You are on page 1of 3

DATA SHEET

BP Series
The Merit Sensor Blood Pressure sensor is ideal for low cost, high-volume, disposable medical applications including blood pressure monitoring. The BP Series sensor is temperature compensated, dielectric gel protected and can easily be installed into a customers pressure device housing.
COMPANY: Merit Sensor is a leader in piezoresistive
pressure sensing and partners with clients to create high performing solutions for a variety of applications and industries. Design, engineering and manufacturing of Merit Sensor products takes place in state-of-the-art wafer fabrication facility in Utah, US.

TECHNOLOGY: Merit Sensor utilizes a piezoresistive


Wheatstone bridge with a chemically etched silicon diaphragm. All products are RoHS compliant.

CAPABILITIES: Merit Sensor designs, engineers, fabricates, dices, assembles, tests, sells and services die and packaged products from a state-of-the-art facility near Salt Lake City, Utah. FEATURES
Pressure range of -30 to 300 mmHg Excellent burst pressure Compliant with AAMI BP22 specications Dielectric gel barrier Fully tested Small, disposable, low-cost package Temperature compensated Solid state piezoresistive sensor Top side pressure entry Easy to install into customers disposable blood pressure device housing Compatible with automated assembly equipment Compatible in air, gas and liquid Shipped as 120UP snapstrates or on tape and reel Automated testing and assembly Custom options available Engineering, design and manufacturing under one roof in Utah, US.

APPLICATIONS
Disposable blood pressure measurement Kidney dialysis machines Infusion pumps Surgical procedures
Sales +1 801.208.4722 Customer Service +1 801.208.4700 Fax +1 801.208.4798 sensors@merit.com www.MeritSensor.com 1600 W. Merit Parkway South Jordan, Utah 84095 USA
BPseriesDS.006

BPSeries
SPECIFICATIONS
Parameter General Pressure Range Overpressure Input Excitation Risk Current Input Impedance Output Impedance Dielectric Breakdown Environmental Temp (Comp/Operating) Temperature (Storage) Humidity Light Sensitivity Operating Product Life Shelf Life Sterilization (ETO) Mechanical Weight Volume Displacement Media Interface Gel Tube Interface Material Performance 1 Offset Sensitivity Calibration Symmetry Accuracy (-30 to 50 mmHg) -25 4.95 97.5 -5 -1 mmHg - 1% 0 of Reading 0 0 0 -0.3 -0.1 1,200 5 1 0 5 100 25 5.05 102.5 5 +1 mmHg + 1% of Reading 3 0.3 0.1 mmHg/C %/C Hz degrees mmHg/8 hrs mmHg V/V/mmHg mmHg % Zero 2 .02 Grams mm3 Medical grade, dielectric gel Polycarbonate 168 3 3 15 -25 10 40 70 90 1 C C % mmHg Hours Years Cycles 7 5 - Per AAMI BP22 1,200 285 10,000 -30 125 1 10 2 3000 315 300 mmHg PSI VDC A VDC Typical burst of +800 PSI Calibrated to 6 VDC Minimum Typical Maximum Units Notes

Electrical (22C unless otherwise stated)

2 6 - Per AAMI BP22

Accuracy (50 to 300 mmHg) -3 Temp Coeff Zero Temp Coeff Sensitivity Frequency Response Phase Shift Offset Drift
NOTES:

% of Reading 6 - Per AAMI BP22

3 3 5 5 4, 5

1. Assumes 6 VDC, 22C and after ve second warm-up unless otherwise specied. 2. Output of sensor with no pressure applied and a150 K resistor shorted across + IN to + OUT. 3. Over a temperature range of 15C to 40C. 4. Normalized offset/bridge voltage 8hours after 20 second warm-up. 5. Previously qualied, not tested in production. 6. Combined effect of sensitivity, repeatability, nonlinearity and hysteresis errors. 7. Sterilization performed by customer.

Sales +1 801.208.4722 Customer Service +1 801.208.4700 Fax +1 801.208.4798 sensors@merit.com www.MeritSensor.com 1600 W. Merit Parkway South Jordan, Utah 84095 USA

BPSeries
DIMENSIONS (millimeters)

Standard Bond Pad Metallization = Ag/Pt

0 14.99 25.53 36.07 46.61 57.15 67.69 78.23 88.77 99.31 109.86 114.30 114.30 105.92 97.79 89.66 81.53 73.41 65.28 57.15 49.02 40.89 32.77 24.64 16.51 8.38 0

NOTE: When sensors are shipped as 120UP Note: *Sensors are shipped as 120UP snapstrates snapstrates, the ceramic plates must be which must be singulated by the customer. singulated by the customer. *Each plate may include units that have failed electrical or visual test. Bad units are identified Each may include units that with a redplate dot on the backside of thehave unit. failed *Plates are shipped in dust electrical or visual test. free Bad anti-static units are identied containers to prevent damage to the gel. with a red dot on the backside of the unit.
*Dimensions are in Inches

Plates are shipped in dust free anti-static containers to prevent damage to the gel.

UNLESS OTHERWISE SPECIFIED:


DIMENSIONS ARE IN INCHES TOLERANCES: FRACTIONAL ANGULAR: MACH BEND TWO PLACE DECIMAL THREE PLACE DECIMAL
PROPRIETARY AND CONFIDENTIAL THE INFORMATION CONTAINED IN THIS INTERPRET GEOMETRIC TOLERANCING PER: MATERIAL FINISH

NAME DRAWN CHECKED ENG APPR. MFG APPR. Q.A. COMMENTS:

DATE

TITLE:

DBPS 120up assembly


SIZE DWG. NO.

DRAWING IS THE SOLE PROPERTY OF Sales +1COMPANY 801.208.4722 <INSERT NAME HERE>. ANY Customer Service +1 801.208.4700 Fax +1 801.208.4798 sensors@merit.com www.MeritSensor.com REPRODUCTION IN PART OR AS A WHOLE WITHOUT THE WRITTEN PERMISSION OF <INSERT COMPANY NAME HERE> IS PROHIBITED.

NEXT ASSY

APPLICATION

1600 W. Merit Parkway South Jordan, Utah 84095 USA


USED ON DO NOT SCALE DRAWING

SCALE: 1:1 WEIGHT:


1

REV

SHEET 1 OF 1