GasAlertMicro 5

Quick Reference Operations Key


ON: Press A. Audible/visual/vibrator alarms activate. The
current TWA, STEL, low, and high alarm setpoints are
displayed. A self-test is performed during the startup
sequence.
OFF: To turn off the detector, press A and hold until the
countdown is complete.
BACKLIGHT: In alarm (auto); in low-light (auto, if
enabled); and on demand (press and release any key in
low-light conditions to reactivate.)
GAS ALARMS: GasAlertMicro 5 is equipped with TWA,
STEL, low, high, multi-gas, and over range (OL) alarms.
Audible/visual/vibrator alarms activate; affected gas
readout blinks advising which gas hazard is present;
L icons light advising alarm level and type present.
PUMP ALARM: Audible/visual/vibrator alarms pulse
slowly. ALARM and J icons flash.
LOW BATTERY ALARM: flashes and audible/visual
alarms pulse. Change the batteries in a non-hazardous
atmosphere.
VIEW TWA, STEL, and PEAK (MAX) EXPOSURES:
Press and hold C. TWA, STEL, and peak exposures are
displayed in ppm or %.
TWA = accumulated exposure to toxic gases averaged
over a work day based on the OSHA or ACGIH method
(Not applicable to O2 and LEL)
STEL = short-term exposure to gas based on a 5-15 min.
user selectable period (Not applicable to O2 and LEL)
PEAK (MAX) = maximum concentration encountered*
*For Oxygen, it is the highest or the lowest concentration
encountered.

iERP:
Order Number:
Size:
Card Stock:
Laminate:
Print Color:
Punched Hole:

Leave the detector ON for the entire work shift.


TWA, STEL, and maximum concentration
readings reset if the detector is left off for more
than 5 minutes.
To clear TWA, STEL, and maximum gas exposure
readings, press and hold C and G simultaneously.
CALIBRATION:
1. In a clean atmosphere, press and hold C and H until
the countdown is complete.
2.
flashes as the detector zeroes the gas
sensors and calibrates the oxygen sensor.
3. When K flashes, connect the calibration bottle and
apply gas at a flow rate of 250-500 ml/min. to span the
sensor(s). Repeat until all sensors have spanned. When
complete, remove the gas.
4. Press C to set the next calibration due date. Press H
or G to set the due date and press C to save. Once all
due dates are set, press C to set the alarm setpoints.
5. Press H or G to change the alarm setpoint and press
C to accept the value. Once all the setpoints have been
set, the display reads Saving calibration indicating
calibration is complete.
APPROVALS:

n ATEX X g

and
others

For a full understanding of the GasAlertMicro 5s


features, operation, calibration and service, read the
user manual.
To contact BW Technologies call:
USA: 1-888-749-8878
Europe: +44 (0) 1869 233004

Canada: 1-800-663-4164
Other: 1-403-248-9226

iERP: 117359

117359
D5613/1
3.00 x 4.4 inches
10 point white
5 mm
Two: 1. Pantone Reflex Blue C ; 2. Black
0.315 inches; upper left as shown

D5613/1

You might also like