Product Information Load cell for tensile and compression from 330 kN to 600 kN

D2 d2

d1 D1

Figure: Load cell 600 kN

Figure: Drawing of load cell with dimenions

Key Benefits

No influence through torsional moments Low creep error even at abrupt load removal
(specimen break)

through holes for mounting bolts simplify installation


and allow a wide range of testing situations to be accommodated

The inside of the extensometer is filled with protective hermetically sealed, corrosion-resistant design for high
long-term reliability gas, so no reaction with oxygen (from air or moisture) is possible

All hollows are laser welded, and so a long-time stable


Further advantages und features and very high consistency will be reached towards the environment

Rotational symmetrical measurement body (computer


optimised), thus measurement properties are independent of the mounting location

In fatigue tests controlled firmness means quasiunlimited life of the materials testing machine

High static overloading gives more protection to the Independent of the stiffness (and so deflection) of the
force transmission user (tensile strength 300% nominal force)

Insensitive to eccentric force transmission and lateral


forces
PI 651 2.0114

PI651_KA330-600.pmd

Product Information Load cell for tensile and compression from 330 kN to 600 kN
Fmax (tensile/compression) [kN] Accuracy grade 1 from .. on [kN] Accuracy grade 0.5 from .. on [kN] Nominal measurement path [mm] Dimensions H Height [mm] D1 Upper mounting stud dia [mm] d1 Upper mounting stud dia [mm] D2 Lower mounting stud dia [mm] d2 Lower mounting stud dia [mm] Eccentricity influence / mm Lateral force influence at 0.1 x Fnom on the load cell upper side Reference temperature [C] Nominal temperature range [C] Use temperature range [C] Storage temperature range [C] Limit force Static limit lateral force Tensile force Protection system to DIN 40050-1 Item no. 3301 0.66 3.3 0.17 160 250 220 220 184 0.02 % 3301 2 0.66 3.3 0.17 160 250 220 220 184 0.02 % 4001 0.8 4.0 0.17 160 250 220 220 184 0.02 % 4001 2 0.8 4.0 0.17 160 250 220 220 184 0.02 % 600 1 1.2 6.0 0.17 160 250 220 220 184 0.02 % 6001 2 1.2 6.0 0.17 160 250 220 220 184 0.02 %

0.02 % 22 (1) +10 ... +60 -30 ... +85 -50 ... +85 150 % 100 % 300 % IP67 630383

0.02 % 22 (1) +10 ... +60 -30 ... +85 -50 ... +85 150 % 100 % 300 % IP67 089308

0.02 % 22 (1) +10 ... +60 -30 ... +85 -50 ... +85 150 % 100 % 300 % IP67 358453

0.02 % 22 (1) +10 ... +60 -30 ... +85 -50 ... +85 150 % 100 % 300 % IP67 358454

0.02 % 22 (1) +10 ... +60 -30 ... +85 -50 ... +85 150 % 100 % 300 % IP67 358455

0.02 % 22 (1) +10 ... +60 -30 ... +85 -50 ... +85 150 % 100 % 300 % IP67 628347

(BPC-L0330KN.G03) (BPC-L0330KN.T03) (BPC-L0400KN.G03) (BPC-L0400KN.T03) (BPC-L0600KN.G03) (BPC-L0600KN.T03)


1 2

with calibration certificate according to DIN EN ISO 7500-1 with additional mounting stud M 52 x 1.5 for second test area

All Data at ambient temperature.

All rights reserved.