under a constant strain Nur Amirah binti Shaharom Supervised by: i) Dr. Hendra Hermawan ii) Assoc. Prof. Ir. Dr. Mohammed Rafiq bin Abdul Kadir Abstract Magnesium-based alloys are currently attracted much interest owing to its potential for biodegradable implants application. However, despite many of its advantages which are superior to those of current implant materials, tackling the rate of degradation of magnesium alloys is one of the grand challenges highlighted. Mg-0.8Ca is believed to have optimum qualities to serve as implant material, thus have been chosen as the testing materials in present work. Failure mechanism of tested material is proposed to be evaluated to further understand its degradation behaviour under synergistic presence of mechanical loading along with corrosive physiological environment. It is believed that under such condition the materials are highly susceptible to stress corrosion cracking (SCC) which might lead to catastrophic complications and sudden failure. This study aims to evaluate the susceptibility Mg-0.8Ca towards SCC in physiological environment, determine the mode of SCC crack propagation and recognise the causes of failure which will be depicted from the propose testing methods. The understanding of failure mechanism will give contributions in the improvement of material properties in order to be successfully applied as biodegradable implants.
Introduction Metallic biomaterials are predominately applied for impaired bone tissue fixture and replacement [1] due to comparable monotonic strength they exhibit for such purposes. However, evaluation on current metallic materials used [1-5], i.e. stainless steel, titanium and cobalt-chromium-based alloys, revealed a deleterious complications due to debris produced from loading-related scenarios, in combination with corrosive environment of the implantation site. Stiffness ratio of current metallic implants toward the adjacent bones is relatively huge leading to various clinical issues due to stress shielding effects [5, 6]. Table 1 below shows typical properties of current metallic materials in comparison to those of human bone.
Table 1 Mechanical properties of commonly used metallic biomaterials and human bone [4]
Development of orthopedic implants had recently shifted from classic standard of material selection to a rather socially and economically beneficial. The selection is now demanded the materials to be biologically active in enhancing physiological response and implant-tissue interactions while progressively degrade during the healing period of damaged tissue [5]. Designated to avoid secondary surgical procedure for implant removal, biodegradable magnesium-based implants which generate mainly soluble, non-toxic degradation products have been generally approved to serve as bone fixation devices [2, 5, 7]. With sufficient mechanical strength and load-bearing capacity, magnesium-based implants exceeded cotemporary polymer based implants for orthopaedic applications. Furthermore, Youngs modulus of magnesium-based implants is a lot closer to human bone compared to permanent metallic implants currently used [4] which may contribute to the reduction of stress shielding effects [5]. However, the rate of degradation of magnesium-based alloys is in biggest concern [1, 2, 4, 5, 8, 9] as the degradation process evolves intolerant hydrogen gas accumulation and the implant might lose its mechanical integrity before the fractured bones completely heal [4, 10]. Generally, poor corrosion resistance in magnesium alloys is believed to be influenced by a few factors which include [11] (a) corrosion product films are too weak to inhibit the high intrinsic dissolution tendency of magnesium, and (b) speedup of micro-galvanic localized corrosion due to the presence of second phases serving as local cathodes. It is believed that understanding degradation mechanism of magnesium alloys is the best perspective in establishing a material with optimised properties most suited the implant applications. Alloying is one of the methods currently developed to enhance magnesium mechanical properties and corrosion resistance [2, 4, 5, 7, 11-16]. Addition of alloying elements has been observed analytically producing magnesium alloys with a lower corrosion rate than that of high purity magnesium in 3% NaCl [11]. Among various alloying elements available, calcium has been favourably chosen and considered as a unique alloying addition owing to Mg 2 Ca phase features exhibited in low alloyed Mg-Ca systems which further affect its microstructure, mechanical properties, electrochemical behaviour and degradation kinetics [5, 7, 14, 15]. Addition of calcium in Mg-Ca alloys reduces grain size and improves general and pitting corrosion resistance as well [5, 14]. Li et al. [14] highlighted a few considerations in choosing calcium as alloying element in magnesium for biodegradable material which include the composition of calcium in human body, its close density to those of human bones, and the need of magnesium in calcium incorporation into the bone have been reported [17] with potential benefit for bone healing. Corrosion in Mg-Ca alloys with calcium content up to 0.8 wt% is well distributed and homogenous in texture while those with higher calcium content demonstrate uncertain, dissipated corrosion. Magnesium with 0.6 0.8 wt% calcium has been proven exhibited slowest degradation rate [9, 15] and good biocompatibility [7].
Figure 1 Hydrogen evolution of Mg-Y in 0.1M NaCl. The rate of degradation can be represented by instantaneous value from hydrogen evolution curve or by average value over a particular time period [11]. Most assessments done in vitro considered only on general practice of magnesium alloys for orthopaedic implants, instead of simulating the implantation site condition specifically. Improvement in corrosion resistance and mechanical properties focused on the effect of alloying element and mechanical processes only, and despite of failure reported experienced by current metallic biomaterial devices, no further assessment on the mechanism of failure in corrosive environment for biodegradable implants has been made despite of the fact that SCC in magnesium alloys would give severe effects to the materials when in service. Thus, the present work utilizes every aspects of Mg-0.8Ca which have been evaluated as the absolute choice of material to be applied specifically for craniofacial fixation devices. Since the magnesium alloys corrode very fast in chloride environment, the occurrence of SCC needs to be ascertained whether the loss of mechanical properties is resulted from the SCC or just the reduction in specimen cross-section due to anodic dissolution [18]. Understanding fracture mechanism endures by Mg-0.8Ca under constant strain in simulated physiological media is expected to contribute further improvement on its corrosion resistance and service life.
Literature Reviews Current developments of metallic materials encompass groups of materials with fine interaction towards biological environment and capable to progressively degrade during the healing process of damaged tissue [5]. Instead of relying upon polymers to be applied as implant materials, biodegradable metallic substitutes are worthier to be established on account of better load-bearing capacity and mechanical strength they demonstrate [5]. Magnesium and magnesium-based alloys are currently drawing much interest as promising candidates to be applicable for such purposes. Current research development on magnesium-based biodegradable implants is focusing mostly on Mg-Al and Mg-RE alloy systems [14]. In vitro assessment [19] of degradation rate and mechanism for magnesium has been summarized by Kirkland et al. detailing their advantages and limitations. Witte et al. [20] have conducted an investigation to compare the corrosion rates of AZ91D and LAE442 and concluded that corrosion rates measured in vivo were orders of magnitude lower than those measured in substitute ocean water prepared according to ASTM-D1141-98 [21]. In vitro test has been accomplished using various solutions which include; (i) simulated body fluid (SBF), (ii) artificial plasma (AP), (iii) phosphate buffer saline (PBS), and (iv) minimum essential medium (MEM, Invitrogen). In vitro test [16] conducted to evaluate the corrosion rate of various magnesium alloys reported that high purity magnesium had the lowest corrosion rate which increased with increasing amount and effectiveness of second phase. Song [16] concluded that Zn-containing magnesium alloy is the most promising material to be developed into a biodegradable material with some modifications such as anodised coating to delay the degradation and purification to reduce the rate. A study [22] conducted to understand failure mechanisms of commercial Kntscher nails of type 316L stainless steel under the combined action of corrosion and stress has been done by G. Bombara and M. Cavallini. Evaluation on damage morphology of the device showed transgranular, branched and unblunted cracks which were first identified as SCC behaviour rather than corrosion fatigue (CF). This was corroborated with appearances of minor and indeterminate fretting fractographs using scanning electron microscope (SEM). The results obtained agreed with investigation done by M. Sivakumar and S. Rajeswari [23] which concluded SCC of failed intramedullary nail has been worsened by high inclusion content and large grain size of the implant. Fractography of failed commercial AISI stainless-steel 316L presented by S. Bandyopadhyay and P. Brockhurst [24] has shown evidence of SCC failure with crack propagation both in intergranular and transgranular modes which determined the susceptibility of austenitic stainless steel at a rather low temperature than mostly reported. Kannan and Raman [25] had claimed that SCC susceptibility of Mg-Al-Zn alloy is not substantial and should not be a concern for implant applications despite of stress- assisted failures that have been reported [2, 12, 26, 27] for current orthopaedic implants. However, a critical review done by Atrens et al. [11] highlighted a number of grand challenges in development of magnesium-based implants, which include consideration of SCC due to corrosion-assisted stresses that is known existed in implant-bone systems. This review is in agreement with investigation by Choudhary and Raman [18] which had confirmed the SCC susceptibility of both smooth and notched specimens of a magnesium alloy in physiological environment. In a different investigation that has been done by Choudhary et al. [28] on AZ91D in modified SBF and air at different strain rates revealed that the alloy is susceptible to SCC.
Materials and Methods Sample preparation Samples will be used are cast Mg-0.8Ca samples whose geometry and dimensions presented in Figure 2. The sample will be ground successively to 1200 grit using SiC paper, ultrasonically cleaned in acetone, and dried to room temperature for 24 hours prior to testing.
Figure 1 Dimension of the proposed sample. Microstructural characterisation The samples which will be used for microstructure characterisation will be mechanically ground with SiC paper up to 2500 grit and polished, follow by chemical etching in 2% Nital solution. Surface morphology of the samples before and after the introduction of strain will be observed using optical microscope. Furthermore, microstructure of samples surface will be examined using SEM before and after the samples failure. Failed samples will be first cleaned chemically using 20% CrO 3 and 10% AgNO 3
solution. The changes of microstructure morphology will be observed to further understand the influence of strain towards failure of samples and is believed to be closely related to the fracture mechanism. The characterisation of failed samples is expected to have features of brittle and/or ductile fracture, and show the mode of cracking either transgranular or intergranular. Immersion test The samples will be assembled to a custom-made apparatus for immersion test in simulated body fluid (SBF) with pH value will be adjusted to 7.4 and strain variation of 0.25% and 0.5% elongation. The test will be done for 1, 2, 4, and 6 hours to evaluate the rate of degradation under such condition from the weight loss percentage, and then prolonged until failure. Each test will be performed in triplets in order to examine the reproducibility. Another set of immersion test will be conducted as control experiment. The samples for the control experiment will not be put under strain, but evaluate the same measurement of weight loss for the same time intervals, which will then be prolonged until failure. All samples will be taken out, left to dry at room temperature for 24 hours and be measured of the weight loss, W a before chemical cleaning. Then, the samples will be cleaned with chromic acid solution to remove corrosion products and let dry to room temperature before the final weight loss, W b will be recorded. W b from failed samples from immersion test with introduction of strain will be compared to W b of the samples from immersion test without strain at the same failure time. The microstructure of both samples will be examined and is expected to verify the effects of strain on the samples microstructure and their influences on degradation rate. In situ observation A sample from the immersion test with strain value of 0.5% will be randomly picked for in situ observation of failure using stereo microscope. Fractography and failure mechanism are expected to be better visualised using this method.
Expected results and discussion The degradation rate will be computed from experiments conducted in different conditions; (i) sample will be put under strain in SBF, (ii) sample will be put under strain in air, and (iii) sample will be immersed in SBF for 24 hours then be put under strain in air. Each of the experiments however, will be repeated thrice. The variation in testing condition is to recognise the causes of failure [18]; either it is indeed a result of the synergistic effect of stress and corrosive environment (i.e. SCC) or simply attributed to the continuously reducing cross-sectional area of the sample due to high corrosion rate of magnesium alloys. Table 2 Design of experiment that is expected to be achieved with some data obtained from current conducted experiments. Media Displacement Time to failure Weight Loss pH Degradation Rate SBF EXPERIMENT 1 - Degradation Rate of MgCa Under Tension Stress 0.50% Time (min) W W W - W W b W - W b pH pH i (W - W b )/t 60 1.87 1.869 0.001 1.838 0.032 7.2 7.9 0.000533333 120 1.89 1.886 0.004 1.852 0.038 7.2 7.9 0.000316667 240 1.929 1.927 0.002 1.835 0.094 7.2 8.2 0.000391667 360 1.803 1.804 -0.001 1.656 0.147 7.4 8.4 0.000408333 t f - 960 1.94 1.86 0.08 1.654 0.286 7.3 9.2 0.000297917 EXPERIMENT 2 - Degradation Rate of MgCa Under Tension Stress 0.25% Time (s) W W W - W W b W - W b pH pH i (W - Wb)/t 60 1.75 1.78 -0.03 1.655 0.095 7.1 7.6 0.001583333 120 1.78 1.82 -0.04 1.727 0.053 7.1 7.6 0.000441667 240 1.88 1.9 -0.02 1.788 0.092 7.1 7.8 0.000383333 360 1.75 1.73 0.02 1.64 0.11 7.1 7.9 0.000305556 T2 W20
EXPERIMENT 3 - Degradation Rate of MgCa Under Tension Stress 0%
W W W - W W b W - W b pH pH i
1h W21
2h W22
4h W23
6h W24
T5 W25
Results obtained in Table 3 are from the first round of experiments. From the data, some graphical evaluation could be made as shows in Table 3. Computing stress vs. time curves will show changes in mechanical properties of the sample tested for each assessed conditions. It is expected that maximum reduction in mechanical properties which might fail the fastest among the samples tested is the one under strain in SBF. It is expected that stress-time curve for the samples that will be pulled in air indicate some similar behaviours in comparison to those pre-immersed samples. Table 3 Degradation rate evaluated for current conducted experiments.
Fractograph of the sample that will be tested under strain in SBF is expected to show common mode for SCC crack propagation of magnesium alloys, i.e. localised attack at the sample circumference which in higher magnification, will reveal transgranular cracking and localized cracks (Figure 3). 60 120 240 360 tf - 960 0.50% 0.00053333 0.00031667 0.00039167 0.00040833 0.00029792 0.25% 0.00158333 0.00044167 0.00038333 0.00030556 0 0.0002 0.0004 0.0006 0.0008 0.001 0.0012 0.0014 0.0016 0.0018 W e i g h t
L o s s
( g )
Rate of Degradation (gm-1) Degradation rates of Mg-0.8Ca under various strain values in SBF
Figure 2 Factograph of failed sample with common mode of SCC cracks propagation for magnesium alloys: (a) overall fracture surface, and (b) evidence of trransgranular cracking [18]. The experiment will use weight loss method in evaluation of degradation rate of Mg- 0.8Ca sample. At least two from three types of measurement, i.e. weight loss rate, hydrogen evolution rate and rate of Mg 2+ leaving the metal surface, are recommended by Atrens et al. [11] to be implemented to better clarify the rate of corrosion on magnesium alloy in service. However, since this study is not focus on the rate of corrosion, only weight loss rate will be measured to give a framework of service and learn the impact of such simulated condition on the specimen towards fracture. It is believed that the rate of corrosion is closely related to the failure mechanism of implants. There are two types of apparatus for transgranular SCC evaluation, the Linearly Increasing Stress Test (LIST) and the Constant Extension Rate Test (CERT). Review done by Atrens et al. [11] described that both of these methods are useful to identify the occurrences of SCC and have capabilities to measure the threshold stress and crack velocity from the final crack size divided by the time of cracking when coupled in identification of crack initiation. The increased crack propagation time under CERT conditions may be important in determining the mechanism for SCC in magnesium alloys. However, the experimental setup presented in this study will be custom-made, an apparatus that simulate the condition which initiate conventional SCC. There may be a room for improvement by applying either LIST or CERT, or both to better evaluate failure mechanism of proposed material. Choudhary and Raman [18] had done SSRT under different testing conditions which included continuous cathodic charging and simultaneous strain in SBF for indication of possible combined effect of hydrogen-induced cracking and anodic dissolution. Furthermore, circumferential notch tensile (CNT) testing was used in their work to determine the threshold stress intensity factor for SCC (K ISCC ) of the tested samples in SBF. SEM fractographs (Figure 4) of the failed samples from CNT test had provided better description on the failure mechanism.
Figure 3 SEM fractographs obtained from CNT: (a) the overall fracture surface showing machined notch, fatigue pre- crack, SCC and mechanical failure zones, (b) fatigue pre-cracked and SCC zones (arrow indicate crack propagation direction), and (c) mechanical failure zone showing dimples [18]. Intergranular in magnesium alloys occurs typically by micro-galvanic corrosion, when there is a presence of continuous or nearly continuous second phase along the grain boundaries, a microstructure typical of many cast creep-resistant magnesium alloys. In contrast, transgranular SCC occurs through the -phase magnesium matrix. From the review done by Atrens et al. [11], they concluded that SCC is associated with environmental conditions leading to the local breakdown of a partially protective surface film which is usually caused by localized corrosion by chloride ions. However, load may be influenced the breakdown as well since the occurrence of SCC is observed on pure magnesium and magnesium alloys (AZ91, AM60, AS41, ZK60A-T5) in distilled water. The influenced of hydrogen to the fracture behaviour and the effect of alloying elements on SCC propagation are not well understood currently. In advance, the understanding of microstructure influences on SCC is required to better understand magnesium transgranular SCC. Furthermore, the understanding of effective hydrogen diffusion rate to the fracture process zone needs to be included which requires an understanding of hydrogen-trap interactions in magnesium alloys. Although Kannan and Raman [13] found insignificant decreases of strength and ductility for AZ91 in simulated body fluid using CERT which indicate SCC, such failure is still largely unexplored.
Conclusion Current studies on the possibilities to implement magnesium-based alloys as degradable implants demanding a much deeper understanding of the degradation behaviour, particularly involving microstructural analysis, to tackle rapid corrosion of the material in physiological environment. SCC is one of many failures recognised among metallic materials, thus it is demanded to have absolute evaluation on such issues to provide crucial information on whether the materials will be tolerable for the given service time of the device in practical. Such studies have been done on industrial magnesium alloys, which are less likely applicable as implant materials. Thus, the present work emphasised the SCC susceptibility assessment and the mechanism of SCC involved, if applicable, leading to failure of the proposed implants material, Mg- 0.8Ca.
References 1. Niinomi, M., Recent metallic materials for biomedical applications. Metallurgical and Materials Transactions A, 2002. 33(3): p. 477-486. 2. Antunes, R.A. and M.C.L. de Oliveira, Corrosion fatigue of biomedical metallic alloys: Mechanisms and mitigation. Acta Biomaterialia, 2012. 8(3): p. 937-962. 3. Blackwood, D.J., 2.27 - Corrosion in Body Fluids, in Shreir's Corrosion, J.A.R. Editor-in- Chief: Tony, Editor. 2010, Elsevier: Oxford. p. 1308-1322. 4. Staiger, M.P., et al., Magnesium and its alloys as orthopedic biomaterials: A review. Biomaterials, 2006. 27(9): p. 1728-1734. 5. Salahshoor, M. and Y. Guo, Biodegradable Orthopedic Magnesium-Calcium (MgCa) Alloys, Processing, and Corrosion Performance. Materials, 2012. 5(1): p. 135-155. 6. Nagels, J., M. Stokdijk, and P.M. Rozing, Stress shielding and bone resorption in shoulder arthroplasty. Journal of Shoulder and Elbow Surgery, 2003. 12(1): p. 35-39. 7. Erdmann, N., et al., Evaluation of the soft tissue biocompatibility of MgCa0.8 and surgical steel 316L in vivo: a comparative study in rabbits. Biomed Eng Online, 2010. 9: p. 63. 8. Singh Raman, R.K., N. Birbilis, and J. Efthimiadis, Corrosion of Mg alloy AZ91 – the role of microstructure. Corrosion Engineering, Science and Technology, 2004. 39(4): p. 346-350. 9. Drynda, A., et al., Development and biocompatibility of a novel corrodible fluoride-coated magnesium-calcium alloy with improved degradation kinetics and adequate mechanical properties for cardiovascular applications. Journal of Biomedical Materials Research Part A, 2010. 93A(2): p. 763-775. 10. Witte, F., et al., In vivo corrosion of four magnesium alloys and the associated bone response. Biomaterials, 2005. 26(17): p. 3557-3563. 11. Atrens, A., M. Liu, and N.I. Zainal Abidin, Corrosion mechanism applicable to biodegradable magnesium implants. Materials Science and Engineering: B, 2011. 176(20): p. 1609-1636. 12. Gu, X.N., et al., Corrosion fatigue behaviors of two biomedical Mg alloys AZ91D and WE43 In simulated body fluid. Acta Biomaterialia, 2010. 6(12): p. 4605-4613. 13. Kannan, M.B. and R.K.S. Raman, In vitro degradation and mechanical integrity of calcium-containing magnesium alloys in modified-simulated body fluid. Biomaterials, 2008. 29(15): p. 2306-2314. 14. Li, Z., et al., The development of binary MgCa alloys for use as biodegradable materials within bone. Biomaterials, 2008. 29(10): p. 1329-1344. 15. Rad, H.R.B., et al., Microstructure analysis and corrosion behavior of biodegradable Mg Ca implant alloys. Materials & Design, 2012. 33(0): p. 88-97. 16. Song, G., Control of biodegradation of biocompatable magnesium alloys. Corrosion Science, 2007. 49(4): p. 1696-1701. 17. Serre, C.M., et al., Influence of magnesium substitution on a collagenapatite biomaterial on the production of a calcifying matrix by human osteoblasts. Journal of Biomedical Materials Research, 1998. 42(4): p. 626-633. 18. Choudhary, L. and R.K. Singh Raman, Magnesium alloys as body implants: Fracture mechanism under dynamic and static loadings in a physiological environment. Acta Biomaterialia, 2012. 8(2): p. 916-923. 19. Kirkland, N.T., N. Birbilis, and M.P. Staiger, Assessing the corrosion of biodegradable magnesium implants: A critical review of current methodologies and their limitations. Acta Biomaterialia, 2012. 8(3): p. 925-936. 20. Witte, F., et al., In vitro and in vivo corrosion measurements of magnesium alloys. Biomaterials, 2006. 27(7): p. 1013-1018. 21. ASTM, Standard Practice for the Preparation of Substitute Ocean Water. 2008. 22. Bombara, G. and M. Cavallini, Stress corrosion cracking of bone implants. Corrosion Science, 1977. 17(2): p. 77-85. 23. Sivakumar, M. and S. Rajeswari, Investigation of failures in stainless steel orthopaedic implant devices: pit-induced stress corrosion cracking. Journal of Materials Science Letters, 1992. 11(15): p. 1039-1042. 24. Bandyopadhyay, S. and P. Brockhurst, Fractographic evidence of mixed mode stress corrosion cracking in stainless steel surgical implants. Journal of Materials Science, 1979. 14(12): p. 3002-3003. 25. Bobby Kannan, M. and R.K. Singh Raman, Evaluating the stress corrosion cracking susceptibility of MgAlZn alloy in modified-simulated body fluid for orthopaedic implant application. Scripta Materialia, 2008. 59(2): p. 175-178. 26. Teoh, S.H., Fatigue of biomaterials: a review. International Journal of Fatigue, 2000. 22(10): p. 825-837. 27. Fleck, C. and D. Eifler, Corrosion, fatigue and corrosion fatigue behaviour of metal implant materials, especially titanium alloys. International Journal of Fatigue, 2010. 32(6): p. 929-935. 28. Choudhary, L., et al., Investigations into stress corrosion cracking behaviour of AZ91D magnesium alloy in physiological environment. Procedia Engineering, 2011. 10(0): p. 518-523.