You are on page 1of 8

Energy Dispersive X-ray

Fluorescence Spectrometer
(EDX-2)

Users Manual
Global Office of Lisun Electronics Inc.
http://www.Lisungroup.com
Lisun Group (Hong Kong) Limited
Add: Room C, 15/F Hua Chiao Commercial Center, 678 Nathan Road, Mongkok, Kowloon,
Hong Kong
Tel: 00852-68852050
Fax: 00852-30785638
Email: SalesHK@Lisungroup.com
Lisun Electronics (Shanghai) Co., Ltd
Add: Room 1210, No. 9 Building, Jiangqiao Wanda Plaza, Hexuan Road, Jiading District,
Shanghai, 201803, China
Tel: +86(21)5108 3341
Fax: +86(21)5108 3342
Email: SalesSH@Lisungroup.com

Lisun Electronics Inc. (USA)


Add: 445 S. Figueroa Street, Los Angeless, CA 90071, U.S.A.
Email: Sales@Lisungroup.com
Lisun China Factory
Add: NO. 37, Xiangyuan Road, Hangzhou City, Zhejiang Province, China
Tel: +86-189-1799-6096
Email: Engineering@Lisungroup.com

Lead in CFL & LED Test Instruments

Lisun Electronics Inc.

Sales@Lisungroup.com

www.Lisungroup.com

The quotation includes the following items:

For the purpose of restricting the pollution caused by the dispose of electronic and electrical
products, the RoHS or Lead Free directive was announced by EU on Feb 13th of 2003 and
became effective on Jul 1st of 2007. Lisun Group developed an high accuracy EDXRF
spectrometer. The EDX-2 can fully meet the RoHS test requirements

-Page 1 -

Lisun Electronics Inc.

Sales@Lisungroup.com

www.Lisungroup.com

Part 1 Specification:

Nondestructive measurement, no chemical and physical damage on measured object

High accuracy which can be similar as chemical test

Detect and analyze more than 75 chemical elements from Sulfur to Uranium, test Range:
2ppm~99.99%

Measure the details for the 5 restricted elements in RoHS Directive: Cd/Pb/Hg/Cr/Br)

esting time: 60~300s

Adopt world-famous electrical cooled detector, reliable and stable, energy resolving power
is less than 160eV

X-ray tube voltage: 5~50kV; Tube current: 50~1000uA

High definition camera built-in; USA made high pressure module built-in; Light
enhancement system

SNR design enhanced, quick test circuit and precise numbering system

Multi-function software and multi-substrate effect calibration module built-in

The system includes a PC and printer which have already installed the English software

Part 2 Installation
2.1 Structure
2.1.1 Front view:

: Cover
: Strut Bar
: Control Column
: Sample Hole
: Body
: Power
: Indicator
: Detector
: Collimator

-Page 2 -

Lisun Electronics Inc.

Sales@Lisungroup.com

www.Lisungroup.com

2.1.2 Rear view


: Fan
: RS-232C port
: USB port
: Outlet

2.2 Installation
2.2.1 Preparation for operation. The RoHS test system needs enough space as follows:

-Page 3 -

Lisun Electronics Inc.

Sales@Lisungroup.com

2.2.2 Connection

-Page 4 -

www.Lisungroup.com

Lisun Electronics Inc.

Sales@Lisungroup.com

www.Lisungroup.com

Part 3 Operation
3.1 RoHS testing process

START

CHECK
CONNECTION

FIRST TEST NO

PREHEAT 30min

YES

SET 200s

NO

TESTING
TIME 200s
YES

INITIALIZATION?

YES

CHOOSE CURVE

CHOOSE COLILIMATOR

TEST

START TO
PRINT

YES

PRINT
NO

END
-Page 5 -

NO

PUT IN SILVER TO CALIBRATE

Lisun Electronics Inc.

Sales@Lisungroup.com

www.Lisungroup.com

3.2 Preparation
Before switching on EDX, check all the connections. Then power on PC, printer, EDX-2,
and preheat 30min.Initiallize EDX-2, put in silver slice, choose 1# collimator, click

system calibrate automatically.


3.3 Sample test
1. According element needed of sample, choose the right rohs testing mode (choose
the wrong mode lead to a bad accuracy)
2. Put in matching collimator (choose the wrong collimator lead to a bad accuracy)
3. Choose testing time, 200s is recommended(too long or too short lead a bad
accuracy)
4. Put the unknown sample into the testing bed, observe the cross of camera is in the
center of unknown sample, or you can adjust it by moving the sample and control
column.
5. Close the cover, the protective light on
6. Click Start
7. After testing, you can save the data and name
8. If need to print, click Total analysis report, then go to report info setting
Notice:

EDX-2 need to calibrate after switching on every time

Wrong mode and collimator and too short RoHS test time lead to additional error.

Too small unknown sample or not in the center of he cross of camera lead error.

Next page is the test report

-Page 6 -

Lisun Electronics Inc.


www.Lisungroup.com
Tel:+86(21)51083342

ANALYSIS REPORT FOR RoHS REGULATIONS


Sample NameCFL Tube
Sample SourceCFL11W
Analysis institutionLisun EDX-2
Analysis Duration200s
Tube Current(uA)400uA

Sample BatchNo.1250
OperatorJacky Jiang
Analysis Date2010-03-14
Tube Voltage(KV) 45kV
Work Curve UsedPE

Y-
4500
4360
4000

3500

3000

2500

2000

1500

1000

Hg
As BrPb

2048

500
Cd

0
0

Element
As
Hg
Pb
Br
Cd

200

400

600

Intensity
0.0020
0.0014
0.0054
0.1454
0.0024

800

1000

1200

Content(ppm)
28.02
25.26
94.04
538.68
121.97

1400

1600

Limitation
1000
1000
1000
1000
100

1800

2000

Results

X-

Pass
Pass
Pass
Pass
False

Note:1)ND stands for contents not larger than 2ppm.


2)All Data are based on sample surface analysis with this X-ray fluorescence Instrument.
3)The Data for Cr&Br are the summation contents respective, their above-limitation
doesn't mean the same thing with VI-valence Cr,PBB and PBDE.