1.0 SCOPE
This document presides over the use of Arrays 2905 laser film
digitization.Hardware and software criterion are clearly defined and specified
in order to standardize and maintain good quality control of digitized
radiographic image.This document only covers on the process procedure to
digitize conventional radiograph films.Hence, this document does not
include any guidance or specifications with regard to the evaluation of the
digitized images for acceptance or rejection to any criteria pertaining to flaw
sizing, defect recognition or any other structural integrity evaluation of the
subject or part imaged.NOTE: Other systems that exhibit equivalent
capabilities enlisted in this document may be used upon consent from a
Malaysia Nuclear Agency personnel who is qualified, and certified in the film
digitization process.
2.0 PROCEDURE GUIDELINES FORTHE DIGITIZATION OF RADIOGRAPHS
This procedure provides guidelines to fully perform digitization radiograph
process in accordance to industrial standard.
3.0 PURPOSE
The aim of this procedure is to clarify the essential items required for
preparing, handling, digitizing and storing industrial radiographs. This
procedure shall also define quality control standards that will allow for
uniformity of data integrity and image quality. This is done in an effort to
allow for the digitization of existing radiographs for subsequent
interpretation and long term storage.
4.0
1.
REFERENCED DOCUMENTS
Applicable Codes
1. ASME Section V, Article 2, Mandatory Appendix III: Digital Image
Acquisition, Display, and Storage for Radiography and Radioscopy,
2010 Edition
4.1.2 ASME Section V, Article 2, Mandatory Appendix VI and
Supplement A: Digital Image Acquisition, Display, Interpretation,
and Storage of Radiographs for Nuclear Applications, 2010
Edition, as applicable.
4.1.3 EN 14096-1,
Non-destructive
testing
Qualification
of
radiographic film digitization systems Part 1: Definitions,
quantitative measurements of image quality parameters,
standard reference film and qualitative control.
4.2 Malaysia Nuclear Agency Level III who is qualified, and certified in the
film digitization
process shall determine Arrays 2905 Start Systems
User's Manual.
4.3 Array 2905 Start Systems User's Manual or equivalent as determined by
a Arrays 2905 Instruction Manual (for model Arrays 2905 film digitized)
or equivalent as determined by a Malaysia Nuclear Agency Level III who
is qualified, and certified in the film digitization process.
PRO PERTY OF
PREPARED BY
DATE
Arrays 2905
APPROVED BY
DR KHAIRUL-
DATE
12 FEB 2015
01
PAGE 02 OF
20
Weight
Approx. 104kg
Environment
conditions
-20 -600C
10-90% Rh (noncondensing)
PROPERTY OF
PREPARED BY
DATE
Arrays 2905
APPROVED BY
DR KHAIRUL-
DATE
12 FEB 2015
01
PAGE 03 OF
20
Interface
Optical Resolution
Sampling Pitch
Density Range (optical
density)
Bits per pixel
LUT
Memory
Power
Power consumption
Environment conditions
Dimensions
Weight
Max 365mm
Photomultiplier
He-Ne laser (633nm)
Laser source: Class 3B Laser product:
Class 1
USB 2.0
50 m
50-200m
Standard 0.00 4.0
12bits(4096 gradation) 14bits (16384
gradation)
12/14 bits
Standard 24MB
AC 100-120 / 200-240
300W
In operation
15-300C
30-75% RH (non-condensing)
In storage
-20-600C
10-90%RH (non-condensing)
W526xH330xD764 (mm)
Approx. 56kg
PROPERTY OF
PREPARED BY
DATE
Arrays 2905
APPROVED BY
DR KHAIRUL-
DATE
12 FEB 2015
01
PAGE 04 OF
20
IF2002CP(GU20IUX)
Screen Diagonal
498mm(19.6)
Display Technology
Response Time
Viewable Image Size
(HXV)
Display Resolution
(HXV)
Pixel pitch
398.4mm X 298.8mm
1600 X 1200
0.249mm X 0.249mm
Brightness
Contrast Ratio
600 : 1
Viewing angle
PROPERTY OF
PREPARED BY
DATE
Arrays 2905
APPROVED BY
DR KHAIRUL-
DATE
12 FEB 2015
01
PAGE 05 OF
20
IF2002CP(GU20IUX)
Power Supply
Power
consumpti
on
Vendor
SKY NET
Model
SNP-A127-M
Input
Output
100-250Vac
3-1.5A, 50/60Hz
12VDC, 9A
Typical
85W
DPMS
10W
IF2002CP(GU20IUX)
Input Connector
DVI-D Connector
Input Signal
Sync 1 : H-sync
Sync 2 : V-sync
60 Hz
162 MHz
USB
Standa
rd
PORT
Mini
Port
Plug & Play
Model
IF2002MP(KT20IUX)
Dimensi Portrait
on (with Landsca
pe
standard Unpack
Weights
ed
Packed
PROPERTY OF
PREPARED BY
APPROVED BY
DR KHAIRUL-
DATE
Arrays 2905
DATE
12 FEB
01
PAGE 06 OF
20
PROPERTY OF
PREPARED BY
APPROVED BY
DR KHAIRUL
DATE
Arrays 2905
DATE
12 FEB
01
PAGE 07 OF
20
Figure :
8.4.4 The collected data shall be tabulated in a continuous manner. The
gray values shall be matched up with its correspondent strip
designations and optical density in accordance to EPRI sheet
reference.
Figure :
PROPERTY OF
PREPARED BY
DATE
Arrays 2905
DATE
APPROVED BY
PAGE 08 OF
20
DR KHAIRUL
12 FEB
01
8.4.5 A polynomial graph with the most fitted line shall be created. The
4th order of polynomial type graph is recommended.
Figure :
8.4.6 Extracted the graph gradient formula. New set of values shall be
tabulated.
Figure :
8.4.7 For 12-bit optical density, the formula, - log10 (1/4095) to - log10
(4095) shall be applied to give overall possible optical density
values.
PROPERTY OF
PREPARED BY
APPROVED BY
DR KHAIRUL
DATE
Arrays 2905
DATE
12 FEB
01
PAGE 08 OF
20
8.4.8 The resultant data shall now be used to re-calibrate the gray value
of the ISee!.exe viewer.
8.5 Modulation Transfer Function verification
8.5.1 The reference digitizing EPRI film shall be loaded.
8.5.2 Profile explorer window shall be opened.
9.1
THE RADIOGRAPHIC COLLECTION VALUE AND CONDITION
ASSESSMENT
9.2Degradation problems should be spot checked on the radiographic
collection. Any loss of image due to improper processing or storage is
categorized as degradation. This includes discoloration, emulsion
separation, leaching, stains, pressure marks, mold, etching, crimps,
creases, abrasions, and fingerprints, fibers embedded in the emulsion,
roller marks and other condition which would obscure the visible image.
9.3The radiographic collection shall be resealed and reported to the Analyst for
disposition if there is evidence of having gross damage that makes it unfit
for ASi-Flat Panel.
PROPERTY OF
PREPARED BY
APPROVED BY
DR KHAIRUL
DATE
Arrays 2905
DATE
12 FEB
01
PAGE 08 OF
20
PROPERTY OF
PREPARED BY
APPROVED BY
DR KHAIRUL
DATE
Arrays 2905
DATE
12 FEB
01
PAGE 09 OF
20
Figure 1: Show that the power and imager data cable must be
connected.
12.1.4Click the ASi-Flat Panel software, AVI100 and wait until the
new window screen appeared
PROPERTY OF
PREPARED BY
APPROVED BY
DR KHAIRUL
DATE
Arrays 2905
DATE
12 FEB
01
PAGE 10 OF
20
12.1.5
10.2.1BALTEAU NDT view will appear on the background
PROPERTY OF
PREPARED BY
APPROVED BY
DR KHAIRUL
DATE
Arrays 2905
DATE
12 FEB
01
PAGE 11 OF
20
Figure 5: The arrow show that mode can be chosen. Accumulation Mode or
Continuous Mode
PROPERTY OF
PREPARED BY
APPROVED BY
DR KHAIRUL
DATE
Arrays 2905
DATE
12 FEB
01
PAGE 13 OF
20
Figure 7: At Continuous Mode the real time image will appear on the screen.
Figure 8: The arrow show that the image will be captured by pressing the
button Start Grab and button Stop Grab to captured the new image.
PROPERTY OF
PREPARED BY
APPROVED BY
DR KHAIRUL
DATE
Arrays 2905
DATE
12 FEB
01
PAGE 14 OF
20
Figure 9: Drag the saved file to ISee!.exe software to view the saved image.
PROPERTY OF
PREPARED BY
APPROVED BY
DR KHAIRUL
DATE
Arrays 2905
DATE
12 FEB
01
PAGE 15 OF
20
Figure 10: Using this software, user can zoom and adjust the saturation and
brightness of the saved image. User also can zoom in and zoom out to
detect the defects on the image saved.
for:
13.3.1 Record keeping (e.g. proper file name and other data input
detailing),
13.3.2 Ensuring that all of the desired radiographic region is
digitized,
13.3.3 Ensuring that the essential range of optical densities are
captured in the radiographic image
13.3.4 Ensuring that there are no obvious digitization artifacts (e.g.
horizontal and/or vertical streaks or blurs).
PROPERTY OF
PREPARED BY
APPROVED BY
DR KHAIRUL
14.0
DATE
Arrays 2905
DATE
12 FEB
01
PAGE 15 OF
20
DOCUMENTATION
14.1 Documentation shall be provided to validate:
14.1.1 That the equipment and digitization requirements of this
procedure have been met.
14.1.2 Calibration
information
and
system
performance
demonstration including digitization standard scans or
references thereto.
14.1.3 That all film degradation and artifact have been noted.
14.1.4 Any missing or additional films encountered.
14.1.5 Any missing or damaged reader sheets.
14.2 Radiographic Documentation Report:
A radiographic documentation report shall be completed for each
digitized image or set of digitized images. A report similar to the
format of this procedure is required and the overall report package
shall include the following as a minimum condition.
14.2.1 Malaysia Nuclear Agency part number, production number,
or purchase order number applicable to the original film and
film report.
14.2.2 Weld identification
14.2.3 Date on radiograph