The Normal Heart-Beat Cycle

How Does a Defibrillator


Work?
In essence the shock circuit in a
defibrillator has three key components: a
high voltage source, a capacitor and
switches.

The Voltage Source

Energy Considerations

Modern defibrillators use direct current


(dc) rather than the alternating current
(ac) which earlier models used. This poses a problem for designers of
battery-operated devices. Transformers cannot step up direct current.
The problem is solved as follows.

The capacitance of a capacitor is the amount of electric charge it can


store for every volt applied to it. The unit of electric charge is the coulomb
(symbol: C). The unit of capacitance is called a farad (symbol: F). One
farad = 1 coulomb per volt.

These events are triggered by the sinoatrial node a collection of


modified myocytes that acts as the hearts pacemaker.

Ventricular Fibrillation
Normally the pulse rate (heartbeat rate) is appropriate for the bodys
oxygen demand. Problems can arise if the heart rate is too low or too
high. If the speed is very high it can lead to ventricular fibrillation (VF) in
which the heart muscle quivers and does not pump efficiently, if at all.
This condition is generally fatal if not treated quickly.
Normal heart-beat can be restored by delivery of a controlled electric
shock. This process is called defibrillation.

What is a Defibrillator?
Defibrillators have been in use for
about sixty years. The earlier
machines were comparatively large
and not really portable. Emergency
portable defibrillators (also called
AEDs
or
automatic
external
defibrillators) are today available in
many public buildings, schools,
clubs etc. and small defibrillators
can now be implanted surgically in
patients with certain chronic heart
problems.

Intelligent
Defibrillators
Over recent decades there have
been major advances in the design
of defibrillators. Much of this work Implantable defibrillator
stemmed from efforts to produce
implantable devices. Modern defibrillators include sensors that can
detect the cardiac rhythm and are programmed to decide whether a
shock is required and to deliver it correctly.

A battery drives an oscillator circuit; in effect this produces a current that


is switched on and off at a high frequency (e.g. 1000 times per second),
although it is still going in one direction only. If this pulsed direct current
is fed into a suitable transformer the required output voltage can be
generated. The factor by which the voltage is stepped up is the ratio of
the number of turns on the input and output coils of the transformer. For
example, if the input coil (primary) has 200 turns and the output coli
(secondary) has 20,000 turns then the voltage is stepped up by a factor
of 100. A 5 V input would then result in a 500 V output.
The alternating output voltage is rectified by means of a diode and fed
into a capacitor which stores the high voltage charge.

The Capacitor
A capacitor consists of two flat conductors or plates (usually of
aluminium foil) with an insulator between them. A conducting lead is
attached to each plate. In practice the whole capacitor assembly is often
rolled and inserted in a can with two connections.

AND

Circuit
The diagram shows a simplified version of a defibrillator circuit. With all
switches open the paddles are attached across the patients chest. S1
is then closed in order to charge the capacitor. S1 is then opened and S2
is closed; this causes the capacitor to discharge through the patient,
hopefully restoring normal cardiac rhythm.

The cardiac cycle has two distinct phases the diastole phase and the
systole phase. In the diastole phase, the heart ventricles are relaxed and
fill with blood. In the systole phase, the ventricles contract and pump
blood around the body through a system of arteries.

SCIENCE TECHNOLOGY
ACTION SEVENTH EDITION
IN

Defibrillators and How They Work

With regard to defibrillators the amount of energy stored in a capacitor is


very important. It can be calculated as follows: E = C V2, where E is the
energy in joules, C the capacitance in farads and V the electric potential
(voltage) measured in volts. This energy, which may be more than 100 J,
is dissipated in the patients body over a small time interval (about 10
milliseconds or one hundredth of a second). For example, if the
capacitance is 1000 F (microfarad) and the voltage is 500 V then the
stored energy is 125 J [E = C V2 = (1000 106)(5002) = 125 J].
Early defibrillators delivering about 400 joules sometimes caused further
cardiac injury.

Electric Current
The electrical resistance of the skin is the main contributor to the human
bodys resistance. If the skin is dry the resistance from one hand to the
other might be over 100, 000 ohms. This is dramatically reduced if the
area of contact is large and the skin is moistened with a suitable
conductive paste or gel. The electrodes provided with AEDs are generally
self-adhesive and are pre-coated with conductive gel. They can reduce
the bodys resistance (across the chest) to about 20 ohms.
Using V=IR we can calculate the peak defibrillation current. If V = 500
volts then the current is 25 amperes. The pulse lasts only about a
hundredth of a second (10 ms) and so the risk of surface burns to the
skin is reduced.

Waveforms
As a capacitor discharges its voltage falls and so does the current
through the patient. Plotting the voltage or the current against time gives
a characteristic graph or waveform. The waveform resulting from a single
capacitor discharge is monophasic, i.e. the current is on one direction
only. Modern defibrillators are generally biphasic; successive current
pulses are in opposite directions. Biphasic defibrillators are considered
to be more effective and today virtually all new defibrillators are of this
type.

Boston Scientific is the worlds largest medical device


company dedicated to less-invasive medicine. It has more than
28,000 employees and 26 manufacturing, distribution and
technology centres delivering more than 15,000 products in
over 45 countries. For more than twenty five years, Boston
Scientific has improved the quality of patient care and the
productivity of health care delivery through the development
and advocacy of less-invasive medical devices to help clinicians
improve patient care by reducing risk, trauma, cost, procedure
time and the need for aftercare.
With corporate headquarters in Natick, Massachusetts and
major operations in North and South America, Boston Scientific
has established three key strategic sites in Ireland Clonmel,
Cork and Galway employing in excess of 3,000 employees
across the broad range of activities in which the company is
involved.
Boston Scientific in Clonmel:
The Clonmel operation is responsible for the production of
implantable pacemakers and defibrillators that offer life saving
therapy for patients suffering from cardiac arrhythmias and
heart failure. It currently employs more than 600 people, many
of whom specialise in a range of engineering disciplines e.g.
Electrical, Mechanical, Biomedical, Quality, R&D and Industrial
along with other functional areas such as Supply Chain, Human
Resources, Finance, Process Development, IT and
Manufacturing.
Boston Scientific Clonmel actively promotes and participates in
the local community through initiatives such as their Schools
Programme where BSC facilitates local students on work
experience and sponsors a Student of the Year Award in local
schools, culminating in an opportunity for summer work. BSC is
actively involved in Junior Achievement activities, with over 16
employees delivering programmes in the region.
You can find this and other lessons at www.sta.ie.

Find out more about the work of Boston Scientific at


www.bostonscientific.com.
Monophasic (left) and Biphasic (right) waveforms

Syllabus References

Student Activities

Examination Questions

Leaving Certificate Biology

Requirements: low voltage variable power supply (012 V), dc voltmeter


(010 V), capacitor (e.g. 100 F), two switches, selection of resistors (e.g.
1 k, 10 k, 100 k) and connecting leads.

Leaving Certificate Biology (HL) 2009

H.3.2.4 Heartbeat Control: An awareness of specialised heart


muscle.
Tissue and the existence and location of pacemaker nodes (SA and
AV).

Leaving Certificate Physics


Electricity (p. 32 - 38).
Electric potential, potential difference, current, circuits, resistance,
capacitance, inductance.

Learning Outcomes
On completion of this lesson, students should be able to:
Understand what is meant by the cardiac cycle and the meaning of
ventricular fibrillation.
Describe the basic operation of a defibrillator.
Outline how a defibrillator circuit works.
Outline the dangers associated with the use of defibrillators.
Calculate the amount of energy stored in a charged capacitor.
Describe the difference between monophasic and biphasic
defibrillators.

Set up the circuit shown


in the diagram. Open
both switches. Turn on
the power supply. Set
the voltage to 8 V. (If the
power supply does not
have a voltage display
then you might check it
with another voltmeter.)

The following information can be used to revise the lessons main


learning points and inform discussion.
The average annual rate of premature death in Ireland due to
cardiovascular disease (CDV) is about 52 per 100,000. Many lives
could be saved by proper use of a defibrillator.
If the heart rate becomes abnormally high the heart muscle may
produce very rapid but ineffective contractions; this condition is
called ventricular fibrillation - a condition that is generally fatal if not
treated quickly.
The main electrical components of a defibrillator are: an oscillator,
a transformer, a rectifier, a capacitor and a switch.
A capacitor stores electric charge. The unit of capacitance is a
coulomb per volt (known as a farad, symbol: F).
In monophasic defibrillators the electrical output is in one direction.
In biphasic defibrillators successive pulses are of opposite phase
(i.e. opposite electrical direction).

Write a short note on each of the following: 1. Pulse 2. Blood


pressure.

(ii)

Comment on the effect of each of the following on the


circulatory system: 1. Diet 2. Exercise.

(iii) Give two ways, other than colour, in which a red blood cell
differs in structure or composition from a typical body cell
such as one in the cheek lining.
(iv) What is the role of the SA (sinoatrial) and AV (atrioventricular)
nodes in the heart?

Close switch S1 and observe the voltmeter. You should see the voltage
rise quickly at first and then more slowly. Why do you think this
happens?

(v)

Give the precise locations of both the SA and the AV nodes in


the heart.

The capacitance of a parallel plate air capacitor is 5 pF. If the plates of


the capacitor are 2 cm apart, what is the common area of the plates?
Take air = 0.

When there is no further change in the voltage open S1 and close S2. Now
the capacitor discharges through R2. Estimate how long it takes for the
voltage to drop to half its value (8 V to 4 V to 2 V to 1 V).

What is the positive charge stored on a 5 F capacitor when it is


connected to a 120 V d.c. supply?

Leaving Certificate Physics (HL) 2010, Q. 5(e)

Leaving Certificate Physics (HL) 2005, Q. 11b

How does this time relate to the value of the variables (C1, R2)?.

Read the following passage and answer the accompanying questions.

1. The cardiac cycle has two distinct phases


the diastole phase and the systole phase.

T F

2. The cardiac cycle is triggered by the sinoatrial node.

T F

3. Normally the heart rate is independent of the bodys


oxygen requirement.

T F

4. Ventricular fibrillation occurs when the heart rate is too low. T F


5. Defibrillation involves the delivery of a controlled
electric shock.

T F

8. The capacitance of a capacitor is inversely proportional


to the area of the plates.

T F

9. The energy delivered by the shock from a typical defibrillator


is about 125 joules.
T F

Check your answers to these questions on www.sta.ie.

(Adapted from Milestones of Science; Curt Supple; 2000)


(i)

7. A transformer is used to convert ac to dc.

10. Defibrillators are also known as AEDs.

The scientist whose research would unite electricity and magnetism


was Michael Faraday. He developed the first electric motor in 1821,
showing that a current carrying conductor could be made to revolve
around a magnet. He went on to expand on Oersteds observation that
an electric current produces a magnetic effect. Perhaps, Faraday
thought, the opposite was also true: a moving magnetic field could
generate an electric current. This was to be called electromagnetic
induction. Soon he had created the first electric generator, and
everyday life would never be the same again. His experiments with
induced currents produced the transformer.

T F

6. Sensors in modern defibrillators can detect the cardiac rhythm


and can deliver the most appropriate shock, if required.
T F

T F

AND

Did You Know?


In 1842 John Erichsen first described the condition known as
defibrillation. He found that a dogs heart went in fibrillation when its
coronary artery was closed off. In 1849 Carl Ludwig and M. Hoffa
showed that an electric shock could induce fibrillation.
In 1889 John A. MacWilliam published a more detailed study of the
condition. He realised that people died as a result of fibrillation; at the
time there was no known way to restore the heart to a normal rhythm.
In 1899 Jean Louis Prevost and Frederic Batelli showed that an electric
shock could restore a fibrillating dogs heart to normal rhythm.
In the 1920s devices were invented to measure the electrical activity of
the heart (electrocardiograms). Dr Carl J. Wiggers identified the different
stages in ventricular fibrillation and greatly improved defibrillation
techniques. In 1947 the first successful defibrillation of a human heart
was carried out by Claude Shaeffer Beck with advice from Wiggers.

Leaving Certificate Physics (HL) 2011, Q. 5(e)

The capacitor stores electric charge and as it does the voltage across it
rises. As a result the voltage difference across R1 drops and so does the
rate of flow of charge through it. Therefore the charging of the capacitor
becomes progressively slower.

True/False Questions
General Learning Points

(i)

SCIENCE TECHNOLOGY
ACTION SEVENTH EDITION
IN

Defibrillators and How They Work

(ii)

List three factors that affect the force on a current-carrying


conductor placed near a magnet.

Claude Schaeffer Beck (18941971)


Claude Beck was the first surgeon to
successfully carry out defibrillation on a human
patient. (As mentioned above, others had
performed defibrillation on dogs.)
He had previously (in the 1930s) developed a
technique to treat heart failure in which a piece
of pectoral muscle (chest muscle) was
implanted into the pericardium a type of sac
that surrounds the heart; this technique is no
longer used.
The also pioneered (in 1935) a procedure called cardiopericardiopexy,
in which a vein was inserted between the aorta and the coronary sinus
a major vein near the surface of the heart.
Towards the end of an operation on the heart of a fourteen-year-old boy
in 1947 the boys heart stopped. Beck reopened the boys chest and
tried unsuccessfully for forty five minutes to restart the heart by
massaging it. He then used a defibrillator and normal heartbeat was
restored. The boy went on to make a full recovery.
Claude Beck died of a stroke in 1971.

What energy transformation takes place in an electric motor?

(iii) What is the function of a commutator in a dc motor?


(iv) Draw a sketch of the output voltage from an ac generator.
(v)

Biographical Notes

Revise The Terms

How are the slip rings connected to an external circuit in an ac


generator?

Can you recall the meaning of the following terms? Reviewing


terminology is a powerful aid to recall and retention.

(vi) A transformer and an induction coil can both be used to


change a small voltage into a larger voltage. What is the basic
difference in the operation of these two devices?

AED, alternating, arteries, biphasic, capacitor, cardiac cycle,


cardiovascular disease (CDV), chronic, coulomb, defibrillation, diode,
electric charge, farad, heart attack, implantable, insulator, monophasic,
myocytes, oscillator circuit, premature, ratio, rectified, sinoatrial node,
transformer, ventricles, ventricular fibrillation (VF).

(vii) Name the Irish physicist who invented the induction coil.
(viii) Give two factors that affect the efficiency of a transformer.

Check the Glossary of Terms for this lesson on www.sta.ie.

You might also like