You are on page 1of 5

ID*:

ID*:
ID*:
s
30 1 15 60 120 180 240
30 1 15 60 120 181 240
0 0 0 0 0 1 0
OK OK OK OK OK OK OK
Table 2. Needle inspection:
d1 d2 d3
1.11 1.11 1.11
max
min
OK OK OK
min
299.2 297.1 298.3
max
min
OK OK OK
R011/1 CALIBRATION REPORT, VICAT DEVICE
Range test:
Test pass
OK
minute
Measured diameter (mm)
OK
Needle length
(45mm)
Required (mm)
Measured length (mm) 47
45
EN196-3
APPROVED
Units:
Time (Vicat):
Time device:
Calibration tools:
Timer device:
Length device:
Weight device:
OK
1.09
1.21
Weight (needle +
holder)(g):
Needle position (central longitudinal) visual
description
Required (mm)
Range test:
Test pas
Table 1. Timer inspection:
Table 3. Needle inspection:
Needle diameter
(mm):
OK
Difference:
Range test:
Test passed
Date of calibration:
Instrument name:
ID:
VICAMATIC, Automatic setting time device
Referent norm:
Final report:
Table 4. Needle inspection:
Table 5. Needle inspection:
Yes , needle is in longitudinal center.
No , needle is bended.
Test passed: OK
YES
OK
295
305
Measured weight (g)
Required (g)
Range test:
Test pass
1
L1 L2 L3
69.0 70.0 40.0
max 65 65 39.5
min 85 85 40.5
OK OK OK
YES
NO
YES
YES
If "NOT", this device Is FORBIDDEN of be used in laboratory (detailed explanation has to be given in annex to this
document)
This device is properly calibrated and is approved to use in laboratory testing.
APPROVED
Dimension (mm)
Mold
Dimension
(mm)
Test pass
R011/1
Range test:
YES
1)
2)
Approved by:
OK
Required
(mm)
Yes , needle ZERO position is set
No , needle ZERO position is not set
YES
Device is clean
Any visible damage (if YES, give description in annex)
Does the room condition suitable for this device
Electricity connector is proper and safe
Yes , parameters are set good
No , parameters are not set good
Team for calibration:
Signature:
Test
passed:
Test
passed:
Test
passed:
Software parameters
Check in the last version of LI001
Needle ZERO position, auto
calibration
OK
Final Conclusion:
OK
OK
YES
Yes , all base plates are same thickens
No , If not, remove all which are different OK
All base plates used for setting time
test has to be same thickness
Table 6. Mold inspection:
Table 7. Software inspection:
Table 8. Software inspection:
Table 9. Base plate dimension:
Table 10 Visual inspection:
2
ID*:
ID*:
ID*:
s
30 1 15 60 120 180 240
-- -- -- -- -- -- --
-- -- -- -- -- -- --
Table 2. Needle inspection:
d1 d2 d3
max
min
-- -- --
min
max
min
-- -- --
No , needle is bended.
Test pass
45
R011/1 CALIBRATION REPORT, VICAT DEVICE
Table 5. Needle inspection:
Table 4. Needle inspection:
Needle position (central longitudinal) visual
description
305
Test passed: --
Weight (needle +
holder)(g):
Measured weight (g)
Test pass
Required (g)
295
minute
Range test: --
Yes , needle is in longitudinal center.
Required (mm)
1.21
Range test: --
Table 3. Needle inspection:
Needle length
(45mm)
Measured length (mm)
Range test: -- --
Range test: --
Needle diameter
(mm):
Measured diameter (mm)
Test pas
Required (mm)
1.09
Difference:
Table 1. Timer inspection:
Units:
Test passed
Length device:
Time device:
Time (Vicat):
Weight device:
Timer device:
ID:
Referent norm: EN196-3
Date of calibration:
Instrument name:
Calibration tools:
--
VICAMATIC, Automatic setting time device
Final report:
3
L1 L2 L3
max 65 65 39.5
min 85 85 40.5
-- -- --
Team for calibration:
1)
2)
Approved by:
Signature:
Electricity connector is proper and safe
Final Conclusion:
--------
--
This device is properly calibrated and is approved to use in laboratory testing.
If "NOT", this device Is FORBIDDEN of be used in laboratory (detailed explanation has to be given in annex to this
document)
Device is clean
Any visible damage (if YES, give description in annex)
Does the room condition suitable for this device
--
Table 10 Visual inspection:
No , parameters are not set good --
Table 9. Base plate dimension:
Test
passed:
No , needle ZERO position is not set
Yes , all base plates are same thickens
No , If not, remove all which are different
All base plates used for setting time
test has to be same thickness
--
Table 8. Software inspection:
Test
passed:
Software parameters
Check in the last version of LI001
Yes , parameters are set good
Needle ZERO position, auto
calibration
Test
passed:
Required
(mm)
Yes , needle ZERO position is set
Range test: OK
Table 7. Software inspection:
Table 6. Mold inspection: R011/1
Mold
Dimension
(mm)
Dimension (mm)
Test pass
4
Calibration description (EN 196-3)
VICAMATIC, Automatic setting time device
R011/1
Table 3
L = 45mm
Table 5
Weight
Needle and holder
together
295-305(g)
T
a
b
l
e

2
d3
d2
d1
L
3
= 40.0 2mm
Table 6
L
1
= 65-85mm
L
2
= 65-85mm
Use L key
for needle
replacement only
Use L key
for needle
ZERO calibration
Glass base plate
Table 7
Table 9
OK NOT
Table 4
5