Professional Documents
Culture Documents
ALLENGERSMAMVENUS
(Highfrequency(40KHz)mammographyunit)
SpecialFeatures
A)
B)
AutomaticfilterselectionwithKV.
Automatic release of compression with termination of
exposureincaseofpowerfailure.
DisplayofmAsinAECmodeafterexposure.
XRAYGENERATOR
Xraygenerator
Highfrequency40KHz.xraygenerator
Maximumpower
3.5KW*
MaximummAoutput
70mA
MaximumKVoutput
35KV
Highvoltagecable
1No.
XRAYTUBE
Xraytube
Rotatinganodexraytube.
Focalspot
Targetmaterial
Smallfocus=0.1mm
Largefocus=0.3mm.
Molybdenum.
Tubeangle
12
Inherentfiltration.
Ba0.5mm
Anodeheatstoragecapacity
300KHU
Anoderotation
2700RPM
C) CONTROLPANEL
Controlpanel
KVrange
22to35KVinstepsof0.5KVeach.
mAsrange
1mAsto700mAs
Techniqueselection
Manualtwopointtechnique(KV,mAs).
Anatomicprogrammedradiography
Filmscreencombination
5filmscreencombinationsareprovided.
Automaticexposurecontrol(AEC)
InbuiltthreechamberAECforbetterexposurecontrol
anddiagnosis
Filmdensitycontrol
5stepfilmdensitycontrolisprovided
Filterselection
Digitaldisplays
FilterisautomaticallyselectedaspertheKVselected
(molybdenumfilter(for29.5to35KV)andaluminium
filter(for22to29KV)
LCDdisplayoncontrolpanel
x
KV
x
mAs
x
Interlocksindicatingthefaultinthemachine
x
Filterselected
LCDdisplayonthestand
x
CompressionforceinKg
x
Compressedbreastthickness
x
Degreescale
Switches
Indicators
D)
Focalspotselectionswitch.
Machineon/offswitch.
Xrayswitch.
XrayONindicator.
STANDASSEMBLY
Acompactstandonwhichcarmcontainingxraytubeandbuckyassemblyismounted.
Verticalmovement(motoroperated)
550mm
Angularmovementofxraytubeand 180
bucky
Focusfilmdistance
650mm.
Breastcompression
Compressionpaddles
E)
Magnificationdevice
1.8X
Bucky
18x24cmbucky,motoroperatedoscillatinggridofsize
18X26cm,5:1,30lines/cmfocaldistance60to70cm.
Collimator
Collimationplate.
18X24cmcollimationplate.
Cone
Coneforlocalizationandradiationprotection.
AutomaticExposureControl(AEC)
InbuiltthreechamberAECforbetterexposurecontroland
diagnosis
CArmmovement.
Exposureswitch
Filmmarkingdevice.
Filmmarkingdevice.Alphanumericidentificationsystem.
Protectivebarrierforoperator
POWERSUPPLYREQUIREMENT
PowerSupply
Singlephase,230volts10%,AC,50Hz,15amps.with
independentearthingonthewallsocket.
OPTIONALACCESSORIES&DEVICES.
x
24x30cmbuckywithgrid(motoroperated).
24x30cmcollimationplateandcompressionpaddle.
Rhodiumfilter.
Stereotaticbiopsysystem
ForALLENGERSMEDICALSYSTEMSLTD.
MEDILINES LIFECARE TECHNOLOGY
AUTHORISEDSIGNATORY