You are on page 1of 2

PRODUCT INFORMATION

LABORATORY ANALYSIS
SENSION 7

SENSION 7 - Measurement of Conductivity,


TDS and Salinity in the lab
4-electrode conductivity cell for
measurement of the complete range
from pure water to waste water
Linear temperature compensation
Non linear (natural water) temperature
compensation according to DIN 38 404
and EN 27 888
One point calibration or manual input
of cell constant
Data memory

LABORATORY_ANALYSIS_SENSION7

Technical data

TDS
Salinity
Temperature
Resolution
Conductivity

TDS

Salinity
Temperature
Accuracy
Conductivity
TDS
Salinity
Temperature
Outputs
Inputs
Power Requirements
Autoread
Data memory
Display
Conversion Factor for TDS
Temperature Compensation

Calibration
Calibration Rewiev
Power Requirements
Waterproof
Dimensions

DR. BRUNO LANGE


GMBH & CO.KG
Willsttterstrae 11
D-40549 Dsseldorf
Tel. +49 (0)2 11 52 88-0
Fax +49 (0)2 11 52 88-143
info@hach-lange.de
www.hach-lange.com

0 to 19.99 S/cm; 20 to 199.9 S/cm;


200 to 1999 S/cm;
2 to 19.99 mS/cm; 20 to 199.9 mS/cm
0 to 50000 mg/l as NaCl
0 to 42 g/l (ppt)
- 10.0 to 110.0C
0.01 S/cm
0.1 S/cm
1 S/cm
0.01 mS/cm
0.1 mS/cm
0.1 mg/l
1 mg/l
0.01 g/l
0.1 g/l
0.1 ppt
0.1C
0.5% of range
0.5% of full scale
0.1 ppt (-2 to 35C)
0.3C at 0-70C;
1.0C at 70-110C
One way, RS-232
One 5-pin with temperature
Line power from 110 VAC
to 220 VAC line power adapter
On/off
Interval data transfer to Printer/PC
99 point memory
Custom LCD
Adjustable automatic or
user-adjustable
Non linear (natural water, according
to DIN 38 404 and EN 27 888)
Linear coefficient [numeric value]%/C
No compensation
Reference temperature 20 or 25C
Conductivity Test: one point
calibration with standard solution.
Manual input of cell constant.
User-activated; shows last calibration
date, standards used
Line power from 110 VAC to 220 VAC
line power adapter
IP32
15 x 25.4 x 8.37 cm (10.15 x 6 x 3.5")

HACH LANGE LTD


Lennox Road
Basingstoke
Hampshire, RG22 4AP
Tel. +44 (0) 12 56 33 34 03
Fax +44 (0) 12 56 33 07 24
info@hach-lange.co.uk
www.hach-lange.com

INSTRUMENT
SENSION 7 Benchtop Conductivity Meter without
conductivity cell
SENSION 7 Benchtop Conductivity Meter with 4-electrode
conductivity cell (1 m), 115 V
SENSION 7 Benchtop Conductivity Meter with 4-electrode
conductivity cell (1 m), 230 V
CONDUCTIVITY STANDARD SOLUTIONS
180 10 S/cm, 90 5 mg/l TDS (85.47 mg/l as NaCl),
100 ml
1000 10 S/cm, 500 5 mg/l TDS (491 mg/l as NaCl),
100 ml
1990 20 S/cm, 995 5 mg/l TDS (1000 mg/l as NaCl),
100 ml
18000 50 S/cm, 9000 25 mg/l TDS (10246 mg/l
as NaCl), 100 ml
53000 S/cm, 35.0 ppt salinity, 500 ml
SINGLET 1000 10 S/cm, 500 5 mg/l TDS (491 mg/l
as NaCl), pk/20, 25 ml
SINGLET 18000 50 S/cm, 9000 25 mg/l TDS
(10246 mg/l as NaCl), pk/20, 25 ml
SINGLET 53000 S/cm, 35.0 ppt salinity, pk/20, 25 ml
OPTIONAL ACCESSORIES
4-electrode conductivity cell (1 m)
4-electrode conductivity cell (3 m)
LIS chamber (low ionic strength samples)
HachLink 2000 software for PC
Computer serial (RS232) cable
Kyoline Brand Thermal Printer with cable, 230 V
Kyoline Brand Thermal Printer with cable, 115 V
Thermal Paper for Kyoline Printer, pk/10 rolls

ART. NO.
5450001
5450010
5450011
2307542
1440042
210542
2307442
2714349
2770520
2770720
2770820
5197500
5197503
5189900
4966500
4967100
A70P020
A70P021
A70P025

HACH LANGE Services


Orders, information and
consultation without delay:
Tel. + 44 (0)12 56 33 34 03
www.hach-lange.com
- up-to-date and secure,
with downloads, information
and shop.

DOC052.52.00417

Range
Conductivity

Ordering information

P.O. Box 16 (Palokorvenkatu 2)


FI-04261 Kerava, Finland
Tel. +358 10 417 4500
Fax +358 10 417 4501
hyxo@hyxo.fi www.hyxo.com

You might also like