Professional Documents
Culture Documents
an appropriate place and has to be always accessible and available for the
personnel!
Instruction Manual
Technical Description
General information for installation, start-up, and measuring mode
Software:V1.035
Version:V1.35_R0
e-mail: info@ams-dielheim.com
web-site: www.ams-dielheim.com
10/2014
Instruction Manual
RHADOX AMS 7300
ATEX Ex II 3G IIB+H2 T3
Page 2
Table of Contents
page
1. INTRODUCTION .......................................................................................................................................... 3
1.1 PHYSICAL PRINCIPLES OF THE ANALYSER RHADOX 7300 ......................................................................... 3
1.2 MEASURING PRINCIPLES ........................................................................................................................... 5
1.3 AS BUILT .................................................................................................................................................. 7
1.4 TECHNICAL DATA RHADOX 7300............................................................................................................. 8
2. INFORMATION FOR THE OPERATOR .................................................................................................... 10
2.1 HOW TO USE THE INSTRUCTION MANUAL.................................................................................................. 10
2.2 APPROVED USE ...................................................................................................................................... 11
2.3 SAFETY INFORMATION............................................................................................................................. 12
3. RELEVANT INFORMATION FOR INSTRUMENTS PROTECTED AGAINST EXPLOSIONS ................. 13
3.1 IGNITION PROTECTION TYPE EEX-P (PRESSURISED ENCLOSURE) .............................................................. 13
3.2 GENERAL INFORMATION .......................................................................................................................... 15
3.3 PURGE UNIT F840 / BY GNNHEIMER .................................................................................................. 15
3.4 CATALYTIC COMBUSTION OF BYPASS-GAS................................................................................................ 16
3.5 IN CASE OF PURGE AIR FAILURE .............................................................................................................. 16
4. INSTALLATION AND START-UP ............................................................................................................. 17
4.1 TECHNICAL DRAWINGS RHADOX 7300 .................................................................................................. 17
4.2 ELECTRIC CONNECTIONS ........................................................................................................................ 17
4.2.1 Supply Voltage ............................................................................................................................ 17
4.2.2 Signal Outputs ............................................................................................................................ 18
4.3 PNEUMATIC CONNECTIONS ..................................................................................................................... 19
4.3.1 Gas Inlets .................................................................................................................................... 19
4.3.2 Gas Outlets ................................................................................................................................. 20
5. START-UP ................................................................................................................................................. 21
5.1 GENERAL OPERATION AND OPERATION/SERVICE CODES.......................................................................... 21
5.2 STARTING THE ANALYSER ....................................................................................................................... 23
5.3 TEMPERATURE MONITORING OF THE PNEUMATIC UNIT (*OPTION) ............................................................. 24
5.3.1 Setting the Cabinet Temperature.................................................................................................. 24
5.4 MEASURING MODE ................................................................................................................................. 24
6. SETTING THE USER PARAMETERS ...................................................................................................... 25
6.1 ANALYSER OPERATION ........................................................................................................................... 25
6.2 SETTING THE SECOND LINE..................................................................................................................... 26
6.3 SUBMENU DIAGNOSIS ............................................................................................................................. 27
6.4 SUBMENU PARAMETER ........................................................................................................................... 27
6.5 SUBMENU ANALOGUE OUT ...................................................................................................................... 28
6.6 SUBMENU CALIBRATION .......................................................................................................................... 29
6.6.1 Adjusting of Auto-calibration ......................................................................................................... 33
6.6.2 Start of Auto-calibration ................................................................................................................ 33
6.6.3 Calibration Errors .......................................................................................................................... 33
6.7 SUBMENU SERVICE................................................................................................................................. 34
6.7.1 Fixed Density (*Option)................................................................................................................ 37
6.7.2 Service Request .......................................................................................................................... 37
6.7.3 Cabinet Temperature (*Option) .................................................................................................... 38
6.7.4 Expert Mode.................................................................................................................................. 38
7. SERIAL INTERFACE ................................................................................................................................. 39
8. ERROR MESSAGES ................................................................................................................................. 40
I. GAS ALARM (*OPTION) ............................................................................................................................ 41
II. MEASURING CABINET COOLER (*OPTION) ......................................................................................... 42
III. SERVICE SOCKET (*OPTION) ................................................................................................................ 42
IV. EXTERNAL CATALYTIC COMBUSTION OF BY-PASS GASES (*OPTION) ........................................ 43
V. INSIDE CATALYTIC COMBUSTION OF BY-PASS GASES (*OPTION) ................................................ 44
Version: V1.35_R0
10/2014
Instruction Manual
RHADOX AMS 7300
ATEX Ex II 3G IIB+H2 T3
Page 3
1. Introduction
1.1 Physical Principles of the Analyser RHADOX 7300
Due to environmental protection and as a cost-saving alternative to natural gas fuelling, more and
more gases used in industrial combustion processes that produced in other operations, such as in
coking plants. However, the composition of these exhaust gases is often subject to strong
variations due to their generating process.
On the other hand, a constant energy supply at the combustion of the off-gases is an economic
necessity and minimizing the emission desired. The reliable and continuous measuring of the air
demand and the Wobbe index of the gas used is necessary in order to adjust the combustion
process to the changing gas qualities and to the specific operating conditions.
Field of Application
For the production of synthetic gases, even with a very low Wobbe index
Systems Characteristics
RHADOX - measuring systems for air demand, Wobbe index, and calorific value,
specific gravity
Customized systems
Burner or catalytic reactor for adjusting to the measuring gas quality
Calibrateable with the PTB certification (special model)
Automatic and manual 2-point calibration
ZrO2 sensor for the oxygen measurement
menu-driven parametrizing
Diagnostic system for the instrument functions and alarms
Continuous operating mode
Version: V1.35_R0
10/2014
Instruction Manual
RHADOX AMS 7300
ATEX Ex II 3G IIB+H2 T3
Page 4
RHADOX
7300
The RHADOX 7300 is a compact system for the continuous and reproducible determination of
the air demand at a total combustion of a gas. The instrument built into an insulated wall cabinet.
The mixing cell consists of two separated gas chambers, i.e. the mixing chamber and the fitting
chamber. This mixing cell plays a key part in the homogenous and reproducible process of the
sample gas with the combustion air. Due to built-in gas jets (Venturi-jets) for the probe gas and
the combustion air, the simple measuring method makes use of the law of Bernoulli (see
measuring principles).
Figure 1.1:
RHADOX 7300
Basic Gas Flow Diagram
From the mixing chamber, the gas/air mixture flows to the fitting chamber, which might design as
a supersonic jet or a flame barrier (depending on the application involved). Hence, a reaction from
the burner on the performance of the mixing chamber has prevented and the ultimate mixing of
the two gases will achieved.
Gas and air supplied to the corresponding apertures located in the mixing cell.
The electronic controller ensures that the pressure of both gases in the two separated gas
volumes is equal and constant (controller accuracy + 0,01 mbar). For this purpose, a pressure
controller for the combustion air as well as a differential pressure transmitter is used in the
instrument.
The front panel features two pressure meters: the controller to adjust the correct pressure of the
combustion air and a meter for the sample gas pressure. The latter will adjust to the correct air
pressure via the electronic controller.
Programming and operating the instrument needs only three push buttons: one for the operation
mode (measurement, calibration, and signal outputs), two for the cursor. A display with two LCD
rows indicates the corresponding values and user fed data (alarms, residual Oxygen). A key
operated switch locks the programming level if required.
Version: V1.35_R0
10/2014
Instruction Manual
RHADOX AMS 7300
ATEX Ex II 3G IIB+H2 T3
Page 5
Using the semi-automatic calibration mode a given and correct value compared to the
corresponding measuring value of the test gas via the electronic controller. Deviations show on
the display.
If a two-point calibration is need, a second calibration point can obtain via an auxiliary airflow to
dilute the reference gas properly. Of course, also two different reference gases can use to cover
the measuring ranges. The actual calibration mode depends on the specific AMS application.
With: p
= A * v = A * 2 * p / p = Const.
(1)
p = [1 + f (Re)] + * v 2 / 2
(2)
Re = . v . D/
(3)
D
V
A
Apart from temperature and pressure, the volume flow depends only on the density and the
viscosity of the gas considered.
For Re < 2000 a laminar gas flow can be assumed, the function f(Re) tends to zero. This
prescription will met through the corresponding layout of the gas jets for measuring gas and air as
well as by maintaining pressure and temperature of both the components sufficiently constant and
equal:
p
T
p
T
~
~
~
~
0.2 mbar
0.25 C
100 mbar
2 C
The dimensions of the aperture(s) in the mixing chamber in order to achieve Reynolds number
following eq. (3) are set during the application at AMS GmbH.
Version: V1.35_R0
10/2014
Instruction Manual
RHADOX AMS 7300
ATEX Ex II 3G IIB+H2 T3
Page 6
(4)
rel.
(5)
(6)
(7)
The instrument computer can determine the calorific value of a gas with the help of an additional,
separately installed density measurement.
There are two systems for the oxidation of a measure gas, a burner with a flame resp. with a
catalytic reactor. If the sample gas cannot burn in a stable flame due to its physical or chemical
characteristics, the catalytic reactor developed by AMS will installed instead of the burner.
The reaction chamber will fill and heat with a noble metal functioning as catalytic converter.
Version: V1.35_R0
10/2014
Instruction Manual
RHADOX AMS 7300
ATEX Ex II 3G IIB+H2 T3
Page 7
1.3 As built
The analyser consists of the following main components:
1. two-part housing,
Left side Pneumatic for gas mixing and analyse
Right side power supply, electronic, evaluation
2. Electronics, wiring terminal, Display und Functional keys
3. Solenoid Valve(s) for Gas shut-off / change
4. Pressure reducer for Instrument Air / Process Gas
5. Oxygen analyse in raw Gas (*Option)
6. Analogue solenoid Valve and Gas mixing with Pressure monitoring
7. Measuring catalyst followed by residual oxygen analysis
8. Density measuring cellblock
9. Catalytic combustion for bypass gas (*Option)
10. Purge Gas Unit for Ex-Zone 2
Version: V1.35_R0
10/2014
Instruction Manual
RHADOX AMS 7300
ATEX Ex II 3G IIB+H2 T3
Page 8
110230VAC / 50/60Hz
Power Consumption
max. 1000 VA
Ambient Temperature
-5 up to +45C
Influence of Temperature
20 up to 40C: 0,04% /C
Display
Analogue Outputs
Measuring Ranges(s)
Accuracy
Repeatability
Wobbe index
Air Demand
Gas Density
Calorific Value
Noise
Drift
Response Times
Wobbe Index and
Air Demand
T90%
:10-40 sec.
depending by application
T90%
: 10-40 sec.
depending by application
Version: V1.35_R0
10/2014
Instruction Manual
RHADOX AMS 7300
ATEX Ex II 3G IIB+H2 T3
Page 9
Analyser Messages
Analyser State
1x
3x
(YES / NO)
(YES / NO)
(YES / NO)
Serial Interface
Calibration
Dimensions
W x H x D ca. 900*1.000*350 mm
Weight
ca. 150 kg
:
:
:
Flow Rates
Instrument Air
Measuring Gas
Calibration Gas
ATEX Ex II 3G IIB+H2 T3
Ambient Temperature
+ 5 Temp + 55 C
Housing protection
IP65
Purge gas
2 / 6 bar (g)
Version: V1.35_R0
10/2014
Instruction Manual
RHADOX AMS 7300
ATEX Ex II 3G IIB+H2 T3
Page 10
Version: V1.35_R0
10/2014
Instruction Manual
RHADOX AMS 7300
ATEX Ex II 3G IIB+H2 T3
Page 11
Warning
While opening the housing of the analyser please observe that certain
parts of this device may carry dangerous voltages. Therefore, only
authorized and qualified personnel may open the instrument.
Note
The conformity declaration of this instrument must renew if the operator,
e.g. using a different sensor type, has carried out modifications in this
instrument.
Version: V1.35_R0
10/2014
Instruction Manual
RHADOX AMS 7300
ATEX Ex II 3G IIB+H2 T3
Page 12
Caution with a warning triangle means that slight personal injury can
occur if the appropriate safety precautions has not observed.
Version: V1.35_R0
10/2014
Instruction Manual
RHADOX AMS 7300
ATEX Ex II 3G IIB+H2 T3
Page 13
Version: V1.35_R0
10/2014
Instruction Manual
RHADOX AMS 7300
ATEX Ex II 3G IIB+H2 T3
Page 14
Please comply all of these technical formalities for the containment system:
Gas inlet pressure: max 5 bar (g)
Air inlet pressure: max. 5 bar (g)
Flow rates during operations
Air: app. 250 - 500 l/h
Gas: app. 50 - 150 l/h
For assessing and establishing, a tight containment considered the norm EN 1127-1 approach. As
permanently sealed containment, devices apply if:
I. They are constructive in such a way, that they remain tight due to technical, or
II. Their technical tightness is ensured by servicing / maintenance and monitoring
Please use only trained and qualified personal to check for leaks in the analyser suitable use in
hazardous areas.
The service and maintenance of the analyser includes the following measures:
1. Visual inspection of the equipment and check for savour, noise due
2. Review of devices with mobile leak detectors and portable gas detection devices
3. Overpressure test of the entire system at 1.5 times of the normal operating pressure
The operator of the system, at least monthly for point 1, shall determine the frequency of
testing. and 2., and quarterly for testing the entire system.
Version: V1.35_R0
10/2014
Instruction Manual
RHADOX AMS 7300
ATEX Ex II 3G IIB+H2 T3
Page 15
Warning The housings of the measuring system (purge unit and calorific
value analyser) may only have opened in a voltage-free state.
Warning If there is explosive dust in the ambient you have to clean the
interior of the housings before closing the lids.
After you have carefully read, the manual you may re-set all values. However, if the analyser used
in the Ex-Zone 2 please make sure that there is no hazardous atmosphere, which might lead to an
explosion.
The measures, which will take if the purge unit and/or purge air fail, are the sole responsibility of
the operator. These should be in accordance with the general regulations of the plant operator.
The purge unit of the analyser for calorific value AMS 7300 protects the analyser via instrument air
against intruding hazardous atmosphere. Further functions, such as shutting down in case of
failing purge air, has not intended.
We again refer to the safety concept of the plant operator. The operator has to guarantee the safe
operation and has to arrange in case of failing instrument air to protect the personnel and the plant
against explosions.
Version: V1.35_R0
10/2014
Instruction Manual
RHADOX AMS 7300
ATEX Ex II 3G IIB+H2 T3
Page 16
Version: V1.35_R0
10/2014
Instruction Manual
RHADOX AMS 7300
ATEX Ex II 3G IIB+H2 T3
Page 17
Danger Only trained personnel may connect the power supply line
since an inappropriate connection may harm the personnel and the
systems. Under certain conditions, the internal parts may carry
hazardous voltages.
Note Always operate the instrument after applying instrument air. This
allows ventilating the catalyst and the optional catalytic combustion and
diminishes the preheat sequence of a cool catalyst as well as the
heating time of the complete system.
Version: V1.35_R0
10/2014
Instruction Manual
RHADOX AMS 7300
ATEX Ex II 3G IIB+H2 T3
Page 18
Assignment
Description
01 02
Analogue Output 1
03 04
Analogue Output 2
05 06
Analogue Output 3
07 - 08
Analogue Output 4
Assignment
09
10
11
RS232
Description
RXD
TXD
GnD
Assignment
Switch
Position
12 - 13
SYSTEM
STATUS
12 - 13
12 - 14
Service Mode
open
12 - 14
12 - 15
Calibration
open
12 - 15
12 - 16
Service Request
open
12 - 16
Measuring Mode
Service Mode
[NO]
Calibration inactive
Calibration active
[NO]
Message inactive
Service request
[NO]
17 -19
Relay 5
20 - 22
Relay 6
Version: V1.35_R0
Description
10/2014
Instruction Manual
RHADOX AMS 7300
ATEX Ex II 3G IIB+H2 T3
Page 19
"Instrument Air"
2. Sample gas
"Process Gas"
3. Calibration Gases
Version: V1.35_R0
10/2014
Instruction Manual
RHADOX AMS 7300
ATEX Ex II 3G IIB+H2 T3
Page 20
4. "Bypass" combustor
Danger This is an inflammable gas/air mixture (ca. 300 Nl/hr)
The by-pass gas has to be disposed of via a back-pressure-free and safe conduit!
Disposal
We recommend leading off the by-pass gas via a stainless steel conduit of 1-2 m with the
appropriate connection size of the lines. If the by-pas gas is lead off via a longer conduit, the
diameter of the lines has to be enlarged.
5. Reference air + purge gas OUT "PurgeGas"
The reference air and purge gas may be discharged freely to the ambient atmosphere.
6. Off-gas "Exhaust/Condensate"
Lead the off-gas away via a continually sloping 1 - 5 m conduit or hose. If the off-gas is lead
off via a longer conduit the diameter of the line has to be enlarged accordingly. The off-gas
conduit leads away max. 20 Nl/hr of a combusted water-saturated gas mixture. On falling
below the dew point, the resulting water has to be disposed of to the rule. The off-gas
temperature at the exhaust fitting of the analyser is below 60 C.
While installing the off-gas conduits make sure that the condensed water can flow off
freely. A congested off-gas conduit can shut down the complete instrument.
Remark:
If you want to install an upward off-gas conduit you have to connect a condense separator
with a corresponding condensate separation at the analysers exit.
Version: V1.35_R0
10/2014
Instruction Manual
RHADOX AMS 7300
ATEX Ex II 3G IIB+H2 T3
Page 21
5. Start-up
Note Before the start-up, all electrical and pneumatic connections have to
connect according to their use. Make sure all pneumatic lines are gas-tight.
Note The inlet valve closed during the preheating time of the pneumatic
unit, of the catalyst, and of the ZrO2 sensor. Please set the gas inlet
pressures to the correct values after all components have reached their
operating temperature and after the main valve has opened. The required
pressure values noted in the test report and next to the manometers.
Note The display will light as soon as you press any key. The lightning will
go out if no key has pressed during the next 20 sec.
Version: V1.35_R0
10/2014
Instruction Manual
RHADOX AMS 7300
ATEX Ex II 3G IIB+H2 T3
Page 22
Via pressing the F-key, you will reach the menu from which you can choose by pressing the F-key
and/or the arrow keys. Sensitive data have protected by codes in the submenu.
1. Operation Code
111
2. Service Code
1111
The chapter submenu outlines when to enter which code. The entered code is valid until you
switch back into the measuring mode.
Software pattern:
Measurement Mode
Menu
Submenu
Measure
Diagnosis
F-key
Parameter
F-key
Analogue Out
F-key
Calibration
F-key
Service
F-key
Note Via pressing the F-key you will get from the measurement mode to the
menu. Now you can select the corresponding submenu via the arrow keys.
After you have quit the submenu by pressing return, select Measure and
you will reach the measurement mode.
Note If you do not press any key for ca. 250 sec. the analyser automatically
switches back to the measurement mode regardless of the menu level.
Version: V1.35_R0
10/2014
Instruction Manual
RHADOX AMS 7300
ATEX Ex II 3G IIB+H2 T3
Page 23
Note The RHADOX 7300 only reaches its normal operating status if all
gases (air, process gas) has connected to the corresponding gas lines.
Please refer to # 5 of the instruction manual!
Note The inlet valve has shut off during the preheating time of the
measuring catalyst and of the ZrO2 sensor. Only after both components
have reached their operating temperature and the main valve is open,
please set the gas pressures of the inlets to the correct values. These
required pressure values can be found in the test report and are marked
next to the manometer.
Display
Description
After powering on the analyser, the start menu displayed.
Among others, the system identification and serial
number has shown.
AMS 7300
Ser.Nr.
Warmup
##,## h
Catalyst Temp.
### C
Startup
Checking
##,## min
Version: V1.35_R0
10/2014
Instruction Manual
RHADOX AMS 7300
ATEX Ex II 3G IIB+H2 T3
Page 24
Version: V1.35_R0
10/2014
Instruction Manual
RHADOX AMS 7300
ATEX Ex II 3G IIB+H2 T3
Page 25
Main Menu
Display
Description
Measure
Diagnose
Parameter
Analog Output
Calibration
Service
Version: V1.35_R0
10/2014
Instruction Manual
RHADOX AMS 7300
ATEX Ex II 3G IIB+H2 T3
Page 26
resp.
.
List of choices for the second line
Display
SG
CARI
Description
Specific Gravity (*Option)
Air demand
Wobbe - Index
O2Mix [%]
O2Gas [%]
O2ZrO2 [%]
dp [mbar]
CAT AB [C]
CAT [C]
Version: V1.35_R0
10/2014
Instruction Manual
RHADOX AMS 7300
ATEX Ex II 3G IIB+H2 T3
Page 27
Note Errors cannot delete as long as their reasons have not been
resolved. Only the message Service Request may delete via the F-key.
Second line
Display
Description
No error message
NO ERROR
Note During measuring, the upper message has indicated. Possible error
messages have described in the corresponding error message chapter.
Version: V1.35_R0
10/2014
Instruction Manual
RHADOX AMS 7300
ATEX Ex II 3G IIB+H2 T3
Page 28
Description
Back to the main menu
extended parameter
A1 ####
A1 420 mA / 0...20 mA
Please select from the list below which measuring value should
be assigned to the analogue output, e.g.
Wi Wobbe Index MJ/m
LCV Lower calorific value[MJ/m] / [MJ/kg] / [kWh/m] / [kcal/kg]
LHV Lower heating value[kcal/kg]
O2 Gas [Vol%] Oxygen concentration in the raw gas
Via the F-key you can choose the limits at the output:
420 mA
420 mA ( E ), user defined constant current in case of error
020 mA
020 mA ( E ), user defined constant current in case of error
020 mA ( C ), with adjustment [value has to be programmed]
OFF output is turned off
frozen, output is frozen to the latest value
Const. I, constant current at the output
Analog Min
Version: V1.35_R0
10/2014
Instruction Manual
RHADOX AMS 7300
ATEX Ex II 3G IIB+H2 T3
Page 29
Analog Max
Const. Curr
Error Current
Version: V1.35_R0
10/2014
Instruction Manual
RHADOX AMS 7300
ATEX Ex II 3G IIB+H2 T3
Page 30
Submenu Calibration
Display
return
Wi LO POINT
Wi HI POINT
SG LO POINT
SG HI POINT
Description
Return to the main menu
CV Low Cal.
CV High Cal.
Submenu Extended
Parameter
return
Set the purge time which
starts and terminates each
calibration
Choose between active and
frozen, i.e. analogue out
status during calibration
Suppresses the monitoring
during the calibration; only
perform if directed
Version: V1.35_R0
Purge Time
Analog Out
Service Cal.
10/2014
Instruction Manual
RHADOX AMS 7300
ATEX Ex II 3G IIB+H2 T3
OFF
Auto-calibration is switched off
Low
Auto-calibration starts with
calibration of the lower
calibration point
High
Auto-calibration starts with
calibration of the higher
calibration point
Low-High
Auto-calibration starts with
calibration of the lower and
higher calibration point
High-Low
Auto-calibration starts with
calibration of the higher and
lower calibration point
Page 31
O2 Low Gas
O2 High Gas
Lo Gas Calib.
Version: V1.35_R0
Hi Gas Calib.
10/2014
Instruction Manual
RHADOX AMS 7300
ATEX Ex II 3G IIB+H2 T3
Page 32
Auto-calibration
mode
Auto-calibration
time interval
Version: V1.35_R0
10/2014
Instruction Manual
RHADOX AMS 7300
ATEX Ex II 3G IIB+H2 T3
Page 33
Cal unstable
Version: V1.35_R0
Description
Calibration point and actual measuring value too distant,
indicating wrong cal gas
Troubleshooting: apply correct cal gas, check calibration
parameters
Fluctuations during sampling sequence are too high, no
calibration possible
Troubleshooting: prolong purge sequence, prevent
pressure fluctuations
10/2014
Instruction Manual
RHADOX AMS 7300
ATEX Ex II 3G IIB+H2 T3
Page 34
Submenu Service
Display
return
O2 Air Offset
O2 Air Span
O2 Mix Offset
O2 Mix Span
O2 Gas Offset
O2 Gas Span
Description
Return to the main menu
ZrO2 Offset
ZrO2 Span
Version: V1.35_R0
10/2014
Instruction Manual
RHADOX AMS 7300
ATEX Ex II 3G IIB+H2 T3
Relay Set.
Page 35
return
Service Relay
is not failsafe / is failsafe
Calib. Relay
is not failsafe / is failsafe
S.Req. Relay
is not failsafe / is failsafe
Reserve Relay 5
Relay 5
is not failsafe / is failsafe
Reserve Relay 6
Relay 6
is not failsafe / is failsafe
Service Request
Parameter
Cab. Temp
Submenu Cabinet
Temperature (*Option)
Return to the service menu via
the F-key
return
LM 35 Offset (normally = 0)
Version: V1.35_R0
10/2014
Instruction Manual
RHADOX AMS 7300
ATEX Ex II 3G IIB+H2 T3
Page 36
Check CT ON/OFF
Cab. Temp. LO
Cab. Temp. HI
Expert Mode
Restart
Wi ##,## MJ / m
Service Request
Note The relay Service Request 12 -16 is active according to the adjusted
operation; please refer to # 7.6 of the manual.
Note An active message does not influence the measurement.
Deleting the active message
The operator can delete the active message in the submenu diagnosis at any time. Please select
the menu diagnosis via the arrow keys; now confirm the message Service Request via the F-key.
The timer is re-set to the chosen time interval and the countdown is re-started.
Remark: The active message serves as an indication for the operator of maintenance work at the
instrument (e.g. filter exchange or absorber replacement).
Version: V1.35_R0
10/2014
Instruction Manual
RHADOX AMS 7300
ATEX Ex II 3G IIB+H2 T3
Page 37
Wi ##,## MJ / m
Service Request
Note The relay Service Request 12 -16 is active according to the adjusted
operation; please refer to # 7.6 of the manual.
Note An active message does not influence the measurement.
Version: V1.35_R0
10/2014
Instruction Manual
RHADOX AMS 7300
ATEX Ex II 3G IIB+H2 T3
Page 38
Version: V1.35_R0
10/2014
Instruction Manual
RHADOX AMS 7300
ATEX Ex II 3G IIB+H2 T3
Page 39
7. Serial Interface
Terminal assignment at terminal-row -X4:
In order to locate terminal-row -X4 please refer to the drawing in the appendix.
Terminal
Name
9
10
11
RS232
Description
RXD
TXD
GnD
Only potential-free units may connect to the serial interface, these are e.g. battery-driven laptops
or galvanically isolated PCs.
Version: V1.35_R0
10/2014
Instruction Manual
RHADOX AMS 7300
ATEX Ex II 3G IIB+H2 T3
Page 40
8. Error Messages
Note The instrument features an extensive self-monitoring. In case of an
error message the second line will display the cause of the shutdown.
Note In case of an error, message the instrument switches to the standby mode. The main valve has shut. The combustion air down regulated
and the catalyst(s) is/are shut off.
System is OFF
AB < 250C ERROR
System is OFF
CAT < 250C ERROR
System is OFF
APP Over Range
System is OFF
CAB TEMP ERROR
Note The instrument can restart any time. Before a restart the operator,
however, has to troubleshoot the error cause or to contact AMS GmbH.
System restart
Calibration
Service
* Restart
Enter the main menu via the F-key. Now choose the
message Restart via the arrow keys. Confirm via the Fkey. The system has restart.
Note If the cause of the shutdown still exists, the instrument again
switches off the system right after the start-up sequence.
Version: V1.35_R0
10/2014
Instruction Manual
RHADOX AMS 7300
ATEX Ex II 3G IIB+H2 T3
Page 41
Transgressing Al2:
ATTENTION: After the measuring system has stopped by a gas alarm, you have to search
for the cause before restarting the analyser. Make sure that you eliminate the cause of the
error. The user has to undertake adequate measures for which he himself exclusively is
responsible!
Furthermore, you have to check the gas protector and its alarm features in a quarterly
interval. Please do not apply any cal gas corresponding with the alarm limits. Please refer
to the attached instruction for the calibration of the central gas warning and of the gas
detector.
Version: V1.35_R0
10/2014
Instruction Manual
RHADOX AMS 7300
ATEX Ex II 3G IIB+H2 T3
Page 42
In order to check the gas protector with its central gas warning, a key switch has installed in the
inner electronic housing; a key switch bridges Al2 of the central gas warning, which means that
the gas alarms function can monitored although the Al2 limit has been crossed yet the analyser
not turns off.
However, before using the key switch you have to make sure that the atmosphere is nonexplosive during the deactivation of the alarm limit and during the monitoring.
For checking, switch the key switch to Pos. I. You are not able to take out the key in this position.
Now you can check the functions of the gas protector, of the central gas warning, of the horn and
of the flashing light (as described above with the help of the corresponding instruction manual
which has enclosed to the users manual).
After checking the functionality, set the key switch back to Pos. 0 and remove the key from
the lock! This is for your own safety and impedes unintended altering of any values!
Test gases necessary:
Alarm limit 1:
Alarm limit 2:
:
:
:
5,5 bar-g
max 100 m/hr at max. cooler power
> 15 C
The temperature switch has also installed into the housing (15). If you wish to reset the switching
point, please see below:
During the normal operating of the analyser, the socket must not used! It means only for a
temporary use.
Version: V1.35_R0
10/2014
Instruction Manual
RHADOX AMS 7300
ATEX Ex II 3G IIB+H2 T3
Page 43
FUNCTION DESCRIPTION:
If the excess bypass gas (sample gas mixed with air) cannot lead off into a safe and back
pressure-free tube, it has to be burntdue to environmental reasons. Please make sure that the
by-pass gas not burns via a naked flame due to safety reasons.
The catalytic reactor for the catalytic combustion has structured for a minimum flow resistance and
for a complete combustion of the by-pass gas. The off-gas consists only of carbon dioxide and
water at the outlet of the catalytic combustion. Therefore, the exhaust gas tube has to be sloping.
The separate temperature controller stabilises the temperature of the catalytic combustion at 400
C. The electronic unit monitors exceeding or falling below this temperature. If the temperature
constantly exceeds or falls below the alarm values, the RHADOX will turn off. No sample gas will
reach the analyser. Only instrument air still be applied and will be lead off via the catalyst
protecting the catalyst at the same time.
ELECTRIC CONNECTION
Please refer to the technical drawings in the appendix for the electric connection of the catalytic
combustion.
Please make sure that:
1. the cross section of the heat line measures min. 2.5mm
2. the temperature sensor is connected with thermal wire (suitable for thermo couple type
NiCrNi) to the heat controller.
Version: V1.35_R0
10/2014
Instruction Manual
RHADOX AMS 7300
ATEX Ex II 3G IIB+H2 T3
Page 44
Version: V1.35_R0
10/2014