- GLOSSARY ACCURACY: The degree of absolute conformity to a standard. NIST: National Institute of Standards and Technology PRECISION: The accuracy with which a number can be represented; sharply defined; conforming to a pattern; tolerance REPEATABILITY: The ability of a probe or instrument to give the same output or reading under repeated identical conditions. RESOLUTION: The least significant digits shown on a digital display. RS-232: A standard computer interface protocol. S E N S O R S E L F - H E AT I N G : T h e i n t e r n a l h e a t generated in the sensor by passing a current through the sensor to measure its resistance. TEMPERATURE STANDARD: An instrument whos calibration is directly traceable to the NIST. STABILITY: The ability of a device to maintain a constant output with the application of a constant input. SYSTEM ACCURACY: The total accuracy of the probe plus the accuracy of the thermometer. SYSTEM CALIBRATION: Calibration of a thermometer and probe combination. THERMISTOR: A temperature sensing element composed of semiconductor material whose resistance varies greatly, in a known manner, with the change of temperature. (The 4600 uses negative temperature coefficient thermistors - as temperature increases, resistance decreases). P/N 139225 REV. A
28
A L P H A T E C H N I C S P R E C I S I O N T E M P E R AT U R E A L P H A T E C H N I C S P R E C I S I O N T E M P E R AT U R E
8/13/2008 11:19:35 AM
4 6 0 0
S E R I E S T H E R M O M E T E R 4 6 0 0 S E R I E S T H E R M O M E T E R
- TABLE OF CONTENTS INTRODUCTION ........................................ 1 OPERATING INSTRUCTIONS....................... 3 RS-232 INTERFACE................................... 11 THERMAL VIEW SOFTWARE ...................... 12 BATTERY REPLACEMENT............................ 13 CARE AND CLEANING ............................. 14 RECALIBRATION ....................................... 15 REPAIR ..................................................... 16 LIMITED WARRANTY ................................ 17 SPECIFICATIONS; 4600-I & 4600S ........... 18 SPECIFICATIONS; 4600-2 ........................ 23 ACCESSORIES .......................................... 26 ACCESSORIES: PROBES ............................ 27 GLOSSARY
*Conical shape; dry measurement (no oil needed) **For mercury thermometer replacement
ALPHA TECHNICS 2400 SERIES PROBES for Model 4400-1 & 4600-1 Thermometers
(Compatible with YSI 400 Series probes)
EUT2401 EUT2402 EUT2404 EUT2406 EUT2410 EUT2416 EUT2418 EUT2408 EUT2409A EUT2309B EUT2421 EUT2427 EUT2405 EUT2441 EUT2451
Vinyl Tip General-Purpose Probe Small Vinyl Tip General-Purpose Probe Glass Tubular Probe Stainless Steel Tubular Probe Tubular Probe with NPT Fitting Autoclavable Tubular Probe Pointed Tubular Probe Banjo Surface Probe High-Temperature Surface Probe Surface Probe Autoclavable Surface Probe Small Surface Probe Air/Gas Probe Airway Probe 1mm Flexible General-Purpose
27
A L P H A T E C H N I C S P R E C I S I O N T E M P E R AT U R E
i
A L P H A T E C H N I C S P R E C I S I O N T E M P E R AT U R E
8/13/2008 11:19:35 AM
4 6 0 0
S E R I E S
T H E R M O M E T E R
4 6 0 0
S E R I E S
T H E R M O M E T E R
- ACCESSORIES -
EUT4650 Thermal View Software See page 12 for details. EUT4651 RS-232 Cable 10 cable to connect the 4600 to a PC. EUT4652 Carrying Case Space for the 4600 and a probe. EUT4652D Deluxe Carrying Case Space for the 4600, several probes, power supply, Thermal View software, Cal plugs, etc. EUT4653 Calibration Plug Kit Allows the user to field check the calibration of the 4600 at several temperatures. EUT4654 Tripod Screws into an insert on the back of the 4600 thermometer. EUT4660 Power Cable Kit Battery eliminator plug with 6-foot lead and instructions for connection to any 12VDC regulated power supply. EUT4661-4663 US, UK, EURO Power Supplies Eliminates the need for batteries when the 4600 is used in a lab, or when doing extended data logging.
The Model 4600 Series are multi-purpose, economical, ultra-high accuracy highresolution, hand-held digital thermometers which accept standard YSI Series 400 thermistor probes and other probes. A very high system accuracy can be achieved by calibrating a thermometer to a probe. Along with HOLD, degrees F/C selection, MIN/MAX, STORE, DELTA T, and low battery indication features, the RS-232 port can also down-load data from the thermometer to a standard PC. Data can be manipulated and sent to other programs using the optional software package Thermal View. Thoughtful design has resulted in a userfriendly enclosure and key pad format. All functions can be easily accessed through the MODE and FUNCTION keys. Functions are user-assigned to desired keys.
1
A L P H A T E C H N I C S P R E C I S I O N T E M P E R AT U R E
26
A L P H A T E C H N I C S P R E C I S I O N T E M P E R AT U R E
8/13/2008 11:19:36 AM
4 6 0 0
S E R I E S
T H E R M O M E T E R
4 6 0 0
S E R I E S
T H E R M O M E T E R
- NOTES -
-INTRODUCTIONLarge Probe Selection The model 4600 accepts standard YSI Series 400 thermistor probes with phone jack termination. Alpha Technics high-accuracy probes are also available.
Computer Interface An RS-232 computer interface allows the user to download information from the 4600 to their own PC to create reports and graphs. PC software, called Thermal View, is available. Friendly Key Pad A user-friendly KEY PAD allows the operator to program the most frequently used functions into the unit. The functions include: HOLD, MIN/MAX, F/C selection, STORE for recording readings, delta T for measuring the difference between a stored value and the current reading.
Big Display Large .6 LCD numbers allows for easy viewing. All functions such as HOLD, STORE, Delta T, F/C, etc., are easily readable and logically arranged.
Low Noise Circuitry Proprietary circuitry allows for very clean low noise signals. Even background electronic noise is substantially reduced. Self-heating, lead resistance, thermistor anomalies and other factors are corrected for, and taken into account.
25
A L P H A T E C H N I C S P R E C I S I O N T E M P E R AT U R E
2
A L P H A T E C H N I C S P R E C I S I O N T E M P E R AT U R E
8/13/2008 11:19:38 AM
4 6 0 0
S E R I E S
T H E R M O M E T E R
4 6 0 0
S E R I E S
T H E R M O M E T E R
-OPERATING INSTRUCTIONS1. Determine that a probe is connected to the unit. 2. Press the green colored rubber key marked ON/OFF to turn the unit ON. (Pressing this key again will turn the unit OFF.) 3 READOUT: Display C/F symbols -40.00 to +150.00* (-40.00 to +299.98) LO* Explanation Reading is displayed in degrees C or F Actual temperature in degrees Celsius (Fahrenheit) Temperature is below -40.00C (-40.00F) or a probe is not inserted Temperature is above 150.00C (299.98F) Battery needs replacement**
- NOTES -
*Some versions of 4600 Series thermometers will have different temperature ranges. See the special specifications in the back of this manual or packed separately with the unit. **To ensure that erroneous results are not displayed, the numerical segments are disabled (all turned on or all off) one hour, of running time, after the battery symbol starts flashing.
3
A L P H A T E C H N I C S P R E C I S I O N T E M P E R AT U R E
24
A L P H A T E C H N I C S P R E C I S I O N T E M P E R AT U R E
8/13/2008 11:19:38 AM
4 6 0 0
S E R I E S
T H E R M O M E T E R
4 6 0 0
S E R I E S
T H E R M O M E T E R
- NOTES -
213.97
+
chk cal
LCD Readout
HOLD: Freezes the display at the present temperature when the HOLD function is activated. MIN/MAX: Displays the lowest (MIN) and highest (MAX) temperature from the time of turning on the meter. STORE: Records a temperature when the STORE function is activated. (The STOREd value can be retrieved and displayed in the DELTA T function.) The value is kept in memory until it is replaced or the unit is turned off. DELTA T: Calculates and displays the present temperature minus the STOREd temperature. *CAL: Allows for user calibration of the unit at specific temperature. C/F: Selects the temperature display units to be displayed in degrees Celcius or Fahrenheit. *CHK CAL: CHECK CALIBRATION allows the user to verify the calibration accuracy of the thermometer at specific temperatures by using certified calibration plugs.
*These functions are available on special 4600 versions only. See instructions with Calibration and Cal. Plug Kits.
23
A L P H A T E C H N I C S P R E C I S I O N T E M P E R AT U R E
4
A L P H A T E C H N I C S P R E C I S I O N T E M P E R AT U R E
8/13/2008 11:19:38 AM
4 6 0 0
S E R I E S
T H E R M O M E T E R
4 6 0 0
S E R I E S
T H E R M O M E T E R
F1
Key Pad
F2
INSTRUMENT ACCURACY VS. TEMPERATURE** (Not including probe error) TEMPERATURE (C) (F) -40 (-40) -30 (-22) -20 (-4) -10 (14) 0 (32) 10 (50) 20 (68) 30 (86) 40 (104) 50 (122) 60 (140) 70 (158) 80 (176) 90 (194) 100 (212) 110 (230) 120 (248) 130 (266) 140 (284) 150 (302) ACCURACY (C) (F) 0.100 (0.180) 0.060 (0.108) 0.035 (0.063) 0.025 (0.045) 0.020 (0.036) 0.015 (0.027) 0.015 (0.027) 0.015 (0.027) 0.017 (0.031) 0.019 (0.034) 0.020 (0.036) 0.020 (0.036) 0.025 (0.045) 0.030 (0.054) 0.035 (0.063) 0.040 (0.072) 0.050 (0.090) 0.060 (0.108) 0.080 (0.144) 0.100 (0.180)
All functions can be accessed through the MODE and FUNCTION (F1 & F2) keys. The user can assign any of the aforementioned functions to either the F1 or F2 keys by scrolling through the functions using the MODE key, then pressing either F1 or F2.
LCD returns to normal
CHK CAL
Check Cal point 2
point 1
MODE
C or F
Cal point 4
CAL
Cal point 1
point 2
Cal point 3
Cal
DELTA T
C/F
F or C
Check Cal
MIN/MAX
Shows Min temp.
HOLD
STORE
22
A L P H A T E C H N I C S P R E C I S I O N T E M P E R AT U R E
8/13/2008 11:19:39 AM
4 6 0 0
S E R I E S
T H E R M O M E T E R
4 6 0 0
S E R I E S
T H E R M O M E T E R
- SPECIFICATIONS MODEL 4600-2 SPECIFICATIONS: THERMISTOR: 10k Ohms @25C used in Alpha Technics Series 2400 probe housings. PROBE CONNECTION: 1/4 phone jack, permanently attached, or custom MEASUREMENT RANGE: -40.00C to +150.00C (40.00F to +299.98.00F) RESOLUTION: 0.01C (0.01F) from -40 to 102C (-40 to 215F) 0.02C (0.05F) from 102 to 150C (215 to 299.98F) REPEATABILITY: 0.002C to 0.01C at -20 to 70C (0.004 to 0.02F) at (-4 to 158F) typical for one week at constant ambient temperature SYSTEM ACCURACY: (probe + thermometer) When used in combination with a 0.1C probe, the thermometer/probe system will have an overall system accuracy of 0.13C or better when the temperature to be measured is between 0C and +70C. High system accuracy over a range of 0C to 100C is available by multiple point temperature calibration. Accuracy equaling that of the meter itself, at those temperatures, are possible. 21
A L P H A T E C H N I C S P R E C I S I O N T E M P E R AT U R E
-OPERATING INSTRUCTIONS5. KEY PAD OPERATION: When the MODE key is pressed, the active function icon flashes and the other function icons turn on solid. Pressing MODE again scrolls the display to the next function. When the desired function icon is flashing, pressing F1 or F2 assigns that function to the selected key and enters the thermometer into that selected function mode. If a function has not been selected, the MODE operation will automatically end after four flashes of the display (about 4 secs.) or if MODE is scrolled past C/F or CHK CAL*.
*Some versions do not have the CHK CAL function.
6
A L P H A T E C H N I C S P R E C I S I O N T E M P E R AT U R E
8/13/2008 11:19:39 AM
4 6 0 0
S E R I E S
T H E R M O M E T E R
4 6 0 0
S E R I E S
T H E R M O M E T E R
- SPECIFICATIONS MODEL 4600-1 & 4600S* SPECIFICATIONS: INSTRUMENT ACCURACY VS. TEMPERATURE** (Not including probe error) TEMPERATURE (C) (F) -40 (-40) -30 (-22) -20 (-4) -10 (14) 0 (32) 10 (50) 20 (68) 30 (86) 40 (104) 50 (122) 60 (140) 70 (158) 80 (176) 90 (194) 100 (212) 110 (230) 120 (248) 130 (266) 140 (284) 150 (302) ACCURACY (C) (F) 0.100 (0.180) 0.050 (0.090) 0.025 (0.045) 0.020 (0.036) 0.015 (0.027) 0.015 (0.027) 0.015 (0.027) 0.015 (0.027) 0.015 (0.027) 0.015 (0.027) 0.020 (0.036) 0.025 (0.045) 0.030 (0.054) 0.045 (0.081) 0.065 (0.117) 0.080 (0.144) 0.110 (0.198) 0.150 (0.270) 0.240 (0.432) 0.320 (0.576)
**Ambient temperature: 18 to 28C (64 to 82F) *The 4600S is a standard 4600 which is calibrated at a specific temperature or temperatures, up to four. The thermometer and probe are calibarted as a system and must stay together to maintain the system level accuracy specification shown on page 19.
7
A L P H A T E C H N I C S P R E C I S I O N T E M P E R AT U R E
20
A L P H A T E C H N I C S P R E C I S I O N T E M P E R AT U R E
8/13/2008 11:19:39 AM
4 6 0 0
S E R I E S
T H E R M O M E T E R
4 6 0 0
S E R I E S
T H E R M O M E T E R
- SPECIFICATIONS MODEL 4600-I & 4600S SPECIFICATIONS: PROBE: YSI 400 Series (2,252 Ohms @25C) CONNECTOR: 1/4 phone jack MEASUREMENT RANGE: -40.00C to +150.00C (40.00F to +299.98.00F) RESOLUTION: 0.01C (0.01F) from -40 to 102C (-40 to 215F) 0.02C (0.05F) from 102 to 150C (215 to 299.98F) REPEATABILITY: 0.002C to 0.01C at -20 to 70C (0.004 to 0.02F) at (-4 to 158F) typical for one week at constant ambient temperature 4600-I SYSTEM ACCURACY: When used in combination with an Alpha Technics Series 1400 or YSIs Series 400 0.1C probes, the thermometer/probe system will have an overall system accuracy of 0.13C or better when the temperature to be measured is between 0C and +70C. 4600S SYSTEM ACCURACY: When calibrated between -20 to 100C over a 70C user specified range, accuracy equaling the meter itself plus 0.020C is possible. 19
A L P H A T E C H N I C S P R E C I S I O N T E M P E R AT U R E
-OPERATING INSTRUCTIONS5. KEY PAD OPERATION: First, you need to assign the STORE function to the F1 key, and the DELTA T function to the F2 key. (You can assign them differently if you wish.) Press the mode key 3 times. (The STORE icon is flashing.) Press the F1 key. (F1 is now the STORE function.) Press F1 again to return to normal operation. Press the MODE key 4 times. (The DELTA T icon is flashing.) Press the F2 key. (F2 is now the DELTA T function) Press F2 two more times to get back to normal operation. Now, F1 is STORE, and F2 is DELTA T. TAKING THE DELTA T READING: Put your probe into beaker A. Allow the reading to stabilize. Press F1 (the STORE function). The STORE icon and the temperature numbers are flashing on and off. This lets you know that you are about to put a value into memory (erasing any previous value which had been stored). When F1 is pressed a second time, the value on the display is put into memory and the thermometer returns to normal operation. 8
A L P H A T E C H N I C S P R E C I S I O N T E M P E R AT U R E
8/13/2008 11:19:39 AM
4 6 0 0
S E R I E S
T H E R M O M E T E R
4 6 0 0
S E R I E S
T H E R M O M E T E R
-OPERATING INSTRUCTIONS5. KEY PAD OPERATION: Put the temperature probe into beaker B. Press F2 (DELTA T). The DELTA T icon will appear and the numerical reading will show the difference between the stored value, beaker A, and the temperature of the present reading, beaker B. This is a real-time delta measurement. As the temperature of beaker B changes, you will see the difference between beaker B and the earlier stored value of beaker A. Press F2 again. The stored value is displayed (beaker A). Press F2 again, and the thermometer returns to normal operation. Press F2 again, and you will see the delta measurement again, and so on.
ENVIRONMENTAL CONDITIONS: Operating: Temperature 15 - 30C (59 to 86F) Humidity 0 - 85% Storage: Temperature 0 - 60C (32 to 140F) Humidity 0 - 70% READING RATE: 2/sec. (one reading every 524msec) DISPLAY: 4 1/2 digit LCD POWER: 9 VDC battery, alkaline. Auto shut off time: 10 minutes. Auto shut off can be disabled or changed via the RS-232 port. Optional 110VAC adapter. BATTERY LIFE: 20 hours typical; battery symbol flashes when less than 5% of life, one hour running time remains. All LCD segments turn on or off after the one hour of running time. DIMENSIONS: 210H x 108W x 44D mm (8.25H x 4.25W(max) x 1.75D(max)) WEIGHT: 405g (14oz.)
9
A L P H A T E C H N I C S P R E C I S I O N T E M P E R AT U R E
18
A L P H A T E C H N I C S P R E C I S I O N T E M P E R AT U R E
8/13/2008 11:19:40 AM
4 6 0 0
S E R I E S
T H E R M O M E T E R
4 6 0 0
S E R I E S
T H E R M O M E T E R
-WARRANTYALPHA TECHNICS THERMOMETER 1 YEAR LIMITED WARRANTY The thermometer manufacturer, Alpha Technics, warrants your thermometer to be free from defects in material and workmanship under normal use and service. The warranty period for the Thermometer is 1 YEAR from the date of purchase and DOES NOT apply to batteries, or when the instrument has been misused, altered or damaged by accident or abnormal conditions of operation. For warranty service, send the instrument with a description of the difficulty, postage prepaid, to Alpha Technics. Alpha Technics assumes no risk for damage in transit. Alpha Technics will, at its option, repair or relace the defective instrument free of charge. However, if we determine that the failure was caused by misuse, alteration, accident, or abnormal condition of operation,you will be billed for the repair. The repaired instrument will be returned, transportation prepaid. ALPHA TECHNICS MAKES NO WARRANTY INCLUDING IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE, OTHER THAN THE WARRANTY STATED ABOVE. ALL WARRANTIES ARE LIMITED TO A PERIOD OF 1 YEAR. FROM THE DATE OF PURCHASE, ALPHA TECHNICS SHALL NOT BE LIABLE FOR ANY SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES, WHETHER IN CONTRACT, TORT, OR OTHERWISE.
Note (U.S.A. only): Some states do not allow limitations of implied warranties, or exclusion of incidental or consequential damages, so the above limitations or exclusions may not apply to you. This warranty gives you specific legal rights and you may have other rights which vary from state to state.
-OPERATING INSTRUCTIONS6. AUTO SHUTOFF: The thermometer is factory set to automatically turn itself off 10 minutes after being turned on or 10 minutes after the last function key entry. You may turn the unit off at any time by pressing the ON/OFF switch. When using the RS-232 interface or the optional external power supply, the auto shutoff feature is automatically disabled. Optional PC software called Thermal View allows the user to vary the auto shutoff time period. The thermometer can be configured to shutoff from 3 to 998 minutes, or not auto shutoff. Thermal View is a Windows based data acquisition and manipulation package which allows information from the thermometer to be sent, via RS-232, to a standard Personal Computer. See the Thermal View manual for additional information and other configuration features.
17
A L P H A T E C H N I C S P R E C I S I O N T E M P E R AT U R E
10
A L P H A T E C H N I C S P R E C I S I O N T E M P E R AT U R E
8/13/2008 11:19:40 AM
4 6 0 0
S E R I E S
T H E R M O M E T E R
4 6 0 0
S E R I E S
T H E R M O M E T E R
-RS-232 INTERFACEThe RS-232 interface allows temperature data from the 4600 Series thermometers to be sent to a standard Personal Computer. An optional software package known as Thermal View allows for easy data capture and manipulation in a Windows environment. Other data capture programs may be used. When using non-Thermal View data capture programs, the following information may be necessary:
The 4600 Series thermometers enter the HOST mode when the RTS line becomes active true. When the RTS line becomes FALSE, the 4600 returns to the normal mode of operation. Baud Rate: Data Length: Parity: Stop Bits: 1 9600 8 Bits None
-REPAIRIf the Thermometer fails, forward it, postage paid, to Alpha Technics. Include a description of the difficulty, place the Thermometer in its carrying case or box and pack the unit securely. Alpha Technics shall assume NO responsibility for damage in transit. IN WARRANTY: Instruments covered by the limited warranty will be promptly repaired or replaced, at Alpha Technics option, and returned at no charge. SEE LIMITED WARRANTY, PAGE 17, FOR ITEMS COVERED AND COMPLETE WARRANTY TERMS. (Other manufacturers probes are not included under Alpha Technics warranty unless otherwise stated.) OUT OF WARRANTY: The Thermometer will be repaired and returned for a fixed fee. (Repairs needed because of abuse or accidental damage will be quoted.) Contact Alpha Technics for current prices. Include a check, money order, or purchase order with the unit.
The 4600 has a standard DB9 female connector with the following pins: PIN 1 2 3 4 5 6 7 8 9 DIRECTION to Host From Host From Host SIGNAL NAME No Connection RX (Receive Data) TX(Transmit Data) No Connection GND No Connection RTS (Request to Send) No Connection No Connection
11
A L P H A T E C H N I C S P R E C I S I O N T E M P E R AT U R E
16
A L P H A T E C H N I C S P R E C I S I O N T E M P E R AT U R E
8/13/2008 11:19:40 AM
4 6 0 0
S E R I E S
T H E R M O M E T E R
4 6 0 0
S E R I E S
T H E R M O M E T E R
-RECALIBRATIONThe Thermometer should be recalibrated at least once a year to maintain its accuracy. (Some 4600 versions may require a 6 month recalibration cycle). Check the Report of Calibration, which came with the unit, or the recalibration label on the unit to determine the next recalibration date.) Recalibration should be performed at Alpha Technics. When the Thermometer requires recalibration, forward it, postage paid, to: Alpha Technics 17151 Gillette Avenue Irvine, CA 92614-5602 Phone (949) 250-6578 Fax: (949) 271-2300
-THERMAL VIEW SOFTWAREThermal View Basic is a WINDOWS based software package which allows the user to control the capture and presentation of temperature data from RS-232 equipped Alpha Technics thermometers. The information can be displayed or printed in Thermal View or exported to other programs. Savable Protocol windows allow the user to select start/stop times and temperatures, set alarms, select units, set duration and other parameters. The capture protocol can be defined to start and stop data collection based on the time of day or on temperature. Status and Alarm windows allow the user to monitor real-time readings and user-set alarm triggered pop up windows while working in other WINDOWS programs.
Contact Alpha Technics for current prices and include with the unit a purchase order, a ship to/bill to address, a phone number, and a contact name. Ship the unit in its original carrying case or box enclosed in a suitable shipping container.
15
A L P H A T E C H N I C S P R E C I S I O N T E M P E R AT U R E
12
A L P H A T E C H N I C S P R E C I S I O N T E M P E R AT U R E
8/13/2008 11:19:42 AM
4 6 0 0
S E R I E S
T H E R M O M E T E R
4 6 0 0
S E R I E S
T H E R M O M E T E R
-BATTERY REPLACEMENTThe Model 4600 Series uses a standard 9 Volt alkaline battery. To replace the battery, turn the unit over, onto its face, and remove the battery door at the bottom of the case. The old battery may be pried out from the mounting terminals by using your thumb or fore finger. When putting in the new 9 Volt alkaline battery, be sure that the terminals line up and snap firmly into the terminal on the circuit board. (The unit will not be damaged if you accidentally try to put the battery in backwards.) Replace the battery door. To increase battery life, a 9 Volt lithium battery can be used
-CARE & CLEANINGExtensive testing has proven this case to be rugged and reliable. It is splash resistant, but not waterproof. The case may not be submerged! Liquid will flow in and damage or ruin the thermometer! DO NOT AUTOCLAVE THE UNIT! If it is necessary to decontaminate the thermometer, a weak chlorine solution may be gently wiped on to the case by a well rung out soft cloth. If your 4600 version has any open connector jacks, do not let any liquid go into the jacks. Your readings will be affected. The clear plastic LCD lens can be easily scratched and/or fogged by inappropriate cleaning products. Use only plastic approved lens cleaners. Alcohol, acetone, lacquer thinners and other harsh chemicals will fog the lens and damage the case and rubber key pad. If your 4600 version has a detachable probe with a plug termination, periodically wipe off and clean the probe plug contacts. Dirty contacts will cause reading errors.