4616 North Mingo Road, Tulsa, Oklahoma. USA Tel: (1) 918-834-2337 Fax: (1) 918-834-9887 Email: CLI.tech@corelab.com
ITEM QUANTITY CATALOG NO. DESCRIPTION
J une13, 2010 All instruments and instrumentation sub-systems were manufactured or assembled in the United States of America. Specifications subject to change without notice. RPTA-200 WELL BORE CONDITIONS RETURN PERMEABILITY TEST APPARATUS is a high pressure, high temperature development of the 100 series designed specifically for the evaluation of drilling mud and mud additives and their affect on reservoir rock samples. A re-designed layout improves operator access to the core holder, accumulators allowing for increased efficiency of operation.
The system allows a number of tests to be conducted including liquid permeability, return permeability; mud lift-off, leak-off rates and the evaluation mud additives at reservoir temperature and pressure. A custom reciprocating mud pump allows mud to be re-circulated across the upper end face of the core sample, at a con trolled shear rate, to simulate filtrate build up and to allow the evaluation of leak off rates. Differential pressure transducers measure the pressure drop across the core material allowing investigation of the depth of invasion of the filtrate and a PC linked digital balance monitors the leak-off rate. The core holder in encapsulated in a custom designed heating jacket and all inlet lines are fitted with block heaters allowing the accumulators to be kept at ambient conditions. Allowing easy cleaning and loading of the accumulators to maximize the number of fluids evaluated daily.
With the addition of extra fluid accumulators the system can be used to evaluate non-acid clean-up treatments. Whilst with the optional construction of wetted parts in Monel allows the evaluation of non-inhibited acids.
Scope of Supply:
Fluid delivery system: HPLC Pump System is capable of 10cc/min to a maximum pressure of 6,000 psig. To obtain higher flow rates up to 25 cc/min requires an additional, interchangeable head, which lowers the maximum pressure to 2,500 psig. Four Accumulators Capacity 1,000 cc, Pressure Rating 10,000psi, Core Holder PPCH400 see below 1.5-inch diameter. Two Temperature Controllers Two Back Pressure Regulators Pressure Measurement System consisting of: Two High Accuracy Transducers 0-7,500 psi, 0.05% Technical Specification
Core Lab Instruments
4616 North Mingo Road, Tulsa, Oklahoma. USA Tel: (1) 918-834-2337 Fax: (1) 918-834-9887 Email: CLI.tech@corelab.com
ITEM QUANTITY CATALOG NO. DESCRIPTION
J une13, 2010 All instruments and instrumentation sub-systems were manufactured or assembled in the United States of America. Specifications subject to change without notice. full scale accuracy and Three Differential Pressure Transducers 0-25, 0-100 and 0-500 psid, 0.2% full scale accuracy Flow rate measurement system consisting of (1) Electronic Balance 0-2100 grams, 0.001 gm accuracy. Computer with Minimal Configuration: Core Duo, 500 GB Hard Drive, 2 GB RAM, 40x CD- ROM, 21 LCD Monitor with Windows Vista Professional Operating System RPTA-200 Operating Software with real time display, data acquisition and safety monitoring system. Operating and Maintenance Manual Toolkit
PPCH-400 PPCH PRESSURE PORTED HASSLER COREHOLDER (1.5 inch Diameter) 316SS body with Viton sleeves and seals. 10,000psig confining pressure and 7,500psig pore pressure, with a maximum circulating pressure of 8,000 psig. Maximum temperature is 300 Deg. F. with Viton seals and sleeve, sample diameter is 1.5 inches and up to 10 inches in length. Two pressure ports are incorporated for determination of the invasion profile. The core holder is provided with a 3-ported dynamic flow by head, which allows the mud fluid or completion fluids to be flowed parallel across the face of the core sample allowing for the establishment of a mud filter cake simulating well bore conditions. The head also allows for the flow of fluid through the sample for return permeability tests.
Options:
Monel wetted parts for use with non-inhibited acids
Mixing Unit - MU100 mixing module complete with fluid reservoirs, paddle stirrer, mixing pump, high pressure mud/completion fluid pump, shear loop, heated formation simulator and temperature controller.
Pipe Rheometer - comprising 4 varying diameter tubing with digital mass flow meter, and gage and differential pressure transducer array. Complete with oil bath heating system to maintain sytem temperature.
Pump Options
Quizix QX6000 allows continuous flow pulse free flow with precision control of pressure, flow rate and volume; can operate in constant pressure or constant flow rate mode.
Isco FDS dual pump systems allows continuous flow pulse free flow with precision control of pressure, flow rate and volume can operate in constant pressure or constant flow rate mode.
FDS- 260D pump system allows for pulse free flow of fluids and improved accuracy in flow determinations. Allows pulse free flow rates of up to 100 cc/min min at pore pressures of up to 7,5000 psig. Pump control and data acquisition features can be integrated into the system operating software.
Technical Specification
Core Lab Instruments
4616 North Mingo Road, Tulsa, Oklahoma. USA Tel: (1) 918-834-2337 Fax: (1) 918-834-9887 Email: CLI.tech@corelab.com
ITEM QUANTITY CATALOG NO. DESCRIPTION
J une13, 2010 All instruments and instrumentation sub-systems were manufactured or assembled in the United States of America. Specifications subject to change without notice. FDS 500D pump system allows for pulse free flow of fluids and improved accuracy in flow determinations. Allows pulse free flow rates of up to 200 cc/min min at pore pressures of up to 3,750 psig. Pump control and data acquisition features can be integrated into the system operating software.
FDS 1000D pump system allows for pulse free flow of fluids and improved accuracy in flow determinations. Allows pulse free flow rates of up to 400 cc/min min at pore pressures of up to 2,000 psig. Pump control and data acquisition features can be integrated into the system operating software.