Specifications Summary
Resolution with Newton DU940 CCD 1200 l/mm @ 500 nm 2400 l/mm @ 300 nm Aperture Focal length Magnification @ centre of CCD Vertical, spectral flange Vertical, multi-track flange Gratings Slit width range (input/output) 0.10 nm 0.05 nm F/4 303 mm
1 1 Interchangeable indexed triple turret Motorised 10 m to 2.5 mm Wide aperture option to 12 mm USB 2.0
Silver-protected coated optics option Most efficient for Near-Infrared detection when used in conjunction with Andor InGaAs cameras
Communication
www.andor.com
Page 1 of 12
Shamrock 303i
Step-by-Step System Configuration
How to customise the Shamrock 303i :
(B)
Step 2. Resolution & band-pass
(A)
Step 3. Input light coupling interface Step 1. - Chassis configuration a) Select either a single output port (A) or dual output port (B) option. b) Select type of optics coating required (aluminium + MgF2 is standard, silver-protected coated optics available on request for NIR detection). c) Select purge port option (for extended detection below 180 nm), or shutter for background acquisition and detectors protection Step 2. - Resolution & band-pass a) Select the appropriate Shamrock spectrograph platform, giving due consideration to bandpass and spectral range requirement. b) Select gratings and detector to fulfill resolution requirements. c) Select gratings for suitable wavelength coverage. Step 3. - Input light coupling interface Refer to accessory tree for available configurations (direct coupling, fibre coupling or 3rd party hardware connectivity). Step 4. - 2nd exit port configuration Refer to accessory tree for available configurations, including camera flanges. Step 5. - Software interface Select either state-of-the-art Solis software or Software Development Kit (SDK) option please refer to appropriate section for futher information.
www.andor.com
Page 2 of 12
Shamrock 303i
Step 1 - Chassis Configuration Ordering Information
Model
SR-303i-A SR-303i-B SR-303i-X-SIL
Description
Shamrock 303i spectrograph base unit with single output port Shamrock 303i spectrograph base unit with dual output port Silver-protected coated optics for models shown above (replace X with relevant model number)
Chassis Accessories
Shutter (SR-SHT-9001)
www.andor.com
Page 3 of 12
Shamrock 303i
Green Aberration-free region Orange Possible impact on system resolution Red Likely impact on system resolution
Czerny-Turner spectrographs are designed to provide the best optical performance for a range of grating angles as reflected on the green parts of the graph above. Outside this range, the spectral lines may exhibit a degree of optical aberration (such as coma), which will become more prominent at the steeper angles. These configurations are reflected by the orange to red scales on the graph. In these regions, consideration should be given to higher spectrograph focal length models with lower groove density gratings to achieve the desired resolution.
Grating (l/mm) Shamrock 163 Bandpass (nm)2,4 Resolution (nm)3,4 Shamrock 303i Bandpass (nm)2,4 Resolution (nm)3,4 Shamrock 500i Bandpass (nm)2,4 Resolution (nm)3,4 Shamrock 750 Bandpass (nm)2,4 Resolution (nm)3,4 245 0.36 123 0.18 61 0.09 30 0.04 19 0.03 155 0.025 364 0.53 183 0.27 92 0.13 46 0.07 29 0.04 235 0.035 600 0.88 297 0.43 144 0.21 67 0.10 39 0.06 325 0.055 150 1072 1.57 300 529 0.77 600 256 0.37 1200 117 0.17 1800 (Holo) 68 0.10 2400 (Holo) 565 0.085
Where aberration is a concern for a particular experimental set-up, the table above shows resolution and band-pass performance for a variety of alternative configurations. This should be used in conjunction with the graph above to assist in selecting the most appropriate Shamrock spectrogrph platform to meet resolution and band-pass needs, whilst minimising the risk of potential aberration.
www.andor.com
Page 4 of 12
Shamrock 303i
Nominal Lines/mm Blaze (nm) dispersion (nm/mm)6 150 150 150 150 150 300 300 300 300 300 600 600 600 600 600 1200 1200 1200 1200 300 500 800 1250 2000 300 500 1000 1200 1700 300 500 1000 1200 1900 (@1600)7 300 500 1000 (@ 800)7 Holographic (500 nm peak) Holographic (250 nm peak) Holographic (500 nm peak) 300 Holographic (220 nm peak) Holographic (400 nm peak) UV-VIS VIS-NIR 21.79 21.70 21.56 21.32 20.87 10.83 10.73 10.43 10.30 9.91 5.34 5.22 4.82 4.62 3.71 4.15 2.58 2.41 1.77 2.07 2.41 1.68 1.43 1.16 1.23 1.04 -
Bandpass (nm)2,6 602 600 596 589 577 299 297 288 285 274 148 144 133 128 102 115 71 67 49 57 67 46 39 32 34 29 -
Resolution (nm)3,6,9 0.88 0.88 0.87 0.86 0.85 0.44 0.43 0.42 0.42 0.40 0.22 0.21 0.20 0.19 0.158 0.17 0.10 0.10 0.078 0.08 0.10 0.07 0.06 0.05 0.05 0.04 -
Andor part number SR3-GRT-0150-0300 SR3-GRT-0150-0500 SR3-GRT-0150-0800 SR3-GRT-0150-1250 SR3-GRT-0150-2000 SR3-GRT-0300-0300 SR3-GRT-0300-0500 SR3-GRT-0300-1000 SR3-GRT-0300-1200 SR3-GRT-0300-1700 SR3-GRT-0600-0300 SR3-GRT-0600-0500 SR3-GRT-0600-1000 SR3-GRT-0600-1200 SR3-GRT-0600-1900 SR3-GRT-1200-0300 SR3-GRT-1200-0500 SR3-GRT-1200-1000
6925
11300
3450
5651
1725
2826
865
1413
SR3-GRT-1200-EH SR3-GRT-1800-FH 540 SR3-GRT-1800-DH SR3-GRT-2400-0300 SR3-GRT-2400-BH SR3-GRT-2400-GH SR3-GRT-MR-AL-MGF2 SR3-GRT-MR-SILVER 405 706 942
www.andor.com
Page 5 of 12
Shamrock 303i
Step 2c - Selecting The Correct Grating Efficiency Option
All graphs shown below represent efficiency for 45 polarisation
www.andor.com
Page 6 of 12
Shamrock 303i
Part No.
Size
SR-ASM-0010
6 x 8 mm (W x H)
SR-ASM-0011
6 x 14 mm (W x H)
SR-ASM-001512
15 mm
Neutral Density Filters13 Long Pass Filters13 Short Pass Filters13 Raman Edge Filters13
SR-ASM-0016
11
6 x 4 mm (W x H)
SR-ASM-0017
6 x 6 mm (W x H)
Fixed FC Fibre Adapter (SR-ASM-8011) Fixed SMA Fibre Adapter (SR-ASM-8003) Y Adjustable Fast Kinetics Fibre Adapter, SMA Input (SR-ASM-8012)
Motorised Slit Assembly14 Motorised Wide Aperture Slit15 X-Y Adjustable Fibre Adapter, (FC: SR-ASM-8053)13 (Standard) (SR-ASZ-0034) (FC-APC: SR-ASM-805513 (Ferrule: SR-ASM-8057) (SMA: SR-ASM-8054) Cover Plate for Motorised Slit Assembly (Please see table on left)
X Adjustable Fibre Adapter, Ferrule Input (SR-ASM-8006) Fixed Fibre Adapter, Ferrule Input (SR-ASM-8001)
X-Y Adjustable Fibre Adapter, (Ferrule: SR-ASM-8010) (FC: SR-ASM-8056)13 (SMA: SR-ASM-8052)
1.5 Flange Adapter C-Mount Adapter for Newport Oriel (SR-ASM-0021) Accessories (SR-ASM-0002)
F-Mount Lens13
www.andor.com
Page 7 of 12
Shamrock 303i
X-Y Adjustable Fibre Adapter, (FC: SR-ASM-8053)13 (FC-APC: SR-ASM-8055)13 (Ferrule: SR-ASM-8057) (SMA: SR-ASM-8054) Exit Port Motorised Slit Assembly14 (SR-ASZ-0005)
Cover Plate for Motorised Slit Assembly (Please refer to table on right)
Cover Plate Apertures for Motorised Slit Part No. SR-ASM-001611 SR-ASM-0017 SR-ASM-0010 SR-ASM-0011 Size 6 x 4 mm (W x H) 6 x 6 mm (W x H) 6 x 8 mm (W x H) 6 x 14 mm (W x H) SR-ASM-001512 15 mm
X-Y Adjustable Fibre Adapter, (Ferrule: SR-ASM-8010) (FC: SR-ASM-8056)13 (SMA: SR-ASM-8052) C-Mount Adapter (SR-ASM-0021) Sample Chamber
(ACC-SR-ASZ-0056)13
Fibre Ferrule
(SR-OPT-80XX)13
iXon Mounting Flange (MFL-SR303i-IXON)17 iXon 888 Mounting Flange (MFL-SR303i-888)17 iXon 888 Mounting Flange Crop Mode & Fast Kinetics (MFL-SR303i-CROP)17
www.andor.com
Page 8 of 12
Shamrock 303i
Step 5 - Selecting A Software Option
The Shamrock 303i requires at least one of the following software options: 1 - Solis Spectroscopy A 32-bit application compatible with 32 and 64-bit Windows (XP, Vista and 7) offering rich functionality for data acquisition and processing, as well as Andor cameras, spectrograph and motorised accessories simultaneous control. AndorBasic provides macro language control of data acquisition, processing, display and export. 2 - Standalone Solis Spectroscopy GUI for standalone spectrograph operation 3 - Andor SDK A software development kit that allows you to control the Andor range of cameras from your own application. Available as 32 and 64-bit libraries for Windows (XP, Vista and 7). Compatible with C/C++, C#, Delphi, VB6, VB.NET, LabVIEW and Matlab.
(a)
(b)
(c)
(d)
(e)
(a) Shutter: Synchronisation mode selection for shutter operation (b) Slit drive: Control the spectrograph slit width - drag blades on icon or type in required slit width (c) Filter wheel: Used to select a particular filter on the filter wheel - just click on the desired filter position (d) Grating turret: Used for setting grating turret to a new position and bringing desired grating in the optical path - just click on the desired grating (e) Flipper motor: Used to select the appropriate exit port
Wavelength drive: Sets the wavelength range for the current grating - drag slider to desired wavelength or just type in appropriate value
Exposure time: Sets the exposure time for the detector - quick access for easy acquisition optimisation
www.andor.com
Page 9 of 12
Shamrock 303i
Product Drawings
Dimensions in mm [inches]
Optical Axis 127 mm [5] with pad feet The optical path height is shown with standard feet attached. The height with adjustable feet SR-ASM-0022 varies from 143 mm [5.6] to 173 mm [6.8]
Screw Type Requirements CCD flange to Spectrograph flange Camera to CCD flange 4 off, M4 x 16 4 off, M3 x 10
Applications Guide
Absorption-Transmission-Reflection Raman (Stimulated, Resonance, CARS, SERS, SORS) Fluorescence -Luminescence Micro-Fluorescence Photon Counting Single Molecule Spectroscopy
Shutter Specifications
Maximum repetition rate Minimum open/close time Minimum lifetime 2 Hz 15 ms Better than 100K cycles
Optical Property
Focal plane size (mm, W x H) 28 x 14
www.andor.com
Page 10 of 12
Shamrock 303i
Typical Setup - Spectroscopy
Lens
PC Newton EMCCD Lamp Software CD Microscope Feet Newton EMCCD Cage System
www.andor.com
Page 11 of 12
Shamrock 303i
Order Today
Need more information? At Andor we are committed to finding the correct solution for you. With a dedicated team of technical advisors, we are able to offer you one-to-one guidance and technical support on all Andor products. For a full listing of our local sales offices, please see: www.andor.com/contact Our regional headquarters are:
Europe Belfast, Northern Ireland Phone +44 (28) 9023 7126 Fax +44 (28) 9031 0792 North America Connecticut, USA Phone +1 (860) 290 9211 Fax +1 (860) 290 9566 Japan Tokyo Phone +81 (3) 3518 6488 Fax +81 (3) 3518 6489 China Beijing Phone +86 (10) 5129 4977 Fax +86 (10) 6445 5401
Items shipped with your spectrograph 1x 3 m USB 2.0 cable Type A Type B 1x Power supply with 3 m mains cable 1x PCI / I2C cable 1x Camera / spectrograph I2C cable 1x CD containing Andor user guides 1x Individual system performance booklet 1x CD containing either Solis software or SDK (if requested at time of order) 1x Allen key set (2 mm, 3 mm & 5 mm)
Regulatory Compliance Compliant with the requirements of the EU EMC and LV Directives, compliant with the international EMC and safety standards IEC 61326-1 and IEC 61010-1.
Minimum Computer Requirements: 3.0 GHz single core or 2.4 GHz multi core processor 2 GB RAM 100 MB free hard disc to install software (at least 1 GB recommended for data spooling) USB 2.0 High Speed Host Controller capable of sustained rate of 40MB/s Windows (XP, Vista and 7)
Operating & Storage Conditions Operating Temperature 0C to 30C ambient Relative Humidity < 70% (non-condensing) Storage Temperature -25C to 50C
ISO14001: 2004
EMS91062
ISO9001: 2008
FM40523
Windows is a registered trademark of Microsoft Corporation. Labview is a registered trademark of National Instruments. Matlab is a registered trademark of The MathWorks Inc.