Professional Documents
Culture Documents
Catastrophic error can be easily detected when the measurement is repeated and new readings are
compared with the other readings. For instance, assume that a measurement process results in the
following dimensions on a component: 20.2 mm, 20.1 mm, 19.9 mm, 20.1 mm, 23.0 mm and 20.0
mm. In this set of measurement it is obvious that the reading having the value 23.0 mm is
significantly different from the others and this could be due to errors made by the instrument
reader. This reading can, therefore, be removed from the set of measurements before further
analysis is carried out, such as determining the average dimension.
2.3.2 Alignment error
Alignment error occurs because the instrument is wrongly aligned relative to the workpiece or
part being measured. An example of a situation that causes alignment error in the use of a dial
indicator is shown in Figure 2.1. The dimension of interest is D, but due to misalignment the dial
indicator gives a reading L which is larger than D. The error e in the measurement due to the
misalignment is, therefore,
Error , e = L D = L L cos = L (1 cos )
(2.1)
For example, if the angle due to misalignment = 10 and the dial indicator shows a reading on
4.52 mm, then the error e = 4.52(1 cos10) = 0.069 mm.
Since this type of error is a function of the cosine of the angle of misalignment, hence it is
also known as cosine error.
Dial indicator
L D
b
a
D
b
L = L ( 1t 1 2t 2 )
(2.2)
where,
L is the dimension measured,
t1 is the temperature of part being measured,
t2 is the temperature of measuring instrument,
1 is the thermal expansion coefficient of part material,
2 is the thermal expansion coefficient of instrument material,
t1 = t1 20C
t2 = t2 20C
Contraction (m)
From eq.(2.2) we can see that in order to eliminate error L in measurement both the part
and instrument temperatures must be 20C. In this case, t1 = t2 = 0 and, therefore, L = 0. This
requires the part and the instrument to be kept in a controlled laboratory maintained at 20C.
However, if the temperature is not 20C but is at some arbitrary but equal value, e.g. t1=t2=t, then
L = 0 if 1 = 2. This implies that both the part and instrument must be made from the same
material since they have the same expansion coefficient. In practice, the part and the instrument are
usually made from different materials and their temperatures are difficult to measure. Neither is it
possible to maintain their temperatures at exactly 20C.
Thus, the most effective way of overcoming error due to the temperature changes is to leave
both the part and instrument to achieve stable ambient temperature. Then, the effect of the
deviation of the actual temperature from 20C is taken into account in the evaluating the error. As
an example the effect of temperature on a block gage as a function of time after being handled is
shown in Figure 2.2. The figure shows that although the time has exceeded 30 minutes, the
dimension of the block gage has not returned to its original value.
2.0
0.25
30
15
Time (min.)
10
Stylus
Deformation of cylinder
Deformation of stylus
Cylinder
2
3
= 1.774W (k1 + k
2
2 3
1
1 3
+
R1 R 2
(2.3)
where,
R1 = radius of stylus,
R2 = radius cylinder,
W = pressure at the stylus,
(1 12 )
k1 =
E1
k2 =
(1 )
2
2
E2
where E1 and E2 are the modulus of Young for the stylus and cylinder material and 1 and 2 are the
corresponding Poissons ratios.
Another form of deformation occurs when a body deforms due to its own weight, known as
dead weight. For a beam the error caused by this type of deformation can be minimized by placing
supports at position shown in Figure 2.4. Under this condition, the deflection of the beam, and
hence the error, is minimum.
11
Beam
Support
0.577 L
Normal
direction
Pointer
Scale
Reading taken
Parallax
error
Actual reading
(b)
(a)
x 1 + x 2 + x 3 + ....... + x n
n
(2.4)
12
(x x )
(2.5)
n 1
If the measurement of a value x is repeated a large number of times, we will get slightly
different values due to the random errors. A frequency distribution of the measurement can be
obtained by plotting the measured value against the number of times each measurement occurs or
number of readings that fall within a range. The distribution will take the familiar bell-shaped
(Gaussian) profile shown in Figure 2.6 where the variable z is given by
z=
(2.6)
2
1
e z
2
(2.7)
The function in eq.(2.7) and shown plotted in Figure 2.6 using =2 is also known as the
probability density function. An important property of the curve shown in Figure 2.6 is that 99.8%
of the data lies within 3 of the mean value. For instance, if a sample of 2000 parts, say nails, is
taken and their diameters measured, we can say with 99.8% level of confidence that the diameter
will lie within 3 of the mean diameter, where is the standard deviation of the 2000 samples.
Similarly, 95% of the data lie within 2 of the mean.
f(z)
0.25
0.2
0.15
0.1
0.05
0
-5
-4
-3
-2
-1
upper limit of 20.1 mm if the measuring instruments shows a reading of 20.12 mm, the product will
be rejected although the reading may be due to the deviation of the measurement process. If the
standard deviation of the measurement process (m) is large compared to the standard deviation of
the manufacturing process (p) more products will be rejected by mistake at the upper limit and
more will be accepted at the lower limit.
Manufacturing process
distribution
Measurement process
distribution
(6m)
Manufacturing tolerance
(6p)
M =
where
M
M
M
a +
b +
c
a
b
c
M
the partial derivative of M with respect to a
a
M
the partial derivative of M with respect to b and so on.
b
The error M is known as compound error. Exercise 2.3 illustrates a problem where the
compound error is to be determined.
14
Revision Exercises
Exercise 2.1
Figure below shows an arrangement for measuring the dimension L on a block using a dial gage. If
the dial gage gives a reading of 5.212 mm, calculate the error in the measurement. Given that angle
= 5.
Dial gage
L
block
Exercise 2.2
The diameter of an aluminium rod is measured using a micrometer in a laboratory where the
temperature is 25C. The reading shown by the micrometer is 12.54 mm. Given that the coefficient
of thermal expansions of aluminium is 2310-6/C and that of the micrometer material is 11.7
10-6/C determine the error in the measurement and the true diameter of the rod.
Exercise 2.3
The volume V of a cylinder is given by the expression: V = r 2 h , where r is the radius and h is the
height of the cylinder. If r = 50 mm and h = 200 mm, and the errors in the measurement of r and h
are, respectively, 0.5 mm and 1.0 mm, determine the error in the volume V.
15