INTRODUCTION
The Thermocouple Calibration Laboratory (TCL)
of the National Institute of Standards and Technology
(NIST) provides a complete description of its
thermocouple calibration facilities and services in its
NIST SP 250 document. This document includes
descriptions of the existing calibration facilities, the
traceability of the temperature references to the current
temperature scale, and uncertainty budgets for the
calibration services provided. This document is
intended to be periodically updated as the facilities and
the methods of the TCL evolve.
The first version of the NIST TCLs SP 250
document was SP 250-35. [1]. It was supplemented in
1994 by NISTIR 5340 [2], which provided new
uncertainty budgets that were compatible with the
newly-published
Guide
for
Uncertainty
in
Measurement (GUM) [3]; these new uncertainty
budgets superseded those provided in the older SP
250-35. Refs. [12] only included uncertainty
estimates for the most frequently used calibration
services. Also, uncertainty estimates were provided in
electromotive force (emf) rather than temperature,
leading to confusion for many customers.
The NIST TCL is preparing a new version of the
SP 250 document, SP 250-92 [4]. This new version
describes the current facilities and calibration
For
independent
(uncorrelated)
uncertainty
components ui, the expanded total uncertainty U is
related to these n uncertainty components ui (expressed
as standard uncertainties) by
n
U2 k
1)
i 1
DESCRIPTION OF UNCERTAINTY
COMPONENTS
The uncertainty analysis for thermocouple
calibrations is based on the ISO and NIST guidelines
for the expression of uncertainty in measurement [3,6].
The total uncertainty is obtained by first determining
the values of the individual uncertainty components.
Uncertainty components are categorized as type A,
which can be evaluated using statistical analysis, or
type B, which must be evaluated using other methods.
Type B
Calibration Repeatability
Uncertainty of Quadratic Function
Emf Measuring System
Fixed-Point Temperature
Fixed-Point Drift
urepeat
uquad
uemf
ufp-temp
ufp-drift
uheat-mj
uheat-rj
Reproducibility of extrapolation
Linearity of emf deviation
Bias of slope at 1064.18 C
Type B
Bias of emf at 1064 C
Combined Expanded (k=2) Extrapolation
Uncertainty
u, C
0.23
0.20
0.14
0.12
0.72
Type
A
Calibration Repeatability
urepeat
Type
B
uemf
ufurn
uheat-rj
urt-cal
urt-inhom
UNCERTAINTIES OF CALIBRATIONS
AND TESTS
The numerical values of the uncertainties for all
NIST thermocouple calibrations and thermoelement
tests are given in Ref. [4]. Some examples of these
values are shown below. We provide the uncertainties
for calibrations of type S, type R, and type B
thermocouples at the Zn, Al, Ag, and Au fixed points
in Table 4. We also provide the uncertainties for
calibrations of type B thermocouples at the Pd fixed
point in Table 5. The propagated combined expanded
uncertainty U of the deviation curve and its
extrapolation above 1064 C for type S and type R
thermocouples as a function of temperature (resulting
from the calibration uncertainties at the fixed points) is
shown in Table 6. Here, the two columns on the right
include the uncertainty for the extrapolation of the
deviation curve. Table 7 shows the propagated
expanded total uncertainty U of the deviation curve
and its extrapolation above 1064 C for type B
thermocouples as a function of temperature (resulting
SUMMARY
We have presented here brief descriptions and
some current numerical values of uncertainties and
their components for thermocouple calibrations and
thermoelement tests performed at the NIST TCL.
Complete descriptions of these uncertainties and the
sources of their estimates will be found in the new
NIST SP 250-92 [4]. A listing of the numerical values
of these uncertainties and their components for all
thermocouple calibrations and thermoelement tests
may also be found in this document.
Table 4. Values of standard uncertainty components u and the total uncertainty U for calibrations
of thermocouples at the gold, silver, aluminum, and zinc fixed points. These components are
described in Table 1. The combined expanded uncertainty U at these fixed points is given at the
right of the table. Uncertainties are expressed in C. The total uncertainty does not consider the
inhomogeneity of the test thermocouple.
Type S and type R thermocouples
Fixed
point
Zn
Al
Ag
Au
urepeat
uquad
uemf
ufp-temp
ufp-drift
uheat-mj
uheat-rj
U(k=2)
0.027
0.030
0.031
0.035
0.021
0.019
0.018
0.017
0.002
0.003
0.004
0.004
0.001
0.001
0.002
0.003
0.006
0.006
0.008
0.009
0.006
0.006
0.006
0.006
0.002
0.002
0.002
0.002
0.071
0.075
0.075
0.081
Type B Thermocouples
Fixed
point
Zn
Al
Ag
Au
urepeat
uquad
uemf
ufp-temp
ufp-drift
uheat-mj
uheat-rj
U(k=2)
0.063
0.059
0.040
0.044
0.047
0.031
0.023
0.021
0.005
0.005
0.005
0.005
0.001
0.001
0.002
0.003
0.006
0.006
0.008
0.009
0.006
0.006
0.006
0.006
0.002
0.002
0.002
0.002
0.16
0.12
0.09
0.10
Table 5. Values of the standard uncertainty components u and total uncertainty U of calibrations
of type B thermocouples at the palladium fixed point. These components are described in Table 1.
Uncertainties are expressed in C. The total uncertainty does not consider the inhomogeneity of
the test thermocouple.
Fixed
point
Pd
urepeat
uquad
uemf
ufp-temp
ufp-plat
ufurn
uheat-rj
U(k=2)
0.043
0.017
0.008
0.4
0.2
0.006
0.002
0.90
U(k=2)
0.06
0.06
0.07
0.06
0.06
0.06
0.06
0.06
0.05
0.06
0.07
t, C
1100
1150
1200
1250
1300
1350
1400
1450
U(k=2)
0.07
0.13
0.22
0.32
0.42
0.52
0.63
0.72
U(k=2)
0.17
0.11
0.09
0.12
0.11
0.08
0.06
0.12
0.22
0.40
0.56
0.60
Pd Point Used
t, C
400
500
600
700
800
900
1000
1100
1200
1300
1400
1450
1500
1600
1700
1750
U(k=2)
0.17
0.10
0.11
0.10
0.11
0.09
0.06
0.14
0.23
0.32
0.54
0.61
0.68
0.9
1.2
1.4
urepeat
0.048
0.046
0.047
0.050
0.053
0.055
0.057
0.060
0.062
0.063
0.066
uemf
0.002
0.002
0.004
0.005
0.006
0.007
0.008
0.010
0.011
0.012
0.013
ufurn
0.001
0.002
0.003
0.004
0.005
0.006
0.007
0.008
0.009
0.010
0.011
REFERENCES
1. Burns, G.W. and Scroger, M.G., The Calibration of
Thermocouples and Thermocouple Materials, NIST
SP 250-35, National Institute of Standards and
Technology, Gaithersburg, MD, USA,1989.
2. Ripple, D.C., Burns, G.W. and Scroger, M.G.,
Assessment of Uncertainties of Thermocouple
Calibrations at NIST, NISTIR 5340, National Institute
of Standards and Technology, Gaithersburg, MD, 1994.
3. ISO, Guide to the Expression of Uncertainty in
Measurement, International Organization for
Standardization, Geneva , 1993.
4. Garrity, K.M., Cross, C.D. and Meyer, C.W., The
Calibration of Thermocouples and Thermoelements,
NIST SP 250-92, National Institute of Standards and
Technology, Gaithersburg, MD, USA, in preparation.
uheat-rj
0.003
0.003
0.003
0.003
0.003
0.003
0.003
0.003
0.003
0.003
0.003
urt-cal
0.031
0.030
0.033
0.031
0.030
0.029
0.028
0.028
0.027
0.030
0.034
urt-inhom
0.03
0.04
0.05
0.06
0.06
0.07
0.08
0.09
0.09
0.10
0.11
U(k=2)
0.13
0.14
0.15
0.16
0.18
0.19
0.20
0.22
0.23
0.25
0.27