Professional Documents
Culture Documents
DOC316.53.01095
Direct Method1
Method 10128
Test preparation
Light shield
Adapter
DR 5000
DR 2800
LZV646
DR 2700
LZV646
DR/2500
DR/2400
5945700
Quantity
Total Organic Carbon Direct Method High Range Test N Tube Reagent Set
DRB200 Reactor
pH Paper
Magnetic Stirrer
Quantity
Water, organic-free
0.3 mL
1
Direct method
2. Use a graduated
cylinder to add 10 mL of
sample to a 50-mL
Erlenmeyer flask that
contains a stir bar.
Stored Programs
426 Organic Carbon HR
Start
Zero
14.
Read
Interferences
The Interfering substances table lists substances that have been tested for interference and found
not to interfere up to the levels indicated.
If the sample contains greater than 1000 mg/L CaCO3 alkalinity, lower the sample pH to less than
7 before testing by adding Sulfuric Acid Solution.
Most sample turbidity is either dissolved during the digestion stage or settled during the cooling
period. Sample turbidities up to 50 NTU have been tested without interference.
Interference level
Aluminum
10 mg/L
Ammonia Nitrogen
1000 mg/L as N
ASTM Wastewater
No effect
Bromide
500 mg/L Br
Bromine
25 mg/L Br2
Calcium
Chloride
5000 mg/L
Chlorine
10 mg/L Cl2
Chlorine Dioxide
6 mg/L ClO2
Copper
10 mg/L
Cyanide
10 mg/L CN
Iodide
50 mg/L
Iron (II)
10 mg/L
Iron (III)
10 mg/L
Magnesium
Manganese (VII)
1 mg/L
Monochloramine
Interference level
Nitrite
Ozone
2 mg/L O3
Phosphate
Silica
Sulfate
Sulfide
20 mg/L S2
Sulfite
50 mg/L SO32
Zinc
5 mg/L
Rinse the sample bottle several times with the sample to be collected.
Accuracy check
Standard additions method (sample spike)
Required for accuracy check:
TenSette Pipet
1. After reading test results, leave the sample cell (unspiked sample) in the instrument.
2. Select standard additions from the instrument menu:
Instrument
Navigate to:
DR 5000
OPTIONS>MORE>STANDARD ADDITIONS
DR 2800
OPTIONS>MORE>STANDARD ADDITIONS
DR 2700
OPTIONS>MORE>STANDARD ADDITIONS
DR/2500
OPTIONS>STANDARD ADDITIONS
DR/2400
OPTIONS>STANDARD ADDITIONS
3. Accept the default values for standard concentration, sample volume and spike volumes. After
the values are accepted, the unspiked sample reading will appear in the top row. See the user
manual for more information.
Organic Carbon, Total
Page 5 of 8
Navigate to:
DR 5000
OPTIONS>MORE>STANDARD ADJUST
DR 2800
OPTIONS>MORE>STANDARD ADJUST
DR 2700
OPTIONS>MORE>STANDARD ADJUST
DR/2500
OPTIONS>STANDARD ADJUST
DR/2400
OPTIONS>STANDARD ADJUST
5. Turn on the Standard Adjust feature and accept the displayed concentration. If an alternate
concentration is used, enter the concentration and adjust the curve to that value.
Method performance
To test for TOC levels below 100 mg/L C, use Method 10173 or Method 10129.
Program
Instrument
Standard
Precision95% Confidence
Limits of Distribution
SensitivityConcentration
per 0.010 Abs
426
DR 5000
350 mg/L C
337363 mg/L C
4 mg/L C
DR 2800
350 mg/L C
337363 mg/L C
4 mg/L C
DR 2700
350 mg/L C
337363 mg/L C
4 mg/L C
DR/2500
DR/2400
100 mg/L C
14 mg/L C
4 mg/L C
250 mg/L C
13 mg/L C
400 mg/L C
15 mg/L C
550 mg/L C
23 mg/L C
700 mg/L C
34 mg/L C
Summary of method
The total organic carbon (TOC) is determined by first sparging the sample under slightly acidic
conditions to remove the inorganic carbon. In the outside vial organic carbon in the sample is
digested by persulfate and acid to form carbon dioxide. During digestion, the carbon dioxide
diffuses into a pH indicator reagent in the inner ampule. The absorption of carbon dioxide into the
indicator forms carbonic acid. Carbonic acid changes the pH of the indicator solution which, in
turn, changes the color. The amount of color change is related to the original amount of carbon
present in the sample. Test results are measured at 598 and 430 nm.
Unit
Catalog number
50 vials
2760445
50/pkg
0.4 mL
25 mL
45233
each
2584335
TOC1
10/pkg
50/pkg
5/pkg
39133
0.3 mL
500 mL
2641549
Quantity/Test
Quantity/Test
Unit
Catalog number
each
50838
each
LTV082.53.40001
each
LTV082.52.40001
each
50541
Magnetic Stirrer
each
2881200
each
1970001
50/pkg
2185696
each
1970010
50/pkg
2199796
each
4531500
13, 5.2 N
each
1864100
280/pkg
2097000
Description
Unit
Catalog number
500 g
31534
5/pkg
2791505
Unit
Catalog number
Pipet,
TenSette,
0.1 to 1.0 mL
TenSette,
1.0 to 10.0 mL
Kimwipes
Recommended standards
100 mL
244932
1000/pkg
2185628
250/pkg
2199725
115 VAC
2936701
Weighing Paper, 76 x 76 mm
500/pkg
1473800
50 mL
1457441
1000 mL
1457453
HACH COMPANY
WORLD HEADQUARTERS
Telephone: (970) 669-3050
FAX: (970) 669-2932