Indirect ophthalmoscopy Practical Learning objectives

To become aware of several types of indirect ophthalmoscope. To establish a basic understanding of the optical principles of the different instruments. To be able to set up each instrument accurately and attempt to carry out an ocular examination. To prepare for an indirect ophthalmoscopy assessment.

Introduction There are four indirect and one direct ophthalmoscopy technique that will be covered in these four weeks. You will rotate through each technique in groups of three or four. The rota will be located on the notice board. The techniques are slit lamp indirect, head mounted indirect, monocular indirect, modified monocular indirect and Hruby lens direct ophthalmoscopy. Refer to the relevant instruction sheets for clinical procedures. Slit lamp indirect ophthalmoscopy Purpose To evaluate the health of the posterior segment of the eye and in particular to detect the presence of macular oedema. A non-contact auxiliary condensing lens is used in conjunction with the slit lamp to provide an inverted, wide field, stereoscopic image with good resolution. Views out to the peripheral retina may be obtained with some lenses. This is useful in order to determine the presence of retinal neovascularisation. Equipment Numerous condensing lenses are available. VOLK, Ocular Instruments, Nikon and Haag Streit manufacture the main types of lens. They are plus powered with two convex aspheric surfaces. The +60D version has the greatest magnification and is best used for the disc and macular. The +78D version is a good general diagnostic lens, the +90D is good with small pupils, as is the SuperPupil, and the +66D produces a high definition view of the disc and the macular. The SuperField, which is approx. +90D, is a good all rounder. They are available in clear or blue-free, 'yellow retina protector glass'. This is considered to be more comfortable for the patient and minimises the risk of phototoxic retinal damage due to prolonged exposure to the focussed beam. The procedure that

follows applies to any high plus condensing lens (e.g., +60D, +66D, +78D or +90D, SuperPupil, and the SuperField NC). Haag Streit produce an 81D lens which provides a view similar to that of a 78D when held one way round, and a view similar to that produced by a 90D lens when held the other way round. Slit lamp. Dilating agents (0.5% or 1.0% tropicamide-for maximum dilation used in conjunction with 2.5% phenylephrine) Set-up Minimal lamp intensity can be used with a darkened room. Set the illumination arm of the slit lamp to be coaxial with the viewing system (i.e. straight-ahead). This will help to ensure proper lens alignment and will facilitate initial fundus detection and observation. Narrow the beam down 1 to 2 mm width with a height corresponding to the vertical height of the dilated pupil. Magnification should be set to 10x. Ensure that the condensing lens surfaces are clean. Hold the lens vertically between the thumb and index finger of the left hand to examine the patients right eye. Hold the lens in the right hand to examine the patients left eye. Some lenses may be used with a lens holder mounted on the upright support of the headrest or an adapter that rests against the patients eyelids. Either of these accessories may make fundus observation easier for the novice. Procedure 1. Instruct the patient to fixate straight ahead, to stare wide and to blink normally. 2. Centre the beam in the patients right pupil and focus on the cornea. 3. Place the lens in front of the patients eye, directly in front of the cornea so the back surface just clears the lashes (approximately 11 mm for the +60D, 7 mm for the +78D, and 6.5 mm for the +90D and the SuperField NC, from the patients cornea). These condensing lenses can be used either way round. Avoid touching the patients eyelashes if possible. Lens positioning may be made easier by placing the other fingers on either the brow bar or the patients forehead. When the lens is properly positioned, a blurred red fundus reflex will appear when looking through the oculars of the slit lamp.

4. Using the joystick, focus on the fundus image by slowly moving away from the cornea, keeping the beam centred in the pupil. When scratches on the surface of the condensing lens come into view the slit lamp needs to be pulled back only a little more (about 1 cm) for the retina to be in focus. 5. Once the retinal image is focussed, widen the beam to observe a greater area of the fundus. The magnification may also be changed to a higher setting at this time. 6. Scan across the entire lens. At this stage keep the lens still. 7. In order to view the peripheral retina ask the patient to change fixation into the nine cardinal positions of gaze. It will be necessary to realign the lens and refocus the slit lamp. 8. To reduce interfering reflections, tilt the lens or increase the setting of the illumination arm to 10 degrees either nasal or temporal once the fundus has been located and focussed. 9. Depending upon the design of the slit lamp, a +60D or a +78D lens may require the patient to move away from the headrest slightly to be able to focus the retina. 10. Centre the slit lamp beam in the left eye and repeat steps 3 through 6 on the patients left eye. Fine tuning the fundus view Lateral and longitudinal adjustments of the lens may be made to optimise the field of view. When viewing finer fundus details, temporally increase the lamp intensity and magnification. A larger field can be obtained by slightly moving the lens and the slit horizontally in the same direction. Recording Note which condensing was used as the view obtained may appear different than the view obtained with other methods of retinal evaluation. The condensing lens forms a real, inverted and laterally reversed image between the lens and the practitioner, approximately at the focus of the lens Descriptions and illustrations should be anatomically correct. Indicate where the lesion is on the retina, not where it appears through the lens.

Hruby Lens direct ophthalmoscopy Introduction The use of the slit-lamp biomicroscope allows a stereoscopic view of the retina, which is useful in assessing the elevation of interesting features. The auxiliary lenses provide high magnification with excellent resolution. Many practitioners experienced with Hruby lens suggest that it surpasses a +78 or +90D indirect examination in detecting subtle abnormalities. Although several aspects of this type of procedure prevent it from becoming part of a routine examination, there is no other way to achieve a comparable sense of depth. The Hruby lens (-55D) produces an upright virtual image that is not laterally reversed. How to perform the procedure Patient co-operation is extremely important and can be enhanced by attention to comfort before hand. Focus the oculars to accommodate any practitioner refractive error, set the pupillary distance, remove all filters and set the magnification to the lowest setting, usually 6-10x. The illumination of the slit-lamp should be adjusted for an intermediate slit height and a 2 mm width, and then placed in the straight ahead position between the oculars (zero degrees). Check to make sure neither eyepiece is occluded. The slit-lamp rheostat should be at a medium setting. As always the slit-lamp should be adjusted for the individual practitioner. The height of the patient is adjusted with the chin rest so that the fixation device is in the centre of the patient's field in front of the contra-lateral eye. After the slit-lamp is moved into position so that the slit is imaged in the patient's pupil, the Hruby lens is introduced in front of the patient's eye as close as possible without contacting the cornea or lashes. Advantages and disadvantages Direct ophthalmoscopy using a slit-lamp and negative auxiliary lenses can provide a very high level of magnification-even greater than that of the monocular hand held direct ophthalmoscope. The actual level of magnification depends on that available through the slit-lamp. Stereopsis is provided to a greater degree than all other examination techniques. The main disadvantage of this technique is the field of view. It is smaller than all other examination methods with the exception of direct monocular ophthalmoscopy (less than two disc diameters for an emmetropic patient). More dilation is required than in other binocular techniques. The quality of the image is easily degraded by media opacities; however this problem can be reduced by increasing the slit-lamp illumination.

As the magnification is so high, small movements of the practitioner, lens, or patient have an immediately noticeable negative effect on image quality. Patient co-operation is extremely important and can be enhanced with the use of a fixation device present on most slit-lamps. Head mounted binocular indirect ophthalmoscopy Introduction Binocular indirect ophthalmoscopy (BIO) is a technique used to evaluate the entire ocular fundus. It provides for stereoscopic, wide-angled, high-resolution views of the entire fundus and overlying vitreous. Its optical principles and illumination options allow for visualisation of the fundus regardless of high ametropia, hazy ocular media, or central opacities. Light beams directed into the patient's eye produce reflected observation beams from the retina. These beams are focused to a viewable, aerial image following placement of a high plus-powered condensing lens at its focal distance in front of the patient's eye. The resultant image is real, magnified, reversed left to right, inverted top to bottom and located between the examiner and the condensing lens. The observer views this image through the oculars of the head-borne indirect ophthalmoscope. Equipment 1) Head band for comfortable instrument placement. 2) Light source with variable intensity illumination beams directed downward and reflected laterally by an adjustable mirrored surface in the main housing. 3) Knobs to align the low plus powered eyepieces (+2.00 to +2.50 D) with the examiners interpupillary distance. Procedure Head mounted BIO adjustment 1) Place the loose BIO onto the head and position the bottom of the front headband one index finger width above the eyebrows. 2) Tighten the crown strap until this headband position begins to stabilise. 3) Position the back head strap on or below the occipital notch and tighten until secure. 4) Loosen the knobs that control the instruments main housing (oculars and light tower).

5) Fixating straight ahead and level, vertically position the oculars to within eye lash distance from your uncorrected eyes aligned tangential to or slightly angled downward from the ocular surface; this should maximise your visual field and minimise horizontal diplopia. 6) Horizontally align each ocular by closing one eye and fixating a centrally positioned thumb at 40 to 50 cm. 7) Adjust the ocular alignment knob or slide the oculars to place an identical centrally positioned thumb in each ocular's field of view. 8) Turn on the light source and fixate straight ahead on a wall at 40 to 50 cm looking at the projected light source. 9) Use the mirror knob to vertically place the light source at the upper one-half to onethird of the field. Head mounted BIO technique 1) Best with dilated pupils. 2) Headset in place and with the voltage set to mid-range 3) Hold a +20 D condensing lens with the thumb, index, and middle fingers of the right hand 4) The more convex surface should be toward the observer and the white-ringed edge closest to the patient 5) From a working distance of 18 to 20 inches, direct the light beam into the pupil, producing a complete red pupillary reflex 6) Pull backward on the lens, maintaining the central position of the pupil reflex, until the entire lens fills with the fundus image 7) Make fine adjustments in the lens tilt and vertex distance to produce a distortion-free full lens view. Monocular indirect ophthalmoscopy Introduction Monocular indirect ophthalmoscopy combines the advantages of increased field of view (indirect ophthalmoscopy) with erect real imaging (direct ophthalmoscopy). By collecting and redirecting peripheral fundus-reflected illumination rays, which cannot be

accomplished with the direct ophthalmoscope, the indirect ophthalmoscope extends the observer's field of view approximately four to five times. An internal lens system then reinverts the initially inverted image to a real erect one, which is then magnified. This image is focusable using the focusing lever/eyepiece lever. Procedure 1) Best carried out on a dilated pupil. 2) To examine the right eye, remove any patient spectacle correction, stand to the patient's right side, and ask them to fixate straight ahead and level with the left eye; the observer should keep on any refractive correction. 3) Dim the room lighting and turn up the instrument rheostat. 4) Push the iris diaphragm lever fully to the left to maximally increase the aperture size 5) Centre the red dot on the filter dial to position the open aperture for normal viewing 6) Observers head against the forehead rest and align eye through the instrument with the patient's right eye, holding the handle with the right hand 7) When positioned several inches in front of the patient and focus through the pupil onto the fundus using the thumb and focusing lever. 8) Approach the patient and stop at 10 to 12 cm from the eye. 9) Adjust the focusing and iris diaphragm to produce a clear maximally illuminated fundus view. 10) Continue to approach the patient until the observer's knuckle lightly touches the patient's cheek, as the working distance decreases, fundus magnification will increase. 11) Angle the light slightly nasally to illuminate the optic disc. Modified (monocular) indirect ophthalmoscopy Introduction A thorough fundus examination is important and required in all young patients with strabismus or amblyopia in order to rule out organic causes of amblyopia prior to the initiation of treatment. Examination of very young children can be difficult especially when a detailed view of the macula and optic nerve is required. The magnification and patient co-operation obtainedwith head mounted binocular indirect ophthalmoscopy using a 20D lens, and slit lamp biomicroscopy using a 90D or other similar lens means that

fundus examination is usually difficult or impossible on younger children. Also the magnification may be inadequate to allow accurate evaluation of posterior pole details. The direct ophthalmoscope is often the best available instrument for detailed retinal examination in young patients. However children often become frightened as the examiner approaches closely, as is necessary with the direct ophthalmoscope and co-operation is lost. Additionally children often fix the ophthalmoscope light and track it as it is moved by the examiner, allowing examination of the macula but not of the disc. The field of view is small and the magnification is more than is usually necessary. This will prevent the examiner from seeing the "big picture". To avoid these difficulties the direct ophthalmoscope can be used in conjunction with a 20D condensing lens, the type used with head mounted binocular indirect ophthalmoscopes. This combination provides a moderately magnified and wider angle, view of the posterior pole. This avoids the close proximity between the patient and examiner required when using a direct ophthalmoscope alone. The technique is called modified (monocular) indirect ophthalmoscopy and has been noted for its ability to provide a good view of the retina through a small pupil. Procedure To begin the examination a red reflex is visualised through the direct ophthalmoscope held approximately 18 cm from the patient's eye. A 20D lens is then placed 3 to 5 cm in front of the patient's eye in the path of the ophthalmoscope light beam, the examiner then needs to move slightly toward or away from the patient until a clear image of the retina is observed. An inverted, aerial image of the retina is produced, located between the observer and the lens. The apparent magnification will gradually increase as the examiner moves closer to this image (i.e. closer to the patient), allowing more detailed examination. A magnification of 4 to 5x is obtained by moving closer to the image. As the examiner moves closer additional lenses in the ophthalmoscope are needed, to keep the image clear depending on the accommodative needs of the examiner. Experience has shown that a viewing distance of approximately 18 cm from the patient is optimal, providing suitable magnification and a wide field of view A disadvantage of the technique, as with conventional direct ophthalmoscopy is the lack of a true stereoscopic view, however, lateral movement and rotation of the direct ophthalmoscope during the examination gives good parallax clues to depth.