Professional Documents
Culture Documents
The First Red switch is the Main switch which On or Off the
power supply.
The Second green switch is the cooling switch which can off the
compressor when you feel any malfunction. You need to keep it
ON continuously in the any temp.you set between 5 to 60 degree
C.
The third one is the illumination switch which gives the better
visibility inside the chamber.
The PID temp. controller having dual display. The Upper display
shows the actual temp. inside the chamber and the lower display
shows the set temp.
It also has the three key at the bottom , the first one is the set
button , the next one is Up button and the last one is the down
button.
When you want to set any temp. first you press set button , Now
the lower display shows SP and upper display shows some value
, you can set any desired value by pressing up or down button
and press set button.
Now the lower display shows OP 2 and upper display shows
Auto. Keep it in auto mode always.
Now you press set button, the lower display shows OUT.L and
upper display shows 100. Keep it 100.0 always.
Now you press set button , the lower display shows tune and
upper display shows no.The tune function we use for better
accuracy, whenever you want the best accuracy , you set the
desired temp and put the tune yes by pressing up or down key
and press the set button. Dont disturb the unit until the light
blinking on tune button glows completely.
Safety Instruction ;