Professional Documents
Culture Documents
0197
REP
MDSS GmbH
Schiffgraben 41
30175 Hannover,
GERMANY
0602-50001-007
Table of Contents
Pages
2
Key Features
.....................................................................................................................................................................................................
2
Expectations of Use
.....................................................................................................................................................................................................
2
General Cautionary Notice
.....................................................................................................................................................................................................
2
Battery Precautionary Note
.....................................................................................................................................................................................................
2
Disposal of Lithium lon Battery
.....................................................................................................................................................................................................
3
User Assistance
.....................................................................................................................................................................................................
3
Identifying Parts and Controls
.....................................................................................................................................................................................................
3
Control symbols and their functions
.....................................................................................................................................................................................................
4
Battery Installation
.....................................................................................................................................................................................................
4
Instruction for Use
Turning the ViScope ON
Initial screen after turning on the ViScope
Screen icons and symbols with descriptions
.....................................................................................................................................................................................................
Headset Positioning
Turning the ViScope ON
5
6
.....................................................................................................................................................................................................
Auscultating on Different Patients
.....................................................................................................................................................................................................
6
7
.....................................................................................................................................................................................................
To Transfer Data to a Computer
System Requirements for the ViScope software
To Record Heart Sound Waveforms in the ViScope/PC
To View the Saved Waveforms
.....................................................................................................................................................................................................
General Instructions
9
Cleaning Instruction
9
Diaphragm Replacement
10
Storage
.....................................................................................................................................................................................................
10
Accessories
.....................................................................................................................................................................................................
10
Technical Specifications
.....................................................................................................................................................................................................
11
International Regulatory Standards
.....................................................................................................................................................................................................
11
Trouble shooting
.....................................................................................................................................................................................................
.....................................................................................................................................................................................................
12
Error codes, causes and solutions
12
User Assurance Tests
14
Auscultation Points
.....................................................................................................................................................................................................
14
Auscultation Points for Specific Abnormality
.....................................................................................................................................................................................................
15
Indicative Waveforms of Different Anomalies
.....................................................................................................................................................................................................
.....................................................................................................................................................................................................
Congratulations and thank you for choosing the ViScope Visual Stethoscope empowering
medical professions with dynamic auscultation the ability to see what you hear
Manufactured by HD Medical (HDM), ViScope delivers the following powerful features:
...............................................................................................................................
Key Features
Integrated visual display showing the hearts activity in real-time
Easily switch between Bell, Diaphragm and Wide modes for audio
Adjustable Audio with multiple levels of amplification
Screen freeze options to capture, save and transfer heart sound
waveforms - ViScope MD
Rechargeable Li-ion battery for continuous use
Expectations of Use
ViScope uses a custom made Li-ion battery and charger. Using any other battery and/or
charger will void the warranty. Observe proper precautions when
handling batteries. Use only batteries approved by HD Medical.
Do not expose the battery to a flame or excessive heat, immerse in or expose to
water, short the terminals or disassemble the battery as doing so could
damage the battery cause fire, injury or environmental contamination.
Liquid leaking from the battery can cause skin burns or damage the ViScope.
If the battery leaks inside the instrument, return the ViScope to the nearest
service center.
Remove the battery during shipment or if the device is going to be stored or
unused for an extended period of time
For disposal of the battery, follow the procedures as stipulated by the local
governing authority.
User Assistance
If you experience any problem with the ViScope Visual Stethoscope, do not attempt to
repair it yourself. Please notify:
HD Medical
India
USA
Email:
website:
service@hdmedicalgroup.com
www.hdmedicalgroup.com
.................................................................................................................................................
2
3
4
7
8
5
6
1
2
3
4
5
6
7
8
9
LCD
Battery charging status indicator
Audio toggle key
Freeze key
Power key
Chestpiece / Diaphragm
Charger connection slot (mini USB)
3 - Way key
Only for ViScope MD
.................................................................................................................................................
10
11
10 Reset key
11 Battery Screw
Function
Battery Installation
Be sure the battery is
inserted correctly into the
battery compartment and
has been charged for 5 hours
before using the ViScope
Visual Stethoscope for the
first time.
Volume
Battery
Status
Icons &
Symbols
will be displayed
Headset Positioning
The headset of the ViScope is specifically designed to be comfortable to wear and use.
The eartubes are set at an angle to fit the anatomy of the typical ear canal.
The ear tips should be pointing forward as you insert them into your ear canal.
The tension of the spring in the headset can be reduced by holding each eartube at the
bend and gradually pulling apart until they are fully extended.
The tension of the spring in the headset can be increased by holding the headset
where the eartubes join the plastic tubing and squeezed until one eartube touches the
other. Repeat if necessary.
A-1
A-2
A -3
A-4
A-5
Press and hold the power key for 3 seconds to turn the device ON
Note:
2.
3.
4.
5.
6.
7.
7.1
7.2
8.
Note:
It is a requirement that the patient should not breathe heavily for 3-4 seconds during auscultation.
Even when the display is in the freeze mode, the audio can still be heard.
Place the chest piece as you would normally, at one of the appropriate positions
to listen to heart sounds (refer to Auscultation Points on page 14)
2.
Once you have a good pickup either when listening, or a visual display, press and
release the Record/Freeze/Unfreeze key.
3.
Hold the device in that position till the screen changes (should be about 10
seconds) to MURMUR / NO MURMUR / PLEASE RECORD AGAIN.
4.
5.
If the recording signal is not suitable for analysis, it will state PLEASE RECORD
AGAIN, in which case repeat step 2.
6.
Once the 10 seconds recording time is up, you can take the device off the patients
chest. The device will complete the analysis, which once done will display a message
on screen. If it states MURMUR-PLEASE CONFIRM WITH CLINICAL
FINDINGS then, the device has detected a possible anomaly; otherwise it will state
NO MURMUR.
7.
3. The number appearing between the arrows indicates the waveform that will
be saved
4. Using the 3 - way key, scroll to the transfer icon and press the 3 - way key to
transfer the waveform to the PC/Laptop
5.
6.
3 The number shown between the arrows is the memory location where the wave
form is saved
4
Each press of 3 - way key reveals the stored waveforms in ascending or descending
order depending on which arrow is selected
5 When ready, press the freeze/unfreeze key and the device is ready to continue
auscultating
General Instructions
Recharging Batteries
1 You may operate the ViScope uninterrupted as long as the battery icon shows at
least one bar. Once the icon has no bars and starts blinking, charge the scope
2 Before charging the ViScope switch it off. Using the mini USB cable Insert the
connector into the USB slot in the ViScope and plug it into an AC power outlet.
The battery charge indicator will glow RED
3 The ViScope takes about 4 to 5 hours to get completely charged, which can be
observed by the GREEN glow of the battery charge indicator
4 Once the battery power is completely drained, the scope will turn completely off.
Should this happen, charge the ViScope as described above, and restart by
pressing and holding the power key for 3 seconds
HD MEDICAL RECOMMENDS YOU DO NOT USE THE VISCOPE FOR AUSCULTATION
WHEN THE BATTERY IS BEING CHARGED. ATTEMPTING TO USE THE SCOPE WHILE
CHARGING MAY NOT BE SAFE.
CAUTION USE ONLY HDM SUPPLIED / AUTHORIZED BATTERY CHARGERS FOR CHARGING THE
BATTERY.
Cleaning Instruction
You may be using your ViScope on a number of different patients each day.
Cleaning the chestpiece and the body of the device is imperative since failure to do so
may cause infections or allergies to be transferred to susceptible patients.
1.
2.
3.
Ideally the chest piece diaphragm should be cleaned after each use while the
body of the ViScope should be cleaned at the end of each day
Use only a soft clean cloth or alcohol wipe to clean
Do not immerse the ViScope in water or any other form of liquid sterilization
for cleaning
Diaphragm Replacement
Storage
1. Always store the device in a cool, dry place
2.
It is advisable to place the device on a soft surface to avoid damage to the device
in general and to the chestpiece in particular
Accessories
USE OF ACCESSORIES NOT AUTHORISED BY HDM MAY DAMAGE THE UNIT
CAUTION
1.
2.
Earplugs (VSC060-003)
a. Use only HDM earplugs.
b. Use of non-HDM earplugs can result in pain in ears due to material
hardness and may degrade audio.
3.
Battery (VSC060-004)
a. Rechargeable Li-Ion 3.7V, 1800 mAh
4.
Technical Specifications
Display
Electrical
Environmental
Operating temperatures: 5 C to 40 C
10
Basic
Standard
Regulatory
Acceptable Limits
ViScope
Capability
Radiated
Emission
30 230MHz,
30dBV/m
30 230MHz
25.18dBV/m
(Max)
230 1000MHz
Harmonics
Current
Emission
IEC 61000-3-2:2005
230 1000MHz,
37dBV/m
Flicker
Emission
IEC 61000-3-3:2005
Passed and No
malfunction observed
Electrostatic
Discharge
IEC 61000-4-2
Passed and No
malfunction observed
Radiated
Susceptibility
IEC 61000-4-3:2006
Contact discharge
up to 6KV
Air discharge
up to 8KV
Passed and No
malfunction observed
Electrical
Fast Transient
IEC 61000-4-4:2004
80MHz 1000MHz
3V/m
1-2.5GHz 3V/m
Passed and No
malfunction observed
High Energy
Surge
IEC 61000-4-5:2005
Passed and No
malfunction observed
Conducted RF
IEC 61000-4-6:2004
Pulse amplitude up to
a 1KV Differential Line to
Neutral
Passed and No
malfunction observed
Power Frequency
Magnetic Field
IEC 61000-4-8:2001
Passed and No
malfunction observed
IEC 61000-4-11:2004
Short
interruption:
0%
of
AC mains for 5 sec
Voltage
dips:
40%
of
AC mains for 5 cycles
70%
of
MC
mains
for
25
cycles
0%
of
AC
mains
voltage for 0.5 cycles
Passed and No
malfunction observed
Trouble shooting
ViScope does not turn on when the power key is pressed.
1. The battery could be completely drained. Charge the battery
2. If this does not work, it is possible that the battery is faulty;
Replace with a HDM approved battery available directly from HDM or an
HDM authorized reseller
In case the ViScope is subjected to large mechanical shock, it is possible that the
LCD may be damaged internally though it may not be readily visible
Please contact HDM or an HDM authorized reseller
Unexpected behavior
1.
11
Error Code
Possible Reason
Solution
Key Stuck
ViScope will be turned off
RAM failure
Contact HDM
FLASH failure
Contact HDM
LOW BATTERY
ViScope will be turned off
Battery is completely
drained and should be
recharged
12
13
Expected Outcome: If the left arrow is scrolled to and highlighted then the number
between the arrows will change in a descending order from 4 to 1 each time the 3 - way
key is pressed
If the right arrow is scrolled to and highlighted then the number between the arrows
changes in the ascending order from 1 to 4 each time the 3 - way key is pressed
When the transfer icon is scrolled to and highlighted and the 3 - way key is pressed, the
stored 4 (four) waveforms in the ViScope will be transferred to the computer that is
connected via the mini USB port
See More Patients and Improve More Lives
HD Medical promotes personal and community health through exceptional healthcare
diagnostic solutions, reinforcing its commitment to delivering Higher Dimensions in
Healthcare. The companys solutions help populations gain access to healthcare
through more effective and rapid screening, while reducing personal healthcare costs
by eliminating unneeded tests such as echocardiograms. With HD Medicals products,
healthcare professionals see more patients and improve more lives.
Auscultation Points
Aortic
area
Pulmonic
area
2
3
4
5
Tricuspid
area
Mitral
area
Abnormality:
Aortic Stenosis S2 is loudest here
Location:
2nd ICS R
Sternal border
Location:
2nd ICS L
Sternal border
Aortic valve
3
4
Tricuspid valve
Pulmonic valve
Mitral valve
Location:
L Lower
Sternal border
Location:
5th ICS
Abnormality:
Mitral stenosis or regurgitation
S1 is loudest here
Abnormality:
Tricuspid
14
Normal
Aortic Regurgitation
Mitral Stenosis
Pulmonary Stenosis
S3
S4
Tricuspid Regurgitation
15