You are on page 1of 24

JUTTI LEVITA, 2011

JUTTI LEVITA-UNIVERSITAS
PADJADJARAN

When the instrument readout is totally free of


matrix effects, sample loss effects, or any other
potential error-causing effects, this method
works well.
The usual procedure is to accurately prepare a
stock standard solution of the analyte and then
to prepare a series of standards by diluting this
stock.

JUTTI LEVITA-UNIVERSITAS
PADJADJARAN

JUTTI LEVITA-UNIVERSITAS
PADJADJARAN

120

100

Signal (mAbs)

80

60

40

20

0
0

10

12

JUTTI
PbLEVITA-UNIVERSITAS
Concentration (ppb)
PADJADJARAN

14

16

18

20

The standard additions method (often referred


to as "spiking method") is commonly used to
determine the concentration of an analyte that
is in a complex matrix, such as biological fluids,
soil samples, etc.

The reason for using the standard additions


method is that the matrix may contain other
components that interfere with the analyte
signal, causing inaccuracy in the determined
concentration.
JUTTI LEVITA-UNIVERSITAS
PADJADJARAN

The idea is to add analyte to the sample ("spike"


the sample) and monitor the change in instrument
response.

The change in instrument response between the


sample and the spiked samples is assumed to be
due only to change in analyte concentration.

The procedure for standard additions is to split the


sample into several even aliquots in separate
volumetric flasks of the same volume.

The first flask is then diluted to volume with


the selected diluent.
JUTTI LEVITA-UNIVERSITAS
PADJADJARAN

A standard containing the analyte is then added in


increasing volumes to the subsequent flasks and
each flask is then diluted to volume with the
selected diluent.

The instrument response is then measured for all


of the diluted solutions and the data is plotted with
volume standard added in the x-axis and
instrument response in the y-axis.

Linear regression is performed and the slope (m)


and y-intercept (b) of the calibration curve
are used to calculate the concentration of analyte
in the sample.
JUTTI LEVITA-UNIVERSITAS
PADJADJARAN

JUTTI LEVITA-UNIVERSITAS
PADJADJARAN

JUTTI LEVITA-UNIVERSITAS
PADJADJARAN

Vx = volume of the sample aliquot


cx = concentration of the sample
cs = concentration of the standard
JUTTI LEVITA-UNIVERSITAS
PADJADJARAN
m = slope

An internal standard is a known amount of a


compound, different from the analyte, added to the
unknown.

The signal from the analyte is compared (ratio) with the


signal from the internal standard to determine analyte
concentration.

Internal standards are used when the quantity of


sample analyzed or the instrument response varies
from run to run and difficult to control. Also used when
sample loss is unavoidable during sample preparation
steps prior to analysis.
JUTTI LEVITA-UNIVERSITAS
PADJADJARAN

If you are injecting or pipetting very small


volumes, any variation from run to run can lead
to very large relative errors.

Instruments or techniques that are fully or


partially automated tend to reduce these
variations and thus improve precision
(decrease the standard deviation).

The curve is made by preparing a set of


standard solutions for analyte (A) with the
addition of a constant amount of a second
species (B) to each solution.
JUTTI LEVITA-UNIVERSITAS
PADJADJARAN

Run 1:

Run 2: You injected a


slightly smaller volume.

Both signals, analyte and internal standard, are affected the


same way. Even though the analyte signal is smaller in run 2,
the ratio to the standard will be the same as run 1.
JUTTI LEVITA-UNIVERSITAS
PADJADJARAN

To use an internal standard, we prepare a


known mixture of standard and analyte to
measure the relative response of the detector
to the two species.

An instrument may not respond exactly the


same way to both compounds.

F is the response factor.


Thus, if F is 1.56, then we would say that the instrument responds
to the analyte 1.56 times greater than the standard.
Use F to determine an unknown analyte concentration with a
JUTTI LEVITA-UNIVERSITAS
known amount of internal
standard.
PADJADJARAN

Each Pb solution contains 100 ppm Cu.

Signal
[Pb]
(ppm)
20
40
60
80
100

Pb

Cu

112
1347
243
1527
326
1383
355
1135
JUTTI LEVITA-UNIVERSITAS
558 PADJADJARAN1440

Pb/Cu
0.083
0.159
0.236
0.313
0.388

JUTTI LEVITA-UNIVERSITAS
PADJADJARAN

No Internal Standard Correction


600

Pb Emission Signal

500

400

300

200

100

0
0

20

40

60

JUTTI LEVITA-UNIVERSITAS
[Pb] (ppm)
PADJADJARAN

80

100

120

Internal Standard Correction

JUTTI LEVITA-UNIVERSITAS
PADJADJARAN

A series of Standard Solutions has been prepared, each having a different amount of "A"
in it. The volume% of "B" and of the
Internal
Standard "IS" is held constant in each
JUTTI
LEVITA-UNIVERSITAS
PADJADJARAN
sample.

JUTTI LEVITA-UNIVERSITAS
PADJADJARAN

JUTTI LEVITA-UNIVERSITAS
PADJADJARAN

JUTTI LEVITA-UNIVERSITAS
PADJADJARAN

JUTTI LEVITA-UNIVERSITAS
PADJADJARAN

JUTTI LEVITA-UNIVERSITAS
PADJADJARAN

You might also like