225 SOFTWARE ADDENDUM SERIAL SCALE SETUP

April 30, 2009 Rev. 1.0

Cardinal Scale Manufacturing Co.


P.O. Box 151 - 203 E. Daugherty Webb City, Missouri 64870 Phone: 417-673-4631 - Fax: 417-673-5001 http://www.cardinalscale.com

SERIAL SCALE SETUP:

225 Addendum

REMOTE INDICATOR SETUP:


Refer to the 225 WEIGHT INFICATOR Installation and Technical Manual. Hold down the SHIFT key and then press the Navigation ENTER key (red square key in center of Navigation arrows). The display will change to show the SETUP/REVIEW MENU. Press the ENTER key to go to SETUP MENU #1.

SETUP MENU #1 1. USA=YES 6. CLR TARE=NO 2. NSC=NO 7. CLEAR ID=NO 3. LFT=NO 8. NO OF SCALES=1 4. OIML=NO 9. TOTALIZE=NO 5. TIME=24 10. MODE OF OP=1 IDS Enter Selection: 0 vNEXT ^EXIT
Press the NEXT navigation key, going to the SETUP MENU #2 screen.

SETUP MENU #2 1. SERIAL 2. DIO 3. PRINT TABS 4. SETUP SCALE 1 5. CALIBRATE 1 Enter Selection: 0

vNEXT

^PREV

Press 4 at the prompt, SCALE SETUP MENU #1 will be displayed.

Page 1 of 2

SERIAL SCALE SETUP:

225 Addendum

SCALE 1 SETUP MENU #1 1. BASE UNITS=3 lb 6. SAMPLES=4 2. CNVT UNITS=5 kg 7. STABLE=3 3. Z TRACK=0.0 8. STBL CNT= 3 4. Z LIMIT=MO 9. WT INTERVALS=1 5. PWR UP Z=NO 10. TYPE=ANALOG Enter Selection: 0 vNEXT ^EXIT

Select item 10, and for the type enter 2 for SERIAL. To ensure proper operation, the settings in SCALE SETUP MENU #2 MUST match the settings of the local indicator. Exit back to SETUP MENU #2 and press the NEXT key, SETUP MENU #3 will be displayed.

SETUP MENU #3 1. G/N ACCUMS=YES 6. KEY LOCKOUT 2. BACK LITE=30 7. LRPort=COM3 3. PASSWORD

Enter Selection: 0

^PREV vEXIT

The system defaults to COM3 for the Local/Remote connection. If a different port is desired, press 7 and select the desired port from the list. Exit back to the weight display mode, turn indicator off and then back on.

Page 2 of 2

2XX-AB ADDENDUM For 225 Indicator

May 11, 2009

Cardinal Scale Manufacturing Co.


P.O. Box 151 - 203 E. Daugherty Webb City, Missouri 64870 Phone: 417-673-4631 - Fax: 417-673-5001 http://www.cardinalscale.com

2XX-AB OPTION CARD

225 Addendum

SETUP:
With the indicator ON, hold the SHIFT key down and press the Navigation ENTER key (red square key in center of the Navigation arrows). The display will change to show the SETUP/REVIEW MENU.

SETUP/REVIEW MENU 1.ENTER CALIBRATION AND SETUP 2.VIEW AUDIT TRAIL COUNTERS 3.CALIBRATE SCALE 1

Enter Selection: 1

^EXIT

Press the ENTER key, then press the NEXT Navigation arrow until SETUP MENU #3 is displayed.

1. 2. 3. 4.

SETUP MENU #3 G/N ACCUMS=XXX 6. KEY LOCKOUT BACK LITE=XX PASSWORD 2xx-AB ^PREV vEXIT

Enter Selection: 0

Press the 4 key and then the ENTER key, the 2xx-AB SETUP MENU will be displayed.

Page 1 of 5

2XX-AB OPTION CARD

225 Addendum

1. 2. 3. 4.

2xx-AB SETUP MENU BAUD=XXX RACK=XX GROUP=X LAST=XXX ^EXIT

Enter Selection: 0

1. BAUD=X The current baud rate setting will be shown, to change the setting press the 1 key and then the ENTER key.

2xx-AB BAUD RATE SELECTION 57=57600 115=115200 230=230000

BAUD=XXX

If the displayed baud rate value is acceptable press the ENTER key, otherwise, key in the value for the desired baud rate and press the ENTER key.

2. RACK=XX The rack address is shown, to change the address, press the 2 key and then the ENTER key.

2xx-AB RACK ADDRESS SELECTION VALID ADDRESS: 0-7, 10-17, 20-27, 30-37

RACK=XX

Page 2 of 5

2XX-AB OPTION CARD

225 Addendum

If the address is correct press the ENTER key, otherwise, enter the desired address and press the ENTER key. 3. GROUP=X The current group number will be shown, to change the group number press the 3 key and then the ENTER key.

2xx-AB GROUP NUMBER SELECTION VALID NUMBERS: 0, 2, 4, 6

GROUP=X
If the displayed group number is acceptable press the ENTER key, otherwise, key in a valid group number and press the ENTER key. 4. LAST=XXX The last rack value is shown, to change, press the 4 key and then the ENTER key.

2xx-AB SETUP MENU

LAST=XXX YES

NO

If the value is correct press the ENTER key, otherwise, press the appropriate soft-key (YES or NO) and then press the ENTER key.

Press the Navigation EXIT arrow until the indicator goes into the weight display mode.

Page 3 of 5

2XX-AB OPTION CARD

225 Addendum

COMMUNICATIONS Date Sent From the XX To The PLC: Word0 Word1 Word2 Word3
Float point weight. Byte 0 Float point weight. Byte 2 Scale Number (Bits 4-7) Cmd1. Returned by 225 when complete Float point weight. Byte 1 Float point weight. Byte 3 Weight Status. See bit definitions below Cmd0. Returned by 225 when complete

Date Sent From the PLC To The 225: Word0 Word1 Word2 Word3
Cmd1. See definitions below N/A Float point value for Cmd. Byte 0 Float point value for Cmd. Byte 2 Cmd0. See definitions below N/A Float point value for Cmd. Byte 1 Float point value for Cmd. Byte 3

COMMANDS CMD0 and CMD1 are combined to make the command word. Command words are broken into command bits for PLC ease of programming. To invoke a command, set the bit to one. The PLC should continue to send a command until the command is returned by the indicator. If the indicator cannot perform a command, the returned command will include bit 7. The scale number is encoded as a four bit binary number, 1 through the number of scales available. NOTE: ALL COMMAND BITS SHOULD BE RELEASED AFTER THEY ARE PROCESSED BY THE INDICATOR. Bit0 Gross/Net Bit1 not defined Bit2 Zero Bit3 PB Tare Bit4 KB Tare, requires float point value be sent in words 2 and 3 Bit5-6 not defined Bit7 This bit is the error bit from the 225 when a command fails. Bit8-11 not defined Bit12 Select scale number, bit 0 (least significant) Bit13 Select scale number, bit 1 Bit14 Select scale number, bit 2 Bit15 Select scale number, bit 3 (most significant)

Page 4 of 5

2XX-AB OPTION CARD

225 Addendum

SCALE NUMBER The scale number is encoded as a four bit binary number, 1 through the number of scales available. Eg. Scale #1 (00010000), Scale #2 (00100000), Scale #3 (00110000), and Total scale (01000000). Bit0 Bit1 Bit2 Bit3 Bit4 Bit5 Bit6 Bit7 (not used) (not used) (not used) (not used) Select scale number, bit 0 (least significant) Select scale number, bit 1 Select scale number, bit 2 Select scale number, bit 3 (most significant) WEIGHT STATUS Bit0 Bit1 Bit2 Bit3 Bit4 Bit5 Bit6 Bit7 Motion Below Zero Center of Zero Over Capacity Weigh Units 0=unit 1 (PRIMARY UNITS), 1=unit 2 (ALTERNATE UNITS) Weigh Mode 0=Gross, 1=Net Not Defined Error. Data is invalid. FLOATING POINT WEIGHT DATA Example for weight data 5000 = Floating-point 459C4000h Byte0 Byte1 Byte2 Byte3 45h 9Ch 40h 00h

Some PLCs may require Byte 1 and Byte 0 to be swapped to be compatible with the Allen-Bradley floating point data format.

Page 5 of 5

2XX-DAC ADDENDUM For 225 Indicator

May 11, 2009

Cardinal Scale Manufacturing Co.


P.O. Box 151 - 203 E. Daugherty Webb City, Missouri 64870 Phone: 417-673-4631 - Fax: 417-673-5001 http://www.cardinalscale.com

2XX-DAC OPTION CARD

225 Addendum

SETUP:
With the indicator ON, hold the SHIFT key down and press the Navigation ENTER key (red square key in center of the Navigation arrows). The display will change to show the SETUP/REVIEW MENU.

SETUP/REVIEW MENU 1.ENTER CALIBRATION AND SETUP 2.VIEW AUDIT TRAIL COUNTERS 3.CALIBRATE SCALE 1

Enter Selection: 1

^EXIT

Press the ENTER key, then press the NEXT Navigation arrow until SETUP MENU #3 is displayed.

SETUP MENU #3 1. G/N ACCUMS=XXX 6. KEY LOCKOUT 2. BACK LITE=XX 3. PASSWORD 5. 2xx-DAC Enter Selection: 0

^PREV

vEXIT

Press the 5 key and then the ENTER key, the 2xx-DAC SETUP MENU will be displayed.

2xx-DAC SETUP MENU 1. GROSS ONLY=XXX 6. ADJUST LOW 2. LOW WT= XXXXXX 7. SCALE=X 3. HIGH WT= XXXXXX 4. VOLTS OUT=XX.XX 5. ADJUST HIGH Enter Selection: 0 ^EXIT

Page 1 of 6

2XX-DAC OPTION CARD

225 Addendum

1. GROSS ONLY=XXX The output can be set to follow the Gross weight only or the displayed weight (gross or net), to change the setting press the 1 key and then the ENTER key.

2xx-DAC SETUP MENU

GROSS ONLY=XXX YES

NO

If the displayed setting is acceptable press the ENTER key, otherwise, press the YES key and then the ENTER key for GROSS weight only, If the NO key is pressed then the output will follow the displayed weight (gross or net).

Page 2 of 6

2XX-DAC OPTION CARD

225 Addendum

2. LOW WT= XXXXXX This is the weight value which outputs zero volts (or 4 mA) from the DAC. To change this value, press the 2 key and then the ENTER key.

2xx-DAC SETUP MENU

LOW WT= XXXXXX


Key-in the weight and then press the ENTER key at the prompt. If the address is correct press the ENTER key, otherwise, enter the desired address and press the ENTER key. If the weight is to be negative, press the +/- key to toggle the sign and then press the ENTER key. Allowable values are: -99999 to 999999. Ex. If the low weight is to be -1000, then key-in 1 0 0 0 +/- ENTER keys.

3. HIGH WT= XXXXXX This is the weight value which outputs maximum selected voltage and current (see VOLTS OUT) from the DAC. To change this value, press the 3 key and then the ENTER key.

2xx-DAC SETUP MENU

HIGH WT= XXXXXX


If this weight is acceptable press the ENTER key, otherwise, key in a weight and press the ENTER key. This weight must be a positive value, up to capacity of scale, and above the LOW WEIGHT value. Allowable values are: 1 to 999999.

Page 3 of 6

2XX-DAC OPTION CARD

225 Addendum

4. VOLTS OUT=XX.XX This is the maximum output value in volts (00.01 to 10.00). All weight values equal to or greater than HIGH WEIGHT will output this value. Note, that if the scale goes over capacity, this value is used also. To change this value, press the 4 key and then the ENTER key.

2xx-DAC SETUP MENU

VOLTS OUT=XX.XX
If the setting is acceptable, press the ENTER key, otherwise, key-in the desired voltage and press the ENTER key. Allowable values are; 00.01 to 10.00. NOTE: If the 4-20 mA output is used, set VOLTS OUT = 10 X (max current-4)/16. Current= (VOLTS OUT/10) X 16 + 4(mA) (Can not exceed 20 mA)

Page 4 of 6

2XX-DAC OPTION CARD

225 Addendum

5. ADJUST HIGH This sets the DAC output to the VOLTS OUT level for adjusting the level. Adjustment potentiometers (pots) (10V and 20 mA) on the option board are used. Press the 5 key and then the ENTER key.

2xx-DAC SETUP MENU

Adjust HIGH value. Press ENTER when finished.

Adjust the 10 V pot for the maximum voltage output entered for VOLTS OUT. For 4-20 mA current output, adjust the 20 mA pot for the calculated maximum current.

6. ADJUST LOW This sets the DAC output to zero for adjusting the level, use the adjustment pots (10V and 4 mA) on the option board are used. Press the 6 key and then the ENTER key.

2xx-DAC SETUP MENU

Adjust LOW value. Press ENTER when finished.

There is NO adjustment for zero volts out. For 4-20 mA output, adjust the 4mA pot for the low (4 mA) current output. NOTE: Cycling between the ADJUST HIGH/LOW is necessary when adjusting the current out. This MUST be repeated until no adjustment is necessary.

Page 5 of 6

2XX-DAC OPTION CARD

225 Addendum

7. SCALE=X In a multi scale configuration the scale to be tracked by the DAC card can be selected. Press the 7 key and then the ENTER key.

2xx-DAC SETUP MENU

SCALE=X
Valid values are: 0 Current selected scale (scale that is in reverse video mode) 1 Scale #1 2 Scale #2 3 Scale #3 4 Total Scale

Press the Navigation EXIT arrow until the indicator goes into the weight display mode.

Page 6 of 6

225 SOFTWARE ADDENDUM DIGITAL FILL CONTROL

May 18, 2009 Rev. 1.0.D

Cardinal Scale Manufacturing Co.


P.O. Box 151 - 203 E. Daugherty Webb City, Missouri 64870 Phone: 417-673-4631 - Fax: 417-673-5001 http://www.cardinalscale.com

225-DFC:

225 Addendum

INTRODUCTION:
The Digital Fill Control (DFC) feature has been designed to control the filling or discharge of material automatically. The DFC feature controls relays in an external enclosure. NOTE: The DFC feature is ONLY available in a single scale configuration. Refer to the 225 WEIGHT INFICATOR Installation and Technical Manual for setup and calibration of the indicator.

SETUP:
With the indicator ON, hold the SHIFT key down and press the Navigation ENTER key (red square key in center of the Navigation arrows). The display will change to show the SETUP/REVIEW MENU.

SETUP/REVIEW MENU 1.ENTER CALIBRATION AND SETUP 2.VIEW AUDIT TRAIL COUNTERS 3.CALIBRATE SCALE 1

Enter Selection: 1

^EXIT

Press the ENTER key, SETUP MENU #1 is displayed.

SETUP MENU #1 1. USA=XXX 6. CLR TARE=XXX 2. NSC=XXX 7. CLEAR ID=XXX 3. LFT=XXX 8. NO OF SCALES=X 4. OIML=XXX 9. TOTALIZE=XXX 5. TIME=XX 10. MODE OF OP=X XXX Enter Selection: 0 ^PREV vEXIT
Key-in 10 and press the ENTER key, the Mode Of Operation Menu will be displayed.

Page 1 of 11

225-DFC:

225 Addendum

Mode Of Operation Menu 0. Normal Indicator 5. Batcher 1. ID Storage 6. Package Weigher 2. Digital Fill Ctrl 3. Preset Weight Comp 4. Check Weigher Enter Selection: 0 ^EXIT
Enter 2 and press the ENTER key, the Digital Fill Control Menu will be displayed.

Digital Fill Control Menu 1. SPEED=X 6. DUMP GATE=XXX 2. GATE SEQ=XXX 7. AUTO DUMP=XXX 3. AUTO TRIM=XXX 8. DCUMLATE=XXX 4. AUTO PRNT=XXX 9. AUTO TARE=XXX 5. MULT DROP=XXX Enter Selection: 0 ^PREV
1. SPEED=X SPEED=1 Single Speed Operation SPEED=2 Two Speed Operation 2. GATE SEQ=XXX If speed is 1, then this item is not applicable (GATE SEQ=N/A). GATE SEQ=1 (AB then B): The AB then B selection will begin the operation with both outputs on until the weight reaches the FAST= value. At that weight, the A (fast-PWC1) output will be turned off. The B (slow-PWC2) output will remain on until the SLOW= weight is reached. At that weight, the B (slow-PWC2) output will be turned off. GATE SEQ=2 (A then B) With A then B selected, the operation will begin with only the A (fast-PWC1) output on until the weight reaches the FAST= value. At that weight, the A (fast-PWC1) output will be turned off and the B (slow-PWC2) output will be turned on. The B (slow-PWC2) output will remain on until the SLOW= weight is reached. At that weight, the B (slow-PWC2) output will be turned off. GATE SEQ=3 (Chatter Gate)

Page 2 of 11

225-DFC:

225 Addendum

The Chatter Gate selection and will control one relay, this selection requires two speed operation. It begins the operation with the A (fast-PWC1) output on until the weight reaches the FAST= value. At that weight, the A (fast-PWC1) output will be turned off. The A (now as slow-PWC1) output will then be turned on a preset time (CHATTER= 0 to 99.9 seconds). The A (slow-PWC1) output will be turned off and the weight value will be compared to the SLOW= value less the TRIM= value. This will be repeated until the weight is equal to or greater than the SLOW= value less the TRIM= value. 3. AUTO TRIM=XXX AUTO TRIM=YES AUTO TRIM=NO Automatic Trim Compensation ON Automatic Trim Compensation OFF

Automatic trim compensation may be selected in which case the 225 indicator will automatically recalculate the required trim compensation based on the difference between the material preset and the amount actually filled. The 225 indicator will allow the manual entry of a trim value for the material. If automatic trim compensation is selected, this trim value will be changed automatically with each fill based on the error between the weight of the material discharged and the corresponding preset. If automatic trim is not selected, the entered trim values remain unchanged. 4. AUTO PRNT=XXX AUTO PRNT=YES AUTO PRNT=NO

Automatic Print Enabled Automatic Print Disabled

The automatic print feature will cause an optional printer to record the total weight of the fill automatically at the conclusion of the fill. The time and date is also included in the printed record. 5. MULT DROP=XXX MULT DROP=YES MULT DROP=NO

Multiple Drop Enabled Multiple Drop Disabled

This feature allows up to 9 (equal) drops to be made to achieve the preset weight value. Note that although the drops are equal amounts, the last drop will compensate for the freefall of the previous drops and the trim value to equal the total preset weight value. 6. DUMP GATE=XXX DUMP GATE=YES DUMP GATE=NO

Enable Dump Gate Disable Dump Gate

This feature is used when filling a hopper instead of a truck on a scale. The optional dump gate can be manually, automatically or remotely controlled (uses output PWC16).

Page 3 of 11

225-DFC:

225 Addendum

7. AUTO DUMP=XXX AUTO DUMP=N/A AUTO DUMP=YES AUTO DUMP=NO

DUMP GATE=NO (above) Automatic Dump Enabled Automatic Dump Disabled

Selection of this feature will cause the completed batch to discharge automatically. The word DUMP appears in the weight display to signal the operator that the batch discharge is taking place. The discharge will continue until the scale weight returns to a value within the predetermined range of zero when the discharge or dump output will be turned off. If the automatic dump feature was not selected, the indicator will halt after discharging the last enabled ingredient and will wait for the manual discharge of the material after which the next batch will be initiated if the automatic recycle feature was selected. 8. DCUMLATE=XXX DCUMLATE=YES DCUMLATE =NO

Weigh Out Weigh In

In a decumulative filling operation, the weigh hopper is filled first until the required amount of material has been placed in the hopper. When the operation is started, the material is dumped until the proper amount has been discharged. 9. AUTO TARE=XXX AUTO TARE=N/A AUTO TARE=YES AUTO TARE =NO

DCUMLATE=YES (above) Automatic Tare Enabled (ON) Automatic Tare Disabled (OFF)

This feature is a two-position toggle that controls whether the fill operation fills to the gross or net weight on the scale. If the automatic tare is selected, the weight on the scale will be used as the TARE weight before beginning the fill operation. Note that with automatic tare enabled (ON), the operation will fill to the target net weight and with automatic tare disabled (OFF), the operation will fill to the gross weight.

Press the EXIT Navigation key until the indicator returns to the weight display mode.

Page 4 of 11

225-DFC:

225 Addendum

SETTING THE DEFAULT PWC OUTPUT STATES: Go to the SETUP MENU #2.

SETUP MENU #2 1. SERIAL 6. SETUP SCALE 2 2. DIO 7. CALIBRATE 2 3. PRINT TABS 8. SETUP SCALE 3 4. SETUP SCALE 1 9. CALIBRATE 3 5. CALIBRATE 1 Enter Selection: 0 vEXIT ^PREV
Press the 2 and then the ENTER key, the DIO SETUP MENU will be displayed.

SETUP MENU #2 PWC1-OFF PWC6-OFF PWC11-OFF PWC16-OFF PWC2-OFF PWC7-OFF PWC12-OFF 17.ALL OFF PWC3-OFF PWC8-OFF PWC13-OFF 18.ALL ON PWC4-OFF PWC9-OFF PWC14-OFF PWC5-OFF PWC10-OFF PWC15-OFF Enter PWC # to Toggle: 0 ^EXIT
Enter the PWC # to toggle the default state of the PWC relay between ON and OFF. Note: While this screen shows all of the PWC outputs only PWCs 1,2 and 16 are used by the DFC.

Page 5 of 11

225-DFC:

225 Addendum

OPERATION:
The Digital Fill Control (DFC) feature has been designed to control the filling or discharge of material automatically. The DFC feature controls relays in an external enclosure. NOTE: The DFC feature is ONLY available in a single scale configuration.

Entering Filling Parameters: At the weight display screen, press the PRESET soft-key. The DFC PRESET PARAMETERS MENU will be displayed.

DFC PRESET PARAMETERS MENU 1. FILL= XXXXXX 6. GATE TIMER=N/A 2. SLOW= XXXXXX 7. CHATTER = N/A 3. TRIM= XXXXXX 4. DROP CNT=X 5. ZERO TOL=XXX ENTER SETTING TO CHANGE: 0
1. FILL/FAST= XXXXXX If single fill operation was selected, enter the target fill weight at the FILL= prompt. If 2-speed was selected then enter the fast fill cutoff weight at the FAST= prompt. 2. SLOW= XXXXXX This item is only for 2-speed operation, enter the slow fill cutoff weight (target weight) at the SLOW= prompt. 3. TRIM= XXXXXX Enter the TRIM weight at the TRIM= prompt. 4. DROP CNT= X If the MULT DROP feature was enabled, enter the number of equal drops at the DROP CNT= prompt. If the MULT DROP feature was disabled, the number of drops will ALWAYS be 1.

Page 6 of 11

225-DFC:

225 Addendum

5. ZERO TOL= XXX If the Dump Gate was enabled, enter the zero tolerance weight at the ZERO TOL= prompt. The zero tolerance is a range of weight within which the scale is considered empty. If the Dump gate was disabled, this prompt is not applicable and the display will show ZERO TOL=N/A. 6. GATE TIMER= XXX If the Dump Gate was enabled, enter the amount of time (in seconds) that the dump gate is to be left open AFTER the weight falls below the zero tolerance weight. Allowable times are 0 through 99 seconds. If the Dump gate was disabled, this prompt is not applicable and the display will show GATE TIMER =N/A. 7. CHATTER = XXX If 2-speed operation and gate sequence = 3 were chosen in setup, enter the amount of time (in tenths of a second) that the chatter gate is to open. Allowable times are 0.0 through 99.9 seconds. If single speed operation was chosen or the gate sequence is not 3, this prompt is not applicable and the display will show CHATTER = N/A.

To return to the weight display mode, press the ENTER key.

Page 7 of 11

225-DFC:

225 Addendum

SINGLE SPEED OPERATION: To Start a Fill Operation 1. Press the START key, the START=YES prompt will be displayed. 2. Press the ENTER key to start the fill operation, or press the NO key and then the ENTER key to abort the operation. 3. PWC1 annunciator comes on and the PWC1 output relay is turned on. 4. The weight will increase to the FILL= weight minus the TRIM= weight. When the displayed weight reaches that value, the PWC1 and annunciator are turned off. 5. When Motion stops, the weight will be printed (if a printer is attached and the automatic print was enabled). 6. The trim weight will be recalculated, and stored in memory (if automatic trim was enabled) and the displayed weight will be added to the GROSS or NET accumulators (if the automatic print was enabled).

To Suspend (Pause) a Fill Operation in Progress The fill operation can be suspended (paused) by the operator if needed by pressing the PAUSE key. 1. Press the PAUSE key. 2. The fill operation is suspended (paused), and the PAUSE key will be replaced with the RESTRT key. 3. To continue the fill operation where it was before it was suspended, press the RESTRT key. 4. To stop (abort) the fill operation, press the STOP key. 5. The 225 will return to normal operation. To Stop (Abort) a Fill Operation in Progress The fill operation can be stopped (aborted) by the operator if needed by pressing the STOP key. 1. Press the STOP key. 2. The filling operation will immediately be aborted. 3. The 225 will return to normal operation.

Page 8 of 11

225-DFC:

225 Addendum

TWO SPEED OPERATION: To Start a Fill Operation GATE SEQ=1 1. Press the START key, the START=YES prompt will be displayed. 2. Press the ENTER key to start the fill operation, or press the NO key and then the ENTER key to abort the operation. 3. PWC1 and PWC2 annunciators come on and the PWC1 and PWC2 output relays are turned on. 4. The weight will increase to the FAST= weight. When the displayed weight reaches that value, the PWC1 output relay and annunciator are turned off. 5. The fill operation continues until the weight reaches the SLOW= weight minus the TRIM= weight. When the weight reaches that value, the PWC2 annunciator and PWC2 relay output will be turned off. 6. When Motion stops, the weight will be printed (if a printer is attached and the automatic print was enabled). 7. The trim weight will be recalculated, and stored in memory (if automatic trim was enabled) and the displayed weight will be added to the GROSS or NET accumulators (if the automatic print was enabled).

GATE SEQ=2 1. Press the START key, the START=YES prompt will be displayed. 2. Press the ENTER key to start the fill operation, or press the NO key and then the ENTER key to abort the operation. 3. PWC1 annunciator comes on and the PWC1 output relay is turned on. 4. The weight will increase to the FAST= weight. When the displayed weight reaches that value, the PWC1 output relay and annunciator are turned off and the PWC2 output relay and annunciator are turned on. 5. The fill operation continues until the weight reaches the SLOW= weight minus the TRIM= weight. When the weight reaches that value, the PWC2 annunciator and PWC2 relay output will be turned off. 6. When Motion stops, the weight will be printed (if a printer is attached and the automatic print was enabled). 7. The trim weight will be recalculated, and stored in memory (if automatic trim was enabled) and the displayed weight will be added to the GROSS or NET accumulators (if the automatic print was enabled).

Page 9 of 11

225-DFC:

225 Addendum

GATE SEQ=3 1. Press the START key, the START=YES prompt will be displayed. 2. Press the ENTER key to start the fill operation, or press the NO key and then the ENTER key to abort the operation. 3. PWC1 annunciator comes on and the PWC1 output relay is turned on. 4. The weight will increase to the FAST= weight. When the displayed weight reaches that value, the PWC1 output relay and annunciator are turned off. 5. The fill operation continues by turning PWC1 annunciator and output relay on and off until the weight is equal or greater than the SLOW= weight minus the TRIM= weight. When the weight reaches that value, the PWC2 annunciator and PWC2 relay output will be turned off. 6. When Motion stops, the weight will be printed (if a printer is attached and the automatic print was enabled). 7. The trim weight will be recalculated, and stored in memory (if automatic trim was enabled) and the displayed weight will be added to the GROSS or NET accumulators (if the automatic print was enabled).

To Suspend (Pause) a Fill Operation in Progress The fill operation can be suspended (paused) by the operator if needed by pressing the PAUSE key. 1. Press the PAUSE key. 2. The fill operation is suspended (paused), and the PAUSE key will be replaced with the RESTRT key. 3. To continue the fill operation where it was before it was suspended, press the RESTRT key. 4. To stop (abort) the fill operation, press the STOP key. 5. The 225 will return to normal operation. To Stop (Abort) a Fill Operation in Progress The fill operation can be stopped (aborted) by the operator if needed by pressing the STOP key. 1. Press the STOP key. 2. The filling operation will immediately be aborted. 3. The 225 will return to normal operation.

Page 10 of 11

225-DFC:

225 Addendum

DECUMULATIVE FILL OPERATION The decumulative operation is used when weighing material out of a weigh hopper. The fill signal will be used to open and close a weigh hopper gate. The settings for the DFC decumulative operation to function properly are as follows: MODE OF OP = 2 DUMP GATE = NO DCUMLATE YES CLR TARE = NO To Start a Decumulative Fill Operation The weigh hopper must be filled with material that equals or exceeds the amount of the preset value to be removed; otherwise an error message -CHECK MATL- will be displayed. This message indicates that the weigh hopper does not have enough material to meet the requirement of the preset weight value. Either more material has to be added to the weigh hopper or the preset weight value must be reduced. 1. With the indicator showing Gross Weight, press the START/STOP key. 2. The display will show START=YES. 3. Press the ENTER key to start the fill (discharge) operation, or press the NO key to abort the operation. 4. The indicator will switch to the Net Weight mode displaying zero weight, and the PWC1 annunciator and relay will turn on beginning the operation. 5. The weight will increase on the display. The weight is actually being discharged from the weigh hopper. 6. When the displayed weight reaches the preset value, the operation will stop. PWC1 t annunciator and relay will turn off. 7. When all motion stops, the weight will be printed (if a printer is attached and automatic print was enabled in setup). 8. The indicator will remain in the NET mode. The Gross Weight can be viewed by pressing the NET/GROSS key to select the Gross Weight. 9. The operation is repeated until the weigh hopper weight goes below the preset weight value. At that time, more material must be placed into the weigh hopper. NOTE! If the decumulative operation is used for multiple materials in a compartmentalized weigh hopper, the 225 can only check the total weight of material to verify there is sufficient amount to meet the preset values.

Page 11 of 11

2XX-DN ADDENDUM For 225 Indicator

May 11, 2009

Cardinal Scale Manufacturing Co.


P.O. Box 151 - 203 E. Daugherty Webb City, Missouri 64870 Phone: 417-673-4631 - Fax: 417-673-5001 http://www.cardinalscale.com

2XX-DN OPTION CARD

225 Addendum

SETUP:
With the indicator ON, hold the SHIFT key down and press the Navigation ENTER key (red square key in center of the Navigation arrows). The display will change to show the SETUP/REVIEW MENU.

SETUP/REVIEW MENU 1.ENTER CALIBRATION AND SETUP 2.VIEW AUDIT TRAIL COUNTERS 3.CALIBRATE SCALE 1

Enter Selection: 1

^EXIT

Press the ENTER key, then press the NEXT Navigation arrow until SETUP MENU #3 is displayed.

1. 2. 3. 4.

SETUP MENU #3 G/N ACCUMS=XXX 6. KEY LOCKOUT BACK LITE=XX PASSWORD 2xx-DN ^PREV vEXIT

Enter Selection: 0

Press the 4 key and then the ENTER key, the 2xx-DN SETUP MENU will be displayed.

2xx-DN SETUP MENU 1. BAUD=X 2. ADDR=XX

Enter Selection: 0
1. BAUD=X

^EXIT

Page 1 of 4

2XX-DN OPTION CARD

225 Addendum

The current baud rate setting will be shown (0, 1 or 2), to change the setting press the 1 key and then the ENTER key.

2xx-DN BAUD RATE SELECTIONS 0=125k 1=250k 2=500k

BAUD=X
If the displayed baud rate value is acceptable press the ENTER key, otherwise, press the 0, 1 or 2 key and then the ENTER key for the desired baud rate. 2. ADDR=XX The current address is shown, to change the address, press the 2 key and then the ENTER key.

2xx-DN ADDRESS SELECTION VALID ADDRESS: 0-63

ADDR=XX
If the address is correct press the ENTER key, otherwise, enter the desired address and press the ENTER key. Press the Navigation EXIT arrow until the indicator goes into the weight display mode. COMMUNICATIONS Date Sent From the XX To The PLC: Word0 Word1 Word2 Word3
Float point weight. Byte 0 Float point weight. Byte 2 Scale Number (Bits 4-7) Cmd1. Returned by 225 when complete Float point weight. Byte 1 Float point weight. Byte 3 Weight Status. See bit definitions below Cmd0. Returned by 225 when complete

Date Sent From the PLC To The 225: Word0 Word1 Word2 Word3
Page 2 of 4

Cmd1. See definitions below

Cmd0. See definitions below

2XX-DN OPTION CARD

225 Addendum

N/A Float point value for Cmd. Byte 0 Float point value for Cmd. Byte 2

N/A Float point value for Cmd. Byte 1 Float point value for Cmd. Byte 3

COMMANDS CMD0 and CMD1 are combined to make the command word. Command words are broken into command bits for PLC ease of programming. To invoke a command, set the bit to one. The PLC should continue to send a command until the command is returned by the indicator. If the indicator cannot perform a command, the returned command will include bit 7. The scale number is encoded as a four bit binary number, 1 through the number of scales available. NOTE: ALL COMMAND BITS SHOULD BE RELEASED AFTER THEY ARE PROCESSED BY THE INDICATOR. Bit0 Gross/Net Bit1 not defined Bit2 Zero Bit3 PB Tare Bit4 KB Tare, requires float point value be sent in words 2 and 3 Bit5-6 not defined Bit7 This bit is the error bit from the 225 when a command fails. Bit8-11 not defined Bit12 Select scale number, bit 0 (least significant) Bit13 Select scale number, bit 1 Bit14 Select scale number, bit 2 Bit15 Select scale number, bit 3 (most significant)

SCALE NUMBER The scale number is encoded as a four bit binary number, 1 through the number of scales available. Eg. Scale #1 (00010000), Scale #2 (00100000), Scale #3 (00110000), and Total scale (01000000). Bit0 Bit1 Bit2 Bit3 Bit4 Bit5 Bit6 Bit7 (not used) (not used) (not used) (not used) Select scale number, bit 0 (least significant) Select scale number, bit 1 Select scale number, bit 2 Select scale number, bit 3 (most significant)

Page 3 of 4

2XX-DN OPTION CARD

225 Addendum

WEIGHT STATUS Bit0 Bit1 Bit2 Bit3 Bit4 Bit5 Bit6 Bit7 Motion Below Zero Center of Zero Over Capacity Weigh Units 0=unit 1 (PRIMARY UNITS), 1=unit 2 (ALTERNATE UNITS) Weigh Mode 0=Gross, 1=Net Not Defined Error. Data is invalid. FLOATING POINT WEIGHT DATA Example for weight data 5000 = Floating-point 459C4000h Byte0 Byte1 Byte2 Byte3 45h 9Ch 40h 00h

Some PLCs may require Byte 1 and Byte 0 to be swapped to be compatible with the Allen-Bradley floating point data format.

Page 4 of 4

2XX-EIP ADDENDUM For 225 Indicator

May 11, 2009

Cardinal Scale Manufacturing Co.


P.O. Box 151 - 203 E. Daugherty Webb City, Missouri 64870 Phone: 417-673-4631 - Fax: 417-673-5001 http://www.cardinalscale.com

2XX-EIP OPTION CARD

225 Addendum

SETUP:
With the indicator ON, hold the SHIFT key down and press the Navigation ENTER key (red square key in center of the Navigation arrows). The display will change to show the SETUP/REVIEW MENU.

SETUP/REVIEW MENU 1.ENTER CALIBRATION AND SETUP 2.VIEW AUDIT TRAIL COUNTERS 3.CALIBRATE SCALE 1

Enter Selection: 1

^EXIT

Press the ENTER key, then press the NEXT Navigation arrow until SETUP MENU #3 is displayed.

1. 2. 3. 4.

SETUP MENU #3 G/N ACCUMS=XXX 6. KEY LOCKOUT BACK LITE=XX PASSWORD 2xx-EIP ^PREV vEXIT

Enter Selection: 0

There are no setup variables for the EIP option card, going to this menu just assures the operator that the 225 can detect the presence of the option card. Press the Navigation EXIT arrow until the indicator goes into the weight display mode.

Page 1 of 3

2XX-EIP OPTION CARD

225 Addendum

COMMUNICATIONS Date Sent From the XX To The PLC: Word0 Word1 Word2 Word3
Float point weight. Byte 0 Float point weight. Byte 2 Scale Number (Bits 4-7) Cmd1. Returned by 225 when complete Float point weight. Byte 1 Float point weight. Byte 3 Weight Status. See bit definitions below Cmd0. Returned by 225 when complete

Date Sent From the PLC To The 225: Word0 Word1 Word2 Word3
Cmd1. See definitions below N/A Float point value for Cmd. Byte 0 Float point value for Cmd. Byte 2 Cmd0. See definitions below N/A Float point value for Cmd. Byte 1 Float point value for Cmd. Byte 3

COMMANDS CMD0 and CMD1 are combined to make the command word. Command words are broken into command bits for PLC ease of programming. To invoke a command, set the bit to one. The PLC should continue to send a command until the command is returned by the indicator. If the indicator cannot perform a command, the returned command will include bit 7. The scale number is encoded as a four bit binary number, 1 through the number of scales available. NOTE: ALL COMMAND BITS SHOULD BE RELEASED AFTER THEY ARE PROCESSED BY THE INDICATOR. Bit0 Gross/Net Bit1 not defined Bit2 Zero Bit3 PB Tare Bit4 KB Tare, requires float point value be sent in words 2 and 3 Bit5-6 not defined Bit7 This bit is the error bit from the 225 when a command fails. Bit8-11 not defined Bit12 Select scale number, bit 0 (least significant) Bit13 Select scale number, bit 1 Bit14 Select scale number, bit 2 Bit15 Select scale number, bit 3 (most significant)

Page 2 of 3

2XX-EIP OPTION CARD

225 Addendum

SCALE NUMBER The scale number is encoded as a four bit binary number, 1 through the number of scales available. Eg. Scale #1 (00010000), Scale #2 (00100000), Scale #3 (00110000), and Total scale (01000000). Bit0 Bit1 Bit2 Bit3 Bit4 Bit5 Bit6 Bit7 (not used) (not used) (not used) (not used) Select scale number, bit 0 (least significant) Select scale number, bit 1 Select scale number, bit 2 Select scale number, bit 3 (most significant) WEIGHT STATUS Bit0 Bit1 Bit2 Bit3 Bit4 Bit5 Bit6 Bit7 Motion Below Zero Center of Zero Over Capacity Weigh Units 0=unit 1 (PRIMARY UNITS), 1=unit 2 (ALTERNATE UNITS) Weigh Mode 0=Gross, 1=Net Not Defined Error. Data is invalid. FLOATING POINT WEIGHT DATA Example for weight data 5000 = Floating-point 459C4000h Byte0 Byte1 Byte2 Byte3 45h 9Ch 40h 00h

Some PLCs may require Byte 1 and Byte 0 to be swapped to be compatible with the Allen-Bradley floating point data format.

Page 3 of 3

2XX-IP ADDENDUM For 225 Indicator

May 11, 2009

Cardinal Scale Manufacturing Co.


P.O. Box 151 - 203 E. Daugherty Webb City, Missouri 64870 Phone: 417-673-4631 - Fax: 417-673-5001 http://www.cardinalscale.com

2XX-IP OPTION CARD

225 Addendum

SETUP:
With the indicator ON, hold the SHIFT key down and press the Navigation ENTER key (red square key in center of the Navigation arrows). The display will change to show the SETUP/REVIEW MENU.

SETUP/REVIEW MENU 1.ENTER CALIBRATION AND SETUP 2.VIEW AUDIT TRAIL COUNTERS 3.CALIBRATE SCALE 1

Enter Selection: 1

^EXIT

Press the ENTER key, then press the NEXT Navigation arrow until SETUP MENU #3 is displayed.

1. 2. 3. 4.

SETUP MENU #3 G/N ACCUMS=XXX 6. KEY LOCKOUT BACK LITE=XX PASSWORD 2xx-IP ^PREV vEXIT

Enter Selection: 0

Press the 4 key and then the ENTER key, the 2xx-IP SETUP MENU will be displayed.

2xx-IP SETUP MENU 1. CONT=XXX 2. TYPE=X XXXXX 3. SCALE=X

Enter Selection: 0

^EXIT

Page 1 of 3

2XX-IP OPTION CARD

225 Addendum

1. CONT=XXX To turn the continuous data output ON or OFF, press the 1 key and then the ENTER key.

2xx-IP SETUP MENU

CONT=XXX YES

NO

Press the YES and ENTER keys to turn continuous data output ON, press the NO and ENTER keys to turn it off.

Page 2 of 3

2XX-IP OPTION CARD

225 Addendum

2. TYPE=X To select the format of the data transmitted from the 2xx-IP card, press the 2 key and then the ENTER key.

2xx-IP CONTINOUS OUTPUT SELECTION 0. SMA 5. NUMERIC WT 1. SB400 6. SB500 2. SB200 7. SB250/500M 3. IQ355 4. TOLEDO SHORT TYPE=X
Enter the 1-digit value for the desired format and press the ENTER key. For a complete description of the Continuous Output and the Weight On Demand formats, refer to the CONTINUOUS OUTPUT FORMATS section in the 225 Weight Indicating Instrument Technical and Operation Manual (8200-M538-O1). 7. SCALE=X In a multi scale configuration the scale to be tracked by the 2xx-IP card can be selected. Press the 7 key and then the ENTER key.

2xx-IP SETUP MENU

SCALE=X
Valid values are: 0 Current selected scale (scale that is in reverse video mode) 1 Scale #1 2 Scale #2 3 Scale #3 4 Total Scale Press the Navigation EXIT arrow until the indicator goes into the weight display mode.

Page 3 of 3

2XX-RS232 ADDENDUM For 225 Indicator

May 11, 2009

Cardinal Scale Manufacturing Co.


P.O. Box 151 - 203 E. Daugherty Webb City, Missouri 64870 Phone: 417-673-4631 - Fax: 417-673-5001 http://www.cardinalscale.com

2XX-RS232 OPTION CARD

225 Addendum

SETUP:
With the indicator ON, hold the SHIFT key down and press the Navigation ENTER key (red square key in center of the Navigation arrows). The display will change to show the SETUP/REVIEW MENU.

SETUP/REVIEW MENU 1.ENTER CALIBRATION AND SETUP 2.VIEW AUDIT TRAIL COUNTERS 3.CALIBRATE SCALE 1

Enter Selection: 1

^EXIT

Press the ENTER key, then press the NEXT Navigation arrow until SETUP MENU #3 is displayed.

1. 2. 3. 4.

SETUP MENU #3 G/N ACCUMS=XXX 6. KEY LOCKOUT BACK LITE=XX PASSWORD 2xx-RS232 ^PREV vEXIT

Enter Selection: 0

Press the 4 key and then the ENTER key, the 2xx-RS232 SETUP MENU will be displayed.

2xx-RS232 SETUP MENU 1. BAUD=XXX 6. TYPE=X XXXXX 2. PARITY=X XXXX 7. SCALE=X 3. BITS=X 8. CTS ENABLED=XXX 4. STOPS=X 5. CONT=XXX Enter Selection: 0 ^EXIT

Page 1 of 5

2XX-RS232 OPTION CARD

225 Addendum

1. BAUD=XX To change the BAUD setting press the 1 key and then the ENTER key.

24= 48= 96=

2xx-RS232 BAUD RATE SELECTION 2400 19= 19200 76= 76800 4800 38= 38400 9600 57= 57600

BAUD=XX
Key-in the 2-digit value for the desired baud rate and press the ENTER key. 2. PARITY=X To change the PARITY setting press the 2 key and then the ENTER key.

2xx-RS232 PARITY SELECTION 0. NONE 1. ODD 2. EVEN

PARITY=X
Key-in the 1-digit value for the desired PARITY and press the ENTER key.

Page 2 of 5

2XX-RS232 OPTION CARD

225 Addendum

3. BITS=X To change the number of data BITS setting, press the 3 key and then the ENTER key.

2xx-RS232 SETUP MENU

BITS=X
Enter the number of data bits at the prompt, allowable values are 7 or 8. 4. STOPS=X To change the number of stop BITS setting, press the 4 key and then the ENTER key.

2xx-RS232 SETUP MENU

STOPS=X
Enter the number of STOP bits at the prompt, allowable values are 1 or 2.

Page 3 of 5

2XX-RS232 OPTION CARD

225 Addendum

5. CONT=XXX To turn the continuous data output ON or OFF, press the 5 key and then the ENTER key.

2xx-RS232 SETUP MENU

CONT=XXX YES

NO

Press the YES and ENTER keys to turn continuous data output ON, press the NO and ENTER keys to turn it off. 6. TYPE=X To select the format of the data transmitted from the 2xx-RS232 card, press the 6 key and then the ENTER key.

2xx-RS232 CONTINOUS 0. SMA 5. 1. SB400 6. 2. SB200 7. 3. IQ355 4. TOLEDO SHORT TYPE=X

OUTPUT SELECTION NUMERIC WT SB500 SB250/500M

Enter the 1-digit value for the desired format and press the ENTER key. For a complete description of the Continuous Output and the Weight On Demand formats, refer to the CONTINUOUS OUTPUT FORMATS section in the 225 Weight Indicating Instrument Technical and Operation Manual (8200-M538-O1).

Page 4 of 5

2XX-RS232 OPTION CARD

225 Addendum

7. SCALE=X In a multi scale configuration the scale to be tracked by the 2xx-RS232 card can be selected. Press the 7 key and then the ENTER key.

2xx-RS232 SETUP MENU

SCALE=X
Valid values are: 0 Current selected scale (scale that is in reverse video mode) 1 Scale #1 2 Scale #2 3 Scale #3 4 Total Scale 8. CTS ENABLED=XXX The Clear To Send handshaking can be enabled or disabled, press the 8 key and then the ENTER key.

2xx-RS232 SETUP MENU

CTS ENABLED=XXX
Press the YES and ENTER keys to enable CTS, press the NO and ENTER keys to disable CTS.

Press the Navigation EXIT arrow until the indicator goes into the weight display mode.

Page 5 of 5

2XX-USB ADDENDUM For 225 Indicator

May 11, 2009

Cardinal Scale Manufacturing Co.


P.O. Box 151 - 203 E. Daugherty Webb City, Missouri 64870 Phone: 417-673-4631 - Fax: 417-673-5001 http://www.cardinalscale.com

2XX-USB OPTION CARD

225 Addendum

SETUP:
With the indicator ON, hold the SHIFT key down and press the Navigation ENTER key (red square key in center of the Navigation arrows). The display will change to show the SETUP/REVIEW MENU.

SETUP/REVIEW MENU 1.ENTER CALIBRATION AND SETUP 2.VIEW AUDIT TRAIL COUNTERS 3.CALIBRATE SCALE 1

Enter Selection: 1

^EXIT

Press the ENTER key, then press the NEXT Navigation arrow until SETUP MENU #3 is displayed.

SETUP MENU #3 1. G/N ACCUMS=XXX 6. KEY LOCKOUT 2. BACK LITE=XX 3. PASSWORD 5. 2xx-USB Enter Selection: 0

^PREV

vEXIT

Press the 5 key and then the ENTER key, the 2xx-USB SETUP MENU will be displayed.

Page 1 of 3

2XX-USB OPTION CARD

225 Addendum

2xx-USB SETUP MENU 1. CONT=XXX 2. TYPE=X XXXXX 3. SCALE=X

Enter Selection: 0

^EXIT

1. CONT=XXX To turn the continuous data output ON or OFF, press the 1 key and then the ENTER key.

2xx-USB SETUP MENU

CONT=XXX YES

NO

Press the YES and ENTER keys to turn continuous data output ON, press the NO and ENTER keys to turn it off.

Page 2 of 3

2XX-USB OPTION CARD

225 Addendum

2. TYPE=X To select the format of the data transmitted from the 2xx-USB card, press the 2 key and then the ENTER key.

2xx-USB CONTINOUS OUTPUT SELECTION 0. SMA 5. NUMERIC WT 1. SB400 6. SB500 2. SB200 7. SB250/500M 3. IQ355 4. TOLEDO SHORT TYPE=X

Enter the 1-digit value for the desired format and press the ENTER key. For a complete description of the Continuous Output and the Weight On Demand formats, refer to the CONTINUOUS OUTPUT FORMATS section in the 225 Weight Indicating Instrument Technical and Operation Manual (8200-M538-O1). 7. SCALE=X In a multi scale configuration the scale to be tracked by the 2xx-USB card can be selected. Press the 7 key and then the ENTER key.

2xx-USB SETUP MENU

SCALE=X

Valid values are: 0 Current selected scale (scale that is in reverse video mode) 1 Scale #1 2 Scale #2 3 Scale #3 4 Total Scale Press the Navigation EXIT arrow until the indicator goes into the weight display mode.
Page 3 of 3

225 SOFTWARE ADDENDUM PRESET WEIGHT COMPARATOR

May 12, 2009 Rev. 1.0.D

Cardinal Scale Manufacturing Co.


P.O. Box 151 - 203 E. Daugherty Webb City, Missouri 64870 Phone: 417-673-4631 - Fax: 417-673-5001 http://www.cardinalscale.com

225-PWC:

225 Addendum

INTRODUCTION:
The Preset Weight Comparator software in the 225 has been modified to allow PWCs to be assigned to a scale in a multi-scale installation. Refer to the 225 WEIGHT INFICATOR Installation and Technical Manual for setup and calibration of the indicator.

SETUP:
With the indicator ON, hold the SHIFT key down and press the Navigation ENTER key (red square key in center of the Navigation arrows). The display will change to show the SETUP/REVIEW MENU.

SETUP/REVIEW MENU 1.ENTER CALIBRATION AND SETUP 2.VIEW AUDIT TRAIL COUNTERS 3.CALIBRATE SCALE 1

Enter Selection: 1

^EXIT

Press the ENTER key, SETUP MENU #1 is displayed.

SETUP MENU #1 1. USA=XXX 6. CLR TARE=XXX 2. NSC=XXX 7. CLEAR ID=XXX 3. LFT=XXX 8. NO OF SCALES=X 4. OIML=XXX 9. TOTALIZE=XXX 5. TIME=XX 10. MODE OF OP=X XXX Enter Selection: 0 ^PREV vEXIT
Key-in 10 and press the ENTER key, the Mode Of Operation Menu will be displayed.

Page 1 of 6

225-PWC:

225 Addendum

Mode Of Operation Menu 0. Normal Indicator 1. ID Storage 3. Preset Weight Comp Enter Selection: 1 ^EXIT

Enter 3 and press the ENTER key, the Preset Weight Comparator Menu will be displayed.

Preset Weight Comparator Menu 1. OUTPUTS=XX 2. BAL PRNT=XXX

Enter Selection: 0

^PREV

1. OUTPUTS=XX The number of Preset Weight Comparator (PWC) outputs can be specified, by pressing the 1 key and then the ENTER key. At the OUTPUTS= prompt enter the number of PWCs desired and press the ENTER key. Valid entry is 1 thru 16. 2. BAL PRNT=XXX With only one PWC selected (OUTPUTS=1), the automatic print on balance feature may be enabled. If more than one PWC has been selected, this feature is disabled. To enable/disable, press the 2 key and then the ENTER key, at the BAL PRNT= prompt press either the YES or NO soft-key and the ENTER key to enable/disable this feature.

Press the EXIT Navigation key until the indicator returns to the weight display mode.

Page 2 of 6

225-PWC:

225 Addendum

SETTING THE DEFAULT PWC STATES: Go to the SETUP MENU #2.

SETUP MENU #2 1. SERIAL 6. SETUP SCALE 2 2. DIO 7. CALIBRATE 2 3. PRINT TABS 8. SETUP SCALE 3 4. SETUP SCALE 1 9. CALIBRATE 3 5. CALIBRATE 1 Enter Selection: 0 vEXIT ^PREV
Press the 2 and then the ENTER key, the DIO SETUP MENU will be displayed.

SETUP MENU #2 PWC1-OFF PWC6-OFF PWC11-OFF PWC16-OFF PWC2-OFF PWC7-OFF PWC12-OFF 17.ALL OFF PWC3-OFF PWC8-OFF PWC13-OFF 18.ALL ON PWC4-OFF PWC9-OFF PWC14-OFF PWC5-OFF PWC10-OFF PWC15-OFF Enter PWC # to Toggle: 0 ^EXIT
Enter the PWC # to toggle the default state of the PWC relay between ON and OFF.

Page 3 of 6

225-PWC:

225 Addendum

OPERATION:
The weight indicator has sixteen outputs, which can be configured during the setup of the weight indicator to perform as Preset Weight Comparators. If the preset weight comparator mode of operation was selected, the indicator will compare each enabled preset weigh value with the displayed weight and then output a signal for each preset based on the results of that comparison. Entering the Preset values: At the weight display screen, press the PRESET soft-key. The Preset values screen will be displayed (only the values for the enabled presets will be displayed, in this example all sixteen factory default preset values are shown).

1.P= 2.P= 3.P= 4.P= 5.P= 6.P= ENTER

500 7.P= 3500 13.P= 6500 1000 8.P= 4000 14.P= 7000 1500 9.P= 4500 15.P= 7500 2000 10.P= 5000 16.P= 8000 2500 11.P= 5500 17. SET TRIM 3000 12.P= 6000 18. SET SCALE to EXIT, or PWC# TO CHANGE: 0

To change a specific PWC weight value, enter the PWC# (1 thru 16) and press the ENTER key and key-in the new weight at the PWC X= prompt. To return to the weight display mode, press the ENTER key.

Page 4 of 6

225-PWC:

225 Addendum

Entering the Trim values: At the weight display screen, press the PRESET soft-key. The Preset values screen will be displayed, at the prompt key-in 17 and press the ENTER key. The trim values for the PWC will be displayed (only the values for the enabled PWCs will be displayed, in this example all sixteen factory default trim values are shown).

1.T= 2.T= 3.T= 4.T= 5.T= 6.T= ENTER to

0 7.T= 0 13.T= 0 0 8.T= 0 14.T= 0 0 9.T= 0 15.T= 0 0 10.T= 0 16.T= 0 0 11.T= 0 17. SET WGT 0 12.T= 0 18. SET SCALE EXIT, or PWC# TO CHANGE: 0

To change a specific PWC trim weight value, enter the PWC# (1 thru 16) and press the ENTER key and key-in the new weight at the TRM X= prompt. To return to the weight display mode, press the ENTER key.

Page 5 of 6

225-PWC:

225 Addendum

Entering the Preset Scale: At the weight display screen, press the PRESET soft-key. The Preset values screen will be displayed, at the prompt key-in 18 and the press the ENTER key. The scale that each PWC is associated with will be displayed (only the values for the enabled PWCs will be displayed, in this example all sixteen factory default preset values are shown).

1.S=1 7.S=1 13.S=1 2.S=1 8.S=1 14.S=1 3.S=1 9.S=1 15.S=1 4.S=1 10.S=1 16.S=1 5.S=1 11.S=1 17. SET TRIM 6.S=1 12.S=1 18. SET SCALE ENTER to EXIT, or PWC# TO CHANGE: 0
To change a specific PWC scale association, enter the PWC# (1 thru 16) and press the ENTER key and key-in the scale association at the SCL X= prompt. To select the TOTAL scale, enter a 4 at the above prompt. To return to the weight display mode, press the ENTER key.

Page 6 of 6