Professional Documents
Culture Documents
Approved By:
Revised By:
Date of Approval:
Date of Implementation:
SANAS
Ron Josias
Mpho Phaloane
Neville Tayler
Page 1 of 7
TR 44-02
CONTENTS:
1.
2.
3.
Abbreviations ......................................................................................................................... 4
4.
5.
General Requirements........................................................................................................... 5
6.
7.
Appendix 1:
Accreditation Schedule........................................................................................... 6
SANAS
Page 2 of 7
TR 44-02
1.
2.
References
[1]
[2]
[3]
NPL, UK 1996
[4]
ISO/IEC Guide
99-12:2007
2007.
Definitions
Absolute Humidity
The mass of water vapour present in a unit volume of moist air of a given temperature and
-3
pressure. SI (metric) units are grams of water per cubic metre of air (g.m ). [3]
Dew Point
The temperature at which dew, or condensation forms, on cooling a gas. This is in effect the
temperature at which air becomes saturated in equilibrium with water, expressed in degrees
Celsius (C) or occasionally other units of temperature.[3]
Repeatability
repeatability
measurement precision under a set of repeatability conditions of measurement. [4]
Relative Humidity
Relative humidity is the ratio of the amount of water vapour, e, in the air to the amount of water
vapour, es, that would be in the air if saturated at the same temperature and pressure, and can
be expressed
relative humidity (in %) = e/es X 100
Where e is the water vapour pressure, and es is the saturation vapour pressure of the prevailing
ambient temperature [3]
SANAS
Page 3 of 7
TR 44-02
3.
Abbreviations
aw
CMC
MSDS
rh
SANAS
4.
water activity
Calibration and Measurement Capability
Material Safety Data Sheets
Relative Humidity
South African National Accreditation System
Environmental Requirements
Significant changes in relative humidity may occur with even slight variations in temperature and without
change in absolute humidity. For example, a 1C variance in temperature at 20C and 50% rh can
introduce an error of 3,2 %rh. [3]
Therfore an accredited laboratory working in the field of humidity measurements shall operate under the
following environmental conditions:
4.1
The laboratory should be maintained at a temperature of 23C 3C. Where measurements are
performed outside this specified requirement, the laboratory must ensure that appropriate
corrections are applied.
4.2
Rates of change and gradients of the ambient temperature and humidity must be kept to below
1 deg C per hour, and 10% rh per hour.
4.3
A high degree of cleanliness shall be maintained in the laboratory and adequate lighting shall be
provided. [1]
4.4
Localized isolation in the form of chambers, cabinets and enclosures may be utilized in order to
minimize the effects of thermal currents and temperature gradients and in so doing provide a
stable and homogeneous environment conducive to reduced measurement uncertainties.
4.5
The laboratory shall consider and include the stability of the environmental conditions on the
relative humidity when estimating the uncertainty of measurement.
SANAS
Page 4 of 7
TR 44-02
5.
6.
7.
General Requirements
5.1
Laboratories operating in the field of humidity may be accredited for the calibration of
dewpoint hygrometers, relative humidity hygrometers, humidity generators, water activity
meters, saturated and unsaturated salt solutions, environmental chambers, and limited
temperature calibration.
(Refer to the Schedule of Accreditation Appendix 1)
5.2
5.3
The laboratory will measure and report the measured value in the units applicable for each
specific parameter. Generally, the units will be C (dew point), C (frost point), %rh relative
humidity, aw (water activity).
5.4
The laboratory is required to have MSDS for all standards fluids, and chemicals, including
saturated and unsaturated salt solutions, and cleaning materials used within the laboratory. [2]
Technical Requirements
6.1
The laboratory is required to have procedures in place for the storage and usage of standards in
order to ensure their ongoing integrity.
6.2
The laboratory shall have and apply procedures for all calibrations.
6.3
The laboratory shall have the necessary standards and equipment required to perform the
calibration measurements in accordance with their accreditation schedule.
6.4
All measurements shall be taken only after adequate stabilization time. This period shall be
addressed in the calibration procedures, taking in to account possible differences in response
time due to different time constants of the reference standard and the hygrometers or salt
solutions under test.
6.5
The calibration of environmental and climatic chambers requires the availability of suitable
instrumentation capable of data storage. Laboratory procedures shall address the number of
temperature and relative humidity sensors required, and the spatial distribution thereof, and this
shall be dependant on the volume of the chamber. Guidance may be obtained from IEC 68-266:1994 Environmental testing. The calibration procedure shall address the time period over
which the measurements are to be performed
6.6
It is recommended that the humidity be measured at each point in the chamber where the
temperature is measured. The application of corrections based on spatial temperature
measurements, will result in humidity values with a larger uncertainty.
6.7
6.8
Laboratories shall record and report the temperature at which the measurement of water activity
or relative humidity has been made.
SANAS
Page 5 of 7
TR 44-02
SCHEDULE OF ACCREDITATION
Permanent Address of Laboratory:
SuperLab
25 Popular Street
Tree Park
Port Southern Africa
Technical Signatories
: Dr I. Dolittle
Mr D Green
Postal Address:
P O Box 22222
Port Southern Africa
X000
Nominated Representative
: Mrs M Brown
Issue No.
Date of issue
Expiry date
: 01
: October 200X
: October 200X
ITEM
RANGE OF MEASURED
QUANTITY
CALIBRATION AND
MEASUREMENT
CAPABILITIES EXPRESSED
AS AN UNCERTAINTY ()
- 75 C to - 60 C
- 60 C to 17 C
0 % rh
11 % rh
33 % rh
53 % rh
75 % rh
90 % rh
95 % rh
2 % rh
2 % rh
2 % rh
2 % rh
2 % rh
2 % rh
2 % rh
Dew-point Hygrometer
Digital Hygrometers /
Thermo hygrometers
15 C to 30 C
10 to 90 % rh
15 C to 50 C
2 % rh
1C
Analogue RH Indicators
10 to 90 % rh
5 % rh
Humidity Recorders
Thermo-Hygrograph
10 % rh to 90 % rh
15 C to 50 C
5 % rh
1C
Data Loggers
10 % rh to 90 % rh
15 C to 50 C
3 % rh
1C
0,100 aw to 0,750 aw
0,750 aw to 0,950 aw
0,020 aw
0,025 aw
5 % rh to 95 % rh
4 % rh
Water Activity
Environmental Chambers
SANAS
Page 6 of 7
TR 44-02
Proposed By:
Section
Change
FM / STC
FM
FM / STC
FM / STC
FM / STC
FM
References
Definitions
Abbreviations
General
Technical
Appendix 1
SANAS
Page 7 of 7