Professional Documents
Culture Documents
Aim: The aim of this experiment is to analyze reactive phosphorus in south quay lake and
monash drain by spectrophotometer measurement.
Outline of method:
Calibration standards
of known
concentrations were
prepared
Concentration of
phosphate was then
determined by
plotting a calibration
graph.
Summary of results:
Blank data
1 drop of
phenolphthalein and
dropwise of sufficient
Naoh were added to
Monash drain and
South Quay samples
and 5 replicate blank
samples
The absorbance of
all samples, samples
and blanks were
measure at 880nm
using a
spectrophotometer
The acidified
phosphomolybdenum
developed into a blue
solutions after 10
minutes.
Abs 1
0.000
0.013
0.001
-0.001
0.010
Abs 2
0.008
0.009
0.010
0.014
0.000
Abs 3
0.006
0.008
0.006
0.013
0.001
Abs 4
0.004
0.004
-0.003
-0.010
0.009
Mean Abs
n-1
0.0050
0.0052
Calibration data
Table 2. The absorbance, mean and standard deviation of the phosphate
Phosphate
(gP/L)
0
50
100
150
200
Abs 1
0.045
0.070
0.116
0.138
Abs 2
0.040
0.077
0.121
0.143
Abs 3
Mean Abs
n-1
0.037
0.067
0.106
0.142
0.0050
0.0407
0.0713
0.1143
0.1410
0.0052
0.0040
0.0223
0.0076
0.0026
Calibration graph
0.16
0.14
f(x) = 0x + 0.01
R = 0.99
0.12
0.1
Absorbance at 880nm
0.08
0.06
0.04
0.02
0
0
50
100
150
200
250
Table 3. The absorbance at 880nm, concentration and mean concentration for the water samples.
Replicat
e
number
Abs 1
Abs 2
Abs 3
Conc 1
Conc 2
Conc 3
Sample
A1
Sample
A2
Sample
A3
0.030
0.035
0.033
35.2857
42.4286
39.5714
Sample
B1
Sample
B2
Sample
B3
QC
Mean
conc
(gP/L)
Mean
Conc
(gP/L)
dilution
corr.
78.1904
39.0952
0.029
0.028
0.024
33.8571
32.4286
26.7143
0.019
0.022
0.025
19.5714
23.8571
28.1429
62.0000
31.0000
47.7142
0.018
0.021
0.018
18.1429
22.4286
23.8571
Sample A mean
(gP/L)
n-1
18.1429 19.5715
0.014
0.017
0.013
12.4286
16.7143
11.0000
13.3810
26.7620
0.017
0.020
0.019
16.7143
21.0000
19.5714
19.0952
38.1905
Sample A mean
(gP/L)
n-1
Conc 3 Mean
Conc
(gP/L)
34.6984
Abs 1
Abs 2
Abs 3
Conc 1
Conc 2
0.077
0.080
0.073
102.428
6
n-1
5.0170
% RSD
106.714
3
4.9209
96.7143
101.952
4
% Relative Error
62.6349
15.2480
39.1430
6.8897
Mean
Conc
(gP/L)
dilution
corr.
1.9524
Sample calculations
Preparation of standards :
Using C1V1 = C2V2 where C1 is 50 gP/L, C2 is 1 gP/L and V2 is 250ml.
50 x V1 = 1 (250)
V1 = 5 ml of 50 gP/L
The same method was used to calculate the working stock at concentration of 100, 150 and 200
gP/L
The equation obtained from the graph is y = 0.0007x + 0.0053 where y is the absorbance value at
880nm and x is the phosphate concentration (gP/L)
Concentration for sample A1:y = 0.0007x + 0.0053
0.030 = 0.0007x + 0.0053
x = 0.030 0.0053
= 35. 2857 gP/L
All samples were diluted by a factor of 2. So the original concentration for sample A is:
Mean concentration sample A x 2
= 39.0952 x 2
= 78.1904 gP/L
Sample A mean:
(Original concentration A1 + Original concentration A2 + Original concentration A3) / 3
(78.1904 + 62.0000 + 47.7142) / 3
= 62.6349 gP/L
Sample A standard deviation:
31
= 15.2480
All the concentration, mean concentration, mean and standard deviation for sample B were
calculated using the same method.
% Relative Error:
[(Actual Value Measured Value) / Actual Value] x 100
[(101.9524 100)/100] x 100
= 1.9524 %
Limit of detection:
Cm = (Sbl + ksbl )/ m
= (0.023 0.005) / 0.0007
= 25.714 g/L
Summary Table
Table 4. Summary of the results.
Sensitivity
Minimum detectable absorbance
Limit of detection
Reproducibility
0.0007 g-1L
2.1679 x 10-8 mol-1L
0.023
25.714 g/L
n-1
% RSD
Accuracy
South
lake
15.2480
24.3443
Monash
drain
6.8897
19.8560
QC
5.0170
4.9209
% RE: 1.9524
101.9524 g/L while the actual value is 100 g/L, so the relative percentage error is 1.9524%.
The relative error of the measurement shows how large the error is in relation to the correct value
which compares the size of the error to the size of the object being measured. Since, the
percentage error is lower than 5%, the data is good because there is lesser error and the value is
almost close to the correct value. If the percentage error is more than 5%,
Precision is the reproducibility of results obtained in identical manner. The relative standard
deviation is widely used in analytical chemistry to express the precision and repeatability of an
assay (Robinson, Skelly Frame and Frame II, 2012). A lower percentage of RSD indicates a
lower variability in the data set. Equally, a higher percentage indicates the data set is more
varied. In this experiment monash drain has a lower percentage RSD which is 19.8560 %
compared to the South Quay lake which is 24.3443%. Hence, Monash drain has a lower
variability and is more precise compared to South Quay.
The sensitivity of this experiment is 0.0007 g-1L. This spectrophotometer has a minimum
detectable absorbance 0.023which means that the device can detect up to 0.023 for the blank.
Limit detection is used to detect the lowest concentration of the spectrophotometer which is
25.714 g/L. The higher concentration may be due to the reading value of the blank and the
standard solution. Colour of phosphomolybdenum blue which will fade in a long period of time
can also contribute to the error in the experiment as there is a prolog waiting time to take the
result for all the sample data. The % RSD for south quay, monash drain and quality control is
24.3443 %, 19.8560%, 4.9209% respectively.
Error in measurement may be represented by a tolerance interval (margin of error). The
spectrophotometer used in manufacturing often set tolerance intervals, or ranges in which
product measurements will be tolerated or accepted before they are considered flawed.
To avoid errors, make the measurement with an instrument that has the highest level of precision.
The smaller the unit, or fraction of a unit, on the spectrophotometer, the more precisely the
device can measure. The precision of a measuring instrument is determined by the smallest unit
to which it can measure (Kenkel, 2003). Besides that, randomizing samples or experiment could
also increase the accuracy by increasing the signal to noise ratio. Other than that, repeat the same
measure several times to get a good average value.
Hence it can be concluded that, the concentration of phosphate in South Quay (Sample A) is
62.6349 gP/L 15.2480 and the concentration of phosphate in Monash drain (Sample B) is
34.6984 gP/L 6.8897. This shows that monash drain is cleaner as there is lesser concentration
of phosphate present it water.
Questions:
1. Is there evidence from the sample measurements that the dissolved reactive phosphorus
concentrations are deemed to be above the Interim National Quality Standards (INWQS)
guideline values issued by Department of Environment (DOE) for a Class III water source?
From the website, the concentration of phosphate in class III is 0.1 mg/L which is 100 g/L
There is no evidence that the dissolved reactive phosphorus concentrations are deemed to be
above the INWQS because the mean concentrations for South Quay (62.6349 gP/L) and
Monash drain (34.6984 gP/L) are lower than the concentration of phosphate.
2. Estimate the improvement in sensitivity if a 4cm cell rather than a 1cm cell was used. Would
this change the detection limit?
The sensitivity of the sample will increase if a 4cm cell is used rather than 1 cm cell.
According to the Beer Lambert Law, the signal will increase if the path length used is
increased. This would change the detection limit. Higher sensitivity gives rise to a lower
detection limit as the intensity of the transmitted light decreases exponentially as the path
length of the sample increases (Koch, 2012).
4. Silicate also undergoes reactions with acidic molybdate to form silicomolybdic acid which
can also be reduced to molybdenum blue. Briefly describe an experiment that would enable
you to determine whether silicate is a chemical interference in the measurement of
phosphate.
Silicate can be tested through acid treatment and reaction with ammonium molybdate to give
a yellow coloured complex or silicomolybdate complex. Sodium sulphate is added as well to
reduce the complex. If the reaction turns blue, there may be possibility that silicate may be
present in the solution. To determine the absorption, absorption spectrum need to be tested
using spectrophotometer. The absorbance readings are then compared. High value of
absorbance reading indicates that silicate is a chemical interference in the measurement of
phosphate which leads to changes in intensity of the analyte signal in spectroscopy
(Henriksen, 2001).
References
Henriksen, A. (2001). Interference from silica in phosphate analysis. Analyst, 91(1081), pp.290291
.
Kenkel, J. (2003). Analytical chemistry for technicians. 1st ed. Boca Raton: Lewis Publishers.
Koch, W. (2012). Ten years of Analytical and Bioanalytical Chemistry. Analytical and bioanalytical
chemistry, 402(1), pp.186 -189.
Robinson, J., Skelly Frame, E. and Frame II, G. (2012). Undergraduate Instrumental Analysis, Sixth
Edition. 1st ed. Hoboken: Taylor and Francis.
Quevauviller, P. (2008). Quality Assurance in Environmental Monitoring. 1st ed. Wiley VCH.