GRS-16/GRS-10/GRS-2 GAMMA SPECTROMETER

Calibration Program

PEICalib

Operation Manual

Version 5.0 March 2006

Rev:

5.0

GRS-16/GRS-10/GRS-2 Operation Manual

WARNING TO USERS!
BE ADVISED THE CONTENTS OF THE CRYSTAL DETECTOR ASSEMBLIES REPRESENT A THERMALLY STABLE MASS. DO NOT OPEN THE CRYSTAL DETECTOR BOXES UNLESS THE INTERNAL TEMPERATURE OF THE DETECTOR ASSEMBLY IS THE SAME AS THE AIR TEMPERATURE OUTSIDE THE BOX. A TEMPERATURE DIFFERENCE OF MORE THAN 5 DEGREES CELCIUS BETWEEN INTERNAL AND EXTERNAL BOX TEMPERATURES CAN CAUSE THE CRYSTAL MASS TO CRACK. SIMILARILY A TEMPERATURE GRADIENT OF MORE THAN 10 DEGREES PER HOUR OF OUTSIDE AIR TEMPERATURE WILL EXCEED THE DETECTOR PACKAGE ABILITY TO MAINTAIN A SAFE TEMPERATURE GRADIENT INSIDE THE DETECTOR CONTAINER. BEFORE OPENING THE DETECTOR ASSEMBLY ENSURE THE OUTSIDE AIR TEMPERATURE HAS BEEN MAINTAINED AT A CONSTANT LEVEL FOR AT LEAST 24 HOURS. PICO ENVIROTEC ASSUMES NO RESPONSIBILITY FOR DECTECOR ARRAYS DAMAGED BY THERMAL SHOCK

2/29

Pico Envirotec Inc.

GRS-16/GRS-10/GRS-2 Operation Manual

Rev:

5.0

Content:
1. GRS-16/GRS-10/GRS-2 INTELLIGENT GAMMA SPECTROMETER
1.1. PURPOSE OF THIS PROGRAM AND MANUAL 1.2. OVERVIEW 1.3. DIFFERENCES AMONG GRS-16/GRS-10/GRS-2 GAMMA SPECTROMETERS 1.4. NOTE ON THIS MANUAL

4
4 4 5 5

2.

PEICALIB - CALIBRATION PROGRAM FOR GRS-16/GRS-10/GRS-2 GAMMA SPECTROMETER


2.1. GENERAL 2.2. PROGRAM REQUIREMENTS 2.3. PROGRAM LOADING 2.4. FINDING THE CONCENTRATOR 2.5. DOWN/UP DISPLAY
2.5.1. RESOLUTION CALCULATION 2.5.2. INDIVIDUAL DETECTOR DISPLAY

6
6 6 6 7 8
10 11

3.

CALIBRATION PROCEDURE
3.1. INITIAL GAIN ADJUSTMENT
3.1.1. EASY GAIN ADJUSTMENT 3.1.2. USING TH SAMPLE FOR INITIAL GAIN ADJUSTMENT 3.1.3. CALIBRATION START

12
12
12 12 12

3.2. LINEARITY CALIBRATION

13

4.

OTHER OPTIONS
4.1. DATA DISPLAY 4.2. DATA RECORDING 4.3. PROGRAM TERMINATION 4.4. FRONT-END ELECTRONIC TEST 4.5. SPECTROMETER VERIFICATION

14
14 14 14 14 14

5.

PRINCIPLES OF AIRBORNE GAMMA RAY SPECTROMETRY


5.1. GAMMA RADIATION
5.1.1. RADIOACTIVE DECAY 5.1.2. GAMMA RAY SPECTRA 5.1.3. INTERACTION OF GAMMA RAYS WITH MATTER

16
16
16 16 17

6.

QUALITY CONTROL
6.1. INSTRUMENTAL VARIABLES
6.1.1. 6.1.2. 6.1.3. 6.1.4. SPECTROMETER RESOLUTION SPECTRAL STABILITY DIGITAL DATA FIDELITY TEST LINE

20
20
20 21 22 22

6.2. OPERATIONAL VARIABLES


6.2.1. FLYING HEIGHT 6.2.2. FLIGHT PATH SPACING 6.2.3. FLYING SPEED

23
23 23 23

6.3. ENVIRONMENTAL VARIABLES


6.3.1. PRECIPITATION 6.3.2. ATMOSPHERIC RADON

24
24 24

APPENDIX A: GRS-16/GRS-10/GRS-2 TECHNICAL PARAMETERS APPENDIX B: RECOMMENDED DATA ACQUISITION SYSTEMS (IRIS/AGIS) APPENDIX D: REFERENCES APPENDIX E: CONTACTS

25 27 28 29

Pico Envirotec Inc.

3/29

Rev:

5.0

GRS-16/GRS-10/GRS-2 Operation Manual

1. GRS-16/GRS-10/GRS-2 INTELLIGENT GAMMA SPECTROMETER


This program operates the Pico Envirotec Inc. Intelligent GRS-16/GRS-10/GRS-2 Gamma Spectrometer supporting up to sixteen/ten/two detectors. GRS-16 or GRS-10 is intended to be used with or without the IRIS (Integrated Radiation Information System) or AGIS (Airborne Geophysical Information system) specific hardware. GRS-2 is used in PGIS Portable Geophysical Information System. Intelligent GRS-16/GRS-10/GRS-2 Gamma Spectrometer is designed to communicate via RS232 communication link. It is providing all necessary spectra correction (gain tracking and linearization) separately for each Gamma detector in real time. The calibration program allows the user to view individual detectors, adjust their parameters, calibrate them and verify the spectrometer operation. Simple communication protocol allows the GRS-16/GRS-10/GRS-2 to be connected to any data acquisition system via RS232 port. Simple calibration procedure and fully automated individual detector tracking simplifies the operation and reduces a possibility of an error introduced by the operator.

1.1.

PURPOSE OF THIS PROGRAM AND MANUAL

All tests, adjustments and calibrations described in this manual are related to the operational performance of the GRS-16/GRS-10/GRS-2 Gamma spectrometers. PEICALIB program provides means for verification of proper operation of the digital spectral detection. It does not provide physical description of properties of the Gamma spectrometer such as sensitivities and stripping constants. These parameters must be established by physical measurements on defined calibration places (calibration pads etc.).

1.2.

OVERVIEW

The GRS-16/GRS-10/GRS-2 Gamma spectrometer is an advanced Spectrometer utilizing NaI(Tl) detectors with individual detector handling. It is hardware-software designed system, exhibiting simplicity, easy interfacing and substantial versatility. Because of individual detector processing and use of the Digital Peak Detector that reduces nonlinearity and almost eliminates "zero base shift" and the "dead time"(see note 2.8.4.). This is achieved through digital processing of each detected Gamma particle (photon). Elimination of internal DC coupling further reduces above mentioned potential problems. New - natural peak detection algorithm provides safe and fast system stabilization without detector housing temperature stabilization and without implanted radioactive sources in the detector housing. Elimination of implanted sources (usually Cs137) for stabilization means no spectra pollution on low energies and therefore better sensitivity of the system for man-made isotopes. When calibrated (with Th source about once a year) linearity and zero offset of the each detector are measured and mathematical correction coefficients are calculated. When operating in real time (collecting data), the gain, linearity and offset of each detector is mathematically corrected for each measurement. Individual detector tracking (tuning) and linearity correction provide better fit of the individual spectra that are being summed and therefore sharper (better resolution) spectrum is obtained. Optionally the GRS-16/GRS-10/GRS-2 system can be controlled by the altitude of the aircraft and calculate absolute values of contamination by individual radionuclei related to the ground and provide the dose rate related to 1meter above the ground. Interfacing via single RS232 communication channel makes the system very flexible. This manual describes PEICALIB program.

4/29

Pico Envirotec Inc.

GRS-16/GRS-10/GRS-2 Operation Manual

Rev:

5.0

1.3. DIFFERENCES AMONG GRS-16/GRS-10/GRS-2 GAMMA SPECTROMETERS


GRS-16 is able to handle maximum sixteen detectors with summed output for in real time recording of 256 or 512 channels. GRS-10 is able to handle maximum ten detectors with same output as GRS-16 but can as well produce individual spectra for in real time recording on 256 or 512 channels or summed output for 256 channels two times a second. GRS-2 is able to handle maximum two detectors with summed or individual output for in real time recording of 256 or 512 channels. PEICALIB program always uses 256 channels mode.

1.4.

NOTE ON THIS MANUAL

Continuous work on the GRS-16/GRS-10/GRS-2 performance improvement may cause that the manual may slightly differ from the delivered software version. Any major changes in operation will be indicated and supplied with the delivered product. Should there be some differences found in the manual and real operation, manufacturer would appreciate reporting such discrepancies. Generally the hardware of GRS-16/GRS-10/GRS-2 is referred as the CONCENTRATOR.

Pico Envirotec Inc.

5/29

Rev:

5.0

GRS-16/GRS-10/GRS-2 Operation Manual

2. PEICALIB - CALIBRATION PROGRAM FOR GRS-16/GRS10/GRS-2 GAMMA SPECTROMETER


2.1. GENERAL
PEICALIB program is designed to verify performance of the "black box" type of an intelligent spectrometer. It is supplied together with GRS-16/GRS-10/GRS-2 hardware containing a Concentrator unit. Program allows the operator to: Set remotely High voltage for each sensor Turn remotely each sensor on or off Appoint it as an "downward" or "upward" looking detector Set the system threshold Calculate the resolution of the system Observe: Each detector independently All detectors in two groups -down/up Uncorrected/Corrected/Tracking spectra can be displayed Individually collected (one second) or time summed spectra may be selected Calibrate individual detectors for linearity with the help of a Tl208 radioactive source In case of verification of the system operation on calibration pads (portable or stationary) collected data may be stored for verification by standard program All calculated or pre-set parameters are automatically stored and used at any time after when the GRS-16/GRS-10/GRS-2 is used either in operation with this program or any data acquisition system.

2.2.

PROGRAM REQUIREMENTS

PEICALIB calibration program is designed to run on IBM-PC compatible computers equipped with Windows 98 or XP operating system operating at minimum 66 MHz. The computer running calibration program has to be connected via COM port with GRS-16/GRS-10/GRS-2 hardware containing a Concentrator unit.

2.3.

PROGRAM LOADING

For operation safety a copy of the program is supplied on a CD. Keep it in a safe place. The PEICALIB.EXE may be copied into any directory and run from that directory. Once initiated the Fig. 1 appears. The form shows which COM port and baud rate is used to detect the Concentrator. Those parameters were restored from previous work session. PEICALIB.INI file is used to store parameters for the next run of the program. If there is no PEICALIB.INI file detected the program will use default parameters: port COM1, 115200 baud.

6/29

Pico Envirotec Inc.

GRS-16/GRS-10/GRS-2 Operation Manual

Rev:

5.0

Fig. 1

Fig. 2

2.4.

FINDING THE CONCENTRATOR

If the Concentrator is connected to the different COM port or if it is using different baud rate press <Cancel> button, otherwise press <OK> button. If <OK> button was pressed the program would look for the Concentrator unit connected to the selected s baud rate erial communication port and the Fig. 2 would appear. For GRS-2 type Concentrator baud rate 57600 baud required, for GRS-16/GRS-10 baud rate 115200 baud required. The Concentrator communicates with the host and at the same time communicates with individual (up to sixteen) detectors as well as the optional four-channel analog to digital converter. The search number is indicated at the top of the display box "Times nn". When the Concentrator is detected the revision of the Concentrator will be displayed (see Fig. 3), otherwise the Fig. 2 will stay on until <Cancel> button is pressed. If <Cancel> button is pressed then the revision Fig. 3 will be displayed but <EMULATION> checkbox will be checked. This means that the program will simulate the Concentrator actions. If the Concentrator is connected to another port, it is possible to select another COM port and baud rate on the form, uncheck <EMULATION> checkbox and click <UPDATE> button. The system will return to the Fig. 2 screen. This can be repeated till the Concentrator is detected.

Fig. 3 When the program is switched to simulation mode the program does not perform any data corrections or controls of the Concentrator (no serial communication are not physically executed). The detector orientation, high voltage and the threshold can be changed using visual controls. To transfer changes to the Concentrator <UPDATE> button has to be applied. The box <Sp ON> is used to power off GRS-2 spectrometer via serial port. It is not used for GRS-16/GRS-10 spectrometers. To proceed to spectra data collection display press <START> button. The button label will be changed to <STOP>. To terminate data collection apply <STOP> button.

Pico Envirotec Inc.

7/29

Rev:

5.0

GRS-16/GRS-10/GRS-2 Operation Manual

2.5.

DOWN/UP DISPLAY

All down and up looking detectors are summed into separate spectra and displayed as shown on Fig.4. The blue color trace represents down looking spectra and the red color trace represents up looking spectra. Total counts for both spectra are printed with appropriate colors. Traces are updated every second with new data. To stop communicating between the host computer and the Concentrator press <STOP> button. <SCALE> control allows changing visual scale of the traces. The checkbox allows applying an auto scale mode. <Single-Accumulate> control allows switching between accumulation (data is summed every second) and single second measurements.

Fig. 4 Clicking on the mouse left button close to the graph draws a black cursor. It can be moved left or right by the <CURSOR> [channel] control. If the cursor is visible the appropriate channel value is displayed and updated each second when the traces are updated (Fig.5). If <RESOLUTION> checkbox is checked the peak resolution is displayed beside.

8/29

Pico Envirotec Inc.

GRS-16/GRS-10/GRS-2 Operation Manual

Rev:

5.0

Fig. 5 Cursor is inhibited from channel 1 to channel 4, since those channels are used for internal functions. PEAK ID checkbox indicates significant peaks positions visible (Fig.6 shows the single detector spectra measuring Th (Tl208) sample in one-second readings).

Pico Envirotec Inc.

9/29

Rev:

5.0

GRS-16/GRS-10/GRS-2 Operation Manual

Fig. 6

2.5.1.

Resolution calculation

Resolution is based on half peak width calculation, with background removed. If the statistics of the peak is not acceptable the resolution will show 0.0 and if the peak is not defined properly it will show 99.9. For proper calculation adequate number of pulses should form the peak. Background is subtracted and the peak is analyzed. Peak position with background subtracted and the resolution are displayed in the last two columns. Caution should be exercised because of the automated simple background removal. The resolution automatically calculated does not represent absolute resolution of the detector. Resolution of the same detector measured with different instruments may vary. Consistency of the measurement assures that if the detector is measured with the same instrument any changes in the quality of the detector can be detected. It should be used for comparison only. See definition of resolution in 6.1.1. Differences of the resolution from individual measurements within +- 0.3% are acceptable, as well as slightly lower resolution than the average of individual detector resolutions when summed together. STATUS tab on the form shows details of the crystal parameters and if equipped with analog converter the analog channel readings (Fig. 7). STATUS column indicates the state to the gain control algorithm for each detector

10/29

Pico Envirotec Inc.

GRS-16/GRS-10/GRS-2 Operation Manual

Rev:

5.0

Fig. 7 Status T U K A N n 0 Meaning Gain latched on Thorium Gain latched on Uranium Gain latched on Potassium Gain controlled from almanac No gain adjustment - wait Error in spectra last spectrum repeated Detector not connected

2.5.2.

Individual detector display

In order to be able to adjust and verify individual detectors the program must be switched to Individual Detector Display by selecting menu item View/Individual/All. Each detector is presented by its own trace. All controls are the same as for Down/Up display. The table displayed beside traces describes some detector parameters or statuses: the X column shows detector number; the TC column shows total count for the detector; the State column shows the state of gain adjustment If only one detector trace needs to be shown select View/Individual/DetectorX menu item, where X is a number of the appropriate detector (Fig.5). Menu items Source/Input, Source/Output, Source/Tune are used to switch between data displayed. To return to Down/Up display menu item View/ View UpDown has to be applied.

Pico Envirotec Inc.

11/29

Rev:

5.0

GRS-16/GRS-10/GRS-2 Operation Manual

3. CALIBRATION PROCEDURE
Begin data collection by pressing <START> button. The screen will display spectra trace from data collected (if Tab <SPECTRA> is selected) and will be updated once a second. To calibrate individual detectors it is necessary to check <Std.Clbr.> checkbox for standard crystals or <Small Clbr.> checkbox for small crystals. The screen will be switched the program to the Individual Detector Display (Fig.8). It is necessary to use Th (Tl208) sample (approximate 5 micro Currie) with low self-absorption of low energies. Note: Small detector is considered detector volume of less than approximately 1litre (60 cci). If the box contains DOWN looking detectors only than one sample placed above or beneath the box and the distance adjusted to Total count measuring about max 4000(+/-200) for large detectors and about 2000(+/-140) cps for small detectors. is adequate. If the box contains Up looking detectors two samples may be necessary to assure that all the detectors are well exposed to the radiation of the calibrating samples. Th sample is used because of its multi energy radiation peaks. The calibration procedure consists of two parts: Initial Gain adjustment Linearity calibration.

3.1.

INITIAL GAIN ADJUSTMENT

Before the program is switched into calibration mode, individual input spectra mode should be selected. Experienced operator may skip the next paragraph.

3.1.1.

Easy gain adjustment

Because of single energy peak, Cs137 sample (less than 5 microCu) may be used to establish the setting of the detector. The Cs peak of the input (uncorrected) spectra should be located approximately around the channel 50. The initial gain adjustment is achieved by the High Voltage (HV) change. When the HV is adjusted up or down in the Status Window, it has to be transmitted to the Concentrator. After the high voltage application you will see the shift of the peak and readjustment may be necessary. Once the Cs peak is located in the approximate position <48..52> Cs sample should be removed.

3.1.2.

Using Th sample for initial gain adjustment

The Th sample should be placed close to the detector container that the Th (Tl208) 2.82MeV sample is visible. The position of this peak should be located between <180...185> by the adjustment of the HV. The operating range of the gain correcting algorithm is from channel 160 to 230 referring to the Th sample.

3.1.3.

Calibration start

The Th sample should be placed (observe max total cont.) close to the detector container in such a position that the total counts on individual detectors are similar and limited to the counts mentioned in 3. After a short period of time the Th peak will be detected. Once stabilized the process can system will detect parameters of the detector and proceed with automated calibration.

12/29

Pico Envirotec Inc.

GRS-16/GRS-10/GRS-2 Operation Manual

Rev:

5.0

3.2.

LINEARITY CALIBRATION

Message Calibration: CALIBRATING is displayed on the top of the form. For each crystal being calibrated a progress bar is displayed with label X1...X16 (Fig.8). When all detectors are calibrated screen will display message Calibration: FINISHED (Fig.9). When the Coef3 coefficient reaches 1.00 the detector is calibrated. Unchecking <Std.Clbr.> (or <Small Clbr.>) checkbox terminates the coefficient calculation process. Calculated coefficients are stored in an almanac file in the Concentrator unit and used in real time spectrum tracking and correction. Correction of each detector provides better fit of each detector improving resolution mainly on the low energy side. After the calibration, spectral adjustment should be checked with a Cs sample. With spectra switched to accumulation and output mode. Calibration Progress bar

Fig. 8

Fig. 9

Pico Envirotec Inc.

13/29

Rev:

5.0

GRS-16/GRS-10/GRS-2 Operation Manual

4. OTHER OPTIONS
4.1. DATA DISPLAY
Program may request and display three different spectra: Raw input spectra, Corrected output spectra and Tracking spectra. Raw and corrected spectra may be displayed as accumulated or one-second spectra. The Tracking spectra are already accumulated in the Concentrator. This spectrum serves for detection of natural reference peaks.

4.2.
PEIVIEW

DATA RECORDING

Acquired data can be recorded in the standard PEI data file format and reviewed by the program using File/Record data menu item. Recording will be stopped if an operator is changing data requesting format (All UP-Down Individual). To stop recording manually simply click File/Record data menu item again. In all cases the data file will be closed. File name is generated with same rules as it is done by AGIS application.

4.3. 4.4.

PROGRAM TERMINATION FRONT-END ELECTRONIC TEST

To terminate the calibration program simply close main form.

This program may be used by the expert users to test the front-end detector electronics. For this program application please contact the manufacturer.

4.5.

SPECTROMETER VERIFICATION

Characteristic of the Spectrometer calibration can be verified with Verify menu item. The verification can be based on Cs137 reference (Cs137 sample is required) or K40 reference (no samples required). To perform the verification it is necessary to create a reference file after the standard calibration is done by clicking <Verify / Create new statistics> menu item. The form Fig.10 will appear. Select a reference element and fill up the Client Name and location fields and click OK.

Fig. 10

14/29

Pico Envirotec Inc.

GRS-16/GRS-10/GRS-2 Operation Manual

Rev:

5.0

The form Fig.11 will appear (the K reference was selected to produce the screenshot).

Fig. 11 If you have a Cs sample and want to create a new statistics file with it you have to place your sample close to the detector container now. First the program will wait for all crystals to be Thorium tuned, and then it will collect the appropriate statistics during several minutes. Wait until the verification form is closed. The appropriate reference file CS_STAT_{YYMMDD}.txt or K_STAT_{YYMMDD}.txt will be created on the hard drive, where {YYMMDD} is a string containing two last digits of the year, two digits of the month and two digits of the day of the file creation. The message Statistics was stored will come. After the file was created with spectrometer properly calibrated it is possible to check periodically the spectrometer status using Verify menu item. To do it is necessary to select <Verify / Cs sample> menu item and place a Cs sample close to the detector container if the file was created with Cs sample, or choose <Verify / No samples> menu item if K40 reference was used to create a reference file. A form with progress bar will reflect the verification progress. The message Verification is successful or Verification failed! will notify about verification result. In case of verification failure it is necessary to recalibrate the spectrometer.

Pico Envirotec Inc.

15/29

Rev:

5.0

GRS-16/GRS-10/GRS-2 Operation Manual

5. PRINCIPLES OF AIRBORNE GAMMA RAY SPECTROMETRY


(Extract from the document: Airborne Gamma Ray Spectrometer Surveying 1991.) This section outlines briefly the physics of gamma rays and the principles of instrumentation used to detect them. The emphasis is on aspects relevant to the practical details of AGRS. For detailed accounts of these topics the reader should consult the texts noted in the Bibliography. Pico Envirotec advanced spetral stabilization technique and advanced QC software makes some of the below described techniques redundant, but for the completness of the understanding of AGRS they are included and it is upon the user or contract to ensure properly collected data.

5.1.

GAMMA RADIATION
5.1.1. Radioactive decay

There are many naturally occurring radioactive elements. However, only three have isotopes that emit gamma radiation of sufficient intensity to be measured by AGRS. These three major sources of gamma radiation are: (a) Potassium-40 which is 0.011 8% of total potassium, (b) Daughter products in the 238U decay series, (c) Daughter products in the 232Th series.

Many man-made radioactive isotopes also emit gamma radiation which can be measured by AGRS. These man-made isotopes are produced by nuclear reactors or are the result of atomic weapons testing programs. High energy cosmic rays produced outside the Earths atmosphere can also be detected by AGRS. This cosmic radiation interacts with the molecules of the atmosphere, the aircraft structure and the detector itself to produce a variety of high energy radiation. This cosmic ray component increases exponentially with the height above sea level.

5.1.2.

Gamma ray spectra

The energies of gamma rays produced by radioactive decay are characteristic of the decaying nuclide. For example 40K decays to 40Ar with the emission of gamma rays at 1460 keV. Gamma ray spectrometers are designed to measure the intensity and energies of gamma rays and hence the abundance of particular radioactive nuclides. Figures 1, 2 and 3 show the gamma ray spectra for potassium, the uranium series and the thorium series. The spectra were obtained with a typical airborne spectrometer system on large concrete calibration pads at Walker Field, Grand Junction, Colorado in the USA. The concrete potassium, uranium and thorium. Figure 4 is a typical airborne spectrum showing gamma ray peaks from all three radioelements. The energy windows used to detect the gamma rays from potassium and the uranium and thorium decay series are shown in Figs 14 and it can be seen that each window contains some contribution from all three radioelements. Owing to gamma ray scattering in the ground, the aircraft structure and the detector, some counts from 2614 keV 208T1 photons from a pure thorium source
16/29 Pico Envirotec Inc.

GRS-16/GRS-10/GRS-2 Operation Manual

Rev:

5.0

are recorded in the lower energy potassium and uranium windows. Counts in these lower energy windows can also arise from low energy gamma ray photons in the thorium decay series. Similarly, counts will be recorded in the lower energy potassium window from a pure uranium source and can also appear in the high energy thorium window owing to high energy gamma ray photons of 214Bi in the uranium decay series. As a result of the poor resolution of sodium iodide detectors, counts can also be recorded in the uranium window from a pure potassium source. A correction procedure, known as stripping, must be made to gamma ray spectrometer data to compensate for this spectral overlapping.

5.1.3.

Interaction of gamma rays with matter

It is clear from Fig. 1 that the monoenergetic spectral lines emitted during decay have been smeared and broadened by the time they are recorded by an airborne spectrometer. These broadened lines are generally called photopeaks and are the result of the limited resolution of the spectrometer. The gamma rays also interact with material in the ground and in the intervening air before reaching the detector. These interactions, as well as those within the detector itself, have a significant effect on the measured gamma ray spectrum. Gamma rays interact with matter by several mechanisms including the photoelectric effect and Compton scattering. In the photoelectric effect the whole energy of a low energy gamma photon is given up to an atomic electron. In Compton scattering, gamma rays lose part of their energy to electrons and are scattered at an angle to their original direction. Because both these effects involve electrons, the attenuation of gamma rays in a particular material is proportional to its electron density. A third effect is pair production, in which the whole energy of a gamma ray is lost in the production of an electronpositron pair. This process predominates at high energies, particularly in materials with high atomic numbers, and is a significant process in the absorption of high energy gamma rays in sodium iodide detectors. Because most materials (rocks, soils, air and water) encountered in airborne radioactivity measurements have a low atomic number and because most natural gamma rays have moderate to low energies (less than 2614 keV), Compton scattering and the photoelectric effect are the predominant absorption processes occurring in the ground and in the air. Since both these processes involve interactions with electrons, the attenuation of gamma rays in most materials is proportional to the electron density of the materialIn airborne spectrometry,the absorbtion of the gamma rays from the ground by the air beneath the aircraft must be taken into account.

Pico Envirotec Inc.

17/29

Rev:

5.0

GRS-16/GRS-10/GRS-2 Operation Manual

18/29

Pico Envirotec Inc.

GRS-16/GRS-10/GRS-2 Operation Manual

Rev:

5.0

Pico Envirotec Inc.

19/29

Rev:

5.0

GRS-16/GRS-10/GRS-2 Operation Manual

6. QUALITY CONTROL
(Extract from the document: Airborne Gamma Ray Spectrometer Surveying 1991.) Three types of variables must be monitored or checked periodically to ensure high quality radiometric data. Instrumental variables include detector gain settings, spectral stability and the fidelity of digital records, Operational variables, which include flight path position and flying height, are similar to the operational variables of any airborne geophysical survey. The main environmental variables which affect spectrometer surveys are weather conditions. The quality control goals and procedures described in this section apply mainly to spectrornetric mapping of the distribution of natural radioelements or of fallout. Search procedures are not likely to require the same rigorous quality checking. Sample specifications for a typical radioelement mapping survey are given in the Appendix.

6.1.

INSTRUMENTAL VARIABLES
6.1.1. Spectrometer resolution

Resolution is a measure of the precision with which the energies of gamma rays can be measured by the spectrometer. The resolution will be poor if the gain setting of any of the detectors is faulty or if one of the detectors is damaged. Resolution is measured using the 662 keV gamma rays from a 137Cs source. A spectrum is plotted as shown in Fig. 10. The amplitude of the peak due to YCs is found and the width of the peak (as a number of channels) at half maximum amplitude is measured. This is defined as the full width at half maximum, or FWHM. The resolution is then calculated as: R% = 100 (keV per channel) FWHM/662 keV For quality control purposes during survey operations, the resoluon should be found each morning immediately after any detector gain adjustments have been made. The resolution should also be determined after work each day, without making any further gain adjustment. The second value will show if there is excessive drift in any detector, owing to instrument problems such as poor temperature stability, or an electronic fault. Resolution should be 8.5 to 9.5% and must never exceed 12%. The results of tests should be recorded in a table or on a graph as the survey progresses and should be included in the operations report.

20/29

Pico Envirotec Inc.

GRS-16/GRS-10/GRS-2 Operation Manual

Rev:

5.0

Spectrum showing the 137CS peak used for determining the resolution of a system (FWHM full width at half maximum).

6.1.2.

Spectral stability

Airborne spectrometers are very stable and it is unusual for sufficient drift to occur which would affect the results significantly. However, drift can occur if the temperature of the crystals changes or if electronic faults occur in the instrument. For this reason it is important to monitor spectral stability to ensure data quality. If full spectral, 256 channel data, are recorded, the best way to check spectral stability is to plot spectra summed over segments of the survey data. Typically, spectra summed over bOOs are used. During this period the flight line will cover a range of geology, so peaks of all three main radioelements should occur. The spectral plots should be made on a field computer if possible, to provide the check as soon as possible. Each plot should be checked to see that the K and Th peaks lie in the correct channels (2 channels) and that the peaks are not unusually wide. If any one of the criteria is not met, the instrument should be thoroughly checked as a fault has probably occurred. Any flight lines affected should be reflown.

Pico Envirotec Inc.

21/29

Rev:

5.0

GRS-16/GRS-10/GRS-2 Operation Manual

A second check of stability is provided by daily source checks made on the ground before survey work each day. These also provide a check of the spectrometer sensitivity. Source checks are mandatory if only four window data are being recorded and are normal procedures for all natural radioelement mapping surveys. The sources must always be placed at exactly the same point relative to the detectors in the aircraft, using a rigid locating frame if possible. The radioactive homogeneity of the airfield apron where the source checks are to be carried out should be investigated and, if necessary, a particular location occupied for every source check. A set of dead time determinations should be made at this location, for the background and for each source. Dead time is found in the manner described in Section 4 for calibration pads. A standard daily source check procedure is composed of: a 60 s background recording, with sources placed at least 30 m from the aircraft, Th source recording, U source recording and 60 s background recording (repeated). The U and Tb recordings should be carried out over sufficient time to accumulate at least 10 000 counts in the U window for the U source and in the Tb window for the Th source. The digital average of each recording should be determined, on a field computer if possible, and dead time corrected. The average of the two background recordings must be subtracted from each source reading, and the results plotted against time over the survey period for inclusion in the operations report. If a source check gives results which differ by more than 5% from the mean of checks to date, the cause of the change should be investigated.

6.1.3.

Digital data fidelity

The digital data should be checked as soon as possible to ensure that all instruments are functioning and the data system and recording system are working correctly. If possible, profiles should be plotted from the digital data and examined to identify any spikes, gaps or other problems in the data. If a field computer is available, it should be used for this checking so that any faults can be identified as early as possible.

6.1.4.

Test line

If upward detectors are not used, regular flights should be made at survey altitude over water or over a repeatable test line over land, in order to monitor atmospheric radon variations. An overland test line also provides a daily dynamic check of instrument performance and some indication of the effects of environmental variables such as rainfall. If a body of water at least 2 km long and 0.5 km wide is available, then flights over water can be used to provide a direct estimate of radon background. If upward detectors are used, then flights over are required to determine the coefficients for the calculation of radon background. Separate overwater flights are not necessary if many lakes are crossed during the survey as, for example, in northern Canada. If no suitable body of water is available, a test line should be chosen to be logistically convenient, easily repeatable and as far as possible typical of the survey area. These requirements are usually fulfilled by using a straight road or railway close to the aircraft base. The line should be about 5 to 8 km long, equivalent to a flying time of about 100 s. The start and end of the line should be marked by features which can be clearly identified on the tracking film or video. If possible the radio-element concentrations should be more or less uniform along the length of the line. It is worth taking a little trouble to get the best line possible. When flying the test line at survey altitude, one should take care to maintain constant flying height and the correct flight path. When the data are returned to the field base, the fiducials of the start and end points of the line should be determined. and the averages of the various radiometric
22/29 Pico Envirotec Inc.

GRS-16/GRS-10/GRS-2 Operation Manual

Rev:

5.0

and altimeter data determined over the. interval using the field computer. The data should be corrected for dead time, as well as cosmic and aircraft background. The results should be plotted against time over the survey period and the plots included in the operations report. Use of test line results to estimate radon background is described in the section on data processing. High level test lines, flown at 800 m above ground level, have been used to estimate radon background, but this procedure cannot be recommended. The radon measured at this level may not be representative of the value at survey height and there may also be a significant contribution from the ground if the area contains granites or other radioactive rocks,

6.2.
6.2.1.

OPERATIONAL VARIABLES
Flying height

Flying height at AOL is an important operational variable, because gamma rays are attenuated by air, and corrections must be made for variations in flying height. The normal acceptable variation in flying height is 20% of the nominal height, that is from 110 to 135 m for a nominal survey height of 120 in. In hilly or~ mountainous terrain, it may not be safe to remain within these limits, and pilots must~ use skill and judgement to provide the best results without endangering the aircraft. As a general rule, spectrometric data obtained at a height greater than 250 m over a distance greater than 12 km will be of little value. To monitor flying height, one should examine the profiles of radar altitude and any areas where the flying height is out of specification should be discussed with the pilot. Provided aircraft safety permits, these tines should be retlown or infill lines introduced. In some cases, deviation of the flight line may be preferable to a large deviation from nominal flying height.

6.2.2.

Flight path spacing

Each survey flight should be plotted onto the navigation maps as soon as possible. Any out of tolerance flying should be identified quickly so that infill or repeat flights can be specified. The tolerance normally permitted for flight path spacing for natural radio-element mapping is 150% of the nominal spacing over a maximum distance of 5 kin, or 200% of nominal spacing at any point. For 1 km spacing, this means that any gap greater than 1.5 km x 5 km between two adjacent lines must be infilled, and that any flight lines more than 2 km apart are out of tolerance. Safety considerations override the specifications for flight line spacing. Flight path spacing is of importance in a search for radioelements as poor navigation can result in a target being missed. The same tolerances as those used for mapping surveys can be set and must be taken into account when deciding the optimum search pattern.

6.2.3.

Flying speed

The aircraft speed is rarely the cause of problems in spectrometric surveying. The area of ground sampled by the detector during each second will be greater as the speed increases and a point source anomaly will be reduced in amplitude. For highly detailed mapping surveys, a maximum acceptable speed may be set.

Pico Envirotec Inc.

23/29

Rev:

5.0

GRS-16/GRS-10/GRS-2 Operation Manual

6.3.

ENVIRONMENTAL VARIABLES
6.3.1. Precipitation

Rain affects the results of a natural radioelernent mapping survey because waterlogged soil attenuates radiation from the ground. Areas of recent heavy rain ~ should therefore be avoided during surveying. Rain during a fallout mapping survey will bring down more radioactive dust f4~ onto the ground. Contamination of the survey aircraft by fallout in rain, either while ~ the plane is in flight or on the ground, should be avoided. Snow forms a radiation attenuating blanket over the ground: l0 cm of fresh snow is equivalent to about 10 m of air. Mapping surveys should be discontinued if there are more than I or 2 cm of snow on the ground.

6.3.2.

Atmospheric radon

As discussed in the section on calibration, there are ways to estimate and remove the effects of atmospheric radon. However, the problems caused by radon are greatest when temperature inversion conditions occur, as the radon becomes trapped beneath the inversion. If possible. these conditions should be avoided, particularly if no upward detectors are used.

24/29

Pico Envirotec Inc.

GRS-16/GRS-10/GRS-2 Operation Manual

Rev:

5.0

APPENDIX A: GRS-16/GRS-10/GRS-2 TECHNICAL PARAMETERS


256 channels optional 512. 1sec and longer. 0.5sec optional 47keV to 3MeV with threshold adjustable from 47keV to 300keV. @255 channels 11.7keV. All energies above 3MeV are detected as Cosmic Rays in channel 255. Anticoincidence: For improvement peak-valley ratio on lower energies, coincidental pulses detected among neighboring detectors are removed and placed in channel 0. Spectra tracking: Fully automatic on natural radionuclei. Independent tracking for individual detectors - Extended range via fine control of high voltage (resolution 0.3V). Time to stabilization: Usually less than 30 seconds on the ground and less than 3 minutes in the air at 100m altitude (based on 4 liters of individual detector volume). In case of a power failure old tracking parameters (almanac) are used till new tracking is reestablished. Spectra correction: Automatic after system calibration. Calibration is required once a year or when a detector is replaced. Max number of detectors: 10 detectors controlled by one GRS-10 concentrator, 16 detectors controlled by one GRS-16 concentrator, 2 detectors controlled by one GRS-2 concentrator. Each detector can be specified as down or up looking or it can be remotely turned off. Signal sampling: 25 MHz by an internal 12bit A to D for each detector. Quantizing error: The least significant bit at 25MHz sampling flash 12 bit analog to digital converter Peak detector: Digital - time resolution 40nsec. Dead time: Insignificant for less than 2000cps per detector (2.8.4) Maximum pulse rate: > 30000cps per each detector. If more than one detector is used substantial number of coincidental events is recorded in channel 0. These events are part of the maximum pulse rate per detector. Maximum channel capacity: 65500 counts Optional Outputs: Absolute activities and radiation exposure can be calculated in real time. (Aircraft altitude must be transmitted to the GSR410.) Communication: Serial from the detector unit to Concentrator. Serial from Concentrator to Data acquisition system. Serial from Concentrator to Superconcentrator and Data acquisition system. Version 12V voltage: 9 to 18V, Version 24V voltage: 18 to 36V, 3.5W per detector 12W per concentrator. Spectra resolution: Data sampling: Energy spectra: Threshold: Energy channel width: Cosmic Rays:

Power requirements:

Pico Envirotec Inc.

25/29

Rev:

5.0

GRS-16/GRS-10/GRS-2 Operation Manual

Operating temperature: Recommendation:

-10 to 55 C. For one-second measurements it is not recommended to exceed 10,000 counts per each input detector. In multidetector systems higher counts produce substantial coincidental events detected in channel 0, therefore hiding increased performance of the system. Very high counts may cause some instability in HV power supply producing nonstable system calibration. This process is not a destructive one and returns to normal when returning to normal count rates.

26/29

Pico Envirotec Inc.

GRS-16/GRS-10/GRS-2 Operation Manual

Rev:

5.0

APPENDIX B: RECOMMENDED SYSTEMS (IRIS/AGIS)

DATA

ACQUISITION

Airborne or truck-borne system IRIS (Integrated Radiation Information System) and AGIS (Airborne Geophysical Information System) support fully the Gamma Spectrometer GRS-16/GRS10/GRS-2. Both systems include precise DGPS navigation with flight path guidance and complete real time data acquisition.

This picture presents gamma spectrometers connected to the IRIS/AGIS main console. Appendix D: 44

Pico Envirotec Inc.

27/29

Rev:

5.0

GRS-16/GRS-10/GRS-2 Operation Manual

APPENDIX D: REFERENCES
Aviv, R., and Vulcan, U., 1983. Airborne gammaray survey over Israel: the methodology of the calibration of the airborne system. Israel Atomic Energy Commission, Report No. Z.D. 58/82 Dickson, B.L., and Lovborg, L-, 1984. An Australian fiicility for the calibration of portable gammaray spectrometers. Exploration Geophysics, 15(4), pp. 260263. Grasty, Ri., 1979. Gamma ray spectromete methods in uranium exploration theory and operational procedures; in Geophysics and Geochemistry in the Search for Metallic Ores; Peter 5. Hood, editor; Geological Survey of Canada, Economic Geology Report 31, pp. 14716 1. Grasty, R.L., 1982. Utilizing experimentally derivrd multichannel gaxnn]aray spectra for the analysis of airborne data. Uranium Exploration Methods, OECD, Paris, pp. 653-669. Grasty, RE, 1987. The design, construction and application of airborne gammaray spectrometer calibration pads Thailand; Geological Survey of Canada, Paper 8710, 34p. Grasty, R.L., Wilkes, P.O., and Kooyman, R., 1988. Background measurements in gammaray surveys. Geological Survey of Canada, Paper 8811. Grasty, R.L., Holman, P.B., and Blanchard, 1, 1991. Transportable calibration pads for ground and airborne gammaray spectrometers, Geol Sun. Can., Paper 9023, 26 p. Green, Ak, 1987. Levelling airborne gammaradiation data using betweenchannel correlation information. Geophysics, 52(11), 15571562. IAEA, 1976. Reporting Methods and Calibration in Uranium Exploration. Technical Report Series no. 174, International Atomic Energy Agency, Vienna. IAEA, 1991. Airborne gamma ray spectrometer surveying. Technical report series, no. 323, International Atomic Energy Agency, Vienna Kirkegaard, P. and Lovborg, L. , 1974. Computer modelling of terrestrial gammaradiation fields. Riso Report No. 303. Lovborg, L., 1984. The calibration of portable and airborne gamma-ray spectrometers - Theory, problems, and facilities. Riso National Laboratory, DK-4000 Roskilde, Denmark, Report RisoM-2456. Lrbcrg, L., Kirkegaard, P., and RoseHansen, 3., 1972. Quantitative interpretation of the gammaray spectra from geologic formations; Proceedings of the Second International Symposium on the Natural Radiation Environment, Houston, Texas, edited by J.A.S. Adams, W.M. Lowder, and TI. Gesell, p. 155180. Markkanen, M., At-vela, H., 1992. Radon emanation from soils; Radiation Protection Dosimetry, Vol. 45 No. 1/4 pp. 269272. Minty, B.R.S., 1992. Airborne gammaray spectxometric background estimation using full spectrum analysis. Geophysics, 57(2), 279-287. Minty B.R.S. and Kennett B.L.N., 1995. Optimum channel combinations for multichannel airborne gammaray spectrometxy. Exploration Geophysics, 25, 173178. Rogers, Vt., Nielson, ICK and Kalkwarf DR., 1984. Radon attenuation handbook for uranium mill tailings cover design. National Technical Information Service, U.S. Dept. of Commerce, NUREG/CR-3533. Wilde, S.A. and Low, G.E., 1978. Perth, Western Australia, 1:250,000 Geological Series Explanatory Notes, Sheet SHSO14, Geological Survey of Western Australia.

28/29

Pico Envirotec Inc.

GRS-16/GRS-10/GRS-2 Operation Manual

Rev:

5.0

APPENDIX E: CONTACTS
Pico Envirotec Inc. 752 Madeline Heights, Newmarket, Ontario, L3X 2J7 Canada Tel: +1 905 853 8536, Fax: +1 905 853 9668, e-mail: picoenv@cs.com. www.picoenvirotec.com AURA, s.r.o. Uvoz 56 602 00 Brno Czech Republic Tel: +420 5 43 245 111 Fax: +420 5 43 245 111 e-mail: engelsmann@aura.cz http://www.aura.cz/

Pico Envirotec Inc.

29/29