3.0 INTRODUCTION
Floating Carriage Diameter is also commonly known as Effective Diameter Measuring
Micrometer. It use Bench micrometer to hold the workpiece as they look like they are
floating. This instrument is used for accurate measurement of Screw Plug Gauges
dimension such as Major Diameter, Minor Diameter and Effective Diameter. All these
dimensions such as Major Diameter, Minor Diameter and Effective Diameter. All these
dimensions have an important role in the screw plug gauges, since the accurancy and
interchangeability of the component depends on the gauges used. To minimize the effect
of slight errors in the micrometer screws and measuring faces, this micrometer is
fundamentally used as comparator.
R1
R2
=
=
D1
D major =
D 1+(R1 + R2 )
R3
The micrometer reading over Screw Pug Gauge and Core Prisms
R4
The diameter of Standard Cylinder
D1
Therefore,
Minor
Diameter
of
Screw
Plug
Gauge,
D minor =
D 1+(R3 + R 4)
Figure 3.1
c) Effective diameter : Also known as Pitch Diameter. It is the Diameter
of imaginary Cylinder,co-axial with the axis of the thread and intersects
the flank of the thread and intersects the flanks of the thread such that
width of the threads and width of spaces between threads are equal.
Referring to the Figure 3
R5
The micrometer reading over Screw Pug Gauge and Wires
R6
The diameter of Standard Cylinder
D3 =
Deff =
D3 + P
D1
D 1+( R5 + R6 )
where P is the
The P calue is a constant depending on the diameter of the wires (d), the pitch (p) and the
type of screw thread, it is found as follows:
Metric Threads (60)
P = 0.86602p d
P = 0.9604p 1.16568d
4.0 APPARATUS
2. Vtd Image Processing Measuring Device for Internal Screw Thread Measurement
Major diameter
1. The position of the fiducial indicator and the micrometer anvil and the gauge to
be measure is between or within the range of the micrometer.
2. A suitable Standard Cylinder was placed between the center the Standard
Cylinder should have diameter within 1mm of the actiual major diameter of
workpieace. The micrometer deum reading R1 is been noted
3. The standard was removed and was replaced with the Screw Plug Gauge to be
measured. The micrometer drum R2 is been taken.
4. The major diameter, Dmajor was evaluated
Minor diameter
1. The position of the fiducial indicator and the micrometer anvil and the gauge to
be measure is between or within the range of the micrometer.
2. A suitable Standard Cylinder was placed between the center. The selected
prism was been inserted between the standard and the anvils. The hanging
prism should be vertical. The reading on micrometer drum R3 was taken.
3. The standard was removed and was replaced with the Screw Plug Gauge to be
measured. The prism was inserted between the thread grooves and the anvils
and the micrometer reading R4 was taken.
4. The Minor Diameter, Dminor was evaluated
Effective Diameter
1. A suitable Standard Cylinder was placed between the center. The Standard
Cylinder should have diameter within 1mm of the actual diameter of
workpiece.
2. The wires selected were inserted in position between Standard Cylinder and
micrometer anvils. The wires were selected accordance N.P.L recommendation.
The reading on micrometer drum R3 was taken.
3. The Standard Cylinder was replaced with Screw Plug Gauge. Note the
micrometer reading R2 with the wre inserted in the thread grooves.
4. The Effectice Diameter, D eff was evaluated.
Micromete
r
Micromete
reading r
Micromete
reading r
Micromete
reading r
Micromete
reading r
Micromete
reading r
reading
over
over screw
standard
plug
and
thread and
standard
(mm)
and
core standard
prisms, R4 cylinder
core (mm)
prisms, R3
sleve
Thimble
Sub
3.5
0.47
0.002
3.5
0.35
0.002
Thimble
Sub Sleve
Total
0.0016
3.9756
0.0010
3.8530
(mm)
10.5
0.12
0.000
0.0002
10.6202
and
wire,
R6
wire, (mm)
R5 (mm)
7.5
0.04
0.006
7.5
0.46
0.004
4.5
0.18
0.002
0.0010
7.5470
0.0020
7.9660
0.0014
4.6832
Sample calculation
To determine R1
= Sleve + Thimbe + Sub-Thimble + Sub-Sleve
= 3.5 + 0.47 + 0.004 + 0.0016
=3.9756
*Repeat the same calculation method to find R2, R3, R4,R5,R6
= 30 + (3.8530 3.9756)
= 29.8774 mm
Minor Diameter Of Screw Plug Gauge
Dminor = D1 + (R4 R3 )
= 30 + (7.5470 10.6202 )
= 26.9268 mm
Effective Diameter
Diameter under the wire, D3
D3 = D1 + (R6 R5 ).
= 30 + (4.6832 7.9660 )
= 26.7172 mm
Metric Threads (60o), P = 0.86602p d
Where is p = 3.5 mm and d = 2.083 mm
P = 0.86602(3.5) (2.083)
= 0.9481 mm
Therefore,
Deff = D3 + P
= 26.7172 + 0.9481
= 27.6653 mm
Major Diameter
Minor Diameter
Effective Diameter
Theoritical Value ( mm )
30.000
26.211
27.727
Percentage Error =
ExperimentalTheoritical
Theoritical
gauge parameters that are needed to measure are major diameter, minor diameter and effective
diameter. From the specimen, the dimension of ISO metric thread and identification of threads
can be known. For instant, M30-3.5-ISO 6H where M is metric standard thread angle 60, 30
is major diameter (mm), 3.5 is pitch of thread (P) and ISO 6H is matrix standard table
(tolerance).
During the experiment, the major diameters have been measured three times and the
average major diameter can be calculated. The average diameter is 29.959 mm and the actual
major diameter is 30.00 mm. The experimental major diameter varies slightly from the actual
major diameter. The percentage error can be calculated which is 24.83%. The average minor
diameters have been measured and calculated which is 24.89 mm but the actual minor
diameter is 25 mm. By comparing the actual minor diameter and the measured diameter the
percentage error is 19.94%. The effective diameter has been measured which is 27.67 mm
whereas the actual effective diameter is 27.5 mm.
There are many possibility ways that errors might occur in readings for each time the
readings are taken. There may occur due to miss alignment of the specimen since there is no
fix location to set the specimen in a correct coordinate. Other possible errors are parallax error
which occurred when eyes positioning is misplaced during setting the alignment. The error
also may be involved due to the system error where apparatus or machine is not well
calibrated. Besides, students familiarity towards the machine may affect the error where
students may not handle properly during the experiment.
Other than that, for the result obtained from the coins, we found 0.178mm diameter
marbles in each hole of the congkak on the 10 cents coin. As we counted all the marbles, the
total marbles that can be found on the 10 cents coin are 45. For the new 20 cents coin, we
found the B,N and M letter between the big flower and small flower. Other than that, for the
new 50 cents we also found the B,N and M letter at the floral design.
There are necessary precautions which must be implementing to minimize the error
during the experiment. Students need to know how to handle properly throughout the
experiment and full attention must be given when conducting the experiment.
12
1. The surface of the specimen is needed to be clean from all contaminant such as oil,
dust, moisture and many more to ensure the accurateness of each reading.
2. The specimen must be placed at the correct position where the projector can show the
image clearly so that readings can be taken correctly.
3. Eyes level must be placed perpendicular to the readings to avoid parallax error.
8.0 CONCLUSION
In a nutshell, we can conclude that the experiment is successfully conducted as we has
fulfill the objective of the experiment which is to check the profile and measure the
geometrical parameters of the given specimen (screw thread) by using profile projector.
For the coins experiment, the purpose of the result obtained from the coins is for the
national security so that people cannot duplicate our coins from our country easily.
From the profile projector, major diameter, minor diameter and effective diameter of
the specimen can be determined. By comparing between the experimental data and actual
13
data, the value that we obtained was closed to the standard value. Hence, we consider the
experiment is successful as the percentage errors are more likely below than one percent.
9.0 RECOMMENDATION
1. Other optical system than using profile projector is Coordinate Measuring Machine
(CMM) where the measurement data taken from CMM is more precise and less error
contribute to the experiment. The measurements are more likely defined by a probe
attached to the third moving axis. This machine can be manually controlled or it may
computer controlled.
14
2. It is recommended for the profile projector to have a high resolution and a more
focusing lens to have a sharp view on the screw thread so that when alignment is being
adjusted, there would be no flaw or error exists throughout the experiment.
3. Before given to the students, it would be recommended if the technician could check
the screw plug gauge quality if any defects could detect prior to experiment. Thus, it
can minimize the errors.
10.0 REFERENCES
http://www.engineersedge.com/screw_threads_chart.htm
15
4. http://www.threadcheck.com/the-three-wire-method-of-measuring-pitchdiameter/technicalinfo/