5 KW VENUS SERIES HIGH FREQUENCY MAMMOGRAPHY SYSTEM

ALLENGERSMAMVENUS
(Highfrequency(40KHz)mammographyunit)

SpecialFeatures

A)

B)

AutomaticfilterselectionwithKV.
Automatic release of compression with termination of
exposureincaseofpowerfailure.
DisplayofmAsinAECmodeafterexposure.

XRAYGENERATOR
Xraygenerator

Highfrequency40KHz.xraygenerator

Maximumpower

3.5KW*

MaximummAoutput

70mA

MaximumKVoutput

35KV

Highvoltagecable

1No.

XRAYTUBE
Xraytube

Rotatinganodexraytube.

Focalspot
Targetmaterial

Smallfocus=0.1mm

Largefocus=0.3mm.
Molybdenum.

Tubeangle

12

Inherentfiltration.

Ba0.5mm

Anodeheatstoragecapacity

300KHU

Anoderotation

2700RPM

C) CONTROLPANEL
Controlpanel

Micro processor controlled feather touch control panel


withLCD

KVrange

22to35KVinstepsof0.5KVeach.

mAsrange

1mAsto700mAs

Techniqueselection

Manualtwopointtechnique(KV,mAs).

Anatomicprogrammedradiography

Auto selection of exposure factors with APR for


small,mediumandlargebreasts.

Filmscreencombination

5filmscreencombinationsareprovided.

Automaticexposurecontrol(AEC)

InbuiltthreechamberAECforbetterexposurecontrol
anddiagnosis

Filmdensitycontrol

5stepfilmdensitycontrolisprovided

Filterselection

Digitaldisplays

FilterisautomaticallyselectedaspertheKVselected
(molybdenumfilter(for29.5to35KV)andaluminium
filter(for22to29KV)
LCDdisplayoncontrolpanel
x
KV
x
mAs
x
Interlocksindicatingthefaultinthemachine
x
Filterselected
LCDdisplayonthestand
x
CompressionforceinKg
x
Compressedbreastthickness
x
Degreescale

Switches

Indicators
D)

Focalspotselectionswitch.

Machineon/offswitch.

Xrayswitch.

Emergency stop switches mushroom type


emergency switch provided on both sides of C.
This switch releases the breast by moving the
compressionpaddleincaseofanypowerfailure

XrayONindicator.

STANDASSEMBLY
Acompactstandonwhichcarmcontainingxraytubeandbuckyassemblyismounted.
Verticalmovement(motoroperated)

550mm

Angularmovementofxraytubeand 180
bucky
Focusfilmdistance

650mm.

Breastcompression

Motorised compression upto 15Kg with the switch for


activationandrelease.Adjustablecompressionforce.
Compression paddles for normal and magnification mode
(spotcompression).Compressionscale:0to10cm.

Compressionpaddles

E)

Magnificationdevice

1.8X

Bucky

18x24cmbucky,motoroperatedoscillatinggridofsize
18X26cm,5:1,30lines/cmfocaldistance60to70cm.

Collimator

Light beam collimator with halogen lamp with auto shut


offtimer

Collimationplate.

18X24cmcollimationplate.

Cone

Coneforlocalizationandradiationprotection.

AutomaticExposureControl(AEC)

InbuiltthreechamberAECforbetterexposurecontroland
diagnosis

CArmmovement.

Switches for up/down movement of C Arm, placed


convenientlyonbothsidesofCArm.

Exposureswitch

Hand switch with retractable cord for initiation of


exposure.

Filmmarkingdevice.

Filmmarkingdevice.Alphanumericidentificationsystem.

Protectivebarrierforoperator

Free standing fully transparent lead glass screen for


operatorprotection.

POWERSUPPLYREQUIREMENT
PowerSupply

Singlephase,230volts10%,AC,50Hz,15amps.with
independentearthingonthewallsocket.

OPTIONALACCESSORIES&DEVICES.
x

24x30cmbuckywithgrid(motoroperated).

24x30cmcollimationplateandcompressionpaddle.

Rhodiumfilter.

Stereotaticbiopsysystem

ForALLENGERSMEDICALSYSTEMSLTD.