Professional Documents
Culture Documents
lOMoARcPSD
INHOUD
1
Distributing prohibited | Downloaded by Jan Maty (hanzus1@gmail.com)
lOMoARcPSD
References .................................................................................................................................................. 35
2
Distributing prohibited | Downloaded by Jan Maty (hanzus1@gmail.com)
lOMoARcPSD
An estimated 2,5 million of laparoscopic surgeries are performed in USA. Of all surgeries to the digestive system
almost 100% is done laparoscopic these days. These numbers are drawn from the interview of MD. Ronald A.
Rhodes [4]. Because of lack of other numbers we will use these numbers also for Europe in order to have a realistic
estimation based on real market needs.
In 2010 Europe counted on average 2,7 hospitals per 100.000 inhabitants [5]. With a population of nearly 500
million people, this makes a total of 13.500 hospitals. Lets say 70% [own estimation] is active in minimally invasive
surgery, leaving out small private clinics, and hospitals in less wealthy countries. This makes a total of 9450
hospitals. However, not all hospitals will be using PortShield so we estimate that 20% (1890 hospitals) of these
9450 hospitals will use it from the beginning.
An estimated average of more or less 265 [2.500.000/9450] laparoscopic surgeries are done annually in each
hospital.
This product will not be used for all laparoscopic surgeries since not all of the surgeries will be in the abdominal
area, and in the first instance hospitals might be holding on to older devices used in these kinds of procedures. An
estimation is done of an average 133 annually supplied laparoscopic surgeries per hospital [50% of estimated
laparoscopic surgeries].
Make a rough estimation of the potential annually sold numbers of the chosen medical device prototype in
Europe. Use estimates of the number of surgical procedures performed, number of surgeons or hospitals in
European countries, number of residents in Europe and an estimation of the expected market share.
3
Distributing prohibited | Downloaded by Jan Maty (hanzus1@gmail.com)
lOMoARcPSD
With 1890 hospitals and 3 trocars per surgery as stated by MediShield, this gives an estimate of 754.110 annually
sold items.
The number of clinicians is not taken into account, because this device is used only once and thrown away after
the procedure. Therefore it does not matter how many surgeries are done per day and by how many physicians
precisely.
Unit market share: The units sold by a particular company as a percentage of total market sales, measured in the
same units. [6]
So for 2,5 million laparoscopic surgeries where 3 trocars are used for each surgery, we estimate a total market unit
sale of 7,5 million items.
Unit market share (%) = 100 * Unit sales (#) / Total Market Unit Sales (#) =>
Unit market share (%) = 100 * 754.110 / 7.500.000 = 10%
All surgically invasive devices intended for transient use are in Class IIa as stated in [7] [8] (M5 paragraph 2.2 Rule
6).
4
Distributing prohibited | Downloaded by Jan Maty (hanzus1@gmail.com)
lOMoARcPSD
D | ISO 14971
Use the ISO 14971 standard and try and answer all questions stated in Appendix C. Skip the questions that do
not concern your chosen medical device prototype. Be critical in answering the questions, if you do not know
the answer or need additional information or calculations to support an answer please state this. It is a first
exploration of the current prototype is already suited to pass the CE-marking application. Based on the answer
of question 1b, choose whether the device will be a disposable or a reusable product.
Table 1 Answers to ISO 14971, Appendix C
2
3
4
8
13
14
15
20
22
23
24
25
26
27
Answer
This device is used to prevent gas leakage and shifting of the used trocar. The PortShield consists of
a balloon that can be rolled down the trocar tube before surgery. The trocar tip can then be
inserted in the patient using the surgeons own preferred technique. When the trocar has been
placed, the balloon is inflated, fixating the trocar in the port site. The variable wall thickness ensures
the anchor shape is maintained when an upward force is applied on the trocar. A solid wire inside
the balloon ensures the balloon can be fully deflated before removal indications of use (na vraag b)
The device is created to be used one procedure.
In case of breaking of the device in the body it can be possible that parts of the balloon have to be
retrieved from within the body.
if the device refuses to work no special intervention is necessary.
no, the device is not intended to be implanted
the device will be in contact with the opening in the human body during the endoscopic procedure.
This can be considered as invasive contact, since part of the device is used inside the body.
the device is made of rubber and plastic. It will be in contact with materials from used trocars.
The medical device is transient and used only once. It is supplied sterile and no sterilization by the
user is needed
the device is intended to be used in combination with regular trocars of sizes 3-5 and 10-12 mm.
Problems that can occur are bad fixation of the rolled down balloon to the shaft of the trocar. Also
penetration of the balloon when fixating it to the trocar can occur.
if not sufficiently inflated, the balloon may apply too much pressure on the incision or even burst.
rubber can dry when stored too long in a dry or hot environment. With very high temperatures
plastics may melt or burn.
it may be necessary to set an expiration date in terms of life expectancy of the balloon. The shelf life
of the balloon is significantly shorter than that of the body of the device. The body can be recycled.
forces applied are under control of the user. the balloon is inflated using a manual inflation device.
the expiration date of the balloon should be considered as an ageing factor for preventing the use
of the device.
Yes, device is intended for single use. The device has to be manually removed from the trocar. The
device will then probably be dirty and directly thrown away. The device should be used only right
after the package is removed.
The device has been in contact with the inside of a human body and it is possible that it has been in
contact with dangerous microorganisms. Therefore it has to be properly disposed of. Materials used
are recyclable.
A small instruction should be given but the basic insertion technique is the one surgeons were
trained to use in minimally invasive surgeries.
Instructions for use are provided directly to the end user by the manufacturer. Since the use only
considers three main actions specifically for this device. The rest are actions the surgeon is already
familiar with, depending on the procedure done.
Question *
1
5
Distributing prohibited | Downloaded by Jan Maty (hanzus1@gmail.com)
lOMoARcPSD
29.3
The only connecting part to the device is the pump used to inflate the balloon. Lack of feedback
might burst the balloon inside the patient.
The device will be used only by trained surgeons.
not inflating the balloon
deflating the balloon/not closing it properly during the procedure
not rolling the balloon down before inserting the trocar
over inflating the balloon
reusing it on another patient
29.7
31
E | USAGE
Describe step by step how the medical device should be used. To start this process, first identify all people who
will have to handle the device at some point.
Table 2 Usage of the medical device
Manufacturing
Step 2
transport to hospital
Step 3
storage in hospital
Step 4
unpacking
Step 5
attach to trocar
Step 6
Step 7
Step 8
Inserting trocar in
body
Attach air tube
Step 9
Keep sterility
quality control
Keep sterility
safety measures, so package will not
be damaged
Keep sterility
safety measures, so package will not
be damaged
keep sterility
follow instructions
keep sterility
mind not tearing the balloon
Step 1
6
Distributing prohibited | Downloaded by Jan Maty (hanzus1@gmail.com)
lOMoARcPSD
insertion.
Step 10
Step 11
Step 12
Performing planned
procedure
Deflating balloon
Step 13
Step 14
Rolling up balloon
Step 15
Detaching device
Step 16
Disposing device
--
to over inflation
7
Distributing prohibited | Downloaded by Jan Maty (hanzus1@gmail.com)
lOMoARcPSD
A | MISUSE
Read Appendix A.2.4, A.2.5, B and E of ISO 14971 standard. Use the same list as set up in Assignment 1e, but
now describe for each step/action the possible misuse of the device. Take into account the different levels of
misuse e.g. the intended use, intended users, process of use, and involved equipment.
Table 3 Possible misuse of the medical device
Description
Intended user R#
Assembly
Packaging
transport to
hospital
storage in
hospital
Risk
Level of misuse
R1
R2
involved
equipment
R3
process of use
R4
process of use
device has to be
packaged sterile
and vacuum
R5
intended
R6
involved
equipment
R7
involved
equipment
R8
intended
R9
>mishandled during
transportation
intended
R10
intended
R11
intended
factory
worker
hospital
personnel
Step # Name
8
Distributing prohibited | Downloaded by Jan Maty (hanzus1@gmail.com)
unpacking
O.R.
personnel
Nurse or
surgeon
rolling down
balloon
Nurse or
surgeon
the surgeon or
nurse rolls the
balloon over the
shaft of the trocar
Inserting
the surgeon inserts Surgeon
trocar in body the trocar into the
body of the patient
R12
intended
R13
process of use
R14
R15
process of use
R16
unintended
R17
intended use
R18
R19
unintended
R20
intended
R21
R22
process of use
10
R23
unintended
11
Inflating
balloon after
insertion.
R24
intended use
R25
intended use
R26
intended use
R27
intended use
R28
intended use
R29
Unintended use
R30
R31
12
Removing air
tube (?)
The nurse or
Surgeon
surgeon uses a
pump to inflate the
balloon inside the
body
the tube is
surgeon or
disconnected to
nurse
avoid it being in the
way of mobility of
intended use
lOMoARcPSD
9
Distributing prohibited | Downloaded by Jan Maty (hanzus1@gmail.com)
lOMoARcPSD
13
14
Performing
planned
procedure
Deflating
balloon
R32
R33
R34
material hazard
R35
process of use
R36
unintended
R37
material hazard
Device is passively
present during the
procedure
surgeon
failure at
manufacturing /
equipment
15
Removing air
tube (?)
the tube is
surgeon or
disconnected to
nurse
avoid it being in the
way of mobility of
the instruments. (?)
R38
process of use
16
Removing
trocar from
body
the trocar is
removed from the
body after the
minimal invasive
surgery has come
to an end
R39
process of use
R40
R41
R47
nurse
R42
Nurse or
Surgeon
R43
Nurse
R44
R45
R46
18
19
20
Detaching
device
Disposing
device
Recycling
by nurse (or
surgeon?)
surgeon
Process of use
Unintended use,
be sure to
dispose/destroy it
so it wont be
reused
10
Distributing prohibited | Downloaded by Jan Maty (hanzus1@gmail.com)
lOMoARcPSD
Severity* >>
negligible minor
Probability* >>
frequent
probably
serious
critical
catastrophic
R35
R38
R47
R10, R36
remote
R14, R16
R17, R18
R29
R43
Severity
Probability
Frequent
10
Probable
10
Occasional
10
Remote
10
10 >
10 >
10 >
10 >
Improbable
11
Distributing prohibited | Downloaded by Jan Maty (hanzus1@gmail.com)
lOMoARcPSD
C | QUANTIFICATION OF OVER-INFLATION
Some of these risks could be determined on a semi-quantitative probability level using calculations, simulations
and estimations. Indicate one risk that can be quantified in this more or less scientific manner. Describe in detail
how you would execute such quantification measurements to indicate the probability of that risks. The
quantification could for example be performed by describing an experiment to determine required data or by
describing technical calculations.
Definition of risk and hazard
The Irish Health and Safety Authority defines the term risk as follows: risk is the likelihood that a person may be
harmed or suffers adverse health effects if exposed to a hazard [9]. A hazard in this case is described as: a
potential source of harm or adverse health effect on a person or persons [9]. In other words: the risk is the
possibility that a person is harmed by a hazard and can be quantified on a probability scale. Where the hazard is
the direct cause of an injury.
Risk of over-inflation of the balloon (unintended use)
Over-inflation of the balloon is seen as a risk since it can cause the balloon to break, maybe even into multiple
pieces. This can lead to loose fractions of the balloon remaining in the abdominal cavity which is seen as a hazard
since it can directly do harm to the patient. Therefore the probability of over-inflating the balloon must be
determined to be able to take the next steps in controlling this risk.
Experiment to determine the risk of over-inflation
This experiment should consist of two parts. One is a technical experiment, to see at which pressure the balloon
will break. Second is a clinical experiment to see to what extent it is possible that the surgeon indeed inserts this
amount of gas into the balloon. From the outcome of these two experiments it would be possible how high the risk
of over-inflation really is.
st
12
Distributing prohibited | Downloaded by Jan Maty (hanzus1@gmail.com)
lOMoARcPSD
- multiple tubes
Test setup
The PortShield device will be clamped so the experimenter will have free hands to conduct the experiment. The
balloon will be connected to the air pump using tubes, with the barometer in between. (Fout! Verwijzingsbron
niet gevonden.)
Method
The balloon of the PortShield will be inflated until it breaks. This will be done a significant amount of times until a
proper conclusion can be formed about the maximum applicable pressure to the device. This maximum applicable
pressure will now be called breaking pressure (BP).
Results
From these two test, three results are obtained:
- Breaking pressure of the balloon
- Can the surgeon reach this breaking pressure when inflating the balloon manually
- How much differs the necessary pressure of the balloon from the breaking pressure: how big is the
safety margin?
13
Distributing prohibited | Downloaded by Jan Maty (hanzus1@gmail.com)
lOMoARcPSD
Conclusions
With these results it can be concluded how big the risk of over-inflation really is. When comparing the actual
breaking pressure with the pressure needed for proper fixation and the actual pressure that can be applied
manually it can be determined whether the risk on over inflation really is significant.
D | HAZARDS
Select two hazards from Assignment 2b. Indicate for each risk what type of measure is required: inherent safety
by design; protective measures in the device; and information for safety. Additionally, give for each hazard a
possible solution based on the selected measure.
Hazard 1
R25:
Measure required:
Protective measures in the device are interpreted as changes that are not immediately visible in the exterior of the
product as it is, but do contribute to the safety. In contrast to safety by design which is interpreted as a redesign
with clear visible changes in the design (like addition of parts)
To minimize the risk of over-inflation, a smart valve can be introduced to the device. Since the device already is
equipped with a valve this can be considered as a protective measure. The idea of the smart valve is releasing
pressure when pressure gets too high. This way the breaking pressure of the balloon will never be reached. It has
to be determined what the ideal release pressure would be which has enough range to breaking pressure but also
to necessary pressure for fixation and anti-gas-leaking.
Hazard 2
R10:
Measure required:
Warnings regarding proper environmental properties could be addressed on the packaging of the device. These
can include fragility, environment temperature, light and humidity warnings (Figure 3). Also an expiration date
should be added to the packaging (although his moreover includes storage).
14
Distributing prohibited | Downloaded by Jan Maty (hanzus1@gmail.com)
lOMoARcPSD
With this in mind it is thought of the add a safety slide to the device. This part will assure that when the trocar is
pulled out of the abdominal wall the valve to the balloon is automatically opened completely to quickly deflate the
balloon. This way the surgeon will not be able to pull out the trocar when the balloon is still inflated, since this
could cause rupturing of the abdominal wall.
There is a fine line between inherent safety by design, and protective measures in the device. Although this
addition does not change the final goal of the product, it is a large change in the looks and size of the product,
therefor we interpret this as inherent safety by design.
15
Distributing prohibited | Downloaded by Jan Maty (hanzus1@gmail.com)
lOMoARcPSD
As seen in Image 4 the safety slide is added to the top part of the PortShield. When the safety slide is pulled the
valve to deflate the balloon is broken so the balloon immediately deflate due to pressure induced by pulling it out
of the abdomen. The safety slide covers the top part of the trocar in such a way that the surgeon has to hold it
when pulling out the trocar. Since the trocar gets wider at the top, as seen in Figure 4, the safety slide is funnel
shaped. It has a cutout where the valve of the trocar is positioned.
F | CONCLUSIONS ON RISKS
Based on all information that you gathered for this assignment give conclusions on the risks of the medical
device.
Argue if the medical device prototype is safe to use after your redesign
As we already know, PortShield was designed in order to prevent gas leakage from the trocar and enhance fixation
during minimal invasive surgeries. Another driver for the design of the PortShield is costs, since fully disposable
trocars are very expensive [11]. Yet the use of a balloon in the abdomen brings risks that would not be present
when using a normal naked trocar. Creating a safety stop reduces one of the risks the balloon brings, namely:
pulling the trocar out before deflation. This risk could cause the hazard of rupturing the abdominal wall. However
this is not the only risk the balloon brings. In Table 3 more risks regarding the balloon can be seen. These are all
minor risks, but it does mean that the product is not risk free yet.
Suppose that you succeeded to reduce all risks to an acceptable probability level, argue if these residual risks
weigh up to the benefits of the device
In case we were able to reduce risks as stated in our table in 2b to an acceptable probability level, we believe that
residual risks would not only weigh up to the benefits of the device, but they would be overcome by the benefits.
This can based on that e.g. all catastrophic risks would become improbable and also remaining leakage probability
would be reduced significantly. Many other risks would be deleted and so the remaining risks, considering their
small severity, would not be able to weigh up to the benefits of the device. The benefit of this device is to make
minimal invasive surgeries safer by preventing as much as possible gas leakage from the trocar and providing the
same time flexibility to the surgeon without damaging the tissue or abdominal wall.
16
Distributing prohibited | Downloaded by Jan Maty (hanzus1@gmail.com)
lOMoARcPSD
PART 3 | MANUFACTURING
In this assignment we continue by exploring several aspects of ISO 13485 norm with a focus on the
manufacturing process of your selected medical device prototype and the quality control of that process.
ISO standard:
The ISO standard with guidelines on how to fulfill the chosen requirement is the ISO 18113-4:2009 [12]
Part 3 | Manufacturing
We decided to choose requirement 13.3 because of b) c) e) f) i) & l) of the requirement cover decisions which are
pointed out in Assignment 1d and 1e, like reusability, sterility, date on packaging.
17
Distributing prohibited | Downloaded by Jan Maty (hanzus1@gmail.com)
lOMoARcPSD
B | MATERIAL
Use the results of Assignment 1, estimated annual sales, to determine the main construction material of the
medical device and answer the following questions. Note: At this stage, you are allowed to make modifications
to the design of the prototype to improve the manufacturability, but keep the focus on the material for this
assignment.
Since the PortShield is a disposable device, it is favorable to be as cheap as possible, but still of high quality. The
estimated annual sales were estimated on 7.5 million items, which is high and thus allows it to produce the
product by mass production. The material which should be used has to be non-toxic and should be safe to use
inside the human body. To choose a right material, the Cambridge Engineering Selector (CES) is used. One of the
most important factors in choosing the material is that the material should be biocompatible. This means that it is
not causing injury, toxic or immunologic reaction to living tissue [CES]. If all the biocompatible materials are
selected in CES they can be categorized in different orders.
Part 3 | Manufacturing
The different biocompatible rubbers are shown in Figure 7, selected on their price. Here we can see that Ethylene
vinyl acetate is a cheap material. To see if this material has also good properties the yield strength and youngs
modulus are checked (Figure 8) with on the x-axis the yield strength, and on the y-axis the youngs modulus. Here
we can conclude that EVA has as good elastic properties as
18
Distributing prohibited | Downloaded by Jan Maty (hanzus1@gmail.com)
lOMoARcPSD
Part 3 | Manufacturing
Certifications and approvals may include but are not limited to:
19
Distributing prohibited | Downloaded by Jan Maty (hanzus1@gmail.com)
lOMoARcPSD
Phthalate-free
Change management
What companies can deliver this material?
Ateva Medical EVA polymers [14] offer medical device companies flexibility in product design. They are
biocompatible, stable and display excellent resistance to tissue growth. Moreover, they can be processed at low
temperatures using a variety of manufacturing methods. EVA is optically transparent, offers good adhesion and
flexibility, boasts excellent tear, puncture, impact and water resistance and is breathable and clinically accepted
around the world.
How would you monitor the consistency of the material once you have chosen a supplier?
To be sure that the material is of the right consistency a chemical analysis should be performed. This can be done
by chemical reactions or by x-ray [15].
By mechanical testing the material properties can be checked, for example yield strength, viscosity to make sure
that it is the right material. By X-ray the material structure can also be checked. [16]
C | MANUFACTURING FLOWCHART
Read paragraph 7.5 of ISO 13485 standard and use previous answers to questions. Make a flowchart from the
raw material to the final medical device product in which all necessary production machines/steps are present.
Discuss which one of these machines is most likely to cause risks/hazards or fluctuations in the production.
Propose a measure of control to monitor this risk or fluctuation.
Manufacturing process
The body component and the valve case are produced by injection molding. After manufacturing the different
components the device have to be assembled, including the ring, balloon, valve and valve case.
1.
Ethylene-vinyl acetate (also known as EVA) is obtained. Eva is the copolymer of ethylene and vinyl
acetate. The weight percent vinyl acetate usually varies from 10 to 40%, with the remainder being
ethylene.
2.
The compound is then added to the dipping machine. The dipping machine is a long, hooded machine
approximately 100 feet (30.5 m) in length. Thick tempered glass rods move along a closed belt between
two circular gears. The belt drags the rods, which are called mandrels, through a series of dips into the
EVA compound. The mandrels rotate to spread the compound evenly. Several coats are required to build
the condom to its required thickness. Between each dip, the compound is hot air dried.
3.
After the final dipping and drying, the balloons automatically roll off the mandrels. A machine shapes and
trims the ring of EVA at the base of each balloon.
4.
Vulcanization of the rubber, forming crosslinks (bridges) between individual polymer chains. Vulcanized
material is less sticky and has superior mechanical properties. [18]
5.
Part 3 | Manufacturing
Following manufacturing steps similar to those of balloons and condoms [17], the following steps for
manufacturing are present:
20
Distributing prohibited | Downloaded by Jan Maty (hanzus1@gmail.com)
lOMoARcPSD
6.
7.
Part 3 | Manufacturing
21
Distributing prohibited | Downloaded by Jan Maty (hanzus1@gmail.com)
lOMoARcPSD
Part 3 | Manufacturing
can be used. In this testing procedure the balloons will be inflated with air so it enlarges. In this case it should be
around a cylinder (like a trocar). If the balloon bursts before it reaches a certain volume (which is bigger than the
volume which is used during surgery), it will fail the test and will be rejected. Also a randomly test should be
performed to see if the assembly of the device is done.
22
Distributing prohibited | Downloaded by Jan Maty (hanzus1@gmail.com)
lOMoARcPSD
Aspect
Sterility
Sharp
Edges
Air-tight
Part
Whole device
Entire device
Criterium*
Device should be delivered sterile
parts should not contain sharp edges
(sealing) balloon,
valve,
Toxicity
all parts
Inflatable
Deflatable
Grip
balloon
balloon
shaft/balloon
Handling
Grip
Materials
Disposable
Valve
holder
all
all parts
Read paragraph 7.3 of ISO 13485 standard (see Blackboard). Formulate all design criteria of the medical device
based on the available information on Blackboard and the demands presented in paragraph 7.3 in a wellorganized scheme or template.
23
Distributing prohibited | Downloaded by Jan Maty (hanzus1@gmail.com)
lOMoARcPSD
Aspect
Sterility
Tested
yes/no
Result
no
Disposable
Toxicity
no
yes,
material
check
air tight
nowhere
to be
found
no
Sensitization studies determine whether a material contains chemicals that may
cause problems after exposure. We recommend the Guinea pig maximization test by
Magnusson & Kligman where: the test animals with a technically enhanced immune
system are exposed intradermally to the test material, and if 15% of the used animals
have a negative reaction to the material then the test is considered positive.
The bubble test is the easiest method for leak detection. A container is filled with
water and the inflated balloon is dipped under the water. From a leak there will be
some bubble emission. To measure the leak rate is not impossible but needs
collection of the bubbles, measure their volume and divide this volume by the
measuring time.
Another way is to use a gas detector.
What is CE marking?
CE stands for Conformit Europenne. [19] This mark states that the product is assessed before being placed on
the market and meets EU safety, health and environmental protection requirements. [20] When this mark is
aquired it means that the product may be sold in the European Economic Area (EEA). Does not mean that the
product has to be produced within the EEA. This regulation creates two standards: one for businesses that all have
to relate to the same rules, and one for consumers who are sure that all products they buy in the EEA are of the
same quality and safety standard. A mistake often made is that CE marking means the product is stated safe by
authorities like the EU, but CE markings can be assigned by a notified body (NB or NoBo), which a third party that is
accredited by the authorities. You can recognize a true CE mark from being shaped out of two colliding circles.
Often accompanied by four numbers noting the notified body it was assessed by.
More information to be found at:
Rijksoverheid [21]
DEKRA [22]
Some definitions. Describe the following terms in your own words and give at least two references where more
information can be found .
24
Distributing prohibited | Downloaded by Jan Maty (hanzus1@gmail.com)
lOMoARcPSD
D | CE MARKING ROUTE
Determine based on the classification of your medical device, which CE marking route should be followed.
Present a flow chart. Discuss at what stage in the route the medical device currently is and indicate it in the flow
chart.
The classification of the PortShield is in class IIa. The steps that should be followed based on this classification are:
1. The conformity Assessment Route needs to be chosen. This can be done by the flowchart shown below in Figure
10. There are two ways to do this, however we would choose the first way due to the simplicity in steps but also
the completeness of this choice in addition to the other:
Follow the procedure of the EC declaration of conformity found in Annex II, the full quality assurance.
or follow the procedure of the EC declaration of conformity set out in Annex VII coupled with either the
25
Distributing prohibited | Downloaded by Jan Maty (hanzus1@gmail.com)
lOMoARcPSD
Stage
The medical device is currently at the step where there are some minor risks, still we are not planning to make any
further design changes and have to start preparing the technical documents that are required to be compiled.
These documents are important to obtain the declaration of conformity. These technical documents should be
very detailed and should contain everything about design, development, manufacturing of the device and the
conformity of the device to the harmonized standards.
procedure relating the EC declaration of conformity of verification of each product or batch of products
(Annex IV), the procedure of production quality assurance (Annex V) or the Product Quality Assurance
(Annex VI)
2. The Technical File has to be compiled
3. The Declaration of Conformity
4. Appoint an Authorised Representative. (Hold the Tech Files for inspection by the Competent Authority)
5. Vigilance and Post Market Surveillance. (affix CE marking & market the products)
26
Distributing prohibited | Downloaded by Jan Maty (hanzus1@gmail.com)
lOMoARcPSD
Figure 11 Relation CE, MDD and ISO 13485 and ISO 14971
Indicate the relation between the CE marking, Medical Device Directive, ISO 14971 and ISO 13485. You can a
give description, but you can also create a flow chart or drawing with some additional comments.
27
Distributing prohibited | Downloaded by Jan Maty (hanzus1@gmail.com)
lOMoARcPSD
A | CLEANING
Let us assume that your medical device product turned out to be a reusable product. This implies that the device
should be sterilized, but more importantly should allow thorough cleaning. Gather the SolidWorks drawings of
your redesigned device (see Assignment 2) and argue if the device in its current form can be cleaned properly. If
not, redesign one aspect of the medical device that overcomes one of the cleaning problems.
What if PortShield is a reusable device, can it be cleaned?
The PortShield is intended for single use. Therefore now the only thing that is really important is sterile packaging,
to provide sterility when using the device. After the procedure the PortShield will be thrown away and thorough
cleaning after the procedure is not needed.
If the PortShield turned out to be a reusable product, the device should be cleaned and sterilized. With the current
device this will not be that easy, mostly because of two reasons:
1 Some parts cannot stand high temperatures
2 accessibility of small parts and groves
Accessibility of parts applies to the inflation tube that connects the bottom of the balloon to the valve. Since this is
a relatively long and very narrow tube it is very hard to clean. A solution for this problem should be found as well,
or this also has to be a disposable part. This would be easiest if this tube is integrated within the balloon for easy
attachment and detachment. Accessibility also applies to different parts that do qualify for sterilization. Since the
PortShield consist of different parts and these parts are attached to each other, small slots at connection places
occur. This is also something that should be considered when the PortShield would be reusable. A possibility is to
make all parts detachable so they can be sterilized separately
Since most common sterilization techniques use high temperatures [33] (from 120C) is will not be able to sterilize
the balloon properly. Either proper materials should be used for the balloon so sterilization will be possible. Or the
PortShield should be designed in such a way that some parts are reusable and some are for one time use. The
balloon could be such a part.
28
Distributing prohibited | Downloaded by Jan Maty (hanzus1@gmail.com)
lOMoARcPSD
nd
29
Distributing prohibited | Downloaded by Jan Maty (hanzus1@gmail.com)
lOMoARcPSD
As Bhoyrul et. al. [34] mentioned the slipping of the trocar is the most annoying problem in minimally invasive
surgeries. According to Tsivian et. al. [35] port site metastases is influenced to some extent by surgeon and OR
team, therefore it could be prevented by fixating trocar to prevent dislodgment or prevent gas leakage along and
around the trocar. So they came up with a cheap disposable solution by which the trocar is fixed to the skin with
an 00 suture and tied around the insufflation nipple to prevent extraction of the trocar [34]. Also tape is wound
around the basis of the trocar to prevent pushing the trocar inside. This method has almost eliminated trocar
fixation and extraction but also prevented gas leakage because the trocar fits closely to the fascia. This fixation
method uses simple, cheap and readily available materials.
C | CLINICAL EVALUATION
Use the results of Assignment 1d, 1e and 3b to formulate a study goal for clinical evaluation in the operating
room. Include at least one objective measure and if applicable formulate a hypothesis.
Goal:
There are two clinical aspects that have to be tested. First there is the impact the device has on the current
procedure of laparoscopic surgery. This impact can be interpreted in different ways, but most logically would be
time of the procedure. The difference in time the procedure takes with and without usage of the PortShield could
indicate whether using this device interferes with usability or maneuverability of the other instruments or if it
might cause the risk of over exposing the patient to anesthesia. Second, the clinical goal of the PortShield in the
O.R. should be tested: prevention of gas leakage. We would want the gas leakage to be significantly less than
without the use of this device.
Hypotheses:
1) Abdominal laparoscopic surgery takes longer when using the PortShield than without making use of the
PortShield
2) Gas leakage during abdominal laparoscopic surgery is reduced when using the PortShield in the procedure,
relative to abdominal laparoscopic surgery without use of the PortShield.
Based on all available information perform a sample size calculation of the number of patients required to
achieve the research goal. Describe the data that were used to perform the calculation.
Using the power calculation program G*Power version 3.1.9.2 the ideal sample size for these studies is calculated.
Since it is clinical research performed in vivo on human subjects it is very important that a secure estimation for
sample size is made.
The first hypothesis, abdominal laparoscopic surgery takes longer when using the PortShield than without making
use of the PortShield, can be tested with a two-tailed t-test, measuring the difference from a constant value. Since
there is a lot of literature available on the duration of laparoscopic surgery. This way there is no need for two
sample groups which will reduce the amount of subjects needed. With a desired effect size of 0.5, significance level
of 0.05, power of 80%, G*Power estimates the needed sample size at 34 subjects.
The second hypothesis, gas leakage during abdominal laparoscopic surgery is reduced when using the PortShield in
the procedure, relative to abdominal laparoscopic surgery without use of the PortShield, can be tested using a two
tailed t-test measuring the difference between two independent means. Lack of prior research in the amount of
D | SAMPLE SIZE
30
Distributing prohibited | Downloaded by Jan Maty (hanzus1@gmail.com)
lOMoARcPSD
gas leakage in this device creates the need for multiple samples. With an effect size of 0.5, significance level of
0.05, power of 80% and equal group size, G*Power estimates a sample size of 64 subjects per group.
Disadvantages:
Possibility of device failure increases risk of the procedure
Extra post-procedure monitoring to keep track of possible post-operative problems
Risk of allergic reaction to materials used in PortShield
F | DEVICE RECOMMENDATION
Use the current knowledge to argue if you would recommend to bring this the device on the market?
We believe that PortShield is a win-win product. We would highly recommend to bring PortShield on the market
based on its purpose of use and its revolutionary design which will eliminate trocar fixation, gas leakage and make
minimally invasive surgeries more safe and easy while using a way more simple procedure of use, than other
known devices until now. However, we have seen that there is for sure still ground for optimization in design,
performance or material which will all make it more cost efficient which is probably the most important factor to
success and sustain on the market.
Advantages:
The procedure may be performed better because of the reduction of gas leakage
Patient contributes to medical research which could improve overall health care
Better monitoring during procedure because of usage of new device
31
Distributing prohibited | Downloaded by Jan Maty (hanzus1@gmail.com)
lOMoARcPSD
PART 6 | IMPLEMENTATION
You have now identified multiple medical, technical, economic and safety criteria to which the new medical
device product has to adhere to. However, in practice we see that even if you fulfil these criteria 70%-90% of
inventions still fail to be actually utilized by healthcare organizations. In case healthcare organizations do use it,
widespread adoption can take as long as 15 years. This is called the Implementation Gap. Implementation is
the pathway towards having the invention actually utilized in daily care practice; to become the new standard.
To achieve this, we need to determine the inventions Implementation Strength during the design process and
pro-actively shape the invention to maximize this strength. The Implementation Analyzer is a software tool that
supports this complex task. In this assignment you will be taught the methodology behind the implementation
analysis and how to use the tool on your own case. You will receive an individual username and password to use
the tool at the start of the course.
Provide the results of the implementation analysis
Figures 12, 13 and 14 show the results of the HIP analysis. The stakeholders mentioned in this analysis are the
surgeon, nurse in hospital, patient in hospital (three core users), surgery assistant (indirectly operationally
involved) and the healthcare insurer (internal and external conditional agents). From left to right the dots in Figure
12 represent the following: surgery assistant, surgeon, healthcare insurer, patient and finally nurse in the bottom
right. This distribution means that the surgery assistant has the largest investment burden, and the patient has the
lowest. Which is logical when you realize that the patient actually is a passive user of this product. It does not
directly contribute to his/her healing or health, while the surgeon or surgery assistant gets extra handlings during
the surgery. The average investment in the change comes down to 2.75 (Figure 14) which is on the positive side of
Part 6 | Implementation
32
Distributing prohibited | Downloaded by Jan Maty (hanzus1@gmail.com)
lOMoARcPSD
the scale. At first glance it is visible that distribution of the stakeholders at appreciation level is not right. This
means somewhere in the HIPP tool information was processed wrongly, since it is not logical that all users have
exactly the same appreciation for the device (which in this case even is negative, -1, Figure 12). That can also be
seen due to the nurse in hospital-dot in the bottom right corner of Figure 12 Embracing level.
The overall negative appreciation is probably caused by the fact that the information that is used to perform the
HIP-tool analysis implies that the device is more expensive than the regular procedure. This can causes the
appreciation of the healthcare insurer, and the direct users to drop. But in spite of the device being more
expensive than naked laparoscopy (without balloon), it is a lot cheaper than usage of entirely disposable trocars
[3] that are designed to be used with an addition such as this balloon. Therefor this data can be interpreted
differently depending on the perspective of the users: do they come from naked laparoscopy, or from expensive
disposable laparoscopy.
Discuss next steps that need to be taken to facilitate the implementation
Relative to other devices, like the trocar, used during laparoscopic procedures, it is easy for the surgeon and nurses
to use this device since it needs very little training or extra knowledge to use. But the PortShield does bring extra
steps before and after to unpack, attach, detach and dispose the device. These extra handlings bring the most
inconvenience for the surgeon and nurse. Therefore a redesign for reduction of total user steps could be beneficial
for the sales of this product.
Also the insures have to be convinced that this device will have a positive effect on mostly the financial side of this
procedure. It is feasible to introduce the product, but a good estimation of cost efficiency has to be made. This can
be done by making a good business model.
Part 6 | Implementation
For the health care suppliers the benefits of the device should be clear. So first good tests of the benefits (less gas
leakage, less slipping of the trocar) should be performed to convince them. Here the price relative to totally
disposable trocars is also a very beneficial aspect of the device in the market.
33
Distributing prohibited | Downloaded by Jan Maty (hanzus1@gmail.com)
lOMoARcPSD
APPENDICES
A | ISO 14971 ANNEX C
8
13
14
15
20
22
23
24
25
26
27
29.3
29.7
31
What is the intended use and how is the medical device to be used?
Is the medical device intended to be implanted?
Is the medical device intended to be in contact with the patient or other persons?
What materials or components are utilized in the medical device or are used with, or are in contact
with, the medical device?
Is the medical device supplied sterile or intended to be sterilized by the user, or are
other microbiological controls applicable?
Is the medical device intended for use in conjunction with other medical devices, medicines or other
medical technologies?
Are there unwanted outputs of energy or substances
Is the medical device susceptible to environmental influences?
Does the medical device have a restricted shelf-life?
To what mechanical forces will the medical device be subjected?
What determines the lifetime of the medical device?
Is the medical device intended for single use?
Is safe decommissioning or disposal of the medical device necessary?
Does installation or use of the medical device require special training or special skills?
How will information for safe use be provided?
Does the medical device have connecting parts or accessories?
Will the medical device be used by persons with special needs?
In what way(s) might the medical device be deliberately misused?
Appendices
1
2
3
4
34
Distributing prohibited | Downloaded by Jan Maty (hanzus1@gmail.com)
lOMoARcPSD
REFERENCES
[1] N. M. A. Bax en D. C. van der Zee, Trocar Fixation During Endoscopic Surgery in Infants and Children,
Springer-Verlag, 1997.
[3] MediShield, Detachable Anchor for Guiding Tool of Minimally Invasive Surgery, 2015, pp. 1-22.
[5] P. Garel, Hope: European Hospital and Healthcare Federation, 2012. [Online]. Available:
http://www.hope.be/03activities/quality_eu-hospitals/eu_country_profiles/00-hospitals_in_europesynthesis.pdf. [Geopend 15 March 2015].
[6] P. W. Farris, N. T. Bendle, P. E. Pfeifer en D. J. Reibstein, Marketing Metrics: The Definitive Guide to Measuring
Marketing Performance, Upper Sadle River, New Jersey: Person Education Inc., 2010.
[7] Council Directive 93/42/EEC concerning medical devices, European Commission, 1993.
[8] Guidelines for the Classification of Medical Devices, European Commission, DG Enterprise, Directorate G,
Unit 4 - Pressure Equipment, Medical Devices, Metrology, 2001.
[11 M. W. Nerkens, Blackboard TU Delft, MediShield B.V., 5 March 2014. [Online]. Available:
https://blackboard.tudelft.nl/bbcswebdav/pid-2416047-dt-content-rid-8212793_2/courses/31830]
141503/PortShield_05_03_2014_modified.pdf. [Geopend March 2015].
References
[10 Health Devices, Top 10 health technology hazards for 2014, ECRI Institute, 2013.
]
35
Distributing prohibited | Downloaded by Jan Maty (hanzus1@gmail.com)
lOMoARcPSD
[13 EVA, ateva medical eva polymers, 1 October 2013. [Online]. Available:
http://www.qmed.com/sites/default/files/EVA-medical-brochure-English--VF.pdf. [Geopend 28 February
]
2015].
[16 Unkown, Materials Testing, Laboratory Testing Inc., 2015. [Online]. Available:
]
http://www.labtesting.com/services/materials-testing/. [Geopend March 2015].
[17 Unknown, How Products are Made, Advameg Inc., 2015. [Online]. Available:
]
http://www.madehow.com/Volume-2/Condom.html. [Geopend March 2015].
[24 Unknown, Basic Information about the European Directive 93/41/EEC on Medical Devices, 2009. [Online].
References
[23 Unknown, European Medical Device Directives, BSI., 2015. [Online]. Available:
]
http://medicaldevices.bsigroup.com/en-GB/our-services/european-mdd/. [Geopend March 2015].
36
Distributing prohibited | Downloaded by Jan Maty (hanzus1@gmail.com)
lOMoARcPSD
[32 Unknown, Steps for Class IIa medical devices compiance, CE-marking, [Online]. Available:
http://ec.europa.eu/enterprise/newapproach/nando/index.cfm?fuseaction=country.notifiedbody&cou_id=52
]
8 . [Geopend March 2015].
[33 W. H. Organization, The International Pharmacopoeia Fourth Edition, WHO, 2015. [Online]. Available:
]
http://apps.who.int/phint/en/p/docf/. [Geopend 3 March 2015].
[35 A. Tsivian en A. Ami Sidi, Metastases in usological laparoscopic surgery, Journal of Urology, nr. 169, pp.
]
1213-1218, 2003.
References
[34 S. Bhoyrul, T. Mori en L. W. Way, Radially expanding diliatation: s superior method of laparoscopic trocar
]
acces, Surgical Endoscopy, vol. 10, pp. 775-778, 1996.
37
Distributing prohibited | Downloaded by Jan Maty (hanzus1@gmail.com)
References
lOMoARcPSD
38
Distributing prohibited | Downloaded by Jan Maty (hanzus1@gmail.com)