Professional Documents
Culture Documents
I.
INTRODUCTION
II.
EXPERIMENTAL
A. Reagents
The following reagents were used: Lix 65N and Lix 70
(Henkel Corporation, lot no. 82B265010 and 8J16029,
respectively), lauric acid (Eastman Kodak Co.), and dinonylnaphthalene sulfonic acid (Pfaltz and Bauer Inc.).
Other chemicals used were Kermac 470B (Triangle
Refineries Co.), Escaid 200 and Aromatic 150 (Exxon
Co.), and N i S O a ' 6 H 2 0 and KNO3 (analytical reagent
grade).
VOLUME 23B, FEBRUARY 1992--13
of 30 mg Ni2+/1 and 1 M KNO3. The standard temperature was 25 ~ and the stirring speed was 2000 min -1.
III.
B. Effect of pH
Figures 3 and 4 illustrate the effects of pH on nickel
extraction by the Lix 65N and Lix 70 systems, respectively, both with and without ultrasound. The pH represents initial values, ranging from 4 to 7. The extraction
rate increased with the increase of initial pH until pH =
6.02, above which pH did not have further effect.
The apparent forward reaction rate constant (kf) was
calculated from the slope of the line in the plot [(Co Ce)/Co] In [(Co - C~)/(C, - C~)] vs time, the method
Fig. 1 - - S c h e m a t i c representation of the experimental apparatus for solvent extraction with ultrasound.
14--VOLUME 23B, FEBRUARY 1992
METALLURGICAL TRANSACTIONS B
100 I
100
RPM
I
/
[] 9 1000
o 9 1500
o 9 2000
* * with
& 9 2500
LIX70
ultrasound
(at 2000 rpm)
8 0 -1
I@I 8O
Iii
}- 6O
r
x~ 40
I-- 4 O
X
ILl
9--
U.I
20
"
no ultrasound
with ultrasound
pH
kf
pH
kfund
20
o 4.02
zx 5.05
[] 6.02
o 7.03
solid, I.IxT0 s y s t e m
0
0
20
40
60
80
100
120
140
160
180
20
40
TIME, rain
developed by Flett et al. ,[12] where C = the nickel concentration in the aqueous phase, inferior t = the extraction time, inferior 0 = the initial state, and inferior e =
the equilibrium state. The solvent extraction rate constants, kl, for each extraction system, are presented in
tables in corresponding figures (Figures 3 and 4).
A comparison of the rates of conventional and ultrasound extractions shows that ultrasound increases the extraction rate of nickel by as much as 4 to 7 times. The
extraction rates in the Lix 65N system were faster than
those in the Lix 70 system, which required about 1 hour
to reach equilibrium even with ultrasound. However, the
uptake of nickel was higher with the Lix 70 system than
with the Lix 65 system. The slower kinetic performance
of the Lix 70 system was caused by steric effects. The
C. Effect of Temperature
The effect of temperature on the solvent extraction of
nickel was studied in the range of 25 ~ to 50 ~ Both
extraction systems, Lix 65N (Figure 5) and Lix 70
(Figure 6) were examined. According to the results presented in Figures 5 and 6, temperature had a positive
60 1
LIX65N
LIX65N
40
40
LU
!-- 30
O
0
I--" 20
X
till
with ultrasound
p.
k,
-i/7,/?'
o 402
322
~ff
zx 5.05
[] 6.02
o 7.02
3.46
3.79
4.27
10
7.83
10.1
12.0Z0
13.0
80
9 4.02
9 5.05
9 6.02
9 7.03
50
t
60
1.43
2.67
2.85
2.79
TIME, rain
~4~
zx
20
40
60
9
9
9
9
80
pH
4.02
5.05
6.02
7.02
100
kf
13.7
16.5
19.8
33.4
20
nd
It//6
I~r
~/
,LF"
o 25
zx 40
m 50
O~
120
TIME, rain
Fig. 3 - - E f f e c t of pH on nickel extraction with Lix 65N-lauric
acid-Kermac 470B. Organic phase: 10 pct (v/v) Lix 65N; 0.1 M lauric
acid; Kermac 470B. Aqueous phase: 30 rng/l Ni-~+; 1 M KNO3. Ratio
of O / A : 100 m l / 1 0 0 ml; T = 25 ~ stirring rate: 2000 rpm; ultrasound: 20 KHz, 47W.
METALLURGICAL TRANSACTIONS B
tU
20
c/
k,
3.46
5.78
13.70
rl~
9 25
9 40
9 50
k,
16.5
27.1
37.1
40
60
80
100
TIME, rnin
Fig. 5 - - T e m p e r a t u r e effect on nickel extraction with Lix 65N-lauric
acid-Kermac 470B. Organic phase: 10 pct (v/v) Lix 65N; 0.1 M lauric
acid; Kermac 470]3. Aqueous phase: 30 m g / l Ni2+; 1 M KNO3; pH~.~,
= 5.05. Ratio of O / A : 100 m l / 1 0 0 ml; stirring rate = 2000 rpm;
ultrasound: 20 KHz, 47 W.
VOLUME 23B, FEBRUARY 1992-- 15
100
1~
trolled b y the rate o f dehydration. Ultrasound can override the contribution o f thermal energy to d e h y d r a t i o n
and the diffusion o f gallate ions. O n the other h a n d , the
rate at which water is r e m o v e d from the shell o f m o l e cules surrounding the nickel ion is several thousand times
faster, but this step is unimportant during the solvent extraction o f nickel. T h e intermediate c o m p l e x o f nickel
and c a r b o x y l i c acid reacts slowly with the m o l e c u l e o f
h y d r o x y o x i m e , and this reaction rate controls the overall
solvent extraction r a t e J 12]
The overall reaction of nickel extraction can be represented b y Eq. [1]:
80
60
40
LII
@/
Ol r
0
o 25
t, 40
t3 50
2.67
7.92
13.30
9 25
9 40
9 50
10.1
23.6
49,3
20
40
60
80
100
120
140
Ni 2~ + 2HOXorg
160
+ 2RHorg
TIME, rain
Fig. 6--Temperature effect on nickel extraction with Lix 70-1auric
acid-Kermac 470B. Organic phase: 10 pct (v/v) Lix 70; 0.1 M lauric
acid; Kermac 470B. Aqueous phase: 30 mg/l Ni2 1 M KNO3; pH~,
= 5.05. Ratio of O/A: 100 ml/100 ml; stirring rate = 2000 rpm;
ultrasound: 20 KHz, 47 W.
K q NiOX2(RH)2org + 2H +
[1]
[2]
where [NiOX2(RH)2]org is the concentration o f nickel in
the organic phase, [HOX]o~g is the concentration o f hyd r o x y o x i m e , and [RH]o,g is the concentration o f lauric
acid.
The value o f K~., can be calculated from the m e a s u r e d
pH values and concentrations o f nickel in the organic
and aqueous phases. The concentrations o f o x i m e s and
lauric acid were a s s u m e d to be constant, b e c a u s e they
were much higher than the concentration o f nickel. The
equilibrium constants as a function o f t e m p e r a t u r e are
given in T a b l e I.
The Rex values in Table I were used to calculate the
enthalpy c h a n g e o f nickel extraction by the V a n ' t H o f f ' s
equation, In Kex = -AH~
+ C. T a b l e II shows the
calculated A H ~ values o f nickel extraction.
The positive enthalpy changes characterized the nickel
extraction reactions with the h y d r o x y o x i m e extractants
as endothermic.
D. Effect of Diluents
K e r m a c 470B was used as a standard diluent in this
study. Because it contains about 17.3 pct aromatic c o m pounds, the contribution o f such c o m p o u n d s to the solvent extraction o f nickel was e x a m i n e d . The use o f
toluene, a representative aromatic c o m p o u n d , as a diluent during the solvent extraction o f nickel with Lix 70
Lix 65N/lauric
acid/Kermac 470B
Temperature
(~
Without
Ultrasound
With
Ultrasound
25
40
50
1.16 10 -5
2.40 x 10 -5
5.53 t0 -5
1.12 x 10 -s
3.03 x 10 -5
5.47 10 -5
Lix 70/lauric
acid/Kermac 470B
Without
Ultrasound
2.73 10 -5
6.63 x 10 -5
1.05 x 10 -4
With
Ultrasound
3.44 x 10 5
8.47 x 10 -5
1.10 x 10 -4
METALLURGICALTRANSACTIONSB
0.6
. 0.5
&H, KJ/mole
Lix 65N/lauric
acid/Kermac 470B
Without
Ultrasound
48.8
With
Ultrasound
50.8
open: No ultrasound
solid: with ultrasound
Lix 70/lauric
acid/Kennac 470B
Without
With
Ultrasound
Ultrasound
43.4
38.1
Lix70
tit
O3
-1- 0.4
o
z
0.3
0.2
decreased extraction rates (Figure 7). The uptake, however, was significantly higher. For example, with ultrasound, the equilibrium uptake of nickel with toluene was
90 pct, as compared to the equilibrium uptake of 63 pct
with Kermac 470B.
The slower extraction rates are explained by the strong
solvency of the aromatic compound. The aromatic diluent also causes the extractant molecule (Lix 70) to be
predominantly inside the droplets of organic phase instead of at the interface. The increased uptake is explained by the increased solubility of the chelated complex
in the aromatic solvent.
0
._= 0.1
65N
,,,
-
Z
O
0.2
0.4
0.6
0.8
1.0
1.2
1.4
1.6
1.8
Figure 8. The open symbols represent the solvent extraction results without ultrasound, and the closed symbols represent the sonochemical extraction results. No
effect of ultrasound on the equilibrium of extraction was
demonstrated.
100
L,
pHina
open: no ultrasound
solid: with ultrasound
o 9 404
8O
LIX70
t, 9 5105
8o
Soo
o
[]
,,
~x 4o
,~ 40
Ill
u.I
~
20 ~
~[
0
60
120
180
240
300
TIME, rain
Fig. 7 - - D i l u e n t effect on solvent extraction of nickel with Lix 70-1auric
acid. Organic phase: 10 pct (v/v) Lix 70; 0.1 M lauric acid. Aqueous
phase: 30 mg/1 Ni 2 1 M KNO3, pHinit = 5.05. Ratio of O / A : 100
m l / 1 0 0 ml; stirring rate = 2000 rpm; ultrasound: 20 KHz, 47 W.
METALLURGICAL TRANSACTIONS B
20
60
120
180
240
TIME, rain
Fig. 9 - - E f f e c t of pH on cobalt extraction with Lix 70-1auric
acid-Kermac 470B. Organic phase: 10 pct (v/v) Lix 70; 0.1 M lauric
acid; Kermac 470B. Aqueous phase: 30 mg/1 Co2+; 1 M KNO3; pHi,,
= 5.05, Ratio of O / A : 100 m l / 1 0 0 ml; T = 25 ~ stirring rate =
2000 rpm; ultrasound: 20KHz, 47 W.
VOLUME 23B, FEBRUARY 1992--17
reach equilibrium even after 3 hours. Ultrasound substantially increased the cobalt extraction rate, so that an
increase in extraction uptake from about 30 pct in the
regular system to about 70 pct with ultrasound was
achieved in 3 hours. However, cobalt stripping from the
organic phase loaded with ultrasound was difficult, especially when loading was performed at a higher initial
pH. Cobalt stripping was possible only with the addition
of a reductant, such as sodium sulfite. In extraction systems without ultrasound, the cobalt(II) extracted with
substituted hydroxyoxime alone can be oxidized to a
Co(Ill) state. The organic phase in which cobalt(II) is
converted to Co(Ill) becomes very difficult to strip. 118)
The addition of a carboxylic acid as a modifier diminishes the oxidation of cobalt in the organic phase. The
interfacial activity of lauric acid slows down the oxidation of cobalt by minimizing the interfacial oxidation
reaction.
During solvent extraction from solutions containing a
mixture of Ni and Co, the presence of nickel suppressed
cobalt extraction. Figure 10 shows that without ultrasound, the equilibrium of cobalt extraction, 12.8 pct, was
quickly achieved (after 10 minutes), while the extraction
of nickel continued throughout the experiment, increasing the separation of nickel from cobalt. With ultrasound, the situation was reversed (Figure 11). Cobalt
extraction continued throughout the experiment, while
nickel extraction reached equilibrium, 61.4 pct, after only
30 minutes. Comparison of the cobalt extraction results
presented in Figures 9 through 11 indicates that without
ultrasound the reaction equilibrium state for cobalt may
not be completely reached and that the obtained plateau
is only the result of an extremely slow reaction. When
ultrasound was added, the extraction of cobalt corresponding to the plateau in Figure 11 continued toward
the true equilibrium position. Unfortunately, the short
duration of the experiment prevented the true equilibrium position of cobalt from being reached. The cobalt
extraction results without ultrasound slowly approach the
corresponding results with ultrasound.
H. Effect of Surfactants
Surfactants play an important role in the kinetics of
solvent extraction of metals. The addition of DNNSA
surfactant, even at a low concentration, increased the extraction rate of nickel in the hydroxyoxime/lauric acid/
Kermac 470B system through enhancement of the interfacial area between the two immiscible phases during
their mixing. Also, the active sulfonic acid molecule has
a preferential position against other extractant molecules
at the interface and can itself take up nickel. [19]
Table III, which compares the effects of DNNSA surfactant and ultrasound in increasing extraction rate by
the formation of microemulsions, shows that ultrasound
had a stronger effect than the addition of DNNSA. In
general, oil and water systems require an appropriate
surfactant and cosurfactant to form an emulsion.
Bauer and Komornicki [2~ used sodium laurylsulfate and
n-pentanol instead of D N N S A to produce a microemulsion in the Lix 70/Versatic 911/kerosene system, obtaining a very fast nickel extraction kinetic. These
surfactants were not compared with the ultrasound in this
study.
18--VOLUME 23B, FEBRUARY 1992
L,X
100
u.I
I0
<
n"
IX
U.I
80
10
r./)
i'n
"lg
z
-irl
60
--I
0
40
20
Z
B.
3
0
o
0
20
40
60
80
100
120
140
160
TIME. rain
Fig. 1 0 - - S o l v e n t extraction of Ni and Co with Lix 70-1auric
acid-Kermac 470B in the absence of ultrasound. Organic phase: 10 pct
(v/v) Lix 70; 0.1 M lauric acid; Kermac 470B. Aqueous phase: 30
m g / l Co 2 and 30 mg/1 Ni2"; 1 M KNO3; pHio~, = 5.05. Ratio of
O / A = 100 m l / 1 0 0 ml, T = 25 ~
stirring rate = 2000 rpm.
SF = {[Ni]0/[Ni]a}/{[Co]o/[Co]a}.
100 1
ux70
,"
d8~
" -
i .z
60
40
0, 0
20
40
. 60 .
.80
100
1;0
120
~176
T I M E , rain
Fig. l l - - S o l v e n t extraction of Ni and Co with Lix70-1auric
acid-Kermac 470B in the presence of ultrasound. Organic phase: 10 pct
(v/v) Lix 70; 0.1 M lauric acid; Kermac 470B. Aqueous phase: 30
m g / l Co 2+ and 30 m g / l Ni2+; 1 M KNO3; pHi., = 5.05. Ratio of
O / A : 100 rnl/100 ml; T = 25 ~ stirring rate = 2000 rprn; ultrasound: 20 KHz, 47 W.
Time to
Equilibrium
(Min)
Ultrasound
(47 W/cm 2)
No
Yes
No
Yes
Surfactant
(mole/l)
Lix 65N
System
Lix 70
System
No
No
0.005M DNNSA
0.005M DNNSA
60
10
30
--
150
60
140
60
METALLURGICAL TRANSACTIONS B
K. Energy Consumption
All of the described experiments took place in the continuous presence of ultrasound. In such batch applications, the total energy consumed was prohibitively high
and could not be offset even by the substantially lower
extraction time. However, it was found that continuous
ultrasound was not required for the extraction of nickel.
When ultrasound was briefly applied, then briefly discontinued, and these cycles repeated, no deterioration of
solvent extraction was found. This phenomenon can be
explained by the microdroplets formed in the presence
of ultrasound, which did not disappear immediately after
the ultrasound was turned off. The interfacial area produced allowed extraction to continue even in the absence
METALLURGICAL TRANSACTIONS B
of ultrasound. When a significant number of droplets coalesced, ultrasound was again required to form a new
batch of droplets and to continue extraction.
These observations prompted an examination of the
effect of intermittent ultrasound on solvent extraction of
nickel. The intermittent study was characterized by the
time when the ultrasound was "on" (T) and "off" (To).
The total on plus off time was fixed at 5 seconds; thus,
the frequency of on and off cycles was fixed to 12 per
minute. The ultrasound generator was turned on and off
by a specially built electronic controller. The total energy consumed by the ultrasonic apparatus during extraction was measured by a watt meter (the watt meter
was placed in line between the wall outlet and ultrasonic
generator). The intensity of introduced ultrasound was
also varied. The results (Table IV) show that energy
consumption was lowest when the on time was
0.25 seconds, the off time was 4.75 seconds, and the
ultrasonic input energy was 61 W. Under these conditions, the total consumed energy was 2.66 W per hour.
It is important to note that the on time can be substantially decreased by increasing the intensity of ultrasound,
with an overall decrease in energy consumption. Even
though a higher intensity ultrasound was used, the shorter
on time reduced energy consumption. The higher intensity ultrasound apparently forms more droplets, and
probably smaller droplets, which take a longer time to
coalesce, lengthening the off time.
When a similar study was performed with Lix 70, it
was found that intermittent ultrasound was not as effective as with Lix 65N. The stability of droplets with Lix 70
was much lower than with Lix 65N, causing them to coalesce faster and reducing their interracial area. Thus, to
achieve the same level of extraction, longer on times were
required with Lix 70.
The encouraging results obtained from the intermittent
application of ultrasound suggested that continuous-flow
solvent extraction might also reduce energy consumption. The objective of using a continuous flow was to
simulate the addition of a variable amount of ultrasonic
energy (of the same intensity) by varying the flow rate.
For this study, a separate reactor was designed (Figure 12),
consisting of two holding containers, one for the organic
phase and the other for the aqueous phase, an ultrasonic
0.25
0.50
1.00
2.00
0.25
0.25
4.75
4.50
4.00
3.00
4.75
4.75
Energy
Consumed
(Wh)
Equilibrium
Achieved
After (Min)
9,56
9.41
7.76
6.71
3.81
2.66
90
60
30
15
30
15
Fig. 12--Schematic representation of the apparatus for continuous solvent extraction in the presence of ultrasound.
takes place. U n d e r the conditions studied, the lowest ene r g y c o n s u m p t i o n was a c h i e v e d at 26 ~ and at the highest flow rates, 160 m l / m i n . About 33 pet Ni was extracted
under these conditions. H i g h e r flow rates were not exa m i n e d b e c a u s e o f the limitations o f the e x p e r i m e n t a l
setup. Despite this limitation, w h i c h could affect the accuracy o f the calculated c o n s u m p t i o n o f energy, the experiment still clearly indicated h o w energy c o n s u m p t i o n
could be reduced.
The lowest energy consumption obtained, 11 k w h / m 3,
is still too high to c o m p e t e with solvent extraction perf o r m e d b y c o n v e n t i o n a l m i x i n g , where the upper acceptable energy c o n s u m p t i o n limit is about 1 k w h / m 3.
Flow rate
Aqueous
(ml/min)
10
20
40
40
80
160
10
20
40
40
80
160
T (~
15
15
15
26
26
26
Residence
Time (Min)
Ni Extracted
(Pet)
Energy consumed
( K W h / m 3)
10.0
5.0
2.5
2.5
1.25
0.625
15
15
15
33
33
33
176
88
88
44
22
11
The experimental conditions were as follows. Organic: 10 pet (vol.) Lix 65N/0.1 M lauric acid/Kermac 470B. Aqueous: 30 mg Ni:
KNO3; pHini, = 5.05. Ultrasound: frequency = 20 KHz; intensity = 47 W.
20--VOLUME 23B, FEBRUARY 1992
0.5 M
METALLURGICALTRANSACTIONSB
Fig. 1 3 - - S c h e m a t i c representation of the proposed new method of solvent extraction: solvent extraction-in-pipe, possible because the extraction
rates are significantly enhanced with ultrasound.
However, the results from the continuous-flow and intermittent ultrasound experiments clearly suggest that
optimization studies may show how to further reduce energy consumption. The most promising solvent extraction technique may be the one that utilizes a pipe for
mixing the organic and aqueous phases, with properly
spaced ultrasonic horns inserted through the pipe wall
(Figure 13). In this kind of a plant, in addition to the
composition of aqueous and organic phases, the other
important process parameters would be the flow rate in
the pipe, the size of ultrasonic horns, the spacing of
ultrasonic horns, and the ultrasonic input energy. The
pipe length between a leaching reactor and a settler would
serve as a mixer, thus avoiding the need for a mixer as
in the traditional mixer/settler design.
IV.
CONCLUSIONS
ACKNOWLEDGMENT
This work was supported by the United States
Department of the Interior, Bureau of Mines, under
Contract No. J0134035 through Department of
Energy Contract No. DE-AC07-76IDO1570.
VOLUME 23B, FEBRUARY 1992--21
REFERENCES
1. C.E. Jordan and B.J. Scheiner: Proc. Eng. Found. Conf., Palm
Coast, FL, Dec. 1989, S. Chander and R.R. Ktimpel, eds.,
Society of Mining Engineers, Inc., Littleton, CO, pp. 147-54.
2. M.V. Gramatikov and M.D. Ivanovskii: J. Appl. Electrochem.
USSR, 1972, vol. 45 (6), 1220-23.
3. R. Walker: Hydrometallurgy, 1979, vol. 4, pp. 209-15.
4. H.V. Fairbanks and W.I. Chen: Sonochem. Eng., Chem. Eng.
Prog. Syrup. Ser., 1971, vol. 67 (109), pp. 108-16.
5. K.R. Purdy, G.W. Simmons, A.E. Hribar, and E.I. Griggs:
Sonochem. Eng., Chem. Eng. Prog. Syrup. Ser., 1971,
vol. 67 (109), pp. 68-73.
6. H.A. Woodle, Jr. and F.C. Vilbrandt: A1ChE J., 1960, vol. 62,
pp. 296-98.
7. E.C. Chen and W.Y. Chon: CEP Symp. Ser. 61, 1967, vol. 77,
p. 44.
8. F. Flisak and A. Perna: Ultrasonics, 1977, Jan., pp. 27-29.
9. B. Pesic and T. Zhou: J. Met., 1989, June, pp. 42-45.
10. D.S. Flett and D.W. West: Proc. Int. Solv. Extrac. Conf., SCI,
London, 1971, vol. 1, pp. 214-23.
11. D.S. Flett, M. Cox, and J.D. Heels: Proc. Int. Solv. Extrac.
Conf., SCI, London, 1974, vol. 3, pp. 2559-75.
12. D.S. Flea, M. Cox, and J.D. Heels: J. lnorg. Nucl. Chem., 1975,
vol. 37, pp. 2533-37.
13. T. Tammi: Tutkimus Tek, 1976, vol. 4, pp. 44-45 (Chem. Abstr.
89 118498j).
14. B.G. Nyman and L. Hummelstedt: Proc. Int. Solv. Extrac. Conf.,
Lyon, France, 1974, vol. 1, pp. 669-84.
15. L. Hummelstedt, H.E. Sund, J. Karjaluoto, L.O. Berts, and B.G.
Nyman: Proc. Int. Solv. Extrac. Conf., Lyon, France, London,
1974, vol. 1, pp. 829-48.
16. D.S. Flett, M. Cox, and J.D. Heels: J. lnorg. Nucl. Chem., 1975,
vol. 37, pp. 2197-2201.
17. T. Zhou and B. Pesic: Trans. IMM, 1989, vol. 98, C147-C152.
18. M. Cox and D.S. Flett: Chem. Ind. (London), 1987, vol. 6,
pp. 188-93.
19. D.S. Flett: Chem. Ind. (London), 1977, vol. 6, pp. 223-24.
20. D. Bauer and J. Komomiki: Proc. Int. Solv. Extrac. Conf., Denver,
CO, 1983, pp. 315-16.
METALLURGICAL TRANSACTIONS B