Professional Documents
Culture Documents
Section High voltage -> High voltage TSG followed by high voltage test descrip-
tions.
Section Filament -> Filament TSG followed by filament test description..
Section Anode rotation -> Anode rotation TSG followed by RAC test description..
Section XRS hints -> XRS hints.
Section Measuring point -> Table of LEDs and measuring points on XGS/XRS.
Definitions and abbreviations 0
Tab. 1 Definitions and abbreviations
Term Definition / Abbreviation
XDC X-ray deflection control
XGR X-ray generator, rotating
XGS X-ray generator, stationary
XRS X-ray system
XRT X-ray tube
XTA X-ray tube assembly
XTC X-ray tube cooling
MVT Middle voltage transmission
16 XRS
SOMATOM Definition C2-028.840.01.09.02 Siemens, 2006
08.11 H CX CS SD CR-CT
Page 16 of 176
For internal use only
Prerequisites 0
Documents
n.a
Tools and auxiliary equipment
Standard tool kit
HV test plugs
HVT High voltage tank
TSG Troubleshooting guide
Term Definition / Abbreviation
Siemens, 2006 C2-028.840.01.09.02 SOMATOM Definition
08.11 H CX CS SD CR-CT
XRS 17
Page 17 of 176
For internal use only
Overview 2.1
The XRS section contains information on possible XRS tests and the corresponding trou-
bleshooting guides. In the SOMATOM Definition, troubleshooting always starts by calling
up the EventLog and searching for the relevant error message. Follow the instructions
given in this error message for troubleshooting. In some cases, the error message will link
you to troubleshooting instructions as described in this document. This link is normally
used if the troubleshooting instructions are too complex or images are needed for trouble-
shooting.
Additionally, read the information which is stored in the tube history (Local Service
->Report). This information can always be useful in case of XRS problems.
Problem isolation 0
Principle problem isolation in case of XRS problems
1. Find relevant error message in Eventlog
NOTE The XGR and XGS assemblies are continuously communicat-
ing using the CAN bus after the system is switched on. For
this reason, 2 error messages are usually sent to the log-
book after XRS problems. Always check both (XGS/XGR)
error messages when troubleshooting.
2. Follow the instructions given in the error message text for troubleshooting
The information provided in the error message will guide you in finding the prob-
lem. In some cases, you will be linked to this Test and Troubleshooting instruc-
tions, # C2-028.840.01.... In this case, follow the relevant troubleshooting
instructions described in this document.
Generator oscillator procedure (GenOSC) 0
Generator oscillator adaptation (Gen-Osc) is an automatic program. The purpose of the
program is to adapt the inverter frequencies. The program performs very short scans with
different inverter frequencies to find the minimum resonance point. This value is stored to
D700 in XGS.
The frequency adaptation procedure must run successfully; otherwise, XRS errors such
as UT too high or UT max errors (e.g XRS_83 error) may abort scanning. The same errors
may appear if parts of the oscillating circuit (MV cable, XGS_PDC_Control, HVT / MVT)
have been replaced without performing GenOSC after exchange. For this reason, always
use the relevant guided tour after part replacement.
NOTE If GenOsc fails because of arcing, try using dummy plugs (i.e.
test plugs for HVT) and perform GenOsc again with dummy
plugs installed.
18 XRS
SOMATOM Definition C2-028.840.01.09.02 Siemens, 2006
08.11 H CX CS SD CR-CT
Page 18 of 176
For internal use only
Flowchart for HV Troubleshooting 0
The flowchart (Fig. 8 / p. 19) gives a guideluine how to deal with tube arcings and repre-
sents the algorithm behind the tool SOMATOM Definition HV Troubleshooting. For more
information regarding this tool, see the documentation for CT016/09/R.
S
i
e
m
e
n
s
,
2
0
0
6
C
2
-
0
2
8
.
8
4
0
.
0
1
.
0
9
.
0
2
S
O
M
A
T
O
M
D
e
f
i
n
i
t
i
o
n
0
8
.
1
1
H
C
X
C
S
S
D
C
R
-
C
T
X
R
S
1
9
P
a
g
e
1
9
o
f
1
7
6
F
o
r
i
n
t
e
r
n
a
l
u
s
e
o
n
l
y
Fig. 8: HV Troubleshooting Flowchart
20 XRS
SOMATOM Definition C2-028.840.01.09.02 Siemens, 2006
08.11 H CX CS SD CR-CT
Page 20 of 176
For internal use only
XDC/Tube 2.2
XDC/Tube Test 0
INIT Tests:
During INIT, all XRT drivers of the XDC (outputs on X13 and X14) are tested in 2 steps. The
first step is an isolation check for detecting short circuits between the output wires and
ground or power (230 V). This step can detect defective driver gates within the XDC as well.
The second step activates the drivers and can detect short circuits between the output
wires or broken wires. The INIT tests have failed if any warning or error message occurs in
the Eventlog during step 1 or 2, or if the requested currents in step 2 cannot be reached.
XDC/Tube Self-Test within Service SW:
The XDC/Tube self-test is basically the same as the INIT test. Depending on the actual sys-
tem state, the XDC/Tube self-test tolerates some problems. So, if the XDC/Tube self-test
fails, switch the gantry off/on to initialize an INIT test.
Test results:
If the test fails, it results in error message " CT_DCA/B 31" where the parameters give
some information on which part of the test failed. In addition to this information, there can
be additional errors or warnings which indicate the error. Follow the instructions in the
warning/error message for troubleshooting
Usage of the tests:
The XDC/Tube self-test can help detect problems in the output drivers of the XDC, prob-
lems within cabling, or problems within the tube. The test can be helpful if the XDC has
aborted scans with the error message "CT_DCA/B 51 ..." or "CT_DCA/B 50", especially
if parameter 3 is 0x00, 0x01, 0x04, 0x05, 0x08, 0x09, 0x12, 0x17, 0x1D, 0x1E or 0x1F. Be
aware that any of these errors can happen during arcing as well, without any real HW prob-
lem within the XDC or the tube. If the INIT tests fail, you have a problem in the XDC, cabling,
or the tube. If any additional error or warning message occurs in the Eventlog during INIT,
it can be used to help detect where the problem is located. You can disconnect the cable
on the XDC (X13 and X14) and/or the tube (X8 and X4) and then repeat the INIT test. When
the additional errors/warnings disappear, you have disconnected the faulty part from the
XDC.
Remark: Test can be used if the system is in or is not in stand-by status.
XDC/Tube test 0
XDC/Tube test
Performing the filament test
1. Select Local Service > Test Tools > Tube/Generator
2. Select XDC/Tube.
3. Click GO and follow the instructions in the dialog. The test is terminated automatically.
Siemens, 2006 C2-028.840.01.09.02 SOMATOM Definition
08.11 H CX CS SD CR-CT
XRS 21
Page 21 of 176
For internal use only
High voltage 2.3
TSG high voltage 0
The following high voltage TSG has to be used in case of high voltage problems. The TSG
starts at the XRT (X-ray tube) and provides guidance through the HV related components.
NOTE XRS components cannot be exchanged from system A to
system B or vice versa for troubleshooting.
NOTE Always use new silicon insulators at the HV sockets ((HV
plug installation at HV sockets / p. 28)) as well as new silicon
disks at the HVT when installing the cable plugs or the test
plugs.
Tab. 2 High voltage TSG
Step Action
1 Check Eventlog for UT max and /or UT too high errors (e.g XRS_83) or arc-
ing problem (XRS_68).
If there are UTmax and /or UT too high errors, perform GenOSC (Local Ser-
vice -> Tune Up -> Expert Mode -> GenOSC).
a) if GenOSC is aborted due to a problem -> continue with item 2
b) if the HV problem is not solved after GenOSC -> continue with item 3.
c) if the HV problem is solved -> GenOSC tables are stored in D700 XGS.
No further action necessary.
2 If there is any doubt that the oil pressure is correct, check the oil pres-
sure.Too low oil pressure especially in the older1.05bar- typ of Definition
tubes is very critical compared to 1.25bar- tybe of tubes.
Perform GenOSC with installed test plugs. See (HVC plug test / p. 30) for
test plug installation.
a) if the GenOSC still aborts -> check the Eventlog for CT_GSA/GRA
messages and follow the instructions given in the error message.
b) if the GenOSC is ok -> try GenOSC again with installed X-ray tube -> if
problem recurs, change XRT (X-ray tube).
3 Perform Gettering (Local Service -> Tune Up -> Expert Mode -> Getter).
a) if the HV problem is not solved after gettering the tube -> continue with
item 4
b) if the HV problem is solved -> no further action.
22 XRS
SOMATOM Definition C2-028.840.01.09.02 Siemens, 2006
08.11 H CX CS SD CR-CT
Page 22 of 176
For internal use only
4 1. Switch off system (select System -> END) and press service switch S1 in
PDC_A (Service Push-Button S1 in PDC-A / p. 11).
2. Switch off rotation enable switch S301.
3. Remove HV plugs from the HV sockets (beside the tube).
4. Remove insulators from the HV plugs.
5. Check HV plugs and the 2 insulators for signs of arcing (black dots or stripes)
(HV plugs with arcing tracks / p. 26).
a) If there are any signs of arcing -> continue with item 5
b) If there are no signs of arcing -> continue with item 6.
5 If there are any signs of arcing on the plugs, HV sockets, or the insulators.
a) Clean HV plugs and the 2 sockets (inside) with alcohol.
b) If any signs of arcing (cracks) remain on the plug or HV sockets ->
change the relevant part(s).
c) If there are no signs of arcing remaining on the parts -> reinstall HV plugs
using new silicon insulators (HV plug installation at HV
sockets / p. 28).
d) Try scanning again -> if the HV problems reoccur continue with item 6.
6 Perform HVC tube test. Follow the dialog of the HVC tube test (tube test with
HV but without load).
a) If there are no XRS HV error messages found in the logbook -> continue
with item 11.
b) If there are XRS HV error messages found in the logbook -> continue
with item 8.
7 Perform HV plug test with installed dummy plugs ( #10 093 700) at HV sock-
ets (HVC plug test / p. 30). Follow the dialog of the HVC plug test. (Local
Service --> Test Tools --> Generator --> HVC plugs)
The HV test plugs have to be installed using new insulators and have
to tightened with a torque of 25 Nm . (HV plug installation at HV
sockets / p. 28)
a) If the test is aborted after 80 s by the UMAR (scan time exceeded), there
is no HV problem on generator side --> change tube.
b) If there are still XRS HV error messages found in the logbook -> con-
tinue with item 8.
Step Action
Siemens, 2006 C2-028.840.01.09.02 SOMATOM Definition
08.11 H CX CS SD CR-CT
XRS 23
Page 23 of 176
For internal use only
Check/adjust the tube oil pressure 0
An incorrect oil pressure can result in various errors.
If the oil pressure is too high, the tube would activate the oil pressure safety switch or the
oil temperature safety switch, when the tube gets hot.
The more likely case, though, is that the oil pressure is too low.
Especially in the Somatom Definition (dual source), the earlier type of tubes needs to be
adjusted to 1.05 bar, current tubes need a pressure of 1.25 bar.
8 1. Switch off system (select System -> END) and press service switch S1 in
PDC_A (Service Push-Button S1 in PDC-A / p. 11).
2. Switch off rotation enable switch S301.
3. Remove HV plugs from the HVT (high voltage tank).
4. Check HV plugs and the 2 silicon disks for signs of arcing (black dots or
stripes) (HV plugs with arcing tracks / p. 26).
a) If there are any signs of arcing -> continue with item 9
b) If there are no signs of arcing -> continue with item 10.
9 If there are any signs of arcing on the plugs or the disks.
a) Clean HV plugs and HVT sockets inside with alcohol
b) If any signs of arcing (e.g. cracks) remain on the HV cable plugs or HVT
sockets -> change the relevant part(s).
c) If there are no signs of arcing remaining on the parts -> reinstall HV plugs
into HVT using new silicon disks.
d) Try scanning again -> if the HV problems reoccur continue with item 10.
10 Perform HVC plug test scans with installed dummy plugs (#1621791) at HVT
(HVC plug test / p. 30). Follow the dialog of the HVC plug test.
a) If there are no XRS error messages found in the logbook (UMAR errors
are acceptable) -> change both HV cables including the HV sockets.
b) If there are XRS error messages found in the logbook -> change HVT.
11 The HV problem is related to tube load.
Possible reasons: Inverter problem, ground problem, HVT.
a) Check Eventlog for XRS errors related to inverters -> if there are any fol-
low the instructions given in the error message.
b) Check ground connections (bad contact....) of the XRS parts.
Step Action
24 XRS
SOMATOM Definition C2-028.840.01.09.02 Siemens, 2006
08.11 H CX CS SD CR-CT
Page 24 of 176
For internal use only
NOTE From 11/2009 on, new tubes need a pressure of 1.25 bar (+
altitude correction). These tubes are clearly labelled, and
also the expansion unit is labelled with a red sticker.
Set the pressure on old tubes without label to 1.05 bar, and
on newer tubes to 1.25 bar according to the label.
The expansion vessel for the following measurement can be ordered from service stock.
The part number for the vessel is: 10569439.
Cool down the system for at least 10 minutes cooling time after the last scan. During
the cooling time, the system must be switched on.
Be aware that some system components may be very hot, e.g., the Straton tube,
XTC, etc. may reach up to 130 degrees Celsius.
Wait until the tube and the cooling system have cooled down, then switch the system to
COMP/ON status at the control box.
Switch off the gantry power using service push button S1 in PDC A.
Remove the gantry lower front ring segment.
Turn the gantry by hand until the Straton MX P tube (A or B) is in the 6 oclock position.
Secure against unintended rotation by using the front safety bolt.
Prepare the expansion vessel.
Open the quick coupling and disconnect the lower oil hose at the tube. Connect the oil
hose from the expansion vessel to the empty quick coupling of the tube.
Fig. 9: Adjusting unit for expansion membrane
Install the adjustment handle (item 1) on the expansion vessel.
- Press the button (item 2) to lock the thread rod on the membrane.
- Hand-tighten the 2 screws of the adjustment thread rod (item 3).
Siemens, 2006 C2-028.840.01.09.02 SOMATOM Definition
08.11 H CX CS SD CR-CT
XRS 25
Page 25 of 176
For internal use only
NOTE The pressure indicator at the expansion vessel shows the
actual pressure in the oil circuit after the oil hose from expan-
sion vessel has been connected.
Check/adjust the oil pressure.
Fig. 10: Adjusting the tube pressure
Check the pressure at the pressure indicator (item 2) attached to the expan-
sion vessel using the expansion handle (item 1).
The correct pressure for tubes without a label is 1.05 bar +/- 0.05 bar at
20C at sea level. See following table for altitude correction of pressure.
Newer tubes are clearly labelled with a sticker that requests a pressure of
1.25 bar at sea level.
See following table for altitude correction of pressure.
If the system is located at a higher altitude than sea level > add the following
correction factor to the sea level pressure (see following table)
For example, Potosi (Bolivia) is located at an altitude of 4000m.
If an older 1.05bar tube needs to be adjusted there, the correct pres-
sure would be 1.05bar + 0.39bar = 1.44bar
In Mexico City (Mexico, altitude 2300m), the same tube would be
adjusted to 1.3 bar, and in Hamburg (Germany, app. sea level) it would
be 1.05bar.
Likewise, for newer1.25bar tubes, the correction factor has to be con-
sidered.
Altitude [m] Difference [bar]
0 0.00
500 0.06
1000 0.12
1500 0.17
26 XRS
SOMATOM Definition C2-028.840.01.09.02 Siemens, 2006
08.11 H CX CS SD CR-CT
Page 26 of 176
For internal use only
Unlock the safety bolt for gantry rotation.
Rotate the gantry by hand and check for unusual noise.
Close all covers opened before.
Switch on circuit breakers F2, F3, F5, F6, F7, and F11 in the PDC A and F2 in the PDC
B.
First switch the F2, F3, F5, F6 circuit breakers to the 0 position and then to the
1 position.
Switch the system to the SYSTEM/ON status at the control box.
HV plugs with arcing tracks 0
HV plugs and insulators with arcing tracks
Fig. 11: HV plug with signs of arcing
2000 0.22
2500 0.27
3000 0.31
3500 0.35
4000 0.39
4500 0.43
5000 0.47
If the pressure is not correct, adjust the membrane position using the adjust-
ment handle (item 1).
After the pressure adjustment is finished, remove the expansion vessel
(remove the oil quick-snap connection to the XRT).
Attach the second oil hose from the XTC again to the tube.
Altitude [m] Difference [bar]
Siemens, 2006 C2-028.840.01.09.02 SOMATOM Definition
08.11 H CX CS SD CR-CT
XRS 27
Page 27 of 176
For internal use only
Fig. 12: HV plug with signs of arcing
Fig. 13: HV insulator with signs of arcing
28 XRS
SOMATOM Definition C2-028.840.01.09.02 Siemens, 2006
08.11 H CX CS SD CR-CT
Page 28 of 176
For internal use only
HV plug installation at HV sockets 0
1. Prepare the high voltage connectors and test plugs.
Fig. 14: Silicon oil on HV plugs
Fig. 15: Silicon oil on HV plugs
Check to ensure the connector is clean and inspect it for damage.
Apply oil (supplied with the part) to the entire plug surface (left figure, item 1).
Use about half of the oil from the tube for both plugs.
Put oil into the new silicon insulator (item 2).
Install the silicon insulator part over the plugs (left figure, item 1)
Ensure there is no air left in the silicon insulator.
Siemens, 2006 C2-028.840.01.09.02 SOMATOM Definition
08.11 H CX CS SD CR-CT
XRS 29
Page 29 of 176
For internal use only
2. Prepare the high voltage connectors.
3. Tighten the HV plugs.
High voltage tests 0
Description of available HV tests.
Fig. 16: Silicon oil on HV plugs
Fig. 17: HV sockets
Check that there is no air left in the silicon insulator.
Insert the O-ring onto the HV plug (item 1).
Put oil onto the silicon insulator (item 2).
Insert the high voltage plugs into the receptacles.
Verify the position of the plugs (+ or -).
Install the O-ring into the HV socket as shown (right figure, arrow).
Hand-tighten the fastening rings.
Fig. 18: Tightening the HV plugs
Tighten the HV plugs using the special tightening tool (delivered with the
system) with a torque of 25 Nm.
30 XRS
SOMATOM Definition C2-028.840.01.09.02 Siemens, 2006
08.11 H CX CS SD CR-CT
Page 30 of 176
For internal use only
HVC tube test
For high tension arcing or HV breakdown, the test utility HVC tube test is available in the
Tube/Generator test tools platform. The HVC tube test lets you apply high tension to the
HV circuit without tube current. No test plugs are used in this test. No radiation is released
during the test procedures. This test should always be performed before an HVC plug test
is started. If this test does not indicate an error, the HVC plug test will not indicate an error
either. This may lead to wrong part replacements.
Prerequisites
Test needs system in standby status.
Performing the HVC tube test
1. Select Local Service -> Test Tools -> Generator --> HVC tube..
2. Select Ttube A or B.
3. Click Go on the virtual user panel. Follow the instructions in the dialog.
4. Click the Start button at the control box.
The test runs for 80 seconds. The test is terminated automatically with an error mes-
sage from UMAR (scan time exceeded).
If test fails perform the HVC plug test described next to select which XRS part is
defective.
HVC plug test
For high tension arcing or HV breakdown, the primary test utility HVC plug test is available
in the Generator test tools platform. The HVC plug test lets you apply high tension to the
HV circuit without tube current. This allows the HV circuit to be systematically reduced
using 2 types of test plugs to isolate a high tension fault to either the Straton MX tube, HV
tank, or the HV cables. No radiation is released during this test procedure.
Prerequisites
Dummy plugs (mini type) -> part no.10 093 700 types for plug into the HV sockets.
Delivered with the system.
Dummy plugs (normal type) -> part no. 16 21 791 types for plug into the HV tank.
NOTE The HVC plug test is only allowed with test plugs installed.
Tests with the installed tube can result in misdiagnosis of
failures. For the installed tube, use the HCV tube test instead.
Performing the HVC plug test
1. Install the relevant test plugs in the HV sockets (front) or HVT.
a) Switch off system (select System -> END).
Siemens, 2006 C2-028.840.01.09.02 SOMATOM Definition
08.11 H CX CS SD CR-CT
XRS 31
Page 31 of 176
For internal use only
b) Press service switch S1 in PDC_A (Service Push-Button S1 in PDC-A / p. 11).
c) Open the gantry covers as needed.
d) Disable gantry rotation with switch S 301.
e) Remove HV plugs and install the test plugs.
HV sockets at front: When installing test or HV plugs, new insulators must
be used.
HVT sockets: When installing test or HV plugs, new silicon disks must be
used.
f) Switch system on (S301 remains switched off).
S301 has to be switched off because during Init the Rot FW searches for an
index pulse. This may result in a 360 degree rotation of the gantry. In this
case the removed HV cables may damage parts in the gantry.
2. Select Local Service -> Test Tools -> Generator -> HVC plugs.
3. Select HVC plug A or HVC plug B test.
4. Click Go on the virtual user panel. Follow the instructions in the dialog. Important work
steps are listed below:
5. Click the Start button at the control box.
The test runs for 80 seconds. The test is terminated automatically with error message
from UMAR (scan time exceeded).
-> If the test fails, follow the HV troubleshooting guide (High voltage / p. 21).
NOTE New silicon insulators at the HV sockets (HV plug installation
at HV sockets / p. 28)) as well as new silicon disks in the HVT
have to be used for the HV cable and test plugs before the
plugs are installed into the components (HV sockets / HVT).
32 XRS
SOMATOM Definition C2-028.840.01.09.02 Siemens, 2006
08.11 H CX CS SD CR-CT
Page 32 of 176
For internal use only
Filament 2.4
TSG filament 0
The filament function is tested during Init (e.g., COPMP/ON -> SYS/ON). After switching
the gantry off/on, a self-test of the D770 (FIL power) in XGR is performed as step 1. In step
2, a filament current is sent to the filament heating in the XRT (X-ray tube), passing the fil-
ament transformer at XRT. If the self-test fails, the error message XRS_111 (XGR FIL Self-
test Error) is sent to the logbook.
Additionally a test utility "FIL" is available in the "Tube/Generator" test tools platform. During
the "INIT" filament test, the involved D700/D770 board components in the XGR and the fil-
ament current to the XRT are tested (only in/out test). With the "FIL" test, the involved fila-
ment components and the filament regulation function are tested. So the "FIL" test is a
more detailed test option.
Remark: Test can be used if system is in or is not in standby status.
Tab. 3 Filament test
Filament 0
Filament test
This test checks the function of the filament power board (FIL-Power) in XGR box, the cable
connection to the XRT, the filament transformer at XRT, and the filament heating in the tube
itself. No radiation is released during this test procedure.
Performing the filament test
1. Select Local Service -> Test Tools -> Tube/Generator.
2. Select FIL
3. Click GO and follow the instructions in the dialog. The test is terminated automatically.
Step Action
1 Switch gantry off/on using the control box (COMP/ON -> SYS/ON) to per-
form an XGR INIT
2 Check in the EventLog for the error message CT_GSA/GRA_20 (XGR FIL
Selftest Error).
-> if the error message was sent to the logbook during INIT -> follow the trou-
bleshooting instructions given in the error message.
-> if the error message was not sent to the logbook during INIT -> continue
with item 3.
3 Perform the "FIL" test function.
-> if the "FIL" test fails -> change 1. XRT and 2. XGR assembly
Siemens, 2006 C2-028.840.01.09.02 SOMATOM Definition
08.11 H CX CS SD CR-CT
XRS 33
Page 33 of 176
For internal use only
Anode rotation (RAC) 2.5
TSG Anode Rotation (RAC) 0
An RAC frequency test function is available beginning with VA20. Use this test function to
verify the RAC function. There are different anode speed (Hz) selections possible. In the
troubleshooting instructions below, 160 Hz is selected because this is very stressful for the
RAC function (Note: Do not use this selection in quick succession to avoid overload).
An RAC extended test function is available beginning with VA20. Use this test function to
check the rotating anode revolution with respect to the time (acceleration). The following
statements are possible:
- No revolution -> e.g., rotating anode bearing defective, bad cable/plug connection, XGR...
- Revolution in the medium revolution range is too low -> e.g., rotating anode motor defec-
tive, XGR....
- Upper revolution range not reachable -> e.g., rotating anode bearing defective, XGR...
Note: Test needs the system in standby status.
Note: Anode rotation on a Staton MX P tube may not be started without a functioning and
connected cooling circuit (e.g., for troubleshooting).
Tab. 4 Anode rotation (RAC) test
Step Action
1 Perform AnodeRot Frequency Local Service > Test Tools > Tube/Generator
> RAC and select 160 Hz.
-> if the RAC test fails -> continue with item 2.
-> if the RAC test is ok -> no further action.
2 Perform AnodeRot Frequency Local Service -> Test Tools -> Tube/Generator
and select the following RAC extended tests:
1. RAC anode movement test
2. RAC acceleration test
3. RAC maximum speed test
If test 1 fails -> continue with item 3
If test 2 fails -> continue with item 4
If test 3 fails -> continue with item 5
3 Possible cause of error: no revolution. 1. Check cable connection W604
between XGR (X6) > XTA (X2) for broken wire or bad contact. 2. Rotating anode
motor defective > change XRT. 3. XGR assembly defective > change XGR
assembly.
4 Possible cause of error: revolution in the medium revolution range is too low. 1.
Rotating anode bearing defective > change XRT. 2. XGR assembly defective >
change XGR assembly.
5 Possible cause of error: upper revolution range not reachable. 1. Rotating anode
bearing defective > change XRT
34 XRS
SOMATOM Definition C2-028.840.01.09.02 Siemens, 2006
08.11 H CX CS SD CR-CT
Page 34 of 176
For internal use only
Single pulse test 2.6
Single pulse test 0
The single pulse test allows you to identify problems from the inverter assemblies up to the
HV_Tank by triggering the inverters for a single pulse. Use the single pulse test after per-
forming the TSG High voltage (High voltage tests / p. 29) as a last step or if there is
IMAX/UB MAX parameter information identified in an error message. Please use a memory
oscilloscope to measure the output of the single pulse test.
Tab. 5 Single pulse test
Prerequisites
Memory oscilloscope
Step Action
1 Perform the single pulse test (Performing the single pulse test / p. 35) with a nor-
mally connected tube.
In addition to the given test result, check the UT_act value in the relevant
test message in the Eventlog.
If UT_act parameter is E 0x50 0x02 0x57 0x1B 0x... 0x... 0x00 > test was
successful. No problem on inverter part.
If UT_act parameter is E 0x50 0x00 0x4D 0x09 0x... 0x... 0x00 > test was
not successful. Continue with item 2.
2 Perform single pulse test (Performing the single pulse test / p. 35) with installed
test plugs (#1621791) (HVC plug test / p. 30). Follow the dialog of the HVC plug
test to install the test plugs
In addition to the given test result, check the UT_act value in the relevant
test message in the Eventlog.
If UT_act parameter is E 0x50 0x02 0x57 0x1B 0x... 0x... 0x00 > test was
successful > X-ray tube problem > exchange X-ray tube.
If UT_act parameter is E 0x50 0x00 0x4D 0x09 0x... 0x... 0x00 > test was
not successful. Continue with item 3.
3 Compare the output diagram of the scope, the enclosed example diagrams and
tolerance values shown below.
The time for a single pulse has to be 9 microseconds +/- 20%.
The current should have a peak value of +170A +/- 20% or -170A +/- 20% (alter-
nating).
The tube voltage (UT_act) should have a peak value above 10 kV.
If the current is too high and/or the tube voltage is too low, check the
MVT and HVT cabling (short circuit)
If cabling is o.k. > replace parts in the following order:
1. MVT
2. HVT
3. Relevant inverter
Siemens, 2006 C2-028.840.01.09.02 SOMATOM Definition
08.11 H CX CS SD CR-CT
XRS 35
Page 35 of 176
For internal use only
Performing the single pulse test
1. Move the XGS_Control assembly out of the PDC to get better access.
NOTE The power at the XGS control is 24 V and 5 V. Do not touch
any parts on the XGS control. Connect the probes only as
described below.
To get better information, it is recommended to use a 3 chan-
nel scope as described below. If only a 2 channel scope is
available, measure UT_act first and then I_load to get the
necessary information.
2. Connect scope:
- Channel 1
- XGS D701 > X211 UT_act (1 V = 20KV)
- ANA_GND > XGS D700 > X87
- Adjust scope to 1 V/div
- Channel 2
- XGS D701 > X5 I_load (1 V = 50A)
- ANA_GND > XGS D700 > X87
- Adjust scope to 1 V/div
- Channel 4
- XGS D700 > X104 N_Start_inv
- GND > XGS D700 > X112
- Adjust scope to 5 V/div
3. Adjust scope
- Time base 20 microseconds
- Trigger channel 4 (neg. slope, single)
4. Select Local Service > Test Tools > Tube/Generator
5. Select Unset all in the generator application.
6. Select Single Pulse
7. Press the Start button on the virtual user panel.
The test is terminated automatically.
The time for a single pulse has to be 9 microseconds +/- 20%.
The current should have a peak value of +170A +/- 20% or -170A +/- 20% (alter-
nating).
The tube voltage (UT_act) should have a peak value above 10 kV.
36 XRS
SOMATOM Definition C2-028.840.01.09.02 Siemens, 2006
08.11 H CX CS SD CR-CT
Page 36 of 176
For internal use only
Single pulse test diagrams 0
The first 2 images are good examples; the last 2 images are error examples.
Fig. 19: Single pulse test diagram
Siemens, 2006 C2-028.840.01.09.02 SOMATOM Definition
08.11 H CX CS SD CR-CT
XRS 37
Page 37 of 176
For internal use only
Fig. 20: Single Pulse test diagram
Error examples
Short-circuit at the oscillating current cable (MV cable) in the HV tank
38 XRS
SOMATOM Definition C2-028.840.01.09.02 Siemens, 2006
08.11 H CX CS SD CR-CT
Page 38 of 176
For internal use only
Fig. 21: Single pulse test diagram
Short-circuit at inverter output
Fig. 22: Single pulse test diagram
Siemens, 2006 C2-028.840.01.09.02 SOMATOM Definition
08.11 H CX CS SD CR-CT
XRS 39
Page 39 of 176
For internal use only
Arcing 2.7
The Arcing test is used from the lab to simulate arcings to verify that the FW/SW reaction
due to arcings is correct. This test is not service-relevant.
40 XRS
SOMATOM Definition C2-028.840.01.09.02 Siemens, 2006
08.11 H CX CS SD CR-CT
Page 40 of 176
For internal use only
XRS hints 2.8
This is a summary of troubleshooting hints obtained from experience.
Tab. 6 XRS hints
Step Action
1 The XGS_PDC control assembly A/B are identical. XGS_PDC control A assem-
bly controls the complete switch-on process of most system parts. If this part
becomes defective, the system components will not switch on. For trouble-
shooting and to ensure that the system components are switched on again
(in case control assembly A is defective), exchange the XGS_PDC control A
in PDC_A with XGS_PDC control B in PDC_B.
1. Exchange XGS_PDC control A in PDC_A with XGS_PDC control B in
PDC_B in accordance with the Replacement of parts PDC instructions.
After replacement, the guided tour for replacing parts has to be performed
for the XGS_PDC_Control.
2. Always install the Dongle plug (including the 120 ohm CAN bus resistor) at
XGS_PDC control B plug X3.
2 The inverter 1 and inverter 3 cable connections can be exchanged for trouble-
shooting (both inverters are identical).
1. Remove inverter plug X23 (inverter 1) and X43 (inverter 3) from the
XGS_PDC_Control backplane.
2. Install inverter plug X23 (inverter 1) in the X43 backplane plug and X43
(inverter 3) in X23 backplane plug on the XGS_PDC_Control backplane.
If the error moves with the inverter -> inverter is defective.
If the error does not move with the inverter -> inverter and inverter
cable ok, check for other problems such as the cable connection,
ground connection, or check for other error messages produced
along with the inverter error message. Follow the instructions
given in this message.
Siemens, 2006 C2-028.840.01.09.02 SOMATOM Definition
08.11 H CX CS SD CR-CT
XRS 41
Page 41 of 176
For internal use only
Switching on status display of XGS 2.9
The seven-segment display on XGS board D700 (see image below) indicates the XGS sta-
tus from switching on the system up to scan execute. The status information of the display
is to be used to obtain information as to the system state when a problem occurs. This infor-
mation can be used for troubleshooting.
Fig. 23: D700 measuring points and LED status
42 XRS
SOMATOM Definition C2-028.840.01.09.02 Siemens, 2006
08.11 H CX CS SD CR-CT
Page 42 of 176
For internal use only
Fig. 24: XRS Status
Siemens, 2006 C2-028.840.01.09.02 SOMATOM Definition
08.11 H CX CS SD CR-CT
XRS 43
Page 43 of 176
For internal use only
Measuring points 2.10
Measuring points D700, D701, D702, D703 0
D700 measuring points
Fig. 25: D700 main board
Tab. 7 Oscillating current measuring points
Tab. 8 Oscillating current and dose act. measuring points
item signal name short name valid
at
description measur-
ing point
1 I_load_MVT I_L_MVT XGS oscillating current MVT (sec-
ondary) 1 V = 50 A.
X66
1 I_load_INV5 I_L_Inv5 XGS oscillating current Inverter 5 / 1
V = 50 A.
X67
item signal name short name valid
at
description measur-
ing point
2 I_load_INV3 I_L_Inv3 XGS oscillating current Inverter 3 / 1
V = 50 A.
X77
2 Dose_act Dose_act XGR actual dose value from fast link
1 V = 13.107 value
* other similar values with
D703 Fastlink control register
selectable.
X76
44 XRS
SOMATOM Definition C2-028.840.01.09.02 Siemens, 2006
08.11 H CX CS SD CR-CT
Page 44 of 176
For internal use only
Tab. 9 Voltage supply measuring points
Tab. 10 Ground measuring points
Tab. 11 HW interrupt for ignition pulses main inverter
Tab. 12 Tube voltage measuring points
item signal name short name valid
at
description measur-
ing point
3 P15V P15V XGS +15 V Voltage supply X69
3 N15V N15V XGS -15 V Voltage supply X71
3 UDC UDC Intermittent voltage X73
3 VCC VCC XGS + 5 V Voltage supply X75
3 ANA_GND ANA_GND XGS analog ground X78
3 GND GND XGS ground X80
item signal name short name valid
at
description measur-
ing point
4 VCC VCC XGS +5 V voltage power supply X75
4 ANA_GND ANA_GND XGS analog ground X78
item signal name short name valid
at
description measur-
ing point
5 S1 S1 XGS Ignition switch: HW interrupt of
the ignition pulses from the
main inverter. Normal position
is off - V162 must be off. No
function in XGR. Normal posi-
tion is off.
n.a
item signal name short name valid
at
description measur-
ing point
6 P_UT act P_UT act XGR positive tube voltage 1 V = 10
kV.
X81
6 N_UT act N_UT act XGR negative tube voltage 1 V = 10
kV.
X83
6 UT_act UT_act XGR Tube voltage 1 V = 20 kV. X85
6 ANA_GND ANA_GND XGR analog ground . X87
Siemens, 2006 C2-028.840.01.09.02 SOMATOM Definition
08.11 H CX CS SD CR-CT
XRS 45
Page 45 of 176
For internal use only
Tab. 13 IT act. measuring points
Tab. 14 Service switch
Tab. 15 IT act. measuring points
Tab. 16 FPGA voltage measuring points
Tab. 17 FPGA voltage measuring points
item signal name short name valid
at
description measur-
ing point
7 MA_meas_1 1 XGR IT_act measurement point 1
for ampere meter.
X61
7 MA_meas_2 IT_nom XGR IT_act measurement point 2
for ampere meter.
X59
Item Signal
name
short name valid
at
Description Measur-
ing point
8 S5 S5 XGR Switch must be on for IT_act
measurements using a
ampere meter at measuring
points X46, X48, X50, X59,
X61. LED V202 is on if switch
is activated. Normal position is
off.
n.a
item signal name short name valid
at
description measur-
ing point
9 IT_act IT_act XGR actual tube current X48
9 IT_nom IT_nom XGR nominal tube current X50
9 ANA_GND ANA_GND XGR analog ground X46
item signal name short name valid
at
description measur-
ing point
10 VCC3.3 VCC3.3 XGS +3.3 V I/O voltage FPGA X20
10 VCCINT VCCINT XGS +1.5 V core voltage X44
item signal name short name valid
at
description measur-
ing point
11 S2 S2 XGS
XGR
Service switch: Switch has no
function at this time. Neverthe-
less, switch must be in off posi-
tion. LED V6 must be off.
n.a
46 XRS
SOMATOM Definition C2-028.840.01.09.02 Siemens, 2006
08.11 H CX CS SD CR-CT
Page 46 of 176
For internal use only
Tab. 18 FPGA digital out
Tab. 19 FPGA digital out
D701 HVC board measuring points
Fig. 26: D701 measuring points
Item signal name short name valid
at
description measur-
ing point
12 N_Start_FIL
_INV
FIL XGR Start FIL inverter (low active).
Can only be used in XGR.
X100
12 N_Start_RA
C_INV
RAC XGR Start RAC inverter (low active).
Can only be used in XGR.
X102
12 N_Start_HV
_INV
INV XGS Start HV inverter (low active). X104
12 XC XC XGS XC signal (X-ray request) X106
12 XRAY_on XRAY XGS XRAY on signal (> 20 kV) X108
12 INV_Fault Inv_err XGS Summary error HV inverter
Can only be used in XGS.
X110
12 GND GND XGS
XGR
Ground X112
Item signal name short name description measur-
ing point
13 S3/S4 n.a Dip switches for debug and configura-
tion of the FPGA - not for service use.
n.a
Siemens, 2006 C2-028.840.01.09.02 SOMATOM Definition
08.11 H CX CS SD CR-CT
XRS 47
Page 47 of 176
For internal use only
Tab. 20 D701 HVC board measuring points
D702 FIL/RAC board
Tab. 21 D702 FIL measuring points
Item signal name short name valid
at
description measur-
ing point
1 UT_act UT_act XGS Actual tube voltage (from rotat-
ing part over fast link) 1 V = 20
kV
X211
1 U_control U_con XGS Set value for kV regulator X216
1 I_load_act I_L_A XGS Actual oscillating current. 1 V =
50 A
X215
1 USTU USTU XGS Set value for kV regulator.
(Voltage control cycle). -5 V ...
0 V
X208
1 I_load_nom I_L_N XGS Nominal oscillating current. 1
V = 50 A
X207
1 UDC_act UZ_act XGS Actual intermediate circuit volt-
age
X210
1 ANA_GND A_GND XGS analog ground X209
1 I_load I_load XGS Oscillating current. 1 V = 50 A X5
1 UT_nom UT_nom XGS Nominal tube voltage X204
1 ANA_GND A_GND XGS analog ground X206
1 USTI USTI XGS Set value for kV regulator (cur-
rent regulator circuit) 0 .... 5 V
X203
1 Upre_con Upre XGS Set value for kV regulator (pre-
liminary control) 0 ... 5 V
X201
1 ANA_GND A_GND XGS analog ground X200
Item signal name short name valid
at
description mea-
suring
point
D702 IT_act IT_act XGR actual tube current
max. 80 mA -> 1 V = 100 mA
<80 mA -> 1 V = 80 mA
X18
D702 IF_act_1 IF_act_1 XGR actual filament heating 1
1 V = 0.273 A
X19
D702 IF_act_2 IF_act_2 XGR actual filament heating 2
1 V = 0.273 A
X20
48 XRS
SOMATOM Definition C2-028.840.01.09.02 Siemens, 2006
08.11 H CX CS SD CR-CT
Page 48 of 176
For internal use only
Tab. 22 D702 RAC measuring points
D732 HV acquisition
HVT (high voltage tank) measuring points
Tab. 23 D732 HVT measuring points
D702 IF_act_3 IF_act_3 XGR actual filament heating 3
1 V = 0.273 A
X21
D702 ANA_GND A_GND XGR analog ground X17
board signal name short name valid
at
description mea-
sur-
ing
point
D702 Test_DAC_1 DAC_1 XGS FPGA controlled debug DAC X10
D702 Test_DAC_2 DAC_2 XGS FPGA controlled debug DAC X11
D702 Res_sin sin XGS Resolver Sine Signal X12
D702 Res_cos cos XGS Resolver Cosine Signal X13
D702 RAC_P3 P3 XGS RAC Strom Phase 3(1 V = 40
A, Offset 2.5 V)
X14
D702 RAC_P2 P2 XGS RAC Strom Phase 3(1 V = 40
A, Offset 2.5 V)
X15
D702 RAC_P1 P1 XGS RAC Strom Phase 3 (1 V = 40
A, Offset 2.5 V)
X16
D702 ANA_GND ANA_GND XGS analog ground X17
Item signal name short name valid
at
description mea-
suring
point
D732 ANA_GND ANA_GND XGR analog ground X12
D732 P_UT_act P_UT_act XGR Tube voltage (pos.) (1 V = 10
kV)
X13
D732 N_UT_act N_UT_act XGR Tube voltage (neg.) (1 V = 10
kV)
X14
Item signal name short name valid
at
description mea-
suring
point
Siemens, 2006 C2-028.840.01.09.02 SOMATOM Definition
08.11 H CX CS SD CR-CT
XRS 49
Page 49 of 176
For internal use only
LEDs on D700, D701, D702, D703 0
D700 LED status
Fig. 27: D700 measuring points and LED status
Tab. 24 D700 main board
board signal name valid at description LEDs
D700 S_FIL_INV XGR Start FIL Inverter. Only used in XGR. V7
D700 S_RAC_INV XGR Start RAC Inverter. Only used in XGR. V8
D700 S_HV_INV XGR/XGS Start HV Inverter (at XGR signal
delayed due to transport (fastlink) delay.
V9
D700 Cycle LED XGR/XGS FPGA Life Cycle LED V10
D700 XC XGR/XGS XC Signal from system V11
D700 XRAY_On XGR/XGS XRAY On Signal (UT_act > 20 kV) V12
D700 INV_Err XGS Inverter Error V13
D700 FPGA_Err XGR/XGS Error Interrupt from FPGA V14
D700 MB 1000 mA XGR IT_act measurement range 1000 mA V15
D700 MB 100 mA XGR IT_act measurement range 100 mA
(amplification *10)
V16
D700 MB 25 mA XGR IT_act measurement range 25 mA
(amplification *40)
V17
D700 D_UT XGR Delta UT_act monitoring V18
50 XRS
SOMATOM Definition C2-028.840.01.09.02 Siemens, 2006
08.11 H CX CS SD CR-CT
Page 50 of 176
For internal use only
D701 HVC board
Tab. 25 D701 main board
D702 FIL/RAC
Tab. 26 D702 main board
D700 STOP XGR/XGS Stop Report Loop V20
D700 UT_MAX XGR UT_act max. monitoring V21
D700 A XGR/XGS System A on System B off V23
D700 rot XGR/XGS rotating on, stationary off V24
board signal name valid at description LEDs
D701 LED 1 n.a not used V205
D701 LED 2 n.a not used V206
D701 LED 3 n.a not used V207
D701 LED 4 n.a HVC Error Interrupt V208
board signal name valid at description LEDs
D702 n.a XGR FPGA out LED 0lim_1: Lim regulator
FIL_1 active
LED_0
D702 n.a XGR FPGA out LED 1 20peak_1: Peak regu-
lator FIL_1 active
LED_1
D702 n.a XGR FPGA out LED 2 20STOP_FIL/
STOP_RAC.
LED_2
D702 n.a XGR FPGA out LED 3 20RAC_State_0. LED_3
D702 n.a XGR FPGA out LED 4 20RAC_State_1 LED_4
D702 n.a XGR FPGA out LED 5 20HALL_A LED_5
D702 n.a XGR FPGA out LED 6 20HALL_B LED_6
D702 n.a XGR FPGA out LED 7 20FIL_Error_Int /
RAC_Error_Int
LED_7
board signal name valid at description LEDs
Siemens, 2006 C2-028.840.01.09.02 SOMATOM Definition
08.11 H CX CS SD CR-CT
XRS 51
Page 51 of 176
For internal use only
D703 XG board
Tab. 27 D703 main board
D704 dongle board
Tab. 28 D704 dongle board
D790 PDC board
Fig. 28: D790 signal LEDs
Item signal name valid at description LEDs
D703 LinkErr XGS/XGR Link Error LED[5]
D703 LightErr XGS/XGR Light Error LED[4]
D703 CfgErr XGS/XGR Config Error (FPGA not programmed) LED[3]
D703 DefCfg XGR xray_and_not_door (XGR) LED[2]
D703 RunCfg XGS door closed (XGS) LED[1]
D703 Prg XGR door closed (XGR via Fastlink) LED[0]
D703 LED 1 XGS XGS_off (XGS indication when starting
inverter)
LED[6]
D703 LED 2 XGS/XGR LifeCycle (flashes if FPGA is loaded) LED[7]
board signal name valid at description LEDs
D704 VCC XGS/XGR Power supply 5 V V1
D704 P_15V XGS/XGR Power supply 15 V V2
52 XRS
SOMATOM Definition C2-028.840.01.09.02 Siemens, 2006
08.11 H CX CS SD CR-CT
Page 52 of 176
For internal use only
Tab. 29 D790 PDC board LEDs
board signal name valid at description LEDs
D790 P_24V_SNT XGS P_24V from switched power supply
(only used for relays, feedback, and
fans)
V1
D790 VCC_SNT XGS VCC (5 V) from switched power supply V2
D790 VCC3.3 XGS FPGA IO voltage (3.3 V) V3
D790 LED morse XGS FPGA Lifecycle LED (if LED flashes ->
FPGA is loaded and is functioning)
V4
D790 P_15V_SNT XGS P_15V from switched power supply V5
D790 VCC_INT XGS FPGA core voltage (1.5 V) V6
D790 N_15V_SNT XGS N_15V from switched power supply V7
D790 P_15V_SNT
_AUX
XGS P_15V from switched power supply for
inverter und fans.
V8
Siemens, 2006 C2-028.840.01.09.02 SOMATOM Definition
08.11 H CX CS SD CR-CT
XRS 53
Page 53 of 176
For internal use only
Fig. 29: D790 PDC board signal LEDs
Tab. 30 D790 PDC board LEDs
board signal name valid at description LEDs
D790 Main_Sw XGS PDC Main Switch Breaker Signal V105
D790 Cooling_CB XGS Cooling Unit Breaker Signal V117
D790 XGS XGS XGS Breaker Signal V116
D790 PHS_CB XGS PHS Circuit Breaker Signal V138
D790 Gan_Stat XGS Gantry Stationary Breaker Signal V144
D790 Gan_Rot XGS Gantry Rotating Breaker Signal V143
D790 UPS XGS UPS Breaker Signal V157
D790 Insul_Power XGS Insulation Power Breaker Signal V158
54 XRS
SOMATOM Definition C2-028.840.01.09.02 Siemens, 2006
08.11 H CX CS SD CR-CT
Page 54 of 176
For internal use only
Tab. 31 D790 PDC board LEDs
board signal name valid at description LEDs
D790 Cooling XGS Cooling Unit Signal V106
D790 PHS XGS PHS Signal
LED off -> trigger and feedback off; LED
on -> trigger and feedback on; LED
flashing -> trigger and feedback differ-
ent
V118
D790 L2_Perm_Ga
ntry
XGS L2 Gantry Signal
LED off -> trigger and feedback off; LED
on -> trigger and feedback on; LED
flashing -> trigger and feedback differ-
ent
V119
D790 Charg_XGR XGS Charging XGR Signal
LED off -> trigger and feedback off; LED
on -> trigger and feedback on; LED
flashing -> trigger and feedback differ-
ent
V139
D790 PWR_XGR XGS Power XGR Signal
LED off -> trigger and feedback off; LED
on -> trigger and feedback on; LED
flashing -> trigger and feedback differ-
ent
V145
D790 IRS XGS IRS Signal
LED off -> trigger and feedback off; LED
on -> trigger and feedback on; LED
flashing -> trigger and feedback differ-
ent
V146
D790 ICS_Ether XGS ICS Signal
LED off -> trigger and feedback off; LED
on -> trigger and feedback on; LED
flashing -> trigger and feedback differ-
ent
V160
D790 Charg_XGS XGS Charging XGS Signal
LED off -> trigger and feedback off; LED
on -> trigger and feedback on; LED
flashing -> trigger and feedback differ-
ent
V159
Siemens, 2006 C2-028.840.01.09.02 SOMATOM Definition
08.11 H CX CS SD CR-CT
XRS 55
Page 55 of 176
For internal use only
Tab. 32 D790 PDC board LEDs
board signal name valid at description LEDs
D790 PWR_XGS XGS Power XGS
LED off -> trigger and feedback off; LED
on -> trigger and feedback on; LED
flashing -> trigger and feedback differ-
ent
V107
D790 Temp_Trans XGS Temp Transformer Signal V120
D790 Power_CTRL XGS Power Control Signal V121
D790 Diff_Sw XGS Diff Switch Signal V140
D790 Volt_Prot XGS Voltage Protection Signal V147
D790 Current Loop XGS Current Loop Signal V148
D790 Door XGS Door Switch Signal
LED On: Door open (in PDC_B always
on)
V162
D790 Fan Mode XGS Fan Mode
LED On: low noise mode (15 V)
V161
56 DMS
SOMATOM Definition C2-028.840.01.09.02 Siemens, 2006
08.11 H CX CS SD CR-CT
Page 56 of 176
For internal use only
3- 3DMS
General 0
Safety 0
WARNING
[ hz_serdoc_F13G01U12M03 ]
Avoid accident and injury or damage to parts.
Risk of accident and injury!
Read and observe the safety information contained in
the General section of this document and/or the
Product specific safety notes.
NOTE Follow ESD guidelines when handling the DMS.
Notes 0
There are 2 Types of DMS:
TACH- DMS and AMS DMS. Please check SPC for partnumbers, some spare parts are
specific for one or the other type.
Most noticeable are the colours of the handles of the modules: TACH modules have a grey
handle, AMS modules have a pink handle.
1. Before starting troubleshooting, be sure you troubleshoot the right area.
- To distinguish between the DMSs and IRS, use the Datalink Test and check the data
path first
- To distinguish between DMS A and DMS B, run the DMS tests and check if the
errors are related to DMS A or DMS B.
- To distinguish between DMS A and DMS B, especially if image quality problems
appear in Dual source modes, it is useful to have a 30 cm water phantom (Service
Tool, also used for IEC Acceptance).
- To distinguish between module and backplane, it is possible to cross the module
cable at the backplane.
If the error moves when crossing the module cables at the backplane, the problem is
related to the module or module cable.
- To distinguish between module and module cable, it is possible to swap the whole
module cable with a neighboring one.
If the error moves when swapping the whole cable, the problem is related to the mod-
ule cable.
NOTE A second peak may appear in the displayed plot with module
cables crossed or swapped. This is normal if no new calibra-
tion (e.g. a new Checkup or Setup/Calibration) has been per-
formed prior to starting a new scan.
Siemens, 2006 C2-028.840.01.09.02 SOMATOM Definition
08.11 H CX CS SD CR-CT
DMS 57
Page 57 of 176
For internal use only
2. If possible, use the test modes with rotation for troubleshooting especially in case of
noise problems.
NOTE Use rotation modes with the collimations shown below and
close all DMS covers before starting rotation.
The two most useful slice combinations for troubleshooting are:
- 32 x 0.6 mm
- 24 x 1.2 mm
Definitions and abbreviations 0
n.a.
Prerequisites 0
1. Check (Environmental Conditions / p. 96) (especially in case of weak rings)
2. Check for objects in the scan field e.g. contrast agent, labels,...
3. Try to reproduce an image the customer has complained about using the same param-
eters
4. Use (ROI/Ring / p. 69) and note the positions as a reference for the position expected
to contain the defect (two positions are possible).
58 DMS
SOMATOM Definition C2-028.840.01.09.02 Siemens, 2006
08.11 H CX CS SD CR-CT
Page 58 of 176
For internal use only
TSG Image Quality 3.1
This chapter provides a basic overview of different types of image quality problems and a
useful way for starting troubleshooting.
General 0
NOTE Important! There is almost no cooling of the DMS electronics
if the metal DMS cover is removed. If the DMS plastic cover
remains installed, there is a risk of overheating. Always
switch the gantry off completely prior to opening and work-
ing at the DMS, to prevent damaging parts.
NOTE Most of the image quality problems are related to defective
channels within a detector module. That is why troubleshoot-
ing should start with defective channel detection. This chap-
ter is sorted for different types of image quality problems and
should guide you. The DMS tests are described later in this
chapter.
Possible reasons for ring and band (mostly 4, 8, or 16 channels) artifacts are:
Defective detector element
Defective detector cable
Defective backplane
Insufficient calibration (e.g. contrast agent at plexiglas ring)
Insufficient tune-up tables
Problem with balancing (reconstruction)
Insufficient environmental conditions
Defective IRS
Bands covering 16 channels might be caused by the incorrect Z-position of one mod-
ule. Run Z-position check of module.
NOTE Problems related to a single slice appear as permanent rings
in sequence modes, and as partial rings in spiral scans.
Possible reasons for streak artifacts are:
Defective detector module
Patched channels
Alignment adjustment
Siemens, 2006 C2-028.840.01.09.02 SOMATOM Definition
08.11 H CX CS SD CR-CT
DMS 59
Page 59 of 176
For internal use only
Tube arcing
Possible reasons for other artifacts, e.g. pattern in image, inhomogeneous areas
are:
The tube
The tube-side collimator
UHR comb (not supported by VA10)
The tube cooling system
The DMS power supply
NOTE If the DMS covers and parts have to be removed during trou-
bleshooting, follow the instructions in the Replacement of
Parts section for detailed informations how to remove and
install DMS parts. Cooling is reduced when the DMS covers
are removed. Never start gantry rotation with the DMS covers
removed. Close the DMS covers before switching on the gan-
try and starting the tests.
TSG strong rings 0
In most cases we suspect that the detector module is defective.
Strong bands (4, 8, or 16 channels) may also result from a defective backplane or DAS
Controller.
NOTE Offset/noise values of an entire backplane that are out of tol-
erance indicate a defective DAS Controller in most cases.
Tests which typically show problems in case of strong rings are:
Offset value
Signal value
Sinogram
Start troubleshooting as follows:
Tab. 33 TSG
Step Action
1 Use (ROI/Ring / p. 69) to determine the area of the defect in the customer
image
2 Check for entries of defective channels in Reports. Refer to section defective
channel detection.
60 DMS
SOMATOM Definition C2-028.840.01.09.02 Siemens, 2006
08.11 H CX CS SD CR-CT
Page 60 of 176
For internal use only
TSG sporadic strong rings 0
Ask the customer to perform a Checkup in case of problems.
Start troubleshooting as follows:
3 Compare the entries in Reports to the channels you found with ROI/Ring.
a) if entries are related to the customer image -> patch out the channel
manually. Refer to section defective channel detection.
b) If no channel is related to the customer image, -> continue with step 4.
4 Run (Defective channel detection / p. 70) in the Tune up platform.
a) if problem is solved -> Run Quality check (Quality Constancy).
b) if not o.k., e.g., too many bad channels found, -> continue with step 5.
5 Run (Guided tour for image quality / p. 73).
a) If there are errors found in one or more tests, -> continue with step 6.
b) If there are no errors found, run (Sinogram / p. 80) for additional informa-
tion.
6 If only one channel is found to be defective, -> patch out the channel man-
ually and check Reports whether the defect has remained in the sys-
tem.
7 If neighboring channels, channels in detector centers, or channels which
cannot be patched out are defective -> continue with step 8.
8 Cross the module cable on backplane with the neighboring module
cable to distinguish between the module and the backplane.
Close all DMS covers.
Repeat the tests which showed errors before and check if the error moves.
a) If error still in initial position -> replace the backplane.
b) If error moves to neighboring slot -> module or module cable defec-
tive, continue with step 9.
9 Swap the complete module cable with the neighboring module cable to dis-
tinguish between the module and the module cable.
Close all DMS covers.
Repeat the tests which showed errors before and check if the error moves.
a) If error still in initial position -> replace the module.
b) If error moves to neighboring slot -> replace the module cable.
10 Additionally, check the power path.
Especially if the errors are related to an entire backplane, -> check the
power supply of the DMS. Refer to (TSG DMS Power path / p. 88)
Step Action
Siemens, 2006 C2-028.840.01.09.02 SOMATOM Definition
08.11 H CX CS SD CR-CT
DMS 61
Page 61 of 176
For internal use only
Tab. 34 TSG
Step Action
1 Ask the customer to save the image in case of problems.
a) The image is important for further evaluations with ROI /Ring.
2 Ask the customer to perform a Checkup in case of problems.
a) This particular (automatically generated) checkup table can be used for
further defective channel evaluations.
3 Use (ROI/Ring / p. 69) to determine the area of the defect in the customer
image.
4 Check for entries of sporadic channels in Reports. Refer to section defective
channel detection.
5 Compare the entries in Reports to the channels you found with ROI/Ring.
a) If sporadic entries are related to the customer image -> patch out the
channel manually. Refer to section defective channel detection.
b) If no channel is related to the customer image, -> continue with step 6.
6 Run (Defective channel detection / p. 70) in the Tune up platform.
a) if problem is solved, -> Run Quality check (Quality Constancy).
b) If not o.k., e.g. too many defective channels found, -> continue with step
7.
7 Run (Guided tour for image quality / p. 73).
a) If there are errors found in one or more tests, -> continue with step 8.
b) If no errors are found, -> continue with step 9.
8 If one channel is indicated as defective, -> patch out the channel manu-
ally.
9 Use Reports and check for irregularities within the tests, especially in the
area of the defect in the customer image.
If irregularities are present, cross module cable on backplane with neighbor-
ing module cable.
Close all DMS covers.
Repeat the tests which showed irregularities before.
a) Check reports again. If irregularities moved for 16 channels, -> replace
the module including the module cable.
b) If no difference is visible, swap back the cable and close all DMS covers.
c) Run a Checkup and check Reports for defective channels again.
10 Contact HSC for further steps.
62 DMS
SOMATOM Definition C2-028.840.01.09.02 Siemens, 2006
08.11 H CX CS SD CR-CT
Page 62 of 176
For internal use only
TSG weak rings 0
A ring is visible in homogeneous areas such as the cerebrum, if the difference compared
to the neighboring channel is about 1 ppm.
In most cases we expect the stability of the element affected to be insufficient.
Tests which typically show problems in case of weak rings:
BasCal stability
Signal linearity
Start troubleshooting as follows:
Tab. 35 TSG
Step Action
1 Check (Environmental Conditions / p. 96)
2 Ensure that DMS was powered on overnight.
3 Check that the Tune-up was performed correctly (use Report function).
4 Use (ROI/Ring / p. 69)to find out the area of the defect in the customer
image
5 Check for entries of sporadic channels in Reports. Refer to section defective
channel detection.
6 Compare the entries in Reports to the channels you found with ROI/Ring.
a) If entries are related to the customer image, -> patch out the channel
manually. Refer to section defective channel detection.
b) If no channel is related to the customer image, -> continue with step 7.
7 Run (Guided tour for image quality / p. 73) and (Popcorn noise / p. 77) test.
a) Compare test outputs to channels you found in the customer image with
the ROI/Ring function.
b) If only one channel is defective, -> patch out the channel manually.
c) If neighboring channels, channels in detector centers, or channels which
cannot be patched out are defective, -> continue with step 8.
8 Cross module cable on backplane with neighboring module cable.
Close all DMS covers.
Repeat the tests which showed errors before and check if errors move.
a) If error still in initial position -> swap back cable and continue with step
9.
b) If error moves to neighboring slot, -> module or module cable unsta-
ble/defective, continue with step 9 and step 10 prior to replacing the
module.
Siemens, 2006 C2-028.840.01.09.02 SOMATOM Definition
08.11 H CX CS SD CR-CT
DMS 63
Page 63 of 176
For internal use only
TSG sporadic weak rings 0
In case of sporadic weak rings, perform TSG weak rings first.
Additionally, start troubleshooting as follows:
Tab. 36 TSG
TSG rings in outer slices only 0
Rings visible in the outermost slices only might occur in case of collimation problems.
Start troubleshooting as follows:
9 Check the Power path. Refer to (TSG DMS Power path / p. 88)
a) If power path and DMS Power supply are o.k., -> continue with step 10.
10 Perform Guided Table Generation in Tune up platform.
Note: Table Generations for DMS A and DMS B are performed. (approx. 2
hours)
a) If still not o.k. -> contact HSC for further steps.
Step Action
Step Action
1 Ask the customer to save the image in case of problems.
a) The image is important for further evaluations.
2 Ask the customer to perform a checkup in case of problems.
a) This particular checkup table can be used for further basecal evaluations.
3 Use (ROI/Ring / p. 69) to determine the area of the defect in the customer
image.
4 Perform (BasCal Stability / p. 75). A difference between a satisfactory and
unsatisfactory Basecal table will show the problem.
a) If only one channel is noted, -> patch out the channel manually.
b) If neighboring channels, channels in detector centers, or channels which
cannot be patched out are noted -> replace the corresponding mod-
ule.
64 DMS
SOMATOM Definition C2-028.840.01.09.02 Siemens, 2006
08.11 H CX CS SD CR-CT
Page 64 of 176
For internal use only
Tab. 37 High voltage TSG
TSG Streak artifacts 0
1. Intermittent streaks tangential to a circle might be caused by a defective module
Start troubleshooting as follows:
Tab. 38 TSG
2. Streaks tangential to a circle might be caused by too many patched channels.
Start troubleshooting as follows:
Step Action
1 Run (Defective channel detection / p. 70) in the Tune up platform.
a) If more channels or a entire row of channels are out of tolerance, -> con-
tinue with step 2.
b) If only one channel is defective, -> continue with step 5.
2 Run (Module Z-Alignment / p. 83).
a) If out of tolerance, check the mechanical positioning of the module. Refer
to the replacement of parts section for instructions.
b) If in tolerance, -> continue with step 3.
Additionally run modes with 24 x 1.2 mm and 32 x 0.6 mm to check if the
error occurs at the same physical slice position.
3 Run Z-Adjust in the Tune up platform.
a) If out of tolerance, troubleshoot the tube collimator.
b) If in tolerance, repeat defective channel detection. Additionallym con-
tinue with step 4.
4 Check slice thickness to ensure Constancy.
a) Run Quality Constancy, check results for slice measurement.
5 Run TSG strong rings
Step Action
1 Run (Defective channel detection / p. 70) in the Tune up platform.
a) If no channel is found to be defective, -> continue with step 2.
b) If defective channels are detected, -> continue with step 3.
2 Run (Popcorn noise / p. 77) test to isolate the defective channel.
3 Run TSG strong rings
Siemens, 2006 C2-028.840.01.09.02 SOMATOM Definition
08.11 H CX CS SD CR-CT
DMS 65
Page 65 of 176
For internal use only
Tab. 39 TSG
3. Fine streak artifacts not correlated tangentially to any circle might be caused by incor-
rect alignment.
Start troubleshooting as follows:
Tab. 40 TSG
4. streak artifacts not correlated to any object might also be caused by tube arcing.
Start troubleshooting as follows:
Tube arcing
Check the tube history for this paticular scan to see if tube arcing was detected.
Getter tube
TSG Dual Source problems 0
Notes:
Step Action
1 Run (Defective channel detection / p. 70) in the Tune up platform.
2 Check if there are adjacent patched channels and continue with step 3.
3 Check Reports for defective channel detection.
a) If -> change is written in Reports, replace the corresponding module.
b) If -> change is not written in Reports, -> continue with step 4.
4 Run TSG strong rings
Step Action
1 Run (MTF test / p. 78) and check for tolerances.
a) If in tolerance, -> continue with step 2 and step 4.
b) If slightly out of tolerance, -> continue with step 2.
c) If results not plausible, go directly to step 3.
d) If only UHR modes (not supported by VA10) are out of tolerance, -> con-
tinue with step 4.
2 Run (Focus Alignment Check / p. 82) in the Tune up platform.
a) Continue with Focus Alignment in the Tune up platform.
Note: Continue with Focus Alignment as well if the Focus Alignment
check is in tolerance. If Focus Alignment fails, -> continue with step 3.
3 Troubleshoot tube collimator (PSD) and Focus deflection.
4 Run comb ali (UHR is not supported by VA10).
5 Run TSG weak rings.
66 DMS
SOMATOM Definition C2-028.840.01.09.02 Siemens, 2006
08.11 H CX CS SD CR-CT
Page 66 of 176
For internal use only
Image quality problems in dual source modes appear most often as a band in the area
where the scan field of the DMS B is extended by the outer channels of DMS A. In dual
source modes the inner 26 cm of the scan field are measured by both DMSs, the outer
area of the scan field (outside 26 cm) is measured by DMS A only. Measure the HU in
the area of the band. Especially in case of a high absorption (obese patients), a high dif-
ference in HU can be measured in the inner and outer area of the scan field.
NOTE For troubleshooting dual source problems, it is very helpful
to have a 30 cm water phantom as used for the IEC Accep-
tance.