Professional Documents
Culture Documents
Administration Manual
Version 4.0
Date of Revision:
June 18, 2012
Table of Contents
1
2
6
7
2 Technical Aspects
2.1 Requirements for minispec Plus installation
If you have already worked with Bruker the minispec Software on the same
PC please note that ...
Bruker the minispec Software Version 2.59 Rev. 2 or higher is required.
(See also Bruker the minispec Software Help about the
minispec). For correct installation see BeforeFirstRun.txt on the
installation CD and the instructions below.
Every care has been taken that your measurement and calibration
data will not be affected by this installation. However, for safety
reasons, we recommend you to backup all your data before starting
with the installation. Bruker cannot accept any liability for lost data.
The
Default
path
for
installation
of
minispec.exe
is:
C:\Program Files\Bruker the minispec\. You can change the
path as desired. It is also technically possible to install several versions of
the minispec software on the same PC. However this is not
recommended.
When you upgrade from an earlier version of minispec Plus and if you
used one of the Bruker SR applications, please note that the ExpSpel file
sr_server.app was changed to sr_server2.app. Older SR calibrations and
methods will no longer work with minispec Plus 4.0 and are hidden
automatically during the upgrade process. You have to create new
calibrations/methods with the upgraded SR application.
7. minispec Plus automatically prompts you to enter the serial number for
the minispec to connect to (see Figure 4). For mq-series and LF-series this
is NDxxxx, where xxxx is a 4-digit number. In case of a mqone minispec,
the serial number will start with NE, e.g. NE1234. Alternatively you can
quit the dialog box and leave DEMO which allows software
demonstration without the need of being connected to a minispec
instrument.
10
Figure 4: First run: Enter the serial number of the connected instrument.
8. If you want, you can browse through the pages of the configuration
window to adjust minispec Pluss settings to your requirements (see
chapter 4). Click the button Apply to save the configuration data,
then leave the configuration window via Exit button. Now minispec Plus
is ready for use.
11
minispec Plus
mqone
mq-series
BVT 3000
Balance
Figure 5: Devices supported by minispec Plus
12
MS Excel
OPUS
BVT.exe
m+ Front End
m+ Back End
m+ SYS
m+ CTE
MS Access
Figure 6: Architecture of minispec Plus
The architecture of minispec Plus is such that the user interface is separated
from the database storage and from the NMR acquisition program
minispec.exe (see Figure 6).
The minispec Plus front end (C:\Program Files\Bruker the minispec\
minispec Plus\minispec Plus FE.mde) contains the user interface, i.e. the
visible part of the software you are working with. It does not communicate
directly with the hardware, but rather with programs like Bruker the minispec
Software (minispec.exe) to control the minispec instrument and BVT.exe to
control the tempering unit BVT 3000. This is why during measurements the
Windows task bar shows these programs. However as long as minispec Plus is
running, they cannot be accessed directly.
The frond end exports data to Microsoft Excel and communicates with
OPUS for chemometric data processing.
The minispec Plus back end (minispec Plus BE.mdb) is the database where
all data concerning measurement, calibration, and configuration are stored.
Please consider that a loss of the back end means a loss of all measured
data. Bruker therefore recommends a regular backup of the back end file. A
data base can be restored simply by defining the backup file as the new
backend data base in minispec Plus (Figure 6, Data Base Definition).
If you set value on the recorded NMR data, also backup the files in the raw
data folder defined in the calibration section of the configruation window
(Figure 7).
13
14
4.1 Environment
Figure 7 depicts the Environment section, which holds software specific
settings:
Path for archiving the data base: minispec Plus allows you to archive
previously acquired data which is no longer of current interest. The default
path where to store the archive data base is given here. The archiving
process is described in chapter 5.4.
15
Default Path for (Excel) export files defines the default path for exporting
data to Microsoft Excel or ASCII files. This default path is used within the
calibration and also within the measurement menu.
The check box Open Excel File determines whether the exported Excel file is
opened immediately after the export is finished.
Max. Mass defines the maximum allowed mass which can be entered in
measurement and calibration procedures. Its purpose is to avoid serious
typing errors.
For making sample mass dependent minispec measurements more
comfortable, the direct communication with a balance (either Sartorius or
Mettler protocols) is possible, allowing an automatic transfer of sample masses
to minispec Plus. The Balance option has to be selected, and consecutively,
the balance type has to be specified in the appearing box. The balance has
to be connected to the PC. The interface parameters have to be specified in
Bruker the minispec Software. Further information about this step is available
in the users manual of the mq minispec.
Some balances support an animal weighing mode which automatically
averages a number of readings taken during a given time span. This mode is
very convenient in combination with the LF-series for weighing animals which
move during the weighing process and thus make a normal result unreliable.
The animal weighing mode has to be configured directly with the balances
panel. minispec Plus requires no further configuration. To find out whether your
balance supports the animal weighing mode and how to configure it, check
your balances manual.
Decimal Places specifies the number of digits that are displayed after the
decimal point of numerical results. This number depends on the application
and the measurement accuracy. In general 4 digits should not be exceeded.
In addition, the accuracy of the sample mass can be specified, which is
application specific (for example animal weighing compared to weighing of
yarns).
The Remove mass entry after every measurement option refers to the
measurement and calibration windows where a sample mass has to be
entered. You may want to switch this feature off if you want to measure
samples with the same mass repeatedly or set up input-less measurements
when combined with consecutive numbering of samples.
Show Tip Texts defines the appearance of tool tip texts on buttons and test
fields as shown below.
16
Format
Description
SampleName
SampleMass
float
17
Pos. Name
Format
Description
Result1
float
Result2
float
Unit
text
% or g
Batch
Calibration
timestamp of measurement
Comment
sample comment
10
Valid
text
OK = successful measurement
Err = error during measurement
11
Outlier
text
12
User
text, max. 50
characters
File name
<SampleName>_<Batch>_YYYYMMDD_HHMMSS.[xml|csv]
Timestamp of the Export:
YYYY = year
MM = month
DD = day
HH = hours
MM = minutes
SS = seconds
Example of a XML-file
name: tsample_tbatch_ 2005_11_02_102512.xml
18
4.2 minispec
The page minispec (Figure 8) defines some hardware settings:
The Serial Number of the instrument to be connected to has to be specified
(NDxxxx for mq- and LF series / NExxxx for mqone). It can be found on the rear
of the electronic unit).
If you control two or more minispec instruments with one PC you can switch
between them using this serial number box and store all acquired data in one
data base.
The Path is the installation path of the latest used Bruker the minispec
Software (minispec.exe).
It is not recommended to have two installations of minispec Software on the
same PC. However if you use such a setup, please make sure you use the
correct version of minispec.exe. To define the version of Bruker the minispec
Software, close minispec Plus and then start Bruker the minispec Software.
The path in minispec Plus will be set automatically with the next start of
minispec Plus.
19
20
For SFC instruments, also the additional SFC calibration standard samples
have to be measured and the so-called k, f and o factors are determined.
4.3 Calibration
In the Calibration page (Figure 10), the calibration and data processing
mode is chosen. For each of the processing methods a license number is
stored in the corresponding NDxxxx.ini or NExxxx.ini file. Only if such a
license is available, the calibration mode option can be chosen.
21
The first data processing method is the classic calibration mode, where a
characteristic NMR parameter is directly correlated to a material property,
e.g. oil content in seeds. In this case, the minimum number of calibration
samples is to be defined. As classic TD-NMR measurements use linear
regression models, the minimum accepted number of calibration samples is
two. However, it is recommended to increase that number for a better
reliability of the calibration. For some methods a polynomial fit leads to better
results than a linear regression; in this case, 3 is the minimum number of
samples.
In the second mode, the chemometric capabilities of OPUS are used for
deriving material properties from NMR signals (Figure 11). The OPUS program is
opened when required for data exchange and processing. Therefore the
OPUS installation path has to be defined according to the current OPUS
installation, i.e where the OPUS.exe file is located.
mq-series and LF-series instruments support all chemometric applications;
mqones only support simple pulse sequences like FID, Hahn echo and solid
echo.
22
23
In addition, the Path for minispec data files for minispec raw data (*.mdt
files) is configurable here. The main path can be specified in more detail by
adding the sample batch and / or the user name. The raw data of each
minispec measurement are stored in ASCII format in the specified folder using
the sample name and the date of acquisition in order to guarantee both a
simple identification mechanism and an unique data file name, e.g.
C:\Bruker minispec Plus Data\RawData\myBatch\Admin\mySample_200
90607_0942177.mdt. This feature also allows alternative methods of manual
data processing.
In this context the requirements for passwords are set. The window appearing
when clicking Password Properties contains the specifications each
password within minispec Plus has to fulfill (Figure 14). If the first checkbox
24
Password required is un-checked, the log in requires only the name of the
user in order to apply the according rights. Only the electronic signing requires
the entry of a password in any case.
According to these settings, accounts will be disabled automatically when an
account has not been used for more than the defined time span or there
have been too many failed login attempts. Only an administrator can then reactivate the disabled account (see Chapter 5.1.2).
25
Another option is to work with two different back end files one for testing
purposes and the other one for production. With the Data Base Definition
function you can switch between testing and production data.
A third option is to use one PC to control two or more minispec instruments. If
you can either keep the data of both instruments in one back end file or use
separate ones and switch the back end files to connect with the other
instrument.
26
4.6 Application
minispec Plus includes a pool of NMR time domain applications which are
organized in groups of practical scope. The configuration page Application
(Figure 17) lists all application groups in minispec Plus. The ones which are
licensed in the NE/NDxxx.ini file are checked. Only applications in selected
application groups are available for calibration and measurement.
For SFC instruments, the option Solid Fat Content (SFC) in Fat Compositions is
selected. Whenever this entry is activated, the main menu of minispec Plus
shows SFC Method Definition rather than Calibrate, no matter if other
application groups are selected or not. For more details please refer to the
minispec Plus manual for classic applications.
The button Toggle SFC Mode activates or deactivates the entry Solid Fat
Content (SFC) in Fat Compositions on the left hand side. If the
ND/NExxxx.ini file contains a license for SFC (230005) the SFC group will be
activated after each start of minispec Plus.
27
Figure 17: Tab Application in the configuration part of minispec Plus. In this window,
application groups can be chosen which are then available in the calibration part.
28
When a data set is marked by mouse click and the data is edited, a pen
appears, indicating that the underlying data set in a database table is
changed (e.g. Fat and Water Analysis as in Figure 18). Pressing the ESC
button cancels your changes and restores the original data set.
The pulse sequence, the corresponding timing and the processing of data are
characteristic for a certain time domain application. A minispec Plus
application is primarily defined by a fixed ExpSpel file describing the NMR
pulse sequence. The administrator has access to the timing parameters by a
click on Define / Show Parameters.
minispec Pluss NMR applications are defined according to worldwide
standards. These standard applications are locked so that no changes can
be made. However, it may be necessary to apply minor adaptations to
special requirements. This can be achieved by duplicating an existing Bruker
standard application and modifying timing and processing parameters.
Mark the new application by mouse click and by a click on the Edit button a
pen appears, indicating that you can change the applications name, group
and locked flag. Pressing the ESC button cancels your changes.
Only locked applications are available for calibration and measurement. Be
sure to lock only after you have checked the applications timing as
applications cannot be un-fixed. To check or modify the timing click on
Define / Show Parameters (see Figure 18).
29
Figure 19: Define / Show Parameters Menu. The visualized pulse sequence is applied.
This window (Figure 19) shows the NMR signal and the timing windows for its
analysis. Please change the NMR parameters only when you are familiar with
the underlying TD-NMR measurement principles. Bruker cannot guarantee
that an application with new timing will work properly. In order to test the
signal response of a special sequence, a Pilot Scan can be performed with
Check NMR Signal. The pulse sequence and the timing specified in the
application will be used together with values to be entered for gain and
scans to perform a minispec measurement (Figure 20).
30
Figure 20: Define / Show Parameters. A NMR signal is acquired with the parameters previously
defined.
The NMR signal remains visible until you click the Stop button.
Note that the minispec hardware is not equally suited for all applications. It is
possible that an applications sampling window overlaps with the instruments
receiver dead time, so that not enough data points can be acquired. In this
case you are prompted to use an appropriate application, i.e. move the
sampling window so that you can be sure it begins after the receiver dead
time is over.
31
32
User Group
Rights
Administrator
Calibrate
calibration, measurement
Measure
measurement
The next table summarizes the hierarchy of user groups and the primary types
of operations that each user group can perform. For example, a user who
only needs to measure samples can obtain an account with measurement
rights only. A person that needs to make measurements as well as calibrations
could obtain an account with calibration and measurement rights. A plant
managers or the facilitys specialists account might be configured as
Administrator. The administrator has the most control of all user groups. You
will notice that various options of the minispec Plus software are grayed or
hidden when a user groups Calibrate or Measure are used.
Function, type of operation
Measure
Calibrate
Administrator
Sign calibration
33
Upon installing the minispec Plus software there are three default users
corresponding to the three levels of user rights:
User Name
Password
User group
Language
Admin
admin
Administrator
English
Calibrate
calibrate
English
Measure
measure
Operator
English
Figure 22: User Account configuration menu. New accounts are defined here, existing
accounts can be modified or deleted.
According to the password settings (see Chapter 4.4) the Enabled flag is
unchecked automatically when an account has not been used for more
days than the defined time span has or if there have been too many failed
34
Some actions include more detailed information which are shown in the text
box in the windows footer.
The action log can be printed or exported to either Excel or ASCII text files.
35
Figure 24: Error Log in minispec Plus. Errors occurring during operation are recorded and are
listed in this window.
After a click on the View Error Log button, a window listing the errors
occurred during minispec Plus operation is displayed. The date of the error
creation can be used as filter parameter. Enter a date, confirm by return key,
and the filter will be applied. The usual export functions are also available.
36
After clicking OK, you are prompted that the archiving procedure will need
some time. The archiving process can be canceled or started at this point.
If you want to examine the data in the archived data base later on, click on
Configure Environment in the administration window (Figure 21). Then
navigate to data base and Data Base Definition (Figure 15). By entering
the name of the archive database the front end will link it. After logging in you
can have a look at the data, but you cannot modify it. In order to switch
back to the current back end, navigate to Data Base Definition and enter
the corresponding file name.
37
Function
CalibResult.xltx
Classic calibration
CalibResultTwoCalib.xltx
ChemCalibResult.xltx
Chemometric calibration
CalibStability.xltx
CalibStabilityTwoCalib.xltx
MeasResult.xltx
Classic measurement
MeasResultSFC.xltx
MeasResultSFCI.xltx
MeasResultTwoCalib.xltx
ChemResult.xltx
Chemometric measurement
NoCalibResult.xltx
RelaxResult.xltx
Relaxation measurement
2. Create
a
backup
copy
of
the
template
file
C:\Program Files\Bruker the minispec\minispec Plus\
ExcelTemplates\<Chosen Template>.xltx
3. Open Excel from Windows Start Menu (do not double click on
<ChosenTemplate>.xltx)
4. Load your chosen template using the Excel menu.
5. Apply your changes
6. Save the template.
38
Then click the left mouse button and drag it down to as many rows as you
want.
39
6 Installation of OPUS 7
An extensive description of OPUS installation and its features can be found in
the original OPUS documentation. This chaper provides only a hands-on
workflow to set up OPUS for chemometric processing of minispec Plus data.
1. When you install OPUS, start setup.exe with administrator rights:
40
3. Click on Next.
41
42
43
44
11. Now the software is installed, which may take a few minutes. You
can skip the printout of the installation protocol. Finally, on the
next window do not check I want to start OPUS and click on
Finish
12. For the first time start right click on the OPUS Icon on the desktop
and select Run as administrator
45
Select the item Setup Quant 2Method and drag it to the toolbar
at the top of OPUS main window. Do the same with Quantitative
Analysis 2. IMPORTANT: Use the Quant 2 commands, not the
Quant commands.
17. If you want to use Multi Evaluation, also drag the commands
Multi Evaluation Setup (old) and Multi Evaluation Test (old).
IMPORTANT: Take the commands with the suffix (old), not the
ones without the suffix. Finally your toolbar should look like this:
18. If you want to use Multi Evaluation, the necessary folders to store
the methods have to be created. The easiest way to do this, is to
click on the icon Multi Evaluation Setup (old) (the one which
displays the tool tip text Configure ME Method).
Then click on Load Multi Evaluation Method.
Enter an arbitrary method name and click on Open.
46
47
48
7 Troubleshooting
7.1 General
Bruker strives to supply you with instructional and accurate documentation.
Therefore we encourage you to tell us how we are doing. Please send us your
suggestions for improvements, corrections, or bug reports. If there is anything
you particularly liked, tell us as well. With your input and assistance, Bruker can
continually improve its products and documentation. You can send your
messages and correspondence via e-mail, FAX, telephone, or mail. It is
important to include the document name, product name, version number,
and page number in your response. Thank you for your help.
Do NOT click on OK, but on Cancel. Then install Service Pack 1 for Microsoft
Office 2010. You can download it for free from:
http://www.microsoft.com/download/en/details.aspx?id=26622
(as at April, 2012)
If you accidentally clicked on OK, delete the file minispec Plus FE.mde and
re-start the installation and select the Recover Last Known option.
49
7.3 HP Tools
On one occation, repeated crashes of the minispec software occurred, when
a certain load was on the processor and the HP (Hewlett Packard) Tools were
installed on a PC built in 2011.
However, this behavior could not be reproduced on PCs from 2012 with
current versions of HP Tools and on PCs without HP Tools.
If you happen to have an older PC which shows the abovementioned
behavior, please uninstall the HP Tools.
50