TABLE OF CONTENTS
Page
Section 1 Introduction
1.1 Purpose of the Manual _________________________________ 1
1.2 Customer Service _____________________________________ 1
1.3 Disclaimer ___________________________________________ 1
1.4 Copyright ___________________________________________ 2
1.5 Prescribing Information _________________________________ 2
Section 2 Safety
2.1 Safety Overview ______________________________________ 3
2.2 Safety Features ______________________________________ 3
2.3 Explanation of Symbols ________________________________ 4
2.4 Explanation of Warnings _______________________________ 4
2.5 Labels ______________________________________________ 5
Section 3 Product Introduction
3.1 TRIO XT Clinical Theory ________________________________ 6
3.2 TRIO XT Radiofrequency (RF) Modality ___________________ 8
3.3 TRIO XT Components _________________________________ 10
Section 4 Pre-Treatment Considerations
4.1 General Pre-Treatment Considerations and Recommendations _ 15
4.2 Pre-Treatment Consideration and Recommendations for RF
Treatments __________________________________________ 15
Section 5 TRIO XT Treatment Protocols
5.1 TRIO XT RF Pre-Set Parameter Settings __________________ 17
5.2 TRIO XT RF Treatment Procedure _______________________ 17
5.3 Specific Area Treatment Recommendations ________________ 18
Face Treatment Program _______________________________ 19
Forehead Treatment Program ___________________________ 20
Peri-Orbital (Eye) Treatment Program _____________________ 21
Peri-Oral (Mouth) Treatment Program _____________________ 22
Neck Treatment Program _______________________________ 24
Abdomen Treatment Program ___________________________ 25
Cellulite Treatment for Thighs and Buttocks Program _________ 26
Back Treatment Program _______________________________ 28
Buttocks Treatment Program ____________________________ 29
Arms Treatment Program _______________________________ 30
Hands Treatment Program ______________________________ 31
Legs Treatment Program _______________________________ 33
TABLE OF CONTENTS
Page
5.4 Post-RF Treatment Recommendations ____________________ 34
5.5 TRIO XT Recommended Number of Treatments and
Treatment Intervals ____________________________________ 34
5.6 Considerations for All Treatments ________________________ 34
5.7 Post-Treatment Considerations __________________________ 35
5.8 Recommended Maintenance Program _____________________ 35
SECTION 1: INTRODUCTION
1.1
This manual explains how to correctly install, operate and get the best performance from your TRIO
XT System. It will help you understand the basic technological principles and procedures that are
utilized by the TRIO XT method. It may be used as a guide, suitable for all levels of technical
expertise, in determining and servicing mechanical and electrical difficulties. This manual is
organized as a quick reference guide. Any additional updated or supplemental information for this
reference guide will be distributed to you as needed. Please read this manual carefully before
administering treatments with your system and keep it nearby for quick reference.
1.2
Customer Service
We welcome and appreciate your comments, questions, and suggestions in order to offer the highest
quality products, accessories, and services to our customers continually. Our qualified professional
teams are ready to assist you with a prompt response to any question communicated to us by
telephone, e-mail, and/or fax.
1.3
Disclaimer
While every attempt is made to ensure the accuracy and completeness of the information in this
document, some errors may exist. Sybaritic, Inc. does not accept responsibility of any kind for
customer loss due to use of or reliance upon this document.
Sybaritic, Inc. reserves the right to make changes without notice in the product described or
contained herein in order to improve design and/or performance. Sybaritic, Inc. assumes no
responsibility or liability for the use of the product, conveys no license or title under any patent,
copyright, or trademark right to this product or processes and makes no representation or
warranty that this product is free from patent, copyright, or trademark right infringement, unless
otherwise specified. Applications that are described herein for the product are for illustrative
purposes only. Sybaritic, Inc. makes no representation or warranty that such applications are
suitable for the specified use without further testing or modification.
1.4
Copyright
This manual is protected by copyrights. All rights are reserved. No part of this publication may be
reproduced, stored in a retrieval system, or transmitted in any form or by any means: electronic,
mechanical, photocopying, recording, or otherwise without the prior written permission from the
author and publisher. All inquiries should be addressed to Sybaritic, Inc. at the headquarters
address listed inside the front cover of this manual, along with a date from a copy of the original
purchase receipt of the TRIO XT System. No attempt has been made to designate as
trademarks or service marks all words or terms in which proprietary rights might exist. The
inclusion, exclusion, or definition of a word or term is not intended to affect, or to express a
judgment on, the validity or legal status of the word or term as a trademark, service mark, or
other proprietary term.
1.5
Prescribing Information
SECTION 2: SAFETY
2.1
Safety Overview
As with any medical device, appropriate care must be taken to ensure proper and safe use. This
entire manual should be thoroughly reviewed before operating the instrument. Further, the
physician should attend a certified training course before using the TRIO XT System for medical
treatments.
All users should become familiar with the requirements for safe use of medical RF and
ultrasound systems as described in the American National Standards Institute (ANSI)
publications.
2.2
Safety Features
Standby/Ready Controls
The unit will not output RF or ultrasound energy when in the STANDBY mode.
Key Switch
Limits the accessibility of the machine to only those that have access to the operating
key. Only those with appropriate training and knowledge of the safe use of the TRIO
XT System should have access to the operating key.
power the machine down. To resume normal operation, rotate the emergency stop
button in the direction of the arrow (clockwise) the button will return to its normal
position and system power will resume.
2.3
Explanation of Symbols
This symbol indicates power is ON.
2.4
Explanation of Warnings
DANGER
Indicates an imminent situation in which the
possibility of exposure to electrical shock or a fire
hazard could result in serious injury to the
practitioner, patient, or staff.
WARNING
Indicates a potentially hazardous situation in
which the possibility of exposure to electrical
shock may occur that could result in serious injury
to the practitioner, patient, or staff.
CAUTION
Indicates a potentially hazardous situation that, if
not avoided, could result in damage to the
equipment.
2.5
Labels
The labels shown are affixed to the TRIO XT device console. Users should be familiar with the
location and meaning of these labels and should review them regularly.
220V system
Affixed to the rear panel of
SYBARITIC, INC.
9220 James Ave., S.,
Minneapolis, MN 55431 USA
*Notes
MODEL:
TRIO XT
220v/50-60Hz
2010
IFE0911001
FUSE
5x20T
220VAC
(Left): After its three-dimensional helix structure unwinds, collagen fibers start to shrink. The skin
will become smoother and tighter and wrinkles will diminish.
(Right): Skin rejuvenation is the achieved result from the hyperplasia of collagen fibers.
3.2
3.2.1
Radiofrequency (RF) current is formed when charged particles flow through a closed circuit. As
the energy meets resistance in the tissue, heat is produced. The amount of heat will vary
depending on the amount of current, the resistance levels in the targeted tissue and the
characteristics of the electrodes. The amount of RF energy applied can be configured to target
specific tissues.
RF energy is a form of electromagnetic energy ranging from 3 kHz to 300 mHz. RF affects skin
by emitting high-frequency radio waves that mimic the thermal effects of lasers and intense
pulsed light sources. RF energy is similar to optical energy in that it interacts with the tissue to
produce a thermal change.
3.2.2
The TRIO XT System enables you to use either a bi-polar and tri-polar electrode configuration.
With a bi-polar system, the current flows between the positive and negative electrode. With a
tri-polar system, the current flows between three identical electrodes that are set at fixed
distances apart. When using the tri-polar hand piece, the TRIO XT electronic system
continuously switches the polarity between pairs of electrodes. Since the distances between
these electrodes vary, the end result is that different depths of penetration of the RF energy are
reached. The level of current is safely limited by the volume of tissue between each pair of
electrodes. No grounding pad is necessary. Multi-polar systems avoid the unpredictability of
mono-polar (single electrode) systems resulting from current passing through the body to the
grounding electrode. The fact that the distribution of the current in the tissue is safer and more
controlled is a major advantage of multi-polar technology.
3.2.3
3.2.4
A number of factors determine the clinical effect of RF energy. Basic output parameters are power
output, pulse width, and RF frequency.
3.3
TRIO XT COMPONENTS
10
Start
System checking
Select Language
11
PROGRAM ICONS
Custom programming key (set parameters accordingly)
Forehead button
EYE Peri-orbital button
Face button
Neck button
12
PROGRAM ICONS
Flashes to alert for abnormal coolant flow
(return
main menu)
to
Pause key
13
SETTNG ADJUSTMENTS
Press
ALARM
If alarm sounds during Ready mode, system will switch to standby mode automatically,
and
Press
Mini
facial
handpiece
Bi-Polar
Body
handpiece
14
Patient must read, understand and sign appropriate Informed Consent (invite to ask
any questions and provide answers to patients satisfaction)
15
Cleanse face with soapy water (do NOT apply moisturizer after cleansing) and dry
thoroughly.
Take before photos (5 views plus areas of opportunity, following clinic photography
protocol)
Ask if patient has a pacemaker or ICD (internal defibrillator), and confirm whether patient
is, or could be pregnant. If yes, discontinue procedure.
Ask patient to remove all jewelry, shirt or blouse, and put on gown.
Prior to TRIO procedure, if topical was applied, remove topical completely and
cleanse area with acetone. Then apply treatment grid markings (for face and
neck use 4 x 4 cm squares; for the body use 8 x 8 cm squares) with washable
skin marker.
16
Treatment Area
Treatment
Time
(minutes)
Cooling
Setting
Energy
Level
Handpiece
Face
Forehead
20
5
**
**
20%
20%
Facial
Facial
Eyes (Peri-Orbital)
**
20%
Facial
Neck
Abdomen
Legs
Back
Feet
Buttocks
Arm
Hand
10
20
20
20
20
20
20
15
**
NA
NA
NA
NA
NA
NA
NA
40%
60%
60%
50%
40%
60%
50%
40%
Facial
Body
Body
Body
Body
Body
Body
Body
5.2
5. Once the section being treated reaches 40C, this temperature must be maintained for
at least one (1) minute. In order to do this without causing discomfort to the patient, it
is helpful follow these steps:
a. Move to an immediately adjacent area and apply the hand piece for 10- 15
seconds;
b. Move back to the section that you are maintaining at 40C, and continue working the
area until the clinical end point has been maintained for at least a full minute. Continue in
this manner until the entire desired facial or body area has been treated.
6. Sensitivity levels around the eyes may make the treatment more uncomfortable and
should be adjusted to provide lower energy settings than other areas of the face.
7. Note: The average face will require treatment of approximately 16 4 x 4 cm treatment
zones. If you spend approximately one minute on each area to reach the target end point
temperature and maintain for one additional minute, the average face should take from
30 to 40 minutes to treat with a single pass. Multiple passes are not recommended.
Note: Use of the Audible Treatment Metronome (ATM) to help utilize the optimal speed of
movement of the treatment hand piece when performing the RF treatments is optional.
18
Activate Handpiece
5)Select energy %.
6)Set Audible Treatment Metronome (ATM).
7)Press READY key to start program.
8)Press foot pedal to activate handpiece.
Step1: Cleanse
Bi-Polar
Energy level
ATM Setting
40%
50 beats/ minute
60%
55 beats/ minute
Step 3: Move handpiece slowly from center to temple in 4 cm x 4 cm sections at a slow constant speed in a
back and forth motion until skin slowly reaches temperature is 40 degrees C. Treat all sections on the
upper portion of the face.
Step 4: Move handpiece from corners of the mouth to temple at a slow constant speed to treat the lower
portion of the face.
19
Full Face Treatment Time: 40 minutes total (20 minutes each side).
Treatment Tips:
I.
Move the handpiece slowly on the face at a slow constant circular or back and forth motion, in
4 cm x 4 cm sections.
II.
III.
Handpiece must make contact with skin before pressing foot pedal or a shock to the patient
may result.
IV.
V.
Check the surface skin temperature frequently with IR external thermometer (40 degrees C)
During operation, ask the patient if she feels warmth and adjust the energy setting accordingly,
energy setting is a guideline only, the setting may go higher than guideline.
6) Set Audible
Adjust energy level and ATM settings according to the chart on this page:
Bi-Polar
Step 1: Cleanse
Energy level
ATM Setting
40%
35 beats/ minute
60%
55 beats/ minute
Handpiece
Step 2: Spread conductive gel to local treatment area. Step 3: Move handpiece slowly from center to
temple in 4 cm x 4 cm sections at a slow constant speed in a back and forth motion until skin slowly
reaches temperature is 40 degrees C.
20
II.
Move the handpiece slowly on the face at a slow constant circular or back and forth motion, in
small trirants.
III.
Always remove handpiece from the forehead with two poles at the same time.
IV.
Make full contact with treatment area before pressing or depressing foot pedal or a shock to
the patient may result.
V.
Energy setting is a guide only energy may be set higher than these guide lines.
VI.
Check the surface skin temperature frequently with IR external thermometer (goal is to
maintain 40 degrees C for one minute)
VII.
During operation, ask the patient if she feels warmth and adjust the energy setting accordingly,
energy setting is a guideline only, settings may need to be higher to achieve 40 degrees C.).
to Activate
Energy level
ATM Setting
40%
90 beats/ minute
60%
90 beats/ minute
Step 1: Cleanse Step 2: Spread cooling gel to one eye area and then next eye area. Step 3: Move
handpiece around the outside corner of each eye in a small circular motion, concentrating on visible crows
21
feet Move at constant slow speed. Be careful not to stop movement. Eye area is sensitive, adjust energy
level accordingly.
Move the handpiece slowly on the face at a slow constant circular or back
and forth motion, in small quadrants
II.
During operation, ask the patient if she feels warmth and adjust the energy
setting accordingly, The recommended energy settings are guidelines only
(higher settings may be required to reach 40 degrees C.).
III.
Make full contact with treatment area before pressing foot pedal or a patient
shock may result.
IV.
activate handpiece.
Program Settings: Start new patients with the lowest recommended settings;
Continue to adjust energy and ATM settings according to the chart on this
page:
22
to Activate
Energy level
ATM Setting
40%
90 beats/ minute
60%
90 beats/ minute
Step 1: Cleanse
Step 2: Spread cooling gel. Step 3: Move mini handpiece in small circular motions surrounding the mouth
at constant speed. Do not move hand piece too slowly (this can cause sudden hot spots and patient
discomfort).
II.
III.
Move the handpiece slowly on the face at a slow constant circular or back and forth
motion, in small quadrants (4 cm x 4 cm grids).
IV.
Make full contact with treatment area before pressing or depressing foot pedal or a
patient shock may result.
V.
Check the surface skin temperature frequently with IR external thermometer (40
degrees C) for one full minute is the goal.
VI.
During operation, ask the patient if she feels warmth and adjust the energy setting
accordingly, The energy setting is a guideline only (energy may go higher than
guideline to reach 40 degrees C.).
23
Activate Handpiece
Energy level
ATM Setting
40%
50 beats/ minute
60%
55 beats/ minute
Step 1: Cleanse
Step 2: Spread cooling gel to one side of neck and then the other
Step 3: Move the facial handpiece from collar bone to mandible in 4 cm x 4 cm sections at a slow constant
speed (Do NOT apply hand piece directly above the Adams apple.
Time: 20 minutes total time (10 minutes per side).
Treatment Tips:
I.
Move the handpiece slowly on the neck at a slow constant circular or back and forth motion, in
small quadrants. Always move in a parallel direction to the treatment electrodes.
II. Make full contact with treatment area before pressing or depressing foot pedal or a patient shock
may result.
III. Check the surface skin temperature frequently with IR external thermometer (goal is 40 degrees
C.).
24
IV. During the procedure, ask the patient if she feels warmth and adjust the energy setting accordingly.
The energy setting is a guideline only (energy may to be higher to maintain 40 degree skin
temperature).
1) Touch front of AVATARs body on LCD touch screen. 2) Press ABDOMEN key 3) Adjust Focus Depth to
the deepest dermal layer. 4) The body handpiece is pre-set on screen, but use of the quad-polar body
hand piece is optional) 5) Select energy %. 6) Set Audible Treatment Metronome (ATM). 7) Press READY
key to start program. 8) Press foot pedal to activate handpiece.
Program Settings: Start new patients with the lowest setting. Adjust energy and ATM settings according to
the chart on this page:
Quad-Polar Body
Bi-Polar Body
Hand Piece
Hand Piece
Energy level
ATM Setting
Energy level
ATM Setting
40%
30 beats/ minute
40%
50 beats/ minute
60%
40 beats/ minute
60%
60 beats/ minute
80%
50 beats/ minute
80%
Activate Handpiece
Step 1: Cleanse Step 2: Divide abdomen into 8 cm x 8 cm grids, and apply conductive gel one half of
abdomen. Step 3: Move the body handpiece to the first side of the abdomen in a linear 8 cm x 8 cm
sections at a slow constant speed, moving the hand piece in a back and forth motion at a speed set by the
ATM. Step 4: Move the handpiece from outside of the waist to the inside. Step 5: Repeat steps 2 through
4 for the other side of the abdomen.
25
Move the handpiece on the abdomen at a slow constant back and forth motion (8cm x 8cm
sections).
II. Make full contact with treatment area before pressing or depressing foot pedal or a patient shock
may result.
III. Check the surface skin temperature frequently with IR external thermometer (40 degrees C. is
goal).
IV. During the procedure, ask the patient if she feels warmth and adjust the energy setting accordingly.
Energy setting is a guideline only (settings may need to be higher than guideline to achieve 40
degrees C.).
V.
1) Select CUSTOM key. 2) Select the body handpiece button/ plug in quad-polar handpiece/ set time to
20 minutes/ set energy %. 3)
Metronome (ATM). 5) Press READY key to start program. 6) Press foot pedal to activate handpiece.
26
Program Settings: Start new patients with the lowest setting; continue to move energy and ATM settings
according to the
chart on this page:
TRI-Polar Body
Bi-Polar Body
Energy level
ATM Setting
Energy level
ATM Setting
40%
30 beats/ minute
40%
40 beats/ minute
60%
40 beats/ minute
60%
60 beats/ minute
80%
50 beats/ minute
80%
Press Foot
Pedal to
Activate
Handpiece
Step 1: Have patient lay on her stomach, Cleanse treatment area. Step 2: Use washable skin marker to
mark off 8 cm x 8 cm square grids. Step 3: Apply conductive gel to first area. Step 4: Apply the handpiece
to the treatment area and move it using a slow back and forth or circular motion, covering the entire
treatment area until the target skin
temperature of 40 degrees C. is
reached. Step 5: Re set timer to 20
minutes and repeat steps 1-4 in each
remaining treatment area.
Time: 40-80 minutes total (20 minutes per area/total time depends on the number of 8 x 8 cm grids
treated).
Treatment Tips:
I.
Move the handpiece slowly on the thigh and buttocks at a slow constant circular or back and
forth motion, in 8 cm x 8 cm quadrants.
II.
Make full contact with skin before pressing or depressing foot pedal or a patient shock may
result.
III.
Check the surface skin temperature frequently with IR external thermometer (each quadrant
must reach the target temperature of 40 degrees C. and maintain for a full minute).
IV.
During the procedure, ask the patient if she feels warmth and adjust the energy setting
accordingly. The energy setting is a guideline only (the setting may need to be higher to
achieve 40 degrees C.).
27
1) Touch the back of the AVATARs body on LCD touch screen. 2) Press the BACK key. 3) Adjust
Focus Depth to the deepest dermal layer. 4) The body handpiece is pre-set, but use of the
quad-polar handpiece is highly recommended.. 5) Select energy %. 6) Set Audible Treatment
Metronome (ATM). 7) Press READY key to start program. 8) Press foot pedal to activate
handpiece.
Program Settings: Start new patients with the lowest setting.
Quad-Polar Body
Bi-Polar Body
Energy level
ATM Setting
Energy level
ATM Setting
40%
30 beats/ minute
40%
40 beats/ minute
60%
40 beats/ minute
60%
60 beats/ minute
80%
50 beats/ minute
80%
Step 1: Cleanse. Step 2: Divide area of treatment into 8 cm x 8 cm grids. Step 3. Apply conductive gel to
the first grid area (start at the lower back center of one side of back). Step 4: Apply hand piece to the body
and move out to the sides and back to the center at a slow constant speed. Move the handpiece in a linear
motion and treat entire 8cm x 8cm section (do NOT treat directly on the spine). Step 5: Move upward
towards the shoulders and treat additional sections in the same manner. Step 6: Repeat steps 3 through 5
on the other side of the back.
Time: 40- 60 minutes total (20 minutes per 8 cm x 8 cm area). Total time depends on size of area treated.
28
Treatment Tips:
I.
Move the handpiece slowly on the back at a slow constant circular or back and forth motion.
Be sure to treat each entire 8 cm x 8 cm section.
II.
Make full contact with treatment area before depressing foot pedal or a patient shock may
result.
III.
Check the surface skin temperature frequently with IR external thermometer (objective is to
reach and maintain an external skin temperature of 40 degrees C. for one full minute in each
area).
IV.
During the procedure, ask the patient if she feels warmth and adjust the energy setting
accordingly. The energy setting is a guideline only. The setting may need to be higher than the
guidelines to achieve target skin temperature of 40 degrees C.
Program Settings: Start new patient with the lowest settings. Adjust energy and ATM settings according
to the chart on this page:
Bi-Polar Body
Quad-Polar Body
Press Foot Pedal to
Activate Handpiece
Energy level
ATM Setting
Energy level
ATM Setting
40%
40 beats/ minute
40%
60 beats/ minute
60%
55 beats/ minute
60%
80 beats/ minute
Step 1: Cleanse Step 2: Use washable marker to mark off 8 cm x 8 cm grids on treatment area and apply
conductive gel to first area.
Step 3: Apply hand piece to the first treatment area and move from the center out to the sides at a slow
constant speed. Move the handpiece in a slow constant circular or back and forth motion throughout the
entire treatment grid area until the skin surface reaches 40 degrees C.. Maintain this temperature for one
29
full minute. Step 4: Repeat steps 2 and 3 for the other treatment grids until the entire buttocks area has
been treated.
Time: 30 to 60 minutes totaldepending on total size of area being treated (15 minutes per grid area).
Treatment Tips:
I.
Move the body handpiece slowly throughout entire 8 cm x 8 cm grid area in a slow constant
back and forth motion until goal temperature is reached.
II.
Make full contact with treatment area before pressing or depressing foot pedal or a patient
shock may result.
III.
Check the surface skin temperature frequently with IR external thermometer (goal is 40
degrees C.).
IV.
During the procedure ask the patient if she feels warmth and adjust the energy setting
accordingly. The energy setting may need to be higher than the guideline to reach and
maintain the 40 degrees C. goal temperature.
1) Touch AVATARs arm, 2) Press the ARM key, 3) Adjust Focus Depth to deepest level.
4) Select handpiece (facial or body handpiece can be used). 5) Select energy %
6) Set Audible Treatment Metronome (ATM), 7) Press READY key to start program 8) Press foot switch
to start emission of energy.
Program Settings: Start new patients with the lowest energy setting recommended; continue to adjust
energy and ATM settings according to the
chart on this page:
Press Foot Pedal to
Activate Handpiece
ATM Setting
50%
50 beats/ minute
60%
55 beats/ minute
Step 1: Cleanse
30
Step 2: Spread conductive gel to the back of one arm area and treat.
Step 3: Move the handpiece in a back and forth linear motion from elbow to the upper arm on the back and
front of arm in 8 cm x 8cm sections until each section reaches goal temperature of 40 degrees C. Step 4:
Repeat steps 2 and 3 on the other arm. Step 5: Repeat steps 2 and 3 on the lower portions of the arms
(below elbow) if necessary.
Time: 30 45 minutes total, 15 minutes per area (total time required depends on size of total area treated).
Treatment Tips:
I.
Move the handpiece slowly using circular or back and forth motions being sure to treat
entirety of each 8cm x 8cm section.
II. Make full contact with treatment area before depressing foot pedal or a patient shock
may result.
III. Check the surface skin temperature frequently with IR external thermometer (goal is to
maintain 40 degrees C. temperature for one full minute).
IV. During the procedure, ask the patient if she feels warmth and adjust the energy setting
accordingly. The recommended energy setting is a guideline only (higher energy may be
needed to reach 40 degrees C.).
1) Touch AVATARs Arm 2) Press the HAND key 3) Adjust Focus Depth to epidermal/ dermal junction.
4) Select mini probe handpiece and select energy % 5) Set Audible Treatment Metronome (ATM)
6) Press READY key to start program 7) Press foot switch to start emission of energy.
Programs Settings: Start new patients with the lowest setting. Adjust energy and ATM settings
according to the chart on this page:
31
Energy level
ATM Setting
60%
50 beats/ minute
Activate Handpiece
Step 3: Move the mini probe handpiece from the finger base to the tip of the finger of each hand at a
constant speed according to the ATM and continue the application of RF energy until the skin reaches goal
temperature of 40 degrees C.
Time: 20 minutes total (10 minutes per hand).
Treatment Tips:
I.
Move the handpiece slowly on the hand and fingers at a slow constant
back and forth motion.
II.
Make full contact with treatment area before depressing foot switch or a
patient shock may result.
III.
IV.
During the procedure, ask the patient if she feels warmth and adjust the
energy setting accordingly. The recommended energy settings are
guidelines only (the actual energy level used may need to be higher
than guideline to reach 40 degrees C.).
32
1) Touch AVATARs Leg. 2) Press the LEG key. 3) Adjust Focus Depth to deepest layer
4) Body Handpiece selected. Plug in quad-polar handpiece. 5) Select energy %
6) Set Audible Treatment Metronome (ATM) 7) Press READY key to start program 8) Depress foot
switch to begin emission of energy.
Program Settings: Start new patients with the lowest recommended setting. Continue to adjust energy and
ATM settings according to the chart on this page:
Quad-Polar Body Hand Piece
Energy level
ATM Setting
50%
50 beats/ minute
60%
55 beats/ minute
80%
70 beats/ minute
bottom of legs).
Step 3: Starting at the bottom of the leg, move the handpiece in back and forth vertical movements up the
back of the leg toward the knee at a constant speed until the skin temperature of 40 degrees C. is reached.
Step 4: Repeat steps 2 and 3 for other sections of legs. For upper legs, turn patient over and repeat steps 2
and 3 on front of upper legs.
Time: 40 60 minutes total (depending on size of area treated).
Treatment Tips:
Move the body handpiece at a slow, constant speed throughout each area.
Do NOT treat area behind the knee (popliteal).
33
Make full contact with treatment area before pressing or depressing foot pedal or a patient
shock may result.
Check the surface skin temperature frequently with IR external thermometer (goal is to reach
and maintain skin temperature of 40 degrees C.).
During the procedure, ask the patient if she feels warmth and adjust the energy setting
accordingly. The energy setting recommendation is a guideline only (actual settings may
need to be higher to reach 40 degrees C.).
5.4
Post procedure: Cleanse face with gentle cleanser to remove any remaining conductive gel
and grid markings. Follow with application of moisturizer and sunscreen.
Be sure to take post treatment photographs (using the same 5 view as pre photos),
and schedule I month, 3 month, & 6 months follow up photo sessions and
evaluations.
Post-operative care instructions: Moisturizer with sunscreen as aftercare and may use cold
compresses (i.e., ice packs) if experiencing discomfort.
FACE
BODY
Recommended No. of
Treatments
4 to 6
8 to 10
Recommended Treatment
Interval
1 Per Week
1 Per Week
Recommend that your patients stay well hydrated during the treatment program.
(Drink minimum of 8 glasses (8 fl. oz.) of water per day)
Program Settings: Start new patients with the lowest recommended parameter settings;
continue to adjust energy and ATM settings according to the chart (shown above) on
each page
34
During the procedure, ask about the patients comfort and adjust the energy setting to the
highest level that is still comfortable for the patient.
Use the Audible Treatment Metronome (ATM) to help determine the proper speed you
should move the handpiece (optional).
Skin surface temperature levels must be checked frequently to ensure the desired
temperature is achieved. Repeated motions in the same area (about 10 times) on the
skin surface should reach the clinical end point of 40 degrees C. Follow the ATM
guidelines for each indication.
Hand piece should be applied with gentle pressure. Always keep the handpiece moving
at a slow constant speed. Never stop on a treatment area with foot switch depressed or
discomfort may result. Always remove the entire handpiece from the skin in a single
motion or a shock may result (release foot switch first, then remove hand piece).
Move the handpiece in small circular or back and forth motions at the recommended
speed with the assistance of the ATM.
The skin around the eyes and forehead are more sensitive therefore, energy levels
around the eyes and forehead may need to be turned down to a lower level than on other
areas of the face.
After the treatment, patients can immediately resume their normal daily activities. There is no
down time or recuperation period required, and no need for follow up visits except for the next
TRIO XT treatment which should be scheduled approximately one week after the previous
treatment. Patients report that there is no post-treatment pain or discomfort, and that the skin
surface is not disrupted or bruised as a result of the treatment.
35
Treatment
Total
treatment
time in
minutes
Cooling
Setting
Energy
Handpiece
40-60%
Bi Facial
**
***
40-60%
Bi Facial
Peri-Orbital(eye)
10
No
40-60%
Bi Mini
Peri-Oral Mouth
No
40-60%
Bi Mini
20
**
40-60%
Bi Facial
Full Face
40
Forehead
Neck
TRI or Bi
Abdomen
40
No
40-80%
Body
36
TRI or Bi
Back
40
No
40-80%
Body
Arms
30
No
50-60%
TRI Body
Hands
20
No
60%
Bi Mini
Legs
40
No
50-80%
TRI Body
TRI or Bi
Buttocks
30
No
40%-60%
Body
37