Operating instructions

ATMOSoft ENT 31
ATMOSoft ENT 41
English

REF 700.0030.A
REF 700.0032.A
REF 700.0035.A
2013-11 Index: 09

Table of Contents
1.0
1.1
1.2
1.3
1.4
2.0
3.0
3.1
3.2
3.3
3.4
3.5
3.6
3.7
3.8
3.9
3.10
3.11
3.11.1
3.11.2
3.10
3.10.2
3.10.3
3.11
3.12
3.13
3.14
3.15
4.0
4.1
4.1.1
4.1.2
5.0

Introduction
Notes on operating instructions
Function
Explanation of pictures and symbols
Scope of supply
For your safety
Setting up and starting up
System requirements
Installation
Options
user configurations
System configuration
Video Settings
Database Settings
Licence Settings
Licensing with ATMOSoft Dongle
Basic functions
Editing of images and videos
Editing images
Editing videos
Patient management
The patient management form
Optional iHandle software
Modules
System
Functions of the operator buttons on the ATMOS Scope handle
User interface
Printout and documentation
Troubleshooting
Permanent power supply of the USB port
Windows 7 (32 Bit/64 Bit)
Windows XP
Technical specifications

3
3
4
5
5
6
7
7
7
9
10
10
12
12
12
13
14
15
15
16
17
18
19
20
20
21
22
22
23
23
23
24
25

1.0 Introduction
1.1

Notes on operating instructions


DVD

These operating instructions contain important notes on how to operate


the ATMOSoft ENT safely, correctly and effectively. Their reading helps to
avoid risks, and also to reduce repair costs and down-times. That increases, amongst other things, the reliability and service-life of the software.
These operating instructions serve not only for new operating personnel
to be instructed in its use, but also for use as a reference manual. Reproduction of these instructions even in part only with the written permission of ATMOS.
These operating instructions must always be kept available near the
device.
The quality management system applied at ATMOS has been certified
according to international standards EN ISO 9001 and EN ISO 13485.
Prior to start-up please peruse chapter 2.0 For your safety, in order to
be prepared for any possible dangerous situations.

These operating instructions are valid for the following products


-ATMOSoft ENT 31
REF 700.0034.0
-ATMOSoft ENT 41
REF 700.0035.0
-ATMOSoft ENT Remote
REF 700.0032.0
-ATMOS Software Disc
REF 700.0031.0

ATMOS

MedizinTechnik GmbH & Co. KG


Ludwig-Kegel-Strae 16
79853 Lenzkirch
Deutschland
Tel. + 49 7653 689-0
Fax: + 49 7653 689-190
+ 49 7653 689-292 (Service Center)
atmos@atmosmed.de
www.atmosmed.de

ATMOSoft ENT 31
Visualisation and archiving software
Performance features:
-Software for the visualisation of video signals on
personal computers. Up to 8 pictures can be
taken at the same time and saved in either PDF
or Word documents.
Additional features
-Saving of individual images (JPG file format)
-ATMOS patient database for the management
of diagnostic results (compatible with ATMOS
Diagnostic Cube)
-GDT-connection
-Optimisation of the video signal (luminance,
sharpness, contrast)
ATMOSoft ENT 41
Visualisation and archiving software
Performance features:
-Software for the visualisation of video signals on
personal computers. Up to 8 pictures can be
taken at the same time and saved in either PDF
or Word documents.
Additional features:
-Saving of individual images (JPG file format)
-ATMOS patient database for the management
of diagnostic results (compatible with ATMOS
Diagnostic Cube)
-GDT-connection
-Optimisation of the video signal (luminance,
sharpness, contrast)
-Saving videos

1.2

Function

Name:

ATMOSoft ENT 31, ATMOSoft ENT 41

Main function:

Visualisation and documentation of diagnostic results

Med. indication / application:

Standard ENT examination

Specification of the main functions:

Saving and editing of images (JPG file format)


Saving and cutting of video sequences
Saving and amplification of audio signals
Selection camera via key combination is possible
HD recording is possible
ATMOS patient database for the administration
of diagnostic results (compatible with ATMOS
Diagnostic Cube)
GDT interface
Optimisation of the video signal (brightness, sharpness, contrast)

Application:

Patients of all ages - no restrictions - who require a


standard ENT examination

Application organ:

Throat, nose, ear

Application time:

Temporary application

Application site:

Application sites are hospitals, doctor's oces and


phoniatricians

Application restrictions:

None

With the ATMOSoft ENT you can easily and quickly safe and archive diagnostic results.
Application of the ATMOSoft ENT:
Saving, editing, archiving, printing and conversion into PDF files of diagnostic results.

1.3

Explanation of pictures and symbols

Short cuts / symbols contained in this manual


Follow the arrows whilst
proceeding, sequence

General information

Move, plug ... in this


direction

Please press where the dot


indicates

Numeration

Turn, shift ... in this


direction

Use oprinal foot switch

Subnumeration

Replace

Check

Engage, check correct fit

Important information

Dispose of properly

Order number

Manufacturing date

Please read,
important information

Pictures contained in this manual

Warning,
especial diligent notice !

Symbold of the ATMOSoft ENT


SN

Serial number

REF

Please observe operating


instructions

1.4

Scope of supply

Prior to dispatch, this ATMOSoft ENT software was subjected to an extensive functional test and has been carefully packed. Nevertheless, please compare the contents of the shipment on completeness immediately upon
receipt (see delivery note).
Scope of supply:
ATMOS Software Disc
ATMOSoft Dongle
Optional:
ATMOSoft Remote

2.0 For your safety

The ATMOSoft ENT may only be used by trained personnel which has been authorised and trained by
ATMOS. It is only to be used in a supervised working
surrounding (IEC 601-1/EN 60601-1)
Please take care that the pictures taken are assigned
to the matching patients when manually saving the
pictures. It lies within the duty of care of the operator.
It is the operator's responsibility to save the produced data.

3.0 Setting up and starting up


3.1

System requirements

System:
Processor :
Board memory (RAM):
HDD:
USB Interface:
Monitor resolution:

3.2

WinXP, Win7 oder Win8 (not for Windows Vista)


min. 2 GHz
2 GB
15 MB Software only
80 GB Data base storage
2 x USB 2.0
800 x 600 min.

Installation

When the Installation CD is placed in a CD or DVD drive, the software should start automatically. If not, open the
contents of the CD and start the Install.exe manually.
Please make sure that full administrative rights are granted for the user in order to be able to install the software
and drivers. Read and write permission for the working partition C:\ and the data storage partition are essential
for the function of the diagnostic software.
Follow the steps as described below to perform the installation:
1. Choose the desired language.

2. Click at the button "ATMOSoft ENT". The installation


wizard should open automatically

3. Accept the Licence Agreement

4. Click "Next" to configure the software

5. Open the software


Click on the button
"toolbox" to go to the
adjustments menu.

3.3

Options

Click onto the register "camera" to do the necessary adjustments. To do this the camera needs to be installed
and connected to the device.

Select your video input

Selecting the
video signal
(Composite or
S-Video)

Select the
resolution

Click "OK" to close the


window and save the changes.

Window closes without


saving the changes.

The software recognises already installed cameras. Please choose the required camera.

Adjustments to the camera settings should only be done by experts.

3.4

user configurations

Click on the register "Users"

Type the
name of your
facility into
this space.

Standard user
needs to be
selected.

!
3.5

The user configuration will be visible only after a restart!

System configuration

Click onto the register "System". The configuration of the system can be performed as follows:

Complete
file name for
the export of
examination
files into the
EDP.

!
10

Up to 6 different users can


be defined
using an up
to 3 letter or
number code
for identification.

Path to
language file.
It will be set
automatically
if the language of the
Windowssettings is
implemented
into the software.
Path to
folder where
PDF files will
be stored for
documentation.
Complete
file name for
the import of
patient files
into the EDP.

These are only examplary paths. They need to be coordinated with the network administrator.

Menu overview
Click onto the register "Settings"
In "Picturename" eight individual
names can be given to different
images.

Example:
Definiton of the text processing
(e.g. WinWord) for further procession of the documentation results.

Example:
Programme to display PDF-software.

Choose the file name to be represented on the PDF:


Number of patient
Date of examination
Path to logo-file (jpg or bmp) and
text to be added on instant printout for print button.

The template for the parameter field can be a *.doc, *.docx, or *.dot file and should be stored inside the AGHDiag
folder for free access to all users.

These are examplary paths only and need to be coordinated with the network administrator.

Please coordinate the storage space for the AGHDiag with your network administrator.

11

3.6

Video Settings

Camera settings should only be done by experts or


technicians to optimise the camera performance.

3.7

Database Settings

Complete file name of the database

Path to the picture database


Path to the video database
Path to the patient database

3.8

Licence Settings
Hardware specific serial number for the purchase of
a licence (activation code). The activation code will be
supplied by your ATMOS contact person.
By clicking on "Formula" a window which has to be filled
in completely opens up.
By clicking on PDF you can generate a PDF file which
you can send to your ATMOS contact person via e-mail
or post.

!
12

Please note that the individually generated activation code enables you for the use on this
computer.

3.9

Licensing with ATMOSoft Dongle

To perform the licencing via the ATMOSoft Dongle follow the steps below:
1. Install the software according to the steps 1 to 4 of the guide on the pages 7 + 8.

2. Connect the USB-Dongle with a free USB socket on


your computer.

3. Start the ATMOSoft

Note: The software cannot be licenced without a correctly connected dongle.


Note: Prior to every start of the ATMOSoft the Dongle has to be connected as otherwise the licensing cannot
be performed.

13

3.10 Basic functions


Button for freeze
frame or activating the
camera

Selection button will jump one frame each time the button is used.

Open files Loading of archived measurements

Close files Saving of measurements


Open the patient management (only in connection with the "BDE"
software)

Open settings
Cancel the programme

Buttons for increasing or decreasing the picture scale or to return


to the original scale of the camera
settings.

Scale

Creates a Word document according to the programme settings.

Creates and stores a PDF document according to the programme settings


Printout of the documentation

14

Preview button for printout.

3.11 Editing of images and videos


The editing menu for images and videos can be started either by pressing the F10 key on your keybord or by
clicking on the right mouse key and by selecting the oder "editing".

3.11.1 Editing images


Place diagnosis marker
Select colour

Open text field


Adopt editing

Close

To perform editing work on your images click on the


neccessary function.
Click with pressed Shift-key to select the position for
markers and texts.
By turning the scroll-wheel on the mouse and with
pressed Shift-key the size of text and objects can be
scaled.
To change the colour select the "select colour"
button, choose the colour and click onto the the
object you want with pressed Shift-key to change the
colour.
To place text on the image select the "open text
field" button and type in the text. By clicking into the
image with pressed Shift-key the text can be placed.
By clicking on the button "adopt editing" the changes on the image will be taken over. Any objects on
the image that were placed until then cannot be
changed anymore.

15

3.11.2

Editing videos
Selcting the start point
Selecting the end point

Open text field


Select colour

Adopt editing
Close

Note:

To edit videos it is neccessary to save them beforehand.

By pressing the button "Selecting the start point" the beginning of the video can be chosen.
By pressing the button "Selecting the end point" the ending of the video can be chosen.
To place a comment click on the button "Open text field" an type in your text. To place the text on the video
click into the image with pressed Shift-key.
To change the colour of a comment click on the button "Select colour". Choose the colour. By clicking on the
comment with pressed Shift-key the colour of the comment is changed.
By clicking on the button "Adopt editing" the changes on your video will be saved.
Note:

The video is editable as long as it has not been clicked on the button "Adobt editing".

Note: The edited video is not finally saved until the button "save changes" which appears when closing the
application is selected.

16

All spaces need to be filled in completely; choose the desired version of the ATMOSoft under the menu "Version".

3.10 Patient management


By clicking on the "handshake" button the patient
management can be opened out of the measurement
screen.

If the ATMOSoft is not integrated into a practice EDP the software can be managed in the database directly.
If integration into the practice EDP is available (GDT interface) access to the patient management is blocked because all patient-related data is automatically adopted.

If examination results shall be saved a reference to the patient needs to be included.

17

3.10.2

The patient management form


In the register "Table" all currently saved patients are
listed.

Searching via patient number


Searching via name
Searching via first name
If you click on the button "without patient data" you
will continue without a patient database.
Window closes without selecting a patient.
"Ok" opens the previously selected patient.
Via the register "New" patients can be added.

Enter patient number


Enter name

Enter first name

Enter date of birth


If you click on the button "without patient data" you will continue without a patient database.
Window closes without selecting a patient.
"Ok" opens the previously selected patient.

Via the register "change" you can alter previously entered patient data..

Save changes
Undo changes
If you click on the button "without patient data" you
will continue without a patient database.
Window closes without selecting a patient.
"Ok" opens the previously selected patient.

18

3.10.3

Optional iHandle software

The ATMOS Remote (REF 700.0032.0) is required when using the automatic start-up by pressing a pushbutton
at the handle.
To use the iHandle software you have to connect the controller of the ATMOS Scope to the control box of the
ATMOSoft ENT. The software of the ATMOS Diagnostic Cube and the ATMOS Scope needs to be installed.

The control box needs to be connected in order to use the buttons


of the ATMOS Scope. (USB interface)

By using one of the two control


buttons on the ATMOS Scope, the
software will be started.

Click on the "toolbox" button to


do changes on the modules.

Scope31

Scope31

19

Usually the first installation of the iHandle is done by our service technicians or our competent partners.
The technician will also install the required drivers for the USB ports. After manually starting the iHandle software
the setup can be executed in the options menu. The iHandle software will afterwards be placed in the "autostart"file of the Windows operating system. It is then automatically available.

3.11 Modules
The ATMOS Scope needs to be selected in the register "Module".
The selected ATMOSoft ENT file.

Scope 31

These are only examplary paths. They need to be coordinated with the network administrator.

3.12 System
In the register "System the programme settings such as language and ID for the iHandle interface can be defined
for the Windows operating system. This is only possible in connection with the ATMOS Diagnostic Cube.

By pressing one of the two


operator buttons on the ATMOS
Scope the ID of the device will be
displayed. This ID must be adjusted
by the two arrow buttons.

20

These are only examplary paths. They need to be coordinated with the network administrator.

3.13 Functions of the operator buttons on the ATMOS Scope handle


The upper button is used to take / freeze the picture or
to reactivate the camera to take a new picture. Press
the button to freeze, press the button again to reactivate the camera.

The lower button is used to skip frames. Each time the


button is pressed, the software will move to the next
frame.

Option video recording (ATMOSoft ENT 41)

By clicking on the camera button the video function will be activated.

Via the two buttons on the ATMOS Scope the recording can be started and stopped.

21

3.14 User interface


The user interface of the ATMOSoft ENT is very simply structured to enable easy documentation of diagnostic results

The major part of the screen is occupied by the life picture taken by the camera and which can be saved by
pressing the upper button.

If the patient data is important it can be saved in the


ATMOS diagnostic patient data file or it can be exported into a practice EDP.
The documentation can either be stored electronically as data files in PDF or Word format or just as well
be printed out immediately.
Download images, respectively films and press F10 or
right-click and select image processing.

3.15 Printout and documentation


Documentations can either be saved as electronic data files or as paper printouts. The data file can be saved as
WinWord or PDF files.
Mircosoft Windows and Adobe Reader are not included in the scope of supply.
The PDF file will be created automatically when clicking onto the PDF-Button on the screen.
Use Ctrl + v to
paste the pictures
of the ATMOS
Scope 31 Software
into the Word document.

To create a Word document, click the


Word-Button and an empty document
will open.

22

Now the document can be handled like any other


Word document.

4.0 Troubleshooting
Description

Cause

Measure

No picture

Wrong device is selected in the


options

Choose the right device

USB is not connected

Connect USB

USB port is switched-off because


of the energy-saving option

Permanent power supply of the


USB port, make the relevant settings in the operating system (see
page 20-21).

Control box is connected improperly to the ATMOS Scope (R1/


R2)

Connect the control box properly

USB is not connected

Connect USB

Acrobat Reader is not installed

Install Acrobat Reader

Options are not saved correctly

Options are not saved correctly

Word not installed

Install Word

Options are not saved correctly

Options are not saved correctly

Remote output is deactivated in


ATMOS Scope setup

Activate remote output

Buttons on the ATMOS Scope


handle do not work

PDF button does not work


Word button does not work
Buttons on the ATMOS Scope
handle do not work

4.1

Permanent power supply of the USB port

4.1.1 Windows 7 (32 Bit/64 Bit)


Path to the energy options
All control panel Items Power Options Edit Plan
Settings

Over Change advances power settings you will get


to the option Power Options, please select USB settings and switch it off.
USB energy option settings

23

4.1.2 Windows XP
Please activate the system control of Windows XP and
with a double click on system you will get to the system properties.

Activate under system properties hardware and click


on device manager.

Now, please select USB Root Hub.

Remove the checkmark automatic switch-off in order


to save energy.

24

5.0 Technical specifications


System:
Processor :
Board memory (RAM):
HDD:
USB Interface:
Monitor resolution:

WinXP, Win7 oder Win8 (not for Windows Vista)


min. 2 GHz
2 GB
15 MB Software only
80 GB database storage
2 x USB 2.0

25

26

ATMOS General terms and conditions

MedizinTechnik

1. General:

case. Should the delivery delay be caused by a culpable infringement

is limited to damage which is regarded as typical for tthat case. This

Our General Standard Terms and Conditions apply exclusively. Clients

of non-substantial contractual duties, our client is also entitled to claim

also applies in the case of our culpable infringement of substantial

terms and conditions which are contrary to or deviate from our General

a one-off damage compen-sation worth 3 percentage points of the

contractual duties The indispensable conditions of German Liability

Standard Terms and Conditions are not recognised unless their validity

delivery value of the goods for each weeks delay, up to a maximum

Law remain unaffected thereby.

is explicitly confirmed in writing. Our General Standard Terms and

which is no higher than 15 percentage points of the delivery value of

Conditions also apply even if we deliver to clients without reservation,

the goods

- For second-hand equipment, the period of warranty shall be reduced


to a period of twelve months.

in the knowledge of the clients contrary terms and conditions. Our


General Standard Terms and Conditions also apply to all future business
with that client.

7. Delivery - Familiarisation
In the case of the delivery of devices for the medico-technical industry
which require assembly and/or familiarisation for the final customer using

10. Reservation of Ownership


We retain ownership of our goods until the receipt of all payments
arising from the business relationship, including all demands arising

2. Proposal - Order Confirmation

specialist trade personnel (such as Ear, Nose and Throat Apparatus and

from installation orders, subsequent orders, repairs, accessory deliveries

Our proposals are subject to change without notice unless otherwise

Suction Units), we reserve the right to deliver the goods exclusively to

and replacement orders. Should we have agreed upon payment on the

stated in our order confirmation. Each order is only accepted by us

the relevant specialist traders. Should the trader not carry out assembly

basis of cheque and bill transactions, the ownership reservation applies

following our written order confirmation.

and/or familiarisation for the final customer, this is carried out by us. In

until the cheque received by us has been paid in, and does not expire

such cases, we reserve the right to charge the client for the additionally

through our credit upon receiving the clients cheque. In the case of

3. Orders

created costs. Our specialist traders operate a recording system so

a breach of contract by the client, especially payment arrears, we are

Every order requires an exact description of all of our products details.

that, if necessary, our products can be traced to the final customer. The

entitled to repossess our goods. Repossession of our goods repre-sents

We assume no liability for errors and damage caused by inaccurate or

specialist trader undertakes to immediately report to us all events and

a withdrawal from the contract, unless explicitly declared in writing by

incomplete ordering details.

risks which must be reported in connection with our products.

us. We have the right to utilise the product after its repossession, whilst

4. Prices

8. Passage of Risk - Packaging

deducting appropriate utilisation costs.The client is responsible for

the income form such use is balanced against the clients arrears, after
Unless otherwise stated in the order confirmation, our prices in the

Unless otherwise stated in our order confirmation, delivery is agreed

order confirmation are ex factory prices and exclude packaging and

ex factory. The risk of the goods damage or loss is therefore transferred

be necessary, the client must carry these out punctually at his own cost.

value added tax. Packaging is charged separately at cost price in the

to the client as soon as the goods leave the factory or the client is in

Our client is entitled to sell the goods he has bought from us in a proper

handling the goods with care. Should maintenance and inspection work

invoice. Value added tax is charged separately in the invoice according

default of acceptance of the goods. This also applies to cases where we

sale transaction. However, he must immediately assign all outstanding

to the legal rate on the invoice date. We reserve the right to change

confirm prepaid carriage. Transport packaging and all other packaging

claims to the value of the final invoice sum (including value added tax)

prices appropriately should price reductions or increases, especially

according to the packaging regulations is not returnable. Our client is

of our claims to his customers or third parties. The client is entitled to

due to wage settlements, changes in the price of materials or currency

responsible for disposing the packaging at its own cost. Our deliveries are

collect this claim even after such assignment. Our right to collect the

fluctuations, be incurred. Proof of such changes will be provided for the

insured by us at the clients expense unless explicitly otherwise agreed.

claim ourselves remains unaffected thereby.We undertake to release

client on request.

No insurance is arranged in the case of goods which are collected by

the securities to which we are entitled if requested to do so by the

our clients. In the case of transport damage, claims are only handled if

client should the realisable value of the our securities be more than 10

5. Payment Conditions - Balancing

the client receives confirmation of any damage, reduced weight or loss

percentage points higher than the outstanding claims. We reserve the

Unless otherwise stated in the order confirmation, our invoices

by the shipping company before accepting the delivery.

right to choose the securities to be released.

9. Warranty

11. Plans and Illustrations

are payable with a 3% discount within 10 days (except for repair and
assembly services) or within 21 days from the invoice date net cash;
money receipts is decisive for complying with this term. We are entitled

The client is responsible for examining the delivered goods

We retain ownership of and copyrights to all plans, illustrations,

to charge interest after the due date at a rate 2% above the relevant

immediately after receiving them to determine any eventual deficiencies

calculations and other documents which are attached to our proposals.

basic interest rate of the German Federal Bank. Should the client have

or delivery errors, and to report these immediately. Should the client

The client must receive explicit written permission before passing these

payment arrears, we are entitled to charge interest on arrears at a rate

fulfil this examining and reporting responsibility, and should payment

on to third parties. Imitating our legally patented products is forbidden

5% above the relevant basic interest rate of the German Federal Bank.

conditions be fulfilled, we shall be liable to the client within the scope

and will be prosecuted.

Should we be able to prove higher damages due to arrears, we are also

of legal regulations. Our period of warranty shall in all cases be two

entitled to claim these. The client only has the right to balance invoices

years. Our client can make use of the warranty as follows, so long as

against its own claims should such claims be confirmed in a court of

he can provide first buyer proof (in the form of an invoice or delivery

Our central office is the place of performance for all disputes in

law or recognised by us. The client does not have the right of retention

note) and provided that the product still has the original, unchanged

connection with these General Standard Terms and Conditions and

due to disputed counterclaims.

serial number:

6. Delivery Periods
Fulfilment of our delivery duties requires the punctual and proper

12. Jurisdiction and Place of Performance

the contracts closed with clients under them. This jurisdiction excludes

a. We choose whether to fulfil our guarantee by providing repair

other jurisdiction relating to persons or subject-matter. Furthermore, our

services free of charge - either on the clients premises or in our factory

client is not entitled to bring charges against us in another court should

- or replacing the product. We can also provide these guarantee

he file counter-charges, carry out counterbalancing or declare retention.

fulfilment of the clients duties. The right to defense on the grounds of an

services through an authorised company;

We, however, are entitled to bring charges against our client at their
general place of jurisdiction or at another relevant court recognised by

unfulfilled contract is reserved.Should the client default in accepting the

b. Should a product be returned to us, the client agrees to send

goods delivery or breach other cooperation duties, we are entitled either

the product in its original or similar packaging, offering the same

German or foreign law.Unless otherwise stated in the order confirmation,

to withdraw from the contract or claim compensation for any increased

protection as the original packaging, to our address or any address

our central office is the place of performance.

costs incurred up to that time without setting a further deadline. The right

notified by us.

to make further claims is reserved. Furthermore, in such cases, the risk

c. Our guarantee ceases to apply if changes of any kind have been

of coin-cidental destruction or a coincidental deterioration in the quality

made to our product, unless such changes have been made by us

Lenzkirch, September 2008

of the delivered goods is transferred to the client in the case of default

or a company authorised by us, or have been previously agreed

ATMOS MedizinTechnik GmbH & Co. KG

in accepting such goods or payment arrears. Acts of God or stoppages

upon in writing by us. Our guarantee also ceases to apply if third

79853 Lenzkirch/Germany

(due to insufficient supplies of material, industrial disputes etc.) entitle

parties have carried out repairs to our products or replaced parts

us either to demand an appropriate extension of delivery periods or to

thereof. This applies regardless of the fact whether these measures

partly or entirely dissolve the delivery contract. This does not give the

individually or collectively led to a deficiency of the product;

client the right to claim damages. We have fulfilled delivery periods if the

d. We accept no responsibility for damage defects caused by

delivery goods have left our factory or the client has been informed of

- operational wear and tear;

the goods readiness for delivery within such delivery periods. Delivery

- incorrect installation or incorrect or insufficient maintenance;

periods stipulated by the client are not recognised by us unless they

- incorrect operation of the product (in contradiction to the operating instructions

form part of our order confirmation. We adhere to legal terms and

delivered with the product); - improper use or operating faults; -

conditions in cases where, as a result of an undue delay in the delivery

inappropriate or negligent handling and care, especially with respect

for which we are liable, the client is entitled to claim that his interests

to dirt, lime, suction of fluids, inappropriate cleaning and sterilisation;

in a continued fulfilment of the contract have ceased. We also adhere

- using accessories and/or replacement parts which are not explicitly

to legal terms and conditions should a delay in delivery be caused by

approved;

deliberate or grossly negligent action by us or our representatives for

- incorrect assembly and/or initial operation by the client or third

which we are responsible. We are also responsible for such actions by

parties; - the clients negligence in handling the product; - unacceptable

our representatives or agents. Should the delivery delay not be caused

operating conditions, such as humidity, temperatures, the power supply,

by our deliberate infringement of contractual duties for which we are

vibrations.

responsible, our liability is limited to damage which is regarded as typical

- accidents, acts of God, especially lightening, water, fire, public

for that case. We are liable according to the legal terms and conditions if

unrest and insufficient ventilation. We are not liable for damage to

and in so far as the delivery delay for which we are responsible is caused

other objects apart from our product itself, except in the case of any

by an infringement of a substantial contractual duty. In such cases, our

deliberate or grossly negligent actions by us or our representatives or

liability is also limited to damage which is regarded as typical for that

agents. Should no deliberate breach of contract be claimed, our liability

This document is copyrighted. Duplication, translations, microfilming and savings on electronic systems, particularly for commercialpurposes
are illegal without prior agreement of the manufacturer. All compiled data are based on manufacturers instructions. All logos,
product names and designations used in this document are property of the respective manufacturer.
We do not take over any warranty and liability in the case of missing inscriptions. Subject to modifications and amendments.

ATMOS MedizinTechnik GmbH & Co. KG


Ludwig-Kegel-Str. 16
79853 Lenzkirch / Germany
Tel: +49 7653 689-370
atmos@atmosmed.de

www.atmosmed.com

P094.xxxx.x 201x-xx Index: xx

A TM O S - S l o g an O p e n S ans R e g u l a r 8p t 17p t D u r c h s c h u L a u f w e i t e 100

You might also like