Professional Documents
Culture Documents
Feb 2004
firmly.
(a)
(b)
(c)
(d)
For elements with analysis wavelengths more than 430 nm, use either Non-BGC mode
(available in AA-6200/6601/6701/6800) or BGC-SR mode (available in AA6601/6701/6800).
In AA-6300, BGC-SR mode can only be used for analysis wavelength more than 600
nm if the AA-6300 is equipped with the optional photomultiplier.
BG signal too large to balance - the hollow cathode lamp intensity smaller than the
background light intensity. Increase the hollow cathode lamp current slightly or use a
smaller slit width (in case of BGC-D2 mode).
BG signal too small to balance - the hollow cathode lamp intensity is larger than the
background light intensity. Decrease the hollow cathode lamp current slightly or use
a bigger slit width (in case of BGC-D2 mode).
Check that the gas pressure supply is correct 0.09 MPa for acetylene and 0.35 MPa
Check that the airflow rate is not too high, i.e. the flow rate should be around 8 L/min.
For nitrous oxide-acetylene flame, check that the high temperature burner head is used
and that the burner recognition plug (for AA-6200) or key (for AA6601/6701/6800/6300) is properly inserted.
For AA-6601/6701/6800, check that the gas leak check has completed:
.
Ensure that the suction force of the ventilation system is appropriate (600 1200
m3/h). Too much suction force can cause the flow of gas from the burner head to be
sucked away before it can be ignited by the pilot flame.
connected.
Ensure that the nitrous oxide-acetylene flame is correctly selected in the software:
Recheck the analytical conditions such as the horizontal position and angle of
burner.
Unstable flame - the gas supply pressure is not consistent. Ensure that the gas
pressure does not fluctuate.
Unstable flame - the quality of gases is not good. Ensure that the gas supplies are
free from water, moisture and dust.
While aspirating water through the nebulizer, check that air-acetylene flame is
lean blue if not, the acetylene gas supply is finishing and organic solvent from
the cylinder tank is flowing to the burner head, causing the flame to turn orange.
The acetylene cylinder should be replaced if the pressure decreases to 5 kg/cm2.
For nitrous oxide-acetylene flame, the red portion height should be at least 1 cm.
Ensure that the water/solvent used in the analysis is clean. This can be checked
by aspirating water/solvent through the nebulizer and checking that air-acetylene
flame is lean blue, i.e. similar to when water/solvent is not aspirated.
Ensure that the drainage is good. Ensure that drain tube is hanging freely in the
air inside the drain container, i.e. the water level does not touch the end of the
tube.
Ensure that the suction force of the ventilation system is appropriate (600 1200
m3/h). Too much suction force can cause the flame to flicker as well as create
excessive noise.
For AA-6200 and AA-6800, the impact bead in the nebulizer is loose. This can
be checked by dismounting the nebulizer:
(a) AA-6200 - section 6.1.2.1 in AA-6200 Instruction Manual.
(b) AA-6800 - section 6.2.2.1 in AA-6800 Instruction Manual.
Note: The user is advised to contact us if not familiar with dismounting the nebulizer
which can be hazardous if not carried out properly.
(a)
(b)
(c)
(d)
(a)
(b)
(c)
(d)
Unstable flame - check that the burner slot is not clogged by carbide (especially
for nitrous oxide-acetylene flame) or salts. Clean the slot with thick paper or
razor. Also, the interior of the burner head might be dirty. For details:
AA-6200 - section 6.1.1 in AA-6200 Instruction Manual.
AA-6601/6701 - section 4.2.1 in AA-6601/6701 Instruction Manual (Users
System Guide).
AA-6800 - section 6.2.1 in AA-6800 Instruction Manual.
AA-6300 - section 8.2.1 in AA-6300 Instruction Manual.
The nebulizer might be clogged. Clean it with cleaning wire:
AA-6200 - section 6.1.2.1 in AA-6200 Instruction Manual.
AA-6601/6701 - section 4.2.2.1 in AA-6601/6701 Instruction Manual (Users
System Guide).
AA-6800 - section 6.2.2.1 in AA-6800 Instruction Manual.
AA-6300 - section 8.2.2.1 in AA-6300 Instruction Manual
The polyethylene tube covering the capillary of nebulizer might be broken and
cause leakage. Replace the polyethylene tube:
(a) AA-6200 - section 6.1.2.2 in AA-6200 Instruction Manual.
(b) AA-6601/6701 - section 4.2.2.2 in AA-6601/6701 Instruction Manual (Users
System Guide).
(c) AA-6800 - section 6.2.2.2 in AA-6800 Instruction Manual.
(d) AA-6300 - section 8.2.2.2 in AA-6300 Instruction Manual
Note: The user is advised to contact us if not familiar with dismounting the nebulizer
which can be hazardous if not carried out properly.
Ensure that the drainage is good to prevent the collection of liquid in the
chamber. The drain tube should be hanging freely in the air inside the drain
container, i.e. the water level does not touch the end of the tube.
Ensure that the water/solvent used in the analysis is clean. This can be checked
by injecting just water into the graphite tube. The absorbance reading obtained
should be very low.
Ensure that the sample is sufficient for the sampling nozzle; otherwise, the nozzle
cannot reach the sample and fails to suck the sample.
The graphite tube has deteriorated as sample may dissolve tube. Change the
graphite tube:
(a) AA-6601/6701 and GFA-6500 - section 4.5 in AA-6601/6701 Instruction Manual
(Users System Guide).
(b) AA-6800 and GFA-6500 - section 6.5 in AA-6800 Instruction Manual.
(c) AA-6800/6300 and GFA-EX7/EX7i section 5.1 in GFA-EX7 Instruction
Manual.
5
The graphite cap, graphite holder, window plate sockets and surfaces of window
plate sockets are contaminated. Clean these parts as:
(a) AA-6601/6701 and GFA-6500 - section 4.6 in AA-6601/6701 Instruction Manual
(Users System Guide).
(b) AA-6800 and GFA-6500 - section 6.6 in AA-6800 Instruction Manual.
(c) AA-6800/6300 and GFA-EX7/EX7i - section 5.2 and 5.3 in GFA-EX7
Instruction Manual.
The graphite cap is misaligned, so that the emission detection hole in the surface
of the cap is not properly facing the light inlet of the photosensor unit. Hence,
the amount of light reaching the temperature sensor is inaccurate, causing the
temperature of the tube to fluctuate. Ensure that the alignment is good.
The pyrolytic tube was not conditioned by heat cleaning before use. This may
cause low sensitivity and poor reproducibility during starting time. The method
for conditioning the pyrolytic tube is:
(a) AA-6601/6701 and GFA-6500 - section 2.5.4.2 in AA-6601/6701 Instruction
Manual (Users System Guide).
(b) AA-6800 and GFA-6500 - section 2.5.4.2 in AA-6800 Instruction Manual.
(c) AA-6800/6300 and GFA-EX7/EX7i - section 2.4.5.2 in GFA-EX7 Instruction
Manual.
The heating program is not suitable, e.g. if the drying temperature is set too high,
the sample boils and scatter. Set a lower drying temperature and longer time.
Some elements such as Cd and Pb atomise instantaneously as they have low
atomising temperature this causes signals to decrease. Higher sensitivity can
also be achieved if the sensitivity mode is set to HIGH.
Ensure that the sample injection is good, i.e. the autosampler pipette tip or
sampling nozzle does not touch the external or internal wall of graphite cap,
graphite holder or the surface of the graphite tube as this might cause the water
break.
The autosampler pipette tip or sampling nozzle has worn out and the volume of
injection is inaccurate. Replace the pipette tip or the sampling nozzle:
(a) ASC-6000 section 4.14 in AA-6601/6701 Instruction Manual (Users System
Guide).
6
The flow line system such as tubing, nozzle, peristaltic pump, syringe and valve
of the autosampler has deteriorated, the tubing especially, can burst after
analysing organic solvents. In order to prevent this, rinse all flow lines with
diluted water so that the organic solvent will not remain inside the peristaltic
pump.
placing a white piece of paper in the atomiser unit. If red or orange light can be seen,
then the hollow cathode lamp is lit. If purple light is viewed, then the D2 lamp is lit.
Ensure that the optical axis of the hollow cathode lamp is correct, especially in the
case of AA-6200.
(a) For flame analysis, first, ensure that the burner head position is correct. Then,
place a white paper in the optical path of the atomiser. If the beam from the
hollow cathode lamp cannot be seen above the burner head, then adjust the
optical axis of the lamp.
(b) For GFA, first, ensure that the atomiser unit position is correct. The beam from
the hollow cathode lamp can be seen if a white piece of paper is placed in front
of the left window plate socket of the GFA unit. The light beam can also be seen
on the white paper when placed on the right window socket.
The surfaces of the light source quartz window plate and monochromator quartz
window plate are dirty (at short wavelength, even adherence of fingerprints will cause
a noise). Wipe off the stain on the surfaces with soft cloth or lint-free tissue paper. If
the stain is hard to be wiped off, the soft cloth or lint-free tissue paper can be soaked
with alcohol.
The optical axis may be distorted at wavelength of 200 nm or less if the burner is
hot. Do not remove the burner head when it is hot as the sample may evaporate
from the chamber. After the burner head has cooled down, carry out the
measurement again.
For AA-6200 and AA-6800, the impact bead in the nebulizer is loose. This can
be checked by dismounting the nebulizer:
(a) AA-6200 - section 6.1.2.1 in AA-6200 Instruction Manual.
(b) AA-6800 - section 6.2.2.1 in AA-6800 Instruction Manual.
Note: The user is advised to contact us if not familiar with dismounting the nebulizer
which can be hazardous if not carried out properly.
Unstable flame - the gas supply pressure is not consistent. Ensure that the gas
pressure does not fluctuate.
Unstable flame - the quality of gases is not good. Ensure that the gas supplies are
free from water, moisture and dust.
Unstable flame - check that the burner slot is not clogged by carbide (especially
for nitrous oxide-acetylene flame) or salts. Clean the slot with thick paper or
razor. Also, the interior of the burner head might be dirty. For details:
AA-6200 - section 6.1.1 in AA-6200 Instruction Manual.
AA-6601/6701 - section 4.2.1 in AA-6601/6701 Instruction Manual (Users
System Guide).
AA-6800 - section 6.2.1 in AA-6800 Instruction Manual.
AA-6300 - section 8.2.1 in AA-6300 Instruction Manual
(a)
(b)
(c)
(d)
While aspirating water through the nebulizer, check that air-acetylene flame is
lean blue if not, the acetylene gas supply is finishing and organic solvent from
the cylinder tank is flowing to the burner head, causing the flame to turn orange.
The acetylene cylinder should be replaced if the pressure decreases to 5 kg/cm2.
For nitrous oxide-acetylene flame, the red portion height should be at least 1 cm.
Ensure that the suction force of the ventilation system is appropriate (600 1200
m3/h). Too much suction force can cause the flame to flicker as well as create
excessive noise.
Ensure that the drainage is good to prevent the collection of liquid in the
chamber. The drain tube should be hanging freely in the air inside the drain
container, i.e. the water level does not touch the end of the tube.
The graphite cap, graphite holder, window plate sockets and surfaces of window
plate sockets are contaminated. Clean these parts as:
(a) AA-6601/6701 and GFA-6500 - section 4.6 in AA-6601/6701 Instruction Manual
(Users System Guide).
(b) AA-6800 and GFA-6500 - section 6.6 in AA-6800 Instruction Manual.
(c) AA-6800/6300 and GFA-EX7/EX7i - section 5.2 and 5.3 in GFA-EX7
Instruction Manual.
The socket window plates of the atomiser might be dirty. Clean off the stain
with soft cloth or lint-free tissue paper. If the stain is hard to be wiped off, the
soft cloth or lint-free tissue paper can be soaked with alcohol.
Hence, the amount of light reaching the temperature sensor is inaccurate, causing the
temperature of the tube to fluctuate. Clean by:
(a) AA-6601/6701 and GFA-6500 - section 4.7 in AA-6601/6701 Instruction Manual
(Users System Guide).
(b) AA-6800 and GFA-6500 - section 6.7 in AA-6800 Instruction Manual.
(c) AA-6800/6300 and GFA-EX7/EX7i - section 5.4 in GFA-EX7 Instruction
Manual.
The graphite cap is misaligned, so that the emission detection hole in the surface of
the cap is not properly facing the light inlet of the photosensor unit. Hence, the
amount of light reaching the temperature sensor is inaccurate, causing the temperature
of the tube to fluctuate. Ensure that the alignment is good.
The guide rods for the right cooling block in the atomiser unit is dirty which prevents
values.
(17) For GFA, during heat atomisation, spark occurs between window plate holder
and cooling block.
Ensure that the window plate holder is securely fastened.
(18) For GFA, the atomiser unit is noisy during the heating process.
Ensure that the left and right cooling blocks are properly aligned.
Graphite tube fails to contact. Ensure that the right cooling block may slide without
difficulty.
10