing pressure at
the primary stage. A link and gear mechanism are used to convert the
movement to rotational motion to be connected the scale and pointer.


Oil and Gas Measuring Instruments
26
3.3.1 Bourdon Tubes

The bourdon tubes are made out of an elliptical flattened bent tube.
One end is sealed and the other is open for fluid to enter. The pressure of
the fluid tends to straighten out the tube. This motion is transferred to the
pointer.

3.3.1.1 C-Type
It is the most used for local indication.

Figure 3.2 Bourdon Type
Oil and Gas Measuring Instruments
27
3.3.1.2 Spiral Type
Increasing the number of turns will increase the displacement of the free
tip without changing the wall thickness.

Figure 3.3 Spiral type

3.3.1.3 Helical Type
The displacement of the tip of the helical type is larger than that of the
spiral one.

Figure 3.4 Helical type
3.3.2 Bellows

A metallic bellows is a series of circular parts, resembling the folds
in an accordion. The parts are designed in such a way that there are
expanded and contracted.

Oil and Gas Measuring Instruments
28


Figure 3.5 Bellows Type
3.3.3 Diaphragms

The operating principle of diaphragm elements is similar to that of
the bellows. The pressure applied causes it to deflect where the deflection
is proportional to the applied pressure.

Figure 3.6 Diaphragm Type
Oil and Gas Measuring Instruments
29
13.4 Temperature Sensing Elements
3.4.1 Bimetallic Thermometer
They are used for local temperature measurements. It is constructed
by bonding two different metals such that they cannot move relative to
each other. All metals try to change their physical dimensions at different
rates when subjected to same change in temperature. The differential
change in expansion of two metals results in bending or flattening the
structure, which in turn moves the pointer via the intermediate element.

3.4.1.1 Strip

Figure 3.7 Strip Type
3.4.1.2 Spiral

Figure 3.8 Spiral type
Oil and Gas Measuring Instruments
30
3.4.1.3 Helical

Figure 3.9 Helical Type
3.4.2 Distance Reading

There are three basic types of distant reading thermometers.
- Liquid filled
- Gas filled
- Combination liquid-vapor filled
The thermometers are filled with fluid at some temperature and sealed.
Almost the entire volume of the fluid is in the sensing bulb.

Figure 3.10 Distance Reading Type
Oil and Gas Measuring Instruments
31
3.5 Level Sensing Elements

Figure 3.11 Installation

3.5.1 Transparent Glass

Sight Glasses for Level Gauges grant the best chemical and
physical properties, holding a very precise place as for chemical
composition within the very large group of "Borosilicate Glass" which is
suitable for many applications.

Figure 3.12 Level Glass
3.5.2 Circular Sight Ports
These are used to allow observation within sealed vessels.

Figure 3.13 Dight Port
Oil and Gas Measuring Instruments
32
3.5.3 Reflex Type

Reflex level gauges working principle is based on the light
refraction and reflection laws. Reflex level gauges use glasses having the
face fitted towards the chamber shaped to have prismatic grooves with
section angle of 90. When in operation, the chamber is filled with liquid
in the lower zone and gases or vapors in the upper zone; the liquid level is
distinguished by different brightness of the glass in the liquid and in the
gas/vapor zone. The reflex level gauges do not need a specific
illumination: the day environmental light is enough. Only during the
night an artificial light must be provided.






Figure 3.14 Reflex Type
3.5.4 Bicolor Type

An illuminator with special red and a green filters is fitted on the
gauge at the opposite side with respect to the observer. This special
illuminator conveys light through the filters obliquely to the back glasses
of the level gauge. Said filters allow crossing only to red and green rays.
Such colored rays reach, through the back glass, the media inside level
body. When the gauge contains steam, green rays are considerably
deviated and prevented from emerging by the observer side; then only red
light, whose rays are smoothly deviated by steam, passes through the
whole internal hole, reaching the observer. Conversely when rays find
Oil and Gas Measuring Instruments
33
water, red rays are considerably deviated and lost inside the internal part
of level gauge, green rays can reach the front glass and seen by the
observer.






Figure 3.15 Bicolor Type

3.5.5 Magnetic Type

Operation of BONT Magnetic Level Gauge is based on some
elementary physical principles:
- The principle whereby liquid in communicating vessels is always
at same level;
- Archimedes's principle according to which a body immersed in a
liquid receives a buoyancy equal to the weight of displaced liquid;
- The principle of attraction between North and South poles of two
permanent magnets and that of repulsion between like poles.
o This principle has two applications in the BONT magnetic
level gauge:
first between the magnet in the chamber float and
every single magnet of the indicating scale:
Second between the magnets of the indicating scale.


Oil and Gas Measuring Instruments
34

Figure 3.16 Magnetic Type
3.5.6 Gamma Level Switching

The transmission of gamma radiation through a container is
affected by the level contents. The intensity of the transmitted radiation is
measured and used to activate switches when pre-set intensity levels are
reached.

Figure 3.17 Gamma Rays Type

3.6 Seismic Transducer (Vibration)

A schematic diagram is shown in figure (3.18). The mass is
connected through a spring and damper arrangement to a housing frame.
The housing frame is connected to the source of vibrations to be
Oil and Gas Measuring Instruments
35
measured. The mass has the tendency to remain fixed in its spatial
position so that the vibration motion is registered as a relative
displacement between mass and housing frame. The seismic transducer
may be used in two different modes. A large mass and a soft spring are
suited for displacement mode, while a relatively small mass and a stiff
spring are used for acceleration mode.

Figure 3.18 Seismic Type

3.7 Role Play

Each Trainee should speak thoroughly about:
- Pressure Sensing
- Level Sensing
- Temperature Sensing
- Vibration Switches.







Oil and Gas Measuring Instruments
36
Chapter 4
Electrical Transducers

4.1 Learning objectives

1. Introduce electrical transducers.
2. Understand the theory of operation of different transducers.

4.2 Introduction

In order to measure non-electrical quantities, a detector is used
usually to convert the physical quantity into a displacement. In electrical
transducers the output is different, it is in electrical form. The output
gives the magnitude of the measurand. The electric signal may be current,
voltage or frequency and production of these signals is based upon
electrical effects which may be resistance, capacitance, induction, etc.
A transducer may be defined as a device, which converts energy
from one form to another. In electrical instrumentation, a transducer may
be defined as a device which converts a physical quantity into electrical
signal. Another name of a transducer is pick up.

4.2.1 Advantages of Electrical Transducers

- Amplification and attenuation may be done easily.
- The mass-inertia effects are minimized.
- The effects of friction are minimized.
- Low power level.
- Use of telemetry.
Oil and Gas Measuring Instruments
37
4.2.2 Classification of Transducers
The transducer consists of two closely related parts:
- Detector Element: It is the part that responds to physical
phenomenon.
- Transduction Element: It transforms the output of the sensing
element to an electrical output.
Classification of transducers is as follows:
- Based on Transduction: like piezoelectric, thermoelectric, etc.
- Primary and Secondary: Example, a primary part that transforms
pressure into displacement and secondary part that transforms
displacement into electrical form.
- Passive and Active: Depends on whether the transducer will derive
power from or to the circuit.
- Analog and Digital: Analog continuous form like voltage or digital
form like pulses.
- Transducers and Inverse Transducers: It depends whether the
transducer convert physical quantity to electrical signal or vice
versa.
4.3 Pressure Sensing Elements
4.3.1 Strain Gauges
If a metal conductor is stretched or compressed, its resistance
changes on account of the fact that both length and diameter are changed.
This property is called piezoresistivity.

Figure 4.1 Strain Gauge
Oil and Gas Measuring Instruments
38
4.3.2 Inductive Type

Figure (4.2) shows an arrangement which uses coils to form the
two arms of an AC bridge. The pressure acts on the diaphragm and
disturbs the reluctance of the paths of magnetic flux for both coils.

Figure 4.2 Inductive Type
4.3.3 Capacitive Type

They convert pressure into displacement which changes the
capacitance value by changing the distance between the two parallel
plates of a capacitor.

Figure 4.3 Capacitive Type
Oil and Gas Measuring Instruments
39
4.3.4 Linear Variable differential Transformer

The LVDT is used as secondary transducer for measurement of
pressure. The pressure is converted into displacement which is sensed by
LVDT and converted into a voltage.

Figure 4.4 LVDT

4.3.5 Photoelectric Type

As shown in figure (4.5) the light path is affected by the applied
pressure which in turn affects the quantity of light received by the
photoelectric transducer.

Figure 4.5 Photoelectric Type
Oil and Gas Measuring Instruments
40
4.3.6 Piezoelectric Type

A piezoelectric material is one in which an electric potential
appears across certain surfaces if the dimensions of the crystal are
changed by the application of mechanical force. The potential is produced
by the displacement of charges. The effect is reversible and is known as
the piezoelectric effect.

Figure 4.6 Piezoelectric Type

4.4 Temperature Sensing Elements

Oil and Gas Measuring Instruments
41
4.4.1 Thermocouple

The thermocouple is one of the simplest of all sensors. It consists
of two wires of dissimilar metals joined near the measurement point. The
output is a small voltage measured between the two wires.

Figure 4.7 The thermocouple
While appealingly simple in concept, the theory behind the thermocouple
is subtle, the basics of which need to be understood for the most effective
use of the sensor.

4.4.1.1 Thermocouple theory

A thermocouple circuit has at least two junctions: the measurement
junction and a reference junction. Typically, the reference junction is
created where the two wires connect to the measuring device. This second
junction it is really two junctions: one for each of the two wires, but
because they are assumed to be at the same temperature (isothermal) they
are considered as one (thermal) junction. It is the point where the metals
change - from the thermocouple metals to what ever metals are used in
the measuring device - typically copper.
The output voltage is related to the temperature difference between
the measurement and the reference junctions. This is phenomena is
known as the Seebeck effect. In practice the Seebeck voltage is made up
of two components: the Peltier voltage generated at the junctions, plus the
Thomson voltage generated in the wires by the temperature gradient.
Oil and Gas Measuring Instruments
42

Figure 4.8 Signal generated by temperature gradient
The Peltier voltage is proportional to the temperature of each
junction while the Thomson voltage is proportional to the square of the
temperature difference between the two junctions. It is the Thomson
voltage that accounts for most of the observed voltage and non-linearity
in thermocouple response.
Each thermocouple type has its characteristic Seebeck voltage
curve. The curve is dependent on the metals, their purity, their
homogeneity and their crystal structure. In the case of alloys, the ratio of
constituents and their distribution in the wire is also important. These
potential inhomogeneous characteristics of metal are why thick wire
thermocouples can be more accurate in high temperature applications,
when the thermocouple metals and their impurities become more mobile
by diffusion.

4.4.1.2 The practical considerations of thermocouples

The above theory of thermocouple operation has important
practical implications that are well worth understanding:
1. A third metal may be introduced into a thermocouple circuit and have
no impact, provided that both ends are at the same temperature. This
means that the thermocouple measurement junction may be soldered,
brazed or welded without affecting the thermocouple's calibration, as long
as there is no net temperature gradient along the third metal.
Further, if the measuring circuit metal (usually copper) is different to that
of the thermocouple, then provided the temperature of the two connecting
Oil and Gas Measuring Instruments
43
terminals is the same and known, the reading will not be affected by the
presence of copper.
2. The thermocouple's output is generated by the temperature gradient
along the wires and not at the junctions as is commonly believed.
Therefore it is important that the quality of the wire be maintained where
temperature gradients exists. Wire quality can be compromised by
contamination from its operating environment and the insulating material.
For temperatures below 400C, contamination of insulated wires is
generally not a problem. At temperatures above 1000C, the choice of
insulation and sheath materials, as well as the wire thickness, become
critical to the calibration stability of the thermocouple.
The fact that a thermocouple's output is not generated at the junction
should redirect attention to other potential problem areas.
3. The voltage generated by a thermocouple is a function of the
temperature difference between the measurement and reference junctions.
Traditionally the reference junction was held at 0C by an ice bath:

Figure 4.9 Traditional Thermocouple Measurement
The ice bath is now considered impractical and is replaced by a reference
junction compensation arrangement. This can be accomplished by
measuring the reference junction temperature with an alternate
temperature sensor (typically an RTD or thermistor) and applying a
correcting voltage to the measured thermocouple voltage before scaling to
temperature.
Oil and Gas Measuring Instruments
44

Figure 4.10 Modern Thermocouple Measurement
The correction can be done electrically in hardware or mathematically in
software. The software method is preferred as it is universal to all
thermocouple types (provided the characteristics are known) and it allows
for the correction of the small non-linearity over the reference
temperature range.
4. The low-level output from thermocouples (typically 50mV full scale)
requires that care be taken to avoid electrical interference from motors,
power cable and transformers. Twisting the thermocouple wire pair (say 1
twist per 10 cm) can greatly reduce magnetic field pickup. Using shielded
cable or running wires in metal conduit can reduce electric field pickup.
The measuring device should provide signal filtering, either in hardware
or by software, with strong rejection of the line frequency (50/60 Hz) and
its harmonics.
5. The operating environment of the thermocouple needs to be
considered. Exposure to oxidizing or reducing atmospheres at high
temperature can significantly degrade some thermocouples.
Thermocouples containing rhodium (B, R and S types) are not suitable
under neutron radiation.

4.4.1.3 The advantages and disadvantages of thermocouples

Because of their physical characteristics, thermocouples are the
preferred method of temperature measurement in many applications.
They can be very rugged, are immune to shock and vibration, are useful
Oil and Gas Measuring Instruments
45
over a wide temperature range, are simple to manufactured, require no
excitation power, there is no self heating and they can be made very
small. No other temperature sensor provides this degree of versatility.
Thermocouples are wonderful sensors to experiment with because of their
robustness, wide temperature range and unique properties.
On the down side, the thermocouple produces a relative low output signal
that is non-linear. These characteristics require a sensitive and stable
measuring device that is able provide reference junction compensation
and linearization. Also the low signal level demands that a higher level of
care be taken when installing to minimize potential noise sources.
The measuring hardware requires good noise rejection capability. Ground
loops can be a problem with non-isolated systems, unless the common
mode range and rejection is adequate.

4.4.1.4 Types of thermocouple

About 13 'standard' thermocouple types are commonly used. Eight
have been given an internationally recognized type designator. Some of
the non-recognized thermocouples may excel in particular niche
applications and have gained a degree of acceptance for this reason, as
well as due to effective marketing by the alloy manufacturer.

Each thermocouple type has characteristics that can be matched to
applications. Industry generally prefers K and N types because of their
suitability to high temperatures, while others often prefer the T type due
to its sensitivity, low cost and ease of use.
A table of standard thermocouple types is presented below. The table also
shows the temperature range for extension grade wire in brackets.

Oil and Gas Measuring Instruments
46
Type
Positive
Material
Negative
Material
Accuracy***
Class 2
Range C
(extension)
Comments
B Pt, 30%Rh Pt, 6%Rh
0.5%
>800C
50 to 1820
(1 to 100)
Good at high temperatures,
no reference junction
compensation required.
C** W, 5%Re W, 26%Re
1%
>425C
0 to 2315
(0 to 870)
Very high temperature use,
brittle
D** W, 3%Re W, 25%Re
1%
>425C
0 to 2315
(0 to 260)
Very high temperature use,
brittle
E Ni, 10%Cr Cu, 45%Ni 0.5% or 1.7C
-270 to 1000
(0 to 200)
General purpose, low and
medium temperatures
G** W W, 26%Re
1%
>425C
0 to 2315
(0 to 260)
Very high temperature use,
brittle
J Fe Cu, 45%Ni 0.75% or 2.2C
-210 to 1200
(0 to 200)
High temperature, reducing
environment
K* Ni, 10%Cr
Ni, 2%Al
2%Mn
1%Si
0.75% or 2.2C
-270 to 1372
(0 to 80)
General purpose high
temperature, oxidizing
environment
M** Ni Ni, 18%Mo 0.75% or 2.2C -50 to 1410 .
N*
Ni, 14%Cr
1.5%Si
Ni,
4.5%Si
0.1%Mg
0.75% or 2.2C
-270 to 1300
(0 to 200)
Relatively new type as a
superior replacement for K
Type.
P** Platinel II Platinel II 1.0% 0 to 1395
A more stable but
expensive substitute for K
& N types
R Pt, 13%Rh Pt 0.25% or 1.5C
-50 to 1768
(0 to 50)
Precision, high temperature
S Pt, 10%Rh Pt 0.25% or 1.5C
-50 to 1768
(0 to 50)
Precision, high temperature
T* Cu Cu, 45%Ni 0.75% or 1.0C
-270 to 400
(-60 to 100)
Good general purpose, low
temperature, tolerant to
moisture.
* Most commonly used thermocouple types, ** Not ANSI recognized types. *** See IEC 584-2 for more details.
Materials codes:- Al = Aluminum, Cr = Chromium, Cu = Copper, Mg = Magnesium, Mo = Molybdenum, Ni =
Nickel, Pt = Platinum, Re = Rhenium, Rh = Rhodium, Si = Silicon, W = Tungsten



Oil and Gas Measuring Instruments
47
4.4.1.5 Accuracy of thermocouples

Thermocouples will function over a wide temperature range - from
near absolute zero to their melting point, however they are normally only
characterized over their stable range. Thermocouple accuracy is a
difficult subject due to a range of factors. In principal and in practice a
thermocouple can achieve excellent results (that is, significantly better
than the above table indicates) if calibrated, used well below its nominal
upper temperature limit and if protected from harsh atmospheres. At
higher temperatures it is often better to use a heavier gauge of wire in
order to maintain stability.

As mentioned previously, the temperature and voltage scales were
redefined in 1990. The eight main thermocouple types - B, E, J, K, N, R,
S and T - were re-characterized in 1993 to reflect the scale changes. (See:
NIST Monograph 175 for details). The remaining types: C, D, G, M and
P appear to have been informally re-characterized.

4.4.1.6 Thermocouple wire grades

There are different grades of thermocouple wire. The principal
divisions are between measurement grades and extension grades. The
measurement grade has the highest purity and should be used where the
temperature gradient is significant. The standard measurement grade
(Class 2) is most commonly used. Special measurement grades (Class 1)
are available with accuracy about twice the standard measurement grades.
The extension thermocouple wire grades are designed for connecting the
thermocouple to the measuring device. The extension wire may be of
different metals to the measurement grade, but are chosen to have a
Oil and Gas Measuring Instruments
48
matching response over a much reduced temperature range - typically -
40C to 120C. The reason for using extension wire is reduced cost - they
can be 20% to 30% of the cost of equivalent measurement grades. Further
cost savings are possible by using thinner gauge extension wire and a
lower temperature rated insulation.

Note: When temperatures within the extension wire's rating are being
measured, it is OK to use the extension wire for the entire circuit. This is
frequently done with T type extension wire, which is accurate over the -
60 to 100C range.

4.4.1.7 Thermocouple wire gauge

At high temperatures, thermocouple wire can under go irreversible
changes in the form of modified crystal structure, selective migration of
alloy components and chemical changes originating from the surface
metal reacting to the surrounding environment. With some types,
mechanical stress and cycling can also induce changes.

Increasing the diameter of the wire where it is exposed to the high
temperatures can reduce the impact of these effects.

The following table can be used as a very approximate guide to wire
gauge:





Oil and Gas Measuring Instruments
49
Type
8 Gauge
4.06mm
16 Gauge
1.63mm
20 Gauge
0.91mm
24 Gauge
0.56mm
28 Gauge
0.38mm
30 Gauge
0.32mm
B 1820 - - 1700 1700 -
C 2315 2315 2315 2315 2315 -
D 2315 2315 2315 2315 2000 -
E 870 620 540 430 400 370
G 2315 2315 2315 2315 2315 -
J 760 560 480 370 370 320
K 1260* 1000* 980 870 820 760
M 1260* 1200* - - - -
N 1260* 1000* 980 870 820 760
P 1395 - 1250 1250 1250 -
R 1760 - - 1480 1480 -
S 1760 - - 1480 1480 -
T 400 370 260 200 200 150
* Upper temperature limits only apply in a protective sheath

At these higher temperatures, the thermocouple wire should be
protected as much as possible from hostile gases. Reducing or oxidizing
gases can corrode some thermocouple wire very quickly. Remember, the
purity of the thermocouple wire is most important where the temperature
gradients are greatest. It is with this part of the thermocouple wiring
where the most care must be taken.

Other sources of wire contamination include the mineral packing
material and the protective metal sheath. Metallic vapor diffusion can be
significant problem at high temperatures. Platinum wires should only be
used inside a nonmetallic sheath, such as high-purity alumna.
Oil and Gas Measuring Instruments
50
High temperature measurement is very difficult in some situations. In
preference, use non-contact methods. However this is not always
possible, as the site of temperature measurement is not always visible to
these types of sensors.

4.4.1.8 Color coding of thermocouple wire

The color coding of thermocouple wire is something of a
nightmare! There are at least seven different standards. There are some
inconsistencies between standards, which seem to have been designed to
confuse. For example the color red in the USA standard is always used
for the negative lead, while in German and Japanese standards it is always
the positive lead. The British, French and International standards avoid
the use of red entirely!

4.4.1.9 Thermocouple mounting

There are four common ways in which thermocouples are mounted
with in a stainless steel or Inconel sheath and electrically insulated with
mineral oxides. Each of the methods has its advantages and
disadvantages.
Oil and Gas Measuring Instruments
51

Figure 4.11 Thermocouple Sheath Options

- Sealed and Isolated from Sheath: Good relatively trouble-free
arrangement. The principal reason for not using this arrangement
for all applications is its sluggish response time - the typical time
constant is 75 seconds
- Sealed and Grounded to Sheath: Can cause ground loops and
other noise injection, but provides a reasonable time constant (40
seconds) and a sealed enclosure.
- Exposed Bead: Faster response time constant (typically 15
seconds), but lacks mechanical and chemical protection, and
electrical isolation from material being measured. The porous
insulating mineral oxides must be sealed
- Exposed Fast Response: Fastest response time constant (typically
2 seconds), depending on the gauge of junction wire. In addition to
problems of the exposed bead type, the protruding and light
construction makes the thermocouple more prone to physical
damage.

Oil and Gas Measuring Instruments
52
4.4.1.10 Conversion Table

ITS-90 Table for type J thermocouple
Thermoelectric Voltage in mV
C 0 1 2 3 4 5 6 7 8 9 10
0 0.000 0.050 0.101 0.151 0.202 0.253 0.303 0.354 0.405 0.456 0.507
10 0.507 0.558 0.609 0.660 0.711 0.762 0.814 0.865 0.916 0.968 1.019
20 1.019 1.071 1.122 1.174 1.226 1.277 1.329 1.381 1.433 1.485 1.537
30 1.537 1.589 1.641 1.693 1.745 1.797 1.849 1.902 1.954 2.006 2.059
40 2.059 2.111 2.164 2.216 2.269 2.322 2.374 2.427 2.480 2.532 2.585
50 2.585 2.638 2.691 2.744 2.797 2.850 2.903 2.956 3.009 3.062 3.116
60 3.116 3.169 3.222 3.275 3.329 3.382 3.436 3.489 3.543 3.596 3.650
70 3.650 3.703 3.757 3.810 3.864 3.918 3.971 4.025 4.079 4.133 4.187
80 4.187 4.240 4.294 4.348 4.402 4.456 4.510 4.564 4.618 4.672 4.726
90 4.726 4.781 4.835 4.889 4.943 4.997 5.052 5.106 5.160 5.215 5.269
100 5.269 5.323 5.378 5.432 5.487 5.541 5.595 5.650 5.705 5.759 5.814
110 5.814 5.868 5.923 5.977 6.032 6.087 6.141 6.196 6.251 6.306 6.360
120 6.360 6.415 6.470 6.525 6.579 6.634 6.689 6.744 6.799 6.854 6.909
130 6.909 6.964 7.019 7.074 7.129 7.184 7.239 7.294 7.349 7.404 7.459
140 7.459 7.514 7.569 7.624 7.679 7.734 7.789 7.844 7.900 7.955 8.010

4.4.2 RTD

Resistance Temperature Detectors (RTDs) rely on the predictable
and repeatable phenomena of the electrical resistance of metals changing
with temperature.
The temperature coefficient for all pure metals is of the same order
- 0.003 to 0.007 ohms/ohm/C. The most common metals used for
temperature sensing are platinum, nickel, copper and molybdenum. While
the resistance - temperature characteristics of certain semiconductor and
Oil and Gas Measuring Instruments
53
ceramic materials are used for temperature sensing, such sensors are
generally not classified as RTDs.

4.4.2.1 How are RTD constructed?

RTDs are manufactured in two ways: using wire or film. Wire
RTDs are a stretched coil of fine wire placed in a ceramic tube that
supports and protects the wire. The wire may be bonded to the ceramic
using a glaze. The wire types are generally the more accurate, due to the
tighter control over metal purity and less strain related errors. They are
also more expensive.

Figure 4.12 RTD

Film RTDs consist of a thin metal film that is silk-screened or
vacuum spluttered onto a ceramic or glassy substrate. A laser trimmer
then trims the RTD to its correct resistance value.

Film sensors are less accurate than wire types, but they are
relatively inexpensive, they are available in small sizes and they are more
robust. Film RTDs can also function as a strain gauge - so don't strain
them! The alumina element should be supported by grease or a light
elastomer, but never embedded in epoxy or mechanically clamped
between hard surfaces.
Oil and Gas Measuring Instruments
54

Figure 4.13 Typical Sheath Mounted RTD Probe

RTDs cannot generally be used in their basic sensing element form,
as they are too delicate. They are usually built into some type of
assembly, which will enable them to withstand the various environmental
conditions to which they will be exposed when used. Most commonly this
is a stainless steel tube with a heat conducting grease (that also dampens
vibration). Standard tube diameters include 3, 4.5, 6, 8, 10, 12 and 15 mm
and standard tube lengths include 250, 300, 500, 750 and 1000 mm.

4.4.2.2 Characteristics of RTDs

Metal RTDs have a response defined by a polynomial:
R(t) = R
0
( 1 + a.t + b.t
2
+ c.t
3
)
Where R
0
is the resistance at 0C, "t" in the temperature in Celsius, and
"a", "b" and "c" are constants dependent on the characteristics of the
metal. In practice this equation is a close but not perfect fit for most
RTDs, so slight modifications are often be made.
Commonly, the temperature characteristics of an RTD are specified
as a single number (the "alpha"), representing the average temperature
coefficient over the 0 to 100C temperature range as calculated by:
Oil and Gas Measuring Instruments
55
alpha = ( R
100
- R
0
) / 100 . R
0
in ohms/ohm/C
Note: RTDs cover a sufficient temperature range that their response needs
to be calibrated in terms of the latest temperature scale ITS90.
It is also of interest to note that the temperature coefficient of an
alloy is frequently very different from that of the constituent metals.
Small traces of impurities can greatly change the temperature
coefficients. Sometimes trace "impurities" are deliberately added so as to
swamp the effects of undesired impurities which are uneconomic to
remove. Other alloys can be tailored for particular temperature
characteristics. For example, an alloy of 84% copper, 12% Manganese
and 4% Nickel has the property of having an almost zero response to
temperature. The alloy is used for the manufacture of precision resistors.

4.4.2.3 Types RTDs

While almost any metal may be used for RTD manufacture, in
practice the number used is limited.

Metal
Temperature
Range
Alpha Comments
Copper Pt -200C to 260C 0.00427 Low cost
Molybdenum Mo -200C to 200C
0.00300
0.00385
Lower cost alternative to platinum in the
lower temperature ranges
Nickel Ni -80C to 260C 0.00672 Low cost, limited temperature range
Nickel - Iron
Ni-
Fe
-200C to 200C 0.00518 Low cost
Platinum Pt -240C to 660C
0.00385
0.00392
Good precision


Oil and Gas Measuring Instruments
56
4.4.2.4 Platinum RTDs

Platinum is by far the most common RTD material, primarily
because of its long-term stability in air. There are two standard Platinum
sensor types, each with a different doping level of 'impurities'. To a large
extent there has been a convergence in platinum RTD standards, with
most national standards bodies adopting the international IEC751-1983,
with amendment 1 in 1986 and amendment 2 in 1995. The USA
continues to maintain its own standard.
All the platinum standards use a modified polynomial known as the
Callendar - Van Dusen equation:
R(t) = R
0
( 1 + a.t + b.t
2
+ c.(t - 100).t
3
)
Platinum RTDs are available with two temperature coefficients or alphas
- the choice is largely based on the national preference in you country, as
indicated in the following table:
Standard
Alpha
ohms/ohm/C
R
0

ohms
Polynomial Coefficients
IEC751
(Pt100)
0.00385055 100
200C < t < 0C
a = 3.90830x10
-3

b = -5.77500x10
-7

c = -4.18301x10
-12

0C < t < 850C
a & b as above, but
c = 0.0
SAMA
RC-4
0.0039200 98.129
a = 3.97869x10
-3

b = -5.86863x10
-7

c = -4.16696x10
-12


The international IEC 751 standard specifies tolerance classes as
indicated in the following table. While only Classes A and B are defined
in IEC 751, it has become common practice to extended the Classes to C
Oil and Gas Measuring Instruments
57
and D, which roughly double the previous error tolerance. The tolerance
classes are often applied to other RTD types.
Tolerance Class Tolerance Equation (C)
Class A ( 0.15 + 0.002.| t | )
Class B ( 0.30 + 0.005. | t | )
Class C ( 0.40 + 0.009. | t | )
Class D ( 0.60 + 0.0018. | t | )
Where | t | indicated the magnitude of the temperature in Celsius (that is
sign is dropped). Some manufacturers further subdivide their RTD
Tolerance Classes into Tolerance Bands for greater choice in price
performance ratios.
4.4.2.6 Characteristics of Platinum RTDs
The IEC751 specifies a number of other characteristics - insulation
resistance, environmental protection, maximum thermoelectric effect,
vibration tolerance, lead marking and sensor marking. Some of these are
discussed below:
Thermoelectric Effect: Platinum RTD generally employs two metals -
the platinum sensing element and copper lead wires, making it a good
candidate for a thermocouple. If a temperature gradient is allows to
develop along the sensing element, a thermoelectric voltage with a
magnitude of about 7 V /C will be generated. This is only likely to be a
problem with very high-precision measurements operating at low
excitation currents.
Wiring Configurations and Lead Marking: There are three wiring
configurations that can be used for measuring resistance - 2, 3 and 4 wire
connections.
Oil and Gas Measuring Instruments
58

Figure 4.14 Wiring configurations
IEC751 requires that wires connected to the same end of the resistor be
the same colour - either red or white, and that the wires at each end be
different.

4.4.3 Thermistor
Thermistor temperature sensors are constructed from sintered metal
oxide in a ceramic matrix that changes electrical resistance with
temperature. They are sensitive but highly non-linear. Their sensitivity,
reliability, ruggedness and ease of use, has made them popular in research
application, but they are less commonly applied to industrial applications,
probably due to a lack on interchangeability between manufactures.
Thermistors are available in large range of sizes and base resistance
values (resistance at 25C). Interchangeability is possible to 0.05C
although 1C is more common.
4.4.3.1 Thermistor construction
The most common form of the thermistor is a bead with two wires
attached. The bead diameter can range from about 0.5mm (0.02") to 5mm
(0.2'').

Figure 4.15Themistor
Oil and Gas Measuring Instruments
59
Mechanically the thermistor is simple and strong, providing the
basis for a high reliability sensor. The most likely failure mode is for the
lead to separate from the body of the thermistor - an unlikely event if the
sensor is mounted securely and with regard to likely vibration. The
sintered metal oxide material is prone to damage by moisture, so is
passivated by glass or epoxy encapsulation. If the encapsulation is
compromised and moisture penetrates, silver migration under the dc bias
can eventually cause shorting between the electrodes.

Like other temperature sensors, thermistors are often mounted in
stainless steel tubes, to protect them from the environment in which they
are to operate. Grease is typically used to improve the thermal contact
between the sensor and the tube.

4.4.3.2 Thermistor characteristics

The following are typical characteristic for the popular 44004
thermistor from YSI:

Parameter Specification
Resistance at 25C 2252 ohms (100 to 1M available)
Measurement range -80 to +120C typical (250C max.)
Interchangeability (tolerance) 0.1 or 0.2C
Stability over 12 months < 0.02C at 25C, < 0.25C at 100C
Time constant < 1.0 seconds in oil, < 60 seconds in still air
self-heating 0.13 C/mW in oil, 1.0 C/mW in air
Coefficients
(see Linearization below)
a = 1.4733 x 10
-3
, b = 2.372 x 10
-3
, c = 1.074 x 10
-7

Dimensions ellipsoid bead 2.5mm x 4mm

Oil and Gas Measuring Instruments
60
4.4.4 Semiconductor

The semiconductor (or IC for integrated circuit) temperature sensor
is an electronic device fabricated in a similar way to other modern
electronic semiconductor components such as microprocessors. Typically
hundreds or thousands of devices are formed on thin silicon wafers.
Before the wafer is scribed and cut into individual chips, they are usually
laser trimmed. Semiconductor temperature sensors are available from a
number of manufacturers. There are no generic types as with
thermocouple and RTDs, although a number of devices are made by more
than one manufacturer. The AD590 and the LM35 have traditionally been
the most popular devices, but over the last few years better alternatives
have become available.
These sensors share a number of characteristics - linear outputs,
relatively small size, limited temperature range (-40 to +120C typical),
low cost, good accuracy if calibrated but also poor interchangeability.
Often the semiconductor temperature sensors are not well designed
thermally, with the semiconductor chip not always in good thermal
contact with an outside surface. Some devices are inclined to oscillate
unless precautions are taken. Provided the limitations of the
semiconductor temperature sensors are understood, they can be used
effectively in many applications. The most popular semiconductor
temperature sensors are based on the fundamental temperature and
current characteristics of the transistor. If two identical transistors are
operated at different but constant collector current densities, then the
difference in their base-emitter voltages is proportional to the absolute
temperature of the transistors. This voltage difference is then converted to
a single ended voltage or a current. An offset may be applied to convert
the signal from absolute temperature to Celsius or Fahrenheit.
Oil and Gas Measuring Instruments
61
In general, the semiconductor temperature sensor is best suited for
embedded applications - that is, for use within equipment. This is because
they tend to be electrically and mechanically more delicate than most
other temperature sensor types. However they do have legitimate
application in many areas, hence their inclusion.

4.5 Level Sensing Elements
4.5.1 Radar Tank Gauging

Figure 4.16 RTG
FMCW radar principle and FFT signal analysis, (FMCW =
frequency-modulated continuous wave). A radar signal is emitted from an
antenna, reflected from the target (in this case, the product surface) and
received back after a delay interval t. The distance of the reflecting
product surface is measured by way of the transit time t of the microwave
signal: for every meter from a target the waves travel a distance of 2 m,
Oil and Gas Measuring Instruments
62
for which they require a time of approx. 6.7 ns. In general, the measured
distance is a = c x t / 2; where c = the speed of light.
The FMCW radar system uses a linear frequency-modulated high-
frequency signal; transmission frequency increases linearly within a time
interval (frequency sweep). Since the transmission frequency changes due
to the time delay during signal propagation, a low-frequency signal
(typically, up to a few kHz), the frequency f of which is proportional to
the reflector distance a, is obtained from the difference between the
current transmission frequency and the received frequency. The product
level is then computed from the difference between tank height and
distance.

Figure 4.17 RTG Signalling
Oil and Gas Measuring Instruments
63
4.5.2 Vibrating Fork

A piezoelectric crystal operated Vibrating Fork type level switch
for detection of level of powders / granules / solids in the hoppers, bins
and silos, etc.

Figure 4.18 Vibrating fork
4.5.3 LVDT

The letters LVDT are an acronym for Linear Variable
Differential Transformer, a common type of electromechanical
transducer that can convert the rectilinear motion of an object to which it
is coupled mechanically into a corresponding electrical signal. LVDT
linear position sensors are readily available that can measure movements
as small as a few millionths of an inch up to several inches, but are also
capable of measuring positions up to 20 inches (0.5 m).

Figure 4.19 LVDT Core
Oil and Gas Measuring Instruments
64
The figure (4.19) shows the components of a typical LVDT. The
transformer's internal structure consists of a primary winding centered
between a pair of identically wound secondary windings, symmetrically
spaced about the primary. The coils are wound on a one-piece hollow
form of thermally stable glass reinforced polymer, encapsulated against
moisture, wrapped in a high permeability magnetic shield, and then
secured in cylindrical stainless steel housing. This coil assembly is
usually the stationary element of the position sensor.
The moving element of an LVDT is a separate tubular armature of
magnetically permeable material called the core, which is free to move
axially within the coil's hollow bore, and mechanically coupled to the
object whose position is being measured. This bore is typically large
enough to provide substantial radial clearance between the core and bore,
with no physical contact between it and the coil.
In operation, the LVDT's primary winding is energized by alternating
current of appropriate amplitude and frequency, known as the primary
excitation. The LVDT's electrical output signal is the differential AC
voltage between the two secondary windings, which varies with the axial
position of the core within the LVDT coil. Usually this AC output voltage
is converted by suitable electronic circuitry to high level DC voltage or
current that is more convenient to use.

4.5.3.1 Advantages

LVDTs have certain significant features and benefits, most of
which derive from its fundamental physical principles of operation or
from the materials and techniques used in its construction.
- Friction-Free Operation
Oil and Gas Measuring Instruments
65
One of the most important features of an LVDT is its friction-free
operation. In normal use, there is no mechanical contact between the
LVDT's core and coil assembly, so there is no rubbing, dragging or other
source of friction. This feature is particularly useful in materials testing,
vibration displacement measurements, and high resolution dimensional
gauging systems.
- Infinite Resolution
Since an LVDT operates on electromagnetic coupling principles in a
friction-free structure, it can measure infinitesimally small changes in
core position. This infinite resolution capability is limited only by the
noise in an LVDT signal conditioner and the output display's resolution.
These same factors also give an LVDT its outstanding repeatability.
- Unlimited Mechanical Life
Because there is normally no contact between the LVDT's core and coil
structure, no parts can rub together or wear out. This means that an LVDT
features unlimited mechanical life. This factor is especially important in
high reliability applications such as aircraft, satellites and space vehicles,
and nuclear installations. It is also highly desirable in many industrial
process control and factory automation systems.
- Over travel Damage Resistant
The internal bore of most LVDTs is open at both ends. In the event of
unanticipated over travel, the core is able to pass completely through the
sensor coil assembly without causing damage. This invulnerability to
position input overload makes an LVDT the ideal sensor for applications
like extensometers that are attached to tensile test samples in destructive
materials testing apparatus.
- Single Axis Sensitivity
An LVDT responds to motion of the core along the coil's axis, but is
generally insensitive to cross-axis motion of the core or to its radial
Oil and Gas Measuring Instruments
66
position. Thus, an LVDT can usually function without adverse effect in
applications involving misaligned or floating moving members, and in
cases where the core doesn't travel in a precisely straight line.
- Separable Coil And Core
Because the only interaction between an LVDT's core and coil is
magnetic coupling, the coil assembly can be isolated from the core by
inserting a non-magnetic tube between the core and the bore. By doing
so, a pressurized fluid can be contained within the tube, in which the core
is free to move, while the coil assembly is depressurized. This feature is
often utilized in LVDTs used for spool position feedback in hydraulic
proportional and/or servo valves.
- Environmentally Robust
The materials and construction techniques used in assembling an LVDT
result in a rugged, durable sensor that is robust to a variety of
environmental conditions. Bonding of the windings is followed by epoxy
encapsulation into the case, resulting in superior moisture and humidity
resistance, as well as the capability to take substantial shock loads and
high vibration levels in all axes. And the internal high-permeability
magnetic shield minimizes the effects of external AC fields.
Both the case and core are made of corrosion resistant metals, with the
case also acting as a supplemental magnetic shield. And for those
applications where the sensor must withstand exposure to flammable or
corrosive vapors and liquids, or operate in pressurized fluid, the case and
coil assembly can be hermetically sealed using a variety of welding
processes.
Ordinary LVDTs can operate over a very wide temperature range, but, if
required, they can be produced to operate down to cryogenic
temperatures, or, using special materials, operate at the elevated
temperatures and radiation levels found in many nuclear reactors.
Oil and Gas Measuring Instruments
67
- Null Point Repeatability
The location of an LVDT's intrinsic null point is extremely stable and
repeatable, even over its very wide operating temperature range. This
makes an LVDT perform well as a null position sensor in closed-loop
control systems and high-performance servo balance instruments.
- Fast Dynamic Response
The absence of friction during ordinary operation permits an LVDT to
respond very fast to changes in core position. The dynamic response of an
LVDT sensor itself is limited only by the inertial effects of the core's
slight mass. More often, the response of an LVDT sensing system is
determined by characteristics of the signal conditioner.
- Absolute Output
An LVDT is an absolute output device, as opposed to an incremental
output device. This means that in the event of loss of power, the position
data being sent from the LVDT will not be lost. When the measuring
system is restarted, the LVDT's output value will be the same as it was
before the power failure occurred.

4.5.3.2 Theory of Operation

This figure illustrates what happens when the LVDT's core is in
different axial positions. The LVDT's primary winding, P, is energized by
a constant amplitude AC source. The magnetic flux thus developed is
coupled by the core to the adjacent secondary windings, S1 and S2 . If the
core is located midway between S1 and S2 , equal flux is coupled to each
secondary so the voltages, E1 and E2 , induced in windings S1 and S2
respectively, are equal. At this reference midway core position, known as
the null point, the differential voltage output, (E1 - E2), is essentially
zero.
Oil and Gas Measuring Instruments
68

Figure 4.20 LVDT Signalling

If the core is moved closer to S1 than to S2 , more flux is coupled to S1
and less to S2 , so the induced voltage E1 is increased while E2 is
decreased, resulting in the differential voltage (E1 - E2). Conversely, if
the core is moved closer to S2 , more flux is coupled to S2 and less to S1 ,
so E2 is increased as E1 is decreased, resulting in the differential voltage
(E2 - E1 ). The top graph shows how the magnitude of the differential
output voltage, EOUT, varies with core position. The value of EOUT at
maximum core displacement from null depends upon the amplitude of the
primary excitation voltage and the sensitivity factor of the particular
LVDT, but is typically several volts RMS. The phase angle of this AC
output voltage, EOUT, referenced to the primary excitation voltage, stays
constant until the center of the core passes the null point, where the phase
angle changes abruptly by 180 degrees, as shown in the middle graph.
This 180 degree phase shift can be used to determine the direction of the
core from the null point by means of appropriate circuitry. This is shown
in the bottom graph, where the polarity of the output signal represents the
core's positional relationship to the null point. The figure shows also that
Oil and Gas Measuring Instruments
69
the output of an LVDT is very linear over its specified range of core
motion, but that the sensor can be used over an extended range with some
reduction in output linearity. The output characteristics of an LVDT vary
with different positions of the core. Full range output is a large signal,
typically a volt or more, and often requires no amplification. Note that an
LVDT continues to operate beyond 100% of full range, but with degraded
linearity.

4.5.4 Servo Motor

A micro-controller based multi-function instrument for precision
level measurement of liquids stored in Cone Roof, Floating Roof tanks,
pressurized Spheres, Mounded Vessels, Bullets and Cryogenic storage
tanks.

Figure 4.21 Servo-motor Type

4.5.5 Pressure Sensing Type

In this type of level gauging, the pressure or differential pressure is
measured converted to level by the following equation.
Oil and Gas Measuring Instruments
70
) (
1 2
h h g P = A
If the tank is open to atmosphere the pressure at the bottom is indication
of level. In closed tanks, differential pressure is the measurand that
indicates the level. The linkage may be direct, liquid filled or sealed
liquid filled.

Figure 4.22 Pressure sensing Type

4.6 Vibration Sensing

4.6.1 Inductive Sensor (Eddy Current)

Inductive sensors use currents induced by magnetic fields to detect
nearby metal objects. The inductive sensor uses a coil (an inductor) to
generate a high frequency magnetic field as shown in Figure 4.23. If there
is a metal object near the changing magnetic field, current will flow in the
object. This resulting current flow sets up a new magnetic field that
opposes the original magnetic field. The net effect is that it changes the
Oil and Gas Measuring Instruments
71
inductance of the coil in the inductive sensor. By measuring the
inductance the sensor can determine when a metal have been brought
nearby. These sensors will detect any metals, when detecting multiple
types of metal multiple sensors are often used.

Figure 4.23 Inductive Sensor
The sensors can detect objects a few centimeters away from the
end. But, the direction to the object can be arbitrary as shown in Figure
4.24. The magnetic field of the unshielded sensor covers a larger volume
around the head of the coil. By adding a shield (a metal jacket around the
sides of the coil) the magnetic field becomes smaller, but also more
directed. Shields will often be available for inductive sensors to improve
their directionality and accuracy.

Figure 4.24 Shielded and Unshielded

Oil and Gas Measuring Instruments
72
4.7 Role Play

Each Trainee should speak thoroughly about one of the electrical
transducers for
- Pressure.
- Temperature.
- Level Gauging and Vibration Sensing.





















Oil and Gas Measuring Instruments
73
Chapter 5
Flow Measurement

5.1 Learning Objectives

1. Review basic properties of fluid flow.
2. To understand the theory of operation of different flow meters.
3. Select the optimum meter according to the application.
4. To avoid pitfalls in flow metering.

5.2 Basic Principles of Fluid Flow and Measurement

5.2.1 Density and Specific Volume

The density of a fluid is the ratio of its mass to its volume. Its
specific volume is the reciprocal of its density. The density of water is
roughly 1000 times that of air at atmospheric pressure.
V
M
=

5.2.2 Thermal Expansion Coefficient

The thermal expansion coefficient, |, is the fractional increase in
specific volume, V
s
, caused by a temperature increase of 1 degree.
dT
dV
V
s
s
1
= |


Oil and Gas Measuring Instruments
74
5.2.3 Compressibility

The compressibility of a fluid, k, is the fractional decrease in
specific volume caused by unit increase of pressure.
dP
dV
V
s
s
1
= k

5.2.4 Viscosity

The viscosity, q, of a fluid is a measure of its resistance to shearing
at a constant rate.

t
q =
where t is the shear stress and is the rate of shear strain. The SI unit of
viscosity is Pascal second, but it is usual to express it in centipoises, cP,
where one cP being 0.001 Pa s. Viscosity is referred to as absolute or
dynamic viscosity to distinguish it from kinematics viscosity, v, which is
the ratio of viscosity to density. The Si unit of which is m
2
s
-1
and
commonly known by centistokes, cSt, where one cSt being 10
-6
m
2
s
-1
.

5.2.5 Air Solubility of Liquids

Air is soluble in liquids, and its solubility is directly proportional to
the absolute pressure. The solubility decreases markedly as the
temperature of the water increases. It is very much soluble in
hydrocarbons where the solubility is not decreased much with increasing
temperature, until quite high temperatures are reached.


Oil and Gas Measuring Instruments
75
5.2.6 Humidity in gases

Gases may be either dry or humid. This is because a gas at a given
temperature is capable of holding a certain maximum amount of water
vapor; this value increases with temperature increase. The relative
humidity is defined as the ratio of the actual partial pressure of the water
vapor to the value of partial pressure that would exist under saturated
conditions at the same temperature.
Sudden changes in humidity may cause errors in gas flow
measurement. In particular, errors easily occur if unsaturated gas is
passed through a wet gas meter, or if a sudden expansion cools a gas
sufficiently to cause precipitation of some of its water vapor.

5.2.7 Reynolds Number

The behavior of fluids flowing through pipes is governed by a
quantity known as Reynolds number which is defined by
q
vD
D
= Re
where v is the mean velocity and D is the pipe diameter. The numerator is
a measure of the flowing fluid's ability to generate a dynamic forces,
while the denominator is a measure of its ability to generate viscous
forces. This means that Reynolds number indicates which kind of forces
predominate the flowing fluid.

5.2.8 Laminar and Turbulent Flow

Laminar flow occurs at Reynolds numbers below about 2000. This
can be likened to the flow of traffic on a busy motorway, with the traffic
Oil and Gas Measuring Instruments
76
in the various lanes traveling on parallel paths at different speeds.
Turbulent flow occurs at Reynolds number above about 2000.

5.2.9 Rotation and Swirl

Bends, flowmeters, valves, etc., produce what is known as rotation
in the flow. The fluid on the outside of the bend has to travel farther than
the fluid in the inside and this distorts the pattern of the flow in highly
complex fashion. On consequence of this is the rotary motion. The flow
returns to steady flow after some distance. For two adjacent bends in
different planes the flow rotates in three dimensions, i.e. swirls. It takes
longer distance for swirl to come back to steady flow.

Figure 5.1 Rotational Flow

5.2.10 Continuity and Bernoulli's Equation

In simple, what goes at one end of the pipe comes out at the other.
This simple fact is the basis of continuity, which holds that the mass flow
rate is the same at all cross-sections of one continuous pipe having no
branches. If the fluid is incompressible, the volumetric flow rate remains
constant also.
The energy possessed by a flowing fluid is the same at every cross-
section along the pipe. Bernoulli's equation expresses this fact in
mathematical terms.
= +
2
2
1
v P constant at all sections.
Oil and Gas Measuring Instruments
77
5.2.11 Velocity Head
The expression v
2
=2g provides a convenient way of indicating the
amount of kinetic energy possessed by the fluid flowing in the pipe. It has
the dimensions of length and is equal to the head to which the fluid would
rise if it were projected vertically upwards. An important use of this
concept is to express the tendency of pipe fittings to dissipate energy in
terms of velocity heads.

5.2.12 Cavitation
It follows from Bernoulli that when the mean velocity increases the
pressure will decrease. In water, volatile hydrocarbons and liquefied
gases cavitation generally occurs only when the pressure at some point
reaches the vapor pressure of the liquid causing bubbles and vapor
pockets to appear. In viscous oil and non-volatile liquid fuels cavitation
generally takes a different form. It begins at pressures somewhat below
atmospheric, but well above the vapor pressure.

5.2.13 Double Block and Bleed Valve

Flowmeters are frequently installed in complex network of piping
containing a number of shut off valves. To eliminate the bypassed flow a
system of double block and bleed valves are installed to confirm the
operator that the valves are sealing perfectly.

Figure 5.2 Double block and bleed valve.
Oil and Gas Measuring Instruments
78
5.2.14 Definitions

- The mean pipe velocity is related to volumetric flowrate, Q
V
, and
pipe cross-sectional area, A.
vA Q
V
=
- Volumetric flowrate, Q
V
, is defined as the rate of change of
volume.
dt
dV
Q
V
=
- Mass flowrate, Q
M
, is the rate of change of mass with time.
- The results of calibration may be plotted as a graph of flowmeter
readout against flowrate. The graphs may be linear or non-linear.
- A more detailed graph is the performance index, which displays
any small deviations from ideal behavior by the flowmeter.

5.2.15 Factors

- Coefficient of discharge, C, is defined by
I
T
Q
Q
C =
where Q
T
denotes true flowrate and Q
I
denotes the flow indicated
by meter.

Figure 5.3 Coefficient of discharge
Oil and Gas Measuring Instruments
79
- Meter correction factor is defined by
I
I T
V
V V
= A
- Meter factor, F, is used in connection with meters for total volume
and is given by
I
T
V
V
F =
- K-Factor is a term used to describe the performance of meters
whose output is in the form of a series of electrical pulses, and
where total pulse count, n, is nominally proportional to the volume
passed, and the pulse frequency, dn/dt, is nominally proportional to
the flowrate.
T
V
n
K =

Figure 5.4 K-factor

5.3 Differential Pressure Meters
5.3.1 Principle of Operation

The meter depends on the fact that when a fluid flows through a
contraction it must accelerate; this causes its kinetic energy to increase,
and consequently its pressure must fall by a corresponding amount. The
volumetric flowrate is given by
Oil and Gas Measuring Instruments
80
( )
2
1
2
1
1
2
2
2
1
|
|
.
|

\
|

o c P
m
A C
Q
V

where c is an empirical coefficient, the expansibility factor. This depends
upon the physical properties of the gas being metered, as well as the
geometry of the flowmeter.

Figure 5.5 Differential Pressure measurement
5.3.2 Advantages

- Simplicity of construction.
- Versatility: used with almost any fluid.
- Economy.
- Experience.

5.3.3 Disadvantages

- Accuracy is not quite enough.
- The output signal is not linear to flowrate

5.3.4 Selecting the Meter

It is usually not difficult to decide which type is better for a
particular job. The lengthy expensive venturi meter has a low head loss
Oil and Gas Measuring Instruments
81
and its high initial cost is justified in situations where large quantities of
liquids are being pumped, i.e. in main water supply pipelines. Gas plants
where head loss is not important the orifice plates are the decision. A
compromise for intermediate cost and size is the nozzle.

5.3.4.1 Venturi Tubes

The venturi tube is the original form of differential pressure meter.
A typical design is shown in figure (5.6). Because energy losses are low
and flow conditions are not far removed from the ideal, the discharge
coefficient of venturi meters is very near unity.

Figure 5.6 Venturi Tube

5.3.4.2 Orifice Plates

An orifice plate is simply a plate with a hole in it, forming a partial
obstruction to the flow. The flowing fluid follows the same kind of path
as it does in venturi tube. However, the narrowest part of the flow stream
is not in the orifice itself, but some distance downstream; this narrowest
section is known as vena contracta. Between the vena contracta and the
pipe wall, numerous eddies form, which dissipate great deal of kinetic
energy that is responsible for the high head loss.
Oil and Gas Measuring Instruments
82

Figure 5.7 Orifice Plate

Concentric orifice plates are made with a circular orifice concentric
with the pipe. In figure 5.7, the tapings are in the adjacent pipes at
distances shown. Another common arrangement is to put the tapings in
the pipe flanges adjacent to the orifice plate. The position of the tapping
affects the discharge coefficient. The orifice plate in figure (5.8) is
described as square-edged because that is the shape of the upstream
although the downstream edge is chamfered. This is used in clean gases
and clean liquids with low viscosity. With viscous liquids it is necessary
to make the edges raduissed or chamfered upstream and square
downstream and they are called quarter-circle or conical-entry.
Concentric orifice plates cannot be used with dirty fluids because
dirt gradually builds up behind the plate until its performance is impaired.
Instead eccentric or chord orifice plates are commonly used but they are
less accurate.

Figure 5.8 Orifice Plate Types
Oil and Gas Measuring Instruments
83
5.3.4.3 Nozzles

Nozzles are more costly than orifice plates but they have three
advantages over them: they have a discharge coefficient very much closer
to unity; they can be used to discharge directly into the atmosphere; and
they have no sharp edge to blunted, i.e. they can be used with dirty and
abrasive fluids.

Figure 5.9 Nozzles

5.3.5 Points to watch when Using

- Install orifice plates correctly watching the edges back and front.
- Installing differential pressure meters, take care the pressure
tapings in acceptable position. These must never be at the bottom
so that would not clog with dirt. With liquids the tapings must not
be positioned at the top so they would fill with bubbles. The best
place is at the side of the pipe.
- Stay within the recommended range of flowrates.
- Cavitation must not be allowed to occur.
Oil and Gas Measuring Instruments
84
- Make periodic inspections for meter and pipe work to trace any
film of dirt, corrosion or organic growth.
- Inspect sharp edges in the orifice if worn or not.
- When used with wet gases, plates are often provided with drain
hole.

5.3.6 Drag Plate

The principle of the drag plate meter is illustrated in figure (5.10)
A circular plate is supported centrally in the pipe by means of hinged
arm. The flowing fluid produces a positive pressure on the upstream side
of the plate and suction on the downstream. This pressure difference
produces forces which tend to move the plate in the direction of flow, but
this force is resisted by a null-balance supporting element at the end of
the support arm. The signal from the null-balance device is proportional
to the force on the plate which is proportional to the square of the
flowrate.

Figure 5.10 Drag plate

5.3.6.1 Advantages

- Dirt cannot be built up.
- There are no pressure tapings to be blocked.
Oil and Gas Measuring Instruments
85
- The flowrate range can be adjusted by a simple range switch.
5.3.6.2 Disadvantages

- Square root characteristics
- For good accuracy, large diameters are used and hence high head
loss.
- The force on large drag plate would be too great to be supported
effectively by null-balance system.

5.3.6.3 When to Use

The drag plate is suitable for liquids containing suspended solids.

5.3.7 Rotameters

In the simplest type of rotameter the body is a tapered transparent
tube of glass or plastic with a scale engraved on it. Inside the tube is a
small solid body with a circular cross-section, the float, when there is no
flow the float rests at the bottom. Flow causes it ot be lift off. Its very low
price is its advantage. The high head loss is its disadvantage.

Figure 5.11 Rotameters
Oil and Gas Measuring Instruments
86

5.3.8 Spring Loaded Variable Aperture Flowmeters

In the differential pressure flowmeter, the area of constriction is
kept constant and the pressure difference is varying. In variable aperture
flowmeters the reverse effect occurs. In this type two degrees of freedom
are possessed which can be used for meter readout. One degree is for the
pressure difference and the other is for the displacement of the member
controlling the aperture.

Figure 5.12 Spring Loaded Variable Aperture

5.3.8.1 Advantages

- Wide range of operation with tolerable accuracy.
- Linear output.
- Less sensitive for viscosity changes.
- Can be installed horizontal, vertical or inclined.
Oil and Gas Measuring Instruments
87

5.3.8.2 Disadvantages
- Larger in diameter than the pipes.
- Expensive.
- High head loss.

5.3.9 Laminar Flowmeters

In turbulent flow, pressure drop is proportional to the square of the
velocity. In laminar flow it is linear. The simplest laminar flowmeter
consists of fine capillary tube with highly sensitive differential pressure
micro manometer connected across it.

Figure 5.13 Laminar Flowmeters
5.3.9.1 Advantages
- Approximately linear output.
- Wide rangeability.
- No moving parts.
- Used for extremely low flow rates.

5.3.9.2 Disadvantages
- Bulky and expensive.
- Calibration is upset by dust particles.
- Sensitive to changes in viscosity.
Oil and Gas Measuring Instruments
88
5.4 Rotating Mechanical Meters

5.4.1 Positive Displacement Meters

In principle, liquids are measured using containers. This technique
is accomplished in continuous process for positive displacement meters.
In gases the mechanism must have very low frictional resistance.

Figure 5.14 Postive displacement meters
5.4.1.1 Advantages
- High accuracy.
- They are not affected by upstream flow disturbances so they can
be used very close to bends.

5.4.1.2 Disadvantages
- Large sizes.
- High head loss.
- Can be damaged by dirt particles.
- If they clutch they will block flow.

5.4.1.3 Points to Watch
- One direction flow only. So installation should be supervised.
- When used with water check internals for non-corrosive materials.

Oil and Gas Measuring Instruments
89
5.4.2 Turbine Meters

It consists of a short length of pipe in the centre of which there are
two bearings supported by spiders. A propeller is mounted so that it can
spin freely on these central bearings. The propeller materials should be
either magnetic or small magnet is inserted in the tip of each blade and a
pick up is installed on the pipe. Meter readout is pulses.

Figure 5.15 Tyrbine Meters
5.4.2.1 Advantages
- They are very accurate.
- The output is digital.
- Moderate head loss.
- Compact in size.
- If they clutch the flow does not block.

5.4.2.2 Disadvantages
- Expensive.
- Need periodic calibration to compensate for wear up.
- Sensitive to viscosity changes.
- Sensitive to flow disturbance and especially swirl.

Oil and Gas Measuring Instruments
90
5.4.2.3 Points to Watch
- Does the application justify cost?
- Examine calibration curves for suitable accuracy.
- Is it applicable for use with the fluid, temperature and pressure?
- Never blow out the line with compressed air or steam because
over speed would damage it.
- For dirty liquids use coarse filters.
- Any nearby electrical signal might introduce errors in pick ups.
- Avoid cavitation.

5.4.3 Bypass Meters
The flowrate in the bypass is approximately a constant fraction of
the flow in the main pipe. The interesting result is that the relationship
between flowrate and pressure drop in the bypass follows square law and
thus it cancels out the square root effect of the orifice plate itself. The
advantage is economical and the disadvantage is that accuracy and
linearity are inferior to those more expensive meters.

Figure 5.16 Bypass meters
5.4.4 Metering Pumps
A metering pump may be regarded as a combination of pump,
flowmeter and flow regulator. It consists of a piston pump with a variable
stroke, a device counting the number of strokes delivered and a pre-
settable mechanism that will stop the pump when the required number of
strokes has been delivered.
Oil and Gas Measuring Instruments
91
5.5 Other Volumetric Flowmeters
5.5.1 Electromagnetic Flowmeters

It utilizes the same basic principle of electrical generator: when a
conductor moves across a magnetic field a voltage is induced in the
conductor, and the magnitude of the voltage is directly proportional to the
speed of the moving conductor. If the conductor is a section of a
conductive liquid flowing in a non-conductive pipe through a magnetic
field, electrodes are mounted in the pipe wall at the positions shown in
figure (5.17). The voltage induced across the electrodes is proportional to
the flowrate.

Figure 5.17 Electromagnetic Flowmeter
5.5.1.1 Advantages
- There is no obstruction whatever to the flow, suitable for
measuring flow rates of heavy suspensions like mud, sewage and
wood pulp.
- Zero head loss.
- Wide range of meter sizes.
Oil and Gas Measuring Instruments
92
- Not affected with upstream flow.
- Not affected by density or viscosity variation.
- Linear output.
- Bi-directional.

5.5.1.2 Disadvantages
- Fluid must be electrical conductive.
- Not very accurate.
- Not cost effective for small pipe sizes.

5.5.1.3 Things to Watch
- Make sure of whole range of duty.
- Is a built in electrode cleaning device needed?
- How the meter is calibrated?
- If installed below ground level make sure it withstands drowning.
- Never to alter meter duty.
- Never install meter with electrodes in vertical diameter because
they would be affected with air bubbles.
- Check zero reading periodically.
- If the pipe system is electro galvanic corrosion prevention system,
then bonding straps are used to bypass the currents around the
meter.

5.5.2 Ultrasonic Flowmeters

Ultrasonic flowmeters use sound waves to determine the flowrate.
Pulses from a transducer travel through a moving fluid at the speed of
sound and provides an indication of fluid velocity.
Oil and Gas Measuring Instruments
93
The first method uses a transit-time method, in which two opposing
transducers are mounted so that sound waves traveling between them are
at 45 degree angle to the direction of the flow. The speed of sound from
the upstream transducer to the downstream transducer represents the
inherent speed of sound plus a contribution due to fluid velocity. The
opposite direction transducer is used to extract the fluid velocity from
speed of sound. It is essential that the fluid is free of entrained gas or
solids to prevent scattering of sound waves.

Figure 5.18 Ultrasonic meters
The second method uses the Doppler Effect. This type uses two
transducer elements mounted in the same side of the pipe. An ultrasonic
sound wave of constant frequency is transmitted into the fluid by one of
the elements. Solids or bubbles within the fluid reflect the sound back to
the receiver element. The Doppler principle states that there will be a shift
in apparent frequency when there is a relative motion between the
transmitter and receiver. Doppler ultrasonic meters require entrained
gases and suspended solids within the flow.
Ultrasonic meters advantages are freedom of obstruction in the
pipe and negligible cost-sensitivity with respect to pipe diameter. The
disadvantages are that performance is very dependent on flow conditions
and that fair accuracy is attainable when properly applied to appropriate
fluids.
Oil and Gas Measuring Instruments
94
5.5.3 Vortex Shedding Meters

The operating principle is based on the phenomenon of vortex
shedding known as the von Karman effect. As a fluid passes a bluff body,
it separates and generates vortices that are shed alternately along and
behind each side of the bluff body. These vortices cause areas of
fluctuating pressure that are detected by a sensor. The frequency of vortex
generation is directly proportional to fluid velocity. Vortex shedding
meters are aimed the section of market as orifice plates. They have the
same moderate accuracy as orifice plates, similar head and the same
sensitivity to upstream flow disturbances. There is no rotating mechanism
so there is no wear. It scores over orifice plates by having linear output.

Figure 5.19 Vortex shedding meters
5.5.4 Thermal Flowmeters

This type of flowmeters is for mass flowrates. The mass flow rate
is given by
( )
1 2
T T c
H
Q
p
M

=
where H is the power supplied in the form of heat and c
p
is the specific
heat capacity at constant pressure. The main use for this type is with
gases at relatively low pressure and flowrates.
Oil and Gas Measuring Instruments
95

Figure 5.20 Thermal Flowmeter

5.5.5 Coriolis Meters

The Coriolis meter uses an obstruction less U-shaped tube as a
sensor and applies Newton's second law of motion to determine flow rate.
Inside the sensor housing, the sensor tube vibrates at its natural
frequency. The sensor tube is driven by an electromagnetic drive coil
located at the center of the bend in the tube and vibrates similar to that of
a tuning fork.

Figure 5.21 Sensor vibration
The fluid flows into the sensor tube and is forced to take on the
vertical momentum of the vibrating tube. When the tube is moving
upward during half of its vibration cycle, the flowing into the sensor
resists being forced upward pushing down on the tube. The fluid flowing
out of the sensor has an upward momentum from the motion of the tube.
As it travels around the tube bens, the fluid resists changes in its vertical
Oil and Gas Measuring Instruments
96
motion by pushing up on the tube. The difference in forces causes the
tubes to twist. When the tube is moving downward during the second half
of its vibrating cycle, it twists in the opposite direction. This twisting
characteristic is called Coriolis effect.

Figure 5.22 Forces on sensor
Due to Newton's second law of motion, the amount of sensor tube
twist is directly proportional to the mass flowrate of the fluid flowing
through the tube. Electromagnetic velocity detectors located on each side
of the flow tube measure the velocity of the vibrating tube. Mass flow is
determined by measuring the time difference exhibited by the velocity
detector signals. During zero flow conditions, no tube twist occurs,
resulting in no time difference between the two velocity signals. With
flow, a twist occurs with a resulting time difference between the two
velocity signals. This time difference is directly proportional to mass
flow.

Digure 5.23 Sensor Twisting



Oil and Gas Measuring Instruments
97
5.6 Velocity Measuring

5.6.1 Pilot Tubes

It is the oldest and simplest form of fluid meter. The fluid in the
mouth of the tube has been brought to rest, and its kinetic energy has
been converted to pressure energy, which creates an enhanced pressure
inside the pilot tube.
2
2
1
v P o =

Figure 5.24 Pilot Tubes

5.6.2 Hot Resistor Anemometers

The basic principle is an electrically heated element is placed
within the stream flow; the higher velocity the more it tends to cool the
element; the change in temperature causes a change in resistance, which
can be measured by some appropriate circuitry.

5.6.3 Laser Doppler Velocity Meters

The schematic arrangement is shown in figure (25). The laser beam
is first passed into a beam splitting prism, and then the two parallel
Oil and Gas Measuring Instruments
98
component beams are passed through a lens which makes them converge
at a point where the flow velocity is to be measured. Whenever a dirt
particle passes through the bright spot where the two beams intersect, it
reflects light in all directions. This reflected light possesses a Doppler
frequency shift. Some of it is picked up by a collecting lens and focused
on a photo detector which reads out the velocity.

Figure 5.25 Laser Doppler velocity meter
5.7 Two Phase Flow

Wherever possible it is better to separate the gas and liquid phases
and meter each one on its own. If it is not practical, there are some
recognized techniques for measuring two phase flow. You will have to
work very hard to obtain accuracy approaching 10%.

Figure 5.26 Two Phase flow behavior
Oil and Gas Measuring Instruments
99
5.8 Choosing the Right Flowmeter



Oil and Gas Measuring Instruments
100
5.9 Calibrating Flowmeters

Calibration of flowmeters can be done using any of the following
techniques depending on how practical the technique is.
- Volumetric Tank

Figure 5.27 Volumetric calibration
- Weighing

Figure 5.28 Dynamic Weighing calibration
- Master Meter

Figure 5.29 Master meter calibration
5.10 Role Play

Each Trainee should speak thoroughly about one of the learning objective
elements.

Oil and Gas Measuring Instruments
101























Oil and Gas Measuring Instruments
102
Chapter 6
Analyzers

6.1 Learning objectives

1. Understand the theory of operation of oxygen, moisture and gas
chromatography analyzers.

6.2 Oxygen Analyzers

The analyzer uses an electrochemical sensor technology to achieve
the measurement of oxygen. See Figure (6.1). The sensor is a self
contained disposable unit which requires no maintenance. The sensor
utilizes the principle of electrochemical reaction to generate a signal
proportional to the oxygen concentration in the sample.
The sensor consists of a cathode and anode which are in contact via
a suitable electrolyte. The sensor has a gas permeable membrane which
covers the cathode allowing gas to pass into the sensor while preventing
liquid electrolyte from leaking out. As the sample diffuses into the sensor,
any oxygen present will dissolve in the electrolyte solution and migrate to
the surface of the cathode. The oxygen is reduced at the cathode.
Simultaneously, an oxidation reaction is occurring at the anode
generating four electrons. These electrons flow to the cathode to reduce
the oxygen. The representative half cell reactions are:
Cathode:
4e- + 2H2O + O2 4OH-
Anode:
4OH- + 2Pb 2PbO + 2H2O + 4e-
Oil and Gas Measuring Instruments
103
The resultant overall cell reaction is:
2Pb + O2 2PbO
This flow of electrons constitutes an electric current which is
directly proportional to the concentration of oxygen present in the sample.
In the absence of oxygen, no oxidation / reduction reaction occurs and
therefore no current is generated. This allows the sensor to have an
absolute zero.

Figure 6.1 Oxygen analyzer sensing element
6.3 Gas Chromatography

The word Chromatography is at present used as a collective
term for a group of methods that at first sight appear somewhat diverse.
These methods, however, have a number of common features. All
chromatographic separations, for instance, involve the transport of a
sample of a mixture through a column. The mixture may be a liquid or a
vapor. The column contains a substance, the stationary phase, which may
consist of a solid absorbing agent or of a liquid partitioning agent
supported by a solid.
The transport of the constituents of the sample through the column
is affected either by a gas or a liquid, the moving phase. Owing to the
Oil and Gas Measuring Instruments
104
selective retention exerted by the stationary phase, the components of the
mixture move through the column at different effective rates. They, thus
tend to segregate into separate bands or zones. The column is designed to
affect this separation at the exit of the column where the individual bands
may be directed to a detector for determination. The separation obtained
with this principle of operation is easily observed by using a piece of
filter paper and putting a drop of oil in the center of the paper. With time,
the light molecules of the oil will travel through the capillaries of the
filter paper faster and farther than the heavy molecules. These light
molecules are small in size and, generally, have fewer side branches or
arms and legs to cause restriction to flow; and therefore, tend to move
through the capillaries in an easier way than the larger molecules. It is
easy to see the heavy, large, dark molecules of the oil restricted and
retained near the center of the filter paper where the original drop was
placed. Some oils will even show slight color bands as the separation of
molecules occurs while traveling toward the edge of the filter paper.

Figure 6.2 Filter Paper
Similar separation takes place in a packed column, (stationary
phase), where the sample molecules are injected at the head of the column
and begin to move through the column under the motive forces of the
carrier gas, (moving phase), where the light molecules travel through the
column faster than the heavy ones. Therefore, the time that the light
Oil and Gas Measuring Instruments
105
molecules are in the column will be shorter than the time the heavy
molecules stay in the column. It is this difference in column retention for
different molecules that provides the separation. A detector is then
employed to measure the relative concentration of each component while
the elution time sequence can be employed to identify each component.

Figure 6.3 Sample column

6.3.1 Thermal Conductivity Detector (TCD)

The Thermal Conductivity Detector used in the GCX utilizes two
filaments, one for sample gas and one for reference gas flow. The
unbalancing of the bridge due to the dilution of the carrier gas by the
sample, and hence the change in Thermal Conductivity offers excellent
sensitivity for most applications.

6.3.2 Flame Ionization Detector (FID)

When a hydrocarbon sample passes through a hydrogen flame, the
molecular structure is altered so that the bond is broken and the carbon
atom becomes a negatively charged and the hydrogen atoms become
positively charged. When placed in an electric field, the ions may be
collected. In the case of the GCX, a positive potential on the polarizing
Oil and Gas Measuring Instruments
106
plate causes all positive ions are collected on the measuring plate or
collector and al negatively charged particles are collected on the
polarization plate. This current may be converted to a voltage for further
processing. The hydrogen/air mixture that supports the flame is converted
to water and exits the burner through the vent. Most of the oxygen is
consumed by the flame with only small amounts of excess hydrogen
remaining. The excess hydrogen (H2) passes through the flame to the
vent without being ionized. The only ionization that occurs is with the
hydrocarbon samples. By placing the burner tip within the effective
electric field, all positive ions will be collected and measured by the
measuring circuit. All negative Ions will be attracted to the positive
polarizing plate. Any extraneous electric fields that exist within the
system will change the performance of the burner. Thus, maintaining a
constant electric field and a clean system is of the utmost importance.
Response of the burner to hydrocarbon components. - In a chromatograph
system, each component to be measured is separated so that there is no
interference between components. Each component is calibrated using a
known concentration to determine response of the system to that
component. The relationship between components does not depend upon
each other but only on the calibration factor.

Figure 6.4 FID
Oil and Gas Measuring Instruments
107
6.4 Moisture Analyzer

The electrolytic moisture sensor consists of a pair of spirally
wound, parallel, electrode wires, partially embedded within the length of
a glass tube. A thin layer of highly absorbent, phosphorous pentoxide
(P2O5) completely coats the interior of the tube. In operation, the sample
gas stream passes through the tube, giving up its entrained water
molecules to the absorbent coating. A current is applied to the electrode
windings, whereby the water molecules are completely and continuously
electrolyzed into their respective hydrogen and oxygen elements.
The stream's moisture level is derived from the current required for
complete electrolysis of the absorbed water. Interpretation of this value is
based on the application of Faraday's Law of Electrolysis, which
describes the quantitative relationship of electrolyte production with the
application of electric current.
H
2
O + e- -> H
2
+ O
2

The current required to electrolyze the absorbed water is directly
proportional to the number of moisture molecules present, as electrolyzed
over a given time; that is the 'mass rate' of water entering the sensor. As
the current measurement is completely dependent upon the mass rate, it
becomes crucial that pressure and flow are strictly regulated.

Figure 6.5 Moisture Sensing element
Oil and Gas Measuring Instruments
108
6.5 Role Play

Each Trainee should speak thoroughly about one of the analyzers.
- Oxygen
- Moisture
- Gas Chromatography






















Oil and Gas Measuring Instruments
109
Chapter 7
Basic Considerations

7.1 Learning objectives

1. Introduce basic considerations for transmitter selection and installation.

7.2 Corrosion Effects

Corrosion is the gradual destruction of a metal by chemical or
electrochemical means. The most generic form of corrosion is galvanic
corrosion. A combination of a cathode, an anode, and an electrolyte must
be present for this type of corrosion.

- Material Selection Guide as for Rosemount
E= Excellent Resistance, Corrosion Rate (CR) < 0.05mm/year.
G= Good Resistance, CR < 0.5mm/year.
F= Fair Resistance, CR < 1.27 mm/year.
P=Poor Resistance, CR > 1.27 mm/year.
-- = Data Not Available.

7.3 Lightning and Static Effects

Lightning is the attraction of a charged cloud to an oppositely
charged earth, another cloud, or another area within the same cloud.
Clouds produce lightning with the help of strong updraft air currents.
These air currents cause rapid freezing of water droplets, which inherit a
charge as they crystallize. Among the many types of lightning, cloud to
Oil and Gas Measuring Instruments
110
ground strikes are the greatest threat to industrial electronic equipment.
Four factors are important in assessing the threat of lightning damage to a
plant or facility.
- Frequency and severity of lightning storms.
- Vulnerability of existing and proposed instrumentation.
- Exposure of systems wiring to possible lightning discharge.
- Potential harmful impact of instrument failure on the process.
Comparing the above factors to the costs of not protecting electronic
equipment will help to decide if protection is beneficial.

Figure 7.1 Globalannual Lightning stroms
Three strategies are effective in minimizing lightning induced
transients on industrial electronics.
- Diversion: Grounded metallic structures form a cone of protection for
equipment and cabling.
Oil and Gas Measuring Instruments
111
- Attenuation: Careful wiring practices, such as metallic raceways, cable
shields, twisted pairs, and extensive grounding and earthing reduce the
magnitude of transients.
- Suppression: Add-on devices limit the magnitude of the transient
appearing at the instrument.

7.4 Winterizing Transmitters

Ensuring that electronic pressure transmitters operate under all
weather conditions requires consideration of three important variable:
installation, protective measures, and cost. First, the transmitter must be
located properly with respect to the process pipe. Second, once optimum
installation is determined, consider the degree of temperature protection
required. Third, the degree of weatherization needed should then be
balanced against bottom line cost. Failures can be caused by the freezing
water or of solutions containing significant amounts of water. A volume
of water will increase about ten percent as it changes to ice at atmospheric
pressure. If the expansion is contained, the pressure exerted by the frozen
fluid increases the magnitude of this increase is large in comparison with
each incremental decrease in temperature.
Temperature (F) Pressure (psia)
32 14.7
30 2100
25 7000
18.5 12660
9.5 20056
5.0 23115
0.5 26103
Oil and Gas Measuring Instruments
112
In any case, proper installation is necessary for good transmitter
performance. In determining the best location, remember the following
guidelines:
- Keep corrosive or hot process material out of contact with the
transmitter.
- Prevent sediment from depositing in the impulse piping.
- Keep the liquid head balanced on both legs of the impulse piping.
- Keep the impulse piping as short as possible.
Avoid ambient temperature gradients and fluctuations

7.4.1 Liquid Service

For liquid flow measurement, mount the transmitters below the
process taps with the drain/vent valves facing downward. This allows the
trapped gases to vent into the process line. Make the taps to the side of
the line to avoid sediment deposits.

Figure 7.2 Liquid service connection
Oil and Gas Measuring Instruments
113
7.4.2 Gas Service

For gas flow measurement, install the transmitter above the process
taps with the drain/vent valves facing upward. This provides automatic
drainage and ensures that no liquid accumulates at the transmitter.

Figure 7.3 Gas service connection.

7.4.3 Protective Measures

Although winterizing a transmitter is relatively easy, protection
should not end there. Impulse lines must also be protected from the point
of measurement to the transmitter this may be accomplished in several
ways. Lines may be protected by tracing, insulation or both.

7.5 Total Probable Error

Total Probable error, TPE, is amore realistic number than would
obtained by simply adding up all the possible errors, since it is unlikely
Oil and Gas Measuring Instruments
114
that all errors would go in the same direction from their means. The root
sum square method, RSS, determines TPE by summing the squares of the
individual errors and taking the root square of the total. Below is a
comparison of two transmitters.

7.6 Discussion

An open discussion is to be opened about different consideration for
selection of transmitters.




Oil and Gas Measuring Instruments
115
References

1. Hugh Jack, "Automating Manufacturing Systems with PLCs",
(jackh@gvsu.edu); version 4.6 December, 2004.
2. A.K. Sawhney, "Electrical and electronics measurements and
instrumentation", J.C. Kapoor for Dhanpat Rai Co, Ltd. Naisarak,
Delhi 1999.
3. Leamington Spa, "Flowmeters", 1979.
4. R.B.Helson, " The HART protocol- a solution enabling
technology", HART communication foundation, 9390 research
blvd., suite II-250, Austin, Texas 78759.
5. Rosemount Measurement Catalog.
6. AEA Technology, "Level gauging", United Kingdom: 329
Harwell, Didcot, OX11 0QJ.
7. KROHNE, "Level Radar BM700".
8. Integrated Publishing Engine Mechanics, www.tpub.com.
9. SBEM, www.sbem-india.com
10. Sensor Network, www.sensornet-work.com.

You might also like