Professional Documents
Culture Documents
337
BBA 11150
a C H U Pitie-Salpetriere, Laboratoire de Physiologie, I N S E R M U3, 91 Boulevard de l'H6pital, F-75013 Paris and b Laboratoire de
Neurobiologie Cellulaire, CNRS, Gif - sur- Yvette (France)
Some novel studies of the properties of the antimony microelectrode used for intracellular pH measurements
are described. First, it is shown that currents in the picoampere range, such as those encountered as leakage
in some electrometers, induce important changes in pH sensitivity. The response time of the electrode has
also been measured and indicates that the electrode exhibits a rapid time course which would be very useful
for dynamic cytoplasmic pH investigations. An example of internal pH recording during cellular acidification
in Xenopus laevis oocyte is also presented.
Introduction
Measurements of intracellular pH using microelectrodes have been greatly advanced in the last
10 years, principally through the work of Thomas
[1 ]. The glass pH-microelectrode, particularly with
the recessed tip, has become the standard for
critical intracellular pH measurements. These electrodes are, however, difficult to fabricate, especially if a very fine tip is desired, and difficult to
use owing to a high resistance. In those instances
where only an indication of the intracellular pH at
rest or rapid response to pH shifts are needed, the
glass electrode may seem to require too great an
effort to be worthwhile. In such cases, the antimony (Sb) microelectrode may prove to be a valid,
easily made alternative.
The antimony electrode has had a long history
in pH measurements [2]. Unlike the selective glass
electrode, the Sb electrode depends on a metalmetal oxide to solution interface for its pH sensi-
[3].
Recently, there has appeared a series of articles
by Fujimoto and coworkers [4-7] in which the
authors have systematically re-evaluated the Sb
microelectrode in relation to (1) the pH-sensitive
glass microelectrode, (2) the physiochemical characteristics and the effects of the buffer used for
calibration, (3) the temperature coefficient and
oxygen effects, and (4) the effects of proteins in
the measured solution. In general, these studies
have shown that under normal conditions the Sb
electrode can provide pH estimated to + 0.2 U or
better against a standard glass microelectrode. Such
a performance is obtained for Sb electrodes having
slopes greater than - 4 5 mV per unit pH and
which did not show drifts greater than 4 mV
before and after calibration in buffers having about
the same ionic strengths as that expected in the
measured medium.
338
We report here the results of some studies relating to the speed of response of these electrodes,
the effects of leakage currents on the p H measurements and some results obtained with intracellular
recordings.
Material and Methods
Microelectrode fabrication
We have used a somewhat simpler procedure
[4,8], for the fabrication of the Sb electrodes:
purified Sb (99.999%) as obtained from most
chemical supply houses was melted in a pyrex
beaker under 1-2 cm of granulated bone charcoal
over an ordinary bunsen flame. The charcoal is
simpler to implement than a nitrogen gas system
to prevent the rapid oxidation of the melted Sb,
and the small quantity of resulting ash does not
appear to affect the performance of the electrodes.
A piece of borosilicate micropipette tubing, 1 m m
inside diameter about 10 cm long, without internal
fiber (Clark Electromedical Instruments CG150T),
was prepared with an appropriate length of flexible tubing slipped over one end. When the Sb was
fluid, the pipette tubing was quickly introduced
into the melt and, with a 10 ml plastic syringe, the
Sb was quickly aspirated. This must be done rather
sharply to be sure to fill a 4 - 5 cm length of the
pipette before the Sb solidifies. Several such tubes
were filled in succession until the molten Sb was
almost exhausted. Some care must be excercised to
be sure that air was not aspirated into the tube, as
this produces voids in the metal column. Such
voids result in separation of the metal column
upon subsequent heating of the pipette to form the
electrode tip.
The microelectrode tip was formed in two steps.
The Sb-filled part of the tube was first pulled by
hand in a small flame, such as the pilot of a
bunsen burner, to a diameter of about 50-100/~m.
It may be necessary to pull more than one time in
the flame to obtain a sufficiently fine drawn segment. The Sb now forms a fine filament and must
be continuous in all parts of the segment. The fine
segment was then broken at 1-1.5 cm from the
point at which the tip of the electrode was to be
formed, and the end reintroduced in the flame at
an angle until a small hook is formed. The tubing
can now be mounted in a microforge, a weight
Electrode calibration
The electrodes must be calibrated before use as
they are not absolutely uniform in responsiveness
and, as pointed out by Fujimoto et al. [4], should
be rejected if they do not show at least - 4 5
m V / p H unit sensitivity or if they show any drift
during the calibration. The test solutions used here
were 50 m M Tris-maleate (Sigma), p H adjusted
with HC1 using a Radiometer model 28 p H meter
and a combination electrode. As pointed out by
several Sb electrode users [4,8,9] the relation between p H of the tested solutions and the potential
of the electrode is almost linear for p H from 3 to
10 with a slight change in slope at p H 7 (see
Fig. 1). Moreover, Hepes (Sigma) appears to be
equivalent to Tris as a buffering system for
calibrations while phosphate introduced a significant change in H+-senstivity of the antimony microelectrode [10].
The tracing shown in Fig. 1A, B and C are
recordings obtained by changing the bathing
medium with a peristaltic pump. It might be expected that the response time should be the same
for each change in pH, but in fact it can be seen in
these tracings that the time to reach steady state
becomes longer as the p H is enhanced. This slowing of the response is interpreted to be due to an
increased time necessary to achieve proton dilution as the p H is increased. For two reasons, first,
in the decreasing direction, steady state is more
339
my.
-5001
,,
0j -~
100G G
87! C .my
8.1 '
G-~en'q~/~
BB3528CM
1Glen
E sb
-- L_I--250
2rnin
pH
ef
Fig. 2. Experimental set-up for testing properties of antimony
7.
/o
j.
6.
QJQ
-360
-3g0
-460
-~0
-s60mv
Fig. 1. Antimony microelectrode calibration Upper trace: Example of voltage recording obtained at different test solutions
from pH 3.8 to 8.7. Parts A and C correspond to the scale on
the left and B to that on the right. Solutions were changed
using a perfusion pump until a plateau was reached Lower
For resistance measurement, and to study the effects of leakage currents in the electrode, a voltage
source was added between the input of the amplifier and ground, as shown in Fig. 2.
All recordings were made with a Brush 280 pen
recorder which has a full scale response time of 10
ms to a square wave input.
Results
rapidly obtained with the same flow conditions,
and secondly, if a large volume is used with the
perfusion system, oscillatory changes are seen due
to the movements of imperfectly mixed media in
the measurement chamber. The slowing of the
response with increased p H is therefore not considered to be due to a p H - d e p e n d e n t response time
of the electrode itself.
Measurement of the Sb electrode potential must
be done with an electrometer type of amplifier as
the resistance of these electrodes tends to be greater
t h a n 10 9 ~. We have found that a good quality
F E T * input o p - a m p having input-bias currents
less than one pA, such as the Analog Devices A D
52J or the Burr Brown 3528CM are quite adequate
even for very long term recordings of up to 5 h.
340
should not exceed 100-200 ms ( R . C). For example, a 2 G O electrode resistance with a 10 pF input
capacitance gives a time constant of 20 ms which,
in turn, should give a response to better than 99%
in 120 ms if Esb were perfectly square.
-50
I
-100
50
I
pA
100
I
150
; ~
. ~ p H 6
-200.
-300"
cr
200msec
pH 7
2//sec
mY
t
-450
o..*
-600
-400
pkl 6
1
5
J
10
I111"11
mV
Fig. 4. Effects of currents on Esb. Current voltage relation for
one electrode at three different pH values (e) and another
electrode at pH 6 (). The inset shows a typical recording
from which the I - V curves were obtained. For large negative
currents, the relation becomes markedly non-linear (see also
Fig. 5B).
341
the electrode decreased with increased size. However, for electrodes in the 1-2/~m diameter range,
the effect of current in the electrode was very
severe (see Fig. 5).
These measurements also show that the resistance of the electrode remains practically linear
with the current at a given pH but became more
non-linear at pH 6 than at pH 7 or 8. At zero
current, however, the potential obtained at a given
pH is very reproducible across electrodes. One
such example is shown in Fig. 4, where two electrodes of quite different resistances (0.9 and 1.8
G~2) give very nearly the same potential at pH 6
and zero current. This was a consistent finding for
all of the electrodes fabricated.
The resistance of newly fabricated electrodes
A
V"'
"1
r
r
:t-
.r
r
500 msec
Intracellular recordings
Intracellular proton activity has been monitored
in oocytes from Xenopus laevis. For this purpose,
the cell was penetrated with an antimony microelectrode and a second one filled with 3 M KC1
for reference and voltage recording. In the present
experiments, internal pH was determined by subtracting the membrane potential (typically - 5 0
mV) from the Sb electrode potential (Fig. 6A and
B). Resting cytoplasmic pH was about 7.45 (Fig.
342
TABLE I
Sb ELECTRODE RESISTANCE CHANGES WITH TIME
Resistance of five antimony microelectrodes measured at 0, 30, 60, 90 and 150 min (one case) in a test solution at pH 7. The measured
resistances were normalized to the maximum value for each electrode and the means determined at the indicated times. The electrode
potentials, Esb, did not show any significant modification, except for number 7, which was initially abnormally low and began with a
high resistance.
Electrode resistance (GO)
Time (min):
Esb
(mY)
30
60
90
No
3
5
7
8
10
0.056
0.30
1.25
0.30
0.21
0.081
0.39
1.30
0.51
0.89
0.081
0.47
1.30
0.64
1.06
0.081
0.51
1.20
0.81
1.13
0.56
0.86
0.95
1.00
C02
100"/.
my
1-,-0
-20
0
mY
~t
~-'-~
-z,8o
-/-,60
-,'-40
-420
pHi
7.5
150
- -
0.65
-- -
413
- 422
- 320
-434
- 432
-
Conclusions
60sec
6.5
P'ig. 6. Intracellular pH monitored with an antimony microelectrode. Potentials recordings from an electrode filled with KCI
(A) and an Sb electrode (B). C: point by point subtraction of A
from B plotted in units of pH, which at rest was 7.45. Superfusion for 90 s with a solution saturated with 100% CO 2 (bar)
induced a large cytoplasmic acidification (0.8 pH unit) which
was reversed with washing.
343
Acknowledgements
We wish to thank Mrs. J. Nicolet for her helpful
assistance throughout this work. This work has
been partially supported by DGRST grant No.
81-E-0576.
References
1 Thomas, R.C. (1978) in Ion Sensitive Intracellular Microelectrodes: How to Make and Use Them (Thomas, R.C.,
ed.), pp. 1-110, Academic Press, London, New York, San
Francisco
2 Ives, D.J.G. and Janz, G.J. (1961) in Reference Electrodes:
Theory and Practice (Ives, D.J.G. and Janz, G.J., eds.), pp.
1-651, Academic Press, London, New York
3 Roos, A. and Boron, W.F. (1981) Physiol. Rev. 61, (2),
296-433
4 Fujimoto, M., Matsumura, Y. and Satake, N. (1980) Jap. J.
Physiol., 30, 491-508
5 Matsumura, Y., Satake, N. and Fujimoto, M. (1980) Jap. J.
Physiol., 30, 509-528
6 Satake, N., Matsumura, Y. and Fujimoto, M. (1980) Jap. J.
Physiol., 30, 671-687
7 Satake, N., Matsumura, Y. and Fujimoto, M. (1980) Jap. J.
Physiol., 30, 689-700
8 Bicher, H,I. and Ohki, S. (1972) Biochim. Biophys. Acta
255, 900-904
9 Vieira, F.L. and Malnic, G. (1968) Am. J. Physiol. 214,
710-718
10 Green, R. and Giebisch, G. (1974) in Ion Sensitive Microelectrodes (Berman, H.J. and Hebert, N.C., eds.), pp. 43-52,
Plenum Press, New York
11 Ammann, D., Lanter, F., Steiner, R.A., Schulthess, P.,
Shijo, Y. and Simon, W. (1981) Anal. Chem. 53 (14),
2267-2270
12 Webb, D,J. and Nuccitelli, R. (1981) J. Cell Biol., 91,
562-567
13 Thomas, R.C. (1974) J. Physiol. 238, 159-180
14 Boron, W.F. and Deweer, P. (1976) J. Gen. Physiol. 67,
91-112
15 Giaume, C. and Korn, H. (1982) Neuroscience, 7 (7),
1723-1730