Professional Documents
Culture Documents
MR Equipment
Geoffrey D. Clarke, Ph.D.
University of Texas Health Science Center at
San Antonio
Radiological Sciences Division
Overview
MRI Standards
MRI Accreditation Program
MRI Quality Control Program
Definitions
Physicists Responsibilities
Technologists Daily & Weekly QC Tasks
Other Tests for the Physicist
ACR
Magnetic
Resonance
Imaging
Quality
Control
Manual
2001
Radiologists Responsibilities
Eleven specific responsibilities including:
Responsibilities of the
Qualified Medical Physicist
or MRI Scientist
Write Purchase Specifications
Perform Acceptance Testing
Baseline Measurements
MRI QC Technologists
Responsibilities
Daily magnetic resonance image
quality control procedures
Weekly quality control of hard copy
and soft copy Images
Routine visual inspection of equipment
Technologist Responsibilities
Designated QC Technologist(s)
QC Notebook
QC policies and procedures
data forms where QC procedure results are
recorded
notes on QC problems and corrective actions
Technologists Tests
Center frequency
Table positioning
Setup & Scanning
Geometric accuracy
High contrast resolution
Low contrast detectability
Artifact analysis
Film quality control
Visual Checklist
Daily
Daily
Daily
Daily
Daily
Daily
Daily
Weekly
Weekly
Documenting the
Quality Control Program
Data form for daily equipment quality
control
ACR MRI QC manual, pg. 64
Inter-Slice RF Interference
Soft Copy Displays (monitors)
FWHM
FWHM
oo
Denotes a totally
uniform magnetic
field.
All signal is at
resonant
frequency, o.
o.
oo
oo
Fourier transform of
signal produces a
Lorentzian peak in
well-shimmed
magnet
Magnet firld
homogenity can be
characterized using
FWHM of resonance
peak
Slice Position
SLICE #1
Slice Spacing
SLICE #11
Slice Thickness
4. Lower display to
one/half the
average.
5. Measure lengths of
top and bottom
ramps and
calculate slice
thickness.
(
top bottom )
Slice Thickness = 0.2
(top + bottom )
Slice Thickness
ACR-MRAP: slice thickness measured should
be + 0.7 mm of prescribed value
Corrective actions:
Check Axial Site Series Images
Replace cables & connectors, look for other sources
of distorted RF pulse shape in RF electronics
Try switching RF coils
Check gradient calibration
Site &
Equipment
Data
Site: _____________________________
MRAP Number: ____________________
Date: ________
Serial Number: ___________
Equipment:
MRI System Manuafacturer: _________________ Model : ________
Processor Manufacturer:
PACS Manufacturer:
Bo
Homogeneity
Method Used (check one): Spectral Peak ___ Phase Difference ___
Other (describe) __________________________
Measured Homogeneity:
Diameter of Spherical
Volume (cm)
Homogeneity
(ppm)
________
________
________
Slice Position
Accuracy
_______
_______
_______
Slice Thickness
Accuracy
Wedge (mm)
=-
=+
Slice Location #1
________
________
=+
Top
______
(fwhm in mm)
Calculated slice
Thickness (mm) ______
Bottom
______
Uniformity Pattern
Birdcage Coil
High Field
Solenoid Coil
Low Field
ACR
Phantom
Slice #7
(high + low)
Measurement Considerations:
Display may not show signal values
Display may not allow user to set signal display
level
There may not be a well-defined high/low
intensity level
Abdomen-pelvis
phased-array
Torso
Phased Array
Max Signal
Min Signal
Signal-to-Noise
Mean Signal
SD of Background Signal
Percent Signal
Ghosting
Ghost Signal
Mean Signal
Background Signal
Maximum signal
SD of Background Signal
Volume Coil
Pulse Sequence:
degrees
Type: ____
2
FOV: _____ cm
Data Recorded
TR: _____
TE: ______
Data Collected:
Mean
Signal
Maximum
Signal
Minimum
Signal
Background
Signal
Calculated Values:
Uniformity
SNR
Ghosting
Noise
Standard
Deviation
Ghost
Signal
Calculated Values:
Signal-to-Noise
Ratio
Percent
Image Uniformity
Percent
Signal Ghosting
Surface Coil
Signal-to-Noise
Noise Standard
Deviation
________
________
No. Slices
11
11
11
11
min
0.5
1.0
Signal-to-Noise
vs. Inter-slice Gap
Signal-to-Noise Ratio
(percent of Maximum)
100%
SE, TE = 20
FE, TE =8
TFE, TE =4
90%
80%
70%
0%
25%
50%
75%
100%
5. Interslice RF Interference
Phantom Description: ___________________________________________
Pulse Sequence:
Series
Number
S lice
Gap
(mm)
100 %
S ignalto-Noise
Ratio
M e a sure d S NR
RF Slice
Interference
Number of slices______
1
2
3
4
90 %
80 %
70 %
0%
25%
50%
75%
100%
Inte r- slice Ga p (pe rce nt of slice thickne ss)
Soft Copy
Displays
-2
-2
______ Cd m .
Review of
Routine QC
Program
ACR MRI QC Manual, pg. 127
(daily)
_________
_________
_________
_________
_________
_________
_________
_________
Summary
Sheet
(pg. 131)
Pass / Fail
_________
_________
_________
_________
_________
_________
_________
_________
_________
______________________________________________
______________________________________________
______________________________________________
______________________________________________
______________________________________________
______________________________________________
______________________________________________
______________________________________________
______________________________________________
Technologist
performs daily tests to assess image quality using
ACR phantom
Weekly checks of hard copy output