AOAC Official Method 2000.18
Fat Content of Raw and Pasteurized Whole Milk
Gerber Method by Weight (Method I)
First Action 2000
Table 2000.18A.
Graduated volume in flat tube at 20EC is 1.000 mL (13.546 g Hg), divided into 80 equal parts, each equivalent in volume to 0.0125 mL.
Error of total calibrated length is not to exceed of smallest graduation. Graduation lines are uniformly centered on flat tube face, with
0.1% lines not less than 3 mm long, the 0.5% lines extending 1 mm
on each side beyond the 0.1% lines, and the 1.0% lines extending not
less than 1 mm on each side beyond the 0.5% lines or extending
across the entire flat face. Each whole percent graduation is identified by pigmented numbers running serially from 0 to 8, at the right
and just above the whole-percent lines, with the zero measurement
nearest the large bulb. Bottles are identified Milk, with the name
or symbol of the manufacturer or distributor permanently inscribed
on the body and a symbol applied after recalibration where required
to attest conformance with regulatory specifications.
(5) Testing.Accuracy of each bottle should be determined before use (usually by manufacturer or calibration laboratory). Mercury is added to the butyrometer by means of a special volumetric
apparatus. Testing is conducted at ambient temperature at 20 2EC.
The small bulb is opened to prevent trapped air bubbles. The apparatus has 2 volumetric settings. The first setting delivers a variable volume that permits filling mercury into the butyrometer until the
meniscus reaches the 0 graduation line. The second setting dispenses
1.000 mL mercury, corresponding to 8% on the graduated scale. The
difference between the expected and observed scale readings must
not exceed division of a graduation line ( 0.05% fat). The accuracy of the testing apparatus is regularly verified with officially calibrated glassware.
(6) Lock stopper and key.Lock stopper is of reagent-resisting
rubber, of standardized dimensions to fit in neck of butyrometer,
molded to provide a seat for the ball or plug, and with a channel for
the key. Rim and channel are reinforced by metal, plastic, or other
firm binding insert. Key is of noncorrodible material suitable for inserting into the stopper.
(b) Butyrometer support rack.To hold butyrometers in a secure, vertical position.
(c) Dispensers.To deliver 10 mL H2SO4 and 1 mL isoamyl alcohol.
(d) Centrifuge.For butyrometers, preferably provided with a
speed indicator to assess 1150 70 rpm. If not equipped with speed
indicator, check operation speed of Gerber centrifuges under full
load regularly with a tachometer. Design of the centrifuge shall ensure that temperature of butyrometer contents after centrifuging is
between 30 and 50EC.
When loaded, the centrifuge must produce, within 2 min, a relative centrifugal acceleration of 350 50 g at the outer end of the
butyrometer stopper. This acceleration is produced by centrifuges
with the following effective radius (horizontal distance between the
Interlaboratory study results for percent fat in milk (g/100 g) determined by Gerber by weight methoda
Sample ID
Mean
No. of labs
sr
sR
RSDr, %
RSDR, %
Raw
4.431
1011
0.023
0.053
0.52
1.20
0.065
0.148
Pasteurized, homogenized
3.677
1011
0.029
0.040
0.78
1.08
0.080
0.111
Mean based on analysis of 7 raw whole milk and 4 pasteurized homogenized whole milk, respectively. Standard deviations (sr, sR) and prediction values (r, R) based
on analyses of 8 raw milks (2 in blind duplicates) and 8 pasteurized homogenized milks (2 in blind duplicates), for a total of 200 tests. Of the total of 200 tests, statistical
outliers were identified for 5 raw milk tests and 5 pasteurized homogenized milk tests. Relative standard deviations (RSDr, RSDR) were calculated as (sr/mean) 100
and (sR/mean) 100, respectively.
Figure 2000.18.
Table 2000.18B.
Centrifuge speed
Effective radius, mm
240
1140
250
1120
260
1100
270
1080
300
1020
center of the centrifuge spindle and the outer end of the butyrometer
stopper) operated at the speed indicated (Table 2000.18B).
(e) Water bath for test butyrometers.With thermometer, device to maintain temperature of fat column at 6063EC, and providing intermittent or constant agitation. Depth of water bath must
permit vertical immersion of butyrometer to terminal (small) bulb.
(f) Water bath for tempering milk samples prior to weighing.With thermometer and device to maintain temperature of milk
at 39 1EC.
(g) Analytical balance.Weighing to nearest 0.01 g. Check accuracy periodically and whenever balance is moved or cleaned.
Keep record of balance calibration checks.
(h) Calibration weights.Class S, standard calibration weights
to verify balance accuracy within weight range to be used for weighing empty milk butyrometers and butyrometers containing test portions.
(i) Reading light.As background when measuring fat columns.
Light should be diffused (soft white) and provide illumination from
angles above and below level of fat column and at eye level. Magnification device to aid reading is useful.
D. Determination