Red (7 mm) 6 6
Blue (9mm) 8
Note: Numeric sizes indicated under manufacturer’s name refer to the instrumentation size labeled by the
manufacturer. Smith & Nephew’s numeric sizes do not correspond to each manufacturer’s sizing. The TRUFIT
CB Plug is uniquely sized for a tight interference fit. Please refer to this chart for all sizing compatibility.
Ordering Information
Additional plug sizes are available for use with other
commercially available osteochondral transplant systems
(see implant systems). The TRUFIT Plug kit contains one of
each of the following:
Kit Components
• Cylindrical Implant, 18 mm long 1. TRUFIT CB Plug
• Delivery Device (outer sleeve and measuring tamp) 2. Delivery Device
a. Measuring Tamp
• Trimming Knife b. Outer Sleeve
3. Trimming Knife
TRUFIT CB Plug Kits
72200930 TRUFIT CB Plug Kit Green (5 mm)
72200931 TRUFIT CB Plug Kit Yellow (6 mm) 2a
72200932 TRUFIT CB Plug Kit Red (7 mm) 1
2b
72200933 TRUFIT CB Plug Kit Tan (7.5 mm)
72200934 TRUFIT CB Plug Kit Blue (9 mm)
3
72200935 TRUFIT CB Plug Kit Orange (9.5 mm)
72200936 TRUFIT CB Plug Kit Purple (11 mm)
Not to scale
Endoscopy www.smith-nephew.com
™Trademarks of Smith & Nephew. Certain marks Reg. US Pat. & TM Office.
Smith & Nephew, Inc. T +1 978 749 1000 All other trademarks acknowledged.
Andover, MA 01810 F +1 978 749 1108 ©2006 Smith & Nephew, Inc. All Rights Reserved.
USA US Customer Service: +1 800 343 5717 Printed in USA 10/06 10600215 Rev. B
TRUREPAIR™ Solutions
9 mm TRUFIT CB Plug immediately post TRUFIT CB Plug 21 months after Biopsy (stained with Safranin 0/Fast
implantation in medial femoral condyle. implantation. Note complete plug Green) 21 months after TRUFIT CB Plug
Note: Blood has wicked into scaffold. integration and defect coverage with Implantation demonstrating healing
normal thickness hyaline-like cartilage. with hyaline-like cartilage and
Small amount of fibrillation due to subchondral bone.
meniscal tear.
Intended Use
The TRUFIT CB Plug is intended to serve as a support for cellular and matrix ingrowth in
osteochondral defect repair such as cartilage and/or bone, that is not intrinsic to the stability
of the bone. The device is a resorbable tissue regeneration scaffold for use in the treatment of
acute focal articular cartilage or osteochondral defects. The product is provided sterile for single
patient use and for implantation into precisely created defects. TRUFIT CB Plug can also be used to
repair osteochondral defects created during autogenous osteochondral transfer procedures. Please
use the size chart to determine the appropriately sized TRUFIT CB Plugs for other commercially
available instrument systems.
Surgical Technique: Osteochondral Repair
Assembly
1 2
Note: If implant
becomes dislodged
from delivery
device, reload into
the window end of
the outer sleeve
with the notched
end of the implant
facing outward.
To assemble the Smith & Nephew TRUFIT At this point, the implant should not extend
Delivery Device, insert the measuring tamp beyond the delivery device.
into the outer sleeve in the direction of the
arrow on the outer sleeve until contact with
the preloaded implant is made.
3 4 5
Note: Caution
should be taken not
to nick the top of
the outer sleeve of
the delivery device.
Insert the measuring tamp of the delivery Slide the outer sleeve down until the lip of Firmly grip the outer sleeve at the windows
device into the defect, ensuring that it the outer sleeve is snugly placed against to secure the implant during cutting.
contacts the bottom of the defect. the surface of the tissue (this automatically Steady the hand and delivery device
adjusts the implant to the correct position against a table or other sterile surface.
for trimming), then carefully remove the Using the lip of the delivery device as a
delivery device from the surgical site. guide, cut the notched end of the implant
with the trimming knife using a firm
downward motion.
Implant Placement
6 7 8
Once the implant has been cut to a flat Carefully seat the implant in the defect and If the implant needs further adjustment or
surface, the delivery device is used to press fit into the defect by pushing on the contouring, separate the measuring tamp
insert the implant into the defect. Slightly measuring tamp manually or lightly tamping from the outer sleeve and use the tamp to
advance the implant beyond the tip of the with a mallet. Remove the delivery device impact the implant. The final placement of
outer sleeve. from the surgical site. the implant should be flush with the
adjacent surface.