You are on page 1of 11

3702 IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, VOL. 58, NO.

10, OCTOBER 2009

Digital Measuring System for Monitoring


Motor Shaft Parameters on Ships
Hrvoje Džapo, Member, IEEE, Zoran Stare, Member, IEEE, and Nenad Bobanac

Abstract—The robust measuring system for the continuous


monitoring of the ship motor shaft average torque, power, and
rotational speed was developed and tested. The system was de-
signed for permanent installation in ships under realistic working
conditions. The strain gage sensor is employed to measure the
torque, whereas the shaft rotational speed is measured by the
inductive proximity switch. The strain gage signal conditioning
circuit design was highly simplified by using a high-resolution
ΣΔ A/D converter (ADC). Such an approach minimizes the
overall number of analog components and the input measuring
chain errors. A contactless power transmission subsystem was
Fig. 1. Classical approach to strain gage signal conditioning circuit design.
developed based on the specifically designed rotary transformer
operating at low frequency. Strain measurement results are trans-
mitted over the near-field low-frequency capacitive link by us- Wheatstone bridge circuit [2]. Torque can also be measured by
ing a specifically designed communication protocol based on the magnetic [3]–[6], capacitive [7], [8], and optical [9] sensors.
digital modulation principle. The receiver microprocessor-based However, resistive strain gage sensors are the preferred choice
unit processes the measured data and transmits the results to for many applications due to their higher accuracy, linearity,
the monitoring human–machine interface over some commonly
used industrial analog and digital communication protocols. The repeatability, etc. The full-bridge configuration is the most com-
principal advantages of the novel measuring system design over mon choice, because it provides the highest sensor sensitivity,
some state-of-the-art commercial solutions are simplicity, higher good temperature compensation, and linearity. The power can
accuracy, stability, reliability, simplified calibration procedure, be calculated from the measured torque and rotational speed by
and contactless signal transmission and processing completely in means of
a digital form, which, hence, reduces the most common shortcom-
ings of similar analog-based systems.
P = 2 · π · M · n. (1)
Index Terms—Digital measurements, marine electronic equip-
ment, mechanical variables measurement, strain measurement, The shaft under load is subject to the biaxial stress condition
torque measurement.
with principal normal stresses at angles of ±45◦ to the lon-
gitudinal axis of the shaft [2]. For the full-bridge strain gage
I. I NTRODUCTION configuration, the analytical formula for torque calculation is
given by
M ODERN SHIP control and monitoring systems regularly
include a subsystem for measuring the most important
mechanical parameters on revolving motor shafts, namely, M=
π G · D3
· ·
um
(2)
torque, power, and rotational speed. The measurement of these 8 Kt UREF
mechanical quantities provides important information about the
where G is the shear modulus, D is the shaft diameter, Kt is the
motor behavior under realistic working conditions and serves
strain gage factor, UREF is the strain gage supply voltage, and
as an input for the closed-loop control system. The shaft torque
um is the voltage measured at the bridge diagonal.
M is determined by measuring the strain ε on the shaft surface,
In practical applications, all measuring circuitry, along with
whereas the shaft rotational speed n is measured by the induc-
the strain gage sensor, must be attached to the revolving shaft
tive proximity switch. Although it is possible to measure the
surface. This means that two principal problems must be solved,
strain by a few different techniques [1], the industry standard
namely, wireless power and data transmission between the
approach is to use resistive strain gage sensors connected in a
stator and rotor parts of a measuring system. The classical
approach [2] to strain gage signal conditioning circuit design
Manuscript received June 29, 2008; revised March 3, 2009. Current version
published September 16, 2009. The Associate Editor coordinating the review
is shown in Fig. 1. Such a configuration typically comprises an
process for this paper was Dr. Izzet Kale. input stage instrumentation amplifier (with adjustable zero and
H. Džapo and Z. Stare are with the Faculty of Electrical Engineering and gain), a low-pass antialiasing filter (usually realized as a higher-
Computing, University of Zagreb, Zagreb 10000, Croatia (e-mail: hrvoje.
dzapo@fer.hr). order analog active filter), and an A/D converter (ADC). In 12-
N. Bobanac is with BOBLab d.o.o., Zagreb 10020, Croatia (e-mail: nenad. to 16-bit ADC systems, the zero level and the amplifier gain
bobanac@zg.t-com.hr). must carefully be adjusted to cover the full input range of the
Color versions of one or more of the figures in this paper are available online
at http://ieeexplore.ieee.org. ADC for the maximum expected torque deflection to achieve
Digital Object Identifier 10.1109/TIM.2009.2019316 acceptable resolution of the measuring system. The Wheatstone
0018-9456/$26.00 © 2009 IEEE

Authorized licensed use limited to: Praveenraja SN. Downloaded on December 11, 2009 at 01:04 from IEEE Xplore. Restrictions apply.
DŽAPO et al.: DIGITAL MEASURING SYSTEM FOR MONITORING MOTOR SHAFT PARAMETERS ON SHIPS 3703

bridge in Fig. 1 is balanced by means of the potentiometer


R0 , whereas the instrumentation amplifier gain is adjusted with
the resistor Rg . The precision shunt resistor Rc is used to
unbalance the bridge to simulate the load in a process of gain
calibration [2].
Such an approach to the input measuring chain design has
a few inherent shortcomings. First, the calibration procedure is
rather complex due to the need for an iterative adjustment of
potentiometers R0 and Rg to achieve the maximum ADC input
dynamic range. Moreover, the careful calculation of torque
values expected under realistic working conditions must be
conducted in advance to correctly predict the voltage levels
at the strain gage sensor output. Furthermore, the calibration
potentiometer’s aging effects may compromise the long-term
input chain stability, which may introduce problems in per-
manent shaft monitoring installations. Finally, it should be
emphasized that the complete analog measuring chain before
ADC (adjustable instrumentation amplifier, resistors, and ana- Fig. 2. Three-dimensional view of the assembled and connected shaft moni-
log anti-aliasing low-pass filter) significantly contributes to toring system.
the overall measuring uncertainty primarily due to component measuring system under laboratory and realistic working envi-
aging effects and temperature drift. Another important problem ronment test conditions.
inherent to the described input measuring chain solution is
the possibility of ADC input saturation due to the excessive
II. D ESCRIPTION OF THE N OVEL S HAFT
long-term strain gage sensor zero creep on the shaft under
M ONITORING S YSTEM
permanent mechanical load, which cannot be compensated by
the autocalibration procedures in the embedded software alone. Nowadays, there are numerous commercial shaft monitor-
Our solution adopts a different approach to input measuring ing systems that employ the classical approach to the strain
chain realization by interfacing the strain gage sensor to the measuring system design (e.g., [22] and [23]). Our previous
integrated circuit with high-resolution ΣΔ ADC (the approach research in this area yielded some solutions [17]–[20] that
previously proposed in [10]). The goal of this paper is to prove were based on the classical measuring principle described in
that such an approach may yield a simpler, more efficient, and Fig. 1. They revealed numerous shortcomings of the classical
accurate shaft monitoring system. Our shaft monitoring system approach in practical long-term shaft monitoring applications,
design was particularly optimized for permanent installations such as complex installation and calibration procedure, tem-
with an emphasis on high measuring accuracy, simplicity, re- perature stability problems, complex and costly circuit design,
liability, and robust behavior in harsh environments. The main ADC input saturation due to long-term zero-creep effects under
advantage of this design is that no zero and gain adjustment po- permanent mechanical load, etc. An alternative approach to
tentiometers are needed on the rotor side, and that all calibration the strain gage signal conditioning circuit design based on the
procedures can be performed on the stator side by means of the sigma–delta modulator has already been proposed [10], [11],
embedded software. Furthermore, in our design, the strain gage [14], [15]. Such an approach to sensor interfacing nowadays
imbalance cannot lead the ADC inputs into saturation under any becomes popular not only with resistive strain gage sensors
circumstances. These advantages are very important in practice. but also with other types of passive sensors, e.g., capacitive
Moreover, we developed a digital contactless data transmission sensors [12]. In particular, the availability of modern precision
subsystem particularly suited for such applications, which is off-the-shelf integrated high-resolution ΣΔ ADCs now allows
less sensitive to the interferences present in noisy environments easy interfacing of the strain gage sensors without the need
than classical analog data transmission solutions, which are for complex analog signal conditioning circuitry. In such an
usually realized by means of carrier-frequency amplifier and approach, the analog input chain measuring uncertainty is
FM modulation [13]. High-precision analog systems are con- primarily determined by the characteristics of the chosen ADC
siderably more difficult to implement to meet the accuracy of and the quality of the strain gage sensor.
digital systems. It was also necessary to develop a continuous The mechanical design of our ship shaft monitoring system
contactless power supply transmission system to accommodate is shown in Fig. 2. The system consists of specifically designed
the solution for permanent installations in ships. Our research rotor and stator parts. A block diagram of the electronic part
showed that it is possible to build a highly sophisticated torque of the system is depicted in Fig. 3. The rotor part includes
monitoring system by such an approach by only using off-the- the strain gage sensor, the measuring transmitter module, and
shelf components, with the overall measuring system charac- the contactless power reception coils, which are all embedded
teristics comparable with or better than most of the present into the compact protective husk. The stator part of the system
commercial solutions. This paper extends the description [16] contains the ac power supply filter, the receiver data-processing
of our approach to the overall measuring system design with module, the tensometer signal pickup, the inductive proximity
an emphasis on the detailed uncertainty analysis of the realized switch, and the power transmission coils.

Authorized licensed use limited to: Praveenraja SN. Downloaded on December 11, 2009 at 01:04 from IEEE Xplore. Restrictions apply.
3704 IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, VOL. 58, NO. 10, OCTOBER 2009

the ADC input range limits, then the long-term zero creep may
cause signal clipping for large torque values. This problem is
avoided in high-resolution systems (> 20 bits), because the
input signal is fed to the ADC with small or no amplification
at all. Although the full 24-bit resolution is not achieved, the
effective resolution of the measuring system, which depends on
the ratio between the maximum sensor output voltage UFS and
the reference voltage UREF , is comparable to 16-bit systems
that utilize the whole ADC input range. For example, for
the sensor with full-scale (FS) voltage output UFS = 10 mV
and bridge excitation UREF = 4.096 V, the best theoretical
effective resolution of the measurement system (which takes
into account the ADC quantization error only) is
 
UFS
Nef = log2 = 15.3 bit (3)
(UREF /224 )

under the ideal conditions and without input signal amplifica-


tion, which is comparable to classical 16-bit systems (Nef in (3)
does not take into account any other errors, such as noise, zero
and gain drifts, nonlinearity, etc.). The measurement system
effective resolution [see (3)] is smaller for sensors with a lower
maximum output voltage (e.g., Nef = 12 bit for sensor with
FS output voltage UFS = 1 mV). In that case, the measurement
system effective resolution may be improved by using a fixed-
Fig. 3. Shaft monitoring system block diagram.
gain input instrumentation amplifier without compromising the
input ADC span for larger zero creeps under the assumption
that the amplifier gain is sufficiently low.
Another major practical improvement of the 24-bit system
is that potentiometers R0 , Rg , and Rc can be left out, thus
eliminating the need for both manual zero and gain calibration
on the rotor side and the influence of these components’ aging
and thermal effects. The system zero need not be manually
calibrated, unlike in the case of the classical input measuring
chain design: it is sufficient to only store the zero output for
the unloaded shaft in the embedded software, whereas the gain
calibration is avoided by employing the precision integrated
reference voltage UREF for ratiometric bridge and ADC supply.
The classical gain calibration procedure, like for the system
shown in Fig. 1, is not necessary in this case because the
voltage um across the bridge diagonal can directly be deduced
Fig. 4. Torque input measuring channel structure.
from the ADC readouts. Hardware offset nulling of the bridge
A block diagram of the shaft monitoring system input mea- output during the zero calibration procedure is not necessary
suring chain is shown in Fig. 4. In this configuration, the strain because the initial bridge imbalance cannot lead the ADC input
gage sensor is powered by the accurate reference voltage source into saturation. However, the inclusion of the resistor Rc in
UREF , whereas the bridge diagonal voltage is fed to the input the application circuit might still be beneficial in a process
differential stage of the integrated circuit. The influence of the of quick verification of the sensor installation. An active anti-
reference voltage UREF variations is eliminated in the conver- aliasing filter can be left out in this configuration due to the
sion results due to the ratiometric measurement approach. The high oversampling frequency of the ΔΣ ADC (a simple passive
adjustable external instrumentation amplifier, similar to the one RC low-pass filter for eliminating the high-frequency noise
shown in Fig. 1, is not necessary because the acceptable mea- is sufficient). The power supply coupling noise (50/60 Hz) is
suring resolution of the strain gage signal can automatically be filtered out by means of digital filtering.
achieved due to the high resolution of the ADC. The reasoning The input measuring chain uncertainty mostly depends on
for going from 16- to 24-bit ADC can be explained as follows: the chosen ADC, which determines the overall measuring
In the 16-bit ADC system, it is necessary to set the zero level characteristics of the system. Design can easily be adapted for
and the gain of the input amplifier by potentiometers in such different applications by choosing the most appropriate ADC
a manner that the whole input analog range of the ADC is for each particular case. In this design, the most important
covered. However, if the output amplifier span is too close to electrical characteristics of the electronic components in the

Authorized licensed use limited to: Praveenraja SN. Downloaded on December 11, 2009 at 01:04 from IEEE Xplore. Restrictions apply.
DŽAPO et al.: DIGITAL MEASURING SYSTEM FOR MONITORING MOTOR SHAFT PARAMETERS ON SHIPS 3705

input measuring chain are the following: 24-bit resolution, The receiver module microprocessor unit handles a few
2 ppm of FS IN L, 2.5 ppm of FS gain error, 0.5-μV offset different real-time tasks: digital signal demodulation, rotational
error, 0.03 ppm of UREF /◦ C gain drift, and 10-nV/◦ C zero speed measurement, calculation of torque, power, and rota-
drift. The results of the measuring system experimental eval- tional speed, digital and analog data transmission to the remote
uation and the measurement uncertainty analysis are presented control unit, automatic zero calibration, and receiver module
in Sections III and IV, respectively. system self-monitoring.
The power transmission subsystem was designed as a low- The rotational speed is measured by counting the pulses
frequency rotary transformer. The primary transmission coils received from the inductive proximity switch (see Fig. 3)
on the stator side are powered by a standard alternating-current in the adaptive time window. The time window bounds are
source (230 VAC, 50–60 Hz), which delivers approximately synchronized with the incoming pulses and measured with
1 W to the secondary rotor side, under the assumption of high precision by using the method described in [17]. It is
6-mm air gap. The secondary coils are compactly embedded implemented by an interrupt-driven approach and a 24-bit
into the protective husk, which is attached to the revolving software timer, which is necessary for longer averaging time
shaft. The transmitter measuring module power consumption windows. The rotational speed measuring uncertainty is mostly
requirements were optimized to only 20% of the delivered determined by the stability of the temperature uncompensated
power, including the 350-Ω sensor consumption. The low- quartz oscillator. The speed measurement resolution of 1/30 s−1
frequency power transmission approach was chosen due to the is determined by the number of marks on the rotor husk.
simplicity of implementation, robustness, and elimination of The lowest priority microprocessor real-time task is dedi-
possible electromagnetic compatibility issues. cated to the calculation of torque, power, and rotational speed
The problem of contactless measuring data transmission can by means of (2). A floating-point arithmetic was employed to
be solved by a few different approaches, e.g., by means of preserve the final calculation result accuracy. The real-time rou-
inductive coupling, capacitive link, infrared communication, tines running at the same priority level transmit the processed
RF transmission, etc. Most of the existing commercial shaft measuring results to the remote ship management system. Our
monitoring systems predominately rely on the analog FM mod- shaft monitoring system provides means for communication by
ulation approach for contactless measuring result transmission the most common analog and digital industrial protocols. The
over the near-field coupling media (e.g., carrier-frequency am- high-accuracy current-loop transmission (4–20 mA) is realized
plifier [13]). The analog FM modulation approach can have by using 16-bit current-mode integrated D/A converters. Al-
a significant impact on the overall measuring uncertainty, re- though the analog transmission deteriorates the overall measur-
gardless of the basic coupling principle and the carrier fre- ing system uncertainty, it may be useful in practice, particularly
quency choice, due to the frequency temperature drift in the for complex supervisory control and data acquisition control
analog communication chain, the electronic components’ aging systems with a large number of analog inputs for monitoring
effects, and sensitivity to noise. The near-field low-frequency different quantities. The RS-232 link is suitable for the easy
data transmission link for torque measurements is a typical connection with the local PC for measuring system configuring,
choice for permanent industrial installations because the RF monitoring, and logging purposes. Remote connection using
communication may be susceptible to interferences caused by RS-485 provides a robust digital link to the remote control
other devices operating at the same license-free industrial, system without measuring data quality loss.
scientific and medical band. All of the digital and analog outputs to the remote ship man-
To avoid the typical problems related to the analog FM agement system are galvanically isolated from the electronics
modulation approach, we implemented a specifically designed around the main processor in the receiver module (see Fig. 3).
low-frequency data transmission link based on the digital This provides higher noise immunity of transmitted data and
frequency-shift-keying (FSK) modulation principle. The digital better safety characteristics of the system. The results of torque,
transmission approach is less sensitive to external influences power, and rotational speed measurements are transmitted to
and noise sources that may be present in realistic industrial the secondary output side of the galvanically isolated interface
environments. The modulated signal is transmitted from the in digital form.
rotor to the stator part of the system by means of near-field Automatic zero compensation is important for long-term
capacitive coupling through the small linkage capacitance of measuring system stability due to the possible zero creep as
an order of magnitude of 0.1–1 pF. Digital modulation is a result of residual stress in the shaft under variable load
performed by the microprocessor in real time without use of conditions. Zero compensation is performed when the shaft is
any external analog electronic components. The FSK signal is stopped, which is checked at the end of the main task loop cycle.
demodulated in the microcontroller unit at the receiver side by The conditions for starting the zero compensation procedure
measuring the time between two consecutive voltage level tran- are satisfied if no pulses from the proximity switch sensor are
sitions in the interrupt-driven procedure. Only valid transitions received for a certain amount of time.
that correspond to T1 = 1/f1 or T0 = 1/f0 are accepted, which
thus eliminates false transitions that might be present due to
III. R ESULTS
electromagnetic interference and noise. We repeat a few periods
of the logical-level signal frequency to minimize the impact of The system was subjected to laboratory tests to experimen-
false signal edges, which thus provides a slower but more robust tally determine the most important technical characteristics and
data communication link. the measuring uncertainty. The results of laboratory tests are

Authorized licensed use limited to: Praveenraja SN. Downloaded on December 11, 2009 at 01:04 from IEEE Xplore. Restrictions apply.
3706 IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, VOL. 58, NO. 10, OCTOBER 2009

TABLE I
OVERVIEW OF THE SHAFT MONITORING SYSTEM TECHNICAL CHARACTERISTICS

the differential measuring input. Two additional 87-Ω resistors


were added to complete the circuit that models the strain gage
output resistance seen from the transmitter module measuring
inputs (R+IN = R−IN = R/2, where R = 350 Ω is the nomi-
nal bridge resistance). The differential voltage ug was fed from
the floating dc voltage source to simulate the bridge imbal-
ance. The floating voltage source was used to prevent ground
loops that would deteriorate measurements by introducing the
influence of external noise. The applied test voltage ug value
was furthermore verified by the high-accuracy digital voltmeter
(DVM) with high common-mode rejection ratio (140 dB). The
bipolar test range ug = ±10 mV was chosen to simulate the
voltage on a strain gage sensor that can be expected under
realistic working conditions. Digital readings were transmitted
over the isolated capacitive near-field data link to the receiver
Fig. 5. Block diagram of the test setup. module and logged on a PC over the serial debug interface.
The accuracy of optional analog outputs (4–20-mA current
summarized in Table I. We designed a configurable measuring loops for torque, power, and rotational speed) was indepen-
setup that very accurately simulates the behavior of the strain dently tested by measuring the current in several points over
gage sensor (output voltage and bridge impedance) and allows the whole working temperature range. The ideal output cur-
fine control of the differential voltage applied to the measur- rent in each test point was defined by a numerical value for
ing input. Finally, we tested the measuring system behavior D/A conversion. The analog outputs were not the subject of
under realistic working conditions on the ship shaft during the detailed uncertainty analysis of the measuring system because
sea trial. the current loop introduces errors that are an order of magnitude
A block diagram of the test setup is shown in Fig. 5. The higher than the accuracy of the digital measuring chain. The
measuring system modules were placed into the temperature rotational speed input was tested in the frequency range of
test chamber. The floating adjustable battery-powered dc volt- 1–500 Hz, which corresponds to the bandwidth of the chosen
age source was used to simulate the signal at the strain gage proximity switch sensor and relates to rotational speeds of up
bridge diagonal seen from the transmitter module measuring to approximately 1000 min−1 (in case of 30 marks on the rotor
input. The actual voltage value for each test point was mon- husk).
itored by the high-accuracy multimeter (HP 3457), whereas The most important part of the test procedure was the de-
the measuring results were logged on a PC over the RS-232 termination of the torque input channel measuring uncertainty.
debug interface. The shaft rotational speed measuring accuracy The measuring system characteristics were tested in the tem-
was tested by the square waveform fed from the high-accuracy perature range from −10 ◦ C to 60 ◦ C by repeating the identical
signal generator (Agilent 33250 A). test sequence four times. Each test sequence contains a series
The torque measuring channel input was modeled by the of individual measurements at three characteristic temperatures
equivalent resistive network and the floating voltage source. (−10 ◦ C, 25 ◦ C, and 60 ◦ C). The test voltage ug that simulates
The Wheatstone bridge has an equivalent impedance R/2 seen the output of the strain gage bridge diagonal was applied in
from each measurement input, whereas the bridge diagonal 21 points over the ±10-mV nominal input range (in 1-mV
voltage can be represented by common-mode UCM and differ- steps) by means of the adjustable floating dc voltage source.
ential ud components. Two 175-Ω resistors in the test circuit The voltage ug in each test point was controlled by DVM, the
are connected in series to simulate the realistic load for the accuracy of which was at least an order of magnitude better than
onboard voltage reference UREF and to provide the common- the system under test. Furthermore, the DVM readings ug for
mode voltage UCM = UREF /2 to achieve optimal conditions at each test point were averaged after repeating ten measurements

Authorized licensed use limited to: Praveenraja SN. Downloaded on December 11, 2009 at 01:04 from IEEE Xplore. Restrictions apply.
DŽAPO et al.: DIGITAL MEASURING SYSTEM FOR MONITORING MOTOR SHAFT PARAMETERS ON SHIPS 3707

Fig. 6. Recorded digital waveform um (t) for the test point ug = −5 mV.
Fig. 7. Total absolute error of the input measuring chain for raw and linearized
(the averaged voltage measured by DVM in some test point measurements.
will simply be denoted as ug ). The voltage um measured by each test point, ug,k is the average voltage measured by DVM
a transmitter module in each test point represents the A/D con- in each point, {a, b} are the linear regression coefficients, and
version result integrated over the 1-s time period and sent to the um,k,lin are the linearized measuring results. The normalized
receiver module in digital form over the galvanically isolated absolute total error of the torque input measuring channel ea
data link. The voltage um in each test point was measured is depicted in Fig. 7 for a single test sequence (at temperatures
60 times to estimate the measurement repeatability. The aver- of −10 ◦ C, 25 ◦ C, and 60 ◦ C). The figure shows the observed
aged measured result in a single test point after 60 measure- absolute error ea for both raw averaged measurements ūm and
ments will be denoted as ūm . An example of the recorded linear least-square fitted data um,lin . The influence of zero drift,
digital waveform um for a single test point is shown in Fig. 6. gain error, and nonlinearity can clearly be noticed in Fig. 7. The
The absolute total error ea normalized on the FS range in total absolute error due to nonlinearity, zero, and gain errors
each test point was calculated by means of does not exceed approximately ±350 ppm of the nominal FS
ūm − ug over the whole working temperature range. This result was
ea = (4) obtained without any kind of gain calibration.
UFS
The zero and gain drift errors were calculated from the
where UFS = 10 mV is the nominal FS range. The nominal FS linearized measuring results. The maximum observed zero drift
sensor output of 10 mV for error analysis was chosen as a good does not exceed 2.5 μV (250 ppm of the nominal FS), which
representative value for most practical situations. However, the corresponds to the change of 36 nV/◦ C (3.6 ppm of the nominal
input measuring chain can easily handle larger sensor outputs FS/◦ C). The contribution of zero drift to the total error is
without saturation (see the explanation in Section II). minimal when zero calibration is performed at the middle of the
Zero calibration of the torque input measuring channel for working temperature range. The gain error behavior is shown in
each test sequence was performed at 25 ◦ C, which corresponds Fig. 8 by plotting the error of the linear least-square-fitted data
to the middle of the working temperature range. Input channel at three different temperatures for a single test sequence with
zero calibration was not performed again at −10 ◦ C and 60 ◦ C removed influence of the zero drift (we get highly repeatable
for the same test sequence to measure the zero drift effects results for the other three test sequences). The maximum total
over the working temperature range and to more accurately gain error is about 250 ppm of the FS (which includes static
model the realistic working conditions. Raw measuring results gain error and temperature drift), whereas the temperature gain
um include the influence of all uncertainty sources (zero and drift alone is about 2.5 ppm/◦ C of the FS.
gain drifts, nonlinearity, random noise, etc.). We performed the The torque measuring channel nonlinearity was estimated
following least-square linearization of the averaged measuring from the raw measuring curve ūm after removing the offset and
results ūm,k for each temperature (separately for every test gain errors. The input measuring channel nonlinearity charac-
sequence) to accurately estimate the zero and gain drift errors teristic for a single test sequence at temperatures of −10 ◦ C,
by minimizing the influence of random noise in each test point: 25 ◦ C, and 60 ◦ C is shown in Fig. 9. One may notice the similar
behavior of the nonlinearity error at all test temperatures. The

Np
(ūm,k − f (ug,k , a, b))2 → min, maximum observed nonlinearity in the four test sequences was
k=1
about 200 ppm of the nominal FS.
um,k,lin = f (ug,k , a, b) = a · ug,k + b (5) All the measurement errors discussed so far were based on
the analysis of averaged results ūm calculated from 60 raw
where NP = 21 is the number of points in a test sequence at measurements um in a single test point. The measurement
fixed temperature, ūm,k is the averaged measuring result in repeatability can be analyzed by plotting standard deviations

Authorized licensed use limited to: Praveenraja SN. Downloaded on December 11, 2009 at 01:04 from IEEE Xplore. Restrictions apply.
3708 IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, VOL. 58, NO. 10, OCTOBER 2009

Fig. 8. Input measuring chain gain error. Fig. 10. Standard deviation of raw measurements for each test point in a single
test sequence at two different temperatures.

Fig. 9. Input measuring chain nonlinearity.

of each series of 60 individual measurements um for every


test point. The standard deviation for one test sequence at Fig. 11. Installation of the shaft monitoring system for the sea trial test.
temperatures of −10 ◦ C and 60 ◦ C is shown in Fig. 10. We measuring system in a few test points. The observed differences
determined the expected standard deviation of the individual were within the 0.5% error limits. Reference measurements
measurements um from all the available measuring results (in and system comparison were conducted and confirmed by
four test sequences) and calculated the value of about σum = an independent accredited institution (“Brodarski Institute,”
1.2 μV (120 ppm of the nominal FS). Croatia). The results show excellent agreement between our
The results presented in Figs. 7–10 and Table I are obtained and the reference measuring systems, because the achievable
for the nominal FS sensor output of 10 mV. However, if the FS accuracy in the field test was limited by both strain gage
output voltage is smaller the given error figures will increase sensor parameter uncertainties and errors related to the sensor
in proportion with the ratio between the nominal and actual installation procedure (strain gage factor uncertainty, angular
FS sensor outputs. This applies to zero drift, nonlinearity, and positional errors during each sensor installation, etc.). The
repeatability errors, but the gain error does not depend on the FS power readout error result takes into account the accuracy of the
sensor output. A more detailed measuring uncertainty analysis whole digital measuring chain (sensor, torque, and rotational
is provided in Section IV. speed measuring channel) and the uncertainty of the reference
The measuring system was installed and subjected to the measuring system.
test in realistic working environment during the sea trial of
the car carrier ship “Grande Angola” in “Uljanik” shipyard,
IV. T ORQUE M EASUREMENT U NCERTAINTY A NALYSIS
Pula, Croatia. The installation of the measuring system is shown
in Fig. 11. We compared the measuring results of the shaft- The value of the shaft torque is calculated by means of
transmitted power measured by our system and the reference (2) from the measured bridge diagonal voltage um and the

Authorized licensed use limited to: Praveenraja SN. Downloaded on December 11, 2009 at 01:04 from IEEE Xplore. Restrictions apply.
DŽAPO et al.: DIGITAL MEASURING SYSTEM FOR MONITORING MOTOR SHAFT PARAMETERS ON SHIPS 3709

user-provided system constants. The torque measurement σum = 1.2 μV). All the other error sources in (6) will be treated
uncertainty analysis will be divided into two parts: 1) the as Type-B uncertainty components with uniform probability
measuring uncertainty analysis of the electronic part of the density distribution. If we denote the observed extremal values
measuring system (including the sensor) and 2) the analysis of some error quantity X by a+ (X) and a− (X) and assume
of the influence of user-provided constant uncertainties in the that the interval [a+ (X), a− (X)] represents 100% containment
torque calculation formula [see (2)]. limits with equal probability of obtaining the value X within
The measuring uncertainty of the electronic part of the these limits, then the Type-B uncertainty u(X) can be calcu-
system is associated with errors arising from the strain gage lated as
sensor and the input measuring chain electronics. The most
a(X) a+ (X) − a− (X)
important error sources related to the strain gage sensor are u(X) = √ , a(X) = (7)
the following: sensor thermal drift, thermocouple effects of 3 2
connecting wires, angular sensor position misalignment errors, where a(X) is a half-width of the uniform probability distri-
output voltage nonlinearity, transverse strain gage sensitivity, bution definition interval. The half-widths of the containment
etc. Some of these error sources can be neglected in the full- intervals for each Type-B error source in (6) were determined
bridge strain gage configuration, namely, sensor thermal drift by measurement:
and nonlinearity, under the assumption of proper sensor in-
stallation. Although the thermocouple effects of the connecting a(Gacc ) = 250 ppm, a(GT ) = 2.5 ppm/◦ C
wires may be significant in some applications [2], the possible
1 ΔZmax 1 2.5 μV
thermal gradients in our design were minimized by placing the a(ZT ) = · = · = 18 nV/◦ C
whole electronic circuitry, along with the sensor, inside a closed 2 ΔT 2 70 ◦ C
thick metal enclosure. In such a design, the complex ac bridge a(IN L) = 2 μV
precision excitation for canceling out the thermocouple effects 1 UREF 1 4.096 V
is not necessary, and a much simpler dc reference voltage a(Q) = · 24 = · = 0.12 μV. (8)
2 2 2 224
supply can be used. Under the assumption that the strain gage
sensor is carefully and properly placed on the shaft surface The estimated standard uncertainties for Type-B error sources
along the biaxial stress principal directions, the sensor angular in (6) were calculated by means of (7) from interval half-widths
misalignment errors [2] can be neglected in most practical given by (8) as
applications. The influence of the uncertainty of the tensometer
u(Gacc ) = 144.3 ppm u(GT ) = 1.44 ppm/◦ C
constant Kt will be covered in the second part of the uncertainty
analysis. The transverse strain gage sensitivity was not taken u(ZT ) = 10.4 nV/◦ C u(IN L) = 1.16 μV
into account in this analysis.
u(Q) = 70.5 nV. (9)
The ADC contribution to the measurement uncertainty can
be analyzed by taking into account the quantization error, zero We do not take into account the absolute zero accuracy of the
temperature drift, gain accuracy, gain temperature drift, noise, input measuring chain because the reference input zero level
and nonlinearity. The contribution of each uncertainty source is determined before measurement by means of the software
can be described by the following error model: zero calibration. The sensitivity coefficients for the calculation
of the combined standard uncertainty can be derived from the
um = (1+Gacc +GT ·ΔT )·um,id +ZT ·ΔT +IN L+Q+N
error model [see (6)] as
(6)
∂um ∂um
where um is the measured voltage, um,id is the ideal voltage cGacc = = um,id cGT = = ΔT · um,id
∂Gacc ∂GT
at the bridge diagonal, Gacc is the gain error, GT is the gain
temperature drift, ZT is the zero offset temperature drift, IN L ∂um ∂um
cZT = = ΔT cIN L = =1
is the integral nonlinearity, Q is the quantization error, and N ∂ZT ∂IN L
is the random noise. All the error coefficients in (6) (except Q) ∂um ∂um
were determined from laboratory test results. The error sources cQ = =1 cN = = 1. (10)
∂Q ∂N
in the analytical error model [see (6)] can be expressed as
Type-A and Type-B uncertainties [21]. The repeatability error The combined standard uncertainty uc (um ) of the readout
N was determined by the statistical analysis of a series of result um over the whole working temperature range can be
observations and represents the Type-A uncertainty component. calculated by means of (11), shown at the bottom of the
The uncertainty due to the measurement nonrepeatability u(N ) page. We assume that all the uncertainty sources in model (6)
was calculated as a standard deviation of voltage values um are uncorrelated. For the strain gage sensor with maximum
measured in the procedure described in Section III (u(N ) = output voltage UFS = 10 mV and working temperature change


uc (um ) = c2Gacc · u2 (Gacc ) + c2GT · u2 (GT ) + c2ZT · u2 (ZT ) + c2IN L · u2 (IN L) + c2Q · u2 (Q) + c2N · u2 (N ) (11)

Authorized licensed use limited to: Praveenraja SN. Downloaded on December 11, 2009 at 01:04 from IEEE Xplore. Restrictions apply.
3710 IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, VOL. 58, NO. 10, OCTOBER 2009

ΔT = 70 ◦ C, the combined standard uncertainty given by (11) drift, nonlinearity, quantization, and nonrepeatability errors in-
yields uc (um ) ≈ 2.53 μV (253 ppm of nominal FS). creases the relative standard uncertainty because their absolute
The strain gage sensor was directly interfaced to the values (in microvolts) are unchanged, and their contribution to
24-bit ΣΔ ADC to achieve good measuring system effective the relative readout error becomes very significant for small
resolution [see (3)] without using an amplifier in the input values of um . For very small values of the measured voltage
stage. However, the input measuring chain noise should be um , the relative standard uncertainty given by (13) can be
considered to estimate the number of significant bits of such approximated by (14), shown at the bottom of the page. In the
high-resolution ADC. The effective number of bits (ENOB) first approximation, the nominator of uc,r (um ) does not depend
provides a convenient measure that represents the number of on the voltage um,id for small measuring values.
bits of an ideal (noise-free) ADC that would exactly produce The combined standard uncertainty uc (M ) of the torque
the same RMS quantization noise as experimentally observed readout value can be calculated by taking into account the con-
by measurements of the system under test. ENOB can be tribution of each parameter uncertainty in the torque calculation
calculated by means of formula [see (2)] and the combined standard uncertainty of the
bridge diagonal voltage measurement uc (um ) as follows:
uc,n (um )
ENOB = Nid − log2  2  2
uQ ∂M ∂M
u2c (M ) = u2 (G) + u2 (D)
1.67 μV ∂G ∂D
= 24 − log2 = 19.4 bit (12)  2  2
70.5 nV ∂M ∂M
+ u (Kt ) +
2
u2c (um ). (15)
where Nid is the ideal number of bits of the ADC, uc,n (um ) is ∂Kt ∂um
the combined measuring uncertainty calculated by (11) without
It should be noted that the uncertainty analysis of UREF in (15)
taking into account linearity errors (zero and gain), and uQ is
is not necessary due to the ratiometric approach to the mea-
the theoretical RMS quantization error of the 24-bit ADC [as
suring chain design. The sensitivity coefficients in (15) for the
calculated in (9)]. The factor uc,n (um ) in (12) corresponds to
estimation of standard uncertainty uc (M ) can be determined
the measured RMS noise of the input measuring chain.
from the torque calculation formula [see (2)]:
It is also interesting to consider the relative standard un-
certainty of the measured bridge diagonal voltage uc,r (um ) ∂M M ∂M 3M
defined as cG = = cD = =
∂G G ∂D D
uc (um ) ∂M M ∂M M
uc,r (um ) = . (13) cKt = =− cum = = . (16)
um ∂Kt Kt ∂um um

For um = UFS , the normalized absolute and relative standard The combined standard uncertainty uc (M ) of the torque read-
uncertainties are identical. For smaller bridge diagonal voltages out result M can be calculated by means of (17), shown at the
um , the contribution of gain accuracy and drift factors Gacc and bottom of the page. The combined relative standard uncertainty
GT in (13) is unchanged, because the relative gain errors do not uc,r (M ) can easily be derived from (17) as (18), shown at the
depend on a readout value. However, the influence of zero offset bottom of the page.

1  2
uc,r (um ) ≈ cZT · u2 (ZT ) + c2IN L · u2 (IN L) + c2Q · u2 (Q) + c2N · u2 (N ) (14)
um


uc (M ) = c2G · u2 (G) + c2D · u2 (D) + c2Kt · u2 (Kt ) + c2um · u2c (um ) =
   2  2
2
1 3 1
= |M | · · u2 (G) + · u2 (D) + · u2 (Kt ) + u2c,r (um ) (17)
G D Kt

 2  2  2
uc (M ) 1 3 1
uc,r (M ) = = · u2 (G) + · u2 (D) + · u2 (Kt ) + u2c,r (um ) (18)
|M | G D Kt

Authorized licensed use limited to: Praveenraja SN. Downloaded on December 11, 2009 at 01:04 from IEEE Xplore. Restrictions apply.
DŽAPO et al.: DIGITAL MEASURING SYSTEM FOR MONITORING MOTOR SHAFT PARAMETERS ON SHIPS 3711

The shear modulus uncertainty u(G) depends on the supplied For a very small torque and a bridge diagonal voltage um =
manufacturer’s data, and the tolerance of this parameter is 1% UFS , the relative standard uncertainty is
usually less than ±0.1%, whereas the shaft diameter uncertainty
depends on the manufacturing process tolerances (for example, c,r (M )
u1%
for shaft diameter D = 500 mm with ±0.1-mm manufacturing   2
tolerance, the relative error is ±0.02%). The strain gage factor 2.53 μV
= 3.36·10−7 +1.19·10−7 +2.08·10−6 +
Kt tolerance (typically ±0.25–0.5%) is known from the sup- 0.1 mV
plied manufacturer’s data and represents the significant source
of torque readout relative error. The experimental determination = 2.54 · 10−6 + 6.4 · 10−4 ≈ 2.54%.
of the true value Kt for the strain gage sensor installed on The results of the torque readout relative standard uncertainty
a ship shaft is not possible because the sensor is typically analysis show that the principal sources of total uncertainty
installed on a large diameter shaft, and the application of the for large torque values are the user-provided constants in (2).
reference calibration moment is not possible. The last error term The highest contribution comes from the parameter Kt , even
in (18) corresponds to the relative standard uncertainty of the under the assumption of high-quality sensor (e.g., a(Kt ) =
measured bridge diagonal voltage uc,r (um ) given by (13). The ±0.25%), whereas the contribution of the electronic part of
user-provided parameters G, D, and Kt in (18) can be treated the measuring system is negligible. The torque readout relative
as Type-B uncertainty components with uniform probability standard uncertainty for small load values primarily increases
density distribution. The corresponding standard uncertainties due to the uncertainties introduced by the input measuring
can be calculated from the estimated parameter tolerances by chain electronics. Our results are comparable to or better than
means of similar existing state-of-the-art commercial solutions to the
authors’ best knowledge (e.g., [22] and [23]), which usually do
u(G) 0.1% not provide such clear figures for relative torque measurement
= √ = 0.058%
G 3 uncertainty over the whole working temperature range and the
full input measuring span (usually, only FS error figures and
u(D) 0.02%
= √ = 0.0115% measuring characteristics such as repeatability or linearity are
D 3 provided).
u(Kt ) 0.25%
= √ = 0.1443. (19)
Kt 3 V. C ONCLUSION
The novel shaft mechanical parameters monitoring system
The torque readout relative combined standard uncertainty for permanent installation in ships has been described. The
uc,r (M ) was calculated from (18) by taking the user-provided torque input measuring channel was realized by directly inter-
parameter standard uncertainties [see (19)] and the voltage facing the strain gage sensor to the high-resolution ΣΔ ADC.
measurement relative standard uncertainty uc,r (um ). For large Outstanding accuracy and stability of the input measuring chain
torque values (um ≈ UFS ), the relative standard uncertainty were accomplished by a particular design approach, on-shaft
uc,r (M ) can be estimated by means of (20), shown at the digital signal processing, and contactless data transmission in
bottom of the page, where the relative standard uncertainty digital form. The observed input measuring chain uncertainty
uc,r (um ) is approximated by the normalized absolute uncer- was about 250 ppm of the FS over the full working temperature
tainty uc (um ) for the FS sensor output. However, for small range for a ±10-mV FS sensor output. The evaluation under
torque values, the contribution of the uc,r (um ) uncertainty realistic working conditions showed excellent agreement of the
term must take into account the value of the measured bridge designed system with the reference shaft torque measuring sys-
diagonal voltage um . For example, for um = 10% UFS , the tem. The software-based input measuring chain calibration, the
torque readout relative standard uncertainty is saturation-free ADC operation (regardless of the initial bridge
imbalance), and the exceptional long-term measuring system
stability are some of the most important practical benefits
c,r (M )
u10% achieved by the proposed measuring system design. It should
  2 be pointed out that excellent measuring system characteristics
2.53 μV were achieved by a very simple design based only on a small
= 3.36·10−7 +1.19·10−7 +2.08·10−6 +
1 mV number of off-the-shelf components.
The relative uncertainty of torque and power readouts
= 2.54 · 10−6 + 6.4 · 10−6 ≈ 0.30%. depends both on measuring chain characteristics and the


uc,r (M ) = (0.058%)2 + 9 · (0.0115%)2 + (0.1443%)2 + (253 ppm)2 =

= 3.36 · 10−7 + 1.19 · 10−7 + 2.08 · 10−6 + 6.4 · 10−8 ≈ 0.16% (20)

Authorized licensed use limited to: Praveenraja SN. Downloaded on December 11, 2009 at 01:04 from IEEE Xplore. Restrictions apply.
3712 IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, VOL. 58, NO. 10, OCTOBER 2009

user-provided system constants, which are necessary to calcu- [19] A. Santic, Z. Stare, R. Magjarevic, B. Banovic, M. Saban, and V. Bilas,
late the actual value of the physical quantity from the mea- “Multichannel telemetry system for measurements on the rotating shaft,”
in Proc. 37th International Annual Gathering KoREMA, Zagreb, Croatia,
sured strain gage sensor output voltage. The influence of the 1992, pp. 360–363.
electronic measuring chain on the torque and power readout [20] A. Santic, B. Banovic, Z. Stare, R. Magjarevic, and M. Cifrek, “A teleme-
accuracy is negligible when compared with the impact of user- try system with microprocessor for measurement of relevant data from the
rotating shaft,” in Proc. 34. Jugoslavenska konferencija ETAN-a, Zagreb,
provided constants over most of the measurement ranges. The Croatia, Jun. 4–8, 1990, pp. 177–184.
input chain measuring uncertainty has a significant impact on [21] Guide to the Expression of Uncertainty in Measurement. Geneva,
torque and power readouts only when measuring very small Switzerland: International Organization for Standardization (ISO), 1995.
[22] McNab Wireless Rotary Shaft Horsepower/Torque Meter SHP 62000.
torque values. Mount Vernon, NY: McNab Corporation.
[23] The Palco (R) Torque Measuring System, VAF Instruments B.V..
Dordrecht, Netherlands.
R EFERENCES
[1] I. J. Garshelis, “Torque and power measurement,” Measurement, Instru-
mentation, and Sensors Handbook CRCnetBase 1999, CRC Press LLC,
1999. Hrvoje Džapo (M’04) was born in Zagreb, Croatia,
[2] K. Hoffmann, An Introduction to Measurements Using Strain Gages. in 1975. He received the Dipl.-Eng. and Ph.D. de-
Darmstadt: Hottinger Baldwin Messtechnik GmbH, 1989. grees in electrical engineering from the University of
[3] I. Sasada, Y. Etoh, and K. Kato, “A figure-of-eight flexible pickup coil Zagreb, Zagreb, in 1999 and 2007, respectively.
for a magnetostrictive torque sensor,” IEEE Trans. on Magnetics, vol. 42, He is currently a Senior Research Assistant with
pp. 3309–3311, 2006. the Department of Electronic Systems and Informa-
[4] I. Sasada and M. Akinaga, “High sensitivity magnetic head for a shaft tion Processing, Faculty of Electrical Engineering
torque sensor,” J. Appl. Phys., vol. 91, pp. 7792–7794, 2002. and Computing, University of Zagreb. His research
[5] H. Ruser, U. Troltzsch, and M. Horn, “Low-cost magnetic torque sensor interests are focused on electronic instrumentation,
principle,” in Proc. of IEEE Sensors 2002, 2002, vol. 2, pp. 901–904. smart sensors, embedded systems design, and imple-
[6] J. Zakrzewski, “Combined magnetoelastic transducer for torque and force mentation of computational inverse techniques for
measurement,” IEEE Trans. Instrumentation and Measurement, vol. 46, nondestructive inspection.
pp. 807–810, 1997.
[7] A. Falkner, “A capacitor-based device for the measurement of shaft
torque,” IEEE Trans. Instrumentation and Measurement, vol. 45, pp. 835–
838, 1996.
[8] R. F. Wolffenbuttel and J. A. Foerster, “Noncontact capacitive torque Zoran Stare (M’77) was born in Zagreb, Croatia, in
sensor for use on a rotating axle,” IEEE Trans. Instrumentation and 1944. He received the Dipl.-Eng., M.Sc., and Ph.D.
Measurement, vol. 39, pp. 1008–1013, 1990. degrees in electrical engineering from the Univer-
[9] W. Y. Li, C. C. Cheng, and Y. L. Lo, “Investigation of strain transmission sity of Zagreb, Zagreb, in 1967, 1978, and 1997,
of surface-bonded FBGs used as strain sensors,” Sensors and Actuators respectively.
A: Physical, vol. 149, pp. 201–207, 2009. He is currently an Associate Professor with the
[10] D. A. Kerth and D. S. Piasecki, “An oversampling converter for strain Department of Electronic Systems and Information
gauge transducers,” IEEE J. Solid-State Circuits, vol. 27, pp. 1689–1696, Processing, Faculty of Electrical Engineering and
1992. Computing, University of Zagreb. His teaching and
[11] “Resistive bridge basics: Part two,” in Application Note 3545. Maxim research interests are focused on electronic instru-
Integrated Products, Inc., 2005. mentation, measurement techniques, industrial elec-
[12] J. O’Dowd, A. Callanan, G. Banarie, and E. Company-Bosch, “Capacitive tronics, embedded systems, and biomedical engineering.
sensor interfacing using sigma–delta techniques,” in Proc. Sensors, 2005
IEEE, 2005, p. 4.
[13] S. Kuhn, “High precision torque measurement system,” in Proc. SENSOR
Conference 2007, Nürnberg, Germany, 2007, vol. 2, pp. 51–56. Nenad Bobanac was born in Zagreb, Croatia,
[14] F. M. L. van der Goes and G. C. M. Meijer, “A simple accurate bridge– in 1951. He received the Dipl.-Eng., M.Sc., and
transducer interface with continuous autocalibration,” IEEE Trans. Instru- Ph.D. degrees in naval architecture from the Uni-
mentation and Measurement, vol. 46, pp. 704–710, 1997. versity of Zagreb, Zagreb, in 1976, 1986, and 1996,
[15] D. McCartney, A. Sherry, J. O’Dowd, and P. Hickey, “A low-noise low- respectively.
drift transducer ADC,” IEEE J. Solid-State Circuits, vol. 32, pp. 959–967, In 1978, he was with the Brodarski Institute,
1997. Marine Research and Special Technologies, Zagreb,
[16] H. Dzapo, Z. Stare, and N. Bobanac, “Digital measuring system for where he was the Head of the R&D Vibration and
monitoring motor shaft parameters on ships,” in Proc. 2008 IEEE In- Shocks Laboratory from 1995 to 2000 and the CEO
ternational Instrumentation and Measurement Technology Conference, from 2000 to 2004. In 2006, he founded BOBLab
Victoria, Canada, 2008, pp. 2079–2084. d.o.o., Zagreb, where he is currently the CEO. He has
[17] H. Dzapo, Z. Stare, and N. Bobanac, “Portable wireless measuring sys- been a national representative in the international standardization organization
tem for monitoring motor shaft parameters,” in Proc. 21th IEEE Instru- ISO/TC108 on mechanical shock and vibration since 1986, a convener of
mentation and Measurement Technology Conference, Como, Italy, 2004, the national standardization committee DZNM TO108 on mechanical shock
pp. 901–906. and vibration, and a convener of the ISO working group for standardization
[18] M. Ivancic, R. Magjarevic, and Z. Stare, “Analysis and optimization (ISO/TC108/SC2/WG2 vibration of ships) since 2003. His research interests
of multichannel measurements at navigable objects,” Elektrotehnika are focused on the different aspects of structure vibration and human suscepti-
ELTHB2, vol. 36, no. 4–5, pp. 197–206, 1993. bility to vibration.

Authorized licensed use limited to: Praveenraja SN. Downloaded on December 11, 2009 at 01:04 from IEEE Xplore. Restrictions apply.

You might also like