Professional Documents
Culture Documents
Headspace
October 1995
Authors
Introduction
Susan M. Brillante
Hewlett-Packard Company
Wilmington, DE 19808-1610
Abstract
The analysis of dissolved gases
in electrical insulating oils is an
efficient diagnostic tool
for routine performance
monitoring of power system
equipment. The analytical
performance of a method using
the HP 5890 Series II GC
system with HP 7694 headspace
sampler with NiCat is evaluated
and compared with that of the
ASTM D 3612, Method A
standard procedure. This
comparison shows that the
headspace technique can be
used to determine dissolved
gases with the same precision
but shorter analysis times and
lower detection limits than the
ASTM method.
Experimental
Instrumentation
A schematic of the system used is
shown in Figure 1. Sampling was
done using an HP 7694 headspace,
equipped with a 44-sample outer
carousel feeding a 6-sample inner
carousel in a thermostatted oven
compartment. This sampler is
equipped with a vibrator to
accelerate analyte equilibrium rate
and improve detection limits. An
equilibration temperature of 70C
was used. All samples were placed
in 20-mL glass vials sealed with
perforated aluminum caps fitted
with a rubber PTFE-lined septum.
All vials were purged with Argon
___________
2
Sample
Transfer Line
Injection Valve
Pressure:
0.4 bar
Vial
Times:
0.25 min
0.25 min
0.25 min
0.90 min
30 min
Pressurization
Pressure Equilibrium
Expansion in Sample Loop
Injection
Mixing
Mixing Speed:
Power 2 (high)
30 minutes
Sample Preparation
Temperatures:
FID
TCD
NICAT II
350C
250C
350C
Oven:
40C for 3 minutes
24C/min to 170C for two minutes
24C/min to 250C for five minutes
Isolation Valve timing:
0-3.6 min
3.6-8 min
8-10 min
10-23 min
columns in series
MolSieve bypassed
columns in series
MolSieve bypassed
Figure 1. Schematic diagram of the Hewlett-Packard GC/headspace transformer oil gas analysis system.
___________
3
Figure 2a & 2b. Set-up used for the preparation of blank and dissolved-gas samples.
Table 2. Residual Gases in the Blank
Sample and Composition of the
Standard at 294.45K and 758.5 Torr
Gas
Blank
Sample
(ppm)
H2
O2
N2
CH4
CO
CO2
C2H2
C2H4
C2H6
C3H8
<0.6
17.0
24.0
<0.6
1.6
8.8
<0.05
<0.04
<0.04
<0.1
Standard
(ppm)
960
3800
960
960
1900
950
960
960
960
___________
4
a-FID
b-TCD
Figure 3a & 3b. FID and TCD chromatograms obtained for the primary oil standard (see
composition in Table 2).
Headspace
H2
O2
N2
CH4
CO
CO2
C2H2
C2H4
C2H6
C3H8
0.6
11.0*
11.2
0.06
0.09
0.1
0.05
0.04
0.04
0.2
5
50
50
1
25
25
1
1
1
1
20
500
500
1
2
2
1
1
1
1
a-FID
Figure 4a & 4b. FID and TCD chromatograms obtained for the intermediate standard
of lowest concentration (1 ppm H2 and CXHY, 11 ppm CO2, 3 ppm CO, 50 ppm O2 and 250
ppm N2).
Analytical Performance
b-TCD
___________
5
360
Headspace (ppm)
340
320
60
50
40
30
20
10
0
n = 31
References
4
Headspace (x104 ppm)
2
1
n = 35
0
0
Conclusion
A comparison of ASTM Method
D 3612 A standard procedure and
the equivalent GC headspace
technique, shows that the GC
headspace sampling technique
compares favorably with the
vacuum gas extraction method.
Thirty-five samples of transformer
oil were analyzed using both
methods and the GC headspace
results showed excellent
correlation with the ASTM method.
The high sensitivity of detection
achieved and the fact that the GC
headspace system is automated are
major advantages. The GC
headspace technique can be used
to assess dissolved gases from any
equipment where the total gas
content is very low. The
mechanical vibration of the
samples using the HP 7694
headspace sampler permitted the
analyses to be performed in a
much shorter time with less
handling than Method A of the
ASTM D 3612 standard.
Copyright 1995
Hewlett-Packard Company
Printed in USA 10/95
(23) 5963-7118E