You are on page 1of 16

GDS500A

Installation of ver. 1.71 GDS500A Software using Win 7


operating system

Installation of ver. 1.71 GDS500A Software


Using Win 7 Operating System

M.Poirier
10-25-10
Ver. 1.71 GDS500A software Installation requires NIDAQ MX 9.0 (NIDAQ
900f2.exe file)
Install NiDaq softwareLoad the Ni Daq CD or .exe file
2 | Page

Installation of ver. 1.71 GDS500A Software using Win 7


operating system

Double click to select Files Currently on the Disc

The software maybe compressed. Select Ok and Unzip

After the process of unzipping you will be asked -"Do you want to allow the following
program to make changes to computer?" Select Yes.
Once the changes are made you will be prompted to pick a Destination Directory,
Select Next

3 | Page

Installation of ver. 1.71 GDS500A Software using Win 7


operating system

Select the default Typical Installation Option and select Next

Uncheck the Product Notifications tab and select Next


4 | Page

Installation of ver. 1.71 GDS500A Software using Win 7


operating system

Select I accept the License Agreement and select Next

Select I accept the above 3 License Agreement(s) and select Next


5 | Page

Installation of ver. 1.71 GDS500A Software using Win 7


operating system

Select Next to start the actual installation

Once installation is complete select Next


6 | Page

Installation of ver. 1.71 GDS500A Software using Win 7


operating system

Select Shut Down and install the 777-917 NiDaq card, 666-511x Filter card and
cables.

Once cards and cables are installed restart the CP.

7 | Page

Installation of ver. 1.71 GDS500A Software using Win 7


operating system
Open the Measurement and Automation software.

Select Devices and Interfaces. Confirm the software is recognizing the NiDaq
hardware by seeing
NI PCI-6110 Dev1 in the drop down tree.

8 | Page

Installation of ver. 1.71 GDS500A Software using Win 7


operating system

Select by double clicking on NI PCI-6110 Dev1 to reveal the test options.

Select Test Panels

9 | Page

Installation of ver. 1.71 GDS500A Software using Win 7


operating system

Use the drop down menu, Channel Name to select the proper channels to be tested
Channel 1 = Dev1/ai0
Channel 3 = Dev1/ai2

Channel 2 = Dev1/ai1
Channel 4 = Dev/ai3

Select Start to enable test.


Voltages for each Channel should be approx. 1.7 to 1.9 Volts

10 | P a g e

Installation of ver. 1.71 GDS500A Software using Win 7


operating system

Once the voltages are confirmed for all 4 Channels, exit the Measurement and
Automation software
and install the GDS500A version 1.70+ software.

If this is a tower upgrade you will need to import the PixNorm.dat, database and
High Voltage Power Supply calibration.
Right click on the Explorer Icon (Start)
Select Open Windows Explorer

11 | P a g e

Installation of ver. 1.71 GDS500A Software using Win 7


operating system

Select Local Disk (C:)

Select Users

12 | P a g e

Installation of ver. 1.71 GDS500A Software using Win 7


operating system

Select All Users


Note- You may have to enable Folder options to Show Hidden Folders and Drives

13 | P a g e

Installation of ver. 1.71 GDS500A Software using Win 7


operating system
Select LECO

Select GDS500A
Import PixNorm.dat and High Voltage Power Supply calibration (Settings.xml) into
the GDS500A folder.

14 | P a g e

Installation of ver. 1.71 GDS500A Software using Win 7


operating system

-On new installs the High Voltage Calibration is not held in the registry and the
settings may need to be manually transposed to the Settings.xml file.
-On software upgrades from earlier versions than 1.70, the settings are transferred
automatically from the registry to the Settings.xml file in the GDS500A folder.
To edit this file, right click on the Settings.xml icon and open with NotePad to add
the needed calibration values.
Note- There may or may not have control rates and proportional gain settings in
your Settings.xml file.
Disregard these settings and just add your voltage settings to the highlighted area
with the proper spacing.

Once editing is complete on the Settings.xml file Select File and select Save.

15 | P a g e

Installation of ver. 1.71 GDS500A Software using Win 7


operating system

In the GDS500A folder select Databases

Inside the Databases folder is the location to place your current database. It must
be named Database to be recognized.

Once installed close all software and re-open the GDS500A software to confirm
proper operation.

16 | P a g e