Data Sheet

10/14-6.84 EN

Thermal Mass Flowmeter FMT700-P (Sensyflow P)


for air, test benches

Direct mass flow measurement of air no additional pressure and temperature compensation required Wide measuring range of 1:40 Highest accuracy over the entire range measured error < 1 % of measured value factory-calibrated, with (optional) DKD calibration certificate Amazingly quick response time <12 ms detection of rapid load changes on engine test benches Low pressure loss No moving parts no wear, maintenance-free Complete system with inlet/outlet runs, air filter and links User-friendly handling immediately ready for use DN 25...DN 200 Separate supply/evaluation unit with display, diagnostic functions and various signal outputs ASAM GDI interface optionally Reference system for intake air measurement of the leading car manufacturers worldwide Used in quality assurance, test bench applications, research & development

High-end reference system Quick response time User-friendly handling

Thermal Mass Flowmeter FMT700-P (Sensyflow P) for air, test benches

10/14-6.84 EN

Function and system design


FMT700-P (Sensyflow P) operates according to the principle of a hot-film anemometer. With this measuring method the gas mass flow rate can be determined directly, without the need for pressure and temperature compensation. FMT700-P (Sensyflow P) comprises a transducer (measuring tube), a supply/evaluation unit, and a measuring section. The transducer is designed as a measuring tube and accommodates the sensor unit and an electronic transmitter circuit. It is available in 6 nominal sizes ranging from DN 25 to DN 200 and is installed in the measuring section by using quick-clamping connectors. The measuring section consists of inlet and outlet runs of the appropriate size and of an air filter or flow conditioner which can be connected e.g. to an air duct via the connection piece on the suction side. The supply/evaluation unit is available as a 19" plug-in version or as a desktop unit. It provides the power supply for the transducer and converts its flow-dependent current signal into standard analog mass flow signals. The supply/evaluation unit with 6-digit display is suitable for all measuring tubes and automatically identifies the appropriate nominal size. Further diagnostic functions allow simple and safe operation. The measuring rate is adjustable according to measuring velocity or signal damping requirements, In case of high measuring velocity requirements the measuring rate can be reduced to 1 ms. Available as option: gas temperature measurement measuring value as standard volume flow serial interface ASAM GDI interface connection for 2. transducer
m Rh RT Ih

Typical applications
In a unique manner, FMT700-P (Sensyflow P) mass flow meters for air combine high accuracy, huge turn-down ratio and extremely fast response time. These features qualify them for the following application fields: Suction air measurements at combustion engines Test benches for turbo chargers Serial testing of flow dependent components like throttle valves, fans, air filters Quality assurance: reference unit for other flow meters Research & Development at universities and institutes

Ordering information
The measuring system consists of the following components: 1. Transducer/measuring tube 2. Supply/evaluation unit with measured value display unit 3. Measuring section with air filter or flow conditioner 4. Cables for connecting the transducer and the supply/ evaluation unit

Measuring principle
FMT700-P (Sensyflow P) operates according to the principle of the hot-film anemometer. This method of measurement is based on the abstraction of heat from a heated body by an enveloping gas flow. The flow-dependent cooling impact is used as the measuring effect. The gas stream flows past two temperature-sensitive resistors Rh and RT which are part of an electrical bridge circuit. Due to the chosen resistance ratio Rh << RT, Rh is heated by the current Ih, and RT adopts the same temperature as the gas. The current Ih is preset by the electronic control circuit to produce a constant temperature difference between the heated resistor Rh and the temperature of the gas. The electrical power generated with resistor Rh exactly compensates its loss of heat to the gas flow. As this loss of heat is dependent on the number of particles which collide with the surface of resistor Rh, Ih represents a measure of the mass flow rate.

Thermal Mass Flowmeter FMT700-P (Sensyflow P) for air, test benches

10/14-6.84 EN

Technical data
Measuring principle
Thermal: hot-film anemometer

Operating conditions
Recommended steadying length
Inlet run 10 pipe diameter (use of air filter or flow conditioner strongly recommended) Outlet run 5 pipe diameter

Input
Measured medium Flow rate of air

Measuring ranges (standard)


Nominal size DN 25 DN 50 DN 80 DN 100 DN 150 DN 200 kg/h 0 ( 1)... 60 0 ( 10)... 400 0 ( 20)... 720 0 ( 40)...1200 0 ( 80)...2400 0 (200)...4000

Environmental conditions
Ambient temperature for transducer -25...+80 C Ambient temperature for evaluation unit -25...+50 C Storage temperature -25...+85 C Degree of protection IP 54 (transducer)

The value in brackets specifies the lower limit of the measuring range for which the given accuracy is achieved.

Output
Output signals
Analog 0... 10 V, 0... 20 mA, 4... 20 mA, (< 4 mA) (load < 500 ) (load < 500 )

Measured medium conditions


Operating temperature Medium -25...+80 C Operating pressure Standard: Option: 0.6...2.5 102 kPa (2.5 bar abs.) up to 0.6...8 102 kPa (8 bar abs.) only DN 25

Digital output serial V24/RS 232 C, electrically isolated ASAM-GDI interface

Construction
Weight Transducer/measuring tube depending on nominal size, see ordering information Evaluation unit 19" plug-in version 7.0 kg 1/2 19" desktop housing 7.3 kg Material Transducer: Steadying lengths: aluminium, black anodized aluminium, black anodized or glass fibre, black coated

Characteristics
Accuracy
Measured error (including hysteresis and non-linearity) < 1 % of measured value Repeatability: < 0.25 % of measured value

Influences
Temperature effect < 0.03 %/K of measured value Pressure effect 0.2 %/100 kPa (/bar) of measured value Response time: T63 12 ms

Process connection Quick-clamping pipe flange, aluminium with quick-clamp chains / rings Electrical connection Transducer: special connecting cable to evaluation unit

Pressure drop transducer/measuring tube

Power supply
Pressure drop under atmospheric conditions 100,00 Pressure drop dp [hPa] 10,00 1,00 0,10 0,01 1 10 100 Mass flow rate Qm 1000 [kg/h] 10000
DN 25 DN 50

DN 80 DN 100 DN 150 DN 200

Power supply/evaluation unit, voltage: 230 V AC 110 V AC Power consumption evaluation unit 38 W Power consumption transducer 10 W Current drain transducer < 600 mA

Thermal Mass Flowmeter FMT700-P (Sensyflow P) for air, test benches

10/14-6.84 EN

Dimensional drawings (dimensions in mm)

Sensyflow P NW 25

170 22.5 125 22.5


52

73
21

DN25KF

24

92

15

95

15

M6 6

12
Z-18960

Sensyflow P NW 50 ... NW 200

L2

L1

L2

di

D1

Z-18961

DN 50 80 100 150 200


Dimensions in mm

D 64 89 118 165 220

D1 80,0 108,5 132,5 180,0 240,0

di 58 80 109 153 200

L 180 185 250 280 330

L1 125

L2 27,5 30,0 62,5 77,5 102,5

B 92

31.5

60.5

H 85,0 93,5 108,0 131,5 159,0

Thermal Mass Flowmeter FMT700-P (Sensyflow P) for air, test benches

10/14-6.84 EN

Dimensional drawings (dimensions in mm)

19 plug-in version

I L

19 desktop housing

required cutout for 19" unit 450 mm x 131 mm

for wiring

Z-17384

Unit Dimensions 19 desktop housing 19 plug-in version


Dimensions in mm

L 310 483

I / 462

H 140 132

h / 58

T / 425

t / 325

X / M6

Thermal Mass Flowmeter FMT700-P (Sensyflow P) for air, test benches

10/14-6.84 EN

Dimensional drawings (dimensions in mm)


Modules DN 25
KF = Small flange
12 13

O-ring

Clamping ring FL-special


15 57

Filter with flange


ca.244 KF KF

10 x D inlet tube with flanges (on both sides)


269 KF

5 x D outlet tube with flanges (on both sides)


KF 144 KF

Hose adapter with flange (on one side)


KF 34

11 Innercenter ring
4

14 Clamping ring FL-Optimal AS


16 73

ca. 110

26.1

26.1

26,1

26,1

26,1

26.1

27

32

32

24

24

30

2,5

2,5
4

2,5

2,5
5

2,5

2,5

10 x D inlet tube with flange (on one side)


260 KF

5 x D outlet tube with flange (on one side)


KF 135

12 6 11 6 sectional detail pipe connection (without clamping ring)


27
KF KF

26,1

26,1

27

32

24

32

24

25

2,5

2,5

25

24

27

10 4

10

Standard measuring sections


ca.830 ca.244 4 269 4 170 4 135

Measuring section 1

573 260 4 170 4 135

Measuring section 2

Individual planning
170

55

46

Thermal Mass Flowmeter FMT700-P (Sensyflow P) for air, test benches

10/14-6.84 EN

Dimensional drawings (dimensions in mm)


Modules DN 50
KOF = Tapered flange (with recessed face and groove for O-ring) ZWF = Intermediate flange (with raised face)
13

Clamping ring
20 102

Filter cartridge (open type) with flange


ca.356 KOF ZWF

10 x D inlet tube with flanges (on both sides)


506 KOF

5 x D outlet tube with flanges (on both sides)


KOF 256 ZWF

Hose adapter with flange (on one side)


KOF 50

10

12

72

11

Intermediate O-ring flange


2 ZWF

Tapered flange
KOF 4

ca. 150

66

58

60

70

66

58

64

64

64

64

64

64

64

64
2

Flow conditioner with flange


X ca.450 KOF

10 x D inlet tube with flange (on one side)


504 KOF

5 x D outlet tube with flange (on one side)


KOF 254

10

12

11

sectional detail pipe connection (without clamping ring)


64

ca. 200

64

66

58

66

64

64

64

58

58

64

25

25

222

Dimensions X and Y depend on customer requirement

Standard measuring sections


ca.1300 ca.356 506 184 254

Measuring section 1 (air filter open) Measuring section 3 Flow conditioner

KOF ZWF

KOF ZWF 942 504 184

ZWF KOF

254

Measuring section 2
KOF ZWF ZWF KOF

Individual planning
184

ZWF

ZWF

64

Thermal Mass Flowmeter FMT700-P (Sensyflow P) for air, test benches

10/14-6.84 EN

Dimensional drawings (dimensions in mm)


Modules DN 80
KOF = Tapered flange (with recessed face and groove for O-ring) ZWF = Intermediate flange (with raised face)
13

Clamping ring
20 145

1 Filter cartridge (open type) with flange


ca.401 KOF ZWF

2 10 x D inlet tube with flanges (on both sides)


806 KOF

3 5 x D outlet tube with flanges (on both sides)


KOF 406 ZWF

6 Hose adapter with flange (on one side)


KOF 80

10

12

114
11

Intermediate O-ring flange


2 ZWF

Tapered flange
KOF 4

ca. 200

89

80

91

80

89

89

89

85

91

95

89

89

89

89
2

7 Flow conditioner with flange


X
ca. 250
ca.450 KOF

4 10 x D inlet tube with flange (on one side)


804 KOF

5 5 x D outlet tube with flange (on one side)


KOF 404

10

12

11

sectional detail pipe connection (without clamping ring)


89

91

91

89

80

89

89

89

80

80

89

25

25

222

Dimensions X and Y depend on customer requirement

Standard measuring sections


ca.1800 ca.401 7 1 806 2 189 404 5

Measuring section 1 (air filter open) Measuring section 3 Flow conditioner

KOF ZWF

KOF ZWF 1397 804 4 189

ZWF KOF

404 5

Measuring section 2
KOF ZWF ZWF KOF

Individual planning
189

ZWF

ZWF

89

Thermal Mass Flowmeter FMT700-P (Sensyflow P) for air, test benches

10/14-6.84 EN

Dimensional drawings (dimensions in mm)


Modules DN 100
KOF = Tapered flange (with recessed face and groove for O-ring) ZWF = Intermediate flange (with raised face)
13

Clamping ring
20 158

Filter cartridge (open type) with flange


ca.526 KOF ZWF

10 x D inlet tube with flanges (on both sides)


1006 KOF

5 x D outlet tube with flanges (on both sides)


KOF 506 ZWF

Hose adapter with flange (on one side)


KOF 100

10

12

127

11

Intermediate O-ring flange


2 ZWF

Tapered flange
KOF 4

ca. 240

118

118

118

118

119

109

118

118

119

118

114

109

122

118
2

Flow conditioner with flange


X ca.550 KOF

10 x D inlet tube with flange (on one side)


1004 KOF

5 x D outlet tube with flange (on one side)


KOF 504

10

12

11

sectional detail pipe connection (without clamping ring)


118

ca. 300

119

109

119

118

118

118

118

109

109

118

25

25

222

Dimensions X and Y depend on customer requirement

Standard measuring sections


ca.2290 ca.526 1006 254 504

Measuring section 1 (air filter open) Measuring section 3 Flow conditioner

KOF ZWF

KOF ZWF 1762 1004 254

ZWF KOF

504

Measuring section 2
KOF ZWF ZWF KOF

Individual planning
254

ZWF

ZWF

118

Thermal Mass Flowmeter FMT700-P (Sensyflow P) for air, test benches

10/14-6.84 EN

Dimensional drawings (dimensions in mm)


Modules DN 150
KOF = Tapered flange (with recessed face and groove for O-ring) ZWF = Intermediate flange (with raised face)
13

Clamping chain
37 ca.220

Filter cartridge (open type) with flange


ca.513 KOF ZWF

10 x D inlet tube with flanges (on both sides)


1518 KOF

5 x D outlet tube with flanges (on both sides)


KOF 768 ZWF

Hose adapter with flange (on one side)


KOF

10

12

11

Intermediate O-ring flange


9 ZWF

Tapered flange
KOF 9

ca. 300

158

158

158

158

158

158

162

158

150

162

150

150

160

158
4

Flow conditioner with flange


X ca.550 KOF

10 x D inlet tube with flange (on one side)


1509 KOF

5 x D outlet tube with flange (on one side)


KOF 759

10

12

11

sectional detail pipe connection (without clamping chain)


155

ca. 350

162

150

162

158

155

158

158

150

150

155

25

25

54 9

Dimensions X and Y depend on customer requirement

Standard measuring sections


ca.3070 ca.513 1518 280 759

Measuring section 1 (air filter open) Measuring section 3 Flow conditioner

KOF ZWF

KOF ZWF 2548 1509 280

ZWF KOF

759

Measuring section 2
KOF ZWF ZWF KOF

Individual planning
280

ZWF

ZWF

10

155

Thermal Mass Flowmeter FMT700-P (Sensyflow P) for air, test benches

10/14-6.84 EN

Dimensional drawings (dimensions in mm)


Modules DN 200
KOF = Tapered flange (with recessed face and groove for O-ring) ZWF = Intermediate flange (with raised face)
13

Clamping chain
37 ca.280

Filter cartridge (open type) with flange


ca.513 KOF ZWF

10 x D inlet tube with flanges (on both sides)


2018 KOF

5 x D outlet tube with flanges (on both sides)


KOF 1018 ZWF

Hose adapter with flange (on one side)


KOF 159

10

12

11

Intermediate O-ring flange


9 ZWF

Tapered flange
KOF 9

ca. 300

208

205

212

205

205

212

200

205

200

205

205

205

200

205
4

Flow conditioner with flange


X ca.550 KOF

10 x D inlet tube with flange (on one side)


2009 KOF

5 x D outlet tube with flange (on one side)


KOF 1009

10

12

11

sectional detail pipe connection (without clamping chain)


205

ca. 350

205

205

205

212

200

205

212

200

200

205

25

25

54 9

Dimensions X and Y depend on customer requirement

Standard measuring sections


ca.3870 ca.513 2018 330 1009

Measuring section 1 (air filter open) Measuring section 3 Flow conditioner

KOF ZWF

KOF ZWF 3348 2009 330

ZWF KOF

1009

Measuring section 2
KOF ZWF ZWF KOF

Individual planning
330

ZWF

ZWF

205

11

Thermal Mass Flowmeter FMT700-P (Sensyflow P) for air, test benches

10/14-6.84 EN

Ordering information
Catalog No. Transducer FMT700-P Comprising of: Sensor unit (built-in), transmitter circuit in terminal box including calibration Transducer, standard characteristic curve and calibration certificate Measuring-point parameters: Type of gas: air Operating temperature: -25...80 C, Pressure (absolute): 0,6..2.5 bar Measuring pipe Measuring range Weight NW 25 0 ( 1)... 60 kg/h 1.100 NW 50 0 ( 10)... 400 kg/h 2.000 NW 80 0 ( 20)... 720 kg/h 2.300 NW 100 0 ( 40)...1200 kg/h 3.100 NW 150 0 ( 80)...2400 kg/h 4.300 NW 200 0 (200)...4000 kg/h 7.900 1) Transducer, special characteristic curve and calibration certificate Measuring-point parameters: Type of gas: air Measuring pipe Weight (kg) NW 25 1.100 NW 50 2.000 NW 80 2.300 NW 100 3.100 NW 150 4.300 NW 200 7.900 Code EUR Deliv. time 4 wks.

7962633 7962634 7962635 7962636 7962637 7962638

4092,00 4265,00 4717,00 5053,00 6461,00 7880,00

7962639 7962640 7962641 7962642 7962643 7962644

4780,00 4953,00 5400,00 5741,00 7149,00 8562,00

Additional ordering information


Code Definition of special characteristic Operating temperature (please specify with clear text) Operating pressure (please specify with clear text) Measuring range (please specify with clear text) Unit (please specify with clear text) Standard state (please specify with clear text) DKD certificate In house calibration with air. DKD = Deutscher Kalibrierdienst (German calibration service), supervised by PTB (Physical and Technical Federal Institute) 512 513 514 515 516 310 EUR 361,00 Deliv. time

2)

Ordering information for equipment packages


Supply/evaluation unit for 1 transducer 19" Plug-in version Weight 7.0 kg Display 6 digit, display unit kg/h Power supply 230 V AC Measuring rate adjustable Serial interface V24/RS 232C, analog output 0...10 V Temperature display 1/2 19" Desktop housing Weight 7.3 kg Display 6 digit, display unit kg/h Power supply 230 V AC Measuring rate adjustable Serial interface V24/RS 232C, analog output 0...10 V Temperature display

3) Catalog No. 7964113 Code EUR 6251,00 Deliv. time 4 wks.

7964114

6572,00

1) Definition of the special characteristic curve, see additional ordering information 2) e. g. Nm3/h related to standard conditions 0 C, 1013 mbar (hPa) 3) 10 % discount in relation to individual combination

12

Thermal Mass Flowmeter FMT700-P (Sensyflow P) for air, test benches

10/14-6.84 EN

Ordering information
Supply/evaluation unit FMT700-P Supply/evaluation unit Display 6 digit, display unit kg/h, measuring rate adjustable for 1 transducer for 2 transducers weight (kg) Design 19" Plug-in card 7.0 1/2 19" Desk-top housing 7.3 2nd Display unit see Code Nos 401-409 Temperature display see Code Nos 411, 412 Power supply 230 V AC 110 V AC Digital interface without V24/RS 232 C, serial ASAM-GDI and V24/RS 232 C, serial Totalizer without Totalizer, only in combination with V24/RS 232 C interface Output signal 0...10 V 0...10 V and 0...20 mA 0...10 V and 4...20 mA Catalog No. V14243Code 0 EUR 4890,00 Deliv. time 4 wks.

1 2 1 2

1639,00 357,00

1 2 0 1 2 0 1 1 4 5

925,00 1450,00 1544,00 468,00 946,00 946,00

Additional ordering information


Code Details only with subsequent or individual order for special characteristic curves or standard volume unit Measuring-point parameter, type of gas air Operating temperature (please specify in clear text) Operating pressure (please specify in clear text) Measuring range (please specify in clear text) Unit (please specify in clear text) Standard state (please specify in clear text) 2nd display unit, switchable to: g/s Nm3/h (at 0 C and 1013 hPa) Nm3/h (at 20 C and 1013 hPa) Nl/s (at 0 C and 1013 hPa) Nl/s (at 20 C and 1013 hPa) Nl/min (at 0 C and 1013 hPa) Switchable to (please note Code-No 516), please specify with clear text Temperature display for 1 transducer for 2 transducers EUR Deliv. time

512 513 514 515 516

401 402 403 404 405 406 409 411 412

315,00 315,00 315,00 315,00 315,00 315,00 315,00 662,00 1030,00

Accessories
Catalog No. Supply/evaluation unit, later implementation of ASAM-GDI including certification (only in combin. with serial interface) 7962840 EUR 525,00 Deliv. time

13

Thermal Mass Flowmeter FMT700-P (Sensyflow P) for air, test benches

10/14-6.84 EN

Ordering information accessories


Catalog No. Measuring section 1, comprising of: Air filter (open), inlet tube 10 x D, 1 tapered flange, 1 intermediate flange, outlet tube 5 x D, 1 tapered flange, 3 clamping rings/chains Nominal size Weight (kg) NW 25 1.500 NW 50 4.100 NW 80 8.250 NW 100 12.600 NW 150 20.000 NW 200 25.600 Measuring section 2, comprising of: Inlet tube 10 x D, 1 tapered flange, outlet tube 5 x D, 1 tapered flange, 2 clamping rings/chains Nominal size Weight (kg) NW 25 0.530 NW 50 2.030 NW 80 5.100 NW 100 7.600 NW 150 12.000 NW 200 16.800 Measuring section 3, comprising of: Flow conditioner, inlet tube 10 x D, 1 tapered flange, 1 intermediate flange, outlet tube 5 x D, 1 tapered flange, 3 clamping rings/chains, hose socket on suction side Nominal size suction side Weight (kg) NW 50 NW 80 9.600 NW 80 NW 100 13.900 NW 100 NW 150 21.200 NW 150 NW 200 32.000 NW 200 NW 250 39.000 4) Air filter (open) , Filter coupling flange, 1 clamping ring/chain Nominal size Weight (kg) NW 25 0.900 NW 50 1.980 NW 80 3.100 NW 100 4.930 NW 150 7.500 NW 200 8.100 4) Flow conditioner , Filter coupling flange, 1 clamping ring/chain, hose socket on suction side Nominal size suction side Weight (kg) NW 50 NW 80 7.500 NW 80 NW 100 8.700 NW 100 NW 150 13.500 NW 150 NW 200 19.500 NW 200 NW 250 21.500 Inlet tube 10 x D , 1 Tapered flange, 1 intermediate flange, 1 clamp. ring/chain 4) Nominal size Weight (kg) NW 25 0.350 NW 50 1.380 NW 80 3.300 NW 100 5.000 NW 150 8.200 NW 200 11.400 4) See dimensional drawings, Data Sheet Code EUR Deliv. time 4 wks.

7962645 7962646 7962647 7962648 7962649 7962650

588,00 715,00 940,00 1188,00 2254,00 2589,00

7962651 7962652 7962653 7962654 7962655 7962656

300,00 388,00 583,00 715,00 1393,00 1628,00

7964108 7964109 7964110 7964111 7964112

1723,00 2001,00 2354,00 3593,00 4008,00

7962657 7962658 7962659 7962660 7962661 7962662

295,00 315,00 452,00 473,00 678,00 788,00

7964102 7964103 7964104 7964105 7964106

1324,00 1513,00 1639,00 2017,00 2206,00

7962663 7962664 7962665 7962666 7962667 7962668

184,00 258,00 405,00 468,00 1003,00 1135,00

14

Thermal Mass Flowmeter FMT700-P (Sensyflow P) for air, test benches

10/14-6.84 EN

Ordering information accessories


Catalog No. 4) Inlet tube 10 x D , 1 Tapered flange, 1 clamping ring/chain Nominal size Weight (kg) NW 25 0.320 NW 50 1.300 NW 80 3.200 NW 100 4.900 NW 150 7.900 NW 200 10.700 Outlet tube 5 x D , 1 Tapered flange, 1 intermediate flange, 1 clamp. ring/chain 4) Nominal size Weight (kg) NW 25 0.320 NW 50 0.800 NW 80 1.850 NW 100 2.800 NW 150 4.600 NW 200 6.700 4) Outlet tube 5 x D , 1 tapered flange, 1 clamping ring/chain Nominal size Weight (kg) NW 25 0.250 NW 50 0.750 NW 80 1.750 NW 100 2.700 NW 150 4.300 NW 200 6.100 4) Hose adapter , 1 tapered flange, 1 clamping ring/chain Nominal size Weight (kg) NW 25 0.120 NW 50 0.320 NW 80 0.500 NW 100 0.630 NW 150 1.600 NW 200 2.000 Code EUR Deliv. time 4 wks.

7962669 7962670 7962671 7962672 7962673 7962674

163,00 210,00 363,00 410,00 788,00 908,00

7962675 7962676 7962677 7962678 7962679 7962680

173,00 231,00 295,00 388,00 908,00 1035,00

7962681 7962682 7962683 7962684 7962685 7962686

152,00 195,00 247,00 325,00 667,00 773,00

7962687 7962688 7962689 7962690 7962691 7962692

99,90 110,00 152,00 220,00 400,00 441,00

Ordering information individual parts


Catalog No. Tapered flange (please order O-ring separately) Nominal size Weight (kg) NW 25 0.020 NW 50 0.067 NW 80 0.112 NW 100 0.114 NW 150 0.353 NW 200 0.600 O-Ring (DN 25 includes centering rings) Nominal size Weight (kg) NW 25 0.012 NW 50 0.004 NW 80 0.005 NW 100 0.006 NW 150 0.015 NW 200 0.031 4) See dimensional drawings, Data Sheet 4) 7962700 7962701 7962702 7962703 7962704 7962705 4) 7962706 7962707 7962708 7962709 7962710 7962711 21,60 5,15 6,18 6,18 11,30 16,50 33,00 38,10 43,30 48,40 91,70 102,00 Code EUR Deliv. time 4 wks.

15

Thermal Mass Flowmeter FMT700-P (Sensyflow P) for air, test benches

10/14-6.84 EN

Ordering information individual parts


Catalog No. Intermediate flange Nominal size Weight (kg) NW 25 0.020 NW 50 0.068 NW 80 0.102 NW 100 0.111 NW 150 0.214 NW 200 0.610 Clamping ring Nominal size Weight (kg) NW 25 0.087 NW 50 0.180 NW 80 0.180 NW 100 0.230 Clamping chain Nominal size Weight (kg) NW 150 0.590 NW 200 0.720 Complete flange coupling, comprising of: 1 tapered flange, 1 intermediate flange, O-ring and 1 clamping ring/chain Nominal size Weight (kg) NW 25 0.137 NW 50 0.320 NW 80 0.430 NW 100 0.480 NW 150 1.210 NW 200 2.020 Filter cartridge Nominal size Weight (kg) NW 25 0.630 NW 50 1.400 NW 80 2.250 NW 100 4.000 NW 150 5.000 NW 200 5.000 Filter coupling flange Nominal size Weight (kg) NW 25 0.198 NW 50 0.420 NW 80 0.670 NW 100 0.740 NW 150 1.900 NW 200 2.450 Transducer cables Length Weight (kg) 3m 0.220 8m 0.490 15 m 0.860 30 m 2.560 Interface cable 25-pin connector (1 piece), length 3 m, weight 0.220 kg 4) See dimensional drawings, Data Sheet 4) 7962712 7962713 7962714 7962715 7962716 7962717 4) 7962718 7962719 7962720 7962721 4) 7962722 7962723 152,00 168,00 33,00 43,30 48,40 86,50 33,00 38,10 43,30 48,40 86,50 91,70 Code EUR Deliv. time 4 wks.

7962724 7962725 7962726 7962727 7962728 7962729

110,00 115,00 126,00 195,00 325,00 410,00

7962730 7962731 7962732 7962733 7962734 7962735

38,10 53,60 70,00 107,00 147,00 147,00

7962736 7962737 7962738 7962739 7962740 7962741

86,50 152,00 142,00 152,00 263,00 352,00

7962693 7962694 7962695 7962696 7962697

126,00 163,00 190,00 220,00 126,00

16

ABB has Sales & Customer Support expertise in over 100 countries worldwide. www.abb.com

The Companys policy is one of continuous product improvement and the right is reserved to modify the information contained herein without notice. Printed in the Fed. Rep. of Germany (11.04) ABB 2004

ABB Ltd. Oldends Lane, Stonehouse Gloucestershire, GL10 3TA UK Tel: +44(0)1453 826661 Fax: +44(0)1453 827856

ABB Inc. 125 E. County Line Road Warminster, PA 18974 USA Tel: +1 215 674 6000 Fax: +1 215 674 7183

ABB Inc. 3450 Harvester Road Burlington Ontario L7N 3W5 Canada Tel: +1 905 681 0565 Fax: +1 905 681 2810

ABB Automation Products GmbH Borsigstr. 2 63755 Alzenau Germany Tel: +49 551 905-534 Fax: +49 551 905-555 CCC-support.deapr@de.abb.com

10/14-6.84 EN Rev. A

You might also like