ZEMA-1004B

Document No.

SHIMADZU UV-VIS SPECTROPHOTOMETER


UV-1700 / UV-1800 / 1650PC / 2450 / 2550 / 3600series
Installation Qualification

Installation Qualification Report


System Name

System ID No.

Installation Site
The undersigned performer reports that the Installation Qualification Protocol has been successfully completed for the
system stated above.

· Performer Signature Date

Print / /

Title

Company

The undersigned reviewer and manager report that the performer has completed the Installation Qualification Protocol
successfully.

· Reviewer Signature Date

Print / /

Title

Company

· Manager Signature Date

Print / /

Title

Company
Installation Qualification Table of Contents

Table of Contents

1. Definitions................................................................................................................................ 3
1-1 Documentation ............................................................................................................................... 3
1-2 Scope .............................................................................................................................................. 4
1-3 Entries and Revisions ..................................................................................................................... 5
2. Preliminary Checks ................................................................................................................ 6
2-1 Reviewer Confirmation .................................................................................................................. 6
2-2 Performer Identification & Approval ............................................................................................. 7
3. Pre-installation Qualification................................................................................................. 9
3-1 Equipment List ............................................................................................................................... 9
3-2 Instruction Manuals ...................................................................................................................... 11
3-3 Installation Environment .............................................................................................................. 14
3-4 Equipment Information ................................................................................................................ 15
4. Installation Qualification ..................................................................................................... 21

5. Change List ........................................................................................................................... 31

6. Attached Documents ............................................................................................................. 34

Rev 3.10 Page 2 of 34


Installation Qualification Definitions

1. Definitions

1-1 Documentation
1. The purpose of this document is to collect, confirm and record the results of the Installation Qualification of the
Shimadzu UV-VIS spectrophotometers.
This document checks the installation site/environment, confirms equipment specifications, checks
documentation such as instruction manuals, and verifies the condition of the installed equipment.

2. The following are established as the responsible parties in the execution of inspections.
· Performer
The Performer is the individual who actually performs the procedures stipulated in the Installation
Qualification Protocol and the Operational Qualification Protocol.
The Performer shall either be an individual who, having been trained by the Quality Assurance Department of
the Analytical & Measuring Instruments Division of Shimadzu Corporation, has been certified as an IQ/OQ
inspector by the relevant manager (General Manager of the Quality Assurance Department), or an individual
recognized by the Manager to be equipped with the ability to conduct this inspection.
· Reviewer
The Reviewer is an individual designated by the Manager charged with responsibility for the system.
The Reviewer verifies the contents of the Installation Qualification Protocol and Operational Qualification
Protocol documentation. In addition, the Reviewer shall attend the inspection procedures together with the
Performer, and evaluate the procedures.
The Manager can serve concurrently as the Reviewer.
· Manager
The Manager is the individual responsible for administering the entire UV-VIS system being inspected.
As well as verifying the contents of the Installation Qualification Protocol and Operational Qualification
Protocol documentation, the Manager shall also grant final approval to the inspection procedures.

3. The Performer shall check each unit and execute inspections using the protocol described in Installation
Qualification Protocol (Document No. ). After checking all items, the Performer shall sign and
date the cover sheet of this document and submit this document as a report.
(IQ must be performed by a trained and knowledgeable individual in accordance with 21 CFR Part 211 Section
211.25 (cGMP, USA promulgated 29 September 1978)).

4. The Reviewer shall first check the content of this entire document. After the performance of each inspection
(Pre-installation Qualification, Installation Qualification and Equipment Management), the Reviewer shall
check their results. After checking all items, the Reviewer shall sign and date the cover sheet of this document
as approval of the Installation Qualification.

5. The Manager shall first check the content of this entire document. After the performance of all installation
qualification protocols, the Manager shall check their results. The Manager shall enter a unique document
control No. in the Document No. space at the top right of the cover sheet, and then sign and date the cover sheet
of this document as final approval of the Installation Qualification.

6. This document is comprised of 1. Definitions, 2. Preliminary Checks, 3. Pre-installation Qualification, 4.


Installation Qualification, 5. Change List, and 6. Attached Documents. All of these sections must be included.

7. This document contains important information and any form of duplication, photography, or reproduction
without the permission of the General Manager of the Quality Assurance Department, Analytical & Measuring
Instruments Division, Shimadzu Corporation, is prohibited. However, the records of the inspections conducted
using this document may be duplicated at the discretion of the Manager.

Rev 3.10 Page 3 of 34


Installation Qualification Definitions

1-2 Scope

This Installation Qualification shall apply to the equipment installed at the following site.

(Address):

(Company):

(Department):

(Installation Site):

(System ID No.):

(Product Model Name)

(Scope of Installation Qualification)


The Installation Qualification Protocol shall include the following.

· Pre-installation Qualification
Confirmation of the specifications of the analytical equipment.
Confirmation and maintenance of documents (instruction manuals, Qualification Protocol and certificates).
Confirmation of installation site and conditions.
Confirmation of the delivered equipment.

· Installation Qualification
Connection, confirmation, and installation qualification of each unit shall be executed by verifying that each
unit is connected correctly.

· Confirmation of Software and Firmware


The scope of the Installation Qualification extends to the verification of software and firmware. Verify that the
versions stated in the certificates etc. provided with the equipment is consistent with the actual versions
displayed when the power is turned on. Verifications with the power turned on shall be executed in accordance
with the OQ (Operational Qualification) Protocol.
During installation, the software and firmware may be upgraded to versions regarded appropriate by Shimadzu
Corporation. In this case, enter the new version in the version information entry column in “4. Installation
Qualification”, and clearly state a comment to this effect in the margin.

Rev 3.10 Page 4 of 34


Installation Qualification Definitions

1-3 Entries and Revisions


1. Making entries
Use a permanent ballpoint pen, or equivalent, when recording results.
Names should be clearly printed.
Confirm that all items, except “Remarks” columns, are completed, and that there are no omissions.
Draw a diagonal line (from bottom left to top right) across all blank columns. (When making no entries, it is
not necessary to enter the “Change List No.__”, “Date” or “Signature”, or to enter the reason for the change in
the Change List in Section 4 of this document.)
Columns in which nothing has to be written in has been filled in with “N/A” in advance.
If choices are available within an item, place a checkmark in the box to select the appropriate decision.

2. Deletion and correction of mistakenly filled items


To correct mistaken entries such as inspection results, delete the entire entry by drawing a diagonal line through
it and enter the correction in the margins. To correct numerical values or symbols (checkmarks), draw 2
horizontal lines across the mistake and re-write the entry. Moreover, write the date, and the signature, and the
reason for the change. (When deletion and correction of mistakenly filled items, it is not necessary to enter the
“Change List No.__”, “Date” or “Signature”, or to enter the reason for the change in the Change List of this
document.)

3. Changing or deleting steps, procedures or acceptance criteria


If there are parts of the steps or procedures which will not be applied, draw a diagonal line through the entire
item and write “Change List No. __” above the line and assign a number to it. Date and sign it. Enter the
necessary information in the Change List of this document.
When making a change, delete the existing description as explained above, enter the change point, and
then write down the reason for the change in the Change List.

4. Revising entries
If the need to alter the contents of an entry arises, change the entry in accordance with “2. Deletion and
correction of mistakenly filled items”, and then write the reason for the change in the Change List of this
document.
When adding a new description, write it in the margins of the relevant page, and write “Change List No.
_”, the date and signature. Then, write the reason for the change in the Change List.

5. Handling Problems
Any problems identified in Installation Qualification must be investigated and appropriate actions must be
taken. All such actions must be documented and be approved by the Manager.

6. In the event of a Fail


If troubles occur during an inspection and the item fails the inspection, place a checkmark in the “Fail” box and
describe the problem in the margins of the relevant page. Write “Change List No. _” and date and sign the entry.
Enter an outline of the problem in the Change List.
After the problem is solved, describe the solution and the result (“Pass”) in the margins of the relevant page.
Write “Change List No.__” and date and sign the entry. Enter the summary of the solution in the Change List of
this document.
If necessary, pages can be added. To add a page, record the “Change List No. ” on the page and date/ sign it.
Add an alphabetical letter (starting from A) to the page number (e.g., 5-A), and insert the added page after the
relevant page. Add the page number of the added page to the Table of Contents and record it in the Change List.

7. Attaching data
Attached data are added by assigning page numbers as “A, B, C…” for each inspection item. If, for example,
there is a total of 3 sheets for attachment A, notate the pages as “32-A-1/3 - A-3/3”, ensuring that no numbers
are omitted. The Performer shall date and sign the attached data.

8. Other
Every item entered into the Change List shall be approved by the Reviewer and Manager.

Rev 3. 10 Page 5 of 34
Installation Qualification Preliminary Checks

2. Preliminary Checks

2-1 Reviewer Confirmation


Before conducting this inspection, enter the name of the Reviewer designated by the Manager.

Reviewer

Name:

Title: Date: / /

Use the spaces below if more than one Reviewer is to be set. If the spaces are not used, place a
checkmark in the “Not Applicable” checkbox.

Reviewer 2 (Not applicable)

Name:

Title: Date: / /

Reviewer 3 (Not applicable)

Name:

Title: Date: / /

Rev 3.10 Page 6 of 34


Installation Qualification Preliminary Checks

2-2 Performer Identification & Approval


Identify and approve the Performer. Choose the entry section according to the way in which the Performer is
selected.
If revisions are to be made to the document by Performers other than the persons certified in the table below,
make a note of the fact that the Performer has been changed (qualification certificate number included) in “5.
Change List”.

 If the Performer is a service engineer authorized by Shimadzu Representative, complete the following.
This is to certify that the below-signed person has received training in the Quality Assurance Department of
the Analytical & Measuring Instruments Division of Shimadzu Corporation. The person has been certified by
the General Manager of the Quality Assurance Department as a qualified IQ/OQ Inspector, as proven by the
certificate. The person performs the Installation Qualification Protocol as the Performer.
Performer
Name: Date: / /
Title:
Company name:
Certificate No.:
Signature:

Use the spaces below if more than one Performer is to be set. If the spaces are not used, place a checkmark in
the “Not Applicable” checkbox.
Performer 2 (Not applicable)
Name: Date: / /
Title:
Company name:
Certificate No.:
Signature:
Performer 3 (Not applicable)
Name: Date: / /
Title:
Company name:
Certificate No.:
Signature:

The above-signed persons are hereby recognized as Certified IQ/OQ Inspectors and approved to perform the
Installation Qualification Protocol as the Performer.
Reviewer
Signature: Date: / /
Manager
Signature: Date: / /

Rev 3. 10 Page 7 of 34
Installation Qualification Preliminary Checks
 If the Performer has been appointed by the Manager, complete the following.
Performer
Name: Date: / /
Title:
Company name:
Signature:

The above-signed person is hereby recognized as a Certified IQ/OQ Inspector and approved to perform the
Installation Qualification Protocol as the Performer.
Reviewer
Signature: Date: / /
Manager
Signature: Date: / /

Rev 3.10 Page 8 of 34


Installation Qualification Pre-installation Qualification

3. Pre-installation Qualification

3-1 Equipment List


The units and items comprising this system are to be recorded in the following equipment list. Place check
marks in the “Applicable” check boxes for the units comprising this system, and also enter the Serial No. for
each unit.
If the unit is not included in the system, place a check mark in the “Not Applicable” box.

<Equipment Configuration>

Not Serial No.


Applicable
Applicable Model Name

  UV-1700
  UV-1800
  UV-1650PC
  UV-2450
  UV-2550
  UV-3600
  UV-1601PC
  UV-2401PC
  UV-2501PC
  PC set
  Control Personal Computer
  Display
  Printer
  Syringe Sipper N
  Syringe Sipper CN
  ASC-5
  DIS-1200
  DIS-8
  CPS-240 A
  CPS-240 B
  TCC-240 A
  HCP-1C
  DPU-414
  ESC/P Compatible Screen Copy Printer
  Time Course (Multi λ) Program Pack
  DNA/Protein Quantitation Program Pack
  Protein Quantitation Program Pack

Rev 3. 10 Page 9 of 34
Installation Qualification Pre-installation Qualification

Not Model Name Serial No.


Applicable
Applicable

  UVProbe Software PKG


  UVProbe-Agent Software
  Validation Software Package N/A
  COL-UVPC (Color Measurement) Software N/A

  FLM-UVPC N/A
(Film Thickness Measurement) Software

Other Options
 
 
 
 
 
 
 

We hereby approve the system configuration specified above.

Performer (signature) Date: / /


Reviewer (signature) Date: / /

Rev 3.10 Page 10 of 34


Installation Qualification Pre-installation Qualification

3-2 Instruction Manuals

Not Rev. Storage Location


Applicable
Applicable Instruction Manual P/N
Symbol A B C

  UV-1700 Series User’s System 206-94783   


Guide
  UV-1700 Series Operation Guide 206-94785   
  UV-1800 System User’s Guide 206-97040   
  UV-1800 Operation Guide 206-97042   

  UV-1601PC Instruction Manual 206-94375 N/A   


User’s System Guide
Information for users who purchased
  the UV-1650PC 206-94647   

  UV-2401/2501 Series Instruction 206-94376   


Manual User’s System Guide

  Information for users who purchased 206-94649   


the UV-2450/2550
  UV-3600 Instruction Manual 206-95518   
  UVProbe Tutorial 206-94459   
  Shimadzu User Authentication Tool   
Easy Operation Procedure for
  206-94653   
Photometric Module
  UV SOP 206-81758   
  PC Set Instruction Manual   
  PC and peripherals N/A N/A   
  Syringe Sipper N/CN 206-94651   
  ASC-5 206-96026   
  CPS-240A/B 206-96225   
  TCC-240A 206-96227   
  DIS-1200 206-93060   
  DIS-8 206-94543   
  HCP-1 206-96368   
  DPU-414 206-94803   
  MPU 206-97124   
  Time Course (Multi λ) Program Pack 206-96956   
DNA/Protein Quantitation Program
  206-93104   
Pack
  Protein Quantitation Program Pack 206-93102   
  CLASS Agent Installation Procedures   

Rev 3. 10 Page 11 of 34
Installation Qualification Pre-installation Qualification

Not Rev. Storage Location


Applicable
Applicable Instruction Manual P/N
Symbol A B C
  UVProbe Agent Software   
  CLASS Agent Manager   
  Program List   
  UVPC Validation Software 206-95012   
  Color Measurement Software
206-94284   
Film Thickness Measurement
  206-94321   
Software
    
    
    
    
    
    
    

Storage Site

A

B

C

The following signees have verified the storage locations of the above instruction manuals.

Performer (signature) Date: / /


Reviewer (signature) Date: / /

Rev 3.10 Page 12 of 34


Installation Qualification Pre-installation Qualification

3-2-1 Certificate of Compliance and Inspection Result Sheet


Shimadzu Corporation provides the following certificates for each unit.

· Certificate of Compliance
· Inspection Report (Spectrophometer unit only)

Place a checkmark in the “Applicable” box if the unit is included in the system to be inspected and verify that the
“Certificate of Compliance” and “Inspection Report” have been supplied. If the unit is not included in the system,
place a checkmark in the “Not Applicable” box.

Not Certificate of
Applicable Applicable Model Name Compliance Inspection Report

  UV-1700  
  UV-1800  
  UV-1650PC  
  UV-2450  
  UV-2550  
  UV-3600 N/A N/A

  UV-1601PC  N/A

  UV-2401PC  N/A

  UV-2501PC  N/A

  UVProbe (Ver. )  N/A

  CLASS-Agent (Ver. )  N/A

  Validation Software (Ver. )  N/A

  COL-UVPC (Ver. )  N/A

  FLM-UVPC (Ver. )  N/A

   
   
   
   
   
   

Store the Certificate of Compliance and the Inspection Report together with this document.

Performer (signature) Date: / /


Reviewer (signature) Date: / /

Rev 3. 10 Page 13 of 34
Installation Qualification Pre-installation Qualification

3-3 Installation Environment


The following environment must be provided for the equipment to function properly.

3-3-1 Installation Environment


(Temperature) Temperature during operation must be within the 15 to 35ºC range with small daily temperature
fluctuations.

(Humidity) Humidity during operation must be within 20 to 80%.


(Above 30ºC, however, humidity must be 70% or less.)

(Bench)
· Space Requirements: Listed in section 3-5 for each component
· Strength Requirements: Listed in section 3-5 for each component

(Other) Avoid exposure to air from air-conditioners.


Avoid exposure to direct sunlight.
Avoid exposure to vibrations.
Clean location, free of dust and corrosive gases.
Avoid exposure to instruments that can generate strong magnetic fields.

3-3-2 Power Supply


(Voltage) 100V - 120V±10%, 220V - 240V±10%

(Frequency) 50/60 Hz

(Connection) The instrument must be grounded.


If the outlet used is the two-prong type, use a power adapter with a grounding lead wire
connected to the grounding terminal. If the outlet used is not equipped with a grounding
terminal, the equipment must be connected to an earth-grounded object. If the outlet used is
the three-prong type, the connection above is not required.
(However, this excludes items that do not require to be grounded, even if the outlet is the
two-prong type.)

Rev 3.10 Page 14 of 34


Installation Qualification Pre-installation Qualification

3-4 Equipment Information

Accessories
Confirm Model Name (Product Name) Component ID
Pass Fail
 UV-1700  No  
 UV-1800  No  
 UV-1601PC/1650PC  No  
 UV-2401PC/2450PC  No  
 UV-2501PC/2550  No  
 UV-3600  No  
 UV Probe  No  
 Syringe Sipper N/CN  No  
 Auto Sample Changer ASC-5  No  
 DIS-1200  No  
 DIS-8  No  
 CPS-240A/B  No  
 TCC-240A  No  
 HCP-1C  No  
 DPU-414  No  
 MPU  No  
 ESC/P Compatible  No  
 PCL Compatible  No  
 Time Course (Multi λ) Program Pack  No  
 DNA/Protein Quantitation Program Pack  No  
 Protein Quantitation Program Pack  No  
  No  
  No  
  No  
  No  
  No  

We hereby approve the equipment information above.

Performer (signature) Date: / /


Reviewer (signature) Date: / /

Rev 3. 10 Page 15 of 34
Installation Qualification Pre-installation Qualification

Computer and Peripherals


 Applicable  Not Applicable
Equipment Information

Product Name: Personal Computer

Manufacturer:

Model Name:

CPU Model Name( ) Clock speed ( Hz)

RAM ( ) MB

HDD ( ) Bytes

Component ID:  No

Power Source: AC V
Frequency: Hz

Power Consumption:

Product Name: Display  Yes  No

Manufacturer:

Model Name:

Component ID:  No

Power Source: AC V
Frequency: Hz

Power Consumption:

Product Name: Printer  Yes  No

Manufacturer:

Model Name:

Component ID:  No

Power Source: AC V
Frequency: Hz

Power Consumption:

Rev 3.10 Page 16 of 34


Installation Qualification Pre-installation Qualification

Equipment Information

Product Name:  Yes  No

Manufacturer:

Model Name:

Component ID:  No

Power Source: AC V
Frequency: Hz

Power Consumption:

Product Name:  Yes  No

Manufacturer:

Model Name:

Component ID:  No

Power Source: AC V
Frequency: Hz

Power Consumption:

Product Name:  Yes  No

Manufacturer:

Model Name:

Component ID:  No

Power Source: AC V
Frequency: Hz

Power Consumption:

We hereby approve the equipment information above.

Performer (signature) Date: / /


Reviewer (signature) Date: / /

Rev 3. 10 Page 17 of 34
Installation Qualification Pre-installation Qualification

UVProbe Spectrophotometer Control Software


 Applicable  Not Applicable
Equipment Information

Product Name: UVProbe


Manufacturer: Shimadzu Corporation
( Not required when provided
Software Product No. (P/N): with the instrument)

UVProbe (Ver. )

Update Disk
 Yes (Ver. )
P/N
 No

Component ID:  No

CLASS-Agent Software
 Applicable  Not Applicable
Equipment Information

Product Name: CLASS-Agent


Manufacturer: Shimadzu Corporation
Software Product No. (P/N):
CLASS-Agent (Ver. )
Install CD-ROM P/N
Update Disk
 Yes (Ver. )
P/N
 No

Component ID:  No

We hereby approve the equipment information above.

Performer (signature) Date: / /


Reviewer (signature) Date: / /

Rev 3.10 Page 18 of 34


Installation Qualification Pre-installation Qualification

Validation Software
 Applicable  Not Applicable
Equipment Information

Product Name: Validation Software Package


Manufacturer: Shimadzu Corporation
Software Product No. (P/N):
Validation Software (Ver. )
Install CD-ROM P/N
Update Disk
 Yes (Ver. )
P/N
 No

Component ID:  No

Color Measurement Software COL-UVPC


 Applicable  Not Applicable
Equipment Information

Product Name: Color Measurement Software COL-UVPC


Manufacturer: Shimadzu Corporation
Software Product No. (P/N):
COL-UVPC (Ver. )
Install CD-ROM P/N
Update Disk
 Yes (Ver. )
P/N
 No

Component ID:  No

We hereby approve the equipment information above.

Performer (signature) Date: / /


Reviewer (signature) Date: / /

Rev 3. 10 Page 19 of 34
Installation Qualification Pre-installation Qualification

Film Thickness Measurement Software FLM-UVPC


 Applicable  Not Applicable
Equipment Information

Product Name: Film Thickness Measurement Software FLM-UVPC


Manufacturer: Shimadzu Corporation
Software Product No. (P/N):
FLM-UVPC (Ver. )
Install CD-ROM P/N
Update Disk
 Yes (Ver. )
P/N
 No

Component ID:  No

We hereby approve the equipment information above.

Performer (signature) Date: / /


Reviewer (signature) Date: / /

Rev 3.10 Page 20 of 34


Installation Qualification Installation Qualification

4. Installation Qualification
This section verifies that the equipment has been installed correctly.
Place a checkmark in the “Applicable” box for the units included in the system to be inspected.

Item
Not
Applicable Applicable Model Name Appearance Installation
Pass Fail Pass Fail Ver.
  UV-1700    
  UV-1800    
  UV-1601PC/1650PC    
  UV-2401PC/2450PC    
  UV-2501PC/2550    
  UV-3600    
  UV Probe    
  Syringe Sipper N/CN     N/A
  Auto Sample Changer ASC-5    
  DIS-1200     N/A
  DIS-8     N/A
  CPS-240A/B     N/A
  TCC-240A     N/A
  HCP-1C     N/A
  DPU-414     N/A
  MPU     N/A
  ESC/P Compatible    
  PCL Compatible    
  Time Course (Multi λ) Program Pack    
DNA/Protein Quantitation Program
     
Pack
  Protein Quantitation Program Pack    
     
     
     
     
     
     

Performer (signature) Date: / /


Reviewer (signature) Date: / /

Rev 3.10 Page 21 of 34


Installation Qualification Installation Qualification

Personal Computer and Peripherals


 Applicable  Not Applicable

Personal Computer Setup

· Checking Computer Appearance, Connection, and Installed Software

(Result) Item Pass Fail


1. Appearance  
2. Installation  

3-1. OS Version  Windows2000 Ver. (SP )


 Windows XP Ver. (SP )

3-2. Printer Driver Installation  

Printer Driver Version Ver.

3-3. Installed Software


 UVProbe Ver.
 Other ( ) Ver.
 Other ( ) Ver.
 Other ( ) Ver.
 Other ( ) Ver.
 Other ( ) Ver.
 Other ( ) Ver.
 Other ( ) Ver.
 Other ( ) Ver.
 Other ( ) Ver.

4. Hard Disk File System


File System Suitability:  Suitable  Unsuitable  No Requirements Specified
 C drive: (Spec. Req.: )
 drive: (Spec. Req.: )
 drive: (Spec. Req.: )
 drive: (Spec. Req.: )

Performer (signature) Date: / /


Reviewer (signature) Date: / /

Rev 3.10 Page 22 of 34


Installation Qualification Installation Qualification

· Security Settings Check

(Result) Policy Settings


Settings Suitability:  Suitable
 Unsuitable
 Not Applicable
 Requirements are not specified
(Reason Not Applicable: )

Item (Settings) (Required Spec.)


User Manager
(Account 1: )
User must change password at next login  Valid  Invalid ( Valid  Invalid)
User cannot change password  Valid  Invalid ( Valid  Invalid)
Password never expires  Valid  Invalid ( Valid  Invalid)
Account is disabled  Valid  Invalid ( Valid  Invalid)

(Account 2: )
User must change password at next login  Valid  Invalid ( Valid  Invalid)
User cannot change password  Valid  Invalid ( Valid  Invalid)
Password never expires  Valid  Invalid ( Valid  Invalid)
Account is disabled  Valid  Invalid ( Valid  Invalid)

(Account 3: )
User must change password at next login  Valid  Invalid ( Valid  Invalid)
User cannot change password  Valid  Invalid ( Valid  Invalid)
Password never expires  Valid  Invalid ( Valid  Invalid)
Account is disabled  Valid  Invalid ( Valid  Invalid)

(Account 4: )
User must change password at next login  Valid  Invalid ( Valid  Invalid)
User cannot change password  Valid  Invalid ( Valid  Invalid)
Password never expires  Valid  Invalid ( Valid  Invalid)
Account is disabled  Valid  Invalid ( Valid  Invalid)

(Account 5: )
User must change password at next login  Valid  Invalid ( Valid  Invalid)
User cannot change password  Valid  Invalid ( Valid  Invalid)
Password never expires  Valid  Invalid ( Valid  Invalid)
Account is disabled  Valid  Invalid ( Valid  Invalid)
(Continued on next page)

Performer (signature) Date: / /


Reviewer (signature) Date: / /

Rev 3. 10 Page 23 of 34
Installation Qualification Installation Qualification

(Result) Item (Settings) (Required Spec.)


Account Policy
Password length ( ) ( )
Password change prohibited period ( ) ( )
Password valid period ( ) ( )
Password log ( ) ( )
Password complexity ( ) ( )
Account lock-out threshold value ( ) ( )
Reset lock-out count ( ) ( )
Lock-out period ( ) ( )
With PIN number returned to original ( ) ( )
state, save all domain user passwords

User Authority Allocation


Change system time
(Permitted account)

Security Options
Interactive login
: Ctrl+Ald+Del not required  Valid  Invalid  Valid  Invalid
Interactive login
: Do not display last user’s name  Valid  Invalid  Valid  Invalid

Audit Policy
Account login event  Pass  Fail ( Pass  Fail
 Not audited  Not audited)
Account management  Pass  Fail ( Pass  Fail
 Not audited  Not audited)
Object access  Pass  Fail ( Pass  Fail
 Not audited  Not audited)
System event  Pass  Fail ( Pass  Fail
 Not audited  Not audited)
Directory service access  Pass  Fail ( Pass  Fail
 Not audited  Not audited)
Process tracking  Pass  Fail ( Pass  Fail
 Not audited  Not audited)
Change policy  Pass  Fail ( Pass  Fail
 Not audited  Not audited)
Login event  Pass  Fail ( Pass  Fail
 Not audited  Not audited)
Use privileges  Pass  Fail ( Pass  Fail
 Not audited  Not audited)

Performer (signature) Date: / /


Reviewer (signature) Date: / /

Rev 3.10 Page 24 of 34


Installation Qualification Installation Qualification

· Network Connections Settings Check

 Applicable  Not Applicable

(Result) 1. Check network connection settings


Settings Suitability:  Suitable  Unsuitable  Requirements not specified
(Reason why not applicable: )
Computer Name: (Required Spec.: )
Computer name duplication:  No duplication  Duplication present
I IP Address: (Required Spec.: )
IP Address duplication:  No duplication  Duplication present
Subnet Mask
 Domain Name: (Required Spec.: )
 Workgroup Name: (Required Spec.: )

2. Automatically Synchronize Time to Reference Computer


Settings Suitability:  Suitable  Unsuitable  Requirements not specified
 Time is synchronized (Ref. Computer Name: )
 Time is not synchronized
(Required Spec.  Synchronized  Not synchronized)

Computer Date and Time

(Result) Item Computer Reference Pass Fail


Computer’s date ( / / ) ( / / )  
Computer’s time ( : : ) ( : : )  
Permitted range ( )

Performer (signature) Date: / /


Reviewer (signature) Date: / /

Rev 3. 10 Page 25 of 34
Installation Qualification Installation Qualification

UVProbe Spectrophotometer Control Software


 Applicable  Not Applicable

Appearance and Installation Check

(Result) Item Pass Fail


Appearance  
Shimadzu User Authentication Tool Installation  
UVProbe Installation  

Shimadzu User Authentication Tool Version:

UVProbe Version:

Installation Mode Check

(Result) Item
UVProbe Installation Mode
Installation Mode Suitability:  Suitable  Unsuitable
 Requirements not specified

Application Mode: (Required Spec.: )

Alteration Check

(Result) Item Pass Fail


 UVProbe Ver.  
 Shimadzu User Authentication Tool  
* Mark the installed program name and whether it passed or failed. Also, attach the printed results.

Performer (signature) Date: / /


Reviewer (signature) Date: / /

Rev 3.10 Page 26 of 34


Installation Qualification Installation Qualification

Security Settings Check

(Result) Item
Shimadzu User Authentication Tool Settings Check
( Not applicable- Shimadzu User Authentication Tool not used)
Settings Suitability: (Required Spec.)
 Suitable  Unsuitable  Requirements not specified
Database connection destination name: ( )
Minimum Password Length: characters or more ( )
Password Validity Period: days ( )
Password Change Lock-out Period: days ( )
Same Password Permitted  Not checked  Checked ( )
If not checked, number of times password stored :_________times
Password meets complexity requirements  Not checked  Checked ( )
User must change password at next login  Not checked  Checked ( )
User Indefinite Lock-out  Not checked  Checked ( )
User Lock-out Period: minutes ( )
Failed User Logon Attempts: Lock-out after attempts. ( )
PC Lock-out Period: minutes ( )
Failed PC Logon Attempts: Lock-out after attempts. ( )

Auto Sample Changer ASC-5


 Applicable  Not Applicable

(Result) Item Pass Fail


Appearance  
Installation  
Checksum Value:
Version:

<Additional · The Relationship Between the Firmware Version and Checksum Value
Information> The relationship between the ASC-5 Auto Sample Changer firmware checksum value and
version number is shown below.
Checksum Value Version
02A ; Ver 1.00

Performer (signature) Date: / /


Reviewer (signature) Date: / /

Rev 3. 10 Page 27 of 34
Installation Qualification Installation Qualification

Class-Agent Software
 Applicable  Not Applicable
Appearance and Installation Check

(Result) Item Pass Fail


Outline  
Agent-Manager Installation  
UVProbe-Agent Installation  
( ) Installation  
( ) Installation  
( ) Installation  

Agent-Manager Ver.  
UVProbe-Agent Ver.  
( ) Ver.  
( ) Ver.  
( ) Ver.  

Installation Mode Check

(Result) 1. CLASS-Agent Installation Mode


Installation Suitability:  Suitable  Unsuitable  Requirements not specified

Installation Mode: (Required Spec.: )

Alteration Check

(Result) Item Pass Fail


 CLASS-Agent Manager  
 UVProbe- Agent  
  
  
  
  
* Enter the installed program name and whether it passed or failed. Also, attach the printed results.

Performer (signature) Date: / /


Reviewer (signature) Date: / /

Rev 3.10 Page 28 of 34


Installation Qualification Installation Qualification

Validation Software Package


 Applicable  Not Applicable
Appearance and Installation Check

(Result) Item Pass Fail


Appearance  
Installation  
Version No.:

Alteration Check

(Result) Item Pass Fail


UVPC Validation Software  
* Attach the printed results.

Color Measurement Software COL-UVPC


 Applicable  Not Applicable
Appearance and Installation Check

(Result) Item Pass Fail


Appearance  
Installation  
Version No.:

Alteration Check

(Result) Item Pass Fail


UVColor  
* Attach the printed results.

Performer (signature) Date: / /


Reviewer (signature) Date: / /

Rev 3. 10 Page 29 of 34
Installation Qualification Installation Qualification

Film Thickness Measurement Software FLM-UVPC


 Applicable  Not Applicable
Appearance and Installation Check

(Result) Item Pass Fail


Appearance  
Installation  
Version No.:

Alteration Check

(Result) Item Pass Fail


UVFilm  
* Attach the printed results.

Performer (signature) Date: / /


Reviewer (signature) Date: / /

Rev 3.10 Page 30 of 34


Installation Qualification Change List

5. Change List
When changing or deleting the contents of this document, or adding new information to this document, record
change list number, page number, reason, date, and name of the Performer in the Change List below, and obtain
the approval of the Reviewer and the Manager.

Change No: Enter the change number assigned in sequence from 1 onwards in accordance with “1.3
Entries and Revisions”.
Page: Enter the page number in which the change, deletion or addition was made.
Reason: Enter the reason for the change, deletion or addition.
Date: Enter the date on which the change, deletion or addition was made.

Change
Page Reason Date Performer Reviewer Manager
No.

Rev 3.10 Page 31 of 34


Installation Qualification Change List

Change
Page Reason Date Performer Reviewer Manager
No.

Rev 3.10 Page 32 of 34


Installation Qualification Change List

Change
Page Reason Date Performer Reviewer Manager
No.

Rev 3. 10 Page 33 of 34
Installation Qualification Attached Documents

6. Attached Documents

Attached Item Description Page


Certificate of Compliance
 3-5 Certificates
Inspection Report
4 Time Course (Multi λ) Program Pack
 Version check Verification data Attachment 34 -
4 DNA/Protein Program Pack
 Version check Verification data Attachment 34 -
4 Protein Quantitation Program Pack
 Version check Verification data Attachment 34 -
4 PC Set/Personal computer and peripheral devices
 Printer driver version check Verification data Attachment 34 -
 Reference computer system time automatic Attachment 34 -
Verification data
adjustment
UV-VIS Spectrophotometer Control Software
4
UVProbe
 Alteration Check Verification data Attachment 34 -
4 Class Agent software
 Alteration Check Verification data Attachment 34 -
4 Validation software
 Alteration Check Verification data Attachment 34 -
4 Color Measurement software COL-UVPC
 Alteration Check Verification data Attachment 34 -
4 Film Thickness software FLM-UVPC
 Alteration Check Verification data Attachment 34 -

* Number the attached data by adding an alphabetical letter (starting from A) to the final page number (e.g., Attachment
34-A-1/3 – A-3/3). Fill in the table above and attach the data to this document.

Performer (signature) Date: / /


Reviewer (signature) Date: / /

Rev 3.10 Page 34 of 34